Omega Products CLDTX-100 User Manual

Page 1
CLDTX-100 Series
SECTION 1.0
THEORY OF OPERATION
1.0 CHLORINE DIOXIDE INTRODUCTION. Chlorine dioxide is considered one of the most powerful biocides available today. It is so effective that it was used to disinfect sites infected with the SARS virus and is used to disinfect areas know to be contaminate with Legionella sp. One of the most important properties of ClO2 that sets it apart from chlorine is its be­havior when placed in water. Not only is ClO2 10 times more soluble in water than chlorine (3.01 grams/Liter at 25 degrees C), it doesn't hydrolyze when placed in solution. It remains as a "true" dissolved gas that retains its useful oxidative and bio­cidal properties throughout the entire 2 to 10 pH range.
1.2 SENSOR OPERATING PRINCIPLE. Chlorine dioxide gas diffuses across the sensor's hydrophobic silicone membrane between the cathode and electrolyte solution. At the applied potential, it is electrochemically reduced at the gold cathode as: ClO2 + 4H+ + 5e- ---> Cl- + 2 H2O At the same time, the silver anode is oxidized to form silver chloride (AgCl) as: Cl- + Ag ---> AgCl + e- The release of electrons at the cathode and acceptance at the anode creates a current flow, which under constant conditions, is proportional to the chlorine dioxide concentration in the medium outside the sensor. The resulting low current output is then conditioned to 4-20mA current by the sensor's on­board electronic circuitry.
PRODUCT INSTRUCTION SHEET
SECTION 2.0
FACTORS INFLUENCING THE SENSOR
2.1 pH. As dicussed in Section 1, there is no significant pH dependence when measuring ClO2.
2.2 Chemical Interferences. The sensor should not be used in water containing surfactants.
2.3 FLOW. The membrance covered chlorine dioxide sensors (CLD series) function at any flow rate. To acheive reproduc­ible measurements, these sensors require a specified constant flow rate. To avoid complications (such as bubbles), it is best to operate the sensors at a flow rate of 0.2-0.6 gpm.
Parts covered by this product instruction sheet include: CLDTX-100 Series
M4678/0708 page 1 of 6
Page 2
PRODUCT INSTRUCTION SHEET
SECTION 3.0
SENSOR PREPARATION
3.0 CHLORINE DIOXIDE SENSOR ASSEMBLY. Your Chlorine Di­oxide sensor is shipped with the membrane cap (CLDA-2016) pre-installed and covered with a cap with water inside to keep the membrane wet. NOTE: IF SENSOR WILL BE STORED DRY OUT OF FLOW CELL, SHAKE BODY DOWNWARD INTO A SINK TO REMOVE THE FILL SOLUTION. TAKE THE MEMBRANE CAP AND IMMERSE IN A CUP OF TAP WATER UNTILL READY TO REUSE. SEE SECTION 9. REPLACE CAP AND ELECTROLYTE BEFORE INSTALLING INTO FLOW CELL (see SECTION 10 for cap and electrolyte change. See SECTION 5 for sensor installation into flow cell).
SECTION 4.0
FLOW CELL INSTALLATION
4.0 FLOW CELL. To obtain accurate Chlorine dioxide reading, the Sensor must be installed into the Flow Cell to prevent air bubbles formation on the membrane, proper spacing between the sensor and the installation wall, and laminar flow across the membrane.
4.1. Using two 1/4” NPT Tube fittings, connect the Flow Cell into your system, noting the inlet (bottom) and outlet (side) orientation. (see FIGURE 1)
4.2. Install clamp with rubber backing as shown in FIG. 2.
4.3 Drill 3/8" diameter hole on the panel.
4.4 Insert bolt as shown in FIG 2
4.5 On back of panel attach lock washer and nut to secure clamp and flow cell to panel.
FIG.1
SENSOR
SNAP RING
GROOVE
THREADED FITTING
SNAP RING (INSTAL L S INTO PROBE GROOVE)
O-RING
BARBED TUBE FITTING
FLOW CELL
BARBED TUBE FITTING
FIG. 2
SECTION 5.0
SENSOR INSTALLATION
5.0 SENSOR INSTALLATION INTO FLOW CELL. a. First install threaded fitting onto sensor body (remove fitting if pre-installed in flow cell) b. Install snap-ring into groove on sensor body c. Next, slide o-ring onto body of sensor until it reaches bottom of threaded fitting. d. Thread sensor assembly into top of flow cell as shown in FIGURE 2. d. Turn on flow and verify the flow through the Flow Cell is at least 0.2gpm (45 liters/hour and no more than 0.6gpm (135 liters/hour).
Parts covered by this product instruction sheet include: CLDTX-100 Series
M4678/0708 page 2 of 6
Page 3
PRODUCT INSTRUCTION SHEET
SECTION 6.0
ELECTRICAL INSTALLATION
6.0 ELECTRICAL INSTALLATION. The sensor produces an ap­proximate output of 4 mA in air and 20mA at the top range of chlorine dioxide output (0-1ppm, 0-2ppm, 0-5ppm).
NOTE: The supply voltage to the Sensor must be 12-24 V DC with minimum of 250 mA. Maximum load is 1 Watt. The sensor has 3 wires, red (+) , black (-) and clear (shield). Twist together or solder black and clear if instrument does not have separate ground. If a separate ground is available such as for PLC’s connect clear (shield) to it. Attach the red wire to the power supply positive terminal (+)and the black wire to the PLC or DVM positive (+) terminal. Connect a wire (customer supplied) from the power suppy negative (-) and the PLC or DVM (-). See FIG 3. The Sensor will require several minutes to stabilize
after power is supplied to it.
SECTION 7.0
SENSOR CONDITIONING
7.0 SENSOR CONDITIONING The sensor requires conditioning prior to generating stable values. a. For new Sensors, allow the Sensor to run for at least hours before calibration. b. If the Sensor will be un-powered for 2 hours or more, run for 3 hours prior to use. c. After membrane/electrolyte replacement, allow the Sensor to run for at least 4 hours.
FIG. 3
Power Supply
sensor red wire
sensor black wire
PLC, DVM,...
Parts covered by this product instruction sheet include: CLDTX-100 Series
M4678/0708 page 3 of 6
Page 4
PRODUCT INSTRUCTION SHEET
SECTION 8.0
CALIBRATION
IMPORTANT NOTE: SENSORS ARE SUPPLIED FACTORY CALIBRATED WITH A 4-20mA SIGNAL OUTPUT CORRESPONDING TO THEIR SPECIFIC RANGE (0-1, 0-2 OR 0-5ppm ClO2). SPAN CALIBRATION IS NECESSARY WHEN RECEIVING A NEW SENSOR SINCE YOUR CONDITIONS MAY VARY FROM THOSE USED AT THE FACTORY FOR SPAN CALIBRATION. THE ZERO POINT CALIBRATION IS NOT NECESSARY SINCE THE ZERO SETTING IS VERY STABLE.
PERIODIC CALIBRATION ( ABOUT ONCE PER WEEK) IS RECOMMENDED. THIS USEFUL IN TRACKING SENSOR FAILURES AS WELL.
8.1 Removal of cord grip to access ZERO and SPAN potentiometers inside the sensor body: a. Remove top nut of cord grip as shown in FIGURE 4. b. Remove remaining cord grip as shown in FIGURE 5. A wrench may be required.
8.2 Slope Calibration: a. Determine the chlorine dioxide (ClO2) concentration using a diethyl-p-phenylenediamine (DPD) colorimeter test kit (see FIGURE 9.), not included with ClO2 sensor and flow cell. b. Measure Chlorine dioxide content with sensor. Make sure that calibration flow rate matches flow rate when measuring sample since probe output is flow rate dependent. c. Adjust the span potentiometer to the desired mA (see FIGURES 9 & 9A) c. Repeat this slope calibration one day after sensor is initially installed. d. Routinely repeat the slope calibration monthly.
FIG. 4
FIG. 6
FIG. 5
FIG. 7
Left potentiometer (Zero adjust)
Right potentiometer (Span adjust)
Cable
SECTION 9.0
SENSOR STORAGE
9.0 STORAGE. Store sensor at 5o C- 50o C ONLY. and maximum humidity of 95% non-condensing. a. Short Term Storage (1 week or less): Store in Flow cell with water to prevent the probe from drying out. b. Intermediate Term (1 week to 1 month): Store in cap, bottle, or beaker with water to keep membrane wet. c. Long Term (1 month or longer): Remove Membrane Cap and store completely immersed in tap water. Turn sensor upright and shake it to remove fill solution from inside the sensor.
FIG. 9
FIG. 8
to Increase
output
to Decrease output
C
2
L
O
ClO2
ZERO
READ
Note: The DPD method for chlorine dioxide (ClO2) measurement is not selective for ClO2 alone. Other oxidants in the water can also be measured as ClO2.
Parts covered by this product instruction sheet include: CLDTX-100 Series
M4678/0708 page 4 of 6
Page 5
PRODUCT INSTRUCTION SHEET
SECTION 10.0
SENSOR MAINTENANCE/
RECONDITIONING
10.1 MEMBRANE CAP REPLACEMENT. If membrane replace­ment is required, a new cap with preinstalled membrane must be used. Order CLD-2016(see Section 12) Cap/membrane replacement. To change membrane cap: a) turn sensor upside down with cap facing upward b) rotate cap counter-clockwise to remove (see FIG 10). c) fill sensor body with electrolyte using needle and bottle of refill solution (see FIG. 11) d) install new membrane cap by threading cap onto sensor rotating cap clockwise (opposite of FIG 10)
SECTION 11.0
SENSOR TROUBLESHOOTING
11.1 CALIBRATION PROBLEMS a) Sensor output HIGHER than DPD test
1) Run in time too short
2) Membrane cap damaged
3) Interference from water contaminants (see Specifications, "Cross Sensitivity")
4) Cable short circuit or damage
b) Sensor output LOW ER than DPD test
1) Run in time too short
2) Deposits on Membrane cap
3) Flow rate too low
4) Air bubbles on membrane
5) Surfactants in water
6) No electrolyte in membrane cap
c) Sensor ouput is 4mA (zero ppm)
1) Run in time too short
2) Chlorine dioxide content below detection limit
3) Sensor not wired correctly (See SECTION 6.0 of this manual)
4) Defective sensor
d) Sensor output UNSTABLE
1) Air bubbles on membrane
2) Membrane damage
3) Non-sensor problem
FIG. 10
FIG. 11
INJECT UNTIL SOLUTION
LEAKS OUT OF FILL HOLE
TROUBLESHOOTING CHART
Symptom Possible Cause Solution/Remedy
The sensor 1) Run in time too short 1) See Sec 5.0 -CONDITIONING cannot be 2) Membrane cap damaged 2) Replace cap - See Sec 8.0 calibrated- 3) Interference from contaminants 3) See SPECIFICATIONS ouput is 4) DPD chemicals bad 4) Use new DPD kit HIGHER than 5)Temperatue increased since cal 5) Match calibration temp. DPD Test
The sensor 1) Run in time too short 1) See Sec 7.0 -CONDITIONING cannot be 2) Deposits on membrane cap 2) Remove deposits or replace calibrated- cap if cleaning ineffective. output is 3) Flow rate too low 3) increase fl ow - See LOWER than SPECIFICATIONS DPD Test 4) Air bubbles on membrane 4) Remove and re-install sen sor to remove bubbles.
5) Surfactants in water 5) Remove surfactants and replace cap
6) No electrolyte in cap 6) Add new electrolyte, run in sensor and re-calibrate
7) Temperature decreased since cal 7) Increase temp to match cal
8) Organic chlorination agents 8) Use chlorinating agents present in water per DIN 19643
Sensor output 1) No ClO2 present` 1) Check for ClO2 with is 4mA (0 ppm) appropriate DPD test.
2) Run in time too short 2) See Sec 7.0 -CONDITIONING
3) ClO2 content below limit 3) Add ClO2, check genera­ tor and repeat calibration
4) No electrolyte in cap 4) Refi ll electrolyte
5) Sensor electrical connection 5) See SECTION 6.0 wrong
Unstable output 1) Air bubbles on sensor membrane 1) Tap to remove bubbles from sensor 2) Membrane damaged 2) Replace membrane, run in sensor and recalibrate.
3) Non-sensor problem 3) check PLC or I/O device
Parts covered by this product instruction sheet include: CLDTX-100 Series
M4678/0708 page 5 of 6
Page 6
PRODUCT INSTRUCTION SHEET
SECTION 12.0
SENSOR SPECIFICATION
12.1 OPERATING SPECIFICATIONS. Follow all operating specifications, especially for pH and flow rate as noted in the specification tables below.
TECHNICAL SPECIFICATIONS
SENSOR Dimensions: 8.2"L x 1" dia Body Material: Black PVC Membrane Material: Silicone O-ring material: Viton Cathode: Gold Anode: Silver chloride (AgCl) Cable: 2 -conductor shielded, 10ft (3mtr) tinned wire leads FLOW CELL Dimensions: 5.58"H x 2.25"DIA Material: Acrylic Connections: 1/4" NPT inlet and outlet
OPERATING SPECIFICATIONS
Operating temperature range: 0-45 degC Maximum operating pressure: 1 bar/14.5 psi/1atm Flow rate minimum: .20 gpm Flow rate maximum: 0.6 ppm pH range: 4-11 Output signal: 4.0+/- 0.2mA in air (zero) 20mA +/- 0.2mA at high range (2, 5 or 10ppm) Power Requirement: 12-24 VDC, 250 mA minimum Cross-Sensitivity:
Do not operate the Sensor in surfactant water. Surfactant water will reduce life of Sensor.
R
FC72 FLOW CELL DIMENSIONS
SENSOR AND FLOW CELL INSTALLATION DIMENSIONS
MAINTENANCE/REPLACEMENT PARTS
FC72 Flow Cell, 1/4 inch FNPT inlet and outlet, includes: 2 each 1/4" barbed tube fittings, clamp, threaded flow cell installation fitting
CLD-ES Chlorine dioxide sensor fill solution, 30mL, 1 each
CLD-RM CLD sensor replacement premembraned cap, 1 each
Parts covered by this product instruction sheet include: CLDTX-100 Series
M4678/0708 page 6 of 6
Loading...