Omega Products CL900A Installation Manual

www.omega.com
e-mail: info@omega.com
User’s Guide
NIST
230 VAC Model Only
Shop on line at
CL900A/CL950A SERIES
®
Dry Block
Probe Calibrator
MADE IN
Servicing North America:
USA: One Omega Drive, P.O. Box 4047
ISO 9001 Certified Stamford CT 06907-0047
TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1, Canada TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378 / 1-800-622-BEST
®
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
®
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico: En Espan˜ ol: (001) 203-359-7803 e-mail: espanol@omega.com
FAX: (001) 203-359-7807 info@omega.com.mx
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
TEL: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: sales@omegaeng.nl
Czech Republic: Frystatska 184, 733 01 Karviná, Czech Republic
TEL: +420 (0)59 6311899 FAX: +420 (0)59 6311114 Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
France: 11, rue Jacques Cartier, 78280 Guyancourt, France
TEL: +33 (0)1 61 37 2900 FAX: +33 (0)1 30 57 5427 Toll Free in France: 0800 466 342 e-mail: sales@omega.fr
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
TEL: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
United Kingdom: One Omega Drive, River Bend Technology Centre
ISO 9002 Certified Northbank, Irlam, Manchester
M44 5BD United Kingdom TEL: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
OMEGAnet®Online Service Internet e-mail
www.omega.com info@omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
Table of Contents
Section ........................................................................... Page
Section 1 Introduction .............................................................................. ... 1-1
1.1 Precautions ............................................................................. 1-1
1.2 Safety Warnings and IEC Symbols .......................................1-1
1.3 Statement on CE Marking .................................................... 1-2
1.4 General Description ............................................................... 1-2
Section 2 Installation ................................................................................... 2-1
2.1 Unpacking and Inspection ................................................... 2-1
2.2 Mounting ................................................................................ 2-2
2.3 Ambient Temperature ........................................................... 2-2
2.4 Power Connection ................................................................. 2-2
Section 3 Operation ...................................................................................... 3-1
3.1 Front Panel Controls and Indicators ................................... 3-1
3.2 Back Panel Connections ........................................................ 3-3
3.3 Changing the Temperature Setpoint ................................... 3-4
3.4 Changing the Controller Parameters .................................. 3-4
3.5 Target Plate Temperature Transition Chart ........................ 3-5
3.6 Replacing Probe Well Inserts in the CL900A ..................... 3-5
3.7 Fanning System ...................................................................... 3-6
3.8 Overhead Reset Switch .......................................................... 3-7
Section 4 Maintenance ................................................................................ 4-1
4.1 Calibration .............................................................................. 4-1
4.2 Cleaning .................................................................................. 4-1
4.2.1 Main Body ....................................................................... 4-1
4.2.2 Target Plate ..................................................................... 4-1
4.2.3 Fan and Vent Guards ..................................................... 4-1
4.3 Fuse Replacement .................................................................. 4-1
Section 5 RC232 Communications ............................................................ 5-1
Section 6 Specifications .............................................................................. 6-1
Section 7 Troubleshooting Guide ............................................................. 7-1
Section 8 Glossary of Terms Used in This Manual ................................ 8-1
Section 9 The OMEGA
®
Family of Blackbody Calibrators .................. 9-1
i
CL900A/CL950A SERIES
Table of
Contents
Table of Figures
Figure Description: ......................................................... Page:
1. IEC Symbols ...............................................................................................1-1
2. Probe Well Inserts for the CL900A ......................................................... 1-2
3. CL900A Probe Well.................................................................................... 2-1
4. CL950A Probe Well ................................................................................... 2-1
5. CL950A-230 Probe Well ........................................................................... 2-1
6. Front Panel ................................................................................................. 3-1
7. Back Panel .................................................................................................. 3-3
8. Controller Menu Hierarchy Showing Factory Default Parameters
Settings ....................................................................................................... 3-4
9. Approximate Target Plate Temperature Transition Times Chart ...... 3-5
10. Removal of Probe Well Insert in CL900A ...............................................3-6
11. Connecting the CL900A to a Computer's Serial Port .......................... 4-3
OMEGAcareSMextended warranty program is available for this model. OMEGAcareSMcovers parts, labor, and equivalent loaners. Ask your sales representative for full details when placing an order.
ii
CL900A/CL950A SERIES
Table of
Figures
1-1
Section 1 - Introduction
Your CL900A/CL950A hot point“ Thermocouple, RTD and thermistor Probe Calibrators have been designed for ease of use and reliability whenever you have the need to test or calibrate temperature probes. It is important that you read this manual completely and follow all safety precautions before operating this instrument.
1.1 Precautions
• Follow all safety precautions and operating instructions outlined in this manual.
• Never leave your calibrator unattended when in use.
• Keep out of reach of children.
• Never touch the probe well area when hot.
• Never place any object within 3 inches of the probe well opening when hot.
• Do not operate in flammable or explosive environments.
• Never operate with a power cord other than the one provided with your unit.
• Remove and or disconnect main power cord before attempting any maintenance or fuse replacement.
• Do not connect and or operate this unit to a non-grounded, non-polarized outlet or power source.
• Do not connect the serial port or reference probe port to equipment with exposed, hazardous, live voltages.
NOTE:
There are no user serviceable parts inside your unit. Attempting to repair or service your unit may void your warranty.
1.2 Safety Warnings and IEC Symbols
This device is marked with international safety and hazard symbols in accordance with IEC 1010. It is important to read and follow all precautions and instructions in this manual before operating or commissioning this device as it contains important information relating to safety and EMC. Failure to follow all safety precautions may result in injury and or damage to your calibrator. Use of this device in a manner not specified by the manufacturer may impair protection provided within the unit.
IEC symbols Description
Introduction
1
Caution, risk of electric shock
Figure 1. IEC symbols
1.3 Statement on Marking
It is the policy of OMEGA to comply with all world-wide safety and EMI/EMC regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon verification of compliance.
1.4 General Description
The CL900A and CL950A are a part of a family of calibrators which allows for calibration of thermocouples, thermisters, RTD and thermistor probes quickly and easily with an accuracy to 0.8°C (1.5°F).
An integral RTD sensor assures high measurement accuracy and repeatability. A built-in microprocessor-based PID digital controller assures exceptional stability. Setpoint and actual temperature are displayed, simultaneously. Temperature can be set in one degree increments from 22°C (40°F) above ambient temperature to 482°C (900°F).
The CL900A differs from the CL950A only with regard to the probe well. The CL900A has a removable, single-port well insert. The CL950A has a built-in multiport well. The CL900A is shipped with the CL906 insert (diameter=1/4", depth=6"). The CL900A-230 is shipped with the CL906-M insert (diameter=4.5 mm, depth= 152.4 mm).
1-2
Introduction
1
Caution, refer to accompanying documents
Caution, hot surface
230 VAC @50Hz (European Models) 115 VAC @60Hz (Domestic Models)
Introduction
1
1-3
Figure 2. Probe Well Inserts for the CL900A
As shown in Figure 1, various inserts may be ordered from OMEGA to meet your specific probe requirements. The insert slides into the probe well of the CL900A.
Figure 3. CL900A Probe Well
The CL950A and CL950A-230 have permanently fixed multi-port wells that accepts 4 different probe sizes. The CL950A and CL950A-230 has 4 standard inch holes with diameters: 1/16", 1/4"x 2, 1/8", 3/16" with 6" hole depth. The CL950A-M and CL950A-M-230 has 4 standard metric holes with diameters: 2 mm, 3 mm, 4.5 mm, and 6 mm x 2 with 152.4 mm hole depth.
Figure 4. CL950A Probe Well Figure 5. CL950A-M Probe Well
Probe Well Inserts with "Inch Sizes" Probe Well Inserts with "Metric Sizes"
OMEGA Probe Hole Probe Hole OMEGA Probe Hole Diameter Hole Part No. Diameter Depth Part No. Probe Depth
CL901 1/8" 4" CL901-M 2 mm 106.6 mm CL902 1/8" 6" CL902-M 2 mm 152.4 mm CL903 3/16" 4" CL903-M 3 mm 106.6 mm CL904 3/16" 6" CL904-M 3 mm 152.4 mm CL905 1/4" 4" CL905-M 4.5 mm 106.6 mm CL906 1/4" 6" CL906-M 4.5 mm 152.4 mm CL907 5/16" 4" CL907-M 6 mm 106.6 mm CL908 5/16" 6" CL908-M 6 mm 152.4 mm CL909 3/8" 4" - - ­CL910 3/8" 6" - - ­CL911 UNDRILLED 6" Max. - - -
1/16"
1/4"
1/8"
1/4"
3/16"
6 mm
3 mm
2 mm
6 mm
4.5 mm
CL900A/CL950A SERIES
1
1-4
NOTES:
2-1
Installation
2
Section 2 - Installation
2.1 Unpacking
Remove the packing list and verify that you have received all your equipment. If you have any questions about the shipment, please call our Customer Service Department at
1-800-622-2378 or 203-359-1660
Website: http://www.omega.com
e-mail: info@omega.com
When you receive the shipment, inspect the container and equipment for any signs of damage. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
NOTE:
The carrier will not honor any damage claims unless all shipping material is saved for inspection. After examining and removing contents, save packing material and carton in the event reshipment is necessary.
The following items are supplied in the box: The following items are supplied in the box:
*A probe well insert is only included with Model CL900A or CL900A-230.
2.2 Mounting
Mount the unit on a bench, table top or shelf in a horizontal position and operate at least ten inches from any air obstructions to the fan, front panel, rear panel, bottom and top of the unit, in an ambient environment between the specified 5°C to 38°C (40°F to 100°F).
2.3 Ambient Temperature
The CL900A/CL950A is designed to operate within an ambient temperature range from 5°C to 38°C (40°F to 100°F). Failure to operate the unit within this range may damage your unit. If the unit is operated in an ambient temperature above 38°C (100°F) a thermal cutoff switch will cut off power to the heater. The process variable (PV) will no longer approach the setpoint variable (SV). The heaters will cool down and cease to operate. If this should occur you should send your unit back to OMEGA for service.
Item: OMEGA Part #:
Hotpoint Calibrator(Removable single port well insert) CL900A /CL900-230
Or Or
Hotpoint Calibrator (Built-in multiport well) CL950A /CL950-230
Probe Well Insert for1/4" Probes (shipped with CL900A)* CL906
Or Or
Probe Well Insert for 4.5 MM Probes (shipped with CL900A-230) * CL906-M
User Manual M-3555
Calibration Certificate CAL-3
Power Cord 200-0028 (Domestic),
200-0107 (International)
Well Insert Removal Tool HP-0037 (CL900A only)