Omega Products CL540 Installation Manual

Page 1
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at www.omega.com
TEMPERATURE
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies Wire: Thermocouple, RTD & Thermistor Calibrators & Ice Point References Recorders, Controllers & Process Monitors Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
Transducers & Strain Gages Load Cells & Pressure Gages Displacement Transducers Instrumentation & Accessories
FLOW/LEVEL
Rotameters, Gas Mass Flowmeters & Flow Computers Air Velocity Indicators Turbine/Paddlewheel Systems Totalizers & Batch Controllers
pH/CONDUCTIVITY
pH Electrodes, Testers & Accessories Benchtop/Laboratory Meters Controllers, Calibrators, Simulators & Pumps Industrial pH & Conductivity Equipment
DATA ACQUISITION
Data Acquisition & Engineering Software Communications-Based Acquisition Systems Plug-in Cards for Apple, IBM & Compatibles Datalogging Systems Recorders, Printers & Plotters
HEATERS
Heating Cable Cartridge & Strip Heaters Immersion & Band Heaters Flexible Heaters Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
Metering & Control Instrumentation Refractometers Pumps & Tubing Air, Soil & Water Monitors Industrial Water & Wastewater Treatment pH, Conductivity & Dissolved Oxygen Instruments
®
MADE IN
User’s Guide
Shop online at
www.omega.com
e-mail: info@omega.com
CL540 SERIES
Thermocouple Simulator
M3979/0603
Page 2
OMEGAnet®Online Service Internet e-mail
www.omega.com info@omega.com
Servicing North America:
USA:
ISO 9001 Certified
Canada:
One Omega Drive, P.O. Box 4047 Stamford CT 06907-0047 TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
976 Bergar Laval (Quebec) H7L 5A1, Canada TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
USA and Canada:
Mexico:
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA Customer Service: 1-800-622-2378 / 1-800-622-BEST Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
En Espan•ol: (001) 203-359-7803 e-mail: espanol@omega.com FAX: (001) 203-359-7807 info@omega.com.mx
®
®
®
Servicing Europe:
Benelux:
Czech Republic:
France:
Germany/Austria:
United Kingdom:
ISO 9002 Certified
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
Postbus 8034, 1180 LA Amstelveen, The Netherlands TEL: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: sales@omegaeng.nl
Frystatska 184/46, 733 01 Karviná, Czech Republic TEL: +420 (0)59 6311899 FAX: +420 (0)59 6311114 Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
11, rue Jacques Cartier, 78280 Guyancourt, France TEL: +33 (0)1 61 37 29 00 FAX: +33 (0)1 30 57 54 27 Toll Free in France: 0800 466 342 e-mail: sales@omega.fr
Daimlerstrasse 26, D-75392 Deckenpfronn, Germany TEL: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
One Omega Drive, River Bend Technology Centre Northbank, Irlam, Manchester M44 5BD United Kingdom TEL: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 37 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal three (3) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR W
ARRANTY RETURNS, please have the
following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2003 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
FOR NON-WARRANTY REPAIRS, for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
consult OMEGA
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the product.
Page 3
CL540 Field Calibration Procedure Continued
CL540 Series Operating Instructions
Basic Keypad Operations
Cold Junction Calibration:
Double click the EZ-Dial knob to enter the Cold Junction Calibration mode. Cal CJ will appear on the top left of the LCD.
1. Connect the Thermistor probe to the + side of the T/C connector.
2. Connect the other side of the Thermistor probe to the OHM meter.
3. Let the temperature settle for 15minutes.
4. Check the Cold Junction Temperature on the top right of the LCD with the temperature reading of the Thermistor probe.
5. If the reading is out of Specification then dial the EZ-Dial to the temperature reading from the Thermistor probe.
6. Press the EZ-Dial Knob down.
7. The display will flash STORED to confirm that the value was stored.
8. Verify the UUT (top right side of the LCD) Cold Junction Temperature is tracking with the temperature reading of the Thermistor probe.
If using and ice bath or ice point calibrator to measure the junction voltage with a Voltage meter, follow the steps below.
1. Connect the CL540 as shown in Figure 1.
2. Let the temperature settle for 15 minutes.
3. Check the Cold Junction Temperature on the top right of the LCD to the Voltage reading on the DVM and T/C chart.
4. If the reading is out of Specification then dial the EZ-Dial to the correct reading form the DVM and T/C chart.
5. Press the EZ-Dial Knob down.
6. The display will flash STORED to confirm that the value was stored.
7. Verify the UUT (top right side of the LCD) Cold Junction Temperature is tracking with the temperature reading of the DVM and T/C Table.
Figure 1
J
Cu
Cu
+
3
_
V
_
Cu
Cu
+
J
4
_
Voltmeter
Isothermal
(same temperature)
Block
Completion of Calibration:
Turn the CL540 off. Next time the unit is turned on the CL540 will be Calibrated and in normal operational mode.
Ice Bath / Ice Point
EZ-Check Switch
Slide the switch to select from three user stored values for the desired calibration points. The User can select HI, DIAL and LO positions. These values can easily be changed to suit the calibration requirements.
ON/OFF Switch
Slide the ON/OFF Switch to turn the CL540 on or off.
EZ-Dial Knob
The EZ-Dial Knob has two adjustment speeds. Simply turning the EZ-Dial Knob will select fine adjustments. While pressing down and turning the EZ-Dial Knob will make course adjustments.
Note: When the EZ-Check Switch is in the HI or LO position, pressing and holding the EZ- DialKnob without turning or LO values.
will store new HI
CL540 Configuration
Instructions for Enabling and Disabling the Configuration Options
1. Turn the CL540 on with the ON/OFF Switch .
2. Press the EZ-Dial Knob while the PRESS EZ-DIAL KNOB FOR CONFIGURATION message is displayed.
3. Select options by turning the EZ-Dial Knob until the arrow points to the desired option.
4. The option can be enabled or disabled by pressing the EZ-Dial Knob.
The CL540 configuration menu will exit automatically after 5 seconds of inactivity and go to normal operation with the options selected. These options are recalled at turn on until they are changed again.
+
CL540 Configuration Menu
Auto Off ON (default)/OFF
If Auto Off is ON, the unit will turn off after 30 minutes to save battery life, if there is no user activity. If Auto Off is OFF the unit will stay on until it is turned off from the keypad. This is typically useful for manual loading or continuous use.
Display Units °C (default)/°F Pressing the EZ-Dial Knob to toggles between °C or °F
TC {tc type} Model CL540: The T/C type is fixed as
ordered from the factory and cannot be changed.
Model CL540Z: To change T/C type, press the EZ-Dial Knob. Turn the EZ-Dial Knob to scroll through the list of available types. Press again to save and return to the configuration menu.
CL540 Configuration
{tc type} is one of:
B, E, J, K, N, R, S, T, or mV
{tc type} is any of:
B, E, J, K, N, R, S, T, or mV
PRESS EZ-DIAL KNOB FOR CONFIGURATION
AUTO OFF ON DISPLAY UNITS °C TC TYPE K
EZ-Dial  K nob
Turning the EZ-Dial Knob to adjust the output up or down. Fine adjustments can be made by turning the EZ-Dial Knob. Coarse adjustments can be made by pressing and turning the EZ-Dial. New values can be stored into the HI and LO EZ-Check positions by pressing down on the EZ-Dial Knob until STORED is indicated on the display.
Page 4
EZ-Check Switch
The EZ-Check Switch has three positions: HI, DIAL, and LO. Its position is shown at the left edge of the display with HI and LO indicators. The output is adjustable in all three positions. The EZ-Check Switch allows user-selected values to be stored in the HI and LO positions when used in combination with the EZ-Dial Knob.
To store new EZ-Check value(s):
1. Dial the display to match the desired stored outputs for the HI or LO positions.
2. Press down on the EZ-Dial Knob until the confirmation message STORED appears.
3. Recall the values by moving the switch between HI, DIAL, and LO.
Operational description: When returning to previously set EZ-Check positions, the DIAL position always recalls the last output value it was dialed to. The HI and LO positions will recall the last STORED value, NOT the last output value it was last dialed to.
Hint: For faster calibrations, the product has been designed so the position of the switch can be felt. This tactile feature allows continuous monitoring of the device being calibrated without looking back at the CL540 display. This is also useful in poor lighting or under difficult operating conditions.
Connection Diagram
THERMOCOUPLE WIRE
Two Wire Connection to Transmitter
Specifications
General Specifications:
(Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from calibration)
Temperature Range -25 to 60 °C (-10 to 140 °F)
Relative Humidity Range 10 % RH 90 % (0 to 35 °C), Non-condensing
Size 4.9 X 3.15 X 1.82 inches (125.5 X 80 X 46.2 mm)
Weight 7.2 oz (204 grams)
Battery 9V Alkaline provides 45 hours of continuous use
Miscellaneous Low battery indication with nominal 1 hour of operation left
Source Thermocouple Specifications (ITS-90 Curves):
Millivolt Uncertainty
Temperature Coefficient of mV Source
Output Noise
Output Impedance
Cold Junction Uncertainty
Cold Junction Sensor Temperature Coefficient
General Temperature Accuracy ±(0.015% of mV setting + 0.009mV) ± 0.25°C (0.5°F)
Output Dial Adjustment Resolution 0.1°C or °F for Model 521/1°C or °F for Model 520
Span -13.000 - 80.000 mV
T/C Type B 594  1820° C (1101.2  3308.0°F)
T/C Type E -260 - 1000° C (-436.0  1832.0°F)
T/C Type J -210  1200° C (-346.0  2192.0 °F)
T/C Type K -245 - 1372° C (-409.0  2501.6°F)
T/C Type N -229 - 1300°C (-380.2  2372.0° F)
T/C Type R 24 - 1768° C (75.2  3214.4°F)
T/C Type S 21 - 1768° C (69.8  3214.4°F)
T/C Type T -251 - 400° C (-419.8  752.0°F)
10 % RH 70 % (35 to 60 °C), Non-condensing
Overload Protected to 60V for 30 seconds or less
High contrast graphic liquid crystal display with 0.357 (9.07 mm) high digits
±(0.015% of mV Setting + 0.009 mV)
±0.005mV/°C Ambient
±5µVpp from 0.1 Hz to 10 Hz
0.2W
±0.25°C (0.5°F)
±0.05°/° in ambient temperature (°C or °F)
RED
TRANSMITTER OR RECEIVER
CL540 Field Calibration Procedure
Suggested Equipment:
1. Precision Voltage Meter with an accuracy of 0.0066% at 80.000mV
2. Precision Thermistor probe with accuracy of ± 0.063°C or
better (YSI 46046 ±0.05°C)
OR
3. A stable ice bath or Thermocouple ice point calibrator and a NIST traceable thermocouple
wire stable to within ±0.063°C.
We recommend using a direct junction temperature measurement technique for the most accurate and reliable calibrations of our equipment. This technique uses accurate RTD or Thermistor probes to measure the reference junction (cold junction) temperature. While measuring the junction voltage with a DVM and ice bath technique will work. This method is less reliable due to complexity and is generally less accurate due to cumulative errors.
Precautions:
Avoid touching thermocouple connections, as this will cause temperature errors in calibration. It is recommended that the CL540 be handled as little as possible during calibration to reduce errors. The heat from your body may cause uneven heating of temperature sensitive components.
Enabling Calibration:
Install a fresh battery and allow 15 minutes for the CL540 to stabilize to ambient temperature. Remove the EZ-Dial Knob, battery cover and the four black Phillips head screws. While holding the CL540 face down in one hand, carefully separate the top and bottom of the housing. Place the unit into calibration mode by shorting the calibration via located on the bottom left side on the PCB with tweezers and turning the UUT on. Verify the CL540 is in Calibration mode by viewing the top Left of the LCD for the word CAL.
Source mV Calibration:
LO Calibration:
1. Connect the CL540 terminals to a Voltage meter.
2. Slide the EZ-Check switch to the LO Position indicated by displaying a LO on the left side of the display.
3. Dial the CL540 so the meter reads 0.000 mV.
4. Press the EZ-Dial Knob down.
5. The display will flash STORED to confirm that the value was stored.
HI Calibration:
1. Slide the EZ-Check switch to the HI Position indicated by displaying a HI on the left side of the display.
2. Dial the CL540 so the meter reads 80.000 mV.
3. Press the EZ-Dial Knob down.
4. The display will flash STORED to confirm that the value was stored.