Data Acquisition & Engineering Software
Communications-Based Acquisition Systems
Plug-in Cards for Apple, IBM & Compatibles
Datalogging Systems
Recorders, Printers & Plotters
En Espan•ol: (001) 203-359-7803e-mail: espanol@omega.com
FAX: (001) 203-359-7807info@omega.com.mx
®
®
®
Servicing Europe:
Benelux:
Czech Republic:
France:
Germany/Austria:
United Kingdom:
ISO 9002 Certified
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that
apply. OMEGA is constantly pursuing certification of its products to the European New Approach
Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts
no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, human applications.
Postbus 8034, 1180 LA Amstelveen, The Netherlands
TEL: +31 (0)20 3472121FAX: +31 (0)20 6434643
Toll Free in Benelux: 0800 0993344
e-mail: sales@omegaeng.nl
One Omega Drive, River Bend Technology Centre
Northbank, Irlam, Manchester
M44 5BD United Kingdom
TEL: +44 (0)161 777 6611FAX: +44 (0)161 777 6622
Toll Free in United Kingdom: 0800-488-488
e-mail: sales@omega.co.uk
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a
period of 37 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month
grace period to the normal three (3) year product warranty to cover handling and shipping time.
This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service
Department will issue an Authorized Return (AR) number immediately upon phone or written request.
Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no
charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser,
including but not limited to mishandling, improper interfacing, operation outside of design limits,
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a result of excessive corrosion;
or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating
conditions outside of OMEGA’s control. Components which wear are not warranted, including but not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any
damages that result from the use of its products in accordance with information provided by
OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be
as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE,
AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of
OMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of the
component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or
activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility
as set forth in our basic WARRANTY/ DISCLAIMER language, and, additionally, purchaser will indemnify
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the
Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE
RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN
(AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return
package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
FOR W
ARRANTY RETURNS, please have the
following information available BEFORE
contacting OMEGA:
1. Purchase Order number under which the product
was PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords
our customers the latest in technology and engineering.
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the
prior written consent of OMEGA ENGINEERING, INC.
FOR NON-WARRANTY REPAIRS,
for current repair charges. Have the following
information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST
of the repair,
consult OMEGA
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems
relative to the product.
CL510 Field Calibration ProcedureCL510 Series Operating Instructions
Suggested Equipment:
1. Precision OHM Meter with accuracy of 0.0275% at 400
W.
Enabling Calibration:
Install a fresh battery and allow 15 minutes for the CL510 to stabilize to ambient
temperature. Remove the EZ-Dial Knob, battery cover and the four black Phillips
head screws. While holding the CL510 face down in one hand, carefully
separate the top and bottom of the housing. Place the unit into calibration mode
by shorting the calibration via located on the bottom left side on the PCB with
tweezers and turning the UUT on. Verify the CL510 is in Calibration mode by
viewing the top Left of the LCD for the word CAL.
Source OHM Calibration:
LO Calibration:
1. Connect the CL510 terminals to an Ohm meter in four wire
Mode.
2. Slide the EZ-Check switch to the LO Position indicated by
displaying a LO on the left side of the display.
3. Dial the CL510 so the meter reads 0.00
W.
4. Press the EZ-Dial Knob down.
5. The display will flash STORED to confirm that the value was
stored.
HI Calibration:
1. Slide the EZ-Check switch to the HI Position indicated by
displaying a HI on the left side of the display.
2. Dial the CL510 so the meter reads 400.00
W.
3. Press the EZ-Dial Knob down.
4. The display will flash STORED to confirm that the value was
stored.
Completion of Calibration:
Turn the CL510 off. Next time the unit is turned on the CL510 will be Calibrated
and in normal operational mode.
Basic Keypad Operations
EZ-Check Switch
Slide the switch to select from three user stored
values for the desired calibration points.
The user can select HI, DIAL, and LO positions.
These values can easily be changed to suit the
calibration requirements.
ON/OFF Switch
Slide the ON/OFF to turn the CL510 on or off.
EZ-Dial Knob
The EZ-Dial Knob has two adjustment
speeds. Simply turning the EZ-Dial Knob will
select fine adjustments. While pressing down
and turning the EZ-Dial Knob will make
course adjustments.
Note: When the EZ-Check Switch is in the
HI or LO position, pressing and holding the EZ-Dial Knob without turning
HI or LO EZ-Check value.
will store a new
CL510 Configuration
Instructions for Enabling and Disabling the Configuration Options
1. Turn the CL510 on with the ON/OFF Switch .
2. Press the EZ-Dial Knob while the PRESS EZ-DIAL KNOB FOR
CONFIGURATION message is displayed.
3. Select options by turning the EZ-Dial Knob until the arrow points
to the desired option.
4. The option can be enabled or disabled by pressing the EZ-Dial
Knob.
The CL510 configuration menu will exit automatically after 5 seconds of
inactivity and go to normal operation with the options selected. These
options are recalled at turn on until they are changed again.
CL510 Configuration
CL510 Configuration Menu
Auto Off ON (default)/OFF
If Auto Off is ON, the unit will turn off after 30 minutes to save battery life, if there is no user activity. If
Auto Off is OFF the unit will stay on until it is turned off from the keypad. This is typically useful for
manual loading or continuous use.
Display Units °C (default)/°F
Pressing the EZ-Dial Knob to toggles between °C or °F
RTD {rtd type}
CL510: The RTD type is fixed as ordered
from the factory and cannot be changed.
CL510-7: To change RTD type, press the
EZ-Dial Knob. Turn the EZ-Dial Knob
to scroll through the list of available
types. Press again to save and return to
the configuration menu.
{rtd type} is one of:
Pt100
a
=3916, Pt100a=3926, Cu10a=427, Ni110 Bristol,
Ni120
{rtd type} is any of:
Pt100
a
=3916, Pt100a=3926, Cu10a=427, Ni110 Bristol,
Ni120
Menu
a
=3850 (default), Pt100a=3902, Pt100
a
=672,
W
a
=3850 (default), Pt100a=3902, Pt100
a
=672,
W
CL510 Series Operating InstructionsCL510 Series Operating Instructions
EZ-Dial Knob
Turn the EZ-Dial Knob to adjust the output up or down. Fine adjustments can be made by turning the
EZ-Dial Knob. Coarse adjustments can be made by pressing and turning the EZ-Dial. New values can
be stored into the HI and LO EZ-Check positions by pressing down on the EZ-Dial Knob until
STORED is indicated on the display.
EZ-Check Switch
The EZ-Check Switch has three positions: HI, DIAL, and LO. Its position is shown at the left edge of the
display with HI and LO indicators. The output is adjustable in all three positions. The EZ-Check
Switch allows user-selected values to be stored in the HI and LO positions when used in combination with
the EZ-Dial Knob.
To store new EZ-Check value(s):
1. Dial the display to match the desired stored outputs for the HI or LO positions.
2. Press down on the EZ-Dial Knob until the confirmation message STORED appears.
3. Recall the values by moving the switch between HI, DIAL, and LO.
Operational description: When returning to previously set EZ-Check positions, the DIAL position always
recalls the last output value it was dialed to. The HI and LO positions will recall the last STORED value,
NOT the last output value it was last dialed to.
Hint: For faster calibrations, the product has been designed so the position of the switch can be felt. This
tactile feature allows continuous monitoring of the device being calibrated without looking back at the
CL510 display. This is also useful in poor lighting or under difficult operating conditions.
Specifications
General Specifications:
(Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from calibration)
Temperature Range -25 to 60 °C (-10 to 140 °F)
Relative Humidity Range 10 % RH 90 % (0 to 35 °C), Non-condensing
Size 4.9 X 3.15 X 1.82 inches (125.5 X 80 X 46.2 mm)
Weight 9.1 oz (258 grams)
Battery 9V Alkaline provides 45 hours of continuous use
Miscellaneous Low battery indication with nominal 1 hour of operation left