Omega Products CL310 User Manual

Documenting Multifunction Calibrator
Portable, High Accuracy
CL310
Standard
U Highest Accuracy in Class
to 0.015% of Reading
U Measure and Simulate T/Cs (13 Types), RTDs
(13 Types), Ω, Current, Voltage, Frequency and Pressure; Source Pulse Trains
U Pressure Module Communication
U Direct Entry of Custom RTD Coefficients
(R0, A, B, C)
U All Simulate Modes Can be Programmed
with Dedicated Setpoints to Speed Calibration and Linearity Tests
U Store Up to 21 Data Points Per
Instrument, and Up to 50 Instruments in Non-Volatile Memory
U Test for Pass/Fail Criteria U Capture Both as Found and as Left Data U Software is Included U Protective Rubber Boot Standard
®
The CL310 documenting multi-function calibrator delivers a feature set unmatched in high accuracy, hand-held calibrators in its price range. The CL310 delivers functions and accuracy associated with fixed installation, laboratory instruments, and has everything needed for virtually any calibration task. What makes this versatile calibrator best in class is the ability to fully document any calibration easily while performing the normal job. Software is included. Some functions will require a Microsoft Excel-compatibility.
CL310 can also measure and source thermocouples, RTDs, current, voltage, frequency, and it can source pulse trains. It has a communications port for pressure modules and an isolated mA/V readback circuit. Arrow keys, direct numeric keypad entry, and three software-driven function buttons combine with a large, backlit, menu-driven graphics display to provide a highly intuitive yet powerful interface. A built-in 250 Ω resistor for Hart® compatibility; compatibility with smart transmitters and PLCs, full fuseless protection, and a serial communications port for control using ASCII commands are additional features that make the CL310 indispensable for calibration tasks. This product is battery operated.
CL310, shown with protective rubber boot (included).
Measuring Pressure
With a PCL-PMA pressure module adaptor, the CL310 will work with all OMEGA® PCL-PM pressure modules. (For information on these see specifications under PCL1200 visit omega.com/pcl1200).
Specifications (23 ±5°C unless otherwise noted)
Voltage: Source: 0.000 to 20.000 Vdc Read:
Isolated: 0.000 to 30.000 Vdc Non-Isolated: 0.000 to 20.000 Vdc
Thermocouple mV: Read/Source: -10.000 to 75.000 mV Current (mA): Source: 0.000 to 24.000 mA Read: Isolated: 0.000 to 24.000 mA Non-Isolated: 0.000 to 24.000 mA
K-43
Frequency (1 to 20V Selectable Amplitude):
CPM Source/Read: 2.0 to 600.0 cpm Hz Source/Read: 1.0 to 1000.0 Hz kHz Source/Read: 1.0 to 10.00 kHz
Pulse (Source Only 1 to 20V; Selectable Amplitude): Pulses: 1 to 30,000; 2 cpm to 10 kHz
Ohms: Source: 5.0 to 4000.0 Ω Read: 0.00 to 4000.0 Ω
Thermocouple Range Read and Source: Type J: -200.0 to 1200.0°C (-328 to 2192°F)
Type K: -200.0 to 1370.0°C (-328 to 2498°F) Type T: -200.0 to 400.0°C (-328 to 752°F) Type E: -200.0 to 950.0°C (-328 to 1742°F) Type R: -20.0 to 1750.0°C (-4 to 3182°F) Type S: -20.0 to 1750.0°C (-4 to 3182°F) Type B: 600.0 to 1800.0°C (1112 to 3272°F) Type C: 0 to 2316.0°C (32 to 4201°F) Type XK: -200.0 to 800.0°C (-328 to 1472°F) Type BP: 0 to 2500.0°C (32 to 4532°F) Type L: -200.0 to 900.0°C (-328 to 1652°F) Type U: -200.0 to 400.0°C (-328 to 752°F) Type N: -200.0 to 1300.0°C (-328 to 2372°F)
RTD Range Read and Source: Ni120 (672): -80.0 to 260.0°C (-112 to 500°F)
Pt100 (385): -200.0 to 800.0°C (-328 to 1472°F) Pt100 (3926): -200.0 to 630.0°C (-328 to 1166°F) Pt100 (3916): -200.0 to 630.0°C (-328 to 1166°F) Pt200 (385): -200.0 to 630.0°C (-328 to 1166°F) Pt500 (385): -200.0 to 630.0°C (-328 to 1166°F) Pt1000 (385): -200.0 to 630.0°C (-328 to 1166°F) Cu10: -100.0 to 260.0°C (-148 to 500°F) YSI400: 15.00 to 50.00°C (59 to 122°F) Cu50: -180.0 to 200.0°C (-292 to 392°F) Cu100: -180.0 to 200.0°C (-292 to 392°F) Pt385-10: -200.0 to 800.0°C (-328 to 1472°F) Pt385-50: -200.0 to 800.0°C (-328 to 1472°F)
Accuracy
Voltage: ±0.015% of reading, ±2 mV Thermocouple: mV ±0.02% of reading, ±10 µV Thermocouple Errors (Typical at 100°C)
(In °C; Add 0.2 for Cold-Junction Compensation Error) Type J: 0.2°C (0.4°F) Type K: 0.3°C (0.5°F)
Type T: 0.2°C (0.4°F) Type B: 1.2°C (2.2°F) Type S: 1.2°C (2.2°F) Type XK: 0.2°C (0.4°F) Type C: 0.6°C (1.6°F) Type L: 0.2°C (0.4°F) Type BP: 0.9°C (34°F) Type N: 0.4°C (0.7°F) Type U: 0.25°C (0.5°F)
RTD Read/Source: Ni120 (672): 0.06°C (0.11°F) Pt100 (385): 0.13°C (0.23°F)
Pt100 (3926): 0.13°C (0.23°F) Pt100 (3916): 0.13°C (0.23°F)
Pt200 (385): 0.45°C (0.81°F) Pt500 (385): 0.21°C (0.38°F)
Pt1000 (385): 0.14°C (0.75°F)
YSI400: 0.05°C (0.09°F) Cu50: 0.20°C (0.36°F) Cu100: 0.13°C (0.23°F) Pt385-10: 0.84°C (2.16°F) Pt385-50: 0.21°C (0.38°F)
Read/Source: Current (mA): ±0.015% of reading, ±2 µA/±0.015% of
reading, ±2 µA
CPM: ±0.05% of reading, ±0.1 Hz: ±0.05% of reading, ±0.1 Hz
kHz: 1 to 10 kHz/±0.250% of reading Ohms (Works with All Pulsed Transmitters Down to 5 ms): 5 to 400 Ω Range: ±0.015% of reading ±0.03 Ω/ ±0.015% of
reading ±0.1 Ω
400 to 1500 Ω Range: ±0.015% of reading ±0.3 Ω
1500 to 4000 Ω Range: ±0.015% of reading ±0.3 Ω
Cu10: 0.82°C (1.48°F)
PCL422-TL, test leads.
DPG4000-C32
RS232 cable
sold separately.
CL310 calibrator,
$2695, complete
with USB cable,
and 4 “AA” alkaline
batteries, protective
rubber boot, software
CD and NIST
calibration certificate.
USB adaptor cable.
Nylon carrying case.
Environmental
Operating Temperature: -10 to 50°C (14 to 122°F) Storage Temperature: -20 to 70°C (-4 to 158°F) Stability: ±0.005% of reading/°C (outside of 23°C ±5°C) Power Requirement: 6 Vdc Batteries: 4 “AA” alkaline (included) or optional rechargeable
Mechanical
Dimensions: 220.9 H x 106.6 W x 58.4 mm D (8.7 x 4.2 x 2.3") Weight: 863 g (30.5 oz)
SM
OMEGACARESM extended warranty program is available for models shown on this page.
Extended Warranty
Program
To Order
Ask your sales representative for full details when placing an order. OMEGACARE covers parts, labor and equivalent loaners.
Visit omega.com/cl310 for Pricing and Details
SM
Model No. Description CL310 Documenting multifunction calibrator
Accessories Model No. Description PCL-PMA Pressure module adaptor DPG4000-C32 RS232 cable with miniature plug PCL422-TL 1.2 m (48") stackable test leads CL20-KIT-2 Thermocouple test lead kit PCL-ACA AC adaptor/charger 120 Vac
Comes complete with instruction manual, test leads, protective rubber boot, nylon carrying case, USB cable, Type K thermocouple, Type K cable, NIST calibration certificate, software CD and 4 “AA” alkaline batteries.
Ordering Examples: CL310, high accuracy multifunction calibrator, and CL20-KIT-2, thermocouple test lead kit.
CL310, high accuracy multifunction calibrator, and PLC-ACA AC adaptor. OCW-1, OMEGACARE to a total of 2 years.
SM
extends standard 1-year warranty
K
K-44