Lab Calibrator
CL3001
Standard
U 0.0025% Calibration Accuracy
U Source/Read Thermocouple, RTD,
Voltage, Current and Pressure
U Custom RTD and SPRT Profiles
U RS232, USB and IEEE-488
Remote Control
U Isolated Measurement Channel
U Includes NIST Certificate with
Calibration Data
The OMEGA® CL3001 calibrator is an accurate fullfeatured temperature, DC and pressure calibrator
intended for R&D, manufacturing and calibration lab
applications. The units simple design and ease of
operation allow users to quickly familiarize themselves
with its operations and features. Time saving functions
like the ability to save, recall and automatically cycle
through setpoints for each output range, the ability
to enter user definable RTD curves, and a complete
remote interface are several key features offered by
the CL3001.
The CL3001 includes an isolated measurement
channel consisting of two voltage ranges: 10V and
100V DC, milliAmp range 0 to 52 mA, milliAmp range.
The isolated measurement includes 24 Vdc power,
and accuracy of 0.005% of reading on voltage ranges.
Measuring pressure with a PCL-PMA pressure module
adaptor, the CL3001 will work with all OMEGA
PCL-PM pressure modules. For information on these
see specifications under PCL1200, please visit
omega.com/pcl1200
CL3001
Specifications
Accuracy: Stated in terms of absolute uncertainty for 1 year
term. Thermocouple values include cold junction
compensation. Values are best in given range and function.
For breakdown of each range please refer to the manual
omega.com/cl3001
Display: Primary input/output display and the isolated
measurement channel display each have 2 line,
16 character fields
Temperature Units: User selectable, °F/°C
Communication: RS232, IEEE-488 (31 available addresses)
Operating Ambient: Less than 80% RH, 0 to 50°C
(32 to 122°F), temperature Cal 18 to 28°C (64 to 82°F)
Power: 100V/120V or 220V/240V, ±10%, line frequency
47 to 63 Hz
Dimensions: 13.3 cm H (5.25") plus 2.9 cm (1.15") for
extended feet width standard rack width 48.3 cm (19")
30.0 cm D (11.81") overall
Weight: 4 kg (9 lb)
Input/Output Functions: Include save, recall and
automatically cycle through setpoints for each output range, the
ability to enter user definable RTD curves, and remote interface
command set.
Thermocouple Specification, Output/Input
Range Uncertainty
Type °C °F °C °F
K -200 1372 -328 2502 0.16 0.29
J -210 1200 -346 2192 0.16 0.29
T -250 400 -418 752 0.14 0.25
E -250 1000 1000 1832 0.15 0.27
N -200 1300 -328 2372 0.18 0.32
L -200 900 -328 1652 0.17 0.31
U -200 600 -328 1112 0.56 1.01
XK -200 800 -328 1472 0.13 0.23
R 0 1750 32 3182 0.33 0.59
S 0 1750 32 3182 0.36 0.65
B 600 1820 1112 3308 0.39 0.70
C 0 2316 324 201 0.26 0.47
BP 0 2500 324 532 0.32 0.58
†
Values are best in range
K-41
Absolute
†
DC Voltage Specifications, Output
Absolute Uncertainty
± (ppm of output µV)
†
Range Ppm output µV Resolution
0 to 100.000 mV 30 3 1 mV
0 to 1.00000V 30 10 10 µV
0 to 10.0000V 30 100 100 µV
0 to 100.000V 30 1 mV 1 mV
Resistance Specifications, Input
Absolute Uncertainty
†
± (ppm of reading Ω)
Range Ppm rdg Ω Resolution
5 to 400.000 Ω 20 0.035 0.001 Ω
5 to 4.00000 kΩ 20 0.35 0.01 Ω
1. Loop power: 24V ±10%
2. HART resistor: 250Ω ±3%
3. Maximum rated loop current: 24mA
DC Voltage Specifications, Isolated Input
Absolute Uncertainty
†
± (ppm of reading mV)
Range Ppm rdg mV Resolution
0 to 10.0000V 50 0.2 100 uV
0 to 100.000V 50 2.0 1 mV
DC Current Specifications, Output
Absolute Uncertainty
†
± (ppm of output µA)
Range Ppm rdg µA Resolution
0 to 100.000 mA 50 1 1 uA
DC Current Specifications, Isolated Input
Absolute Uncertainty
†
± (ppm of reading µA)
Range Ppm rdg µA Resolution
0 to 50.000 mA 100 1 0.1 uA
1. Loop power: 24V ±10%
2. HART® resistor: 250 Ω ±3%
3. Maximum rated loop current: 24 mA
Resistance Specifications, Output
Absolute Uncertainty
†
±Ohms
Nominal
Range Ω Resolution Current
5 to 400.000 Ω 0.015 0.001 Ω 1 to 3 mA
5 to 4.00000 kΩ 0.25 0.01 Ω 0.1 to 1 mA
1. Continuously variable from 0 to 4 kΩ.
2. For currents lower than shown, the specification
becomes:
New Spec. = Stated Spec. x Imin/Iactual.
For example, a 500 µA stimulus measuring 100 Ω
has a specification of:
0.015 Ω x 1 mA/500 µA = 0.03 Ω.
†
Values are best in range
SM
OMEGACARESM extended warranty program
is available for models shown on this page.
Extended Warranty
Program
Ask your sales representative for full details
when placing an order. OMEGACARE
covers parts, labor and equivalent loaners.
SM
RTD and Thermistor Specifications, Output
Range Uncertainty
Absolute
†
Type °C °F °C °F
Pt 385, 100 Ω
Pt 3926, 100 Ω
Pt 3916, 100Ω
Pt 385, 200 Ω
Pt 385, 500 Ω
Pt 385, 1000 Ω
Ni120, 120 Ω
Cu 427, 10 Ω
YSI 400
-200 800 -328 1472 0.04 0.07
-200 630 -328 1166 0.04 0.07
-200 630 -328 1166 0.03 0.05
-200 630 -328 1166 0.38 0.68
-200 630 -328 1166 0.15 0.27
-200 630 -328 1166 0.07 0.13
-80 260 -112 500 0.02 0.04
-100 260 -148 500 0.38 0.68
15 50 59 122 0.007 0.013
RTD and Thermistor Specifications, Input
Absolute
Range Uncertainty
†
Type °C °F °C °F
Pt 385, 100 Ω
Pt 3926, 100 Ω
Pt 3916, 100 Ω
Pt 385, 200 Ω
Pt 385, 500 Ω
Pt 385, 1000 Ω
Ni120, 120 Ω
Cu 427, 10 Ω
YSI 400
SPRT
-200 660 -328 1220 0.06 0.011
†
Values are best in range
To Order
-200 800 -328 1472 0.012 0.07
-200 630 -328 1166 0.011 0.022
-200 630 -328 1166 0.006 0.011
-200 630 -328 1166 0.009 0.016
-200 630 -328 1166 0.008 0.014
-200 630 -328 1166 0.012 0.022
-80 260 -112 500 0.010 0.018
-100 260 -148 500 0.069 0.124
15 50 59 122 0.007 0.013
Visit omega.com/cl3001 for Pricing and Details
Model No. Description
CL3001 Precision lab calibrator
Comes complete with NIST certificate, operator’s manual,
power cord 0.9 m (3'), and thermocouple shorting connector.
Ordering Example: CL3001, precision lab calibrator.
OCW-2 OMEGACARESM extends standard 1-year warranty to a total
of 3 years.
Accessories
Model No. Description
PCL-PMA Pressure module adaptor
CL-300-CABLE-(*)-2
TAC-CAB 0.9 m (3') test leads
PCL422-TL Stackable test leads
Male mini-connector to spade lug, insert one thermocouple type
*
J, K, T, E, R, S, B, N.
Thermocouple extension cables
K
K-42