Electrical Connection Terminal Barr
Cable InletCale Knockout ½“ BSP (2x) and ¼”BSP (2x)
Installation Flat Surface
Relative Humidity 20 to 80 %
Electrical Power90 to 240 VAC (50/60 Hz)
Power Consumption 3.5 VA
Sample Flow250 to 750 ml/min
Sample Pressure
(Min./max.)
Sample Temperature5 to 40°C
Weight 6 Kg (equipment and bubble trapper)
TRANSMITER
Analogic Output 2x 4 to 20 mA
Digital Output RS-485
CONTROLER
Type ON-OFF / PWM One,NO (1A / 250 VAC)
Set-Point from 0 to 100% of scale
ACCESSORIES
Instruction Manual
Installation Hardware
Cast Aluminum SAE 323(BASE)
Analyzer / Transmiter / Turbidity Controller
ABS (Frontal Lid)
10 / 20 psi
2 aluminum base SAE 323, 2 “U” clamps in SS 316,
Nuts and washers in SS 316
3
Page 5
3.
The equipment is offered in a compact cast aluminum case, SAE-323 with lower oxidation power,
treated against corrosion, finished with epoxy-electrostatic paint and expanded polypropylene panels
(PU). With reduced dimensions and very light weight, is build under IP-67 protection.
Under the same case, user will find: Local Indicator, Analyzer, Transmitter and Controller of ease
operation. The control installation can be done on Flat Surface or on 2" tube.
The electrical connection is made by a terminal block located at the lower portion of the controller and
cables will go thru 2 cable knockouts of ½".
Mechanical Description
15
6
9
7
10
11
8
PARTS DESCRIPTION:
1 - Frontal Lid in ABS
2 - Alphanumeric Display 2 lines x 16 characters
3 - 3 keys tactile membrane:
= Selects the desired option, flashing option.
<SELECT>
= Confirms the selected option chosen at SELECT key
<ENTER>
= Goes back one step or moves back to prior screen
<ESCAPE>
4 - Cable knockouts (2x) of ½" BSP and (2x) of ¼“ BSP for instrumentation and power cables
5 - Sensor / Turbidity Detector
6 - Reading Chamber
7 - Bubble Trapper (to remove bubbles)
8 - ½’NPT Female Valve - Cleaning Drainage
9 - 3/4” NPT Female Valve - Sample Inlet
10-Sample Drainage (ø 3/4”)
11-Calibration Cup
2
3
4
4
Page 6
4.
Turbidity Measurement Principles
TURBIDITY MEASUREMENT
The measured Turbidity of determined sample, is the reading referred by light
dispersed and absorption that goes thru the sample.
Turbidity does not mean suspended solids measurement, but a effect determination
of light refracted through solids.
Turbidity measurements are used to evaluate the quantity of water sample. To
understand turbidity it is helpful to think about the characteristics of mixtures.
A mixture is defined as Homogenous or Monophasic, when the components of
the mixture are uniformly dispersed, or dissolved, throughout the mixture, Salt water
is an example of a homogenous mixture or Solution. Heterogenous or Polyphasic
mixtures are those whose components are not well distributed throughout the
mixture. The components may simple be floating in the mixture, given enough time
these components will settle out. This type of mixture is also called Suspension.
The clarity of a water sample is directly related to the amount of material suspended in
the water. Turbidity measurements are a way to measure the amount of material
suspended in a water sample. They are commonly used to monitor the effectiveness of
filtration processes.
The OMEGA Turbidity meter uses the principle of Nephelometry. As light of a known
intensity is passed through a sample some of the light will be scattered by particles in the
sample. A detector is placed at 90° to the sample to collect the scattered light. The
intensity of the scattered light is compared to that of the source. This value is reported as
a Nephelometric Turbidity Unit, or NTU. The meter is calibrated with Formazine
standards of a known NTU.
Water can sometimes have color of its own (from humic sources, for example). This
color may interfere with the light source, absorbing some of the light, and reducing the
amount of light reaching the detector. This can cause the measurement to be skewed.
The OMEGA Turbidity Meter LED light source emits at a range close to infrared. In this
range absorption is minimized, and turbidity measurements are more accurate.
5
Page 7
4.
Turbidity Measurement Principles (cont.)
Vial
90°
Light Emitting
near “ IR”
Reflected Light @ 90°
Detector
As detection involves the difference between the light reaching the vial and
transmitted by the sample placed at the cuvette, it is convenient to minimize the effects
that diminish the light intensity transmitted, among those the most important, it is the
absorption caused by the samples color.
Because of this it is essential to work in a light wavelenght where absorption would be
minimum (close to infrared), if we work in a electromagnetic spectrum visible region, this
color interference would certainly alter the results.
In order to quantify the turbidity, we can use many comparison standards, being NTU
(Nephelometric Turbidity Unit) scale the most common, patronized from formazine
standard suspensions. This way we would have a comparison standard scale between
different materials, in a way we can evaluate turbidity with precision.
Transmitted
Light
Bibliographic References
Bela G. Lipták (chief editor) Analytical Instrumentation;
Howard A. Strobel and William R. Heineman, Chemical Instrumentation, A Systematic
Approach.
6
Page 8
5.
Dimensional
Below you will find panel dimensional for installation in flat surface and distances between support
holes.
5
2
1
4
5
1
Sample
Inlet
Drain
Drain
144
Cable Length
4
Ø14
Electrical
Power
Dimensions in Millimeters
7
Page 9
6.
Equipment Installation
Follow rigorously below instructions:
1- Remove the equipment from its box and verify for any possible damages, caused by the
transportation.
2- Install the controller in a strategic place, of ease access and operation, free of vibrations and
vapors.
3- Avoid direct exposure of the controller to solar rays and in case if necessary, protect the controller.
4- Proceed with terminal connections of cables at terminal barr.
5- Inspect the electrical installation in order to certify that all connections are correct.
6- Verify the power that is being supplied and make sure to connect it the proper place.
7- Connect the circuit breakers from the power distribution box.
3.1. Important Recommendations
3.1.1. The equipment must be Being
Electrical Powerindependent
from other system components.
so, the power cable of Control Valves, Solenoids, Alarms, etc, must be connected directly at the
Distribution Box, and "" at the equipment connection board.
3.1.2. if the equipment are providing the proper sealing to the cable inlets.
Verify cable knockouts
This proceeding is a must in order to preserve enclosure as
3.1.3. . It reduces impedance, generating measuring errors. Verify the
Be careful with Humidity!!
cable knockout and if necessary, using a hair dryer.
3.1.4. s length,
Do not cut or try to attach the connection cable.
never
IP-67.
dry the connection block
In case you need to change it
Contact OMEGA Customer Service.
RF Shield
Meas.
Sensor Cable
Electr. Shield
Yellow
Green
3.1.5. ON-OFF are , offering innumerous advantages for the equipment, such
outputsthyristor
type
as: no presence of sparks, faster commutation, noise practically un-existent, no presence of RF
interference and many more.
The outputs can commute any charge, since they are
250V / 1A
.
powered by alternate tension (VAC), limited to
8
Page 10
7.
CN2 - Connector 2
Slots Connection
1 and 2 Electrical Power for 90/240 Vac 50/60Hz connection
3 Ground
4 and 5 SP2 - Set-Point 2 for alarm or control
6 and 7 SP1 - Set-Point 1 for alarm or control
8 and 9 Digital Transmission Output 4 to 20 mA
F1 General Fuse (3.0A)
F2 Set-Point 2 Fuse (1.0 A)
F3Set-Point 1 Fuse (1.0A)
Interconnection Barr
9
Page 11
8.
S
e
n
s
o
r
R
S
4
8
5
m
A
S
P
1
S
P
2
Electrical Schematic
Cn2
NO
C
NO
C
MAIN BOARD
Cn1
Power
90 - 240Vac
Female
Connetor
Cn2
Shie
Gre.
Connector Description
Yellow
Colorless
RF Shield
Cn1
Meas.
Yellow
Green
Ground
Cable 5
Cable 4
Cable 3
RF Shield
Electr. Shield
Cable 1
Cable 2
Detector
Cable 6
Converter
R
S
T
Box
Electrical
Note: It’s very important to
use phases and cables for
Equipments and valves.
N
Distribution
Controls
RS485/RS232
Microcomputer
or
Alarm
Graphic
SDLC
PLC or
Register or
Sensor / Turbidity
Bubble
Trapper
10
Page 12
9.
Lens and Bubble Trap Cleaning
Sensor Unit Lens and Bubble Trapper Cleaning
LENS CLEANING OF SENSOR UNIT
Remove the Sensor/Turbidity Detector (5) by turning it clockwise (45degrees) and pull it up. Using a
soft sponge and neutral soap, wash the Led and the photocell Lenses. After this, dry it using a soft
Absorbent paper, a type that will not leave residues.
Sensor Unit
Photocell
LED
BUBBLE TRAPPER CLEANING
In order to clean the Bubble Trapper, open the ½”NPT Female Valve (8), so the accumulated dirt can
be remove from the bubble trapper. Leave the process to flow for about 2 minutes, if noticed that the
process has residues it is necessary to remove the bubble trapper lower lid and wash it using
detergent.
Then place the valve back in place, close it and place the equipment in operation.
Bubble Trapper Body
Lower Lid
Turn clockwise
11
Page 13
10.
It is recommended to calibrate this instrument every three months or as programmed routine.
1) Prepare the Standard Solutions based on the values chosen during Set Up Mode.
2) Clean the Calibration Cup (11).
3) Close the Sample Inlet Valve (9).
4) Open the Drainage Valve (8), so the sample located inside the bubble trapper can be removed.
5) Insert the Calibration Cup (11) inside the Reading Chamber at the Bubble Trapper Internal Body
as shown below.
Calibration Information
Calibration
Cup (11)
Reading
Chamber (6)
Bubble Trapper
Internal Body
6) Insert the first Standard Solution into the Calibration Cup (11) until it overflows. Then place the
Sensor Unit (5) at the Reading Chamber (6), wait about 10seconds then follow the Calibration Menu
of the equipment (page 20).
7) Access the equipment calibration menu and follow the steps required.
Note: when replacing the calibration standards from the calibration cup, the cup MUST be washed
and dried, making sure no residues is being left inside the cup, for next calibration points.
8) Once executed the hole calibration process, remove the Calibration Cup (11), close the Drain
Valve(8) and place the equipment in operation.
Note: verify the sample flow (from 700 thru 2000ml.min.).
12
Page 14
11.
Turbidity Standard
Instructions on how to prepare Calibration Solutions
1. Standard Zero
In order to obtain a turbidity close to Zero, use a good quality deionized or distilled water and filter it twice in
a roll, using a 0.2
(Blank).
Note: this water will be used to dilute the standard.
2. Standard Solutions
Formazine Standard must be purchased separate. Digimed offers 100mL bottle of 4000NTU Stabilized
Standard Solution, model# TRS-444H, for dilution and calibration purpose.
1 - Necessary Materials:
1.1 - 1 Volumetric Flask 100 mL
1 Volumetric Pipette 50 mL
2 L Distilled or Deionized Water, Filtered (0.2 µm)
2 - Calibration Solution 1000 NTU
2.1 - In a 100ml Volumetric Flask, using the Volumetric Pipette add 25ml of 4000NTU Stabilized Standard
Solution.
2.2 - Add filtered water up to the mark on the flask.
2.3 - Before using the solution mix it by gently inverting the flask several times. Avoid creating bubbles.
2.4 - The Solution is Valid for 15 days.
Note:
place.
3 - Calibration Solution 500 NTU
2.1 - In a 100ml Volumetric Flask, using the Volumetric Pipette add 50ml of 1000NTU Stabilized Standard
Solution.
2.2 - Add filtered water up to the mark on the flask.
2.3 - Before using the solution mix it by gently inverting the flask several times. Avoid creating bubbles.
2.4 - The Solution is Valid for 15 days.
Note:
place.
4 - Calibration Solution 100 NTU
3.1 - In a 100ml Volumetric Flask, using the Volumetric Pipette add 10ml of 1000NTU Stabilized Standard
Solution.
3.2 - Add filtered water up to the mark on the flask.
3.3 - Before using the solution mix it by gently inverting the flask several times. Avoid creating bubbles.
3.4 - The Solution is Valid for 10 days.
3.5 - After its use, discharge the solution.
Note:
place.
5 - Calibration Solution 10 NTU
4.1 - In a 100ml Volumetric Flask, using the Volumetric Pipette add 1ml of 1000NTU Stabilized Standard
Solution.
4.2 - Add filtered water up to the mark on the flask.
4.3 - Before using the solution mix it by gently inverting the flask several times. Avoid creating bubbles.
4.4 - Stir it manually for before using it.
4.4 - The Solution is Valid for 5 days.
4.5 - After its use, discharge the solution.
Place.
Note:
µm filter and teorically you will obtain a water with 0.12NTU, that can be considered Zero
For best storage conditions place the solution in a dark bottle and store it in a fresh and dark
For best storage conditions place the solution in a dark bottle and store it in a fresh and dark
For best storage conditions place the solution in a dark bottle and store it in a fresh and dark
For best storage conditions place the solution in a dark bottle and store it in a fresh and dark
Attention: for TURBIDITY, this instrument MUST BE calibrated using
Formazine Standard and cannot be calibrated using Polymers!!!
13
Page 15
12.
Equipment Operation
Set Up Procedures
The equipment offers a non-volatil memory (E2PROM), in order to store operations functions (resolution,
reading, Calibration and more). Even when turned off from power, all functions chosen during set up will remain
stored.
Before starting any work with the equipment, it is recommended to to certify
that you have chosen the correct options for the operation .
When at the , press key in order to select the desired function, flashing
option, then press , press key until SET function flashes, then
press . A Password will be requested, press in sequence ,
<ENT>
flashing option, press key until the desired option flashes then press
In order to move to the next screen, user must press .
SELECT FUNCTION
<ENT>
<ENT>
, then follow step by step the options shown at the screen. In case the user desires to change the
key to confirm the option chosen
<ESC>
key. In order to access the
<SEL>
menu
<SEL>
SET UP
<ENT>
key
verify the SET UP parameters,
<SEL>
<SEL>
<ENT>
key to confirm the option.
Read Operation
At this operation user will have options to and. In case the desire is to , press
<SEL>
guide the user step by step on how to proceed with the perfect calibration. In case the desire is to Read, press key
until option key to confirm, then the display show the following form:
<ENT>
key until . From this point on the program will
Read
option flashes, then press key to confirm the option chosen
Cal
flashes, then press
CALIBRATEREADCALIBRATE
<ENT>
4
the Sensor
<SEL>
»»» »»»
-> »»»»» »»»
1
1- The "Prompt" is a signal that flashes every time a reading is performed.
2- The measured value
3- The Unit (NTU)
4- Contact Outputs Conditions
2
3
14
Page 16
12.
Equipment Operation (cont.)
menus are self explanatory with its respective options, that are simply selected by pressing
T
<SELECT>
key. After choosing the selection (flashing option), press key, to confirm
<ENTER>
the slected option.
If a mistake occur or user desires to change option already chosen, press , to go back to
<ESCAPE>
prior step. At every touch the screen will move back one step back. While in Reading Mode,
SCAPE>
<E6 seconds to exit this mode and return to main
key must be pressed and hold for about
menu.
OMEGA
MOD:TRCN441
DISPLAY TEST
»»»»»»»»»»»»»»»»
»»»»»»»»»»»»»»»»
Interpreting the Read menu. The Set Points will be
displayed and by the box on the side of each
contact, user can find out the status of these
contacts.
Indicates that the contact is On and operating at
the moment.
Indicates that the contact is On but not operating
at the moment.
Indicates that this contact was set as Off
X
All contacts are programed during Set Up Mode.
Press <ESC> key and hold in order to exit the
Reading Mode and access the Main Menu.
S1:» S2:»
-> 120.01 NTU
TURBIDIMETER
/SET UPREAD
NOTE:
While during Reading, press key and the instrument will go to , turning off the
<ENTER>STAND-BY
alarm and control outputs.
Note A - Every time you see the symbols “ ” “ ”, that means that the user can
adjust the displayed value up or down.
<SEL>
To increase the value press key until “ ” flashes, then press to
confirm, then press key and at every touch the value will increase by one
unit.
To decrease the value presskey until “ ” flashes, then press to
confirm, then press key and at every touch the value will decrease by one
unit.
If a mistake is made, press key to return and correct the value!
<SEL>
<SEL>
<SEL>
<ESC>
>> and <
<
<ENT>
<ENT>
15
Page 17
12.
Press until flashes then press
<ENT>
Equipment Operation - Set Up
<SEL>Set Up
key.
TURBIDIMETER
READ/SET UP
A Password is required in order to access the
UP<SEL> <ENT><ESC>
. Press in sequence, and
In order to select the desired language, press
<SEL>
press key to confirm.
User can program the instrument, such as Default
Calibration, On Line Calibration and more. If chosen ,
the last configuration will remain in effect. Press
key until the desired option flashes, then press
Key to confirm .
The Range cannot be changed. It is default from
key until the desired option flashes, then
<ENT>
You can calibrate the instrument as factory default.
Choose and confirm and the instrument will
calibrate as factory default! This option is offered in
case the user does have any other way to perform
a calibration procedure.
The user will be allowed to adjust the Turbidity
value read at sample. Press until
flashes then confirm by presssing .
Yes
<SEL> MAN
<ENT>
SET
No
<SEL>
<ENT>
PASSWORD
___
LANG.: PORT./
/SPANISHENGLISH
CONFIG. INSTR.?
/ YESNO
RANGE
0 - 10000 NTU
FACTORY DEFAULT?
/ YESNO
CONFIRM?
/ YESNO
ON LINE CAL.:
MANUAL / AUTO
Go to Page 17
A
User will be able to choose the number of calibration
Points. Presskey until the desired opiton
flashes and confirmed by pressing
User will be able to adjust the value of the
calibration points chosen. Refer to page 16 (Note
A) for instructions on how adjust this value.
User can choose between the following Reading Modes:
Continuous - Read continuously after time is set
Average - Reads the average after time is set
User can program the time between Readings for
Average Reading.
Refer to Page 16 (Note A) for instructions on how to
adjust this time.
<SEL>
<ENT>.
CAL. POINTS:
1 / / 32
CALIBRATION
Pt.1 100.00NTU<>
CALIBRATION
Pt.2 200.00NTU<>
READING MODE:
/CONTIN. AVER.
TIME BETWEEN
READINGS:5s <>
A
Go to Page 17
16
Page 18
12.
Equipment Operation - Set Up (cont.)
From Page 16
A
User can program the display to show information like
Barr graph, Clock and more.
User can choose if desire to have Barr graph shown
above the Reading screen. When Bar Graph is
displayed, the information about Sensibility and Sample
Temperature will not be displayed!
User can now adjust the Minimum and Maximum
values for the Bar Graph. Refer to Page 16 (Note A) for
instructions on how to modify these values.
User has the option to configure Contacts SP 1and SP 2.
Press key to choose the desired option then
<SELECT>
press key to confirm.
<ENTER>
If user chooses Contact S1 as On, while at Reading Mode,
a or a will be displayed after the S1, indicating that
this Contact is On. If user chooses Contact S1 as Off, while
at Reading Mode, a will be
User can adjust the value for SP 1.
Refer to Page 16 (Note A) for instruction on how to adjust
this value. Resolution will depend upon the Resolution
chosen while at Set Up.
User can choose the Acting desired.
Press key to choose the desired option then
<SELECT>
press key to confirm
<ENTER>
X
DISPLAY CONFIG.?
/ YESNO
BARGRAPH?
/ YES NO
Min: 0.00NTU
Max:10000.00NTU
CONTACTS CONFIG?
/ YESNO
CONFIG. CONTACT
S1
CONFIG. S1:
/OFFON
SET POINT S1:
5.00 NTU <>
ACTING:
/REVERSEDIRECT
<
>
<
>
Go to Page 18
A
User can adjust the value for Hystereses.
Refer to Page 16 (note A)for instruction on how to adjust
this value. Resolution will depend upon the Resolution
chosen while at Set Up.
For the Burn Out configuration, user will have three options as:
On - Contact is going to stay On all the time, Off - contact will stay
Off all the time or Hold - contact is going to follow the last situation,
before going to Hold status.
<SEL>
Presskey until the desired option flashes, then press
<ENT>
If user chooses Contact S2 as On, while at Reading Mode,
a or a will be displayed after the S1, indicating that
this Contact is On. If user chooses Contact S1 as Off, while
at Reading Mode, a will be
to confirm.
X
HYSTERESE:
0.00NTU <>
BURN-OUT CONFIG.
STAND-BY: HOLD
CONFIG. CONTACT
S2
CONFIG. S2:
/OFFON
Got to Page 18
A
17
Page 19
12.
User can adjust the value for SP 2.
Refer to Page 16 (Note A) for instruction on how to adjust
this value. Resolution will depend upon the Resolution
chosen while at Set Up.
Equipment Operation - Set Up (cont.)
SET POINT S2:
20.00 NTU <>
User can choose the Acting desired.
Press key to choose the desired option then
<SELECT>
press key to confirm
<ENTER>
User can adjust the value for Hystereses.
Refer to Page 16 (note A)for instruction on how to adjust
this value. Resolution will depend upon the Resolution
chosen while at Set Up.
For the Burn Out configuration, user will have three options as:
On - Contact is going to stay On all the time, Off - contact will stay
Off all the time or Hold - contact is going to follow the last situation,
before going to Hold status.
Presskey until the desired option flashes, then press
<SEL>
<ENT>
to confirm.
User has the option to configure Current.
Press key to choose the desired option then
<SELECT>
press key to confirm.
<ENTER>
User has the option to choose using Output mA.
<SELECT>
Press key to choose the desired option then
<ENTER>
press key to confirm.
User can adjust the value the values for 4 and 20mA
Current Output.
Refer to Page 16 (Note A) for instruction on how to adjust
this value. Resolution will depend upon the Resolution
chosen while at Set Up.
ACTING:
/REVERSEDIRECT
HYSTERESE:
0.00NTU <>
BURN-OUT CONFIG:
STAND-BY: HOLD
CURRENT CONFIG.?
/ YESNO
OUTPUT CONFIG.:
mA
MA OUTPUT:
/ON OFF
VALUE FOR 4 mA
0.00NTU <>
VALUE FOR 20 mA
5.00NTU <>
From Page 17
A
For the Burn Out configuration, user will have three options
as: 4mA, 20mA or Hold.
<SEL>
Presskey until the desired option flashes, then
<ENT>
press to confirm.
User will be able to calibrate mA output.
Press key to choose the desired option then
<SELECT>
press key to confirm
<ENTER>
Connect Amp Meter to output 4-20mA and it will be possible
to adjust them, then press <ENTER> when ready.
Adjust the value as needed by pressing (to
decrease) or (to increase) keys, so the outputs
can be adjusted.
<SELECT>
will increase.
<ESCAPE>
<ESCAPE>
key will decrease the value and key
<SELECT>
BURN-OUT CONFIG:
STAND-BY:4mA
CALIBRATE mA?
/ YESNO
CONFIRM?
/ YESNO
PLACE AMP METER
@ OUTPUT 4-20 mA
READY?
ADJUST 4 mA
<SEL>- <ESC>+
ADJUST 20 mA
<SEL>- <ESC>+
Go to Page 19
18
Page 20
12.
Equipment Operation - Set Up (cont.)
User will have the option to configure digital output RS-485
for Proprietary or Modbus Protocol. Press key
to choose the desired option then press key to
confirm .
User can choose the protocol desired.
User can choose the Speed.
User can choose the Bits Number.
User can choose the Parity.
User will be able to define the instrument identification
number within the network, up to 32 instruments. Refer to
Page 13 for instruction on how to adjust this value.
<SELECT>
<ENTER>
RS-485 CONFIG.?
PROTOCOL
SPEED (100bps)
BITS NUMBER
COMMUNICATION-ID
NUMBER 1 <>
/ YESNO
/MODBUSPROPR.
24 48 9612
/87
PARITY
/EVENODD
WAIT
TURBIDIMETER
READ/SET UP
19
Page 21
12.
Equipment Operation - Calibration
Before initiating the Calibrtion Procedure, preapre the Standard Solutions and have the deionized water
ready. Also clean the Calibration Reservoir using deionized waterat every change of Standard Solution.
Press until flashes then press key.<SEL>Read<ENT>
TURBIDIMETER
/SET UPREAD
Press until flashes then press
<SEL>Calibrate<ENT>
key.
Follow intructions as per page 13, dip the Standard
Solution on the value shown, this Std Solution is chosen
<SEL>
during Set Up Mode. Press until desired opiton
<ENT>
flashes then press key. User will still have the
option to accept this calibration point or not!
<ENT>
Press key to confirm.
Place the sensor unit, as per instruciton on page 13, into
the Standard value shown. This Std Solution is chosen
during Set Up Mode. Press until desired opiton
flashes then press key.
<ENT>
Press key to confirm.
<ENT>
<SEL>
TURBIDIMETER
READ/CALIBRATE
CAL. 100.00 NTU?
/ YES NO
CONFIRM ?
/ YES NO
PLACE SENSOR
@ 100.00 NTU
READY ?
================
WAIT
The equipment will analyze the Standard Solution and
If it does not match the instrument linarity, it will inform
the user about its quality, but if the user decides to use
this solution, he will be able to do so!
Follow intructions as per page 13, dip the Standard
Solution on the value shown, this Std Solution is chosen
<SEL>
during Set Up Mode. Press until desired opiton
<ENT>
flashes then press key. User will still have the
option to accept this calibration point or not!
<ENT>
Press key to confirm.
Place the sensor unit, as per instruciton on page 13, into
the Standard value sown. Tthis Std Solution is chosen
during Set Up Mode. Press until desired opiton
flashes then press key.
<ENT>
Press key to confirm.
Remove the sensor from the bubble trapper, remove the
Calibration cup. Re-insert the sensor into the trapper,
open the process and proceed with Readings.
<ENT>
<SEL>
NON CONFORMANCE
STANDARD <ENTER>
CONFIRM STAND.?
/ YESNO
Standard in conformance
CAL.:200.00 NTU?
/ YESNO
CONFIRM ?
/ YESNO
PLACE SENSOR
@ 200.00 NTU
READY ?
================
WAIT
GO TO SAMPLE!
READY ?
================
->200.49 NTU
Indicates when the
Bargraph was not chosen
during Set Up Mode.
S1:» S2:»
-> 200.49 NTU
20
Page 22
12.
Equipment Operation - Read
Find below instructions on how to perform a Turbidity Reading. Before starting, verify if the bubble
trapper is clean, if the drain hose is connected and “open” and if the sample is flowing correctly.
Finally insert the sensor unit at the bubble trapper and lock it by turning it counter clockwise
(45degrres).
Press until flashes then press key.<SEL>Read<ENT>
TURBIDIMETER
/SET UPREAD
Press until flashes then press key.<SEL>Read<ENT>
TURBIDIMETER
READ/CALIBRATE
Verify if sample is flowing and drain hose is connected,
then press <ENT> key to start Reading Mode.
GO TO SAMPLE!
READY?
================
Presskey to place the equipment in Stand By.<ENT>
-> 200.49 NTU
STAND BY
Presskey to exit Stand By and return to Read Mode.<ENT>
================
-> 200.49 NTU
Presskey to verify the Contact and Current
<SEL>
conditions.
S1:» S2:»
-> 4.83 mA
<SEL>
Press and it will be able to adjust the value
Read, if chosen this procedure during Set Up Mode
(On Line Calibration-page 17). Refer to page 16
(Note A) for instructions on how to adjust this value.
ON LINE CALIBR.
200.49 NTU <>
================
-> 200.00 NTU
Indicates when the
Bargraph was not chosen
during Set Up Mode.
S1:» S2:»
-> 200.49 NTU
In order to exist the reading Mode, press and hold <ESC>
key. User will then access the main menu.
TURBIDIMETER
/SET UPREAD
21
Page 23
13.
Communication Protocol
Communication Protocol (TRCN441):
1) Proprietary:
Order:
ESC ID P CR LF
0x1B 0x50 0x0D 0x0A Hexadecimal
The ID is configured at the instrument from 1 to 32.
Answer:
L > > V V V V V V N T U C C C C C C m A
Example of answer for an un-stable value
L > 250 . 0 N T U 1 2 . 0 0 m A
Example of answer for a stable value
L > > 250 . 0 N T U 1 2 . 0 0 m A
Note: a) When the answer comes a C instead of a L, it means that the equipment is under
calibration function (is being operated in location by the user, executing the calibration
operation at the instrument).
b) When the answer comes an S instead of a L, it means that the equipment is under Set Up
function and it is being operated in location by the user.
2) RS485 – It is a “physical location”, where the proprietary protocol will be “transported”.
As factory default, this communication comes configured as:
Speed = 9600
Parity = none
Number of Bits = 8
Stop Bit = 1
ID = 1
22
Page 24
Page 25
M-4522/0908