Omega RH62C User Manual

Page 1
OMEGA
RH62C
HYGROMETER
Page 2
OMEGAnetSM On-Line Service
Internet e-mail
info@omega.com
For immediate technical or application assistance:
USA and Canada: Mexico and Latin America:
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
SM
Tel: (95) 800-TC-OMEGA
SM
Customer Service: 1-800-622-2378 / 1-800-622-BEST
SM
FAX: (95) 203-359-7807
Engineering Service: 1-800-872-9436 / 1-800-USA-WHENSMEn Español: (203) 359-7803
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
e-mail: espanol@omega.com
Servicing North America:
USA: ISO 9001 Certified Canada:
One Omega Drive, Box 4047 976 Bergar
Stamford, CT 06907-0047 Laval (Quebec) H7L5A1
Tel: (203) 359-1660 Tel: (514) 856-6928
FAX: (203)359-7700 FAX: (514) 856-6886
e-mail: info@omega.com e-mail: info@omega.com
Servicing Europe:
Benelux:
Postbus 8034, 1180 LA Amstelveen,
The Netherlands
Tel: (31) 20 6418405 FAX: (31) 20 6434643
Toll Free in Benelux: 06 0993344
e-mail: nl@omega.com
Czech Republic:
ul. Rude armady 1868, 733 01 Karvina-
Hranice, Czech Repubic
Tel: 420 (69) 6311627 FAX: 420 (69)
6311114
e-mail: czech@omega.com
France:
9, rue Denis Papin, 78190 Trappes
Tel: (33) 130-621-400 FAX: (33)130-699-120
Toll Free in France: 0800-4-06342
e-mail: france@omega.com
Germany/Austria:
Daimlerstrasse 26, D-75392
Deckenpfronn, Germany
Tel: 49 (07056) 3017 FAX: 49 (07056) 8540
Toll Free in Germany: 0800 82 66342
e-mail: germany@omega.com
United Kingdom: ISO 9002 Certified
One Omega Drive
Riverbend Technology Centre Northbank, Irlam,
Manchester, M44 5EX, England
Tel: 44 (161) 777-6611 FAX: 44 (161) 777-6622
Toll Free in England: 0800-488-488
e-mail: sales@omega.com.uk
omega.com
TM
Page 3
1
INTRODUCTION
This instrument is a portable 3½ digit, compact-sized digital Thermo-Hygrom­eter designed for simple one hand operation. Uses Platinum Resistance Tem­perature Detector Pt385/1000W (Alpha=0.00385) as temperature sensor, and uses thin film polymer capacitive type relative humidity sensor as hygrometer sensor.
SAFETY INFORMATION
It is recommended that you read the safety and operation instructions before using the hygrometer.
CAUTION
Do not immerse the hygrometer sensor head into liquids since this causes permanent damage to the sensor.
The meter when not in use, please use protective metal cap cover the sensor head and rotate clockwise it to extend sensors life.
The symbol on the instrument indicates that the operator must refer to an
explanation in this manual.
Page 4
2
SPECIFICATIONS
GENERAL
Display: 3½ digit liquid crystal display (LCD) with maximum reading of 1999. Overtemperature: "OL" mark indication. Low battery indication: The " " is displayed when the battery voltage drops
below the operating level.
Measurement rate: 2.5 times per second, nominal. Accuracy: Stated accuracy at 23°C ± 5°C, <75% relative humidity. Temperature Coefficient: 0.1 times the applicable accuracy specification per
°C from °C to 18°C and 28°C to 50°C.
Operating environment: 0°C to 50°C at <75% relative humidity. Storage environment: -20°C to 60°C at <80% relative humidity. Battery: 4 pcs 1.5V (AAA size). Battery Life: 200 hours typical. Dimensions: 170mm(H) x 44mm(W) x 40mm(D). Weight: 160g (including probe and batteries).
Page 5
3
ELECTRICAL TEMPERATURE
Temperature Scale: Celsius. Temperature Sensor: RTD Pt385/1000W. Measurement Range: -20°C to 100°C. Resolution: 0.1°C. Accuracy: ±0.5°C 0°C to 50°C
±1°C -20°C to 0°C, 50°C to 100°C
RELATIVE HUMIDITY
Humidity Sensor: Electronic capacitance polymer film sensor. (The sensor is
unaffected by water condensate, is immune to most reagent vapors)
Measurement Range: 0% to 100%RH. Resolution: 0.1%RH. Accuracy: ±2.5%RH (10% to 90%RH)
±5%RH (<10%, > 90%RH)
Sensor Response Time for 90% of Total Range: 60sec typical. Sensor Stability: ±2%RH, 2 years typical.
Page 6
4
Sensor Hysteresis (excursion of 10% to 90% to 10% RH): ±1%RH typical. Sensor Temperature Dependence: Negligible between 0°C to 50°C.
OPERATING INSTRUCTIONS
Push Button
Display Back-Light Button
Press " " button to toggle between turn on and turn off the Back-Light.
DATA HOLD Button
Press the DATA HOLD key to enter the DATA HOLD mode, the" H " annunciator is displayed. When DATA HOLD mode is selected, the hygrometer holds the present readings and stops all further measurements. Press the DATA HOLD key again cancels DATA HOLD mode, causing hygrom­eter to resume taking measurements.
Page 7
5
MAX HOLD Button
Press the MAX HOLD key to enter the MAX HOLD mode. The hygrometer then records and updates the maximum absolute values and the MAX annunciator appears on the display. Press the MAX HOLD key again to exit the MAX HOLD recording mode. In the MAX HOLD mode, press DATA HOLD key to stop the recording, press DATA HOLD key again to resume recording.
OPERATION
1. Rotate counter clockwise to remove the protective metal cap.
2. Set the power switch to the desired %RH or °C range.
3. Read the display.
4. Cover sensor head to extend sensor life when not in use.
Page 8
6
SPECIAL CONSIDERATIONS
Before a reliable measurement can be made, the measuring hygrometer and medium to be measured must be in temperature and humidity eguilibrium.
Temperature measurement errors Due to too short measurement time, sunshine during the measurement, heating, cold outer walls, air draft (e.g. fans), radiating hand and / or body heat etc.
Humidity measurement errors Due to steam, water splashes, dripping water or condensation (not water condensate) on the sensor etc. However, repeatability and long-term stability are not impaired by this.
Contamination By dust in the air or measurements in powdery substances. This can be largely avoided by using a corresponding filter. The filter must be cleaned or replaced periodically depending upon the degree of contamination of the measuring site.
Page 9
OPERATOR MAINTENANCE
Battery Replacement
Power is supplied by four 1.5V (AAA size) batteries. The " " appears on the LCD display when replacement is needed. To replace the batteries, remove the screw from the back of the meter and lift off the battery cover. Remove the batteries from battery contacts.
Cleaning
Periodically wipe the case with a damp cloth and detergent, do not use abrasives or solvents.
7
Page 10
WARRANTY
OMEGA warrants this unit to be free of defects in materials and workmanship and to give satisfactory service for a period of 13 months from date of purchase. OMEGA W arr anty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA's customers receive maximum coverage on each product. If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA's Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon e xamination by OMEGA, if the unit is f ound to be def ective it will be repaired or replaced at no charge. Howev er, this WARRANTY is VOID if the unit shows e vidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA's control. Components which wear or which are damaged by misuse are not warranted. This includes contact points, fuses, and triacs.
OMEGA is glad to offer suggestions on the of use of its various products. Nevertheless, OMEGA only warrants that the parts manufactured by it will be as specified and free of defects
OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are e xclusive and the total liability of OMEGA with respect to this order, whether based on contract warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
Every precaution for accuracy has been taken in the preparation of this manual; however , OMEGA ENGINEERING, INC. neither assumes responsibility for any omissions or errors that ma y appear nor assumes liability for any damages that result from the use of the products in accordance with the information contained in the manual.
SPECIAL CONDITION: Should this equipment be used in or with any nuclear installation or activity, purchaser will indemnity OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the equipment in such a manner.
Page 11
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected application.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) T O OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA'S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges , freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please hav e the following
information available BEFORE contacting OMEGA:
1. P.O. number under which the product was
PURCHASED.
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems relative
to the product.
FOR NON-WARRANTY REPAIRS, consult OMEGA for
current repair charges. Have the follo wing information available BEFORE contacting OMEGA:
1. P.O. number to cover the COST of the repair.
2. Model and serial number of product , and
3. Repair instructions and/or specific problems relative to the product.
OMEGA's policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 1999 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readab le from, in whole or in part, without prior wr itten consent of OMEGA ENGINEERING, INC.
Page 12
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA...Of Course!
HEATERS
þ Heating Cable
þ Cartridge & Strip Heaters
þ Immersion & Band Heaters
þ Flexible Heaters
þ Laboratory Heaters
PRESSURE/STRAIN AND FORCE
þ Transducers & Strain Gauges
þ Load Cells & Pressure Gauges
þ Displacement Transducers
þ Instrumentation & Accessories
FLOW/LEVEL
þ Rotameters, Gas Mass
Flowmeters & Flow Computers
þ Air Velocity Indicators
þ Turbine/Paddlewheel Systems
þ Totalizers & Batch Controllers
TEMPERATURE
þ
Thermocouple, RTD & Thermistor
Probes, Connectors, Panels &
Assemblies
þ Wire: Thermocouple, RTD &
Thermistor
þ
Calibrators & Ice Point References
þ
Recorders, Controllers & Process
Monitors
þ Infrared Pyrometers
ENVIRONMENTAL MONITORING AND CONTROL
þ
Metering & Control Instrumentation
þ Refractometers
þ Pumps & Tubing
þ Air, Soil & Water Monitors
þ Industrial Water & Wastewater
Treatment
þ pH, Conductivity & Dissolved
Oxygen Instruments
pH/CONDUCTIVITY
þ pH Electrodes, Testers &
Accessories
þ Benchtop/Laboratory Meters
þ Controllers, Calibrators,
Simulators & Pumps
þ Industrial pH & Conductivity
Equipment
DATA ACQUISITION
þ Data Acquisition & Engineering
Software
þ Communications-Based
Acquisition Systems
þ Plug-in Cards for Apple, IBM
& Compatibles
þ Datalogging Systems
þ Recorders, Printers & Plotters
M-2860/0799
Loading...