United Kingdom:One Omega Drive, River Bend Technology Centre
ISO 9002 CertifiedNorthbank, Irlam, Manchester
M44 5BD United Kingdom
TEL: +44 (0)161 777 6611FAX: +44 (0)161 777 6622
Toll Free in United Kingdom: 0800-488-488
e-mail: sales@omega.co.uk
onn, Germany
AX: +49 (0)7056 9398-29
F
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI
regulations that apply. OMEGA is constantly pursuing certification of its products to the European New
Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any
errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, human applications.
Foreword
Thank you for purchasing the OMEGA RD100B/RD1800B Recorder.
This user’s manual describes the functions of the Ethernet interface and the RS-422A/
485 communication interface. To ensure correct use, please read this manual
thoroughly before beginning operation.
After reading the manual, keep it in a convenient location for quick reference whenever a
question arises during operation.
The following five manuals, including this one, are provided as manuals for the RD100B/
RD1800B Recorder. Please read all of them.
The figures used in this manual are mostly of the RD100B. If you are using the
RD1800B, refer to the figures for reference.
• Paper Manual
Manual TitleManual No.Description
RD100B RecorderM-4232Explains the basic operations of the RD100B
Operation Guiderecorder.
RD1800B RecorderM-4243Explains the basic operations of the RD1800B
Operation Guiderecorder.
• Electronic Manuals Provided on the Accompanying CD-ROM
Manual TitleManual No.Description
RD100B RecorderM-4231Explains all the functions and procedures of the
User’s ManualRD100B recorder excluding the communication
functions.
RD1800B RecorderM-4242Explains all the functions and procedures of the
User’s ManualRD1800B recorder excluding the
communicationfunctions.
RD100B/RD1800BM-4233This manual. Explains the functions of the
Communication InterfaceEthernet interface and the RS-422A/485
User’s Manualcommunication interface.
Notes
Revisions
• The contents of this manual are subject to change without prior notice as a result of
continuing improvements to the instrument’s performance and functions. The figures
given in this manual may differ from those that actually appear on your screen.
• Every effort has been made in the preparation of this manual to ensure the accuracy
of its contents. However, should you have any questions or find any errors, please
contact OMEGA.
• Copying or reproducing all or any part of the contents of this manual without
permission is strictly prohibited.
• The TCP/IP software of this product and the document concerning the TCP/IP
software have been developed/created based on the BSD Networking Software,
Release 1 that has been licensed from the University of California.
• 1st EditionDecember 2004
• 2nd EditionMarch 2005
• 3rd EditionAugust 2005
i
Trademarks
• Microsoft, MS-DOS, Windows, Windows NT, and Windows XP are either registered
trademarks or trademarks of Microsoft Corporation in the United States and/or other
countries.
• Adobe, Acrobat, and PostScript are trademarks of Adobe Systems incorporated.
• For purposes of this manual, the TM and ® symbols do not accompany their
respective trademark names or registered trademark names.
• Other company and product names are trademarks or registered trademarks of their
respective holders.
Functional Enhancement of the RD100B Recorder
Functions have been added or modified on the RD100B Recorder since system version
1.11. You can check the system version on the system display. For details, see the
RD100B User's Manual (M-4231).
Suffix Code Added or Modified Functions
-
-
-
/C3
/C7
(Changed) Expanded the selectable range of alarm values during linear
scaling (including 1-5V and SQRT) to −5% to 105% of the scale.
(Changed) The procedure to set the start/end date and time of Daylight
Saving Time (DST) has been changed. The TD command
can be used on the µR20000 and the RD100B with system
version 1.11. The SS command can be used on the
RD100B with system version 1.02 or earlier.
(Added) The print/display format of the date can be changed.
(Changed) Modbus slave protocol can be used.
Two-wire sstem.
(Changed) Users with the same user name cannot be registered.
Reference
Section 4.4: SA
Section 4.4: TD
Section 4.5:
Section 4.5:
Section 3.2
Section 2.5
XN
YS
command
command
command
command
ii
How to Use This Manual
Structure of the Manual
This user’s manual consists of the following sections.
Chapter 1 Overview of the Communication Functions
Gives an overview of the communication functions.
Chapter 2 Using the Ethernet Interface (/C7 Option)
Explains the specifications of the Ethernet interface and how to use the interface.
Chapter 3 Using the RS-422A/485 Communication Interface (/C3 Option)
Explains the specifications of the RS-422A/485 communication interface and how to use the
interface.
Chapter 4 Commands
Explains each command that is available.
Chapter 5 Responses
Explains the responses that the recorder returns and the output format of the setup data and
measured/computed data.
Chapter 6 Status Information
Explains the registers that indicate the recorder statuses.
Appendix
Provides an ASCII character code table, flow charts for outputting data from the recorder,
login procedure, and a list of error messages.
Index
Index of contents.
iii
How to Use This Manual
Conventions Used in This Manual
• Unit
• k: Denotes 1000. Example: 5 kg, 100 kHz
• K: Denotes 1024. Example: 640 KB
• Note
The following markings are used in this manual.
Improper handling or use can lead to injury to the user or damage
to the instrument.
indicate that the user must refer to the user’s manual for special
instructions. The same symbol appears in the corresponding place
in the user’s manual to identify those instructions. In the manual,
the symbol is used in conjunction with the word “WARNING” or
“CAUTION.”
This symbol appears on the instrument to
WARNING
CAUTION
Note
• Bold Characters
Bold characters are mainly characters and numbers that appear on the display.
• Subheadings
On pages that describe the operating procedures in Chapter 2 and 3, the following
symbols are used to distinguish the procedures from their explanations.
Explanation
Procedure
Calls attention to actions or conditions that could cause serious or
fatal injury to the user, and precautions that can be taken to prevent
such occurrences.
Calls attentions to actions or conditions that could cause light injury
to the user or damage to the instrument or user’s data, and
precautions that can be taken to prevent such occurrences.
Calls attention to information that is important for proper operation
of the instrument.
This subsection describes the setup parameters and the limitations
on the procedures.
Follow the numbered steps. All procedures are written with
inexperienced users in mind; depending on the operation, not all
steps need to be taken.
iv
Names of Parts and Basic Key Operations
Display and Keys
You use the panel keys and the display to configure the communication functions. For a
description of other parts of the recorder, see section 3.1 in the
(The figure below is of the RD100B Recorder.)
FrontRear Panel
Optional terminal block
Display
Keys
Power
switch
Indicator
Ethernet port
Keys
<While setting functions, when the FUNC key/DISP MENU key is pressed>
CHARACTER Key: Changes the character type when entering a character. Press
this key while holding down the SHIFT key to switch the character type in reverse
order.
UP/DOWN Key: Switches the setup item or the value.
Press this key while holding down the SHIFT key to switch the setup item
or the value in reverse order.
LEFT/RIGHT Key: Moves the cursor to the right when entering a value
or character. Press this key while holding down the SHIFT key to move
the cursor to the left..
ESC Key: Cancels the operation.
When pressed with the SHIFT key, the display of the comment
on the setting turns ON/OFF.
SHIFT Key: Used with the key, key, or the
CHARACTER key.
ENTER Key: Confirms the setup item or value.
Recorder User’s Manual
RS-422A/485
communication
terminals
.
RCD
RCD key
Starts/stops recording.
CHARACTERESC/?SHIFT
MENUFEED
DISP
DISP key
Switches the screen in the main display.
MENU key
Hold this key down for 3 seconds to enter Setting mode. Hold this key down for
3 seconds also to exit from Setting mode.
1
FUNC
MENU
DISP MENU key
Hold this key down for 3 seconds to switch to the data display
setup screen. Hold this key down for 3 seconds also to exit
from the data display setup screen.
FUNC key
Used when executing manual printout, message printout, etc.
CH UP
CH UP key
Switches the displayed channel.
(when manual switching is specified)
FEED key
Feeds the chart paper.
<During normal operation>
v
Names of Parts and Basic Key Operations
Basic Key Operations
This section describes basic operations on the front panel keys to change various settings.
• Execution Modes
The recorder has the following execution modes.
• Operation mode: A mode used to perform recording and monitoring.
•
Setting
mode
parameters.
• Basic Setting mode: A mode used to set the basic specifications of functions with
the recording operation stopped.
* In the explanation of commands in (chapter 4), Run mode collectively refers to
Operation mode and Setting mode.
Settings related to communications are configured in Basic Setting mode. You
cannot enter Basic Setting mode while the recorder is recording or while
computation is in progress on the computation function (/M1 option).
• Entering Basic Setting Mode
Hold down the
The Setting mode display appears.
Set=Range
Input range and s
The panel keys are set to the functions marked above the keys as shown below.
CHARACTERESC/?SHIFT
RCD
MENUFEED
: A mode used to set the input range, alarms, chart speed, and other
MENU
key for 3 seconds.
DISP
FUNC
1
MENU
CH UP
Hold down both the (
DISP
) key and the (
FUNC
) key for 3 seconds.
The Basic Setting mode display appears. The top and bottom lines are the setup item
and comment, respectively. The section that is blinking in the setup item that you
change. In this manual, the section that you change appears shaded.
The comment line shows useful information such as a description of the setup item
and the range of selectable values. Read the comment and change the items as
necessary.
Setup item
Comment
Basic=Alarm
Auxiliary alarm
The item to be controlled blinks.
• Selecting the Setup Item and Value
DISP
(
The selected item change each time you press the
item change in reverse order if you press the
SHIFT (
Basic=Ethernet
FEED
) key.
Selections
(
DISP
) key. The selected
) while holding down the
This manual denotes the operation of pressing a key while holding down the SHIFT
FEED
(
) key as SHIFT + the other key (for example: SHIFT + key).
After you make a selection, press the
CH UP
(
) key. The next screen appears.
When the Setting Complete screen appears, the changed item is applied.
Ethernet host
Setting complete
• Using the ESC Key
If you press the ESC (
a higher level menu. If you do not show the Setting Complete screen, the changes
you made up to that point are discarded.
You can show and hide the comment on the bottom line by pressing the ESC (
key while holding down the SHIFT (
vi
1
MENU
) key, the operation is cancelled, and the display returns to
FEED
) key.
1
MENU
)
Names of Parts and Basic Key Operations
• Entering Values
Use the
SHIFT +
A=000.000.000.000
A=000.000.000.000
key or SHIFT + key to move the cursor. Use the key or
key to change a digit value. You repeat these steps to enter the value.
SHIFT
key + key
key
Target digit
key
SHIFT
key + key
When you press the key, the change is applied and the next setup item is
displayed.
• Entering Characters
Use the
key or SHIFT + key to move the cursor.
Use the CHARACTER key or SHIFT + CHARACTER key to select the character type.
Use the
key or SHIFT + key to select a character. You repeat these steps
to set the character string.
Cursor
Host=ABCDefgh2004
CHR:A-Z
Select the character type
CHARACTER
SHIFT
key +
key or
CHARACTER
key
The character type changes in the following order: uppercase alphabet, lowercase
alphabet, numbers, and symbols.
key
Host=ABCDefgh2004
CHR:A-Z
SHIFT
key + key
When you press the key, the change is applied and the next screen is displayed.
Inserting Characters
Press the CHARACTER key or SHIFT + CHARACTER key to show Ins DISP and
then press the
key. A space for one character is inserted. Enter the character.
Deleting a Character
Press the CHARACTER key or SHIFT + CHARACTER key to show Del DISP and
then press the
key. The character is deleted.
Deleting an Entire Character String
Press the CHARACTER key or SHIFT + CHARACTER key to show Clear DISP and
then press the
key. The entire character string is deleted.
Copying & Pasting a Character String
Show the copy source character string.
Press the CHARACTER key or SHIFT + CHARACTER key to show Copy DISP and
then press the
key. The character string is saved to the memory.
Show the copy destination.
Press the CHARACTER key or SHIFT + CHARACTER key to show Paste DISP and
then press the
key. The character string is pasted.
vii
Names of Parts and Basic Key Operations
• Exiting from Basic Setting Mode
Press the ESC key several times to return to the Basic= screen.
Press the
The setup save screen appears.
Basic=End
Save Setting
Press the key to select Store and then press the key.
The setting is applied and the Operation mode screen appears.
If you select Abort and press the
mode screen appears.
End=Store
Save settings and
key or SHIFT + key to select End and then press the key.
key, the setting is discarded and the Operation
viii
Contents
1
Foreword ......................................................................................................................................... i
Functional Enhancement of the RD100B Recorder ........................................................................ ii
How to Use This Manual ................................................................................................................ iii
Names of Parts and Basic Key Operations .................................................................................... v
Chapter 1 Overview of the Communication Functions
1.1Communication Functions Using the Ethernet Interface (/C7 Option) ............................. 1-1
Functional Construction ................................................................................................... 1-1
Setting/Measurement Server ........................................................................................... 1-1
Maintenance/Test Server ................................................................................................. 1-2
Instrument Information Server ......................................................................................... 1-2
Other Functions ............................................................................................................... 1-3
1.2Communication Functions Using the RS-422A/485 Communication Interface
System Errors .............................................................................................................. App-8
1
2
3
4
5
6
App
Index
xi
Chapter 1 Overview of the Communication Functions
1.1Communication Functions Using the Ethernet
Interface (/C7 Option)
The recorder can be equipped with an optional Ethernet interface. For details on how to
use the Ethernet interface, see chapter 2.
Functional Construction
The following figure shows the relationship between the communication function of the
recorder and the Ethernet interface. Perform communication according to the respective
protocol.
*
Protocol
is a set of rules that two computers use to communicate via a communication line (or
network).
Communication functions of the recorder
Maintenance/
Test Server
TCP
Instrument
Information Server
UDP
IP
Application
Upper layer
protocol
Lower layer
protocol
Interface
Setting/
Measurement Server
Login (user authentication/access
privileges granting)
Dedicated protocol
Ethernet (10BASE-T)
1
Overview of the Communication Functions
TCP (Transmission Control Protocol)
UDP (User Datagram Protocol)
IP (Internet Protocol)
Setting/Measurement Server
• You can specify settings that are approximately equivalent to those specified by front
panel key operations. However, you cannot turn the power ON/OFF, set the user
name and password for communications, nor set the key lock.
• The data below can be output.
Data TypeOutput Format
Measured/computed dataBINARY/ASCII
Setup dataASCII
Periodic printout and the most recent TLOG computation dataASCII
Status informationASCII
Information on connected usersASCII
• The commands that can be used are Setting commands, Basic Setting commands,
Control commands, and Output commands.
Connect the recorder and the
PC using an Ethernet cable.
PC
<Related Topics>
• Ethernet interface settings: Section 2.3
• Commands: Section 4.2
• Data output format: Chapter 5
1-1
1.1 Communication Functions Using the Ethernet Interface (/C7 Option)
Maintenance/Test Server
• Outputs Ethernet communication information such as connection information and
network statistics from the recorder.
• The commands that can be used Maintenance/Test commands.
<Related Topics>
• Ethernet interface settings: Section 2.3
• Commands: Section 4.2
Instrument Information Server
• Outputs the serial number, model name, and other information about the recorder
connected via the Ethernet network.
• The commands that can be used Instrument Information Output commands.
<Related Topics>
• Ethernet interface settings: Section 2.3
• Commands: Section 4.2
1-2
1.1 Communication Functions Using the Ethernet Interface (/C7 Option)
Other Functions
• Login Function
Only users that are registered in advance can access the Setting/Measurement and
Maintenance/Test servers.
• Users are identified by their user name and password.
• You can register one administrator and six users.
Administrator privileges
The administrator can use all the functions on the Setting/Measurement and
Maintenance/Test servers
User Privileges
• Setting/Measurement server
Users can output measured data, setup data, scheduled printing, and the most
recent TLOG computation data. Users cannot control the recorder.
• Maintenance/Test server
Users cannot disconnect communications between the recorder and other PCs.
All other operations are allowed.
• There is a maximum number of simultaneous connections that can be established
with the recorder.
<Related Topics>
• Login function settings: Section 2.5
• Maximum number of simultaneous connections: Section 2.1
• Commands available to the administrator and users: Section 4.2
1
Overview of the Communication Functions
• Communication Timeout
This function drops the connection with the PC if there is no data transmission for a
given time at the application level (see “Functional Construction”). For example, this
function prevents a PC from being connected to the recorder indefinitely which would
prohibit other users from making new connections for data transfer.
<Related Topics>
• Communication timeout setting: Section 2.6
• Keepalive
This function drops the connection if there is no response to the inspection packet that
is periodically transmitted at the TCP level.
<Related Topics>
• Keepalive setting: Section 2.6
1-3
1.2Communication Functions Using the RS-422A/
485 Communication Interface (/C3 Option)
The recorder can be equipped with an optional RS-422A/485 communication interface.
For details on how to use the RS-422A/485 communication interface, see chapter 3.
Functional Construction
The following figure shows the relationship between the communication function of the
recorder and the RS-422A/485 communication interface. Perform communication
according to the respective protocol.
*
Protocol
is a set of rules that two computers use to communicate via a communication line (or
network).
Communication functions of the recorder
Application
Interface
Setting/Measurement Server
Dedicated protocolProtocol
RS-422A/485
Connect the recorder and the
PC using a serial cable.
PC
Modbus Slave
Modbus protocol
Setting/Measurement Server
The functions are the same as those of the Setting/Measurement server of the Ethernet
interface. See page 1-1.
<Related Topics>
• RS-422A/485 communication interface settings: Section 3.5
• Commands: Section 4.2
• RS-422A/485 dedicated commands: Section 4.8
• Data output format: Chapter 5
Modbus Slave
• The Modbus protocol can be used to read the measured/computed data on your PC
by reading the input registers of the recorder. The communication input data can be
written or read by writing/reading the hold register of the recorder.
• For details on the Modbus function codes that the recorder supports, see section 3.4.
• This function can be used only when communicating via the serial interface (option).
• For a description on the settings required in using this function, see section 3.5.
Data to slave device
Modbus master device
Data from slave device
Serial communication
Recorder
Modbus slave device
1-4
Chapter 2 Using the Ethernet Interface (/C7 Option)
2.1Ethernet Interface Specifications
Basic Specifications
ItemSpecifications
Electrical and mechanical specificationsConforms to IEEE 802.3
(Ethernet frames are of DIX specification)
Transmission medium type10BASE-T
ProtocolTCP, IP, UDP, ICMP, and ARP
The Maximum Number of Simultaneous Connections and the Number of Simultaneous
Use
The following table shows the maximum number of simultaneous connections, the
number of simultaneous users, and the port numbers of the recorder.
FunctionMaximum Number Number of Simultaneous Users Port Number
Setting/Measurement 312
server
Maintenance/Test111
server
Instrument---34264/udp
Information server
†
The port numbers are fixed.
††
For details on administrator and user privileges, see “Login Function” in section 1.1.
of Connections<Administrators><Users>
††
††
34260/tcp
34261/tcp
2
Using the Ethernet Interface (/C7 Option)
†
2-1
2.2Connecting the Ethernet Interface
When Connecting Only the Recorder and a PC
Connect the recorder and the PC via a HUB as in the following figure.
RD100B/RD1800B Recorder
10BASE-T
straight cable
RD100B/RD1800B Recorder
When Connecting to a Preexisting Network
The following figure illustrates an example in which a recorder and a PC are connected
to the network. When connecting the recorder or the PC to a preexisting network, the
transfer rate, connector type, etc. must be matched. For details, consult your system or
network administrator.
RD100B/RD1800B Recorder
Hub
10BASE-T
straight cable
Hub, router, etc.
PC
Network
PC
Note
• Depending on the reliability of the network or the volume of network traffic, all the transferred
data may not be retrieved by the PC.
• Communication performance deteriorates if multiple PCs access the recorder simultaneously.
2-2
10BASE-T
straight cable
2.3Configuring the Ethernet Interface
Procedure
Set the host name and IP address of the recorder. You do not have to set the DNS
(domain name system).
Setup Items
Ethernet
End
Host
Local IPAM
DNS
HostDomain
Host nameDomain name
IP addressSubnet mask
DNS
DNS On/OffServer
Suffix_P
Domain suffix
(primary)
P
(primary)
Suffix_S
Domain suffix
(secondary)
G
Default gateway
S
Server
(secondary)
For a description of the basic setup operations, see “Basic Key Operations” on page v.
Entering Basic Setting Mode
Hold down the
Next, hold down both the
MENU
key for 3 seconds to display the Setting mode screen.
DISP
(
) key and the (
FUNC
) key for 3 seconds to
display the Basic Setting mode screen.
Note
To cancel an operation, press the ESC key.
Host Name and Domain Name
1.Press the
Basic=Ethernet
key to select Ethernet and then press the key.
2
Using the Ethernet Interface (/C7 Option)
2.Press the key to select Host and then press the key.
Ethernet=Host
3.Set the host name of the recorder and then press the key.
Key operations
• Use the
key to select the digit for entering a character.
• Use the CHARACTER key to select the character type.
• Use the
Host=
key to select the character you wish to enter.
4.Set the domain name and press the key in the same fashion as in step 3.
The setting complete screen appears.
Domain=
Ethernet host
Setting complete
5.Press the ESC key to return to the higher level menu. To save the settings and
exit from Basic Setting mode, proceed to “Saving the Settings.”
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
CHARACTER
2-3
2.3 Configuring the Ethernet Interface
IP Address, Subnet Mask, and Default Gateway
1.Press the
Basic=Ethernet
2.Press the key to select Local IP and then press the key.
Ethernet=Local IP
3.Set the IP address of the recorder and then press the key.
Key operations
• Use the
• Use the
A=
4.Set the IP address of the subnet mask and then press the key in the same
fashion as in step 3.
M=
5.Set the IP address of the default gateway and then press the key in the
same fashion as in step 3.
The local IP setting complete screen appears.
G=
key to select Ethernet and then press the key.
key to select the digit for entering a value.
key to select the value you wish to enter.
Ethernet address
Setting complete
6.Press the ESC key to return to the higher level menu. To save the settings and
exit from Basic Setting mode, proceed to “Saving the Settings.”
DNS (Domain Name System)
1.Press the
Basic=Ethernet
key to select Ethernet and then press the key.
2.Press the key to select DNS and then press the key.
Ethernet=DNS
3.Press the key to select On and then press the key.
DNS=On
4.Set the IP address of the primary DNS server and then press the key.
Key operations
• Use the
• Use the
P=
key to select the digit for entering a value.
key to select the value you wish to enter.
5.Set the IP address of the secondary DNS server and then press the key in
the same fashion as in step 4.
S=
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
2-4
CHARACTER
2.3 Configuring the Ethernet Interface
6.Set the primary domain suffix and then press the key.
Key operations
• Use the
key to select the digit for entering a character.
• Use the CHARACTER key to select the character type.
• Use the
Suffix_P=
key to select the character you wish to enter.
7.Set the secondary domain suffix and then press the key in the same fashion
as in step 6.
The DNS setting complete screen appears.
Suffix_S=
Ethernet DNS
Setting complete
8.Press the ESC key to return to the higher level menu. To save the settings and
exit from Basic Setting mode, proceed to “Saving the Settings.”
Saving the Settings
1.Press the ESC key to return to the Basic= screen.
2.Press the
3.Press the
key to select End and then press the key.
key to select Store and then press the key.
The settings are activated, and the Operation mode screen appears.
2
Using the Ethernet Interface (/C7 Option)
Explanation
For details on the settings, consult your system or network administrator.
• Host Name
Set the recorder’s host name and the domain name of the network to which the
recorder belongs. Be sure to set these items when using the DNS.
• Host
Set the recorder’s host name using up to 64 alphanumeric characters.
• Domain
Set the network domain name to which the recorder belongs using up to 64
alphanumeric characters.
• IP Address, Subnet Mask, and Default Gateway
• IP address
• Set the IP address to assign to the recorder. The default value is 0.0.0.0.
• The IP address is used to distinguish between the various devices connected to
the Internet when communicating using the TCP/IP protocol. The address is a
32-bit value normally expressed with four values (0 to 255), each separated by a
period as in 192.168.111.24.
• M (Subnet Mask)
• Specify the mask that is used to determine the network address from the IP
address. The default value is 0.0.0.0.
• Set this value according to the system or the network to which the recorder
belongs. In some cases, this setting may not be necessary.
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
CHARACTER
2-5
2.3 Configuring the Ethernet Interface
• G (Default Gateway)
• Set the IP address of the gateway (router, etc.) used to communicate with other
networks. The default value is 0.0.0.0.
• Set this value according to the system or the network to which the recorder
belongs. In some cases, this setting may not be necessary.
• Setting the DNS (Domain Name System)
The DNS is a system that correlates the host name/domain name to the IP address.
The host name/domain name can be used instead of the IP address when accessing
the network. The DNS server manages the database that contains the host name/
domain name and IP address correlation.
• On/Off
Select On when using the DNS.
• P (Primary DNS Server)
Set the IP address of the primary DNS server. The default value is 0.0.0.0.
• S (Secondary DNS Server)
Set the IP address of the secondary DNS server. The default value is 0.0.0.0. If
the primary DNS server is down, the secondary server is used to search the host
name and IP address correlation.
When the recorder searches another server using the DNS server, the domain
name of the recorder is appended to the host name as a possible domain name if it
is omitted. If the IP address corresponding to the server name is not found on the
DNS server, then it may be that the system is configured to use another domain
name for searching. This alternate domain name is specified as the domain suffix.
• Set the domain suffix using up to 64 alphanumeric characters.
• Up to two domain suffixes can be specified (primary and secondary).
• Saving the Settings
To activate the settings made in the Basic Setting mode, the settings must be saved.
2-6
2.4Checking the Connection Status
The connection status of the Ethernet interface can be confirmed with the indicator that
is located to the left of the Ethernet port on the recorder.
IndicatorConnection Status of the Ethernet Interface
ON (Green)The Ethernet interface is electrically connected.
Blinking (Green)Transmitting data.
OFFThe Ethernet interface is not electrically connected.
Optional terminal block
Indicator
Ethernet port
2
Using the Ethernet Interface (/C7 Option)
2-7
2.5Registering Users
Users that can access the recorder via the Ethernet network must be registered. This
function is called login function.
Setup Items
Ethernet
Login
LoginSet
Login
Use/Not
Level
Admin/UserOn/Off
Register
User
User name
Procedure
End
Password
For a description of the basic operations, see “Basic Key Operations” on page v.
Entering Basic Setting Mode
Hold down the
Next, hold down both the
MENU
key for 3 seconds to display the Setting mode screen.
DISP
(
) key and the (
FUNC
) key for 3 seconds to
display the Basic Setting mode screen.
Note
To cancel an operation, press the ESC key.
Enabling/Disabling the Login Function
1.Press the
Basic=Ethernet
2.Press the key to select Login and then press the key.
Ethernet=Login
3.Press the key to select Use and then press the key.
The setting complete screen appears.
Login=On
key to select Ethernet and then press the key.
4Press the ESC key to return to the Ethernet menu.
Registering Users
5.Press the
6.Press the key to select Admin (administrator) or User1 to User6, and then
7.Press the key to select On and then press the key.
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
2-8
key to select LoginSet and then press the key.
Ethernet=LoginSet
press the
Level=Admin
Register=On
key.
CHARACTER
2.5 Registering Users
8.Set the user name and then press the key.
Key operations
• Use the
key to select the digit for entering a character.
• Use the CHARACTER key to select the character type.
• Use the
User=
key to select the character you wish to enter.
9.Set the password and then press the key in the same fashion as in step 8.
The setting complete screen appears.
Password=
Ethernet login
Setting complete
To register other users, press the key to return to step 6 and repeat steps 8,
8, and 9.
Saving the Settings
1.Press the ESC key to return to the Basic= screen.
2.Press the
3.Press the
key to select End and then press the key.
key to select Store and then press the key.
The settings are activated, and the Operation mode screen appears.
2
Using the Ethernet Interface (/C7 Option)
Explanation
You can limit the users that can access the Setting/Measurement and Maintenance/Test
servers on the recorder via the Ethernet interface.
• Enabling/Disabling the Login Function
Set whether to use the login function.
• Registering Users
• User level
Select either of the user levels, administrator or user.
• Administrator
One administrator can be registered. An administrator has the authority to use all
Setting/Measurement server and Maintenance/Test server commands.
• User
Six users can be registered. A user has limited authority to use the commands.
See section 4.2.
• Selecting Whether to Register (On/Off) the User
If On is selected, set the user name and password.
• Setting the User Name
• Set the user name using up to 16 alphanumeric characters.
• The same user name can not be registered.
• Since the word “quit” is reserved as a command on the recorder, the user name
“quit” is not allowed.
• Setting the Password
Set the password using up to 4 alphanumeric characters and spaces.
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
CHARACTER
2-9
2.5 Registering Users
Note
• The relationship between the login function and the user name that is used when accessing
the recorder is as follows:
• When the login function is set to “Use”
• The registered user name and password can be used to login to the recorder.
• The user level is the level that was specified when the user name was registered.
• When the login function is set to “Not”
• The user name “admin” can be used to login to the recorder as an administrator.
Password is not necessary.
• The user name “user” can be used to access the recorder as a user. Password is not
necessary.
• There are limitations on the number of simultaneous connections or simultaneous uses of the
recorder from the PC (see section 2.1).
• For a description of the login process of the Setting/Measurement server and Maintenance/
Test server, see appendix 3.
• Saving the Settings
To activate the settings made in the Basic Setting mode, the settings must be saved.
2-10
2.6Setting the Communication Timeout and
Keepalive
Procedure
The communication timeout function and the keepalive function can be configured.
Setup Items
Ethernet
End
K.Alive
Timeout
On/Off
Keep alive
On/Off
DurationTimeout
Timeout time
For a description of the basic operations, see “Basic Key Operations” on page v.
Entering Basic Setting Mode
Hold down the
Next, hold down both the
MENU
key for 3 seconds to display the Setting mode screen.
DISP
(
) key and the (
FUNC
) key for 3 seconds to
display the Basic Setting mode screen.
Note
To cancel an operation, press the ESC key.
Communication Timeout
1.Press the
Basic=Ethernet
key to select Ethernet and then press the key.
2
Using the Ethernet Interface (/C7 Option)
2.Press the key to select Timeout and then press the key.
Ethernet=Timeout
3.Press the key to select On and then press the key.
Timeout=On
4.Set the timeout time and then press the key.
Key operations
• Use the
• Use the
Duration=
Ethernet timeout
Setting complete
key to select the digit for entering a value.
key to select the value you wish to enter.
5Press the ESC key to return to the higher level menu. To save the settings and
exit from Basic Setting mode, proceed to “Saving the Settings.”
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
CHARACTER
2-11
2.6 Setting the Communication Timeout and Keepalive
Keepalive
1.Press the
Basic=Ethernet
key to select Ethernet and then press the key.
2.Press the key to select K.Alive and then press the key.
Ethernet=K. Alive
3.Press the key to select On and then press the key.
The setting complete screen appears.
Keep alive=On
Keep alive
Setting complete
5Press the ESC key to return to the higher level menu. To save the settings and
exit from Basic Setting mode, proceed to “Saving the Settings.”
Saving the Settings
1.Press the ESC key to return to the Basic= screen.
2.Press the
3.Press the
key to select End and then press the key.
key to select Store and then press the key.
The settings are activated, and the Operation mode screen appears.
Explanation
The communication timeout function and the keepalive function can be configured.
• Communication Timeout
• Selecting On or Off
If On is selected, set the timeout time.
• Timeout Time
If communication timeout is enabled, the connection is dropped if no data transfer
is detected over a time period specified here.
Selectable range: 1 to 120 minutes
• Enabling (On)/Disabling (Off) Keepalive
Select On to enable the keepalive function.
• Saving the Settings
To activate the settings made in the Basic Setting mode, the settings must be saved.
<Related Topics>
Keepalive: Section 1.1
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
2-12
CHARACTER
Chapter 3 Using the RS-422A/485 Communication Interface (/C3 Option)
3.1RS-422A/485 Communication Interface
Specifications
This section describes the RS-422A/485 communication interface specifications.
ItemSpecifications
Terminal block typeNumber of terminals: 6, terminal attachment screws: ISO
M4/nominal length of 6 mm
Electrical and mechanicalComplies with the EIA-422A(RS-422A) and
specificationsEIA-485(RS-485) standards
ConnectionMultidrop Four-wire system1:32
Transmission modeHalf-duplex
SynchronizationStart-stop synchronization
Baud rate
Start bitFixed to 1 bit
Data lengthSelect 7 or 8 bits
ParitySelect Odd, Even, or None (no parity).
Stop bitFixed to 1 bit
Received buffer length2047 bytes
Escape sequenceOpen and close
Electrical characteristics6 points consisting of FG, SG, SDB, SDA, RDB, and RDA
Communication distanceUp to 1.2 km
Terminal resistance120 Ω, 1/2 W
Select from 1200, 2400, 4800, 9600, 19200, and 38400 [bps].
The SG, SDB, SDA, RDB, and RDA terminals and the
internal circuitry of the recorder are functionally isolated.
The FG terminal is the frame ground.
Two-wire system1:31 (Modbus slave protocol)
3
Using the RS-422A/485 Communication Interface (/C3 Option)
3-1
3.2Terminal Arrangement and Signal Names and
the Connection Procedure of the RS-422A/485
Communication Interface
Terminal Arrangement and Signal Names
Rear panel
L
ALARM
N
NO C NC
NO C NC
NO C NC
NO C NC
CHART
Connection Procedure
• Cable
NO C NC
NO C
NC
REMOTE
23
1
5C
4
RS-422A
SD SG
SD
BG
A
RD FG
RD
BG
A
Terminal NameDescription
FG (Frame Ground)Case ground of the recorder.
SG (Signal Ground)Signal ground.
SDB (Send Data B)Send data B (+).
SDA (Send Data A)Send data A (–).
RDB (Received Data B)Receive data B (+).
RDA (Received Data A)Receive data A (–).
Use the cable that meets the conditions below.
ItemConditions
CableShielded twisted pair cable
Characteristic impedance100 Ω
Capacitance50 pF/m
Cable lengthUp to 1.2 km*
* The transmission distance of the RS-422A/485 interface is not the straight-line distance, but
rather the total length of the (shielded twisted-pair) cable.
SDA SDB SG
RS-422A
SD
A
RD
A
SD SG
B
RD FG
B
RDA RDB FG
3 pairs AWG24-14 (Four-wire system),
2 pairs 24 AWG or more (Two-wire system)
• Connecting the Cable
As shown in the following figure, attach a crimp-on lug with isolation sleeves for 4 mm
screws to the end of the cable. Keep the exposed section from the end of the shield
within 5 cm.
Four-wire systemTwo-wire system
3-2
SDBSDASG
FG
RDBRDA
Shield
Electric
potential of
the shield
SDBSDASG
FG
RDBRDA
Shield
Electric
potential of
the shield
3.2
Terminal Arrangement and Signal Names and the Connection Procedure of the RS-422A/485 Communication Interface
WARNING
To prevent the possibility of electric shock, connect the cables with the power
turned OFF.
Note
• Connect the RD pin to the SD (TD) pin on the PC (converter) end and the SD pin to the RD
pin on the PC end.
• The two-wire system can be used only when using the Modbus protocol.
Connection Example with a Host Computer
A connection can be made with a host computer having a RS-232, RS-422A, or RS-485
port.
• In the case of RS-232, a converter is used.
• For recommended converters, see “Serial Interface Converter” on the next page.
• The two-wire system can be used only when using the Modbus protocol. For the
configuration procedure, see section 3.5
• Four-Wire System
Generally, a four-wire system is used to connect to a host computer. In the case of a
four-wire system, the transmission and reception lines need to be crossed over.
Host
computer
Terminator (external) 120 Ω 1/2W or greater
RS-422A/485
terminal on the
recorder
Terminator (external)
3
Using the RS-422A/485 Communication Interface (/C3 Option)
SDA( - )
SDB( + )
RDA( - )
RDB( + )
SG
(SDA)
(SDB)
(RD A)
(RD B)
(SG)
SD A
SD B
RD A
RD B
#1
SG
FG
(SDA)
(SDB)
(RD A)
(RD B)
(SG)
SD A
SD B
RD A
RD B
SG
FG
(SDA)
(SDB)
(RD A)
(RD B)
(SG)
#2#n
Do not connect terminators to #1 through #n-1.
SD A
SD B
RD A
RD B
SG
FG
(#n ≤ 32)
3-3
3.2
Terminal Arrangement and Signal Names and the Connection Procedure of the RS-422A/485 Communication Interface
(The following figure illustrates the case when the host computer’s interface is RS-232.)
Host
computer
RS-232
TD( - )
TD( + )
RD( - )
RD( + )
SHIELD
Converter
Host
computer
SDA( – )
SDB( + )
RDA( – )
RDB( + )
SG
Terminator (external) 120 Ω 1/2W or greater
RS-422A/485
terminal on the
recorder
(SDA)
(SDB)
(RD A)
(RD B)
(SG)
SD A
SD B
RD A
RD B
SG
FG
(SDA)
(SDB)
(RD A)
(RD B)
(SG)
#1
Do not connect terminators to #1 through #n-1.
• Two-Wire System
Connect the transmission and reception signals with the same polarity on the RS422A/485 terminal block. The two-wire system can be used only when using the
Modbus protocol.
Terminator (externally attached) 120 Ω, 1/2 W or more
RS-422A/485
terminal on the
recorder
(A)
(B)
(SG)
SD A
SD B
RD A
RD B
SG
FG
(A)
(B)
(SG)
Terminator (external)
SD A
SD B
RD A
RD B
SG
FG
(SDA)
(SDB)
(RD A)
(RD B)
(SG)
#2#n
Terminator (externally attached)
SD A
SD B
(A)
(B)
RD A
RD B
SG
(SG)
FG
SD A
SD B
RD A
RD B
SG
FG
(#n ≤ 32)
SD A
SD B
RD A
RD B
SG
FG
3-4
#1
#2#n
Do not connect terminator to #1 to #n–1
(#n ≤ 31)
3.2
Terminal Arrangement and Signal Names and the Connection Procedure of the RS-422A/485 Communication Interface
(The following figure illustrates the case when the host computer’s interface is RS-232.)
Host
computer
RS-232
RD( – )
RD( + )
TD( – )
TD( + )
SHIELD
Converter
Terminator (externally attached) 120 Ω, 1/2 W or more
RS-422A/485
terminal on the
Recorder
(A)
(B)
(SG)
SD A
SD B
RD A
RD B
SG
FG
#1
(A)
(B)
(SG)
Do not connect terminator to #1 to #n–1
Note
• The method used to eliminate noise varies depending on the situation. In the connection
example, the shield of the cable is connected only to the recorder’s ground (one-sided
grounding). This is effective when there is a difference in the electric potential between the
computer’s ground and the recorder’s ground. This may be the case for long distance
communications. If there is no difference in the electric potential between the computer’s
ground and the recorder’s ground, the method of connecting the shield also to the computer’s
ground may be effective (two-sided grounding). In addition, in some cases, using two-sided
grounding with a capacitor connected in series on one side is effective. Consider these
possibilities to eliminate noise.
• When using the two-wire type interface (Modbus protocol), the 485 driver must be set to high
impedance within 3.5 characters after the last data byte is sent by the host computer.
Terminator (externally attached)
SD A
SD B
RD A
RD B
SG
FG
#2#n
(A)
(B)
(SG)
SD A
SD B
RD A
RD B
(#n ≤ 31)
3
Using the RS-422A/485 Communication Interface (/C3 Option)
SG
FG
• Serial Interface Converter
On some non-recommended converters, the signal polarity may be reversed (A/B or
+/- indication). In this case, reverse the connection.
CAUTION
Some converters have FG and SG pins that are not isolated. In this case, do
not connect anything to the converter’s FG and SG pins (unlike the figure on the
previous page). Especially in the case of long distance communications, the
potential difference that appears may damage the recorder or cause
communication errors. For converters that do not have the SG pin, they can be
used without using the signal ground. For details, see the manual that came
with the converter.
3-5
3.2
Terminal Arrangement and Signal Names and the Connection Procedure of the RS-422A/485 Communication Interface
For a two-wire system, the host computer must control the transmission driver of the
converter in order to prevent collisions of transmit and received data. When using the
recommended converter, the driver is controlled using the RS (RTS) signal on the RS-
232.
• When Instruments That Support Only the RS-422A Interface Exist in the System
When using the four-wire system, up to 32 recorders can be connected to a single
host computer. However, this may not be true if instruments that support only the RS422A interface exist in the system.
When the instrument that support only the RS-422A interface exist in the
system
The maximum number of connection is 16. Some of OMEGA’s conventional recorder
only support the RS-422A driver. In this case, only up to 16 units can be connected.
Note
In the RS-422A standard, 10 is the maximum number of connections that are allowed on one
port (for a four-wire system).
• Terminator
When using a multidrop connection (including a point-to-point connection), connect a
terminator only to the recorder on the end of the chain. In addition, turn the terminator
on the host computer ON (see the computer’s manual). If a converter is being used,
turn ON its terminator. The terminator must be attached externally to the
recommended converters.
3-6
3.3The Bit Structure of One Character and the
Operation of the Receive Buffer
The Bit Structure of One Character
The serial interface on the recorder communicates using start-stop synchronization. In
start-stop synchronization, a start bit is added every time a character is transmitted. The
start bit is followed by the data bits, parity bit, and stop bit. (See the figure below.)
Return to the idle
state (dotted line)
or the start bit of
the next data
character
(solid line)
Stop bit
Circuit idle
state
Start bit
Receive Buffer and Received Data
The data received from the PC is first placed in the receive buffer of the recorder. When
the received buffer becomes full, all of the data that overflow are discarded.
1 character
Data bit
(7 or 8 bits)
Parity bit
Even, Odd,
or None
3
Using the RS-422A/485 Communication Interface (/C3 Option)
3-7
3.4Modbus Slave Protocol Specifications
The Modbus slave protocol specifications of the recorder are as follows:
ItemSpecifications
Transmission mediumRS-422A/485
Flow controlNone only
Baud rateSelect from 1200, 2400, 4800, 9600, 19200, or 38400 [bps]
Start bitFixed to 1 bit
Stop bitFixed to 1 bit
Parity checkSelect odd, even, or none (no parity).
Transmission modeRTU (Remote Terminal Unit) mode only
Maximum number ofFour-wire system: 32 slave devices
connected unitsTwo-wire system: 31 slave devices
The function codes of the Modbus slave protocol that the recorder supports are shown
below. The recorder does not support broadcast commands.
Function CodeSpecificationsOperation
3Read the hold register (4xxxx).The master device can read
4Read the input register (3xxxx).The master device loads the
6Single write to hold register (4xxxx)The master device writes to
8Loopback testThe master device performs a
16Write to the hold register (4xxxx)The master device writes to
* Hexadecimal ”00”
• Data length: 8 bits
• Data interval: 24 bits or less*
• Error detection: Uses CRC-16
* Determines message termination with a time interval to 3.5
characters or more.
the communication input data
written using function codes 6
and 16.
computed, measured, alarm,
and time data of the recordr.
the communication input data
of the recorder.
loopback test of the recorder.
The recorder only supports
message return (test code
0x00*)
the communication input data
of the recorder.
3-8
Registers
3.4
Modbus Slave Specifications
The registers for using the Modbus slave protocol are listed below.
The register data does not contain unit and decimal point position information. Set them
on the Modbus master. Binary values are stored to the register in order from the highest
byte.
Input registerData
30001Measured data of CH01
::
30024Measured data of CH24
• The data is a 16-bit signed integer. The value is the same as the measured data in binary
output format (see page 5-13).
• The decimal point and unit information varies depending on the range setting of each
channel.
• Valid channels vary depending on the device. An error response (code 2) occurs when an
invalid channel is read.
31001Alarm status of the measured data of CH01
::
31024Alarm status of the measured data of CH24
• The data is a 16-bit signed integer. The value is the same as the alarm status in binary
output format (see page 5-13). The data is entered in the "A2A1A4A3" order in the register.
• Valid channels vary depending on the device. An error response (code 2) occurs when an
invalid channel is read.
32001Computed data of CH0A (lower word)
32002Computed data of CH0A (upper word)
32003Computed data of CH0B (lower word)
32004Computed data of CH0B (upper word)
::
32048Computed data of CH1P (upper word)
• Registers corresponding to models with the /M1 computation function option.
• The data is a 32-bit signed integer. Two registers are assigned for each data point. The
value is the same as the computed data in binary output format (see page 5-13).
• The decimal point and unit information varies depending on the span setting of each
channel.
• Valid channels vary depending on the device. An error response (code 2) occurs when an
invalid channel is read.
33001Alarm status of the computed data of CH0A
::
33024Alarm status of the computed data of CH1P
• Registers corresponding to models with the /M1 computation function option.
• Data type and value are the same as those of the alarm status of the measured data.
• Valid channels vary depending on the device. An error response (code 2) occurs when an
invalid channel is read.
36001List of alarms of the measured data of CH01 to CH04
36002List of alarms of the measured data of CH05 to CH08
36003List of alarms of the measured data of CH09 to CH12
36004List of alarms of the measured data of CH13 to CH16
36005List of alarms of the measured data of CH17 to CH20
36006List of alarms of the measured data of CH21 to CH24
Example of Register 36001
1 register (1 word)
3
Using the RS-422A/485 Communication Interface (/C3 Option)
36007 to 36020Always 0
CH4CH3CH2CH1
The bit is set to 1 when an alarm occurs.
Level 1
Level 2
Level 3
Level 4
3-9
3.4
Modbus Slave Protocol Specifications
Input registerData
36021List of alarms of the measured data of CH0A to CH0D
36022List of alarms of the measured data of CH0E to CH0J
36023List of alarms of the measured data of CH0K to CH0P
36024List of alarms of the measured data of CH1A to CH1D
36025List of alarms of the measured data of CH1E to CH1J
36026List of alarms of the measured data of CH1K to CH1P
• The register details are same as the list of alarms of measured data.
• Registers 36001 to 35026 can be read with a single command.
40001Communication input data of C01
::
40024Communication input data of C24
• Pen model: C01 to C08, dot model: C01 to C12 (RD100B), C01 to C24 (RD1800B)
• The data is a 16-bit signed integer.
Modbus Error Response
The recorder returns the following error codes to the master device. For the error
messages related to communications that the recorder displays, see appendix 4.
CodeMeaningCause
1Bad function codeUnsupported function request.
2Bad register numberAttempted to read/write to a register that
3Bad number of registersThe specified number of registers is less
However, no response is returned for the following cases.
• CRC error
• Errors other than those in the table above.
has no corresponding channel.
than or equal to 0 or greater than or
equal to 126 (when reading)/124 (when
writing).
3-10
3.5Setting the Serial Interface
The serial interface must be configured.
Setup Items
RS422/485
Address
Baud rateData lengthParity
Procedure
End
Protocol
For a description of the basic operations, see “Basic Key Operations” on page v.
Entering Basic Setting Mode
Hold down the
Next, hold down both the
MENU
key for 3 seconds to display the Setting mode screen.
DISP
(
) key and the (
FUNC
) key for 3 seconds to
display the Basic Setting mode screen.
Note
To cancel an operation, press the ESC key.
1.Press the key to select RS422/485 and then press the key.
Basic=RS422/485
2.Press the key to select the recorder’s address and then press the key.
Address=1
3.Press the key to select the Baud rate value and then press the key.
Baud rate=38400
4Press the key to select the Data length value and then press the key.
Data length=8
3
Using the RS-422A/485 Communication Interface (/C3 Option)
5Press the key to select the Parity value and then press the key.
Parity=Even
6.Press the key to select the NORMAL value and then press the key.
The setting complete screen appears.
Protocol=NORMAL
RS422/485
Setting complete
Saving the Settings
1.Press the ESC key to return to the Basic= screen.
2.Press the
3.Press the
key to select End and then press the key.
key to select Store and then press the key.
The settings are activated, and the Operation mode screen appears.
3-11
3.5 Setting the Serial Interface
Explanation
• Address
• Baud rate
• Data length
• Parity (Parity check method)
• Protocol
Select the address from the following range.
01 to 32
Select the baud rate from the following:
1200, 2400, 4800, 9600, 19200, or 38400
Select the data length from below. To output data in BINARY format, be sure to set
the data length to 8 bits.
7 or 8
Select the parity check from the following:
Odd, Even, or None
Select the protocol when using the Modbus slave protocol.
NORMAL: Standard protocol
MODBUS: Modbus slave protocol
• Saving the Settings
To activate the settings made in the Basic Setting mode, the settings must be saved.
3-12
Chapter 4 Commands
4.1Command Syntax
The syntax of the setting/basic setting/output commands (see sections 4.4 to 4.7) of the
instrument is given below. ASCII codes (see appendix 1) are used for the character codes.
For the Maintenance/Test command syntax, see section 4.9.
For the Instrument Information server command syntax, see section 4.10.
Command nameParameter
Command example
SR 02,SKIP;SR 03,VOLT,2V,-1500,1800
• Command Name
Defined using two alphabet characters.
?
Terminator
Delimiter
Sub delimiter
4
Commands
Parameter (02 SKIP·····1800)
Delimiter (,)
Command name (SR)
Sub delimiter (;)
• Parameter
• Command parameters.
• Set using alphabet characters or numerical values.
• Parameters are separated by delimiters (commas).
• When the parameter is a numerical value, the valid range varies depending on the
command.
• Spaces before and after of the parameter are ignored (except for parameters that
are specified using an ASCII character string (unit, tag, and message string), when
spaces are valid.)
• You can omit the parameters that do not need to be changed from their current
settings. However, delimiters cannot be omitted.
ExampleSR 01,,2V<terminator>
If multiple parameters are omitted and delimiters occur at the end of the command,
those delimiters can be omitted.
• By placing a query after a command or parameter, the setting information of the
corresponding command can be queried. Some commands cannot execute
queries. For the query syntax of each command, see sections 4.4 to 4.7.
Example 1SR[ p1]?SR? or SR p1? can be executed.
Example 2SA[ p1[,p2]]? SA?, SA p1?, or SA p1,p2? can be executed.
• Delimiter
• A comma is used as a delimiter.
• Parameters are separated by delimiters.
• Sub Delimiter
• A semicolon is used as a sub delimiter.
• By separating each command with a sub delimiter, up to 10 commands can be
specified one after another. However, the following commands and queries cannot
be specified one after another. Use them independently.
• Output commands other than BO, CS, and IF commands.
• YE command
• Queries
* If there are consecutive sub delimiters, they are considered to be single. In addition, sub
delimiters at the front and at the end are ignored.
Example;SR01,VOLT;;;SR02,VOLT;<terminator> is taken to be
SR01,VOLT;SR02,VOLT<terminator>.
Response
• Terminator (Terminating Character)
Use either of the following two characters for the terminator.
• CR + LF (0DH 0AH in ASCII code.)
• LF (0AH in ASCII code.)
Note
• The total data length from the first character to the terminator must be less than 2047 bytes. In
addition, the character string length of 1 command must be less than 512 bytes.
• Commands are not case sensitive (with the exception of user-specified character strings).
• All the commands that are listed using sub delimiters are executed even if one of the
commands is erroneous.
• Spaces that are inserted before and after a parameter are ignored. However, if spaces are
inserted before a command, after a sub delimiter, or after a query, an error occurs.
The recorder returns a response (affirmative/negative response) to a command that is
separated by a single terminator.* The controller should follow the one command to one
response format. When the command-response rule is not followed, the operation is not
guaranteed.
For the response syntax, see section 5.1.
* The exceptions are the RS-422A/485 dedicated commands (see section 4.8).
Note
When using the RS-422A/485 interface, allow at least 1 ms before sending the next command
after receiving a response. Otherwise, the command may not be processed correctly.
4-2
4.2A List of Commands
Execution Modes and User Levels
Execution Modes
The recorder has two execution modes. Each command is specified to be used in a
particular execution mode. If you attempt to execute a command in a mode that is
different from the specification, a syntax error occurs. Use the DS command to switch to
the appropriate mode, and then execute the command. Query commands can be
executed in either mode.
• Basic Setting mode
Measurement/computation is stopped and settings are changed in this mode.
• Run mode
Run mode
User Levels
The administrator and user specifications in the table indicate the user level that is
specified using the login function for Ethernet communications. For details, see section
1.1.
Setting Commands
Command FunctionExecution ModeAdministrator User Page
Name
SRSets the input range.Run modeYesNo4-10
SOSets the computing equation (/M1 option).Run modeYesNo4-11
VBSets the bias.Run modeYesNo4-11
SASets the alarm.Run modeYesNo4-11
SNSets the unit.Run modeYesNo4-12
SCSets the chart speed.Run modeYesNo4-12
SDSets the date and time.Run modeYesNo4-13
VTSets the dot printing interval (dot model).Run modeYesNo4-13
SZSets zone recording.Run modeYesNo4-13
SPSets the partial expanded recording.Run modeYesNo4-13
VRTurns ON/OFF the recording on each channel.Run modeYesNo4-13
STSets the tag.Run modeYesNo4-14
SGSets the message.Run modeYesNo4-14
SESets the secondary chart speedRun modeYesNo4-14
(used by the remote control function (/R1 option)).
SVSets the moving average (dot model).Run modeYesNo4-14
SFSets the input filter (pen model).Run modeYesNo4-14
BDSets the alarm delay time.Run modeYesNo4-14
VFSets the brightness of the display (VFD) and internalRun modeYesNo4-14
illumination.
TDSets the DST.Run modeYesNo4-14
SSSets the DST. (Can be used on the RD100B with system Run modeYesNo4-15
version 1.02 or earlier)
SKSets the computation constant (/M1 option).Run modeYesNo4-15
SJSets the timer used in TLOG computation (/M1 option).Run modeYesNo4-15
CMSets the communication input data (/M1 option).Run modeYesNo4-15
FRSets the acquiring interval to the FIFO buffer.Run modeYesNo4-15
VDSets the data display screen.Run modeYesNo4-16
collectively refers to Operation mode and Setting mode of the recorder.
Yes: Command usable
No: Command not usable
4
Commands
4-3
4.2 A List of Commands
Basic Setting Commands
Note
• In order to activate the settings that are changed using the basic setting commands, the
settings must be saved using the YE or XE command.
• The settings that are returned in response to a query in the basic setting mode will contain the
new settings even if they are not saved.
• When the settings are saved with the XE command, communication is not dropped. The
settings that are changed using the YS/YB/YA/YN/YD/YQ/YK command are activated after
power-cycling the recorder.
• When the YE command is executed, communication is dropped. The response to the YE
command is not returned.
Command FunctionExecution ModeAdministrator User Page
Name
XASets alarm related settings.Basic Setting modeYesNo4-18
XISets the A/D integral time.Basic Setting modeYesNo4-18
XBSets the burnout detection.Basic Setting modeYesNo4-19
XJSets the RJC.Basic Setting modeYesNo4-19
UCChanges the dot color (dot model).Basic Setting modeYesNo4-19
UOSets the pen offset compensation (pen model).Basic Setting modeYesNo4-19
UPSets the items to be printed.Basic Setting modeYesNo4-19
URSets the periodic printout interval.Basic Setting modeYesNo4-19
UMSets the types of report data that are output to theBasic Setting modeYesNo4-20
periodic printout.
UBSets the display mode of the bar graph.Basic Setting modeYesNo4-20
UISets whether to use moving average (dot model).Basic Setting modeYesNo4-20
UJSets whether to use the input filter (pen model).Basic Setting modeYesNo4-20
UKSets whether to use of partial expanded recording.Basic Setting modeYesNo4-20
ULSelects the display/record language.Basic Setting modeYesNo4-20
XNSelects the date format.Basic Setting modeYesNo4-20
XTSelects the temperature unit.Basic Setting modeYesNo4-21
UFSets whether to use the extended functions.Basic Setting modeYesNo4-21
UTSelects the time printout format.Basic Setting modeYesNo4-21
XRSets the remote control input (/R1 option).Basic Setting modeYesNo4-21
YSSets the RS-422A/485 interface (/C3 option).Basic Setting modeYesNo4-21
XQSets the TLOG timer (/M1 option).Basic Setting modeYesNo4-22
UNChanges the assignment of channels to the recording pen Basic Setting modeYesNo4-22
(pen model, /M1 option).
USSets the computation error procedure (/M1 option).Basic Setting modeYesNo4-22
YBSets the host name and domain name (/C7 option).Basic Setting modeYesNo4-22
YASets the IP address (/C7 option).Basic Setting mode YesNo4-22
YNSets the DNS (/C7 option).Basic Setting modeYesNo4-23
YDSets whether to use the login function via communication Basic Setting mode YesNo4-23
(/C7 option).
YQSets the communication timeout (/C7 option)Basic Setting modeYesNo4-23
quitDisconnects the connection of the instrument beingAll modesYesYes4-30
operated.
Yes: Command usable
No: Command not usable
Instrument Information Output Commands
These commands can be used only when using Ethernet communications.
ParameterFunctionPage
serialOutputs the serial number.4-30
hostOutputs the host name.4-30
ipOutputs the IP address.4-30
4-6
4.3Parameter Values
This section explains frequently used parameters.
Input Range
The following tables show the input types (VOLT, TC, RTD, DI, and 1-5V), range types,
and the ranges for the leftmost and rightmost values of the span.
• DC Voltage (VOLT), Square Root (SQRT), Difference between Channels (DELTA)
Range TypeParameter forRange of Leftmost andRange of Leftmost and
the SR Command Rightmost Values of Span Rightmost Values of Span
20 mV20 mV–20.00 to 20.00 mV–2000 to 2000
60 mV60 mV–60.00 to 60.00 mV–6000 to 6000
200 mV200 mV–200.0 to 200.0 mV–2000 to 2000
2 V2 V–2.000 to 2.000 V–2000 to 2000
6 V6 V–6.000 to 6.000 V–6000 to 6000
20 V20 V–20.00 to 20.00 V–2000 to 2000
50 V50 V–50.00 to 50.00 V–5000 to 5000
• 1-5V
Range TypeParameter forRange of Leftmost andRange of Leftmost and
the SR Command Rightmost Values of Span Rightmost Values of Span
1-5V1-5VLeftmost value:800 to 1200
of the SR Command
4
Commands
of the SR Command
0.8000 to 1.200 V
Rightmost value:4800 to 5200
4.8000 to 5.200 V
• Thermocouple (TC)
Range TypeParameter forRange of Leftmost andRange of Leftmost and
the SR Command Rightmost Values of Span Rightmost Values of Span
RR0.0 to 1760.0°C0 to 17600
SS0.0 to 1760.0°C0 to 17600
BB0.0 to 1820.0°C0 to 18200
KK–200.0 to 1370.0°C–2000 to 13700
EE–200.0 to 800.0°C–2000 to 8000
JJ–200.0 to 1100.0°C–2000 to 11000
TT–200.0 to 400.0°C–2000 to 4000
NN0.0 to 1300.0°C0 to 13000
WW0.0 to 2315.0°C0 to 23150
LL–200.0 to 900.0°C–2000 to 9000
UU–200.0 to 400.0°C–2000 to 4000
WReWRe0.0 to 2400.0°C0 to 24000
Difference between channels (DELTA)
R-–1760.0 to 1760.0°C–17600 to 17600
S-–1760.0 to 1760.0°C–17600 to 17600
B-–1820.0 to 1820.0°C–18200 to 18200
K-–1570.0 to 1570.0°C–15700 to 15700
E-–1000.0 to 1000.0°C–10000 to 10000
J-–1300.0 to 1300.0°C–13000 to 13000
T-–600.0 to 600.0°C–6000 to 6000
N-–1300.0 to 1300.0°C–13000 to 13000
W-–1999.9 to 2315.0°C–19999 to 23150
L-–1100.0 to 1100.0°C–11000 to 11000
U-–600.0 to 600.0°C–6000 to 6000
WRe-–1999.9 to 2400.0°C–19999 to 24000
of the SR Command
4-7
4.3 Parameter Values
Miscellaneous
• Resistance Temperature Detector (RTD)
Range TypeParameter forRange of Leftmost andRange of Leftmost and
Pt100PT–200.0 to 600.0°C–2000 to 6000
JPt100JPT–200.0 to 550.0°C–2000 to 5500
Difference between channels (DELTA)
Pt100-–800.0 to 800.0°C–8000 to 8000
JPt100-–750.0 to 750.0°C–7500 to 7500
the SR Command Rightmost Values of Span Rightmost Values of Span
of the SR Command
• ON/OFF input (DI)
Range TypeParameter forRange of Leftmost andRange of Leftmost and
LevelLEVEL0 to 1
ContactCONT0 to 1
†
“0” when less than 2.4 V, “1” when greater than or equal to 2.4 V.
††
“0” when contact is OFF, “1” when contact is ON.
the SR Command Rightmost Values of Span Rightmost Values of Span
Description • Specifies the address of the instrument with
ISOutputs status information.
SyntaxIS p1<terminator>
p1 Output status information (0)
ExampleOutput status information.
IS 0
Description The output status can be masked using the
status filter (IF command).
FUOutputs user information.
SyntaxFU p1<terminator>
p1 Output user information (0)
ExampleOutput user information.
FU 0
Description Outputs the information of the user currently
connected to the recorder.
ESC CCloses the instrument.
SyntaxESC C p1<terminator:CR+LF>
ExampleClose the instrument at address 01.
Description • Clears the current connection with the
Commands
The ASCII code of ESC is 1BH.
p1 Instrument address (01 to 32)
ESC O 01
which to communicate.
• Only one instrument can be opened at any
given time.
• When an instrument is opened with the ESC
O command, any other instrument that is
currently open is automatically closed.
• When this command is received correctly, the
recorder transmits the data “ESC 0 xx” (xx:
Instrument’s address).
• Use CR+LF for the terminator. LF cannot be
used.
The ASCII code of ESC is 1BH.
p1 Instrument address (01 to 32)
ESC C 01
instrument.
• When this command is received correctly, the
recorder transmits the data “ESC C xx” (xx:
Instrument’s address).
• Use CR+LF for the terminator. LF cannot be
used.
4-28
4.9Maintenance/Test
Commands (Available when
using the maintenance/test
server function via
Ethernet communications)
closeDisconnects the connection
between other instruments.
Syntaxclose,p1,p2:p3<terminator>
p1 Port on the recorder side (0 to 65535)
p2 IP address on the PC side
(0.0.0.0 to 255.255.255.255)
p3 Port on the PC side (0 to 65535)
Exampleclose,34159,192.168.111.24:1054
E0
Description This command cannot be used to disconnect a
server port. Also, it cannot disconnect the
recorder being operated. Use the quit command
for this purpose.
conOutputs connection information.
Syntaxcon<terminator>
Example
con
EA
04/10/01 12:34:56
Active connections
Proto Local AddressForeign AddressState
TCP192.168.111. 24:34159 192.168.111. 24:1053
ESTABLISHED
TCP0. 0. 0. 0:341550. 0. 0. 0: 0 LISTEN
TCP0. 0. 0. 0:341590. 0. 0. 0: 0 LISTEN
TCP0. 0. 0. 0:341500. 0. 0. 0: 0 LISTEN
EN
TCP
Protocol used.
Local Address
The recorder’s socket address.
Displays “IP address:port number.”
Foreign Address
The destination socket address.
Displays “IP address:port number.”
State
Connection status.
ESTABLISHED
Connection established.
4.9 Maintenance/Test Commands
ethOutputs Ethernet statistical
information.
Syntaxeth<terminator>
Example
eth
EA
04/10/01 12:34:56
Ethernet Statistics
Name In Pkt In Err Out Pkt Out Err16 Coll
lo000000
mb07406400
EN
helpOutputs help.
Syntaxhelp [,p1]<terminator>
p1 Command name
(close, con, eth, help, net, quit)
Example
help
EA
con- echo connection information
eth- echo ethernet information
help- echo help
net- echo network status
quit- close this connection
EN
netOutputs network statistical
information.
Syntaxnet<terminator>
Example
net
EA
04/10/01 12:34:56
Network Status
APP: power on time = 00/00/00 12:34:56
APP: applalive = disable
APP: genedrops = 0
APP: diagdrops = 0
APP: ftpsdrops = 0
TCP: keepalive = 30 s
TCP: connects = 14
TCP: closed = 0
TCP: timeoutdrop = 0
TCP: keepdrops = 0
TCP: sndtotal = 53
TCP: sndbyte = 0
TCP: sndrexmitpack = 0
TCP: sndrexmitbyte = 1
TCP: rcvtotal = 0
TCP: rcvbyte = 0
DLC: 16 collisions = 0
EN
4-29
4
Commands
4.10 Instrument Information Output Commands (Available when using the instrument information server
function via Ethernet communications)
TCP: keepalive
Keepalive check cycle.
TCP: connects
Total number of connections established.
TCP: closed
Total number of dropped connections.
TCP: timeoutdrop
Total number of dropped connections due to
TCP retransmission timeout. When the
transmitted packet (the unit of transmitted
data) is not received, the packet is
automatically retransmitted at a
predetermined time interval. If the packet is
not received after 14 retransmissions,
timeout occurs and the connection is
dropped.
TCP: keepdrops
Total number of dropped connections due to
TCP keepalive timeout.
TCP: sndtotal
Total number of transmitted packets.
TCP: sndbyte
Total number of transmitted bytes.
TCP: sndrexmitpack
Total number of retransmitted packets.
TCP: sndrexmitbyte
Total number of retransmitted bytes.
TCP: rcvtotal
Total number of received packets.
TCP: rcvbyte
Total number of received bytes.
DLC: 16 collisions
Number of collision incidents. A collision
occurs when two or more instruments on the
network attempt to transmit simultaneously.
The tendency for collisions to occur
increases when the network is congested.
16 collisions would mean 16 consecutive
collision incidents.
quitDisconnects the connection of
the instrument being operated
Syntaxquit<terminator>
4.10 Instrument Information
Output Commands
(Available when using the
instrument information
server function via
Ethernet communications)
The instrument information server function
interprets one UDP packet to be one command
and returns a single packet (containing the
recorder’s information) in response to the
command.
Port number34264/udp
(see section 2.1)
Transfer dataASCII
Received buffer length128
Transmit buffer length512
Maximum number of parameters 32
In the command packet, parameters
corresponding to the desired information are
placed one after another.
Parameter Description
serialOutputs the serial number.
hostOutputs the host name (the host name specified
in section 2.3).
ipOutputs the IP address (the IP address specified
in section 2.3).
ExampleQuery the IP address and host name. (Of the
two frames below, the top frame represents the
command packet, the bottom frame represents
the response packet.)
ip host
EA
ip = 192.168.111.24
host = ABC
EN
Description • Separate each parameter with one or more
blanks (space, tab, carriage return, line feed).
• Parameters are not case sensitive.
• Undefined parameters are ignored.
• Parameters beyond the 32nd parameter are
ignored.
4-30
Chapter 5 Responses
5.1Response Syntax
The following table shows the types of responses for various commands described in the
previous chapter.
The recorder returns a response (affirmative/negative response) to a command that is
separated by a single terminator. The controller should follow the one command to one
response format.
Setting/Measurement
server
Maintenance/Test Server
Instrument Information server
Note
The “
CRLF
” used in this section denotes a terminator.
CommandFunction
Setting command
Basic setting command
Control command
Output
command
See section 4.9.
See section 4.10.
Command TypeAffirmationNegation
Setup, measurement,
and computation data
output
RS-422A/485 dedicated
Response
Affirmative
response
ASCII output
BINARY output
Dedicated response
Single
negative
response
or
Multiple
negative
responses
No response
5
Responses
Affirmative Response
When the command is processed correctly, an affirmative response is returned.
• Syntax
E0
CRLF
• Example
E0
Single Negative Response
When the command is not processed correctly, a single negative response is returned.
• Syntax
E1_nnn_mmm•••m
nnnError number (001 to 999)
mmm•••mMessage (variable length, one line)
_Space
• Example
E1 001 "System error"
Multiple Negative Responses
• If there is an error in any one of the multiple commands that are separated by sub
delimiters, multiple negative responses are returned.
• The response is generated for each erroneous command.
• If there are multiple commands that have errors, the negative responses are
separated by commas.
• The error position number is assigned to the series of commands in order starting with
“1” assigned to the first command.
• Syntax
E2_ee:nnn
E2_ee:nnn,ee:nnn,•••,ee:nnn
eeError position (01 to 10)
nnnError number (001 to 999)
_Space
CRLF
CRLF
CRLF
(When there is only one error)
(When there are multiple errors)
5-1
5.1 Response Syntax
ASCII Output
BINARY Output
• Example
E2 02:001
The following types of ASCII data are available. For the data formats, see section 5.2.
Setting data, basic setting data, decimal position/unit information, measured/
computed data, report data generated by the periodic printout, status information, and
user information
• Syntax
EA
CRLF
•••••••••••••••
CRLF
:
•••••••••••••••
•••••••••••••••
EN
CRLF
CRLF
CRLF
• Conceptual Diagram
1 byte
ASCII
BINARY
BINARY header
(12 bytes)
BINARY data
BINARY footer
(2 bytes)
'E''B'CRLF
Data length
FlagID
Data sum
Header sum
BINARY data
•EBCRLF
Indicates that the data is BINARY.
• Data Length
The byte value of “flag + identifier + header sum + BINARY data + data sum.”
• Header Sum
The sum value of “data length + flag + identifier.”
• BINARY Data
For the output format of various data types, see section 5.3.
• Data Sum
The sum value of “BINARY data.”
Note
The data length of the BINARY header section is output according to the byte order specified
with the BO command.
5-2
5.1 Response Syntax
• Flag
Bit Name (Abbreviation)FlagFlagMeaning of the Flag
01
7BOMSBLSBOutput byte order
6CSNoYesExistence of a checksum
5–––
4–––
3–––
2–––
1–––
0Reserved––Fixed to 1.
• When the BO flag is “0,” the MSB is output first. When the BO flag is “1,” the LSB
is output first.
• If the check sum is enabled (parameter = 1) using the CS command parameter,
each sum value is inserted in the header sum and data sum sections in the
“Conceptual Diagram” on the previous page. If the check sum is disabled
(parameter = 0), a zero is inserted in the header sum and data sum sections. For a
sample program that calculates the sum value, see “Calculating the Sum Value” on
the next page.
• The bits that have “–” for the flag and flag are not used. The value is undefined.
• Identifier
ID NumberBinary Data TypeFormat
0Undefined file–
1Measured/computed dataSection 5.3
1FIFO dataSection 5.3
10Setup data fileUndisclosed
5
Responses
• The table above shows the different types of BINARY Data.
• Measured/computed data can be output using the FD command.
• FIFO data can be output using the FF command.
• The setup data file can be output using the FE command. The setup data file can
be loaded in the cofiguration software.
• The identifier section in the “Conceptual diagram” on the previous page contains
the ID number shown above.
Note
BINARY data that are not indicated in the above table are considered undefined files.
• Calculating the Sum Value
If you set the parameter of the CS command to “1 (enabled),” the checksum value is
output only during serial communications. The check sum is the same as that used in
the TCP/IP and is derived according to the following algorithm.
Buffer on Which the Sum Value Is Calculated
• For the header sum, it is calculated from “data length + flag + identifier” (fixed to 6
bytes).
• For the data sum, it is calculated from “BINARY data.”
1 byte
Padding
0
(1)(2)(3)(4)
(6)(5)
5-3
5.1 Response Syntax
If the data length of the buffer is odd, a “0” is padded so that it is even. (1) through (6)
are summed as unsigned two-byte integers (unsigned short). If the digit overflows a
“1” is added. Finally, the result is bit-wise inverted.
Sample Program
The sum value is determined using the following sample program, and the calculated
result is returned. The sum determined by the sample program can be compared with
the header sum of the output BINARY header section and the data sum of the output
BINARY footer section.
/*
* Sum Calculation Function (for a 32-bit CPU)
*
* Parameters buff: Pointer to the top of the data on which the sum is calculated
*len: Length of the data on which the sum is calculated
* Return value:Calculated sum
*/
intcksum(unsigned char *buff, int len)
{
unsigned short *p;/* Pointer to the next two-byte data word in the buffer
that is to be summed. */
unsigned int csum;/* Checksum value */
inti;
intodd;
csum = 0;/* Initialize. */
odd = len%2;/* Check whether or not the number of data points is
even. */
len >>= 1;/* Determine the number of data points using a
“short” data type. */
p = (unsigned short *)buff;
5-4
for(i=0;i<len;i++)/* Sum using an unsigned short data type. */
csum += *p++;
if(odd){/* When the data length is odd */
union tmp{/* Pad with a 0, and add to the unsigned short data.
csum = csum - 0xffff;/* If the digit overflows again, add a 1. */
return((~csum) & 0xffff); /* bit inversion */
}
RS-422A/485 Dedicated Commands and Responses
The following table shows dedicated commands for the RS-422A/485 interface and their
responses.
Command SyntaxMeaningResponse
ESC
O_xx
CRLF
(_ space)• Response when the instrument with the
ESC
C_xx
CRLF
(_ space)• Response when the instrument with the
* The causes that the condition become “The instrument with the specified address does not
exist” is such as a command error, the address not matching that of the instrument, the
instrument is not turned ON, and the instrument not being connected via the serial interface.
• The “xx” in the table indicates the instrument’s address. Specify the address that
is assigned to the instrument from 01 to 32.
• Only one instrument can be opened at any one time.
• When an instrument is opened with the ESC O command, all commands on the
instrument become active.
• When an instrument is opened with the ESC O command, any other instrument
that is open is automatically closed.
• Use CR+LF for the terminator. LF cannot be used.
Open the instrument.• Response from the instrument with the
Close the instrument. • Response from the instrument with the
5.1 Response Syntax
specified address
ESC
O xx
CRLF
specified address does not exist*
specified address
ESC
C xx
CRLF
specified address does not exist*
5
Responses
Note
• The ASCII code of ESC is 1BH.
5-5
5.2Output Format of ASCII Data
The following types of ASCII data are available.
• Setting data/Basic Setting data
• Decimal point position/unit information
• Measured/computed data
• Report data generated by the periodic printout
• Status information
• User information
Note
The “
CRLF
” used in this section denotes a terminator.
Setting/Basic Setting data
• The FE command (FE0 or FE2) is used to output the data.
• The setting/basic setting data are output in the order of the listed commands in the
table in section 4.2, “A List of Commands.” However, the setting information for the
following commands is not output.
• Setting commands
SD, CM, and FR
• Basic Setting commands
YE and XE
• The output format of the setting/basic setting data is the same as the syntax of each
command.
• Some commands are output in multiple lines. (Example: Commands that are
specified for each channel.)
• Syntax
The two-character command name and the succeeding parameters are output in the
following syntax.
EA
CRLF
ttsss•••s
••••••••••••••
CRLF
EN
ttCommand name (SR, SA•••, XA, XI•••)
sss•••s Setting, basic setting data (variable length, one line)
• Example
EA
SR01,VOLT,20mV,0,20
SR02,VOLT,20mV,0,20
•••••••••••••••••••
EN
CRLF
5-6
Decimal Point Position/Unit Information
• The FE command (FE1) is used to output the data.
• Syntax
The data is output for each channel in the following syntax.
EA
CRLF
s_kccuuuuuu,pp
••••••••••••••••••
EN
CRLF
sData status (N, D, S)
kChannel type
ccChannel number
uuuuuuUnit information (6 characters, left-justified)
ppDecimal point position (00 to 04)
_Space
CRLF
N: Normal
D: Differential input
S: Skip (When the input range is set to SKIP for a measurement
channel or when the channel is turned OFF for a computation
channel)
0: Measurement channel
A: Computation channel
RD100B: 01 to 06, 0A to 0P
RD1800B: 01 to 24, 0A to 1P
mV____: mV
V_____: V
^C____: °C
xxxxxx: (User-defined character string)
No decimal (00000) for 00.
One digit to the right of the decimal (0000.0) for 01.
Two digits to the right of the decimal (000.00) for 02.
Three digits to the right of the decimal (00.000) for 03.
Four digits to the right of the decimal (0.0000) for 04.
5.2 Output Format of ASCII Data
5
Responses
• Example
EA
N 001mV ,01
N 002mV ,01
EN
5-7
5.2 Output Format of ASCII Data
Measured/computed Data
• The FD (FD0) or FY (FYInst, FYTlog1, or FYTlog2) command is used to output the
data.
• Syntax
The measured/computed data are output in the following syntax along with the date
and time information for each channel
EA
DATE_yy/mo/dd
TIME_hh:mi:ss.mmmt_S1S2S3S4S5S
s_kcca1a2a3a4uuuuuuf1dddddEf2pp
••••••••••••••••••••••••••••
EN
CRLF
CRLF
CRLF
6
CRLF
CRLF
yyYear (00 to 99)
moMonth (01 to 12)
ddDay (01 to 31)
hhHour (00 to 23)
miMinute (00 to 59)
ssSecond (00 to 59)
mmmMillisecond (000 to 999. A period is placed between seconds
and milliseconds.)
t‘S’(=summer) or ‘ ’ space(=winter)
S
1S2S3S4S5S6
Data status
Takes on the values below for the FYTlog1 and FYTlog2
commands. For all other cases, they are all spaces.
S
1
Time change during TLOG computation: T (occurred) or space
(not occurred)
S
2
Power OFF and ON during TLOG computation: P (occurred) or
space (not occurred)
S
3
Data reset during the TLOG computation: R (occurred) or
space (not occurred)
S
4S5S6
All spaces
sChannel data status (N, D, S, O, B, E)
N: Normal
D: Differential input
S: Skip
O: Over
B: Burnout
E: Error
kChannel type
0: Measurement channel
A: Computation channel
ccChannel number
RD100B: 01 to 06, 0A to 0P
RD1800B: 01 to 24, 0A to 1P
a
1a2a3a4
a
1
a
2
a
3
a
4
Alarm status (level 1)
Alarm status (level 2)
Alarm status (level 3)
Alarm status (level 4)
5-8
5.2 Output Format of ASCII Data
Each status is set to H, L, h, l, R, r, T, t, or space.
H: high limit alarm, L: low limit alarm, h: difference high-limit
alarm, l: difference low-limit alarm, R: high limit on rate-ofchange alarm, r: low limit on rate-of-change alarm, T: delay
high limit alarm, t: delay low limit alarm, space: no alarm
uuuuuuUnit information (6 characters, left-justified)
mV____: mV
V_____: V
^C____: °C
xxxxxx: (User-defined character string)
f
1
Sign (+, –) of mantissa
• Positive over data, error data, and the burnout data when
“up” is specified are positive.
• Negative over data and the burnout data when “down” is
specified are negaitive.
dddddMantissa (00000 to 99999, 5 digits)
• 8 digits for computed data.
• For error data (channel data status is E), over data (channel
data status is O), or burnout data (channel status data is B),
the mantissa is set to 99999 (99999999 for computed data).
f
2
Sign (+, –) of exponent
ppExponent (00 to 04)
_Space
5
Responses
• Example
EA
DATE 99/02/23
TIME 19:56:32.500
N 001h mV +12345E-03
N 002 mV -12345E-01
S 003
EN
Note
• Data for non-existing channels are not output (not even the channel number).
• For channels set to skip, output values from alarm status to exponent are spaces.
5-9
5.2 Output Format of ASCII Data
Report Data Generated by the Periodic Printout
• The FY command (FYREPORT) is used to output the data.
• Report data generated by the periodic printout is output.
Syntax
EA
CRLF
YY/MO/DD_HH:MI:SS.MMMT
yy/mo/dd_hh:mi:ss.mmmt_S1S2S3S4S5S
s1s2s3s4s5s6kccuuuuuuf1dddddEf2pp_f1dddddEf2pp_f1dddddEf2pp_f1ddddd
Ef
pp_f1ddddddddEf2pp
2
••••••••••••••••••
EN
CRLF
YY/MO/DD HH:MI:SS.MMMTReport start time information
yy/mo/dd hh:mi:ss.mmmtReport end time information
YY, yyYear (00 to 99)
MO, moMonth (01 to 12)
DD, ddDay (01 to 31)
HH, hhHour (00 to 23)
MI, miMinute (00 to 59)
SS, ssSecond (00 to 59)
MMM, mmmMillisecond (000 to 999)
T, t‘S’(=summer) or ‘ ’ space(=winter)
S
1S2S3S4S5S6
S
1
S
2
S
3
S
4S5S6
s
1s2s3s4s5s6
s
1
s
2
s
3
s
4
s
5s6
ccChannel number
kChannel type (0, A)
uuuuuuUnit information (6 characters)
Data status
Time change during reporting: T (occurred) or space
(not occurred)
Power OFF and ON during reporting: P (occurred) or
space (not occurred)
Data clear during reporting: R (occurred) or space (not
occurred)
All spaces
Channel data status
Channel mode at the end of reporting: S (Skipped) or space
(Mode other than skipped)
Range change during reporting: C (occurred) or space (not
occurred)
Error data occurrence during reporting: E (yes) or space (no)
±over data occurrence during reporting: O (yes) or space (no)
All spaces
RD100B: 01 to 06, 0A to 0P
RD1800B: 01 to 24, 0A to 1P
0: Measurement channel
A: Computation channel
CRLF
CRLF
CRLF
6
5-10
5.2 Output Format of ASCII Data
f1dddddEf2pp_f1dddddEf2pp_f1dddddEf2pp_f1dddddEf2pp_f1dddddddd
Ef
pp
2
The data is output in the following order: most recent value,
minimum value, maximum value, average value, and sum
value.
f
dddddEf2pp Most recent value, minimum value, maximum value, and
1
average value of the measurement channel
f
ddddddddEf2pp
1
Sum value of the measurement channel or the most recent
value, minimum value, maximum value, average value, and
sum value of the computation channel
f
1
Sign (+, –) of mantissa
dddddMantissa (00000 to 99999)
dddddddd Mantissa (00000000 to 99999999)
f
• For details on the status information, see section 6.2, “The Bit Structure of the Status
Information.”
• Syntax
EA
ddd.ccc.bbb.aaa
EN
• Example
EA
000.000.032.000
EN
CRLF
CRLF
CRLF
aaaStatus information 1 (000 to 255)
bbbStatus information 2 (000 to 255)
cccStatus information 3 (000 to 255)
dddStatus information 4 (000 to 255)
User Information
• The FU command is used to output the data.
• User name, user level, and other information are output.
• Syntax
EA
CRLF
p_l_uuu•••u
EN
CRLF
pPhysical layer
lUser level
uuu•••uUser name (up to 16 characters)
_Space
• Example
EA
E A admin
EN
CRLF
E: Ethernet
S: RS-422A/485
A: Administrator
U: User
5-12
5.3Output Format of BINARY Data
This section describes the output format of the BINARY data that is disclosed. For the
BINARY output format, see “BINARY Output” on page 5-2. For other BINARY data
types, see “Identifier” on page 5-3.
The measured data is output using signed 16-bit integer; the computed data is output
using signed 32-bit integer. These integers can be understood as physical values by
adding the decimal point and the unit.
Typical Examples to Obtain Physical Values
BINARY ValueDecimal Position CodePhysical Value (Measured Value)
• The FD (FD1) command is used to output the most recent measured/computed data.
• The FF (FEGET, FERESEND, and FEGETNEW) command is used to output the
FIFO data. The decimal point position and unit can be determined using the FE
command.
• The ID number of the output format is 1. See “Identifier” on page 5-3.
5
Responses
2 bytes
Number of blocks
2 bytes
Number of bytes
Block 1
...
Block n
BINARY data
(The BINARY data section
on the “Conceptual
diagram” on page 5-2.)
*The sections indicated as (Reserved) are not used. The value is undefined.
4 bytes
5-13
5.3 Output Format of BINARY Data
• Flag
The meanings of the flags are given on the table below. The flags are valid during
FIFO data output. The flags are undefined for other cases.
BitFlagFlagMeaning of the Flag
6––
5––
4––
3––
2NoYesIndicates that the decimal position or unit information was changed
1NoYesIndicates that the FIFO acquiring interval was changed during
0NoYesIndicates that the internal process took too much time (computation,
The bits that have “–” for the flag column are not used. The value is undefined.
• Block Member
NameBINARY Value
Year0 to 99
Month1 to 12
Day1 to 31
Hour0 to 23
Minute0 to 59
Second0 to 59
Millisecond0 to 999
DST1 (=summer) or 0 (=winter)
Measurement, computation00H: measurement, 80H: computation
ChannelRD100B: 01 to 06 and 31 to 42
Alarm status*
A1 (Bit 0 to 3)
A2 (Bit 4 to 7)0 to 8
A3 (Bit 0 to 3)
A4 (Bit 4 to 7)
Measured data0 to FFFFH
Computed data0 to FFFFFFFFH
* BINARY value 0 to 8 is entered in the upper and lower 4 bits of a byte (8 bits) for the alarm
status. The binary values 0 to 8 correspond to H (high limit alarm), L (low limit alarm), h
(difference high-limit alarm), l (difference low-limit alarm), R (high limit on rate-of-change
alarm), r (low limit on rate-of-change alarm), T (delay high limit alarm), and t (delay low limit
alarm) as follows:
0: no alarm, 1: H, 2: L, 3: h, 4: l, 5: R, 6: r, 7: T, and 8: t.
01
during measurement.
measurement.
for example) and that FIFO dropout occurred.
RD1800B: 01 to 24, 31 to 54
• Special Data Value
The measured/computed data take on the following values under special conditions.
Special Data ValueMeasured DataComputed Data
+ Over7FFFH7FFF7FFFH
– Over8001H80018001H
Skip8002H80028002H
Burnout (when “up” is set)7FFAH7FFF7FFFH
Burnout (when “down” is set)8006H80018001H
Error8004H80048004H
Undefined8005H80058005H
Note
The number of blocks, number of bytes, and measured/computed data are output according to
the byte order specified with the BO command.
5-14
Chapter 6 Status Information
6.1Status Information and Filter
The following figure illustrates the status information and filter on the recorder.
4 bytes
4 3 2 1
Status information
Filter
4 3 2 1
• When a status indicated on the following page is entered, the corresponding bit in the
condition register is set to “1.” The logical AND of the condition register and the filter
becomes the status information.
• The IF command can be used to set the filter.
• The IS command is used to output the status information. Status information 1 and 2
are cleared when they are output. Status information 3 and 4 are not cleared when it
is output, and remains at “1” while the event is occurring.
• When multiple connections are up, filters can be specified for the individual
connection. Therefore, the status information can be held for each connection.
Condition register
6
Status Information
6-1
6.2The Bit Structure of the Status Information
The following four groups of status information are output in response to a status
information output request using the IS command. For the output format, see “Status
Information” in section 5.2, “Output Format of ASCII Data.”
Status Information 1
BitNameDescription
0A/D conversion completeSet to “1” when the A/D conversion of the
1––
2Periodic printout timeoutSet to “1” when the periodic printout timer expires.
3TLOG timeoutSet to “1” when the TLOG timer expires.
4––
5––
6––
7––
Status Information 2
BitNameDescription
0Measurement dropSet to “1” when the measurement process could not
1Decimal point/unitSet to “1” when the decimal point/unit information is
information changechanged.
2Command errorSet to “1” when there is a command syntax error.
3Execution errorSet to “1” when an error occurs during command
4––
5––
6––
7––
measurement is complete.
keep up.
execution.
Status Information 3
Status Information 4
BitNameDescription
0––
1Chart endSet to “1” while the recorder is out of chart paper.
2––
3––
4––
5Chart feedingSet to “1” while the chart is being fed through the
panel key.
6––
7––
BitNameDescription
0Basic settingSet to “1” during Basic Setting mode.
1RecordingSet to “1” while recording is in progress.
2ComputingSet to “1” while computation is in progress.
3Alarm generatingSet to “1” while the alarm is occurring.
4––
5––
6––
7––
6-2
Appendix
Appendix 1ASCII Character Codes
The table below contains the ASCII character codes of characters that can be used. The
table below shows characters each command can use.
Command Used forCharacters
SNUnitAlphanumeric characters, signs, and space
STTagAlphanumeric characters, signs, and space
SGMessageAlphanumeric characters, signs, and space
YBHost/Domain nameAlphanumeric characters and signs
YNDomain suffixAlphanumeric characters and signs
SOComputing equationAlphanumeric characters, signs, and space
Note
“µ”, “Ω”, “2”, “3” , and ”°” are assigned to character codes as follows:
Characters in the parentheses are characters assigned on a keyboard.
Upper 4 bits
7
6
5
4
0
1
3
2
89ABCDEF
0
1
2
3
4
5
6
7
8
Lower 4 bits
9
LF
A
B
C
D
CR
E
ESC
SP
(space)
#
%
(
)
*
+
–
.
m
p
q
a
r
b
s
c
t
d
u
e
v
f
w
g
x
h
y
i
z
j
µ
k
Ω
l
2
3
n
App
Appendix
P
@
0
Q
A
1
R
B
2
S
C
3
T
D
4
U
E
5
V
F
6
W
G
7
X
H
8
Y
I
9
Z
J
K
L
M
°
N
F
/
O
o
App-1
Appendix 2Output Flow of FIFO Data
The recorder has a dedicated internal memory for outputting measured/computed data.
This memory is structured as a FIFO (First-In-First-Out). Measured/computed data are
constantly acquired to the internal memory at the specified acquiring interval (FIFO
acquiring interval, set with the FR command). By using this function, it is possible to
read measured/computed data that have been saved at the specified intervals
regardless of the frequency at which the PC periodically reads the measured/computed
data.
The following example shows the case when the acquiring interval is 1 s and the
buffer capacity is for 8 intervals.
Block 1
2
3
Block 8
4
5
6
7
Most recent acquire
Position: WP
Previous read
Position: RP1
Current read
Position: RP2
• Acquiring of the measured/computed data
• The measured/computed data are acquired to the internal memory at 1 s intervals.
• Measured/computed data are acquired to blocks 1 through 8 in order. After
acquiring to block 8, the next acquiring operation returns to block 1.
• Reading the measured/computed data (FF GET command is used, logging
output)
Outputs the data from the next to the previous read position (RP1) to the most recent
acquire position (WP).
In this example, more than 2 s has elapsed from the previous read operation.
Therefore, data in blocks 5 and 6 are output.
• Reading the measured/computed data (FF GETNEW command is used, output
of the most recent value)
Output the specified number of blocks of FIFO data back starting from the recent
acquire position (WP).
In this example, if you specify the number of blocks to “5,” data in blocks 2 to 6 are
output.
The buffer capacity varies depending on the model.
Pen model: 240 intervals (30 s at an acquiring interval of 125 ms)
Dot model:60 intervals (60 s at an acquiring interval of 1 s)
App-2
Appendix 2 Output Flow of FIFO Data
Example in which the FIFO acquiring interval on the recorder is set to 1 s and the
measured data from channel 01 to 06 are continuously output (logging function)
Command to send
START
Command description
NO
Send command
FF RESET
Receive response
E0
Wait for the FIFO
acquiring interval (1 s)
Send command
FF GET,01,06
Receive response
BINARY header
BINARY data
BINARY footer
Stop measurement?
YES
END
Received response data
Move the FIFO read position to the most recent acquire position.
Output the measured/computed data using the FIFO.
BINARY (see page 5-2 and 5-13)
App
Appendix
Note
• The FIFO acquiring interval must be set using the FR command beforehand.
• The FIFO acquiring interval applies to both serial and Ethernet communications.
App-3
Appendix 3Login Procedure
When using the Setting/Measurement server or the Maintenance/Test server via the
Ethernet interface (/C7 option), you must log into the recorder from the PC. If you
complete the procedure successfully up to login complete in the following figure, the
commands in chapter 4 become functional.
When using the login function
Connect
No
E1 421
Within the number of
connections
†
Disconnect
E1 422
Disconnect
1
E1 420
E1 400
No
Within 2
minutes?
User name
Entered
"quit"?
E1 401
Within 2
minutes?
Password
Verification
match?
Within the number of
simultaneous uses?
E0
Yes
No
1
NoNo
†††
Increment retry count
E1 403
Wait 5 s
Within 3
retries?
No
††
1
E1 404
Within 3
retries?
††
1
Disconnect
E1 403
†
Connections cannot exceed the maximum number of connections (see section 2.1).
††
If you attempt to log in using a wrong password four consecutive times, the communication
is dropped (the number of retries for login is three).
†††
If you attempt to log in causing the number of simultaneous uses at the administrator or
user level to be exceeded (see section 2.1) four consecutive times, the communication is
dropped (even if the password is correct).
App-4
Login complete
StartJudgment
End
User entry
Response from the µR (message omitted)
For a description of the response format, see section 5.1.
Code (for a description of codes and messages, see page App-6)
Transition
When not using the login function
Login as “admin” or “user.”
• The user name “admin” can be used to login to the recorder as an administrator.
• The user name “user” can be used to access the recorder as a user.
No
E1 421
Connect
Within the number of
connections?
Appendix 3 Login Procedure
Disconnect
E1 422
Disconnect
1
E1 420
E1 402
No
Within 2
minutes?
User name
Entered
"quit"?
Verification
match?
Within the number of
simultaneous uses?
E0
Yes
No
1
NoNo
Increment retry count
E1 403
Within 3
retries?
No
1
E1 404
Within 3
retries?
App
Appendix
1
Disconnect
Login complete
App-5
Appendix 4A List of Error Messages
There are cases in which error codes and messages are displayed on the screen during
operation. The error messages and their description are listed below.
Setting Errors
CodeMessageExplanation/Countermeasures
1System error.Contact OMEGA.
2Incorrect date or time setting.Check the setting.
3A disabled channel is selected.The channel does not exist.
4Incorrect function parameter.Incorrect communication parameter.
5The input numerical value exceeds the set range.–
6Incorrect input character string.The entered character cannot be used.
7Too many characters.–
8Incorrect input mode.Incorrect range mode (Volt, TC, Scale, etc.) setting.
9Incorrect input range code.Incorrect range type (2 V, R, PT100, etc.) setting.
10Format error.Incorrect character string format.
11Range settings are not same within the selectedChannels whose range differs cannot be set simultaneously.
12An invalid characters.Contains an invalid character.
13Ref. CH error.Specify a channel whose range is set to voltage, TC, or RTD for
21Cannot set an alarm for a SKIPPED channel.–
22The upper and lower span limits are equal.This is not allowed.
23The upper and lower scale limits are equal.This is not allowed.
24The lower span limit is greater than the upper span–
25The lower scale limit is greater than the upper scale –
26Bias cannot be set to the SKIPPED channel.–
27Bias cannot be set to the DI channel.–
30The partial boundary value exceeds the range of the –
31Partial is invalid on the SKIPPED channel.–
35The upper and lower limits of the printing zoneSet the rightmost value of the zone – the leftmost value ≥ 5
36The lower limit of the printing zone is greaterSet the rightmost value of the zone – the leftmost value ≥ 5
37The printing zone is narrower than the minimumSet the rightmost value of the zone – the leftmost value ≥ 5
38Partial is invalid on the DI channel.–
47All items in DISP menu parameters are set to SKIP. –
48Start = Finish.The DST start time and end time cannot be set to the same time.
49Invalid or missing DST time settings.Since the time gains one hour when the DST starts, the set-up
61There is no channel specified by the MATHSet a computation channel.
62MATH expression grammar is incorrect.Enter the equation correctly.
63MATH expression sequence is incorrect.Enter the equation correctly.
64MATH upper and lower span values are equal.This is not allowed.
70MATH constant description is incorrect.Incorrect computation constant syntax.
71The range of the MATH constant is exceeded.–
72MATH channel is turned off–
App-6
channels.
the reference channel.
limit.
limit.
span.
are equal.mm.
than the upper limit.mm.
width (5 mm).mm.
time does not exist.
expression.
Appendix 4 A List of Error Messages
CodeMessageExplanation/Countermeasures
81All space or ‘quit’ string cannot be specified.–
86The key-lock release password is incorrect.Enter the correct password.
87This key is locked.–
91Password is incorrect.Enter the correct password.
100IP address doesn’t belong to class A, B, or C.–
101The result of the masked IP address is all 0s or 1s.–
102SUBNET mask is incorrect.Set a correct subnet mask.
103The net part of default gateway is not equal toSet the correct default gateway.
151This action is invalid during calculation.–
160This action is invalid during chart end.–
161This action is invalid during pen hold.–
162Cannot set an number for a skipped data.–
163This action is invalid during record.–
164This action is invalid during manual printing.–
165This action is invalid during list printing.–
166This action is invalid during setup list printing.–
167This action is invalid during chart feed.–
169This action is invalid during ribbon hold.–
that of IP address.
Operation Errors
CodeMessageExplanation/Countermeasures
232There is no available data.There is no data for periodic printout or data for calculating TLOG
when the timer expired.
Communication Errors
The messages from 390 to 422 can only be returned via communication, and are not displayed on the recorder.
CodeMessageExplanation/Countermeasures
300Command is too long.–
301Too many number of commands delimited with ’;’.Keep the number of commands separated by sub delimiters
under 10.
302This command has not been defined.–
303Data request command can not be enumerated with –
sub-delimiter.
350Command is not permitted to the current user level. –
351This command cannot be specified in the current–
mode.
352The option is not installed.–
353This command cannot be specified in the current–
setting.
354This command is not available during calculation.–
390Command error.–
391Delimiter error.–
392Parameter error.–
393No permission.–
394No such connection.–
395Use “quit” to close this connection.Attempted to disconnect its own connection.
396Failed to disconnect.–
397No TCP control block.The control block of the specified connection cannot be found.
App
Appendix
App-7
Appendix 4 A List of Error Messages
CodeMessageExplanation/Countermeasures
400Input username.–
401Input password.–
402Select username from ‘admin’ or ‘user’.If the recorder is configured not to use the user name and
403Login incorrect, try again!–
404No more login at the specified level is acceptable.–
420Connection has been lost.–
421The number of simultaneous connection has been–
exceeded.
422Communication has timed-out.–
password, use user names ‘admin’ or ‘user’.
Warning Messages
CodeMessageExplanation/Countermeasures
600Initialized.Settings and measured data have been initialized.
System Errors
CodeMessageExplanation/Countermeasures
902RAM failure.Contact OMEGA.
910A/D error.Contact OMEGA.
921A/D calibration value error.Contact OMEGA.
922A/D calibration is in the wrong order.Contact OMEGA.
types of responses ..............................................................5-1
UPage
unit .................................................................................... 4-12
unit information ................................................................. 4-27
unit information output ........................................................ 5-7
user ..................................................................................... 2-9
user information ................................................................ 4-28
user information output ..................................................... 5-12
user level ..................................................................... 2-9, 4-3
user name ........................................................................... 2-9
ZPage
zone recording .................................................................. 4-13
Index
Index
Index-3
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a
period of
grace period to the normal
ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service
Department will issue an Authorized Return (AR) number immediately upon phone or written request.
Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no
charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits,
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a result of excessive corrosion;
or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating
conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any
damages that result from the use of its products in accordance with information provided by
OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the
company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of
OMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of the
component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or
activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility
as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the
Product(s) in such a manner.
13 months fr
om date of purchase. OMEGA’s WARRANTY adds an additional one (1) month
one (1) year product warranty to cover handling and shipping time. This
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE
RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN
(AR) NUMBER FROM OMEGA
’S CUSTOMER SER
VICE DEP
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return
package and on any correspondence.
The pur
chaser is r
esponsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
FOR
following information available BEFORE
contacting OMEGA:
1. Purchase Order number under which the prod-
2.
3. Repair instructions and/or specific problems
ARRANTY
W
RETURNS, please have the
uct was PURCHASED,
Model and serial number of the pr
warranty, and
oduct under
FOR NON-WARRANTY REPAIRS,
for current repair charges. Have the following
information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST
of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems
elative to the pr
r
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our
customers the latest in technology and engineering.
duced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior
written consent of OMEGA ENGINEERING, INC.