Omega M4502, 707 User Manual

/
HH806U/HH806AU
MULTILOGGER THERMOMETER
INSTRUCTION
SHEET
Shop online at: omega.com e-mail: info@omega.com
For latest product manuals: omegamanual.info
M4502
0707
OMEGAnet® On-Line Service
U.S.A: One Omega Drive, Box 4047
Stamford, CT 06907-0047
ISO 9001 Tel: (203) 359-1660 FAX: (203) 359-7700 Certified
e-mail: info@omega.com
Canada: 976 Bergar Laval (Quebec) H7L 5A1, Canada Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
U.S.A Sales Service: 1-800-826-6342/1-800-TC-OMEGA and Customer Service: 1-800-622-2378/1-800-622-BEST Canada:
Mexico: En Espanol: (001) 203-359-7803
e-mail: espanol@omega.com FAX: (001) 203-359-7807 info@omega.com.mx
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands Tel: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: sales@omegaeng.nl
Czech Frystatska 184, 733 01 Karviná, Czech Republic Republic: Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114
Toll Free: 0800-1-66342 e-mail: info@omegashop.cz France: 11, rue Jacques Cartier, 78280 Guyancourt, France
Tel: +33 (0)1 61 37 2900 FAX: +33 (0)1 30 57 5427 Toll Free in France: 0800 466 342 e-mail: sales@omega.fr
Germany/ Daimlerstrasse 26, D-75392 Deckenpfronn, Germany Austria: Tel: + 49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29
Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
United One Omega Drive, River Bend Technology Centre Kingdom: Northbank, Irlam, Manchester
ISO 9002 M44 5BD United Kingdom Certified
Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
Engineering Service: 1-800-872-9436/1-800-USA-WHEN® TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622
Servicing North America:
Servicing Europe:
Internet e-mail
info@omega.com
INTRODUCTION
The instrument is a portable digital thermometer that measures external thermocouples of type K, J, R, S, T, E, N. The thermocouples types comply with the N.I.S.T. – ITS 90 standard reference temperature/voltage tables. The thermometer features a dual thermocouple input, an adjustable T/C offset and an USB interface for uploading data to a PC using optional software and cable.
SAFETY INFORMATION
To avoid electrical shock, do not use this instrument
when working voltages at the measurement surface
®
®
To avoid damage or burns, do not make temperature
measurement in microwave ovens.
Repeated sharp flexing can break the thermocouple
leads. To prolong lead life, avoid sharp bends in the
leads, especially near the connector.
WARNING
over 24V AC or DC.
WARNING
CAUTION
SPECIFICATIONS
ELECTRICAL
Temperature Scale: Measurement Range:
K-TYPE (0.1°) -200°C to 1372°C (-328°F to 2501°F) J-TYPE (0.1°) -210°C to 1200°C ( -346°F to 2192°F) T-TYPE (0.1°) -200°C to 400°C (-328°F to 752°F) E-TYPE (0.1°) -210°C to 1000°C (-346°F to 1832°F) R-TYPE (1°) 0°C to 1767°C (32°F to 3212°F)
Celsius or Fahrenheit user-selectable
Input Protection:
24V dc or 24V ac rms maximum input voltage on any combination of input pins.
Reading Rate: 2.5 time per second.
ENVIRONMENTAL
Ambient Operating Ranges:
0°C to 50°C (32°F to 122°F) <80% R.H.
Storage Temperature:
-20°C to 60°C (-4°F to 140°F) <70% R.H.
GENERAL
Display:
There are three display areas on the HH806 series LCD (liquid crystal display). The Main and Second displays are 4 ½ digits with a maximum reading of 19999. These are used for displaying the value of T1 or T2. The Third display is for the date, time, or the differential value of T1 to T2.
Overload: “----.-” or “OL” is display. Battery: 1.5V x 4 PCS (SIZE AAA) UM-4 R03. Battery Life: 120 hours typical with carbon zinc battery. Dimensions: 160mm(H) x 83mm(W) x 38mm(D). Weight: Approx. 365g including batteries. Supplied Thermocouple:
1 meter (40”) type K insulated beaded wire thermocouple. Maximum insulation temperature is 482°C (900°F). Thermocouple accuracy is ±1.1°C or 0.4% of reading (whichever is greater) from 0°C to 1250°C. Wire Communication Protocol: 2400 baud rate. (HH806U) 19200 baud rate. (HH806A)
External Connections:
1. USB Port
2. DC power JACK(12V)
1
2
S-TYPE (1°) 0°C to 1767°C ( 32°F to 3212°F) N-TYPE (0.1°) -50°C to 1300°C ( -58°F to 2372°F)
*Based on the ITS-90 temperature standard.
According to temperature standard ITS-90.
Accuracy:
K/J/T/E-TYPE
±(0.05% rdg + 0.3°C) on -50°C to 1372°C
±(0.05% rdg + 0.7°C) on -50°C to -210°C
±(0.05% rdg + 0.6°F) on -58°F to 2501°F
±(0.05% rdg + 1.4°F) on -58°F to -346°F N-TYPE
±(0.05% rdg + 0.8°C) on -50°C to 0°C
±(0.05% rdg + 0.4°C) on 0°C to 1300°C
±(0.05% rdg + 1.6°F) on -58°F to 32°F
±(0.05% rdg + 0.8°F) on 32°F to 2372°F R/S-TYPE
±(0.05% rdg + 2°C) on 0°C to 1767°C
±(0.05% rdg + 4°F) on 32°F to 3212°F
Temperature Coefficient:
0.1 times the applicable accuracy specification per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to122°F).
OPERATING INSTRUCTIONS
1. OPERATIONAL MODE
There are three operation modes-Normal, Shift, and Setup Mode. NORMAL MODE:
This is the default mode, the operating functions for the normal mode are printed on the top of each button in white.
SHIFT MODE:
The operating functions for the shift mode are printed in gray on the buttons. While in the normal mode, press the SHIFT button to switch to shift mode. At the lower-right corner of the display, the word “Shift” is displayed to indicate shift mode. To switch back to normal mode, press the SHIFT button again.
SETUP MODE:
Press the set[ ] button in normal mode to switch to setup mode, the indicator “SET” is shown on the left side of the display. To switch back to normal mode, press SET[ ] button.
2. NORMAL MODE
The following functions can only be used in the normal mode.
(1) “ ” Power Button
The “ ” button turns the thermometer on or off. When the meter is in MAX/MIN record mode, the power off function is disabled.
(2) “[LIMITS]” Button (only Main display)
The limits function will alert the user when a meas­urement exceeds a specified limit. To set the limit val­ues, refer to limits function in the setup mode. Press the [LIMITS] button to activate the limits function; the word “LIMIT” should be displayed on the LCD. When the value of the main display exceeds the Hi limit, the word “Hi” will be displayed and the ther­mometer will beep in a pulsed tone. If the value of the main display is lower then the Lo limit, the word “Lo” will be displayed and the thermometer will beep in a continuous tone. To exit the limits function, press the [LIMITS] button.
(3) “ ” Button
The backlight function is represented by this button
”. Pressing the button will turn on or off the LCD
“ backlight. The backlight will turn off automatically af­ter.
(4) “SAVE/READ” Button
The read data function works in conjunction with the save function in the shift mode, it is used for reading saved data. The save function can be activated in shift mode. Press the SAVE/READ button to read saved data; the word “READ” should be displayed on the LCD. To navigate around the save data table, press the overlay “SECOND” button until the “#” sign is dis­played on the second display. The location of the read pointer within the saved data table will be displayed. The arrow buttons on the overlay are used for scrolling through the saved data. Press the smaller arrows “
or “
” to step through the data one at a time. Press the larger arrows “ ten at a time. Pressing the overlay “ESC” button deac­tivates the read data function.
” or “ ” to step through the data
(5) “LOG/READ” Button
The read log function works in conjunction with the log function. It is used for reading logged data. The log function can be activated in the shift mode. Press the LOG/READ button to activate the log read function; the word “READ” is displayed on the LCD. Press the overlay SECOND button to rotate through following display menus: T1, T2, GRP, and #. T1 and T2: Dis­plays the T1 or T2 saved data. GRP: Displays the cur­rent group number. #: Displays the current location of the read pointer within a selected group. The arrow buttons on the overlay are used for scrolling through the data and groups. Press the smaller arrows “
” to step through the logged data or groups one at a
“ time. Press the larger arrows “ through the data or groups ten at a time. To navigate the logged data and groups, press the overlay SECOND button until GRP appears in the second dis­play panel. Then select the group using the arrows. Press the SECOND button again until the “#” sign is displayed. The location of the read pointer in the se­lected group will be displayed. Use the arrows to scroll through the data. Pressing the overlay “ESC” button deactivates the read data function.
” or “ ” to step
(7) HOLD Mode (only Main display)
When HOLD mode is selected, the thermometer holds the present readings and stops all further measure­ments. To activate the data hold mode, press the HOLD button, and “HOLD” is displayed on the LCD. Pressing the HOLD button again cancels the function, and the instrument will automatically resume meas­urements.
(8) MIN/MAX with Time record Mode
(only Main display)
The MIN/MAX function records the highest and low­est value recorded, and it calculates the average read­ing, and the differences of MAX to MIN. Press MIN/MAX button to enter the MIN/MAX recording mode. The beeper emits a tone when a new minimum or maximum measurement is recorded. Press the MIN/MAX button again to rotate through the current readings: MAX: The highest measurement recorded. MIN: The lowest measurement recorded. MAX-MIN: The difference of the highest and the lowest measure­ment. AVG: The average values of the measurements. This mode works in conjunction with the hold function, pressing the HOLD button will stop the recording and measurements (Previously recorded readings are not erased). Press HOLD button again to resume recording and measurements. To prevent accidental loss of MIN, MAX and AVG data, the MIN/MAX function can only be cancelled by pressing and holding down the MIN MAX key for more then 2 seconds. The automatic power-off feature, and the power, °C/°F, REL, SET, Hi/Lo Limits, TYPE, T1/T2 buttons are also disabled.
” or
(10) “T1/T2” Button (Main display)
The input selection button [T1/T2] selects the input for the main display, T1 thermocouple or T2 thermocou­ple. Press the T1/T2 button to switch between the two inputs. When meter is turned on, it is set to the display that was last in use.
(11) “T1/T2” Button (Second display)
The input selection button [T1/T2] selects the input for the second display, T1 thermocouple or T2 thermo­couple. Press the T1/T2 button to switch between the two inputs. When meter is turned on, it is set to the display that was last in use.
(12) “T1-T2/TIME” Button (Third display)
The input selection button [T1-T2] selects the system time and date, or the differential between the two ther­mocouples (T1-T2) for the third display. Press the T1-T2 button to switch the display options. When me­ter is turned on, it is set to the display that was last in use.
3. SHIFT MODE
The following functions can only be used in the shift mode.
(2) “°C/°F” Button
Press the °C/°F button to select the temperature scale, readings can be displayed in Celsius (°C) or Fahrenheit (°F). When the thermometer is turned on, it is set to the temperature scale that was last in use.
(4) “SAVE” Button
The save function stores the T1, T2 data in a nonvola­tile memory. Press the SAVE button to save the cur­rent data, the word SAVE is displayed to indicate the data has been saved. The built in memory can store up to 128 data (HH806U)/256 data (HH806AU). The data can be read using the read function in the normal mode.
(5) “LOG” Button (HH806U)
The data log function continuously records the data according to a specified time interval. The time inter­val can be set using the interval setup function [INVT] in the set up mode. Press the LOG button to activate the log function; the indicators “LOG” and “MEM” will be displayed on the LCD. There are 16 groups that are used for storing the log data, and each group uses 64 data slots. If the current log session exceeds 64 data, the log function will automatically use the next group to store the following data. A maximum of 1024 data point can be saved in one log session. Press the LOG button again to exit the data log function.
“LOG” Button (HH806AU)
The data log function continuously records the data according to a specified time interval. The time inter­val can be set using the interval setup function [INVT] in the set up mode. Press the LOG button to activate the log function; the indicators “LOG” and “MEM” will be displayed on the LCD. There are 16 groups that are used for storing the log data. A maximum of 16,000 data point can be saved in one log session. Pressing the LOG button again to exit the data log
function.
(6) “CLR ?” Button
The CLR function clears all the saved and logged data in memory. When the CLR button is pressed, indicator “MEM” is displayed and the “CLR” on upper-right of LCD will blink. Pressing the “ENTER” button printed on the overlay in white will clear all saved and logged data. Press “ESC” button to exit this function without clearing data.
(7) “REL” Button (Main display)
The relative value function can be used for comparing the saved reference value with other measurements. Press the “REL” button to store the current measure­ment as the reference value, and “REL” should be dis­played on the right part of the LCD. The next meas­urement will display the current value compared to the reference value. Press “REL” button again to clear the reference value and deactivate the relative value meas­urement function.
(8) “APO” Button
Press the APO button to turn the “Auto power off” function on or off. When this function is enabled, the indicator “APO” is shown at the upper left part of the LCD. When APO (Auto power off) is enabled, it will automatically turn the thermometer off no button is pressed for a period longer than the set time interval (the default time for APO is 5 minutes). Press power button to resume operation.
(10) “TYPE” Button (Main display)
Press this button to change the type of thermocouple in the main display (K/J/T/E/R/S/N). If the inputs of the main and second display are the same, then pressing this button will change the thermocouple type for both displays.
(11) “TYPE” Button (Second display)
Press this button to change the type of thermocouple in the second display (K/J/T/E/R/S/N). If the inputs of the main and second display are the same, then pressing this button will change the thermocouple type for both displays.
4. SETUP MODE
The following functions can only be use in the setup mode.
(2)
“[LIMITS]” Button (Hi/Lo limit setting)
Press the LIMITS button to enter the Hi/Lo limit setup function. The words “LIMIT” and “Hi” will be flash on the LCD along with the previous value for the Hi limit. Enter the new Hi limit value using the number keys printed in white on the overlay. The resolution of Hi/Lo limit setup is 1 count. The “-” button (same button as the ESC) can be used to enter negative val­ues. Press the “ENTER” button to confirm the new limit. You will then be prompted to enter a new Lo limit value. Enter the new value and then press enter to finish setup of limits.
(3) “[INVT]”
To setup the time interval for the log function, press the [INVT] button. The indicator “INV” will blink on
Button
(Interval time setting)