Omega M1982-0902 User Manual

VA Series
Vapor Actuated Dial Thermometer
INSTRUCTION
SHEET
M1982/0902
General Description
The OMEGA®VA Series Vapor Actuated Thermometer is ideal for remote readings, such as panel installations, with capillary lengths from 5 to 100 feet. This thermometer gives excellent readings above and below ambient temperature and, being vapor actuated, is not subject to reading errors due to either RFI/EMI interference or ambient temperature variations along the capillary tube length.
Available Models
Front Flange U-Clamp (Panel Mount) Rear Flange
(Panel Mount) (Surface Mounted)
Figure 1. Case Style
Installing the Vapor Actuated Thermometer
Unpacking Instructions
Remove the Packing List and verify that you have received all equipment, including the following (quantities in parentheses):
Vapor Actuated Dial Thermometer (1) Operator’s Manual (1)
If you have any questions about the shipment, please call the OMEGA Customer Service Department.
When you receive the shipment, inspect the container and equipment for signs of damage. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
VA Series Vapor Actuated Dial Thermometer
The carrier will not honor damage claims unless all shipping material is saved for inspection. After examining and removing contents, save packing material and carton in the event reshipment is necessary.
Case and Mounting Dimensions
Front Flange U-Clamp Rear Flange
(Panel Mount) (Panel Mount) (Surface Mount)
Figure 2. Case and Mounting Dimensions
Installing the Indicating Head
Surface Mount (Rear Flange)
1. Make a circular cutout in the center of the hole mounting pattern for the capillary tube to fit through.
2. Drill the three (3) mounting holes.
3. Using the screw holes that you drilled, screw the back of the head to your wall or panel.
NOTE
2
VA Series Vapor Actuated Dial Thermometer
Flush Mount (Front Flange and U-Clamp)
1. Make a circular cutout in the panel slightly larger than the case.
2. Drill three (3) mounting holes for the Front Flange Model.
3. From the front of the panel, pass the bulb and capillary through the
cutout.
4. Affix the head to the panel:
4a. U-clamp Model: Tighten U-clamp(s) against the back of your
panel.
4b. Front Flange Model: Screw the flange to the front of your panel,
using the screw holes that you drilled.
Installing the Bulb
If the bulb is installed more than 10 feet above or below the dial, it may be necessary to recalibrate the instrument.
NOTE
3
Loading...
+ 5 hidden pages