INTRODUCTION
The Model J DialtempTMis a non electronic thermometer that features a 360 degree swivel neck with a 90 degree tilt range of it’s large easy-to-read 5 inch dial face. The heliarc stainless steel welded construction provides a moisture-proof bond, while preventing the possibility of rust and corrosion. The temperature sensing device consists of sensitive bi-metal coils. They are wound to pre-determined standards and are individually trimmed and tested for full-scale accuracy, contributing minimum pointer vibration and maximum heat transfer. One external reset screw enables fast and simple recalibration. This superior DialTemp instrumentation design combines high precision with durability and dependability, making the Model J Thermometer ideal for a wide variety of applications.
UNPACKING INSTRUCTIONS
Remove the packing list and verify that all equipment has been received. If there are any questions about the shipment, please call the OMEGA Customer Service Department at 1-800-622-2378 or (203) 359-1660.
Upon receipt of shipment, inspect the container and equipment for signs of damage. Take particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material in event reshipment is necessary.
HEAD ADJUSTMENT INSTRUCTIONS
Adjust the angle of the head before installing the Model J Thermometer. Refer to Figure 1 to locate the adjustment screws.
Install the thermometer so that the maximum ambient head temperature remains below 300°F (149°C) at all times.
1. Adjust the rotational position of the head by loosening the two bracket bolts and rotating the bracket until the head is in the desired position.
NOTE
NOTE
NOTE: Observe that the head and stem must be in the full upright position in order to rotate the head. Tighten the bracket bolts securely.
2. Adjust the tilt angle of the head by loosening the two hinge screws and tilting the head to the desired position. Tighten the hinge screws securely.
TO RESET THE DIAL
1. Determine the room temperature by using a reference thermometer of known accuracy.
2. Locate the reset screw on the back of the head (see Figure 1).
3. The numbered dial face inside the head will rotate as the screw is turned. Use a slotted screwdriver or a
1
4" hex wrench to adjust the setting.
4. Align the pointer with the correct temperature on the numbered dial face.
The thermometer is now ready for installation.
INSTALLATION INSTRUCTIONS
The Model J DialTemp Thermometer is supplied with a 1⁄2" -14 NPT pipe thread for mounting purposes. Carefully screw the thermometer into the fitting by hand until slightly tight. Complete tightening by using a
5
8 " wrench
on the wrench flats. When using a thermowell, tighten only FINGER TIGHT.
NEVER TIGHTEN BY TWISTING THE HEAD! AVOID HANDLING THE THERMOMETER BY THE STEM OR BENDING THE STEM.
To obtain a positive seal, coat the threads with pipe joint compound or wrap with Teflon tape.
When inserting the DialTemp into a semi-solid material, hold the stem firmly and insert slowly and carefully to avoid bending. When inserting the thermometer into a material, always hold it by the stem, not by the head.
OPERATING INSTRUCTIONS
For correct temperature measurement, the stem of the DialTemp must be immersed at least 2" in fluids and 4" in gases.
RANGES AND DIVISIONS:
Fahrenheit Celsius
-40-160°F 2 0-100°C 1 *0-200°F 2 -10-110°C 1 *0-250°F 2 0-150°C 1 50-300°F 2 0-300°C 5
*50-500°F 5 150-750°F 5 100-800°F 5
200-1000°F 10
* Off-the-shelf ranges/divisions for the 2
1
2" and 6" stem lengths. Not
recommended for continuous use above 800°F.
SPECIFICATIONS
ACCURACY: ±1% of full scale RESPONSE TIME: Readings are stabilized within 40 seconds