OMEGA ITHX-SD-2 User guide

Page 1
TM
Users Guide
Shop online at
omega.com
e-mail: info@omega.com
For latest product manuals:
www.omegamanual.info
Temperature + Humidity
Page 2
Page 3
TABLE OF CONTENTS
Part 1: INTRODUCTION
1.1 Safety and EMC Considerations .....................................................................2
1.2 Before You Begin .............................................................................................2
1.3 Description .......................................................................................................2
Part 2: HARDWARE
2.1 Dimensions ......................................................................................................5
2.2 Wall Mounting ..................................................................................................6
2.3 Parts of the iServer Unit .............................................................................. 7
2.4 Disassembly Instruction ............................................................................. 8
2.5 Network Communication Interfaces .......................................................... 9
2.5.1 10/100 BASE-T RJ-45 Pinout .......................................................... 9
2.5.2 Connecting iServer to PC/Hub/Switch/Router .............................. 9
2.6 Relay Wiring Connections .......................................................................... 9
PART 3 NETWORK CONFIGURATION
3.1 Network Protocols ..................................................................................... 10
3.2 Ethernet (MAC) Address ........................................................................... 10
3.3 DHCP .......................................................................................................... 11
3.4 DNS .......................................................................................................... 11
3.5 IP Address .................................................................................................. 11
3.6 Port Number ............................................................................................... 12
PART 4 OPERATIONS ......................................................................................................
4.1 IP Address Assignment ............................................................................ 12
4.1.1
4.1.2
4.1.3 Direct Connection (Method 3) ...................................................... 13
4.1.4 iConnect Software (Method 4) ..................................................... 17
4.2 Access and Configuration Using a Web Browser .................................. 19
4.3 LOGIN and ADMINISTRATOR Passwords ............................................... 19
4.4 Setup .......................................................................................................... 20
4.4.1 Overview ........................................................................................ 20
4.4.2 Network .......................................................................................... 21
4.4.2.1 IP Configuration ...................................................................... 21
4.4.2.2 Ethernet Port ........................................................................... 23
4.4.3 Configuration ................................................................................. 23
4.4.3.1 Data and Time ......................................................................... 23
4.4.3.3 Sensors ................................................................................... 25
4.4.3.4 Alarm Relays ........................................................................... 27
4.4.4 Management .................................................................................. 29
4.4.4.1 Setup ....................................................................................... 29
SNMP Simple Network Management Protocol ................................ 29
SMTP Simple Mail Transfer Protocol ............................................... 30
4.4.4.1.1 Sending Text Messages to a Cell Phone ....................... 30
4.4.4.2 Alarms ..................................................................................... 30
DHCP (Method 1) ARP HTTPget Commands
........................................................................... 12
(Method 2) ......................................... 13
4.4.3.2 Server ...................................................................................... 25
i
Page 4
TABLE OF CONTENTS (continued)
4.4.5 Security .......................................................................................... 32
4.4.6 Recording ...................................................................................... 34
4.4.6.1 Start Recording ....................................................................... 34
4.4.6.2 Status ....................................................................................... 36
4.4.6.3 Data Retrieval .......................................................................... 37
4.4.6.4 Format SD Card ...................................................................... 38
4.4.7 System ........................................................................................... 38
4.4.7.1 Reboot ..................................................................................... 38
4.4.7.2 Defaults ................................................................................... 39
4.4.7.3 Upgrade ................................................................................... 39
4.4.7.4 Download Configuration ........................................................ 40
4.4.7.5 Upload Configuration ............................................................. 41
4.4.8 Diagnostics .................................................................................... 41
4.4.9 Log Out .......................................................................................... 41
4.5 Readings .................................................................................................... 42
4.5.1 HTML .............................................................................................. 42
4.5.2 Java ................................................................................................ 43
4.6 Chart .......................................................................................................... 44
4.7 Web Link ..................................................................................................... 45
4.8 Telnet Setup ............................................................................................... 46
4.8.1 Telnet Connection ......................................................................... 46
4.9 HTTPget Program ...................................................................................... 47
4.9.1 HTTPget using Port 2000 .............................................................. 48
4.10 Flash Card Reader ................................................................................... 50
4.11 iLog Software ........................................................................................... 52
PART 5 SPECIFICATIONS ........................................................................................... 54
APPENDIX A GLOSSARY ........................................................................................... 56
APPENDIX B IP Address ............................................................................................. 57
APPENDIX C ARP Commands ................................................................................... 58
APPENDIX D IP Netmask ............................................................................................ 59
APPENDIX E ASCII Chart ............................................................................................ 60
APPENDIX E ASCII Control Codes ............................................................................ 61
APPENDIX F iLog Error Messages ............................................................................ 62
APPENDIX G Java Runtime Environment Setup ....................................................... 63
APPENDIX H Sensor Information ............................................................................... 65
APPENDIX J SNMP ...................................................................................................... 67
Part 6: APPROVALS INFORMATION
6.1 CE APPROVALS ............................................................................................68
ii
Page 5
LIST OF FIGURES:
Figure 1.1 iServer and iLD Big Display on the Ethernet Network .............................. 4
Figure 2.1 Dimensions ................................................................................................... 5
Figure 2.2 Wall Mounting ............................................................................................... 6
Figure 2.3 Parts of the iServer Unit ............................................................................... 7
Figure 2.4 Opening the Unit ........................................................................................... 8
Figure 2.5 RJ45 Pinout ................................................................................................... 9
Figure 2.6 Relay Connections ....................................................................................... 9
Figure 3.1 Labeling .................................................................................................. 10
Figure 4.1 ARP and HTTPget Commands on a DOS Window .................................. 13
Figure 4.2 Connecting Computer Directly to iServer ................................................ 14
Figure 4.3 Network Connections ................................................................................. 14
Figure 4.4 Local Area Connection .............................................................................. 14
Figure 4.5 Changing TCP/IP Properties on Your Computer ..................................... 15
Figure 4.6 iServer Welcome Page ............................................................................... 16
Figure 4.7 Pinging the iServer from MS-DOS Prompt ............................................... 16
Figure 4.8 Assigning an IP Address using iConnect ................................................. 17
Figure 4.9 Accessing the iServer’s using iConnect .................................................. 18
Figure 4.10 iServer Welcome Page ............................................................................. 19
Figure 4.11 LOGIN and ADMINISTRATOR Passwords .............................................. 19
Figure 4.12 Overview .................................................................................................. 20
Figure 4.13 Network: IP Configuration ...................................................................... 21
Figure 4.14 Network: Ethernet Configuration ............................................................ 23
Figure 4.15 Network: Ethernet Configuration ............................................................ 24
Figure 4.16 Configuration Menu: Server .................................................................... 25
Figure 4.17 Configuration Menu: Sensors ................................................................. 26
Figure 4.18 Configuration Menu: Alarm Relays ......................................................... 28
Figure 4.19 Management Menu: Setup ....................................................................... 29
Figure 4.20 Management Menu: Alarms ..................................................................... 30
Figure 4.21 Security Menu ........................................................................................... 32
Figure 4.22 Recording Menu: Start Recording .......................................................... 34
Figure 4.23 Recording Menu: Status .......................................................................... 36
Figure 4.24 Recording Menu: Data Retrieval ............................................................. 37
Figure 4.25 Recording Menu: Format SD Card .......................................................... 38
Figure 4.26 System: Reboot ........................................................................................ 38
Figure 4.27 System: Defaults ...................................................................................... 39
Figure 4.28 System: Upgrade ...................................................................................... 39
Figure 4.29 System: Download Configuration ........................................................... 40
Figure 4.30 System: Upload Configuration ................................................................ 40
Figure 4.31 Diagnostics ............................................................................................... 41
Figure 4.32 Readings: HTML ....................................................................................... 42
Figure 4.33 Readings: Java ......................................................................................... 43
Figure 4.34 Adjustable Chart ....................................................................................... 44
Figure 4.35 Web Link .................................................................................................. 45
Figure 4.36 Tera Term Telnet Connection Screen ..................................................... 46
Figure 4.37 Telnet to port 2002 – p command ........................................................... 47
Figure 4.38 Telnet to port 2002 – ? command ........................................................... 47
Figure 4.39 Recorded Data File Directory .................................................................. 53
Figure 4.40 Example of data recorded on the SD card, standard text (.txt) file format 53
iii
Page 6
LIST OF FIGURES (continued)
Figure 4.41 iLog Software Logging Data .................................................................... 54
Figure C.1 ARP Commands and Responses ............................................................. 58
Figure G.1 Java 1.7 Screen Shot ................................................................................. 63
Figure H.1 RH Accuracy Chart .................................................................................... 65
Figure H.2 Temperature Accuracy Chart .................................................................... 65
Figure H.3 Normal Range ............................................................................................. 65
LIST OF TABLES
Table 2.1 Parts of iServer Unit ....................................................................................... 7
Table 4.1 iLog Excel Applications ............................................................................... 53
Table F-1 iLog Error Messages ................................................................................... 62
Table J-1 SNMP MIB-2 .................................................................................................. 67
Table J-2 SNMP Trap .................................................................................................. 67
iv
Page 7
NOTES, WARNINGS and CAUTIONS
Information that is especially important to note is identified by the following labels:
W*+0! W3ARNING or CAUTION W%),+.0ANT W0%,
NOTE: Provides you with information that is important to successfully setup and use the iServer.
CAUTION: Tells you about the risk of electrical shock.
CAUTION: Risk of danger. Tells you of circumstances or practices
that can affect the instrument’s functionality and must refer to accompanying documents.
TIP: Provides you helpful hints.
FEATURES
W Temperature W Relative Humidity W Web Server W Dual Probe Capability W Virtual Chart Recorder W Two Relay Alarms W 2GB SD Flash Memory Card W Password Protection W Email Alarms W Data Logging W Real Time Clock W Accurate Readings W SNMP Trap W LCD Display W Back-up Battery 9Vdc
1
Page 8
PART 1 INTRODUCTION
1.1 Safety and EMC Considerations
Refer to the CE Approvals Section.
EMC Considerations
W Whenever EMC is an issue, always use shielded cables. W Never run signal and power wires in the same conduit. W Use twisted-pair wires for signal connections. W Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist.
Failure to follow all instructions and warnings may result in injury!
1.2 Before You Begin Inspecting Your Shipment: Remove the packing slip and verify that you have received
everything listed. Inspect the container and equipment for signs of damage as soon as you receive the shipment. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent. The carrier will not honor damage claims unless all shipping material is saved for inspection. After examining and removing the contents, save the packing material and carton in the event reshipment is necessary.
Customer Service: If you need assistance, please contact the Customer Service Department nearest you.
Manuals, Software: The latest Operation Manual as well as free configuration software (iConnect) and datalogging software (iLog) are available at the website listed on the
cover page of this manual or on the CD-ROM enclosed with your shipment.
1.3 Description View Temperature + Humidity with a Web Browser. The iServer let’s you monitor and
record Temperature, Relative Humidity and Dew Point over an Ethernet network or the Internet with no special software -- just a Web browser. The iServer serves Active Web Pages to display real time readings and display charts of temperature, humidity and dew point, or log data in standard data formats for use in a spreadsheet or data acquisition program such as Excel or Visual Basic.
Adjustable Charts. The Java™ Applet chart scales are fully adjustable on the fly. For example, the chart can display one day, one week, one month or one year. Temperature and humidity can be charted across the full span (-40 to 124ºC, and 0 to 100% RH) or within any narrow range (such as 20 to 30ºC).
When a second sensor is added, users can select a chart that records channel 1, channel 2 or the difference of the two channels.
SD Flash Memory Card. The iServer comes complete with a removable 2 GB SD Flash Memory card that can store up to seven years of readings taken at ten second intervals. The data is recorded on widely available SD (Secure Digital) flash cards. The format is a simple text file that is easily imported to spreadsheets and other programs. It can be read on a PC or MAC with a USB card reader. You can also download the data remotely over an Ethernet network or the Internet.
Installation and operation of the iServer requires no special training, tools, or software. The device connects to any Ethernet network with standard cable and is powered by a universal 100 to 240 Vac adapter which is supplied with the product.
2
Page 9
1.3 Description (continued) Back-up Battery. The iServer comes with a universal 100 to 240 Vac power adapter.
A standard 9 Volt Alkaline battery (also included) allows the device to continue recording data for 60 hours (at one minute interval setting) without external ac power (such as a power outage).
Display and Chart Two Channels. The transmitter comes complete with a temperature and humidity probe for measurement of a single location.
With the addition of a second probe, the transmitter can measure and display temperature, humidity and dew point in a second location up to ten feet away.
The transmitter can display and chart absolute measurements in both locations, or a differential measurement between the two locations. The second probe requires no change to the basic transmitter hardware, it can be added at the time of purchase or in the future. A choice of industrial probes in 2” and 5” lengths, and a wand style for ambient indoor applications are available. A simple DB-9 “Y” connector is required for adding a second probe.
Sensor and Calibration. It is not necessary to take the iTHX-SD out of service for routine calibration. The temperature/humidity sensors are interchangeable and can be replaced for about the same cost as a typical calibration. Replacement sensors with optional 3-point NIST traceable calibration certificates are available.
Email Alarms. The device can trigger an alarm if temperature or humidity goes above or below a set point that you determine. Your alarm can be sent by email to a single user or to a group distribution list, including text messages to cell phones and PDA’s.
Alarm Relays. The iServer features two 1.5 Amp relays. With the easy Web-based setup page, the two relays can be programmed for any combination of temperature or humidity, and high or low set points. The relays can also be programmed to remain latched and require a manual reset if a limit is exceeded.
Link to Web CAM or IP Camera. The Web page includes a link to a “Web Cam” or “IP camera” (not included). If you get a message about an alarm condition, you can quickly click on the link to view the actual scene over the Internet.
3
Page 10
1.3 Description (continued)
SMTP SNMP Server
Award-winning Technology. The iServer is simple to install and use. It features award winning technology that requires no special software except a Web browser. The iServer connects to an Ethernet Network with a standard RJ45 connector and sends data in standard TCP/IP packets. It is easily configured with a simple menu using a Web browser and can be password protected. From within an Ethernet LAN or over the Internet, the user simply types its IP address or an easy to remember name such as “Cleanroom5” or “ServerRoom” in any Web Browser, and the iServer serves a Web Page with the current readings. The iServer comes complete with one temperature/humidity wand probe, universal ac power adapter, 9V battery, SD card, and full documentation.
Example:
A standard web browser can be used to monitor and chart temperature and humidity. The browser can also be used to configure the device’s IP address, passwords for access and all configuration parameters. An iLD Big Display can display temperature and humidity received from an iServer over the Ethernet or the Internet.
The following example illustrates how you can hookup an iServer and iLD to your network:
COMPUTER with
Standard Web Browser
COL ON TX RX
and
Figure 1.1 iServer and iLD Big Display on the Ethernet Network
4
Page 11
PART 2 HARDWARE
2.54 [0.10]
2.1 Dimensions
270.4 [10.65]
12.7 [0.50 ]
82.6 [3.25]
19.1 [0.75]
61.6 [2.42]
54.1 [2.13]
88.9 [3.50]
If unit is to be mounted on a flat surface, you may take the bottom rubber feet off the unit.
!
3.0 [0.12]
36.6 [1.44]
90.3 [3.56]
Dimensions are in mm with inches in [ ].
Figure 2.1 Dimensions
5
Page 12
2.2 Wall Mounting
33.3 [1.31]
Position unit where required. Mark and drill the two #4 screw holes.
After bracket is mounted on the wall, align back of unit over the three bracket clips, once engaged, slide downward, the unit will snap in place.
61.6 [2.42]
38.1 [1.50] 11.7 [0.46 ]
REAR WIRE
ENTRY AREA
DRILL 3.17 [0.125]
USE TWO, #4
FLAT HEAD SCREWS
TO MOUNT BRACKET
SNAP
BRACKET CLIPS (3)
It is recommended that you ground your unit by connecting a wire to the Return/ Ground position of the relay connector, see Figure 2.6.
Dimensions are in mm with inches in [ ].
Figure 2.2 Wall Mounting
6
Page 13
2.3 Parts of the iServer Unit
1
The 9V battery is the back-up power for the recording function only.
1
8
Figure 2.3 Parts of the iServer Unit
Table 2.1 Parts of iServer Unit
ETHERNET RJ45 interface for 10/100BASE-T connection.
RESET Button: Momentary (Push and Release) resets power on unit; Push and Hold for
10 seconds to reset unit to factory defaults and reset power.
100 BASE-T LED (Green) On: Indicates 100 Mbps; LED Off: Indicates 10 Mbps.
LINK/ACT LED (Green) On/Blinking: Indicates good network link and network activities
DHCP LED (Yellow): When DHCP is enabled, once the iServer receives the IP
RECORDING LED (Green) Blinking Fast: Indicates that the unit is recording. LED Blinking:
°C/°F Button: Press to change LCD display units of measurement between °C and °F.
TIME/IP Button: Press repeatedly to change LCD display from: 1) Date and Time;
(receiving or sending packets).
parameters from the DHCP server this LED will turn Solid yellow.
during back-up battery operation according to sampling rate, or recording interval if recording is ON.
Press and Hold along with TIME/IP button during power-on, this will enable DHCP mode.
2) Temperature and Humidity; 3) iServer’s IP address. Press and Hold along with °C/°F button during power up, this will enable DHCP mode.
7
Page 14
2.3 Parts of the iServer Unit (continued)
IS NOT BEING USED
STBY Button: 1) Stops the recording 2) Press before ejecting Flash Card.
BKLT Button: Push and Hold to display Backlight on LCD when it is running on the
NOTE: display will show “Safe to Eject SD” after button has been pressed.
back-up battery (backlight is always on while running on the ac adapter).
2.4 Disassembly Instruction
You may need to open the unit for one of the following reasons:
W To wire relay connector. (Refer to Figure 2.6) W To connect or replace the battery.
Disconnect the power supply before proceeding.
W Make sure the flash memory card is fully inserted (or removed), before removing the
cover.
W Remove probe from handle clip (if needed).
Remove cover, by removing 2 mounting screws on each side.
Press STBY before ejecting Flash Card. Display will show “Safe to Eject SD” after button has been pressed.
FLASH
CARD
COVER
MOUNTING SCREWS (2)
RELAY
CONNECTOR
TRAY
POWER ADAPTOR
REMOVE BATTERY CLIP INSULATOR COVER AND
STORE WHERE SHOWN WHEN 9V BATTERY IS BEING USED.
KEEP INSULATOR COVER ON BATTERY CLIP WHEN BATTERY
Figure 2.4 Opening the Unit
8
Page 15
2.5 Network Communication Interfaces
51
Removable Plug
2.5.1 10/100 BASE-T RJ-45 Pinout
The 10/100BASE-T Ethernet network system is used in the iServer for network connectivity. The 10 Mbps or 100 Mbps twisted-pair Ethernet system operates over two pairs of wires. One pair is used for receiving data signals and the other pair is used for transmitting data signals. This means that four pins of the eight-pin
Pin Name Description
1 +Tx +Transmit Data 2 - Tx -Transmit Data 3 +RX +Receive Data 4 N/C Not Connected 5 N/C Not Connected 6 -Rx -Receive Data 7 N/C Not Connected 8 N/C Not Connected
connector are used.
Figure 2.5 RJ45 Pinout
For CE compliance at 100 Mbps: use shielded cable, opposite end of cable must be grounded.
2.5.2 Connecting iServer to PC/Hub/Switch/Router
The iServer’s Ethernet interface can automatically detect the Rx and Tx lines on a twisted pair Ethernet cable (MDI/MDIX Auto Cross). Therefore, to connect an iServer to a PC/Hub/Switch/Router, either a straight-through or a cross-over cable can be used.
On certain devices (like iServer), it is possible for the hardware to automatically correct errors in cable selection, making the distinction between a “straight­through” cable and a “cross-over” cable unimportant. This capability is known as “Auto MDI/MDIX”.
2.6 Relay Wiring Connections
To access the Relay Connector you must remove the cover, refer to Section 2.4. It is recommended that you ground your unit by connecting a wire to the
Ground/Return position of the relay connector.
Connector for Relays under the Cover
1
6
Side or Bottom Wire Entry for Relay Connector
Figure 2.6 Relay Connections
RELAY 2 RELAY 1
6234
9 Vdc
NO2
COM2
RTN
9
NO1
COM1
Page 16
PART 3 NETWORK CONFIGURATION
3.1 Network Protocols
The iServer can be connected to an Ethernet network communicating through standard IP protocols including TCP, UDP, SNMP, SMTP, ARP, HTTP (WEB access), DNS, DHCP, ICMP, SNTP, and Telnet.
3.2 Ethernet (MAC) Address
MAC (Media Access Control) address is a unique hardware number for Ethernet devices like computers, network switches, print servers, etc. When you’re connected to an Ethernet LAN using a computer you can see a table of IP addresses called “ARP table” stored on that computer. The ARP table relates IP addresses of devices on a network to their corresponding MAC addresses. The MAC address can be found on a label attached to your Ethernet device and it contains 6 bytes (12 characters) of hexadecimal numbers XX:XX:XX:XX:XX:XX hex
For example: 0A:0C:3D:0B:0A:0B
Remove the small label with the default IP address of 192.168.1.200 and there will be room to put your IP address. See Figure 3.1
REMOVE FACTORY DEFAULT
IP ADDRESS LABEL.
PUT THE NEWLY ASSIGNED
IP ADDRESS IN THIS LOCATION.
Figure 3.1 Labeling
FIRMWARE VERSION #
10
Page 17
3.3 DHCP
DHCP, Dynamic Host Configuration Protocol, enables computers and network devices to receive their IP configurations from a DHCP server.
If DHCP is enabled on your iServer, as soon as the iServer that is connected to the network is powered on, there will be an exchange of information between the iServer and the DHCP server. As a result, the DHCP server will assign an IP address, a Gateway address, a Subnet Mask, and a DNS address to the iServer. Note that the DHCP server must be correctly configured to make such assignments.
If fixed or static IP address is desired, the DHCP function must be disabled. The iServer is shipped with DHCP disabled (factory default). The DHCP can be enabled by accessing the iServer’s web server and selecting Network
menu (refer to Section 4.4.2) or by pressing and holding the two front buttons °C/°F and TIME/IP during power-on (refer to Section 2.3).
It’s very important to communicate with the network administrator in order to understand DHCP and its existing configurations on the host server before enabling DHCP on the iServer.
The iServer is shipped with a default static IP address of 192.168.1.200 and Subnet Mask of 255.255.255.0.
3.4 DNS
DNS, Domain Name System, enables computers and devices to be recognized over a network based on a specific name instead of IP addresses.
For example, instead of having to use http://192.168.1.200 (IP address), you would use http://ithx03ec or any name up to sixteen alphanumeric characters defined as a Host Name in the iServer’s web server.
The default Host Name for an iServer is “ithx” followed by the last four digits of the MAC address of that iServer unit.
On Windows servers where DHCP and DNS are separate functions it is very important to configure the DHCP server to communicate with DNS in order for the iServer’s Host Name to correctly respond. If you cannot access the iServer using its Host Name, please contact your network administrator to make sure DHCP and DNS servers are linked together.
If DNS server address is setup, all Host Names reported during weblink configuration (i.e. SMTP server IP, SNTP server IP, SNMP trap server IP, etc.) will be translated into IP addresses.
3.5 IP Address
Every active device connected to the TCP/IP network must have a unique IP address. This IP address is used to build a connection to the iServer.
All network devices like computers that use TCP/IP protocol to communicate with each other should have a unique 32-bit address called IP address. The IP address is divided into two portions, the network ID and the host ID. For instance, every computer on the same network uses the same network ID. At the same time, all of them have different host IDs. For more details about the IP address see Appendix B.
11
Page 18
3.6 Port Number
All TCP connections are defined by the IP address and a port number. A port number is an internal address that provides a TCP/IP interface between an application software on a computer and a device on the network or between two devices on the network.
There are two default TCP port (socket) numbers assigned to the iServer:
1) Port 2000: Once a TCP connection is made to the iServer using port 2000 or any port number that is configured on the iServer for Local Port (see Network menu, Section 4.4.2), the iServer can then be polled for variables like temperature, humidity, etc. using commands. For the list of commands see Section 4.9.1.
2) Port 2002: This port is the iServer’s network console port for reading or changing the settings within the iServer. For the list of settings refer to Section 4.8.1. For both of the above cases, for example, a Telnet application can be used.
PART 4 OPERATIONS
Depending on user’s preference and network setup, the iServer can be configured in several ways:
Using the latest versions of Web browsers like Internet Explorer, Firefox, Opera, or Safari (iPhone) the iServer’s Web server can be accessed, see Section 4.2.
It can also be configured using a TCP connection to port 2002 using a command line interface, see Section 4.8.1.
The iCONNECT Configuration Software can also be used to find and configure the iServer over the Ethernet, see Section 4.1.2.
Before you configure the iServer, you will need to assign an IP address to the unit. Refer to Section 4.1.1, 4.1.2 or 4.1.3.
4.1 IP Address Assignment
4.1.1 DHCP (Method 1)
The DHCP server will assign an IP address, see Section 3.3.
4.1.2 ARP HTTPget Commands (Method 2)
You can download the HTTPget program from our web site to use it for this section.
1. From the Command line (DOS prompt) where you keep the HTTPget.exe file, type the
following:
arp -s 128.100.101.76 00-03-34-00-9c-5c press the <Enter> key
128.100.101.76, for example, is the IP address that you want to assign to your iServer
and 00-03-34-00-9c-5c is the MAC address of your iServer labeled on the bottom of the unit.
2. From the same prompt, type the following HTTPget command:
httpget -r -S “00000000” 128.100.101.76:1 press the <Enter> key
00000000 is the default ADMINISTRATOR password in your iServer unit and
128.100.101.76:1 is the IP address that your iServer will accept on TCP port 1. At this
point, you have successfully assigned 128.100.101.76 IP address to your iServer. Continued on the following page.
12
Page 19
4.1.2 ARP HTTPget Commands (Method 2) (continued)
The above IP address of 128.100.101.76 is an example to show how these commands work. To get a valid IP address on your network you need to consult with your IT department.
Figure 4.1 shows the screen shot of an actual DOS window with the above commands executed.
C:\>arp -s 128.100.101.76 00-03-34-00-9c-5c
C:\>httpget -r -S “00000000” 128.100.101.76:1 New IP is assigned. The unit will reset in 10 seconds. C:\>
Figure 4.1 ARP and HTTPget Commands on a DOS Window
4.1.3 Direct Connection (Method 3)
The iServer is shipped with a default IP address of 192.168.1.200 and Subnet Mask of
255.255.255.0.
To start, connect your iServer directly to your PC using a straight or cross-over Ethernet cable and power it on.
Make sure that the Network LINK/ACT LED on the iServer and your Ethernet switch is solid ON. Refer to Figure 4.2.
If you are going to use a Web browser or Telnet program to access the iServer using its default IP address of 192.168.1.200, you need to make sure that the PC from which you’re establishing the connection has an IP address that is in the same range as the iServer’s IP address.
For example, you can assign your PC an IP address of 192.168.1.x, where x is any number from 1 to 254.
Your PC’s IP address cannot be the same as the iServer’s IP address.
You also need to make sure that your PC’s subnet mask is 255.255.255.0. This is a good way to access the iServer over the network and make any configuration
changes needed. Changing your PC’s IP address is temporary and it’s only for the purpose of accessing
the iServer for initial configurations like assigning a new IP address, subnet mask, etc. Once the iServer is configured you can put your PC back to its original IP settings. If 192.168.1.200 is already in use on your network, connect the iServer directly to
your computer using a CAT5 Ethernet cable (either straight or cross-over cable will be detected by the iServer) and proceed as described below.
13
Page 20
4.1.3 Direct Connection (Method 3) (continued)
.
Ethernet Cable
Computer’s Ethernet Port (RJ45 connection)
iServer with
default IP address
of 192.168.1.200
Figure 4.2 Connecting Computer Directly to iServer
1) Click on your Windows “Start” menu and select “Control Panel” from the list options.
Double click on the “Network Connections” icon
After connecting the iServer to computer, power it on LINK/ACT LED should be SOLID green
100 BASE-T LINK / ACT DHCP
1
8
RECORDING
Plug in the Power Adapter
Figure 4.3 Network Connections
2) You now have “Network Connections” window opened.
The “Local Area Connection” icon has all the settings for your ethernet connection.
Double click on this icon.
Figure 4.4 Local Area Connection
14
Page 21
4.1.3 Direct Connection (Method 3) (continued)
Fill out the feilds for the IP
3) Click on “Properties” button
Select “Internet Protocol (TCP/IP)” and click on “Properties” button
address 192.168.1.100 and Subnet mask 255.255.255.0 as indicated below. Press OK and also on all other remaining windows.
Figure 4.5 Changing TCP/IP Properties on Your Computer
4) Now you may need to reboot your computer.
5) Once your PC is back up, open your browser and type the iServer’s default IP address of 192.168.1.200 in the “Address” window of your browser and press the Enter key.
You will soon be inside the Welcome page (see
Figure 4.6)
of the iServers WEB
server and from there you can access all the menus.
iTHX-SD Welcome Page
http://192.168.1.200Address
Readings Chart
SetupWeb Link
Figure 4.6 iServer Welcome Page
15
Page 22
4.1.3 Direct Connection (Method 3) (continued)
To verify a good connection to the iServer, from a DOS prompt on your computer type “ping 192.168.1.200” and press Enter.
You should get a reply as shown in Figure 4.7. If you don’t receive a reply it means that you do not have a network connection
between your PC and the iServer. Check your IP address, subnet mask, and gateway address.
C:\>ping 192.168.1.200 Pinging 192.168.1.200 with 32 bytes of data:
Reply from 192.168.1.200: bytes=32 time=15ms TTL=64 Reply from 192.168.1.200: bytes=32 time=8ms TTL=64 Reply from 192.168.1.200: bytes=32 time=8ms TTL=64 Reply from 192.168.1.200: bytes=32 time=8ms TTL=64
Pinging statistics for 192.168.1.200: Packets: Sent=4, Received=4, Lost=0 (0% loss)
Approximate round trip times in milli-seconds: Minimum=8ms, Maximum=15ms, Average=9ms
Figure 4.7 Pinging the iServer from MS-DOS Prompt
16
Page 23
4.1.4 iConnect Software
(Method 4)
The iServer can be assigned an IP Address by using the iConnect software. a) Download the iConnect software from the website listed on the cover of this manual
(software section).
b) Install iConnect software on a networked PC. This software is compatible with
Windows 95, 98, NT, 2000, and XP.
c) Use iConnect to assign an IP address to the iServer and access its web pages for
configuration. You can also use any standard web browser to access the iServer’s web pages.
Consult with your IT department for obtaining an IP address.
Figure 4.8 Assigning an IP Address using iConnect
1) Place the IP address in this box.
2) Take the MAC address from the label attached to the bottom of the iServer and
place it in this box.
3) Click here to send the above IP address to the iServer.
4) After the IP address is assigned to the iServer, click here to access it’s web pages.
5) Click here to Ping the iServer whose IP address is shown in the IP address box.
6) Click here to find all the iServer’s on your network .
7) The IP addresses for the iServer’s found by the iConnect will be listed here.
8) These fields indicate the IP address and the subnet mask of the PC on which the
iConnect is running.
17
Page 24
4.1.4 iConnect Software
d) To access the iServer for Configuration: Click on the “View Webpage” button, you will access the iServer’s welcome page.
To take advantage of the iServer’s full capability use any standard web browser to access the iServer’s web pages as described in Section 4.2.
(Method 4)
(continued)
Readings Chart
SetupWeb Link
Figure 4.9 Accessing the iServer’s using iConnect
18
Page 25
4.2 Access and Configuration Using a Web Browser
WStart your web browser. WIn the URL field, type http://192.168.1.200 (iServer’s default IP address). WThe Welcome Page, will be displayed.
iTHX-SD Welcome Page
http://192.168.1.200Address
Readings Chart
SetupWeb Link
Figure 4.10 iServer Welcome Page
In order to access the iServer’s web pages, users may be prompted for a password, as shown below.
4.3 LOGIN and ADMINISTRATOR Passwords
Login
http://192.168.1.200Address
Readings
Chart
Password:
Web Link Setup
LOGIN
OK
Administrator Login
http://192.168.1.200Address
Readings
Web Link Setup
Chart
ADMINISTRATOR
Password:
OK
Figure 4.11 LOGIN and ADMINISTRATOR Passwords
There are 2 different access levels:
1. LOGIN Password is required to access the iServer’s Readings, Chart, and Web Link pages unless it’s disabled. The default password is 12345678.
2. ADMINISTRATOR Password is required to access the iServer’s Setup menu unless it’s disabled. The default password is 00000000.
Refer to Section 4.4.5 for Password setup.
19
Page 26
4.4 Setup
Menu BarMenu Panel
Clicking the Setup button on the Welcome page (see Figure 4.10) will provide access to the Menu Panel (see Figure 4.12).
Using this Panel you can configure the iServer entirely.
4.4.1 Overview
Once the Administrator password is entered, the Overview page will appear which provides a summary of important parameters within the iServer.
All the fields are read-only.
Overview
http://192.168.1.200Address
Overview
Network Configuration Management Security Recording System Diagnostics Log Out
Readings
Model Firmware Version DHCP MAC Address IP Address Subnet Mask Gateway Address Hostname Ethernet Port Web Server Port Recording Uptime
Chart
Web Link Setup
Figure 4.12 Overview
x.x Disabled 00:03:34:00:b8:70
192.168.1.200
255.255.255.0
0.0.0.0 ithxb870 Auto 80
OFF
1 day, 00:00:00 hh:mm:ss
20
Page 27
4.4.2 Network
This menu provides network configurations including IP parameters and Ethernet interface options. Fields are described below.
4.4.2.1 IP Configuration
Network
http://192.168.1.200Address
Overview
Network
Configuration Management Security Recording System Diagnostics Log Out
Readings
DHCP
MAC Address
IP Address
Subnet Mask
Gateway Address
DNS Address
Host Name
Protocol
Local Port
Web Server Port
Web Link Title
Web Link Address
Chart
Web Link Setup
IP Configuration Ethernet Port
00:03:34:00:b8:70
192.168.1.200
255.255.255.0
0.0.0.0
0.0.0.0
ithxb870
TCP
2000
80
WebLink
www.abc.com
Save Changes Reset
Figure 4.13 Network: IP Configuration
DHCP – If the box is checked the iServer will dynamically request an IP address, a
subnet mask, a gateway address, and a DNS address from the DHCP server. By default the DHCP option is disabled. For more information about DHCP, see Section 3.3.
DHCP can be enabled by pressing and holding the two front buttons ºC/ºF and
TIME/IP during power-on (refer to Section 2.3).
MAC Address – This Indicates the hardware address of the iServer and it is non- configurable. For more information about MAC Address, see Section 3.2.
IP Address – This indicates the IP address of the iServer. The iServer’s default IP
address is 192.168.1.200. When DHCP is enabled this field will be dimmed. Consult with your IT department for obtaining an IP address.
Subnet Mask – It’s a 32-bit number that is used to determine which part of the IP address is the network portion and which part is the host portion. When DHCP is enabled this field will be dimmed. The iServer’s default Subnet Mask is 255.255.255.0. Consult with your IT department for obtaining a subnet mask.
21
Page 28
4.4.2.1 IP Configuration (continued)
Gateway Address – This points to the router that forwards traffic to a destination
address outside of the subnet on which the iServer resides. This is the IP address of the router which functions as a gateway. When DHCP is enabled this field will be dimmed. The iServer’s default Gateway address is 0.0.0.0. Consult with your IT department for obtaining a gateway address.
DNS Address – In order to use the hostname to access the iServer, the DNS server on your network must be configured, refer to Section 3.4. iServer plays the role of a DNS client, in the sense that the iServer will actively query the DNS server for the IP address associated with a particular domain name. When DHCP is enabled this field will be dimmed. The iServer’s default DNS address is 0.0.0.0. Consult with your IT department for obtaining a DNS address.
Host Name – If DHCP is enabled, the iServer will send this name to the DHCP server. This name is used so that the iServer can be accessed based on a specific name instead of an IP address. For example, instead of using http://192.168.1.200 (IP address), you would use http://ithxA838 or any name up to sixteen (16) alphanumeric characters. The default Host Name for an iServer is “ithx” followed by the last four digits of the MAC address of that particular iServer.
On Windows servers where DHCP and DNS are separate functions it’s very important to configure the DHCP server to communicate with DNS in order for the iServer’s Host Name to correctly respond. If you cannot access the iServer using its Host Name, please contact your network administrator to make sure DHCP and DNS servers are linked together.
Protocol – It’s the network protocol the iServer communicates with the Ethernet Network. Options are TCP and UDP. The default is TCP.
Local Port – The default port is 2000. Refer to Section 3.6. Web Server Port – The default port is 80. This is the primary port number for the HTTP
protocol used for communication between internet browsers and web sites/web servers. Web servers open this port then listen for incoming connections from web browsers. Similarly, when a web browser is given an IP address (like the iServer’s IP address), it assumes that the iServer’s web server is listening for connections on port 80. If this port is changed to anything but 80 then on the browser the new port number must be indicated with a colon ( changed to 500, you will then need to type http://192.168.1.200:500 on the browser to access the iServer’s web server.
One of the applications where the Web Server Port number may need to change is when users want to access the iServer’s web server from outside the local area network (i.e. Internet). By setting up “Port Forwarding” inside a router that is the gateway to that local area network this task can be accomplished. “Port Forwarding” technique uses the Web Server Port number to forward the Internet connection to the iServer on the LAN.
Web Link Title – This is a text field that appears on the button on the iServer’s Web Link Page, refer to Section 4.7. This can describe the Web Link Address assigned below.
Web Link Address – This provides a link to any TCP/IP node on the network or any Web link on the Internet. Examples would be www.123abc.com or if you have a device with an embedded Web server (just like the iServer) once you enter its IP address in this field and click on the Web Link button on the iServer’s Welcome page you’ll be able to access your device using the same browser interface.
:) after the IP address. For example, if the Web Server Port is
22
Page 29
4.4.2.2 Ethernet Port
Ethernet Port
http://192.168.1.200Address
Overview
Network
Configuration Management
Readings
Auto-Negotiation
Speed
Duplex
Chart
Web Link Setup
IP Configuration Ethernet Port
100 Mbps
Full
10 Mbps
Half
Security Recording System
Save Changes Reset
Diagnostics Log Out
Figure 4.14 Network: Ethernet Configuration
Auto-Negotiation – It is the link, in terms of speed and duplex, between the iServer and
another Ethernet device like an Ethernet switch. If Auto-Negotiation box is checked, the iServer will auto-negotiate the speed and duplex
with the attached Ethernet device. If any of the other options are selected, the speed and duplex will be fixed.
It’s important to have the same Ethernet port configuration on the iServer and the attached Ethernet device. If iServer cannot auto-negotiate with the attached Ethernet node it will default to 10 Mbps and Half-Duplex. Once Auto-Negotiation is checked, other fields under this category will be dimmed. By default the Auto-Negotiation is checked.
If the iServer detects the link to be 100 Mbps the 100 BASE-T LED will be solid green.
4.4.3 Configuration
This menu provides configurations for the real-time clock, server parameters, sensors, and alarm relays settings. Fields are described below.
4.4.3.1 Date and Time Current Date – This field indicates the iServer’s real time clock date. The format is yyyy/
mm/dd. When there is no date defined, the iServer will be defaulted to 2099/01/01 and be shown in red to alert you that it has not yet been set.
Current Time – This field indicates the iServer’s real time clock time. The format is military time (24-hour) and it is entered as hh:mm:ss.
Change Date and Time – By clicking on this option the real data and time can be entered.
23
Page 30
4.4.3.1 Data and Time (continued)
Date and Time
http://192.168.1.200Address
Readings
Overview Network
Configuration
Management Security Recording System Diagnostics Log Out
Chart
Date and Time
Current Date
Current Time
Change Date and
Time Server
Network Time Server
NTS Address
Time Zone
Web Link Setup
SensorsServer Alarm RelaysDate and Time
2099/01/01 (yyyy/mm/dd)
00:00:00
Time
(GMT-08:00)Pacific Time(US&Canada)
Save ChangesCancel
Figure 4.15 Network: Ethernet Configuration
Once the AC adaptor is disconnected from the iServer and there is no battery back-up, the iServer will lose its date and time settings.
Once the iServer is rebooted from any web page, the iServer will lose its date and time settings whether or not the back-up battery is installed.
If the iServer is setup to get it’s time from a Network Time Server, then the time will be re-assigned after the reboot.
Network Time Server – If there is a time server on the network or the Internet, the iServer will get the real date and time once the Network Time Server’s IP address is provided. Default is unchecked.
NTS Address – The field to enter the Network Time Server’s IP address. The U.S. National Institute of Standards and Technology (NIST) publishes a list of time
servers on the Internet used by the NIST Internet Time Service (ITS). The list includes each server’s name, IP address, and location in the United States. As of the publication date of this manual, links to the list can be found at http://tf.nist.gov.
Time Zone – If Network Time Server option is checked, the correct time zone must be selected for the correct time display.
24
Page 31
4.4.3.2 Server
Server
http://192.168.1.200Address
Overview Network
Configuration
Management Security Recording System Diagnostics Log Out
Readings
Server
ServerType
Interval
Time Stamp
Disconnect After Data Sent
Web Link Setup
Chart
Save Changes Reset
SensorsServer Alarm RelaysDate and Time
Command
5 secs
Figure 4.16 Configuration Menu: Server
Server Type – Options are Command and Continuous.
If Command is selected the iServer will respond to commands sent from a network host (for the list of commands refer to Section 4.9.1). If Continuous is selected the iServer will send temperature, humidity, dewpoint values and alarm relay status once every time amount selected in the Interval box. The Command and Continuous modes operate when a network host opens a TCP connection to the iServer’s IP address with port 2000. In UDP Command mode, iServer will respond back with a UDP packet (to the _ address and port) if a host application sends a UDP command to iServer’s IP address with port
2000.
In UDP Continuous mode, iServer starts sending continuous data if a host application sends a trigger or UDP packet to iServer’s IP address with port 2000. Interval – This is the time interval (in seconds) between each data transmission when the iServer is in Continuous mode. The default value is set to 5 seconds and the minimum is 2 seconds. Time Stamp – If checked, the iServer will stamp the data with date and time before sending it out to a network host. This will apply to Command and Continuous modes. Default is unchecked. Disconnect After Data Sent – If checked, once the iServer responds back with data, the iServer will close the TCP connection that had been made from the network host. This feature should be used if data acquisition software expects data to be ended by closing TCP connection from the client.
4.4.3.3 Sensors
This is the page where the Sensor Parameters are defined. Reading 1 is for Temperature, Reading 2 is for Humidity, and Reading 3 is for Dewpoint.
Enabled Probes – Choose Both Probes if you have two probes (sensors). Choose Only Probe 1 if you just have one probe, or if you have two probes but only want to show Probe 1. Default is Both Probes.
Degree Unit – This would be the units of measurement for the probe sensor. Examples are F and C for temperature and dewpoint. Once the unit of temperature is selected, either C or F, it will be a global change throughout the web server, as well as on the LCD display. If you use the front C/F buttons to change the unit of temperature it will have no effect on the web pages.
25
Page 32
4.4.3.3 Sensors (continued) End Character – This means that the iServer will send a character (in Hex) after the
value for each variable. This will apply to either Continuous or Command mode. Once the value is selected, it will be a global change for all readings. The default is set for 0D (Hex representation of <CR>). Description – This can be a name for the sensor or where the sensor is located. This field is text only and can have up to 16 alphanumeric characters. This description will appear wherever Reading 1, 2 or 3 is displayed on the web server. Unit – This would be the units of measurement for that particular sensor. Examples are % for humidity, F and C for temperature and dewpoint. Once the unit of temperature is selected, either C or F, it will be a global change through out the web server, as well as on the LCD display. If you use the front C/F buttons to change the unit of temperature it will have no affect on the web pages.
Sensors
http://128.100.101.232Address
The adjusting values must be assigned after the unit of temperature (C or F) is selected. If the unit of temperature is changed, you must readjust the values for a correct result.
Overview Network
Configuration
Management Security Recording System Diagnostics Log Out
Readings
Enabled Probes End CharactersDegree Unit 0x
Probe 1
Reading 1
Description
End Character
Adjust by
Reading 2
Description
End Character
Adjust by
Reading 3
Description
End Character
Chart
Temperature 1
Offset
Humidity 1
Offset
Dewpoint 1
Web Link Setup
SensorsServer Alarm RelaysDate and Time
Both Probes
Unit
0x0A
0x0A
0x0A
Data Format
Offset
Reading Low
Reading High
Unit %
Data Format
Offset
Reading Low
Reading High
Unit C
Data Format
C
C
000.0C
0.0
0.0
100.0
H00.0%
0.0
0.0
100.0
D000.0C
Display
Ref. Low
Ref. High
Display
Ref. Low
Ref. High
Display
0A
0.0
100.0
0.0
100.0
Figure 4.17a Configuration Menu: Sensors
Display – If checked the variable (i.e. temperature) will be displayed on the Readings and Chart applets as well as data transmission in Continuous mode (see Server Section
4.4.3.2). The default is checked. Data Format – This indicates the format of the variable being displayed on a network
host when Continuous or Command transmission is used. Reading 1: T000.0C, 000.0C,
000.0; Reading 2: H00.0%, 00.0%, 00.0 (the C or F will change to the appropriate unit once you press Save Changes button).
Adjust by – If it’s determined that the reading is slightly off this field can be used to adjust the reading. The options are Offset and Scale & Offset. Using the Offset method you can manually assign a numerical value to adjust the reading. For example if the actual reading is 55 where it should be 54 you can enter -1 in the Offset field. The Scale & Offset method uses two points
Reading Low 10.0
Reading High
200.0
Ref. Low
Ref. High
12.0
195.0
to adjust the reading within the entire temperature range. For example, if at the lower point the actual reading is 10 where it should be 12 and at the higher point the actual reading is 200 where it should 195, here is how the numbers are entered. The adjusting numbers can be positive or negative with one decimal point.
26
Page 33
4.4.3.3 Sensors (continued)
Offset – If it’s determined that the reading is slightly off, this field can be used to
manually assign a numerical value to adjust the reading. The offset can be a positive or a negative number with one decimal point.
The temperature offset value must be assigned after the Unit (C or F) is selected. If the unit of temperature is changed, you must reassign the offset value for the
correct result. Sampling Rate – This determines the rate in which the iServer polls the probe for temperature and humidity values. The options are 1, 10, and 60 seconds. The default is set for 10 seconds.
The higher the sampling rate, the more accurate data is transmitted by the probe.
LCD Display – Probe to Display - This dropdown box is only available when Both Probes are enabled (on Enabled Probes drop down box). If you only have 1 probe then probe 2 will be shown as OPEN on the readings page, chart and recorded data.
Sensors
http://192.168.1.200Address
Probe 2
Reading 1
Description
End Character
Adjust by
Description
End Character
Adjust by
Description
End Character
Sampling Rate
One Sample per
LCD Display
Temperature 2
Offset
Reading 2
Humidity 2
Offset
Reading 3
Dewpoint 2
Probe To Display
0x0A
0x0A
0x0A
1 second
Unit Display
C
Data Format
Offset
0.0
Reading Low
Reading High
Unit %
Data Format
Offset
0.0
Reading Low
Reading High
Unit C
Data Format
Both Probes
Save Changes Reset
000.0C
0.0
100.0
H00.0%
0.0
100.0
D000.0C
Ref. Low
Ref. High
Display
Ref. Low
Ref. High
Display
0.0
100.0
0.0
100.0
Figure 4.17b Configuration Menu: Sensors
4.4.3.4 Alarm Relays
This section would allow the two relays on the iServer to be configured as
needed. For physical connections and pin-outs refer to Section 2.6.
Relay 1 and Relay 2 – This configures the first and second relays on the iServer. The options for each are temperature, humidity, and dew point. This means that both Relays can be set for temperature or both Relays can be set for humidity, or one can be set for humidity and one for temperature. Description – This is a text field for naming the device connected to the relay or it could be the location where the device is installed. This field can take up to 16 alphanumeric characters. This name will be displayed anywhere on the web pages associated with alarm relays; for example the Readings applet. Defaults are Alarm Relay1 and Alarm Relay2.
27
Page 34
4.4.3.4 Alarm Relays (continued)
Alarm Relays
http://192.168.1.200Address
Overview Network
Configuration
Management Security Recording System Diagnostics Log Out
Readings
Probe 1
Relay 1
Description
Status
Type
Set Point High
Set Point Low
End Character 0x
Display
Record all alarm events on the flash memory
Chart
Web Link Setup
SensorsServer Alarm RelaysDate and Time
Temperature
R1alarm1
Latch
Clear Latch
101.0
30.0
0A
Save Changes Reset
Probe 2
Relay 2
Description
Status
Type
Set Point High
Set Point Low
End Character 0x
Display
Retrieve
Dew Point
R2alarm2
Low or HighLow or High
Latch
Clear Latch
60.0
0.2
0F
Alarm status with
the associated set point will be shown on the chart applet as well.
Figure 4.18 Configuration Menu: Alarm Relays
Status – The options are Disable, Low, High, and Low or High. The default is Disable.
If Low is selected Set Point Low must be given a value and once the sensor reading falls below that value the relay will change status. If High is selected Set Point High must be given a value and once the sensor reading rises above that value the relay will change status. If Low or High is selected Set Point Low and High must be given values and relay status will change once the sensor reading is outside that range. If Disable is selected relays will be deactivated. Type – The options are Unlatch and Latch. The default is Unlatch. If Unlatch is selected and the relay is activated, once the sensor reading goes back within the Set Point range, the alarm will be deactivated automatically. If Latch is selected, once the sensor reading goes back within the Set Point range, the relay will remain active until the user clears it by clicking on the Clear Latch button.
Set Point High – This is the field for the upper Set Point. Set Point Low – This is the field for the lower Set Point. End Character – This is the character (in Hex) that the iServer will send along with the
alarm status. This applies to either Command and Continuous mode described on the Server page, under the Configuration menu (Section 4.4.3.2). The default is 0D meaning a Carriage Return character will be sent after the value. Display – This would indicate whether to display the status of the alarm relays on the Readings applet as well as the TCP Command and Continuous modes, and any other related fields. The default is unchecked. Record all alarm events on the flash memory – If checked, all the alarm events with stamped date and time will be recorded on the flash memory card. By pressing the Retrieve button the ASCII log file will be opened. The default is unchecked.
Using the web server, the log file can be retrieved if the file size is 32 KB or smaller. For a larger file size you must use a flash card reader to open the log file. Refer to Section 4.10.
28
Page 35
4.4.4 Management
This page provides the iServer’s email, SNMP and alarm settings. SNMP (Simple Network Management Protocol) is the protocol used by network management systems to communicate with network devices that respond to SNMP connections for the purpose of problem detections and corrections. If SNMP and/or SMTP features are desired, please make sure iServer’s network setting as well as SNMP trap server and SMTP mail server are setup properly. Failure to do so may cause iServer to be unresponsive while waiting for these servers to reply back. Fields are described below.
4.4.4.1 Setup -
Setup
http://192.168.1.200Address
Overview Network Configuration
Management
Security Recording System Diagnostics Log Out
Readings
Simple Network Management Protocol
SNMP Response Enabled SNMP Community Contact Location SNMP Trap Server
Simple Mail Transfer Protocol
SMTP Server SMTP Server Port From: To :
Subject: Reminder Interval Transmission Delay
Chart
Web Link Setup
AlarmsSetup
0.0.0.0
25
Save Changes Reset
public
None
None
0.0.0.0
60
1
mins
mins
Figure 4.19 Management Menu: Setup
Simple Network Management Protocol SNMP Response Enabled – If this option is checked the iServer will respond to network
nodes broadcasting SNMP requests. The default is unchecked. Make sure your network is setup properly (gateway, etc) before using SNMP feature SNMP Community – Every SNMP communication takes place using a community
string. The default is Public. Contact – This text field specifies the contact person for this node, together with information
on how to contact this person. This field takes up to 16 alphanumeric characters. Location – This text field specifies the location of the iServer. For example it can be
“control station”, which is the place where the iServer is located. This field takes up to 16 alphanumeric characters.
29
Page 36
4.4.4.1 Setup (continued) SNMP Trap Server – This field contains the IP address of the SNMP trap server located
somewhere on the network. The trap server listens for SNMP traps coming from the iServer when there is an alarm condition (refer to Alarms Section 4.4.4.2 under the Management page).
Simple Mail Transfer Protocol SMTP Server – This field specifies the IP Address of the SMTP server.
1) You must have an email server (SMTP server) on your network in order to receive emails generated by the iServer.
2) iServer does not support SMTP server authentication.
SMTP Server Port – This specifies the TCP port used by the SMTP Server. The default is 25. From – This field specifies the name of the person who sends the email. It can also be
an email address. To – This field specifies the email address of the recipient. This field can take up to 200
alphanumeric characters; limited to 4 addresses of 50 characters each. Each address should be separated by a comma, without spaces. Ex: aaa@bbb.com,ccc@ddd.com
You can create a “distribution group address” if you need to have more email addresses.
Subject – This field specifies the subject of the email. Emails for all the alarms will have this common subject line. Example of a subject can be “Alarm from iServer”.
Reminder Interval – This field sets a reminder interval for email and/or trap to be sent again. The allowed minimum value is 1 minute. If it’s set to 0 the iServer will not send a reminder email. The default is 60 minutes.
Power Reset alarm is only sent once, regardless of reminder interval. Transmission Delay – Once the iServer alarm condition is met, this is the amount of
time the alarm condition is met before iServer sends any email or trap. If it’s set to 0, email and/or trap will be sent immediately. The default is 1 minute.
4.4.4.1.1 Sending Text Messages to a Cell Phone
In the SMTP To field, you can use the following format to have the iTHX-SD send a text message to your cell phone. Since most cell phones are capable of receiving text messages you just need to find the correct email format for your cell phone provider.
T-Mobile phone_number@tmomail.net Virgin Mobile phone_number@vmobl.com AT&T phone_number@txt.att.net Sprint phone_number@messaging.sprintpcs.com Verizon phone_number@vtext.com Nextel phone_number@messaging.nextel.com “phone_number” is your 10 digit cell phone number.
4.4.4.2 Alarms
This includes events or conditions. Once a condition is met the iServer will send an email or/and an SNMP trap if either one or both are checked. Fields are described below.
Multiple alarm events can be checked and for each event an email and/or a trap will be sent.
30
Page 37
4.4.4.2 Alarms (continued)
Alarms
http://192.168.1.200Address
Overview Network Configuration
Management
Security Recording System Diagnostics Log Out
Readings
Events
Temperature Low Low 0.0High 0.0Low 0.00.0
Humidity
Dewpoint
Temp Low High0.00.0
Humi
Dew
Power Reset
Probe 1 Disconnected
Probe 2 Disconnected
Recording Stopped
SD Card 10% Memory Left
Low Battery
Relay 1 Activated
Relay 2 Activated
Web Link Setup
Chart
Low High
Low 0.00.0
Low HighLow 0.00.0
Low or High
Low High
Low or High
Low or High
0.00.0
Low High
0.00.0
Save Changes Reset
AlarmsSetup
EmailSNMP Trap
Probe 1
Probe 2
Device
There is a text box next to each event. The text content will be the body of the email.
Figure 4.20 Management Menu: Alarms
Temperature – The options are Low, High, and Low or High.
If Low is selected the Low field must be given a value (lower than that value is an alarm condition).
If High is selected the High field must be given a value (higher than that value is an alarm condition).
If Low or High is selected the Low and High fields must be given values (outside the band is an alarm condition).
Humidity – The same as Temperature defined above. Dew Point – The same as Temperature defined above. Power Reset – This refers to the iServer being powered off and back on (power reset).If
the option is checked the iServer will automatically send one email or/and an SNMP trap after rebooting, regardless of reminder interval.
Probe Disconnected – If the probe attached to the iServer is disconnected that is considered as an alarm condition.
Recording Stopped – If the iServer is recording, once the recording is stopped (no matter what the cause was) an email or/and a trap will be sent.
SD Card 10% Memory Left – An alarm will be triggered if 10% or less memory left on the flash card.
31
Page 38
Low Battery – If the 9V battery lowers to about 7V, the iServer will prompt an alarm. Relay 1 Activated – If relay 1 is activated the iServer will prompt an alarm. See Alarm Relays section under the Configuration page (Section 4.4.3.4). Relay 2 Activated – If relay 2 is activated the iServer will prompt an alarm. See Alarm Relays section under the Configuration page (Section 4.4.3.4).
4.4.5 Security
This page provides security and access settings for the iServer. Fields are described below.
Security
http://192.168.1.200Address
Overview Network Configuration Management
Security
Recording System Diagnostics Log Out
Readings
Chart
Web Link Setup
Login Password
Admin Password
Local Port Password
Te lnet Console Access
Web Server Access
Secured Readings and Charts
IP Exclusivity
Figure 4.21 Security Menu
12345678
00000000
Enable
No
IP Address
1
0.0.0.0
2
0.0.0.0
3
0.0.0.0
4
0.0.0.0
Save Changes Reset
Change
Change
Change
32
Page 39
Login Password – To access Readings, Chart, and Web Link menus this password is required. The password length can be up to 16 alphanumeric case-sensitive characters. To change the password click on Change.
Empty box means no password is required. The default Login Password is 12345678. Administrator Password – To access Setup menu this password is required. The
password length can be up to 16 alphanumeric case-sensitive characters. To change the password click on Change.
Empty box means no password is required. The default Administrator Password is
00000000. This password will also be prompted when Telnet connection to port 2002 is
made. Port 2002 is the Telnet console port that allows users to configure some of the parameters that can be configured through the web server.
Local Port Password – To access the iServer via TCP or UDP socket connection to port 2000, this password is required. The maximum length can be up to 16 alphanumeric case-sensitive characters.
Empty box means no password is required. The default is none. Telnet Console Access – If it is unchecked, Telnet console connection to port 2002 will
be blocked. The iServer must be rebooted for the change to take place. The default is checked. Web Server Access – If it is unchecked, the iServer’s Web server will not be accessible.
The iServer must be rebooted for the change to take place. The default is checked.
When the Web Server Access is not enabled, the iServer needs to be configured using Telnet console.
If someone needs to access the Web Server again, use the Telnet console to enable it.
Secured Readings and Charts – If it is checked the Login password will be required to display Readings and Chart applets.
The default is checked. IP Exclusivity – This table contains the IP addresses that are allowed to communicate
with the iServer over TCP or UDP connection. This means if the table is empty connections from all IP nodes are allowed to the iServer.
If enabled, all packets from IP addresses not on this list are ignored and thrown away.
33
Page 40
4.4.6 Recording
This section includes all the parameters and settings for data recording, SD flash card and recording status, data retrieval, and formatting the SD card.
Fields are described below.
4.4.6.1 Start Recording
Start Recording
http://192.168.1.200Address
Overview Network Configuration Management Security
Recording
System Diagnostics Log Out
Readings
Chart
Current Date: 2099/01/01 (yyyy/mm/dd) Current Time: 00:00:00
Interval(1-3600)seconds
Type
Wrap
Set Date and Time
Start Date
Stop Date
Set Weekdays (only for scheduled recording)
Mon Tue Wed Thur Fri Sat Sun
Web Link Setup
10
yyyy/mm/dd
yyyy/mm/dd
Start Recording
Format SD CardStart Recording Status Data Retrieval
Category
Start Time
Stop Time
Title
Quick Start Recording
Continuous
hh:mm:ss(24 hours)
hh:mm:ss(24 hours)
Figure 4.22 Recording Menu: Start Recording
Interval – This is the time interval between each data. It can be between 1 to 3600
seconds.
Title – This is the header for each recorded data file on the SD flash card. Quick Start Recording – If this button is clicked recording will start immediately.
Make sure your Current Date and Time are correct. Type – The options are Wrap and Non-Wrap. If Wrap is selected once the SD flash card is full the iServer will delete the old data
stored on the card in order to record the new data. Each time a new recording takes place an old record will be deleted.
If Non-Wrap is selected the iServer will stop recording once the SD flash card is full.
34
Page 41
4.4.6.1 Start Recording (continued)
Category – The options are Continuous and Scheduled.
In Continuous recording Start Date and Start Time fields are required. If there is not an End Date and End Time, the recording will continue non-stop.
In Scheduled recording, any of the days can be selected and the recording will take place only on those days. Here are some important notes about the Scheduled recording:
W The Start Date, Start Time, and Stop Time must be defined. The Stop Date does not
apply to Scheduled recording. This means that the Scheduled recording will never stop until the user decides to stop it.
W If the Start Date is one of the selected days recording will start on that day. W If the Start Date is not one of the selected days recording will start on the next 1st
selected day.
W The Start Time and the Stop Time apply to every single day that is selected. W The Start Time is the time that recording will start on those selected days. W The Stop Time is the time that recording will stop on those selected days. W The Start Time must be earlier (at least 20 seconds) than the stop time. W Let’s say today we’re configuring the unit for Scheduled recording. If the Start Time
is earlier than the Current Time (real time clock) and today is one of the selected days, the recording will start immediately.
W Scheduled recording is suitable for users who want to start and stop recording at
certain times on certain days. Users who want to have continuous recording three days a week, for instance, this feature can also be used. Just make sure to select start and stop times that would link the three days together.
For example, we need to have 3 days (72 hrs) a week continuous recording (Monday, Tuesday, and Wednesday) with recording Interval of 1 minute. The Start Time should be 00:00:00 and the Stop Time should be 23:59:00. This will connect all 3 days together and it will run on weekly bases.
Start Date – This is the date on which the data recording must start. Start Time – This is the time that the recording will start. It’s based on a 24-hour format.
The Start Time must be at least twenty seconds after the Current Time (real time). Stop Date – This is the date on which the data recording will stop. Stop Date does not
apply to Scheduled recording. Stop Time – This is the time that the recording will stop. It’s based on a 24-hour format.
A failure on the Ethernet network will not interrupt data recording.
35
Page 42
4.4.6.2 Status
This page describes the Recording and SD flash card status. All the fields on this page are read-only.
Status
http://192.168.1.200Address
Overview Network Configuration Management Security
Recording
System Diagnostics Log Out
Readings
Current Date: 2099/01/01 Current Time: 00:00:00
Recording Status
Recording StartDate StopDate Category Interval
SD Card Status
SD Card Detected SD Card Write Protected To tal Flash Size Available Flash Size
Battery Status
Check Battery
Uptime 0 days,00:00:00 hh:mm:ss
Chart
Figure 4.23 Recording Menu: Status
Web Link Setup
OFF
(yyyy/mm/dd) (yyyy/mm/dd)
Format SD CardStart Recording Status Data Retrieval
StartTime StopTime Type Title
YES NO 1925456(KB) 1925452(KB)
(hh:mm:ss)(24 hours) (hh:mm:ss)(24 hours)
36
Page 43
4.4.6.3 Data Retrieval
This page can be used to retrieve the recorded data from the SD flash card and save it into an ASCII file.
Using the Retrieve EndDate field select the date (year, month, and day) that is the end date for the retrieved data.
Then use Retrieve Period field to identify the duration of the retrieved data; selections are: 1 Day, 1 Week, 1 Month and 1 Year.
The year, month and day dates are taken from your computer.
Now, you can press on the Start Data Retrieval button to log the data into a file. The file is a text document with extension .csv. It can be opened in MS Excel, MS
Wordpad, Notepad or any software package that can read text files. The fields are separated by a comma ( , ) see Figure Figure 4.40 for a sample text file.
Data Retrieval
http://192.168.1.200Address
Overview
Readings
Retrieve End DateMonth Day
Chart
Web Link Setup
Format SD CardStart Recording Status Data Retrieval
January
12010
Network Configuration
Retrieve Period
1 Day
Management Security
Recording
System
Start Data Retrieval
Diagnostics Log Out
Figure 4.24 Recording Menu: Data Retrieval
Refer to Section 4.10.1 for steps that may help you to open the data file in an ASCII or Excel application.
The data recorded on the SD card can be read on a PC or MAC using a USB or any standard card reader. The file format is text and is easily imported to spread sheets and other programs. For details see Section 4.10.
37
Page 44
4.4.6.4 Format SD Card
Clicking on Format SD Card button will start formatting the flash card inside the iServer. Formatting the SD card will erase all data on the card. Allow up to 2 minutes for format to be completed which then it will display “SD Card Formatted”.
Format SD Card
http://192.168.1.200Address
Overview Network Configuration Management Security
Recording
System Diagnostics Log Out
Readings
Chart
While card is being formatted, please wait for message:
Web Link Setup
Format SD CardStart Recording Status Data Retrieval
Format SD Card
Formatting SD Card will erase all data.
Allow up to 2 minutes to format.
“SD Card Formatted” before continuing.
Figure 4.25 Recording Menu: Format SD Card
4.4.7 System
This menu provides various options to reboot, restore defaults, upgrade firmware and download/upload configuration for the iServer. Fields are described below.
4.4.7.1 Reboot
Reboot
http://192.168.1.200Address
Overview Network Configuration Management Security Recording
System
Readings
Chart
DefaultsReboot Download ConfigurationUpgrade Upload Configuration
Web Link Setup
Reboot
Figure 4.26 System: Reboot
Reboot – This option resets the power on the iServer.
If the iServer is recording data the reboot option will stop the recording. It will also set the real date/time back to default unless the Network Time Server option
from Configuration page, is enabled.
38
Page 45
4.4.7.2 Defaults
Clicking on this button will put the iServer back to its default settings.
Defaults
http://192.168.1.200Address
Readings
DefaultsReboot Download ConfigurationUpgrade Upload Configuration
Overview Network Configuration Management Security Recording
System
Chart
Web Link Setup
Restore Factory Defaults
Figure 4.27 System: Defaults
4.4.7.3 Upgrade
Upgrade Firmware
http://192.168.1.200Address
Overview Network Configuration Management Security Recording
System
Diagnostics
Readings
Chart
DefaultsReboot Download ConfigurationUpgrade Upload Configuration
Please make sure no other clients are connected to the device.
Web Link Setup
Filename: iTHX-SD*.bin
Browse...
Upgrade
Consult manual for complete instructions.
Figure 4.28 System: Upgrade
Upgrade – This option loads a new firmware into the iServer over the network. The
firmware is a binary file with .bin extension. This file must be kept on the same PC from which the iServer is accessed.
If you have an iServer with a firmware version previous to 1.09.0, it is mandatory to set
the iServer to default settings after a firmware upgrade to any version, after 1.09.0.
Once the firmware upload is completed the iServer will power reset itself. This typically takes about 30 seconds.
When doing firmware upgrade on the iServer, there should be NO other on-going or new Web server (HTTP) connections made to the device; otherwise, the iServer will not upgrade the firmware. You can do either of the following to prevent this from happening:
1. Connect the iServer directly to the PC using a straight or crossed Ethernet cable. This
will also help to have a faster firmware upload.
2. From the “Network” menu, change the Web Server Port, for example, to 500 (default
is 80), save the change, and reset the power on your iServer. The next time you will access the iServer’s Web server, you must type http://192.168.1.200:500 where
192.168.1.200 is the IP address of the iServer and 500 is the new Web server port (HTTP port).
39
Page 46
4.4.7.3 Upgrade (continued)
Either one of the above options will make you the only user on that iServer. Set the iServer to default settings after the firmware is upgraded. See Section 4.4.7.2 on how to set the iServer to default settings. Applying default settings will change the iServer’s settings to factory defaults. If those settings are important to you and you do not want to reconfigure the iServer, you
will need to do the following steps. From the “System” menu:
1) “Download” the iServer’s configuration settings, Section 4.4.7.4.
2) “Upgrade” the iServer with the new firmware, Section 4.4.7.3.
3) Set the iServer to “Default” settings. The iServer’s IP address is now
192.168.1.200.
4) “Upload” the same configuration you obtained from step 1 onto the iServer, Section 4.4.7.5.
After you have successfully upgraded your device, you can verify the firmware version number on the “Overview” page (see Section 4.4.1).
4.4.7.4 Download Configuration
Clicking on this button will download the configuration file that contains all the settings stored in the iServer and use it as a reference.
Download
http://192.168.1.200Address
Overview Network Configuration Management Security Recording
System
Diagnostics Log Out
Readings
Chart
DefaultsReboot Download ConfigurationUpgrade Upload Configuration
Web Link Setup
Download Configuration File
Figure 4.29 System: Download Configuration
40
Page 47
4.4.7.5 Upload Configuration
Upload Configuration
http://192.168.1.200Address
Overview Network Configuration Management Security Recording
System
Diagnostics Log Out
Readings
DefaultsReboot Download ConfigurationUpgrade Upload Configuration
Please make sure no other clients are connected to the device.
Web Link Setup
Chart
Upload Configuration File *.cfg
Consult manual for complete instructions.
Browse...
Figure 4.30 System: Upload Configuration
Upload Configuration – Using this option to upload the configuration file into the
iServer. It is recommended to download and store a working copy of this configuration file in case the device is configured improperly.
You can set the iServer to its default settings and upload the configuration file back into the iServer.
4.4.8 Diagnostics
This page provides the number of Ethernet packets received or transmitted. Pressing the Reset Counter button will set the counts to zero.
Diagnostics
http://192.168.1.200Address
Overview Network Configuration Management Security Recording System
Diagnostics
Log Out
Readings
Ethernet Received Transmitted
Packets 4437 6073
Chart
Web Link Setup
Reset Counter
Figure 4.31 Diagnostics
4.4.9 Log Out
Once you click on Log Out, you will be logged out, and taken back to the Welcome page.
41
Page 48
4.5 Readings
This page displays temperature, humidity, dew point, and the status of the alarm relays and recording.
It can be displayed either as an HTML page or a Java applet.
4.5.1 HTML
Figure 4.32 Readings: HTML - Shown with Probe 1 and 2 Enabled
Refresh – Adding a time value in seconds will automatically update the HTML page with
fresh readings.
42
Page 49
4.5.2 Java If Java option is chosen installation of Java Runtime is required (refer to Appendix G).
2Refresh
Seconds
Data Logging: INACTIVE
Figure 4.33 Readings: Java - shown with Probe 1 enabled
Data Logging – This option will require the Java Policy file to be setup.
Press on Help[?] for details. An alternate way to set up the Java Policy can be found in Appendix G. Once the Java Policy file is completed, click on Data Logging button to log the data
displayed on this page into a file. The file is in comma separated values (.csv) format, which can be opened in Excel or as ASCII. The fields are separated by a comma ( , ).
When the button is ACTIVE it means that the data is currently being logged into a file. The value in seconds in the Refresh box determines how often the data is being logged. The default is 2 seconds.
If you want to access the “Chart” and/or the “Readings” applet from outside your local network on which the iServer is connected, you must allow ports 1001 (Readings) and 2003 (Chart) on your Router.
If a blank screen appears without any “java application running” or image of a “Java logo”, please verify you have the latest Java Runtime Environment installed and configured according to the instructions (refer to Appendix G). If you do not have Java Runtime Environment installed, you may download it from our website (in the software section) or contact the Customer Service Department nearest you.
“Java and the Java Coffee Cup Logo are trademarks or registered trademarks of Sun Microsystems, Inc. in the U.S. and other countries.”
43
Page 50
4.6 Chart This page charts the temperature, humidity, and dew point in real-time across the full span (-40 to 124ºC, and 0 to 100% RH) or within any narrow range (such as 20 to 30ºC). The time-base can display one minute, one hour, one day, one week, one month or one year.
Figure 4.34 Adjustable Chart
If a blank screen appears without any “java application running” or image of a “Java logo”, please verify you have the latest Java Runtime Environment installed and configured according to the instructions (refer to Appendix G). If you do not have Java Runtime Environment, you may download it from our website (in the software section) or contact the Customer Service Department
nearest you. High / Low – If Alarm Relays are enabled (in the Configuration menu), the set point values for High and Low alarms will be listed and graphed on the chart. Max. / Min. – This shows the maximum and minimum values for Temperature and Humidity (or Dewpoint) indicated on the Y-axis.
44
Page 51
4.6 Chart (continued) Display Selection – The options are Both Probes, Probe 1, and Probe 2 X-axis – This indicates the time, options includes one minute, one hour, one day, one
week, one month or one year.
Y-axis – The left Y-axis is dedicated to temperature. Either humidity or dewpoint can be selected for the right Y-axis.
Data Source – The options are Live and Recorded. If Live is selected the real-time data will be charted for the time indicated on the X-axis. If Recorded is selected the stored data on the SD card will be charted for the time
indicated on the X-axis. Alarm Relay Set Points – This option will hide or chart (normal) the set point values.
The bold option will thicken the displayed set point lines.
If Alarm Relays are disabled (in the Configuration menu) the set points for High and Low alarms will not be graphed or displayed on the chart.
Sensor Type – Choose Humidity or Dewpoint. Show – To select which probes are charted. Graph Line Style – Choose between Normal and Bold. Save Chart – Save the current graph in PNG (Portable Network Graphics) format. The
filename has extension .png. This option will require the Java Policy file to be setup. Press on Help[?] for details. An alternate way to set up the Java Policy can be found in
Appendix G. Print Chart – The displayed chart will be printed once the button is clicked.
If you want to access the “Chart” and/or the “Readings” applet from outside your local network on which the iServer is connected, you must allow ports 1001 (Readings) and 2003 (Chart) on your Router.
4.7 Web Link This page is configured from the Network menu under IP configuration. Refer to Section 4.4.2.1.
Web Link
http://192.168.1.200Address
Readings
Chart
Web Link Setup
Web Link
Figure 4.35 Web Link
45
Page 52
4.8 Telnet Setup
With a telnet program, you can connect to the iServer using its IP address and TCP Port
2000. In Continuous mode (refer to Configuration:Sensors page Section 4.4.3.2), the telnet session will receive continuous data (temperature and humidity) from the iServer. In Command mode, a command can be sent to query the iServer.
Refer to Section 4.9.1 for list of commands.
4.8.1 Telnet Connection
Telnet stands for Telecommunications Network, it is a protocol that provides a way for users to connect to a node with a valid IP address anywhere on an Ethernet network or the Internet . You can open a Telnet session using a terminal emulation programs like Tera Term (downloadable from the internet), which is a free software for Windows.
You can initiate a telnet connection by typing the IP address of the iServer, and Port 2002 for logging into the iServer’s telnet consol port (see Figure 4.36). By telneting to Port 2000, you will be able to access temperature, humidity, and dew point readings.
The default password for Telnet Login is 00000000 and can be changed if
desired.
Te ra Term: New Connection
192.168.1.200
Host:TCP/IP:
UNSPEC
elpOK
S
erial:
ort#:
2002TCP p
lnet
Te
COM1: Communications Port {COM1
Cancel H
Protocol:
Figure 4.36 Tera Term Telnet Connection Screen
Once the connection is established you will be prompted for an Administrator Password (default is 00000000). Telnet console port (2002) allows users to configure some parameters within the iServer. Figure 4.38 shows the parameters that can be changed.
After connecting to the iServer, you can use the following commands to read, modify, and get help.
p displays the current settings (Figure 4.37) ? displays all the available options as well as an example on how to enter a
command (Figure 4.38) s set a new configuration (see the example in Figure 4.38), sending the
command c is required to permanently store the new settings RESET restart the iServer FACTORY set the iServer to it’s factory default, the RESET command is required after
the FACTORY command
c permanently store the new settings q quit the telnet session
46
Page 53
4.8.1 Telnet Connection (continued)
Figure 4.37 F igure 4.38
Telnet to port 2002 – p command Telnet to port 2002 – ? command
4.9 HTTPget Program
The HTTPget software is used to send a single HTTP or TCP request to an iServer product. In contrast, the telnet or Hyperterminal programs allow a continuous connection with multiple requests to be sent to the iServer product.
Generally HTTPget is used for simply programming an IP address to the iServer or for quickly obtaining a reading of from a device.
The iServer product must be configured from the configuration web page so that the “Server Type” value is set to “Command”, refer to Section 4.4.3.2.
47
Page 54
4.9.1 HTTPget using Port 2000
You can setup and read the information from the iServer by using the HTTPget program. The following program can be used to read data from the iServer by using TCP port
2000. The command string is sent to this TCP port, then the response can be read back from the same socket.
The HTTPget.exe file is used to setup and read information from the iServer. This file can be downloaded from the website listed on the cover of this manual (software section).
Notes on using HTTPget : The HTTPget.exe program should be installed to the windows directory (usually c:\winnt
or c:\windows)
1. Open up a command window (or have a DOS window)
a) Click on start menu b) Click on “Run” c) In the dialog box that appears, type “cmd” or “command” and click on “OK” button. d) A command window should now appear.
2. If you now type “httpget” and press the “enter” key, the program options should be
displayed.
3. Next run HTTPget with the options displayed below
httpget -r -S “*SRTC\r” 192.168.1.200:2000
where:
-r –S are parameters needed for the command string *SRTC is the command:
\r is the carriage return termination character
192.168.1.200 is an IP address 2000 is a socket port number
Response:
076.6 (in Deg.F format)
*SRT/*SRTC Get Temperature Reading in Celsius from Probe 1 *SRT2/*SRTC2 Get Temperature Reading in Celsius from Probe 2 *SRT3/*SRTC3 Get Temperature Reading in Celsius from Probe 1 and 2;
separated by a semi-colon
*SRTF Get Temperature Reading in Fahrenheit from Probe 1 *SRTF2 Get Temperature Reading in Fahrenheit from Probe 2 *SRTF3 Get Temperature Reading in Fahrenheit from Probe 1 and 2;
separated by a semi-colon
*SRH Get Humidity (%) Reading from Probe 1 *SRH2 Get Humidity (%) Reading from Probe 2 *SRH3 Get Humidity (%) Reading from Probe 1 and 2; separated by a
semi-colon
*SRD/*SRDC Get Dewpoint Reading in Celsius from Probe 1 *SRD2/*SRDC2 Get Dewpoint Reading in Celsius from Probe 2 *SRD3/*SRDC3 Get Dewpoint Reading in Celsius from Probe 1 and 2; separated
by a semi-colon
48
Page 55
4.9.1 HTTPget using Port 2000 (continued) *SRDF Get Dewpoint Reading in Fahrenheit from Probe 1 *SRTD2 Get Dewpoint Reading in Fahrenheit from Probe 1 *SRDF3 Get Dewpoint Reading in Fahrenheit from Probe 1 and 2;
separated by a semi-colon *SRB/*SRBC Get Temperature and Humidity Reading in Celsius from Probe 1; separated by a semi-colon *SRB2/*SRBC2 Get Temperature and Humidity Reading in Celsius from Probe 2; separated by a semi-colon *SRB3/*SRBC3 Get Temperature and Humidity Reading in Celsius from Probe 1 and 2; separated by a semi-colon *SRBF Get Temperature and Humidity Reading in Fahrenheit from Probe 1; separated by a semi-colon *SRBF2 Get Temperature and Humidity Reading in Fahrenheit from Probe 2; separated by a semi-colon *SRBF3 Get Temperature and Humidity Reading in Fahrenheit from Probe 1 and 2; separated by a semi-colon
*SRYS Indicate AC or DC power and battery status *SRYRST Reset Power on *SR_XD1 Disable alarm condition check for relay 1 *SR_XD2 Disable alarm condition check for relay 2 *SR_XC2 Reset relay 2 latch condition *SR_XC1 Reset relay 1 latch condition *SR_XC2 Reset relay 2 latch condition *SR_XE1 Re-enable alarm condition check for relay 1 (must be
executed after *SR_XD1) *SR_XE2 Re-enable alarm condition check for relay 2 (must be executed after *SR_XD2)
*SR_XP1 Close relay 1 for < 200 mSec *SR_XP2 Close relay 2 for < 200 mSec *SR_XH1 Close relay 1 *SR_XH2 Close relay 2 *SR_XL1 Open relay 1 *SR_XL2 Open relay 2 *SR_XR1 Read current state of relay 1 *SR_XR2 Read current state of relay 2
Commands developed specifically for DASYLab/LabView
*SRAC Get All Readings in below format in Celsius *SRAF Get All Readings in below format in Fahrenheit Format:
T1[+/-]###.#H1[+/-]###.#D1[+/-]###.#T2[+/-]###.#H2[+/-]###.#D2[+/]###.#;[EndCharacter]
*SRACS Get All Readings in below format in Celsius *SRAFS Get All Readings in below format in Fahrenheit Format:
[+/-]###.# [+/-]###.# [+/-]###.# [+/-]###.# [+/-]###.# [+/-]###.#;[EndCharacter]
49
Page 56
4.10 Flash Card Reader
The data recorded on the SD card can be read on a PC or MAC with a USB or any standard card reader.
The format is a simple “.csv” text file that is easily imported in MS Excel or MS WordPad or any software package that can read text files (see Figure 4.40).
The entire recording session is saved into a file kept in the root directory. The name of this file is S followed by the date the recording was started, so for example:
S20100401.txt is April 1st, 2010. The same recording data is also saved as daily files located in a series of month and
year directories in which the recording took place (see Figure 4.39).
E:\
ile Edit View Favorites Tools Help
F
Name
2010 Events S20100401.txt 15,598KB
E:\2010
ile Edit View Favorites Tools Help
F
Name
04 05
E:\2010\04
F
ile Edit View Favorites Tools Help
Name
20100401.txt
20100402.txt
20100403.txt
20100404.txt
20100405.txt
20100406.txt
20100407.txt
Type Date ModifiedSize
File Folder File Folder Te xt Document
Type Date ModifiedSize File Folder
File Folder
598 KB 598 KB 598 KB 598 KB 598 KB 598 KB 598 KB
4/01/2010 8:08AM 4/01/2010 8:08AM 5/02/2010 1:08PM
Type Date ModifiedSize
Te xt Document Te xt Document Te xt Document Te xt Document Te xt Document Te xt Document Te xt Document
4/01/2010 8:08AM 5/02/2010 1:08PM
4/02/2010 12:00AM 4/03/2010 12:00AM 4/04/2010 12:00AM 4/05/2010 12:00AM 4/06/2010 12:00AM 4/07/2010 12:00AM 4/08/2010 12:00AM
Figure 4.39 Recorded Data File Directory
#2010/04/02;09:00:00;23.3C;35.6%;D6.7C; #2010/04/02;09:00:10;23.3C;35.6%;D6.7C; #2010/04/02;09:00:20;23.3C;35.6%;D6.7C; #2010/04/02;09:00:30;23.3C;35.7%;D6.7C; #2010/04/02;09:00:40;23.3C;35.7%;D6.7C;
Figure 4.40 Example of data recorded
on the SD card
50
Page 57
4.10.1 Opening Older iTHX-SD .txt Data Files
1. Right-click on the data file and select Open with Wordpad. This will display the
ASCII data in WordPad.
2. From the Edit menu select the Replace option to replace all the ; (semicolons)
with , (commas).
3. From the File menu select Save As and save the file with an extension .csv (Comma
Separated Value).
4. This .csv file can now be opened in Excel with the correct format.
Alternatively, you can do the following--
1 Open Excel. Go to File, Open, select Files of Type: Choose Text files (*.prn, *.txt,
*.csv), and then select the data file you wish to open.
2. A Text Import Wizard should pop-up. Select Delimited type of file, then Next. Click on semicolon as the Delimiter type, then Next. The screen should show how the file is being divided into columns, click on Finish. 3 From the File menu select Save As and save the file with an extension .csv. The next time you open this file it will open in Excel automatically.
You may remove the letters C or F and D from the recorded values (using the Replace option of Excel or Wordpad). This will enable you to create charts of the data.
51
Page 58
4.11 iLog Software
This is an Excel application software that can log temperature, humidity and dewpoint from the iServer over the local network (Ethernet) or the internet.
a) Download the iLog software from the website listed in this manual. b) Install iLog software on a networked PC. This software is compatible with Windows
95, 98, NT, 2000, XP, Windows Vista and Windows 7 (32 and 64-bit).
c) If you have Excel 2007 or higher, when installing iLog choose “Custom” installation
option and on the next window check the box for “Excel 2007 Apps” and continue
the installation to the end. d) For complete information of how to use the iLog software, click on the HELP button. e) There is a list of Error Messages in Appendix F. f) For single probe units use iLog -- 1.9.4;
for dual probe units use iLog - iTHX-W Dual Probe
Figure 4.41 iLog Software Logging Data
The Options button will let you select the product model. Choose iTHX-SD from the Model Selection pull-down menu (from General tab).
52
Page 59
iLog Software (continued) Table 4.1 iLog Excel Applications
The iLog application actually consists of several Excel files, though most supported devices can be accessed by the main iLog program. The main program is listed as “iLog”, plus a version number, under the Start Menu program links (those links available by clicking the Start button on the Windows taskbar).
In the following table, the iLog File column shows how many sensors are logged by the main iLog file. If more than 3 sensors are available, then the Alternate column shows how many sensors the device can support. Devices with more than 3 sensors will have their own Excel file. For these files, click on the model device found in the Start Menu.
Networked Product iLog File Alternate
iTHX-W, iTHX-2 3 column iTHX-M, iTHX-SD 3 column iTHX-W Dual Probe 1st probe / 3 column 6 column iSE-TC, iSD-TC 3 column 5 column iSE-TH, iSD-TH 3 column 5 column iBTHX-W, iBTHX-D 3 column 4 column iBTX-W/-D, iBTX-M, iBTX-SD 2 column iPTX-W 2 column iTCX 3 column iTH Controller 3 column iVI 3 column 4 column iSeries 1 column iDRX/iDRN 1 column INF-B 1 column
Wireless Devices
The “Auto” column shows number of columns per remote device that are possibly displayed.
The Full column shows number of columns allocated per device for the “Full” spreadsheet, which will be able to display all data for all active devices.
Networked Product Auto Full
zSeries Receiver and Remotes 1 to 4 column / device 4 column / 32 device wiSeries with zED Remotes 1 to 2 column / device 2 column / 32 device UWTC REC-3 and Remotes 1 or 2 column / device 2 column / 32 device wiSeries with UWTC Remotes 1 to 2 column / device 2 column / 32 device
The active wireless devices, when shown in the Excel application, will be shown with the device number and the units returned.
53
Page 60
PART 5
SPECIFICATIONS
SENSOR SPECIFICATIONS Relative Humidity (RH)
Accuracy/Range at 25ºC ±2.75% for 10 to 90%;
- Non-Condensing:
Hysteresis: ±1% RH Non-linearity: ±3% Response Time: 8 seconds, tau 63% (time for reaching 63% of a step
Repeatability: ±0.1% Resolution: 0.1%, 12bit
Temperature (T)
Accuracy/Range*: Wand Probe: ±0.5°C for 5° to 45°C (±1°F for 41° to 113°F);
Industrial Probe:
Up Refer to chart in Appendix H *Note: extended temperature range is for Industrial Probe
Response Time: 5 to 30 seconds, tau 63% (response time depends on heat
Repeatability: ±0.1°C Resolution: 0.1°C, 14 bit
±3% for 5 to 10% and 90 to 95% ±4% for 0 to 5% and 95 to 100% Refer to chart in Appendix H
function, valid at 25C and 1m/s airflow)
NOTE: Reconditioning of the probe may be necessary if the probe is stored for a period of time in a harsh environment (temperatures below 0°C or above 70°C or exposure to chemical vapors, condensation, etc). To recondition the probe refer to Appendix H.4
Up to ±1°C for 0° to 5°C and 45° to 70°C Up to ±2°F for 32° to 41°F and 113° to 158°F) Refer to chart in Appendix H
±0.5°C for 5° to 45°C ( ±1°F for 41 to 113°F);
Up to ±1.5°C for -40° to 5°C and 45° to 124°C
to ±2.7°F for -40° to 41°F and 113° to 255°F)
only, the iServers operating temperature is 0 to 60°C
capacity of and thermal resistance to sensor substrate)
PROBE SPECIFICATIONS
Wand Probe: 198 mm long x 19 mm diameter (7.8” x 0.75”)
Cable with DB9 connector: 152 mm long (6”) Cable operating temperature: 0 to 80°C (32 to 176°F) Industrial Probe: 137mm or 51mm long x 16mm diameter (5” or 2” x 0.63”) Cable with DB9 connector: 3m or 0.9m long (10 or 3) Cable operating temperature: -40 to 125°C (-40 to 257°F)
54
Page 61
iSERVER SPECIFICATIONS
Interfaces Ethernet (RJ45): Fixed or auto-negotiating 10/100Base-T; Sensor: Digital 4-wire (DB-9) Supported Protocols: TCP, UDP, SNMP, SMTP,
DHCP, HTTP, and Telnet
LED Indicators: 100 BASE-T, Network Link and Activity, DHCP, Recording LCD Display: 16 digits, 6mm (0.23”) Memory Data Flash Card: 2 GB SD Flash Card for 8 months of data storage at
1 second recording intervals, or 7 years at 10 seconds
Relay Outputs: Two relays 1.5A @ 30Vdc Management: Device configuration and monitoring through embedded
WEB server
Embedded WEB Server: Serves WEB pages containing real-time data and live
updated charts within definable time intervals.
POWER
Power Input: 9 to 12 Vdc Consumption: 3.5 W. Safety Qualified ac Power Adapter (included) Nominal Output: 9 Vdc @ 0.5 A
Input: 100 to 240 Vac, 50/60Hz
Back-up Battery: 9 Vdc, alkaline. 30 hours at 10 second recording intervals,
or 60 hours at 1 minute intervals
ENVIRONMENTAL
Operating Temperatures iServer Unit: 0 to 60°C (32 to 140°F) Battery: -18 to 55°C (-0.4 to 131°F) ac Power Adapter: 0 to 40°C (32 to 104°F) Wand Cable: 0 to 80°C (32 to 176°F) Industrial Cable: -40 to 125°C (-40 to 257°F) Storage Temperature: -40 to 85°C (-40 to 185°F) Material: Steel metal case with wall mount bracket iServer Housing Dimensions: 36.6H x 62W x 90D mm (1.44 x 2.93 x 3.56 in) Weight: 389 g (0.86 lbs.), including battery
SNTP, ARP, ICMP, DNS,
Auto MDI/MDIX
GENERAL
Agency Approvals: FCC-B, CE Software: Compatible with Windows operating systems Firmware upgrade from Ethernet OPC Server iConnect: Configuration software for the Ethernet interface iLog: Macro for automatic data logging in MS Excel
55
Page 62
Appendix A GLOSSARY
User of this manual should be familiar with following definitions: ARP (Address Resolution Protocol) is a protocol for mapping an Internet Protocol address
(IP address) to a physical machine address that is recognized in the local network. For example, the IP address in use today is an address that is 32-bits long. In an Ethernet local area network, however, addresses for attached devices are 48-bits long. (The physical machine address is also known as a Media Access Control or MAC address.) A table, usually called the ARP cache, is used to maintain a correlation between each MAC address and its corresponding IP address. ARP provides the protocol rules for making this correlation and providing address conversion in both directions.
Ethernet is a network protocol defined by the IEEE 802.3 standard. Ethernet-based networks use MAC Address rather then IP Address to exchange data between computers. By using ARP and adding TCP/IP support, Ethernet devices may be connected as part of the Internet. An Ethernet LAN typically uses coaxial cable or special grades of twisted pair wires. The most commonly installed Ethernet systems are called 10BASE-T and provide transmission speeds up to 10 Mbps. Devices are connected to the cable and compete for access using a Carrier Sense Multiple Access with Collision Detection (CSMA/CD) protocol.
IP (Internet Protocol) is the method or protocol by which data is sent from one computer to another on the Internet.
IP address (Internet Protocol address) is a 32-bit number that identifies each sender or receiver of information that is sent in packets across the Internet.
IP Netmask is a 32-bit pattern of bits used to determine which part of the IP address is the network portion and which part is the host portion.
MAC (Media Access Control) Address is your computer’s unique hardware number. When you’re connected to the Internet from your computer, a correspondence table relates your IP address to your computer’s physical (MAC) address on the LAN.
Ping is a utility that tests the network connectivity. It is used to determine if the host is capable of exchanging information with another host.
Port number/Socket number is a way to identify a specific process to which an Internet or other network message is to be forwarded when it arrives at a server. It is a predefined address that serves as a route from the application to the Transport layer or from the Transport layer to the application of the TCP/IP system.
Sockets are a method for communication between a client program and a server program in a network and defined as “the endpoint in a connection.” Information transferred across the Internet primarily occurs between sockets.
SMTP Simple Mail Transfer Protocol is an Internet standard for electronic mail (email) transfer across the Internet. SMTP clients usually use SMTP to send email messages by specifying the SMTP server. The email server uses SMTP to both send and receive email messages.
SNMP Simple Network Management Protocol is a network monitoring protocol to monitor devices connected to an Ethernet Network.
TCP/IP (Transmission Control Protocol/Internet Protocol) is the basic communication language or protocol of the Internet. When you are set up with direct access to the Internet, your computer is provided with a copy of the TCP/IP program just as every other computer that you may send messages to or get information from also has a copy of TCP/IP. TCP/IP often is used as a general term to indicate generic access to the Internet.
UDP/IP (User Datagram Protocol/Internet Protocol) is the TCP/IP standard protocol that allows an application program on one machine to send a datagram to an application program on another.
data to every node on the same network. The Directed UDP transmits data to one node only.
The UDP can be either in Broadcast or Directed form. The Broadcast UDP transmits
56
Page 63
Appendix B IP Address
An IP address is a unique 32-bit address assigned to a computer and includes:
a A network ID number identifying a network. a A host ID number identifying a computer on the network.
All IP addresses have been divided into three smaller groups (classes) A, B and C
a Class A addresses have 8-bits of network ID and 24-bits of host ID. They can support
a large number of hosts, approximately 2 = 16,777,216 computers per network.
The IP addresses range in binary from 00000001.xxxxxxxx.xxxxxxxx.xxxxxxxx
to 01111111.xxxxxxxx.xxxxxxxx.xxxxxxxx
The IP addresses range in decimal from 1.x.x.x to 127.x.x.x
Class A network ID’s support a very large number of hosts.
a Class B addresses have 16-bits of network ID and 16-bits of host ID. They can
support approximately 216 = 65,536 computers per network.
The IP addresses range in binary from 10000000 00000000.xxxxxxxx.xxxxxxxx to 10111111 11111111.xxxxxxxx.xxxxxxxx
The IP addresses range in decimal from 128.0.x.x TO 191.255.xxx.xxx
Class B network ID’s support a medium number of hosts.
a Class C addresses have 24-bits of network ID and 8-bits of host ID. They can support approximately 28 = 256 computers per network.
The IP addresses range in binary from 11000000.00000000.00000000.xxxxxxxx to 11011111.11111111.11111111.xxxxxxxx
The IP addresses range in decimal from 192.0.0.xxx to 223.255.255.xxx
Class C network ID’s support a small number of hosts.
The rest of the addresses are divided into two classes, D and E. Class D networks are not assigned to the host. They are used for multicasting. The address range from 224.x.x.x to 239.x.x.x
Class E networks are experimental or reserved addresses. The address range from 240.x.x.x to 247.x.x.x
57
Page 64
Appendix C ARP Commands
ARP is the Internet layer protocol responsible for matching or obtaining the MAC (hardware) address that corresponds to a particular IP address. The ARP command allows the user to view the current contents of the ARP cache of the local computer (residing on the same network). Microsoft includes the ARP.EXE utility for viewing and modifying the ARP cache with its Windows products. The following ARP commands can be used to view cache entries:
a arp –a Use this command to view all ARP cache entries. a arp –a plus IP address Use this command to view ARP cache entries associated
with one particular interface on a network with multiple adapters.
a arp –g Same as arp –a. a arp –N Use this command to display ARP entries for specific network interface. a arp – s plus IP address plus Physical address Use this command to manually
add a permanent static entry to the ARP cache.
a arp –d Use this command to manually delete a static entry.
Ping the destination computer using IP address first before using the arp -a command.
The following window shows examples of arp commands and responses.
a Your computer has an IP address of 192.168.1.118 a The destination computer has an IP address of 192.168.1.96
Figure C.1 ARP Commands and Responses
58
Page 65
Appendix D IP Netmask
IP Netmask or Subnet Mask is a 32-bit pattern of ones and zeros used to determine network portion of an IP address from the host portion of the IP address. Subnet mask is a network ID that is created by borrowing bits from host portion of IP address and using them as part of a network ID. The table below shows a default subnet mask for address Classes A, B, and C. Each bit that is set to “1” in the subnet mask corresponds to the bit in the IP address that is to be used as the network ID. Each bit that is set to “0” in the subnet mask corresponds to a bit in the IP address that is to be used as the host ID.
Address Class Mask Binary Value
or Dotted Notation
Class A 255.0.0.0 Class B 255.255.0.0 Class C 255.255.255.0
11111111 11111111 11111111
00000000 11111111 11111111
00000000 00000000 11111111
Mask Decimal Value
00000000 00000000 00000000
If your network requires more network ID’s, you can extend the default subnet mask to include additional bits from the host ID. This allows for additional network ID’s within the network. The table below shows some examples of subnet masks and bits moved from the hosts ID to create a new subnet.
Mask Dotted Notation Mask Binary Mask Bits
Class A
255.0.0.0 (Default) 0
255.192.0.0 2
255.224.0.0 3
255.240.0.0 4
255.248.0.0 5
255.252.0.0 6
255.254.0.0 7
255.255.0.0 8
255.255.128.0 9
255.255.192.0.0 10
……………......... .
255.255.255.252 22
11111111 11111111 11111111 11111111 11111111 11111111 11111111 11111111 11111111 11111111
........
11111111
00000000 11000000 11100000 11110000 11111000 11111100 11111110 11111111 11111111 11111111
........
11111111
00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 10000000 11000000
........
11111111
00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000
........
11111100
Class B
255.255.0.0 (Default) 0
255.255.192.0 2
……………......... .
255.255.255.252 14
11111111 11111111
........
11111111
11111111 11111111
........
11111111
00000000 11000000
........
11111111
00000000 00000000
........
11111100
Class C
255.255.255.0 (Default) 0
255.255.255.192 2 …………………. .
255.255.255.254
11111111 11111111
........
11111111
11111111 11111111
........
11111111
11111111 11111111
........
11111111
00000000 11000000
........
11111100
6
To determine the number of valid hosts ID’s remaining after subnetting, use the following equation: 2n – 2, where n is the number of octet digits left after the subnet mask.
59
Page 66
Appendix E ASCII Chart
ASCII Dec Hex Binary ASCII Dec Hex Binary
Char No Parity Char No parity
NUL 00 00 00000000 @ 64 40 01000000 SOH 01 01 00000001 A 65 41 01000000 STX 02 02 00000010 B 66 42 01000010 ETX 03 03 00000011 C 67 43 01000011 EOT 04 04 00000100 D 68 44 01000100 ENQ 05 05 00000101 E 69 45 01000101 ACK 06 06 00000110 F 70 46 01000110 BEL 07 07 00000111 G 71 47 01000111 BS 08 08 00001000 H 72 48 01001000 HT 09 09 00001001 I 73 49 01001001 LF 10 0A 00001010 J 74 4A 01001010 VT 11 0B 00001011 K 75 4B 01001011 FF 12 0C 00001100 L 76 4C 01001100 CR 13 0D 00001101 M 77 4D 01001101 SO 14 0E 00001110 N 78 4E 01001110 SI 15 0F 00001111 O 79 4F 01001111 DLE 16 10 00010000 P 80 50 01010000 DC1 17 11 00010001 Q 81 51 01010001 DC2 18 12 00010010 R 82 52 01010010 DC3 19 13 00010011 S 83 53 01010011 DC4 20 14 00010100 T 84 54 01010100 NAK 21 15 00010101 U 85 55 01010101 SYN 22 16 00010110 V 86 56 01010110 ETB 23 17 00010111 W 87 57 01010111 CAN 24 18 00011000 X 88 58 01011000 EM 25 19 00011001 Y 89 59 01011001 SUB 26 1A 00011010 Z 90 5A 01011010 ESC 27 1B 00011011 [ 91 5B 01011011 FS 28 1C 00011100 \ 92 5C 01011100 GS 29 1D 00011101 ] 93 5D 01011101 RS 30 1E 00011110 ^ 94 5E 01011110 US 31 1F 00011111 _ 95 5F 01011111 SP 32 20 00100000 ` 96 60 01100000 ! 33 21 00100001 a 97 61 01100001 34 22 00100010 b 98 62 01100010 # 35 23 00100011 c 99 63 01100011 $ 36 24 00100100 d 100 64 01100100 % 37 25 00100101 e 101 65 01100101 & 38 26 00100110 f 102 66 01100110 39 27 00100111 g 103 67 01100111 ( 40 28 00101000 h 104 68 01101000 ) 41 29 00101001 I 105 69 01101001 * 42 2A 00101010 j 106 6A 01101010 + 43 2B 00101011 k 107 6B 01101011 , 44 2C 00101100 l 108 6C 01101100
- 45 2D 00101101 m 109 6D 01101101 . 46 2E 00101110 n 110 6E 01101110
60
Page 67
Appendix E ASCII Chart Continuation
/ 47 2F 00101111 o 111 6F 01101111 0 48 30 00110000 p 112 70 01110000 1 49 31 00110001 q 113 71 01110001 2 50 32 00110010 r 114 72 01110010 3 51 33 00110011 s 115 73 01110011 4 52 34 00110100 t 116 74 01110100 5 53 35 00110101 u 117 75 01110101 6 54 36 00110110 v 118 76 01110110 7 55 37 00110111 w 119 77 01110111 8 56 38 00111000 x 120 78 01111000 9 57 39 00111001 y 121 79 01111001 : 58 3A 00111010 z 122 7A 01111010 ; 59 3B 00111011 { 123 7B 01111011 < 60 3C 00111100 | 124 7C 01111100 = 61 3D 00111101 } 125 7D 01111101 > 62 3E 00111110 ~ 126 7E 01111110 ? 63 3F 00111111 DEL 127 7F 01111111
ASCII Control Codes
ASCII Dec Hex Ctrl Key Definition ASCII Dec Hex Ctrl Key Definition Char Equiv. Char Equiv.
NUL 00 00 Crtl @ Null Character DC1 17 11 Crtl Q Data Control 1
- XON SOH 01 01 Crtl A Start of DC2 18 12 Crtl R Data Control 2 Header STX 02 02 Crtl B Start of Text DC3 19 13 Crtl S Data Control 3
- XOFF ETX 03 03 Crtl C End of Text DC4 20 14 Crtl T Data Control 4 EOT 04 04 Crtl D End of NAK 21 15 Crtl U Negative Transmission Acknowledge ENQ 05 05 Crtl E Inquiry SYN 22 16 Crtl V Synchronous Idle ACK 06 06 Crtl F Acknowledge ETB 23 17 Crtl W End of Trans Block BEL 07 07 Crtl G Bell CAN 24 18 Crtl X Cancel BS 08 08 Crtl H Back Space EM 25 19 Crtl Y End of Medium HT 09 09 Crtl I Horizontal SUB 26 1A Crtl Z Substitute Tabulation LF 10 0A Crtl J Line Feed ESC 27 1B Crtl [ Escape VT 11 0B Crtl K Vertical FS 28 1C Crtl \ File Separator Tabulation FF 12 0C Crtl L Form Feed GS 29 1D Crtl ] Group Separator CR 13 0D Crtl M Carriage RS 30 1E Crtl | Record Return Separator SO 14 0E Crtl N Shift Out US 31 1F Crtl _ Unit Separator SI 15 0F Crtl O Shift In SP 32 20 Space DLE 16 10 Crtl P Data Link Escape
61
Page 68
Appendix F iLog Error Messages
Table F-1 iLog Error Messages
Error # Description Note
-8003 User stopped logging readings.
-10005 Failed to find the iServer. Ethernet cable is disconnected,
iServer is powered off, connections across the firewall require longer “connection to socket time out” setting.
-10006 Windows socket was closed.
-10007 Windows socket error. Wrong IP or wrong Port number
was used.
-10008 The iServer failed to respond to a request. Wrong IP or wrong Port number
was used.
-10011 Response came empty. No data was sent.
-10012 Device responded with Possibly the iLog is configured for
“Serial Time Out” string. wrong product model.
-10014 Terminal Server Mode when the Port is 1000. Try Port 2000 in iLog configuration.
-15100 Error on obtaining the temperature reading. Possibly the iLog is configured for
wrong product model.
-15105 Error on obtaining the humidity reading. Possibly the iLog is configured for
wrong product model.
-15110 Error on obtaining the dew point reading. Possibly the iLog is configured for
wrong product model.
62
Page 69
Appendix G Java Runtime Environment Setup
If your computer does not have Java installed, please download from http://java.sun.com. You can change the Java setting by clicking its icon in Control Panel. To load the applet, you have to enable the web browser and disable cache.
G.1 Java Runtime Environment 1.7 Setup instructions
1. Go to your computer’s Control Panel. Open the Java Plug-in
Verify that the path is correct.
Figure G.1 Java 1.7 Screen Shot
63
Page 70
G.2 Browser Proxy Selection
Accessing iServer within your internal network
W Usually when the computer and iServers are on an internal network, you will not use
Proxy server access.
W You should un-check the “Use Browser Settings” option on the “Proxy” tab.
Accessing iServers units using the internet
W Often the web browser will use Proxy server access to the internet. In such cases,
the default Java runtime settings on the “Proxy” tab should suffice. The default setting is the “Use Browser Settings” option.
W If the default proxy setting doesn’t work, then you may have a situation where the
proxy settings of the web browser are incorrect.
Diagnostics:
If the web page of the iServer appears, then the HTTP Proxy is working fine. If the data isn’t updated on the iServer upon selecting the “Readings” or “Chart” web
page, there may be a problem with access through a winsock proxy server. In such cases your network administrator will need to provide the winsock proxy server and port #s. .
These values should be entered into the Socks line on the “Proxy” tab (of the Java Plugin control panel) or into the “connections” tab on the View,Internet Options dialog and make sure that the Proxy tab shows that the “Use Browser Settings” option is not selected (i.e. when you specify proxy connections in the Java Plugin control panel).
Accessing iServer units over Peer-to-Peer network
A simple peer-to-peer network is setup by disconnecting from the main network (as users will often do when trying to do initial setup of the iServer) and then connecting the iServer to another computer using an ethernet hub, an ethernet switch, or a Cross-over cable connection.
Often when using a peer-to-peer network, the Java plugin and the web browser (such as Internet Explorer) have internet connections configured to connect through proxy servers. In such case, you will need to simply assign the final IP address on this peer to peer network and then view the iServer charts after connecting the iServer into the regular network. Otherwise you can disable the Java plug-in’s “Use Browser Settings” temporarily and then reconfigure the Java plug-in settings for regular network access after testing the iServer chart access on your peer-to-peer network.
The “Use Browser Settings” should not be selected. And the HTTP and Socks proxy entries should be blank. Also, setup the internet browser so that proxy servers are disabled.
64
Page 71
Appendix H Sensor Information
H.1 Accuracy
RH Accuracy Chart Temperature Accuracy Chart
Figure H.1 Figure H.2
Accuracies are tested at Manufacturer’s Outgoing Quality Control at 25°C (77°F) and
3.3V. Values exclude hysteresis and non-linearity, and is only applicable to noncondensing environments.
H.2 Operating Conditions
Sensor works stable within recommended normal range – see Figure. Long term exposures to conditions outside normal range, especially at humidity >80%RH, may temporarily offset the RH signal (+3 %RH after 60h). After return to normal range it will slowly return towards calibration state by itself. See Section H.4 “Reconditioning Procedure” to accelerate eliminating the offset. Prolonged exposure to extreme conditions may accelerate aging.
Figure H.3 Normal Range
H.3 Storage Conditions and Handling Instructions
It is of great importance to understand that a humidity sensor is not a normal electronic component and needs to be handled with care.
Chemical vapors at high concentration in combination with long exposure times may offset the sensor reading. For these reasons it is recommended to store the sensors in original packaging including the sealed ESD bag at following conditions: Temperature shall be in the range of 10°C – 50°C (0 – 80°C for limited time) and humidity at 20 – 60%RH (sensors that are not stored in ESD bags). For sensors that have been removed from the original packaging we recommend to store them in ESD bags made of PE-HD8.
In manufacturing and transport the sensors shall be prevented of high concentration of chemical solvents and long exposure times. Out-gassing of glues, adhesive tapes and stickers or out-gassing packaging material such as bubble foils, foams, etc. shall be avoided. Manufacturing area shall be well ventilated.
65
Page 72
Appendix H Sensor Information (continued)
H.4 Reconditioning Procedure
As stated above extreme conditions or exposure to solvent vapors may offset the sensor. The following reconditioning procedure may bring the sensor back to calibration state:
Baking: 100 – 105°C at < 5%RH for 10h Re-Hydration: 20 – 30°C at ~ 75%RH for 12h.
(75%RH can conveniently be generated with saturated NaCl solution. 100 – 105°C correspond to 212 – 221°F, 20 – 30°C correspond to 68 – 86°F)
H.5 Temperature Effects
Relative humidity reading strongly depends on temperature. Therefore, it is essential to keep humidity sensors at the same temperature as the air of which the relative humidity is to be measured. In case of testing or qualification the reference sensor and test sensor must show equal temperature to allow for comparing humidity readings.
The packaging of sensor is designed for minimal heat transfer from the pins to the sensor. Still, if the sensor shares a PCB with electronic components that produce heat it should be mounted in a way that prevents heat transfer or keeps it as low as possible. Furthermore, there are self-heating effects in case the measurement frequency is too high.
H.6 Light
The sensor is not light sensitive. Prolonged direct exposure to sunshine or strong UV radiation may age the housing.
H.7 Materials Used for Sealing / Mounting
Many materials absorb humidity and will act as a buffer increasing response times and hysteresis. Materials in the vicinity of the sensor must therefore be carefully chosen. Recommended materials are: Any metals, LCP, POM (Delrin), PTFE (Teflon), PE, PEEK, PP, PB, PPS, PSU, PVDF, PVF. For sealing and gluing (use sparingly): Use high filled epoxy for electronic packaging (e.g. glob top, underfill), and Silicone.
Out-gassing of these materials may also contaminate the sensor (see Section H.3). Therefore try to add the sensor as a last manufacturing step to the assembly, store the assembly well ventilated after manufacturing or bake at 50°C for 24h to outgas contaminants before packing.
66
Page 73
Appendix J SNMP
Table J-1 SNMP MIB-2
Version: SNMPv1 Private Enterprise Number: .1.3.6.1.4.1.34089 MIB-2 (RFC 1213) supported: System Group
Object Identifier Object Description Access Default (OID)
.1.3.6.1.2.1.1.1 sysDescr The name and version Read-only iTHX-SD/ X.XX identification of the product X.XX is the firmware version
.1.3.6.1.2.1.1.2 sysObjectID The authoritative identification Read-only .1.3.6.1.4.1.34089.X.X of the network management X.X is the product subsystem contained in the system object ID product. This value is allocated within the SMI enterprises subtree (1.3.6.1.4.1).
.1.3.6.1.2.1.1.3 sysUpTime The time since the system Read-only was last re-initialized.
.1.3.6.1.2.1.1.4 sysContact The contact person for this Read-write None node, together with information on how to contact this person.
.1.3.6.1.2.1.1.5 sysName The hostname of this node. Read-write ithxXXXX XXXX are the last 4 characters of the MAC address
.1.3.6.1.2.1.1.6 sysLocation The physical location of Read-write None this node.
.1.3.6.1.2.1.1.7 sysServices The set of services that this Read-only 76 product primarily offers. Services include The value is a sum of network application, end-to-end layer services implemented. and internet
Table J-2 SNMP Trap
SNMP Trap (RFC 1157) version: SNMPv1 Generic trap type: enterpriseSpecific(6)
Specific Trap Alarm Description Specific Trap Alarm Description
0 Reading 1 7 SD card 10% memory left 1 Reading 2 8 Low battery 2 Reading 3 9 Relay 1 activated 3 Reserved 10 Relay 2 activated 4 Power reset 11 Reserved 5 Sensor disconnected 12 Reserved 6 Recording stopped
67
Page 74
PART 6 APPROVALS INFORMATION
6.1 CE APPROVAL
This product conforms to the EMC directive 89/336/EEC amended by 93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and laboratory.
Basic Insulation Pollution Degree 2 Dielectric withstand Test per 1 min
W Input Power to Sensor Metal Body: none W Input Power to Ethernet Output: 1500Vac W Input Power to Relays: 1500Vac W Ethernet Output to Relays: 1500Vac
Measurement Category I
Category I are measurements performed on circuits not directly connected to the Mains Supply (power). Unit measures Air Temperature and Humidity.
Transients Overvoltage Surge (1.2/50uS Pulse)
W Input Power: 500V Transients Overvoltage W Ethernet: 1500V Transients Overvoltage
Note: The ac/dc power adapter must have Safety Qualified Agency Approvals for CE with Double Insulation rating.
The ac/dc power adapter is 9Vdc. The minimum output current rating is 500mA.
EMC EN61000-6-1:2001 (Immunity) and EN61000-6-3:2001 (Emissions) Immunity requirements for residential, commercial and light-industrial environments
W EMC Emissions Table 1, Class B W EMC Immunity Table 1: Enclosure
Table 2: Signal Lines Ports Table 3: Dc input/Dc output Ports
EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement, control and laboratory.
W EMC Emissions Table 4, Class B of EN61326 W EMC Immunity Table 1 of EN61326
Note: I/O lines / sensor cables require shielded cables and these cables must be located on conductive cable trays or in conduits.
Refer to the EMC and Safety installation considerations (Guidelines) of this manual for additional information.
6.2 FCC
This device complies with Part 15, Subpart B, Class B of the FCC rules.
68
Page 75
NOTES
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
______________________________________________________________________________
Page 76
NOTES
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
______________________________________________________________________________
Page 77
NOTES
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
______________________________________________________________________________
Page 78
NOTES
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
______________________________________________________________________________
Page 79
Page 80