It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI
regulations that apply. OMEGA is constantly pursuing certification of its products to the European New
Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any
errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, human applications.
One Omega Drive, River Bend Technology Centre, Northbank
Irlam, Manchester M44 5BD England
Toll-Free: 0800-488-488TEL: +44 (0)161 777-6611
FAX: +44 (0)161 777-6622e-mail: [email protected]
PART NUMBERING ...................................................... 18
REPLACEMENT PARTS ...................................................18
3
Page 4
INTRODUCTION
The FTB–1400 Flow Monitor is a state-of-the-art, digital signal processing
ow monitor, designed to provide the user with exceptional exibility at a
very aordable price. Though designed for use with Omega FTB–1400 ow
meters, this display can be used with almost any ow meter producing a
low amplitude AC output or contact closure signal(s).
This ow monitor is capable of accepting a low-level frequency input for
calculating ow rate and total. These calculations can then be displayed
in the desired units of measurement. All FTB–1400 ow monitors come
pre-calibrated, from the factory, if ordered with an Omega FTB–1400 Flow
Meter. If required, however, it can easily be re-congured in the eld. The
monitor’s large 8 digit by .75" numeric liquid crystal display makes extended range viewing practical. The second 8 digit by .38" alphanumeric display
provides for selectable units viewing in run mode and prompts for variables in program mode. Finally, the user can choose between displaying
rate, total, or alternating between both rate and total.
FTB-1400 Series
RUN PROGRAM
RELAY1 RELAY2
ONE OMEGA DRIVE, BOX 4047
PHONE 203-359-1660 FAX 203-359-7700 www.omega.com
STAMFORD, CT 06907-0047
FLOW MONITORING SYSTEM
Figure 1: Flow Monitor Face
4
Page 5
FEATURES
• Displays Rate and/or Total
• Large 0.75 Inch, 8 Digit Display for Easy Viewing
• Simple, Front Panel Programming
• Various Mounting Styles Available
• NEMA 4X Suitable for Outdoor Mounting
• Intrinsically Safe
• Microprocessor Based, Low Power Components
• 4-20 mA Loop Powered
• Automatic Decimal Point Locating
• Lead Zero Blinking
• Surface Mount Technology Use Throughout
5
Page 6
SPECIFICATIONS
Power Supply:
4-20 mA loop power
Power Consumption:
25 mA (maximum)
Alphanumeric Rate and Total Display:
8 digit, .75" high numeric display
8 character, .38" high alphanumeric display
Fixed or toggle modes of operation for ow rate and totalizer display
Pulsed Output Signal:
Outputs one pulse for each increment of the least signicant
totalizer digit
Max. Voltage: 30 Vdc
Pulse Type: Opto-Isolated open collector transistor
Pulse Width ON State: 0.9 V drop @ 5.0 mA or 0.7 V drop @ 0.1 A
Magnetic Pick-up Inputs:
Frequency Range: 0 to 3500 Hz
Trigger Sensitivity: 30 mV p-p
Over Voltage Protected: ±30 VDC
Frequency Measurement Accuracy: ±0.1%
Temperature Drift: 50 ppm / ºC (max)
Transient Overvoltages: Category 3, in accordance with IEC664
Pollution Degree: 2, in accordance with IEC664
Mounting Classication:
Meter Mount: NEMA 4X Enclosure
Remote Mount: NEMA 4X Enclosure
Swivel Mount: NEMA 4X Enclosure
Environmental:
Operating Temperature: -22 ºF to +158 ºF (-30 ºC to +70 ºC)
Humidity: 0-90% Non-condensing
The monitor has two modes of operation referred to as the RUN mode and
the PROGRAM mode. Both the run mode and the program mode display
screen enunciators conrming the state of the monitor. A quick glance at
the lower left hand corner of the LCD screen will conrm operating status.
Normal operation will be in the RUN mode. To access the program mode,
press the MENU button until the rst programming screen is displayed.
After programming the display with the necessary information, a lock out
feature can be turned on to prevent unauthorized access or changing the
meter’s setup parameters.
BASIC PROGRAMMING MODE
Keys:
MENU – Switch between RUN and PROGRAM modes
UP Arrow – Scrolls through programming sub-menus in forward
direction and increments numeric variables
RIGHT Arrow – Scrolls through programming sub-menus in reverse
direction and moves the active digit to the right
ENTER – Used to enter sub-menus, save programming information
and in the reset process
If your monitor was ordered with an Omega ow meter, the two components ship from the factory calibrated as a set. If the monitor is a replacement, the turbine’s K-factor has changed, or the monitor is being used with
some other pulse generating device, programming will be necessary.
PROGRAMMING
Each turbine ow meter is shipped with either a K-factor value or frequency
data. If frequency data is provided, the data must be converted to a K-factor
before programming. K-factor information, when supplied, can usually be
found on the neck of the ow meter or stamped on the body. The K-factor
represents the number of pulses per unit of volume. The K-factor will be
needed to program the monitor.
ENTER PROGRAM MODE – Change to program mode by pressing the MENU button once. The mode indicator will change from
RUN to PROGRAM.
7
Page 8
NOTE: If any input value exceeds the meter's capabilities for that particular
parameter, the LIMIT indicator will begin to ash indicating an invalid entry.
Press ENTER once to return to the parameter’s entry screen to reenter the
value.
SELECT THE METER SIZE – At the METER prompt, press the ENTER button once. The current meter size number will begin to ash. Using the
arrow keys, scroll through the size choices until you nd the bore size of
your meter. Press ENTER once to save the meter size choice.
NOTE: The meter connection size and the bore size are dierent. For example,
many of the 1” NPT turbines have bore sizes that range from 3⁄8” up to 1”. Be
sure to use the correct bore size or the meter will report incorrect ows and
totals.
Mode
Indicator
Programming
Keys
FTB-1400 Series
RUN PROGRAM
RELAY1 RELAY2
ONE OMEGA DRIVE, BOX 4047
STAMFORD, CT 06907-0047
PHONE 203-359-1660 FAX 203-359-7700 www.omega.com
Rate / Total Display
And
Programming Choices
FLOW MONITORING SYSTEM
Units Display
And
Programming Choices
Figure 2: Flow Monitor Controls
8
Page 9
ENTER THE METER’S K-FACTOR UNIT – Directly after the METER size is
selected, the display’s K-factor unit must be chosen. Use the UP arrow
key to select your K-factor unit. For meters calibrated in gallons, use
PUL/GAL (pulses per gallon); for meters calibrated in cubic meters, use
PUL/M3 (pulses per cubic meter), etc. Press ENTER to save the K-factor
unit and advance to the next parameter.
NOTE: Unless otherwise specied, Omega turbine ow meters are supplied
with K-factors measured in pulses per gallon (PUL/GAL) which will
automatically convert to your desired units of measure.
NOTE: The K-factor supplied with the meter or calculated from calibration
data will be needed to complete the next step..
SELECT THE RATE/TOTAL UNITS OF MEASURE – The monitor allows
the choice of ve common rate/total units. The monitor shows the rate/
total unit that the display is currently set for. If the current selection is
correct, press the ENTER key once to advance to the next parameter. To
change to an alternate unit, use the arrow keys to scroll to the desired
rate unit and press ENTER to save the choice.
SelectionRateTotal
GPM/GALGallons per MinuteGallons
LPM/LITLiters per MinuteLiters
M3PH/M3Cubic Meters per HourCubic Meters
M3PD/M3Cubic Meters per DayCubic Meters
BPD/BBLOil Barrels per DayOil Barrels
NOTE: The total unit’s output multiplier cannot be modied in the Simplied
program level. This option is reserved in the Advanced program level.
SELECT THE DISPLAY FUNCTION – The monitor can display RATE or TOTAL
or alternate between BOTH rate and total. At the DISPLAY prompt, press
the ENTER key once. The monitor now shows the display mode currently
in eect. If the current selection is correct, press the ENTER key to advance
to the next parameter. To change to an alternate display mode, use the arrow keys to scroll to the desired display mode and press ENTER to save the
choice.
A TEST function is also available in the Display Function sub-menu. With
the test function selected, the display acts like a frequency counter and
displays the raw input frequency being supplied to the frequency input
terminals. This is very useful when troubleshooting ow problems.
9
Page 10
TOTALIZER PULSE OUTPUT – The pulse output parameter can be either
enabled or disabled. When enabled this output generates 20 mS duration
pulse for every time the least signicant digit of the totalizer increments.
The amplitude of the pulse is dependent on the voltage level of the supply
connected to the pulse output and is limited to a maximum 30 VDC.
FTB1400 SIMPLIFIED 420 MA PROGRAMMING
FLOW 4mA SETTING – When the loop powered option is ordered, the
ow rate that corresponds to 4mA must be set. Zero is default ow rate
for this setting. If the current selection is correct, press the ENTER key
once to advance to the next parameter. If adjustment is required, use
the RIGHT arrow key to select the position of the number you wish to
change. Then use the UP arrow key to increment the number. Once you
have completed this step, press the ENTER key to advance to the next
parameter.
FLOW 20mA SETTING – The ow rate that corresponds to 20mA must
be set next. The turbine meter’s maximum ow rate is the default value.
If the current selection is correct, press the ENTER key once to advance
to the next parameter. If adjustment is required, use the RIGHT arrow
key to select the position of the number you wish to change. Then use
the UP arrow key to increment the number. Once you have completed
this step, press the ENTER key to advance to the next parameter.
4-20mA CALIBRATION – When ordered with a 4-20mA option, this
menu item allows the ne adjustment of the 4-20mA output. The 4mA
setting is typically between 35 and 50. To set the 4mA value, connect
an ammeter in series with the loop power supply. At the 4mA OUT
prompt, adjust the 4mA value to obtain a 4mA reading on the ammeter. The UP arrow key increments the value and the RIGHT arrow key
decrements the value. When a steady 4mA reading is obtained on the
ammeter, press the ENTER key to lock in this value and move to the
20mA adjustment. The 20mA adjustment is performed using the same
procedure as the 4mA adjustment.
4-20mA TEST – The monitor contains a diagnostic routine that allows
the simulation of mA values between 4 and 20 to check output tracking. At the 4-20TEST prompt the arrow keys change the simulated mA
output in increments of 1mA. The ammeter should track the simulated
mA output. If a 4-20mA test in not necessary, pressing the ENTER key
once will escape the testing at any time.
10
Page 11
Mag
J2
Input
Keypad
Connector
DC
1 2 3 4 5
JP2
Pulse
Inuput
+10 - 30 VDC
+1 - 5 VDC
1 2 3 4 5 6
-
+
+ -+
Pulse OutFreq. In4-20 mA
TB1
-
Figure 3: Typical Ammeter Connection
PASSWORD – Password protection prevents unauthorized users from
changing programming information. Initially, the password is set to all zeros. To change the password, simply enter any 4 digit code using the arrow
keys as previously described to enter the password value. Pressing ENTER
once will store the password and take you back to the RST PSWD screen.
NOTE: This password will allow the operator to manually reset totals.
RST PSWD – The reset password screen allows the operator to enter any 4
digit code for the manual reset totals function.
RESET TOTAL – To reset the monitor total display, in RUN mode press the
MENU and ENTER simultaneously until TOTAL RST starts to ash. The TOTAL
RST will stop ashing and the display will return to the RUN mode at the
conclusion of the reset procedure.
STORE TOTAL – The current total can be manually stored in the monitor’s
ash memory. Press and hold the ENTER key for 2 seconds. The display will
respond with a ashing TOTALSVD and then return to the RUN mode.
AUTOMATIC STORE TOTAL – The monitor is equipped with a store total
feature that works automatically, saving the current total to ash memory.
The frequency of saves depends on the power supply option chosen.
11
Page 12
Loop Powered: Once every ten minutes.
+10 - 30 VDC
ADDITIONAL INPUT OPTIONS
The FTB–1400 Flow Monitor is capable of receiving magnetic pick-up input
(small signal sine wave) or a contact closure input (pulse). Since Omega
FTB–1400 Flow Meters utilize a magnetic pick-up, the monitor is shipped
congured for magnetic pick-up input. To change to a contact closure input, remove JP2 from the top two pins and jumper them to the bottom two
pins. See Figure 4.
LOOP POWERED
20 mS Pulse
+1 - 5 VDC
+30 VDC (Max)
10K
250 Typ
R
(Loop Supply Voltage - 5)
=
MAX
0.02
Figure 4: Wiring Diagram
Keypad
Connector
1 2 3 4 5
1 2 3 4 5 6
+ -+ -+
Pulse OutFreq. In4-20 mA
JP2
-
Mag
Input
Pulse
Inuput
TB1
J2
Magnetic Pickup
12
Page 13
K Factor Unit
Units
Pulse Output
Simplied Menu
Figure 5: Basic Programming Menu
13
Page 14
5.75"
7.00"
3.93"
8.93"
7.00"
7.00"
RUN PROGRAM
RELAY1 RELAY2
MENUENTER
METER DISPLAY
14
2.40"
TYP.
2.25"
TYP.
5.75"
RUN PROGRAM
RELAY1 RELAY2
MENUENTER
4.30"
REMOTE DISPLAY
Page 15
3.93"
5.75"
12.25"
RUN PROGRAM
RELAY1 RELAY2
MENUENTER
7.00"
SWIVEL MOUNT
Figure 6: MOUNTING OPTIONS
15
Page 16
WIRE HARNESS
SWIVEL MOUNT
METER MOUNT OR
SCALE: .75 = 1
SWIVEL MOUNT DETAIL
METER MOUNT
OR SWIVEL MOUNT
MAGNETIC PICK-UP
REMOTE CABLE WITH STRAIGHT AMPHENOL
CONNECTOR OR REMOTE CABLE WITH
90° AMPHENOL CONNECTOR
MAGNETIC PICKUP
10 FT. TO MAX. 100 FT.
STRAIGHT TO TURBINE
CONNECTOR 90° OR
OPTIONAL AMPHENOL
WIRE HARNESS
FTB-1400 SERIES
TURBINE METER
16
Figure 7: Installation Drawing
Page 17
TROUBLESHOOTING GUIDE
TroubleRemedy
No LCD Display
No Rate or Total
Displayed
Flow Rate Display
Interprets Reading
Constantly
Flow Rate Indicator
Bounces
• Check power supply voltage. Should be between 10 and
30 Vdc.
• Check connection from meter pick-up to display input
terminals.
• Check turbine meter rotor for debris. Rotor should spin
freely.
• Check programming of ow monitor.
• This is usually an indication of external noise. Keep all AC
wires separate from DC wires.
• Check for large motors close to the meter pick-up.
• Check for radio antenna in close proximity.
• Try disconnecting the pick-up from the monitor pig tail.
This should stop the noise.
• This usually indicates a weak signal. Replace pick-up
and/or check all connections.
• Examine K-factor.
DEFAULT KFACTOR VALUES
Meter SizeDefault K-factorLower LimitUpper Limit
0.37520,00016,00024,000
0.50013,00010,40015,600
0.7502,7502,2003,300
0.8752,6862,1483,223
1.000870.0696.01,044
1.500330.0264.0396.0
2.00052.041.662.0
3.00057.045.668.0
4.00029.023.235.0
6.0007.05.68.0
8.0003.02.44.0
10.0001.61.32.0
17
Page 18
PART NUMBERING
FTB–1400–XD–LP
Mounting Style
M - Meter Mount
R - Remote Mount
S - Swivel Mount
REPLACEMENT PARTS
ComponentPart Number
KeypadB260713
Pick-up CableB222-121
Desiccant BagB260630
PVC UnionB220016
PVC Reducer BushingB220056
Rubber WasherB228207
Steel Lock WasherB220018
Desiccant ShieldB280680
Cord GripB220103
18
Page 19
WARRANTY/ DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and
workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds
an additional one (1) month grace period to the normal one (1) year product warranty to
cover handling and shipping time. This ensures that OMEGA’s customers receive maximum
coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer
Service Department will issue an Authorized Return (AR) number immediately upon phone or
written request. Upon examination by OMEGA, if the unit is found to be defective, it will be
repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting
from any action of the purchaser, including but not limited to mishandling, improper
interfacing, operation outside of design limits, improper repair, or unauthorized modification.
This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows
evidence of having been damaged as a result of excessive corrosion; or current, heat, moistur e
or vibration; improper specification; misapplication; misuse or other operating conditions
outside of OMEGA’s control. Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes
liability for any damages that result from the use of its products in accordance with
information provided by OMEGA, either verbal or written. OMEGA warrants only
that the parts manufactured by the company will be as specified a nd free of
defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY
KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL
IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total
liability of OMEGA with respect to this order, whether based on contract, warranty,
negligence, indemnification, strict liability or otherwise, shall not exceed the
purchase price of the component upon which liability is based. In no event shall
OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1)
as a “Basic Component” under 10 CFR 21 (NRC), used in or with any
activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or
with any nuclear installation or activity, medical application, used on humans, or misused in
any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER
language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from
any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
nuclear installation or
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.
RETURN REQUESTS/INQUIRIES
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN
AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT
(IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number s hould then be
marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to
prevent breakage in transit.
FOR WARRANTY
the following information available BEFORE
contacting OMEGA:
1. Purchase Order number under which
the product was PURCHASED,
2. Model and serial number of the product
under warranty, and
3. Repair instructions and/or specific
problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, w henever an improvement is possible.
This affords our customers the latest in technology and engineering.
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without
the prior written consent of OMEGA ENGINEERING, INC.
RETURNS, please have
FOR NON-WARRANTY REPAIRS,
OMEGA for current repair charges. Have the
following information available BEFORE
contacting OMEGA:
1. Purchase Order number to cover the
COST of the repair,
consult
2. Model and serial number of theproduct, and
3. Repair instructions and/or specific problems
relative to the product.