Omega FMA 4000 User Manual

www.omega.com
e-mail: info@omega.com
User’s Guide
FMA 4000
Digital Mass Flow Meters
Shop online at
Servicing North America:
ISO 9001 Certified Stamford CT 06907-0047
Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@ omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@ omega.ca
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378 / 1-800-622-BEST
®
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
®
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico: En Espan˜ ol: (001) 203-359-7803 e-mail: espanol@ omega.com
FAX: (001) 203-359-7807 info@ omega.com.mx
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: sales@ omegaeng.nl
Czech Republic: Rude´ arm-dy 1868, 733 01 Karvin- 8
Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114 Toll Free: 0800-1-66342 e-mail: info@ omegashop.cz
France: 11, rue Jacques Cartier, 78280 Guyancourt, France
Tel: +33 (0)1 61 37 29 00 FAX: +33 (0)1 30 57 54 27 Toll Free in France: 0800 466 342 e-mail: sales@ omega.fr
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: info@ omega.de
United Kingdom: One Omega Drive, River Bend Technology Centre
ISO 9002 Certified Northbank, Irlam, Manchester
M44 5BD United Kingdom Tel: +44 (0)161 777 6611 Toll Free in United Kingdom: 0800-488-48 FAX: +44 (0)161 777 6622 e-mail: sales@ omega.co.uk
OMEGAnet®Online Service Internet e-mail
www.omega.com info@ omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient-connected applications.
TABLE OF CONTENTS
1. UNPACKING THE FMA 4000 MASS FLOW METER...................................
1.1 Inspect Package for External Damage.................................................
1.2 Unpack the Mass Flow Meter...............................................................
1.3 Returning Merchandise for Repair.......................................................
2. INSTALLATION........................................................................................
2.1 Primary Gas Connections.................................................................
2.2 Electrical Connections......................................................................
2.2.1 Power Supply Connections..............................................................
2.2.2 Output Signals Connections..............................................................
2.2.3 Communication Parameters and Connections...................................
3. PRINCIPLE OF OPERATION...................................................................
4. SPECIFICATIONS...................................................................................
5. OPERATING INSTRUCTIONS..................................................................
5.1 Preparation and Warm Up..................................................................
5.2 Swamping Condition.......................................................................
5.3 FMA 4000 Parameters Settings...........................................................
5.3.1 Engineering Units Settings...............................................................
5.3.2 Gas Table Settings..............................................................................
5.3.3 Totalizer Settings.............................................................................
5.3.4 Flow Alarm Settings........................................................................
5.3.5 Relay Assignment Settings..............................................................
5.3.6 K Factors Settings...........................................................................
5.3.7 Zero Calibration...............................................................................
5.3.8 Self Diagnostic Alarm.......................................................................
5.4 Analog output Signals configuration...................................................
6. MAINTENANCE.........................................................................................
6.1 Introduction......................................................................................
6.2 Flow Path Cleaning...........................................................................
6.2.1 Restrictor Flow Element (RFE)........................................................
6.2.2 FMA 4000 model.............................................................................
1 1 1 1
1 1 3 3 3 4
6
7
9
9 10 11 11 12 12 13 14 14 15 17 17
18 18 19 19 19
7. CALIBRATION PROCEDURES.................................................................
7.1 Flow Calibration...............................................................................
7.2 Gas Calibration of FMA 4000 Mass Flow Meter................................
7.2.1 Connections and Initial Warm Up.....................................................
7.2.2 ZERO Check/Adjustment Adjustment.................................................
7.2.3 Gas Linearization Table Adjustment.................................................
7.3 Analog output Calibration of FMA 4000 Mass Flow Meter..............
7.3.1 Initial Setup.......................................................................................
7.3.2 Gas flow 0-5 Vdc analog output calibration....................................
7.3.3 Gas flow 4-20 mA analog output calibration...................................
8. RS485 / RS232 SOFTWARE INTERFACE COMMANDS.........................
8.1 General............................................................................................
8.2 Commands Structure.........................................................................
8.3 ASCII Commands Set.........................................................................
9. TROUBLESHOOTING................................................................................
9.1 Common Conditions........................................................................
9.2 Troubleshooting Guide.....................................................................
9.3 Technical Assistance.......................................................................
10. CALIBRATION CONVERSIONS FROM REFERENCE GASES...................
APPENDIX I OMEGA FMA 4000 EEPROM Variables..............................
APPENDIX II INTERNAL USER SELECTABLE GAS FACTOR TABLE
(INTERNAL “K” FACTORS)........................................................
APPENDIX III GAS FACTOR TABLE (“K” FACTORS)....................................
APPENDIX IV COMPONENT DIAGRAM......................................................
APPENDIX V DIMENSIONAL DRAWINGS.................................................
APPENDIX VI WARRANTY...........................................................................
20 20 21 21 21 21 23 24 25 25
26 26 26 28
34 34 35 37
37
38
41
42
46
48
50
TRADEMARKS
Buna-N®-is a registered trademark of DuPont Dow Elastomers. Kalrez®-is a registered trademark of DuPont Dow Elastomers.
Neoprene
®
-is a registered trademark of DuPont.
Omega
®
-is a registered trademark of Omega Engineering Inc.
1
1. UNPACKING THE FMA 4000 MASS FLOW METER
1.1 Inspect Package for External Damage
Your FMA 4000 Mass Flow Meter was carefully packed in a sturdy cardboard car­ton, with anti-static cushioning materials to withstand shipping shock. Upon receipt, inspect the package for possible external damage. In case of external damage to the package contact the shipping company immediately.
1.2 Unpack the Mass Flow Meter
Open the carton carefully from the top and inspect for any sign of concealed ship­ping damage. In addition to contacting the shipping carrier please forward a copy of any damage report to Omega7 directly.
When unpacking the instrument please make sure that you have all the items indicated on the Packing List. Please report any shortages promptly.
1.3 Returning Merchandise for Repair
Please contact an OMEGA7 customer service representative and request a
Return Authorization Number (AR).
It is mandatory that any equipment returned for servicing be purged and neutral­ized of any dangerous contents including but not limited to toxic, bacterially infec­tious, corrosive or radioactive substances. No work shall be performed on a returned product unless the customer submits a fully executed, signed SAFETY CERTIFICATE. Please request form from the Service Manager.
2. INSTALLATION
2.1 Primary Gas Connections
Please note that the FMA 4000 Mass Flow Meter will not operate with liquids. Only clean gases are allowed to be introduced into the instrument. If gases are con­taminated they must be filtered to prevent the introduction of impediments into the sensor.
2
CAUTION: FMA 4000 TRANSDUCERS SHOULD NOT BE USED FOR MONITORING OXYGEN GAS UNLESS SPECIFICALLY CLEANED AND PREPARED FOR SUCH APPLICATION.
For more information, contact Omega7.
Attitude limit of the Mass Flow Meter is ±15Ffrom calibration position (standard calibration is in horizontal position). This means that the gas flow path of the Flow Meter must be within this limit in order to maintain the original calibration accura­cy. Should there be need for a different orientation of the meter, re-calibration may be necessary. It is also preferable to install the FMA 4000 transducer in a stable environment, free of frequent and sudden temperature changes, high moisture, and drafts.
Prior to connecting gas lines inspect all parts of the piping system including fer­rules and fittings for dust or other contaminant’s.
When connecting the gas system to be monitored, be sure to observe the direc­tion of gas flow as indicated by the arrow on the front of the meter.
Insert tubing into the compression fittings until the ends of the properly sized tub­ing home flush against the shoulders of the fittings. Compression fittings are to be tightened to one and one quarter turns according to the manufacturer's instruc­tions. Avoid over tightening which will seriously damage the Restrictor Flow Elements (RFE's)!
CAUTION: For FMA 4000 model, the maximum pressure in the gas line should not exceed 500 PSIA (34.47 bars). Applying pressure above 500 PSIA (34.47 bars) will seriously damage the flow sensor.
FMA 4000 transducers are supplied with either standard 1/4 inch, or optional 1/8 inch inlet and outlet compression fittings which should NOT be removed unless the meter is being cleaned or calibrated for a new flow range.
Using a Helium Leak Detector or other equivalent method, perform a thorough leak test of the entire system. (All FMA 4000's are checked prior to shipment for leakage within stated limits. See specifications in this manual.)
3
2.2 Electrical Connections
FMA 4000 is supplied with a 15 pin “D” connector. Pin diagram is presented in
Figure b-1.
2.2.1 Power Supply Connections
The power supply requirements for FMA 4000 transducers are: 11 to 26 Vdc, (unipolar power supply)
DC Power (+) --------------- pin 7 of the 15 pin “D” connector DC Power (-) --------------- pin 5 of the 15 pin “D” connector
CAUTION: Do not apply power voltage above 26Vdc. Doing so will cause FMA 4000 damage or faulty operation.
2.2.2 Output Signals Connections
CAUTION: When connecting the load to the output terminals, do not exceed the rated values shown in the specifications. Failure to do so might cause damage to this device. Be sure to check if the wiring and the polarity of the power supply is correct before turning the power ON. Wiring error may cause damage or faulty operation.
FMA 4000 Mass Flow Meters are equipped with either calibrated 0-5 or calibrat­ed 4-20 mA output signals (jumper selectable). This linear output signal repre­sents 0-100% of the flow meter’s full scale range.
WARNING: The 4-20 mA current loop output is self-powered (non-isolated). Do NOT connect an external voltage source to the output signals.
Flow 0-5 VDC or 4-20 mA output signal connection:
Plus (+) -------------------------- pin 2 of the 15 pin “D” connector Minus (-) -------------------------- pin 1 of the 15 pin “D” connector
To eliminate the possibility of noise interference, use a separate cable entry for the DC power and signal lines.
2.2.3 Communication Parameters and Connections
The digital interface operates via RS485 (optional RS232) and provides access to applicable internal data including: flow, CPU temperature reading, auto zero, total­izer and alarm settings, gas table, conversion factors and engineering units selec­tion, dynamic response compensation and linearization table adjustment.
Communication Settings for RS485 / RS232 communication interface:
Baud rate: ...................... 9600 baud
Stop bit: ...................... 1
Data bits: ...................... 8
Parity: ...................... None
Flow Control: ...................... None
RS485 communication interface connection:
The RS485 converter/adapter must be configured for: multidrop, 2 wire, half duplex mode. The transmitter circuit must be enabled by TD or RTS (depending on which is available on the converter/adapter). Settings for the receiver circuit should follow the selection made for the transmitter circuit in order to eliminate echo.
RS485 T(-) or R(-) ...................... pin 8 of the 15 pin “D” connector (TX-)
RS485 T(+) or R(+) ...................... pin 15 of the 15 pin “D” connector (RX+)
RS485 GND (if available) ...................... pin 9 of the 15 pin “D” connector (GND)
RS232 communication interface connection:
Crossover connection has to be established:
RS232 RX (pin 2 on the DB9 connector) ..... pin 8 of the 15 pin “D” connector (TX)
RS232 TX (pin 3 on the DB9 connector) ..... pin 15 of the 15 pin “D” connector (RX)
RS232 GND (pin 5 on the DB9 connector) ..... pin 9 of the 15 pin “D” connector (GND)
4
5
PIN FMA 4000 FUNCTION
1 Common, Signal Ground For Pin 2
(4-20 mA return).
2 0-5 Vdc or 4-20mA Flow Signal Output. 3 Relay No. 2 - Normally Open Contact. 4 Relay No. 2 - Common Contact. 5 Common, Power Supply
(- DC power for 11 to 26 Vdc).
6 Relay No. 1 - Common Contact. 7 Plus Power Supply
(+ DC power for 11 to 26 Vdc).
8 RS485 (-) (Optional RS232 TX). 9 RS232 Signal GND (RS485 GND Optional). 10 Do not connect (Test/Maintenance terminal). 11 Relay No. 2 - Normally Closed Contact. 12 Relay No. 1 - Normally Open Contact. 13 Relay No. 1 - Normally Closed Contact. 14 Do not connect (Test/Maintenance terminal). 15 RS485 (+) (Optional RS232 RX). Shield Chassis Ground.
Figure b.1 - FMA 4000 15 PIN “D” CONNECTOR CONFIGURATION
IMPORTANT NOTES:
Generally, “D” Connector numbering patterns are standardized. There are, how­ever, some connectors with nonconforming patterns and the numbering sequence on your mating connector may or may not coincide with the numbering sequence shown in our pin configuration table above. It is imperative that you match the appropriate wires in accordance with the correct sequence regardless of the particular numbers displayed on the mating connector.
Make sure power is OFF when connecting or disconnecting any cables in the system.
The (+) and (-) power inputs are each protected by a 300mA M (medium time-lag) resettable fuse. If a shorting condition or polarity reversal occurs, the fuse will cut power to the flow transducer circuit. Disconnect the power to the unit, remove the faulty condition, and reconnect the power. The fuse will reset once the faulty con­dition has been removed. DC Power cable length may not exceed 9.5 feet (3 meters). Use of the FMA 4000 flow transducer in a manner other than that spec­ified in this manual or in writing from Omega, may impair the protection provided by the equipment.
3. PRINCIPLE OF OPERATION
The stream of gas entering the Mass Flow transducer is split by shunting a small portion of the flow through a capillary stainless steel sensor tube. The remainder of the gas flows through the primary flow conduit. The geometry of the primary con­duit and the sensor tube are designed to ensure laminar flow in each branch. According to principles of fluid dynamics the flow rates of a gas in the two laminar flow conduits are proportional to one another. Therefore, the flow rates measured in the sensor tube are directly proportional to the total flow through the transducer.
In order to sense the flow in the sensor tube, heat flux is introduced at two sec­tions of the sensor tube by means of precision wound heater-sensor coils. Heat is transferred through the thin wall of the sensor tube to the gas flowing inside. As gas flow takes place heat is carried by the gas stream from the upstream coil to the downstream coil windings. The resultant temperature dependent resistance differential is detected by the electronic control circuit. The measured temperature gradient at the sensor windings is linearly proportional to the instantaneous rate of flow taking place.
An output signal is generated that is a function of the amount of heat carried by the gases to indicate mass-molecular based flow rates.
Additionally, the FMA 4000 Mass Flow Meter incorporates a Precision Analog Microcontroller (ARM7TDMI7 MCU) and non-volatile memory that stores all hard-
ware specific variables and up to 10 different calibration tables. The flow rate can be displayed in 23 different volumetric or mass flow engineering units. Flow meter parameters and functions can be programmed remotely via the RS485/RS232 (optional) interface. FMA 4000 flow meters support various functions including: programmable flow totalizer, low, high or range flow alarm, automatic zero adjust­ment (activated via local button or communication interface), 2 programmable SPDT relays output, 0-5 Vdc / 4-20 mA analog outputs (jumper selectable), self diagnostic alarm, 36 internal and user defined K-factor. Optional local 2x16 LCD readout with adjustable back light provides flow rate and total volume reading in currently selected engineering units and diagnostic events indication.
6
7
4. SPECIFICATIONS
FLOW MEDIUM: Please note that FMA 4000 Mass Flow Meters are designed to work only with clean gases. Never try to measure flow rates of liquids with any FMA 4000.
CALIBRATIONS: Performed at standard conditions [14.7 psia (101.4 kPa) and 70
F
F
(21.1
F
C)] unless otherwise requested or stated.
ENVIRONMENTAL (PER IEC 664): Installation Level II; Pollution Degree II.
FLOW ACCURACY (INCLUDING LINEARITY): ±1% of FS at calibration temperature and
pressure.
REPEATABILITY: ±0.15% of full scale.
FLOW TEMPERATURE COEFFICIENT: 0.15% of full scale/
F
C or better.
FLOW PRESSURE COEFFICIENT: 0.01% of full scale/psi (6.895 kPa) or better.
FLOW RESPONSE TIME: 1000ms time constant; approximately 2 seconds to within ±2%
of set flow rate for 25% to 100% of full scale flow.
MAXIMUM GAS PRESSURE: 500 psig (3447 kPa gauge).
MAXIMUM PRESSURE DROP: 0.18 PSID (at 10 L/min flow). See Table IV for
pressure drops associated with various models and flow rates.
GAS AND AMBIENT TEMPERATURE: 41
F
F to 122 FF (5 FC to 50 FC).
RELATIVE GAS HUMIDITY: Up to 70%.
LEAK INTEGRITY: 1 x 10-9sccs He maximum to the outside environment.
ATTITUDE SENSITIVITY: Incremental deviation of up to 1% from stated accuracy, after re-
zeroing.
OUTPUT SIGNALS: Linear 0-5 Vdc (3000 ohms min load impedance);
Linear 4-20 mA (500 ohms maximum loop resistance). Maximum noise 20mV peak to peak (for 0-5 Vdc output).
TRANSDUCER INPUT POWER: 11 to 26 Vdc, 100 mV maximum peak to peak output noise.
Power consumption: +12Vdc (200 mA maximum);
+24Vdc (100 mA maximum); Circuit board have built-in polarity reversal protection, 300mA resettable fuse provide power input protection.
WETTED MATERIALS: Anodized aluminum, brass, 316 stainless steel, 416 stainless steel, FKM, O-rings; BUNA-N7, NEOPRENE7 or KALREZ7 O-rings are optional.
CAUTION: Omega makes no expressed or implied guarantees of corrosion resistance of mass flow meters as pertains to different flow media reacting with components of meters. It is the customers' sole responsibility to select the model suitable for a particular gas based on the fluid contacting (wetted) materials offered in the different models.
INLET AND OUTLET CONNECTIONS: Model FMA 4000 standard 1/4" compression fittings. Optional 1/8" or 3/8" compression fittings and 1/4" VCR fittings are available.
DISPLAY: Optional local 2x16 characters LCD with adjustable backlight (2 lines of text).
CALIBRATION OPTIONS: Standard is one 10 points NIST calibration.
Optional, up to 9 additional calibrations may be ordered at additional charge.
CE COMPLIANCE: EMC Compliance with 89/336/EEC as amended. Emission Standard: EN 55011:1991, Group 1, Class A. Immunity Standard: EN 55082-1:1992.
8
*Flow rates are stated for Nitrogen at STP conditions [i.e. 70 FF (21.1 FC) at 1 atm].
For other gases use the K factor as a multiplier from APPENDIX III.
TABLE IV PRESSURE DROPS
MODEL
FLOW RATE
[std liters/min]
MAXIMUM PRESSURE DROP
[mm H2O] [psid] [kPa]
FMA 4000 up to 10 130 0.18 1.275
CODE
scc/min [N2]
CODE
std liters/min [N2]
00 0 to 5 07 0 to 1 01 0 to 10 08 0 to 2
02 0 to 20 09 0 to 5 03 0 to 50 10 0 to 10 04 0 to 100 05 0 to 200
06 0 to 500
TABLE I FMA 4000 LOW FLOW MASS FLOW METER*
FLOW RANGES
MODEL
WEIGHT
SHIPPING WEIGHT
FMA 4000 transmitter 2.20 lbs. (1.00 kg) 3.70 lbs. (1.68 kg)
9
5. OPERATING INSTRUCTIONS
5.1 Preparation and Warm Up
It is assumed that the Mass Flow Meter has been correctly installed and thor­oughly leak tested as described in section 2. Make sure the flow source is OFF. When applying power to a flow meter within the first two seconds, you will see on the LCD display: the product name, the software version, and revision of the EEP­ROM table (applicable for LCD option only).
Figure b-2: FMA 4000 first Banner Screen
Within the next two seconds, the RS485 network address, the analog output set­tings, and currently selected gas calibration table will be displayed (applicable for LCD option only).
Figure b-3: FMA 4000 second Banner Screen
Note: Actual content of the LCD screen may vary depending on the model and device configuration.
After two seconds, the LSD display switches to the main screen with the following information:
- Mass Flow reading in current engineering units (upper line).
- Totalizer Volume reading in current volume or mass based engineering units (lower line).
Figure b-4: FMA 4000 Main Screen
OMEGA FMA 4000 485
S: Ver1.4 Rev.A0
Ad: 11 Out: 0-5Vdc
Gas# 1 AIR
F: 50.0 L/min
T: 75660.5 Ltr
10
During initial powering of the FMA 4000 transducer, the flow output signal will be indicating a higher than usual output. This is an indication that the FMA 4000 transducer has not yet attained its minimum operating temperature. This condition will automatically cancel within a few minutes and the transducer should eventu­ally indicate zero.
For the FMA 4000 transducer with LCD option: If the LCD diagnostic is activated, the second line of the LCD will display the time remaining until the end of the warm up period (Minutes:Seconds format) and will alternatively switch to Totalizer reading indication every 2 seconds.
Figure b-5: FMA 4000 Main Screen during Sensor Warm up period.
5.2 Swamping Condition
If a flow of more than 10% above the maximum flow rate of the Mass Flow Meter is taking place, a condition known as “swamping” may occur. Readings of a “swamped” meter cannot be assumed to be either accurate or linear. Flow must be restored to below 110% of maximum meter range. Once flow rates are lowered to within calibrated range, the swamping condition will end. Operation of the meter above 110% of maximum calibrated flow may increase recovery time.
Note: Allow the Digital Mass Flow Meter to warm-up for a MINIMUM of 6 minutes.
Note: During the first 6 minutes of the initial powering of the FMA 4000 transducer, the status LED will emit CONSTANT UMBER light.
Note: After 6 minutes of the initial powering of the FMA 4000 the transducer, status LED will emit a constant GREEN light (normal operation, ready to measure). For FMA 4000 with LCD option, the screen will reflect flow and totalizer reading. (see Figure b-4).
F: 50.0 L/min
** WarmUp 2:39 **
11
5.3 FMA 4000 Parameters Settings
5.3.1 Engineering Units Settings
The FMA 4000 Mass Flow Meter is capable of displaying flow rate with 23 different Engineering Units. Digital interface commands (see paragraph 8.3 ASCII Command Set “FMA 4000 SOFTWARE INTERFACE COMMANDS”) are provided to:
- get currently active Engineering Units
- set desired Engineering Units.
The following Engineering Units are available:
TABLE VI UNITS OF MEASUREMENT
NUMBER INDEX
FLOW RATE
ENGINEERING
UNITS
TOTALIZER
ENGINEERING
UNITS
DESCRIPTION
1 0 % %s Percent of full scale
2 1 mL/sec mL Milliliter per second
3 2 mL/min mL Milliliter per minute
4 3 mL/hr mL Milliliter per hour
5 4 L/sec Ltr Liter per second
6 5 L/ min Ltr Liter per minute
7 6 L/hr Ltr Liter per hour
87
m
3
/sec m
3
Cubic meter per second
98
m
3
/ min m
3
Cubic meter per minute
10 9
m
3
/hr m
3
Cubic meter per hour
11 10
f
3
/sec f
3
Cubic feet per second
12 11
f
3
/min f
3
Cubic feet per minute
13 12
f
3
/hr f
3
Cubic feet per hour
14 13 g/sec g Grams per second
15 14 g/min g Grams per minute
16 15 g/hr g Grams per hour
17 16 kg/sec kg Kilograms per second
18 17 kg/min kg Kilograms per minute
19 18 kg/hr kg Kilograms per hour
20 19 Lb/sec Lb Pounds per second
21 20 Lb/min Lb Pounds per minute
22 21 Lb/hr Lb Pounds per hour
23 22 User UD User defined
5.3.2 Gas Table Settings
The FMA 4000 Mass Flow Meter is capable of storing calibration data for up to 10 different gases. Digital interface commands are provided to:
- get currently active Gas Table number and Gas name
- set desired Gas Table.
5.3.3 Totalizer Settings
The total volume of the gas is calculated by integrating the actual gas flow rate with respect to the time. Digital interface commands are provided to:
- reset the totalizer to ZERO
- start the totalizer at a preset flow
- assign action at a preset total volume
- start/stop (enable/disable) totalizing the flow
- read totalizer via digital interface
The Totalizer has several attributes which may be configured by the user. These attributes control the conditions which cause the Totalizer to start integrat­ing the gas flow and also to specify actions to be taken when the Total Volume is outside the specified limit.
Totalizer action conditions become true when the totalizer reading and preset “Stop at Total” volumes are equal.
12
Note: Once Flow Unit of Measure is changed, the Totalizer’s Volume/Mass based Unit of Measure will be changed automatically.
Note: By default the FMA 4000 is shipped with at least one valid calibration table (unless optional additional calibrations were ordered). If instead of the valid Gas name (for example NITROGEN), the LCD screen or digital interface displays Gas designator as “Uncalibrated”, then the user has chosen the Gas Table which was not calibrated. Using an “Uncalibrated” Gas Table will result in erroneous reading.
Note: Before enabling the Totalizer, ensure that all totalizer settings are configured properly. Totalizer Start values have to be entered in %F.S. engineering unit. The Totalizer will not totalize until the flow rate becomes equal to or more than the Totalizer Start value. Totalizer Stop values must be entered in currently active volume / mass based engineering units. If the Totalizer Stop at preset total volume feature is not required, then set Totalizer Stop value to zero.
Mode Enable /Disable - Allows the user to Enable/Disable Flow Alarm.
Low Alarm - The value of the monitored Flow in % F.S. below
which is considered an alarm condition.
Note: The value of the Low alarm must be less than the
value of the High Alarm.
High Alarm- The value of the monitored Flow in % F.S. above
which is considered an alarm condition.
Note: The value of the High alarm must be more than the
value of the Low Alarm.
Action Delay- Th e ti me i n se conds that the Flow rate value must remain
above the high limit or below the low limit before an alarm condition is indicated. Valid settings are in the range of 0 to 3600 seconds.
13
Local maintenance push button is available for manual Totalizer reset on the field. The maintenance push button is located on the right side of the flow meter inside the maintenance window above the 15 pin D-connector (see Figure c-1 “FMA 4000 configuration jumpers”).
5.3.4 Flow Alarm Settings
FMA 4000 provides the user with a flexible alarm/warning system that monitors the Gas Flow for conditions that fall outside configurable limits as well as visual feedback for the user via the status LED and LCD (only for devices with LCD option) or via a Relay contact closure.
The flow alarm has several attributes which may be configured by the user via a digital interface. These attributes control the conditions which cause the alarm to occur and to specify actions to be taken when the flow rate is outside the speci­fied conditions.
Note: In order to locally Reset Totalizer, the reset push button must be pressed during power up sequence. The following sequence is recommended:
1. Disconnect FMA 4000 from the power.
2. Press maintenance push button (do not release).
3. Apply power to the FMA 4000 while holding down the maintenance push button.
4. Release maintenance push button after 6 seconds. For FMA 4000 with optional LCD, when FMA 4000 Main Screen appears (see Figure b-4).
The current Flow Alarm settings and status are available via digital interface (see paragraph 8.3 ASCII Command Set “FMA 4000 SOFTWARE INTERFACE COM­MANDS”).
5.3.5 Relay Assignment Settings
Two sets of dry contact relay outputs are provided to actuate user supplied equip­ment. These are programmable via digital interface such that the relays can be made to switch when a specified event occurs (e.g. when a low or high flow alarm limit is exceeded or when the totalizer reaches a specified value).
The user can configure each Relay action from 6 different options:
No Action : (N) No assignment (relay is not assigned to any events and not energized). Totalizer > Limit : (T) Totalizer reached preset limit volume. High Flow Alarm : (H) High Flow Alarm condition. Low Flow Alarm : (L) Low Flow Alarm condition. Range between H&L : (R) Range between High and Low Flow Alarm condition. Manual Enabled : (M) Activated regardless of the Alarm and Totalizer conditions.
5.3.6 K Factors Settings
Conversion factors relative to Nitrogen for up to 36 gases are stored in the FMA 4000 (see APPENDIX II). In addition, provision is made for a user-defined con­version factor. Conversion factors may be applied to any of the ten gas calibra­tions via digital interface commands.
14
Latch Mode- Controls Latch feature when Relays are assigned to
Alarm event. Following settings are available:
0 - Latch feature is disabled for both relays 1 - Latch feature is enabled for Relay#1 and disabled for Relay#2 2 - Latch feature is enabled for Relay#2 and disabled for Relay#1 3 - Latch feature is enabled for both relays.
Note: If the alarm condition is detected, and the Relay is assigned to Alarm event, the corresponding Relay will be energized.
Note: By default, flow alarm is non-latching. That means the alarm is
indicated only while the monitored value exceeds the specified
conditions. If Relay is assigned to the Alarm event, in some cases, the
Alarm Latch feature may be desirable.
15
The available K Factor settings are:
Disabled (K = 1).
Internal Index The index [0-35] from internal K factor table (see APPENDIX II).
User Defined User defined conversion factor.
5.3.7 Zero Calibration
The FMA 4000 includes an auto zero function that, when activated, automatical­ly adjusts the mass flow sensor to read zero. The initial zero adjustment for your FMA 4000 was performed at the factory. It is not required to perform zero calibra­tion unless the device has zero reading offset with no flow conditions.
Shut off the flow of gas into the Digital Mass Flow Meter. To ensure that no seep­age or leak occurs into the meter, it is good practice to temporarily disconnect the gas source. The Auto Zero may be initiated via digital communication interface or locally by pressing the maintenance push button, which is located on the right side of the flow meter inside the maintenance window above the 15 pin D-connector (see Figure c-1 “FMA 4000 configuration jumpers”).
To start Auto Zero locally, press the maintenance push button. The status LED will flash not periodically with the RED light. On the FMA 4000 with optional LCD, the following screen will appear:
Note: The conversion factors will not be applied for % F.S. engineering unit.
Note: Before performing Zero Calibration, make sure the device is powered up for at least 15 minutes and absolutely no flow condition is established.
Note: The same maintenance push button is used for Auto Zero initiation and Totalizer reset. The internal diagnostic algorithm will prevent initiating Auto Zero function via the maintenance push button before the 6 minutes sensor warm up period has elapsed.
16
Figure b-6: FMA 4000 Screen in the beginning of Auto Zero procedure.
The Auto Zero procedure normally takes 1 - 2 minutes during which time the DP Zero counts and the Sensor reading changes approximately every 3 to 6 seconds.
Figure b-7: FMA 4000 during the Auto Zero procedure.
The nominal value for a fully balanced sensor is 120 Counts. If the FMA 4000’s digital signal processor was able to adjust the Sensor reading within 120 ± 10 counts, then Auto Zero is considered successful. The status LED will return to a constant GREEN light and the screen below will appear:
Figure b-7: FMA 4000 during the Auto Zero procedure.
If the device was unable to adjust the Sensor reading to within 120 ± 10 counts, then Auto Zero is considered as unsuccessful. The constant RED light will appear on the status LED. The user will be prompted with the “AutoZero ERROR!” screen.
AUTOZERO IS ON!
AUTOZERO IS ON!
S: 405 DP: 512
AutoZero is Done
S: 122 DP: 544
Note: The actual value of the Sensor and DP counts will vary for each FMA 4000.
Note: For FMA 4000 with RS232 option all Auto Zero status info available via digital communication interface.
17
5.3.8 Self Diagnostic Alarm
FMA 4000 series Mass Flow Meters are equipped with a self-diagnostic alarm which is available via multicolor LED, digital interface and on screen indication (for devices with optional LCD). The following diagnostic events are supported:
5.4 Analog Output Signals configuration
FMA 4000 series Mass Flow Meters are equipped with calibrated 0-5 Vdc and 4­20 mA output signals. The set of the jumpers (J7A, J7B, J7C) located on the right side of the flow meter, inside of the maintenance window above the 15 pin D-con­nector (see Figure c-1 “FMA 4000 configuration jumpers”) are used to switch between 0-5 Vdc or 4-20 mA output signals (see Table VI).
NUMBER
DIAGNOSTIC
ALARM DESCRIPTION
LED COLOR
AND PATTERN
PRIORITY
LEVEL
1
Auto Zero procedure is running
Not periodically flashing RED
0
2
FATAL ERROR (reset or maintenance service is required for return in to the normal operation)
Constant RED 1
3
CPU Temperature too high (Electronics Overheating)
Flashing RED/UMBER 2
4
Sensor in the warm up stage (first 6 minutes after power up sequence, normal operation, no critical diagnostic events present)
Constant UMBER 3
5
Flow Sensor Temperature too low Flashing UMBER/OFF 4
6
Flow Sensor Temperature too high Flashing RED/OFF 5
7
Totalizer Reading hit preset limit Flashing GREEN/UMBER 6
8
Low flow Alarm conditions Flashing GREEN/OFF 7
9
High flow Alarm conditions Flashing GREEN/RED 8
10
Normal operation, no diagnostic events
Constant GREEN 9
Note: [0] - Priority Level is highest (most important). When two or more diagnostic events are present at the same time, the event with the highest priority level will be indicated on the status LED and displayed on the LCD (if equipped). All diagnostic events may be accessed simultaneously via digital communication interface (see paragraph 8.3 “ASCII Command Set”).
18
Analog output signals of 0-5 Vdc and 4-20 mA are attained at the appropriate pins of the 15-pin “D” connector (see Figure b-1) on the side of the FMA 4000 transducer.
Tabl e V I Analog Output Jumper Configuration
See APPENDIX IV for actual jumpers layout on the PCB.
6. MAINTENANCE
6.1 Introduction
It is important that the Mass Flow Meter is only used with clean, filtered gases. Liquids may not be metered. Since the RTD sensor consists, in part, of a small capillary stainless steel tube, it is prone to occlusion due to impediments or gas crystallization. Other flow passages are also easily obstructed.
Therefore, great care must be exercised to avoid the introduction of any potential flow impediment. To protect the instrument, a 50 micron (FMA 4000) filter is built into the inlet of the flow transducer. The filter screen and the flow paths may require occasional cleaning as described below. There is no other recommended mainte­nance required. It is good practice, however, to keep the meter away from vibra­tion, hot or corrosive environments and excessive RF or magnetic interference.
If periodic calibrations are required, they should be performed by qualified per­sonnel and calibrating instruments, as described in section 7. It is recommended that units are returned to Omega
®
for repair service and calibration.
ANALOG SIGNAL
OUTPUT
0-5 Vdc 4-20 mA
Flow Rate Output
Jumper Header J7
J7.A J7.B J7.C
5-9 6-10 7-11
J7.A J7.B J7.C
1-5 2-6 3-7
Note: Digital output (communication) is simultaneously available with analog output.
CAUTION: TO PROTECT SERVICING PERSONNEL IT IS MANDATORY THAT ANY INSTRUMENT BEING SERVICED IS COMPLETELY PURGED AND NEUTRALIZED OF TOXIC, BACTERIOLOGICALLY INFECTED, CORROSIVE OR RADIOACTIVE CONTENTS.
19
6.2 Flow Path Cleaning
Before attempting any disassembly of the unit for cleaning, try inspecting the flow paths by looking into the inlet and outlet ends of the meter for any debris that may be clogging the flow through the meter. Remove debris as necessary. If the flow path is clogged, proceed with steps below.
Do not attempt to disassemble the sensor. If blockage of the sensor tube is not alle­viated by flushing through with cleaning fluids, please return meter for servicing.
6.2.1 Restrictor Flow Element (RFE)
The Restrictor Flow Element (RFE) is a precision flow divider inside the trans­ducer which splits the inlet gas flow by a preset amount to the sensor and main flow paths. The particular RFE used in a given Mass Flow Meter depends on the gas and flow range of the instrument.
6.2.2 FMA 4000 Model
Unscrew the inlet compression fitting of meter. Note that the Restrictor Flow Element (RFE) is connected to the inlet fitting. Carefully disassemble the RFE from the inlet connection. The 50 micron filter screen will now become visible. Push the screen out through the inlet fitting. Clean or replace each of the removed parts as necessary. If alcohol is used for cleaning, allow time for drying.
Inspect the flow path inside the transducer for any visible signs of contaminant. If necessary, flush the flow path through with alcohol. Thoroughly dry the flow paths by flowing clean dry gas through.
Carefully re-install the RFE and inlet fitting avoiding any twisting and deforming to the RFE. Be sure that no dust has collected on the O-ring seal.
CAUTION: DISASSEMBLY MAY COMPROMISE CURRENT CALIBRATION.
NOTE: OVER TIGHTENING WILL DEFORM AND RENDER THE RFE DEFECTIVE. IT IS ADVISABLE THAT AT LEAST ONE CALIBRATION POINT BE CHECKED AFTER RE-INSTALLING THE INLET FITTING. SEE SECTION (7.2.3).
20
7. CALIBRATION PROCEDURES
7.1 Flow Calibration
Omega
®
Engineerings' Flow Calibration Laboratory offers professional calibration support for Mass Flow Meters using precision calibrators under strictly controlled conditions. NIST traceable calibrations are available. Calibrations can also be per­formed at customers' site using available standards.
Factory calibrations are performed using NIST traceable precision volumetric cal­ibrators incorporating liquid sealed frictionless actuators.
Generally, calibrations are performed using dry nitrogen gas. The calibration can then be corrected to the appropriate gas desired based on relative correction [K] factors shown in the gas factor table (see APPENDIX III). A reference gas, other than nitrogen, may be used to better approximate the flow characteristics of cer­tain gases. This practice is recommended when a reference gas is found with ther­modynamic properties similar to the actual gas under consideration. The appro­priate relative correction factor should be recalculated (see section 9).
It is standard practice to calibrate Mass Flow Meters with dry nitrogen gas at
70.0
F
F (21.1 FC), 20 psia (137.9 kPa absolute) inlet pressure and 0 psig outlet pressure. It is best to calibrate FMA 4000 transducers to actual operating condi­tions. Specific gas calibrations of non-toxic and non-corrosive gases are available for specific conditions. Please contact your Omega
®
for a price quotation.
It is recommended that a flow calibrator be used which has at least four times bet­ter collective accuracy than that of the Mass Flow Meter to be calibrated. Equipment required for calibration includes: a flow calibration standard, PC with available RS485 / RS232 communication interface, a certified high sensitivity multi meter (for analog output calibration only), an insulated (plastic) screwdriver, a flow regulator (for example - metering needle valve) installed upstream from the Mass Flow Meter, and a pressure regulated source of dry filtered nitrogen gas (or other suitable reference gas). Using Omega
®
supplied calibration and mainte-
nance software to simplify the calibration process is recommended.
Gas and ambient temperature, as well as inlet and outlet pressure conditions, should be set up in accordance with actual operating conditions.
NOTE: REMOVAL OF THE FACTORY INSTALLED CALIBRATION SEALS AND/OR ANY ADJUSTMENTS MADE TO THE METER, AS DESCRIBED IN THIS SECTION, WILL VOID ANY CALIBRATION WARRANTY APPLICABLE.
21
7.2 Gas Flow Calibration of FMA 4000 Mass Flow Meter
FMA 4000 Mass Flow Meters may be field recalibrated/checked for the same range they were originally factory calibrated for. When linearity adjustment is needed or flow range changes are being made, proceed to step 7.2.3. Flow range changes may require a different Restrictor Flow Element (RFE). Consult Omega
®
for more information.
7.2.1 Connections and Initial Warm Up
Power up the Mass Flow Meter for at least 15 minutes prior to commencing the calibration procedure. Establish digital RS485 / RS232 communication between PC (communication terminal) and the FMA 4000. Start Omega
®
supplied calibra-
tion and maintenance software on the PC.
7.2.2 ZERO Check/Adjustment
Using Omega
®
supplied calibration and maintenance software open Back Door
access:
Query/BackDoor/Open
When software prompts with Warning, click the [YES] button. This will open the access to the rest of the Query menu. Start Sensor Compensated Average reading:
Query/Read/ SensorCompAverage
This will display Device Sensor Average ADC counts.
With no flow conditions, the sensor Average reading must be in the range 120± 10 counts. If it is not, perform Auto Zero procedure (see section 5.3.10 “Zero Calibration”).
7.2.3 Gas Linearization Table Adjustment
All adjustments in this section are made from the outside of the meter via digital communication interface between a PC (terminal) and FMA
4000. There is no need to disassemble any part of the instrument or perform internal PCB component (potentiometers) adjustment.
Note: Your FMA 4000 Digital Mass Flow Meter was calibrated at the factory for the specified gas and full scale flow range (see device’s front label). There is no need to adjust the gas linearization table unless linearity adjustment is needed, flow range has to be changed, or new additional calibration is required. Any alteration of the gas linearization table will VOID calibration warranty supplied with instrument.
22
Gas flow calibration parameters are separately stored in the Gas Dependent por­tion of the EEPROM memory for each of 10 calibration tables. See APPENDIX I for complete list of gas dependent variables.
The FMA 4000 gas flow calibration involves building a table of the actual flow val­ues (indexes 114, 116, 118, 120, 122, 124, 126, 128, 130, 132, 134) and corre­sponding sensor readings (indexes 113, 115, 117, 118, 119, 121, 123, 125, 127, 129, 131, 133).
Actual flow values are entered in normalized fraction format: 100.000 % F.S. cor­responds to 1.000000 flow value and 0.000 % F.S. corresponds to 0.000000 flow value. The valid range for flow values is from 0.000000 to 1.000000 (note: FMA 4000 will accept up to 6 digits after decimal point).
Sensor readings are entered in counts of 12 bits ADC output and should always be in the range of 0 to 4095. There are 11 elements in the table so the data should be obtained at an increment of 10.0 % of full scale (0.0, 10.0, 20.0, 30.0, 40.0,
50.0, 60.0, 70.0, 80.0, 90.0 and 100.0 % F.S.).
If a new gas table is going to be created, it is recommended to start calibration from 100% full scale. If only linearity adjustment is required, calibration can be started in any intermediate portion of the gas table.
Using the flow regulator, adjust the flow rate to 100% of full scale flow. Check the flow rate indicated against the flow calibrator. Observe the flow reading on the FMA 4000. If the difference between calibrator and FMA 4000 flow reading is more than 0.5% F.S., make a correction in the sensor reading in the correspon­ding position of the linearization table (see Index 133).
If the FMA 4000 flow reading is more than the calibrator reading, the number of counts in the index 133 must be decreased. If the FMA 4000 flow reading is less than the calibrator reading, the number of counts in the index 133 must be increased. Once Index 133 is adjusted with a new value, check the FMA 4000 flow rate against the calibrator and, if required, perform additional adjustments for Index 133.
Note: Make sure the correct gas number and name selected are current. All adjustments made to the gas linearization table will be applied to the currently selected gas. Use Gas Select command via digital communication interface (see paragraph 8.3 ASCII Command Set “FMA 4000 SOFTWARE INTERFACE COMMANDS”) or Omega
®
supplied calibration and maintenance software to verify current gas table or select a new gas table.
Note: Do not alter memory index 113 (must be 120 counts) and 114 (must be 0.0). These numbers represent zero flow calibration points and should not be changed.
If a simple communication terminal is used for communication with the FMA 4000, then “MW” (Memory Write) command from the software interface commands set may be used to adjust sensor value in the linearization table (see section 8.3 for complete software interface commands list).
Memory Read “MR” command can be used to read the current value of the index. Assuming the FMA 4000 is configured with RS485 interface and has address “11”, the following example will first read the existing value of Index 133 and then write a new adjusted value:
!11,MR,133[CR] - reads EEPROM address 133
!11,MW,133,3450[CR] - writes new sensor value (3450 counts) in to the index 133
Once 100% F.S. calibration is completed, the user can proceed with calibration for another 9 points of the linearization table by using the same approach.
7.3 Analog output Calibration of FMA 4000 Mass Flow Meter
FMA 4000 series Mass Flow Meters are equipped with calibrated 0-5 Vdc and 4-20 mA output signals. The set of the jumpers (J7A, J7B, J7C) on the printed circuit board is used to switch between 0-5 Vdc and 4-20 mA output signals (Figure c-1 “FMA 4000 configuration jumpers”).
Figure c-1 FMA 4000 Analog Output Configuration Jumpers
23
Note: It is recommended to use Omega®supplied calibration and maintenance software for gas table calibration. This software includes an automated calibration procedure which may radically simplify reading and writing for the EEPROM linearization table.
FUNCTION J7A
J7B J7C
JCD
ANALOG 0-5 VDC 5-9 6-10 7-11
OUTPUT 4-20 mA 1-5 2-6 3-7
RS485 TERMINAL RESISTOR
OFF
8-12
ON 4-8
J7 Jumpers
9
10
11
12
5
6
7
8
1
2
3
4
A
B
CD
AutoZero/Reset push button.
24
The FMA 4000 analog output calibration involves calculation and storing of the offset and span variables in the EEPROM for each available output. The 0-5 Vdc output has only scale variable and 20 mA output has offset and scale variables. The following is a list of the Gas independent variables used for analog output computation:
Index Name Description
25 AoutScaleV - DAC 0-5 Vdc Analog Output Scale
27 AoutScale_mA - DAC 4-20mA Analog Output Scale
28 AoutOffset_mA - DAC 4-20mA Analog Output Offset
7.3.1 Initial Setup
Power up the Mass Flow Meter for at least 15 minutes prior to commencing the calibration procedure. Make sure absolutely no flow takes place through the meter. Establish digital RS485 / RS232 communication between PC (communi­cation terminal) and FMA 4000. The commands provided below assume that cal­ibration will be performed manually (w/o Omega
®
supplied calibration and main-
tenance software) and the device has RS485 address 11. If Omega
®
supplied cal­ibration and maintenance software is used, skip the next section and follow the software prompts.
Note: The analog output available on the FMA 4000 Digital Mass Flow Meter was calibrated at the factory for the specified gas and full scale flow range (see the device’s front label). There is no need to perform analog output calibration unless the EEPROM IC was replaced or offset/span adjustment is needed. Any alteration of the analog output scaling variables in the Gas independent table will VOID calibration warranty supplied with instrument.
Note: It is recommended to use the Omega®supplied calibration and maintenance software for analog output calibration. This software includes an automated calibration procedure which may radically simplify calculation of the offsets and spans variables and, the reading and writing for the EEPROM table.
25
Enter Backdoor mode by typing: !11,MW,1000,1[CR] Unit will respond with: !11,BackDoorEnabled: Y Disable DAC update by typing: !11,WRITE,4,D[CR] Unit will respond with: !11,DisableUpdate: D
7.3.2 Gas flow 0-5 Vdc analog output calibration
1. Install jumpers J7A, J7B and J7C on the PC board for 0-5 Vdc output (see Table VI).
2. Connect a certified high sensitivity multi meter set for the voltage measurement to the pins 2 (+) and 1 (-) of the 15 pins D connector.
3. Write 4000 counts to the DAC channel 1: !11,WRITE,1,4000[CR]
4. Read voltage with the meter and calculate:
5. Save FlowOutScaleV in to the EEPROM: !11,MW,25,X[CR]
Where: X – the calculated AoutScaleV value.
7.3.3 Gas flow 4-20 mA analog output calibration
1. Install jumpers J7A, J7B and J7C on the PC board for 4-20 mA output (see Table VI).
2. Connect a certified high sensitivity multi meter set for the current measurement to pins 2 (+) and 1 (-) of the 15 pins D connector.
3. Write 4000 counts to the DAC channel 1: !11,WRITE,1,4000[CR]
4. Read current with the meter and calculate:
5. Write zero counts to the DAC channel 1: !11,WRITE,1,0CR]
6. Read offset current with the meter and calculate:
7. Save AoutScale_mA in to the EEPROM: !11,MW,27,Y[CR] Save AoutOffset_mA in to the EEPROM: !11,MW,28,Z[CR]
Where: Y – the calculated AoutScale_mA value.
Z – the calculated AoutOffset_mA value.
Note: When done with the analog output calibration make sure the DAC update is enabled and the BackDoor is closed (see command below).
26
Enable DAC update by typing: !11,WRITE,4,N[CR] Unit will respond with: !11,DisableUpdate: N
Close BackDoor access by typing: !11,MW,1000,0[CR] Unit will respond with: !11,BackDoorEnabled: N
8. RS485 / RS232 SOFTWARE INTERFACE COMMANDS
8.1 General
The standard FMA 4000 comes with an RS485 interface. For the optional RS232 interface, the start character (!) and two hexadecimal characters for the address must be omitted. The protocol described below allows for communications with the unit using either a custom software program or a “dumb terminal.” All values are sent as printable ASCII characters. For RS485 interface, the start character is always (!). The command string is terminated with a carriage return (line feeds are automatically stripped out by the FMA 4000). See section 2.2.3 for information regarding communication parameters and cable connections.
8.2 Commands Structure
The structure of the command string:
!<Addr>,<Cmd>,Arg1,Arg2,Arg3,Arg4<CR>
Where:
! Start character ** Addr RS485 device address in the ASCII representation of hexadecimal
(00 through FF are valid).** Cmd The one or two character command from the table below. Arg1 to Arg4 The command arguments from the table below.
Multiple arguments are comma delimited. CR Carriage Return character.
Several examples of commands follow. All assume that the FMA 4000 has been configured for address 18 (12 hex) on the RS485 bus:
1. To get current calibration tables: !12,G<CR> The FMA 4000 will reply: !12,G 0 AIR<CR> (Assuming Current Gas table is #0, calibrated for AIR )
2. To get current Alarm status: !12,A,R<CR> The FMA 4000 will reply: !12,N<CR> (Assuming no alarm conditions)
3. To get a flow reading: !12,F<CR>The FMA 4000 will reply:
!12,50.0<CR> (Assuming the flow is at 50% FS)
4. Set the high alarm limit to 85% of full scale flow rate:
!12,A,H,85.0<CR>
The FMA 4000 will reply: !12,AH85.0<CR>
Note: ** Default address for all units is 11. Do not submit start character and two character hexadecimal device address for RS232 option.
27
The global address can be used to change RS485 address for a particular device with unknown address:
1. Make sure only one device (which address must be changed) is connected to the RS485 network.
2. Type the memory write command with global address: !00,MW,7,XX[CR] where XX, the new hexadecimal address, can be [01 – FF].
After assigning the new address, a device will accept commands with the new address.
Note: Address 00 is reserved for global addressing. Do not assign, the global address for any device. When command with global address is sent, all devices on the RS485 bus execute the command but do not reply with an acknowledge message.
Note: Do not assign the same RS485 address for two or more devices on the same RS485 bus. If two or more devices with the same address are connected to the one RS485 network, a communication collision will take place on the bus and communication errors will occur.
28
8.3 ASCII Commands Set
OMEGA FMA 4000 SOFTWARE INTERFACE COMMANDS
COMMAND
NAME
DESCRIPTION
No.
COMMAND SYNTAX
Command Argument 1
Argument 2 Argument 3
Argument 4 Response
Flow Requests the current flow
sensor reading in current EU
1
F <Value> (Actual flow in
current engineering units)
Diagnostic Enable / Disable LCD
Diagnostic messages (only
for LCD option).
Request current status of
the Diagnostic events, LED
status and LCD diagnostic
mode (enabled/disabled).
2
D
E (enable LCD
** Diagnostic Messages)
D:E
D (disable LCD
** Diagnostic Messages)
D:D
NO ARGUMENT
(read current status of
the diagnostic word)
D:0x0,L:9,E
0x0 – diagnostic word
9 - current LED status
E - LCD mode (enabled)
Roll back to
N
2
feature.
Enable / Disable Roll back
to N
2
feature.
3 N E
(enable Roll back to N
2
)
N:E
D
(enable Roll back to N
2
)*
N:D
NO ARGUMENT
(read current mode of
the N2 Roll back )
N:DOrN:E
Gas Select
Selects one of the ten
primary gas calibration
tables to use. Tables are
entered via the MEM
commands at time of
calibration.
4 G
0 (gas 0)*to9 (gas 9)
G0 through G9,
<Gas Name>
NO ARGUMENT
(read status)
G0 through G9,
<Gas Name>
Note: An “*” indicates power up default settings. An “**” indicates optional feature not available on all models.
29
COMMAND
NAME
DESCRIPTION
No.
COMMAND SYNTAX
Command
Argument 1
Argument 2
Argument 3
Argument 4 Response
Auto
Zero
Starts /reads the status of
the auto zero feature (Note:
The Z,N command can be
used only when absolutely
no flow thru the meter and
no earlier then 6 minutes after
power up. It can take several
minutes to complete. Unit will
not respond to other commands
when this is in progress.)
5 Z
N (do it now) ZN
W (Write Zero to
EEPROM)
ZW (when done)
S (status while auto
zero in progress)
ZNI,<value> while Z,N
is in progress.
V (Display zero value)
ZV,<zero value>
Flow
Alarms
Sets / reads the status of
the gas flow alarms.
Note: High and Low limits
have to be entered in the
%F.S. High alarm value
has to be more than Low
alarm value.
Alarm conditions:
Flow > High Limit = H
Flow < Low Limit = L
Low < Flow < High = N
6 A H (high flow limit) <Value> (0-100%FS) AH<Value>
L (low flow limit) <Value> (0-100%FS)
AL<Value>
A (action delay in seconds)
<Value> (0-3600 sec.)
AA:<Value>
E (enable alarm) AE
D (disable alarm)* AD
R (read current status)
N (no alarm)
H (high alarm)
L (low alarm)
S (Read current settings)
AS:M,L,H,D,B where:
M – mode (E/D)
L – Low settings (%FS)
H – High settings (%FS)
D – Action Delay (sec)
B – Latch mode (0-3)
B Block (Latch) mode
<Value> (0-disabled*)
(1-enabl’d L)
(2-enabl’d H)
(3 –both L,H)
AB:<Value> where:
Value = 0 - 3
30
COMMAND
NAME
DESCRIPTION
No.
COMMAND SYNTAX
Command
Argument 1
Argument 2 Argument 3 Argument 4 Response
Relay
Action
Assigns action of the two SPDT
relays. The coil is energized
when the condition specified by
an Argument 2 becomes true.
Argument 2:
N - no action, relay disabled*
T - totalizer reading > limit
H - high flow alarm
L - low flow alarm
R - Range between High &
Low alarms
M - Manual Relay overdrive
S - Read current status
7
R 1 (relay 1)
2 (relay 2)
N* R1N or R2N
T R1T or R2T
H R1H or R2H
L R1L or R2L
R
R1R or R2R
M
R1M or R2M
S
RxN, RxT, RxH,
RxL, RxR , RxM
Totalizer
Sets and controls action of the
flow totalizer.
NOTE: If Warm Up Delay option
is set to E (enabled) the Totalizer
will not totalize the flow during
first 6 minutes after power up.
8 T
Z (Reset to zero)
TZ
F (start totalizer at flow F.S.)
L (Limit gas volume in current E.U.)
<value>
(flow %FS)
TF<value>
<value>
(gas volume)
TL<value>
D (disable totalizer)
TD
E (enable totalizer)
TE
R (read current totalizer volume) <value> (in current EU)
W (Warm Up Delay)
E – enable
D – disable*
TW:E or TW:D
S (setting status)
TS: Mode, Start,
Limit, Warm Up
31
COMMAND
NAME
DESCRIPTION
No.
COMMAND SYNTAX
Command Argument 1
Argument 2 Argument 3
Argument 4 Response
K Factors
Applies a gas correction
factor to the currently
selected primary gas
calibration table.
(NOTE: does not work with
% F.S. engineering unit.)
See list of the internal
K-factors in the operating
manual.
9 K
D*(disable, sets K=1) KD
I (Internal K-factor) No argument
(enable previously set
internal K-factor)
KI,<value>,<Gas>
Gas Index (0 -35) KI,<Index>,<Gas>
U (user specified factor) No argument
(enable previously set
user K-factor)
KU,<value>
<value>
(decimal correction
factor) (0-1000)
KU,<value>
S (status) SK, <Mode>, <Index>,
<Value> where:
Mode: D, I, U
Index: 0-35
Value: K-Factor value
32
COMMAND
NAME
DESCRIPTION
No.
COMMAND SYNTAX
Command Argument 1 Argument 2 Argument 3 Argument 4 Response
Units
Set the units of
measure for gas
flow and totalizer
reading.
Note: The units
of the totalizer
output are not
per unit time.
10
U
% (% full scale)* U:%
mL/sec
U:mL/sec
mL/min
U:mL/min
mL/hr U:mL/hr
L/sec U:L/sec
L/min
U:L/min
L/hr U:L/hr
m3/sec
U:m3/sec
m3/min
U:m3/min
m3/hr U:m3/hr
f3/sec U:f3/sec
f3/min U:f3/min
f3/hr U:f3/hr
g/sec
U:g/sec
g/min
U:g/min
g/hr U:g/hr
kg/sec U:kg/sec
kg/min
U:kg/min
kg/hr U:kg/hr
Lb/sec U:Lb/sec
Lb/min
U:Lb/min
Lb/hr
U:Lb/hr
USER (user defined)
<value> (conversion
factor from L/min)
S - seconds
M – minutes
H – hours (Time base)
Y - use density
N – do not use
density
U:USER,<Factor>,
<Time base>,
<Density mode>
No Argument
<status>Returns current EU.
U,<EU name>
33
UART Error Codes:
1 - Not Supported Command or Back Door is not enabled.
2 - Wrong # of Arguments.
3 - Address is Out of Range (MR or MW commands).
4 - Wrong # of the characters in the Argument.
5 - Attempt to Alter Write Protected Area in the EEPROM.
6 - Proper Command or Argument is not found.
7 - Wrong value of the Argument.
8 - Reserved.
9 - Manufacture specific info EE KEY (wrong key or key is disabled).
COMMAND
NAME
DESCRIPTION
No.
COMMAND SYNTAX
Command Argument 1 Argument 2 Argument 3
Argument 4
Response
Maintenance
Timer
Hours since last time unit
was calibrated.
11
C
R (read timer) <Value> (in Hours)
C(set timer to zero) CC
Full Scale
Returns the full scale rated flow in
L/min. (Note: This term is not
multiplied by the current K factor)
12
E
<Value>
(in L/min)
LCD Back
Light
LCD Back Light control
(0-100.0%)
0 - off
100 - Maximum Intensity
13 B 0 to 100% B:<Counts>
where:
Counts (0 – 4095)
B:<Value>
where:
Value (0 – 100.0)
No Argument
<Current settings>
Read
EEPROM
Memory
Reads the value in the
specified memory location.
14 MR 0000 to999
(Table Index)
<value>
Write
EEPROM
Memory
Writes the specified value to the
specified memory location. Use
Carefully, can cause unit to
malfunction. (Note: Some addresses
are write protected!)
15
MW 0000 to 999
(Table Index)
Value
MW,XXX,<Value>
where:
XXX=Table Index
34
9. TROUBLESHOOTING
9.1 Common Conditions
Your FMA 4000 Digital Mass Flow Meter was thoroughly checked at numerous quality control points during and after manufacturing and assembly operations. It was calibrated according to your desired flow and pressure conditions for a given gas or a mixture of gases.
It was carefully packed to prevent damage during shipment. Should you feel that the instrument is not functioning properly, please check for the following common conditions first:
Are all cables connected correctly? Are there any leaks in the installation? Is the power supply correctly selected according to requirements? When several meters are used a power supply with appropriate current rating should be selected.
Were the connector pinouts matched properly? When interchanging with other manufacturers' equipment, cables and connectors must be carefully wired for cor­rect pin configurations. Is the pressure differential across the instrument sufficient?
9.2 Troubleshooting Guide
35
NO. INDICATION LIKELY REASON SOLUTION
1 No zero reading after
15 min. warm up time and no flow condition.
Embedded temperature has been changed.
Perform Auto Zero Procedure (see section
5.3.6 “Zero Calibration”).
2 Status LED indicator
and LCD Display remains blank when unit is powered up. No response when flow is introduced from analog outputs 0-5 Vdc or 4-20 mA.
Power supply is bad or polarity is reversed.
Measure voltage on pins 7 and 5 of the 15 pin D-connector. If voltage is out of specified range, then replace power supply with a new one. If polarity is reversed (reading is negative) make correct connection.
PC board is defective. Return FMA 4000 to factory for repair.
3 LCD Display reading or
/ and analog output 0-5 Vdc signal fluctuate in wide range during flow measurement.
Output 0-5 Vdc signal (pins 2–1 of the D-connector) is shorted on the GND or overloaded.
Check external connections to pin 2 – 1, of the D-connector. Make sure the load resistance is more than 1000 Ohm.
4 LCD Display reading
does correspond to the correct flow range, but 0-5 Vdc output signal does not change (always the same read ing or around zero).
Output 0-5 Vdc schematic is burned out or damaged.
Return FMA 4000 to factory for repair.
Analog flow output scale and offset variable are corrupted.
Restore original EEPROM scale and offset variable or perform analog output recalibration (see section 7.3).
5 LCD Display reading
and 0-5 Vdc output voltage do correspond to the correct flow range, but 4-20 mA output signal does not change (always the same or reading around 4.0 mA).
External loop is open or load resistance more than 500 Ohm.
Check external connections to pins 2 and 15 of the D-connector. Make sure the loop resistance is less than 500 Ohm.
Output 4-20 mA schematic is burned out or damaged.
Return FMA 4000 to factory for repair.
6 Calibration is off (more
than ±1.0 % F.S.).
FMA 4000 has initial zero shift.
Shut off the flow of gas into the FMA 4000 (ensure gas source is disconnected and no seepage or leak occurs into the meter). Wait for 15 min. with no flow condition and perform Auto Zero calibration Procedure (see section 5.3.7 “Zero Calibration”).
7 LCD Display reading is
above maximum flow range and output volt age 0-5 Vdc signal is more than 5.0 Vdc when gas flows through the FMA 4000.
Sensor under swamping conditions (flow is more than 10% above maximum flow rate for particular FMA 4000).
Lower the flow through FMA 4000 within calibrated range or shut down the flow completely. The swamping condition will end automatically.
PC board is defective. Return FMA 4000 to factory for repair.
36
NO. INDICATION LIKELY REASON SOLUTION
8 Gas flows through the
FMA 4000, but LCD Display reading and the output voltage 0-5 Vdc signal do not respond to flow.
The gas flow is too low for particular model of FMA 4000.
Check maximum flow range on transducer’s front panel and make required flow adjustment.
FMA 4000 models: RFE is not connected properly to the inlet fitting.
Unscrew the inlet compression fitting of the meter and reinstall RFE (see section 6.2.2). NOTE: Calibration accuracy can be affected.
Sensor or PC board is defective.
Return FMA 4000 to factory for repair.
9 Gas does not flow
through the FMA 4000 with inlet pressure applied to the inlet fitting. LCD Display reading and output voltage 0-5 Vdc signal show zero flow.
Filter screen obstructed at inlet.
Flush clean or disassemble to remove impediments or replace the filter screen (see section 6.2).
NOTE: Calibration accuracy can be affected.
10 Gas flows through the
FMA 4000, output voltage 0-5 Vdc signal does not respond to flow (reading near 1mV).
Direction of the gas flow is reversed.
Check the direction of gas flow as indicated by the arrow on the front of the meter and make required reconnection in the installation.
FMA 4000 is connected
in the installation with back pressure conditions and gas leak exist in the system.
Locate and correct gas leak in the system.
If FMA 4000 has internal leak return it to factory for repair.
11 The Status LED
indicator is rapidly flashing with UMBER color on /off.
Sensor temperature is too low.
Make sure the ambient and gas temperatures are within specified range (above 5
F
C)
12 The Status LED
indicator is rapidly flashing with RED color on /off.
Sensor temperature is too high.
Make sure the ambient and gas temperatures are within specified range (below 50
F
C).
13 The Status LED
indicator is rapidly flashing with RED and UMBER colors.
MCU temperature is too high (overload).
Disconnect power from the FMA 4000.
Make sure the ambient temperature is with in specified range (below 50
F
C). Let the device cool down for at least 15 minutes. Apply power to the FMA 4000 and check status LED indication. If overload condition will be indicated again the unit has to be returned to the factory for repair.
14 The Status LED
indicator is constantly on with the RED light.
Fatal Error (EEPROM or Auto Zero error).
Cycle the power on the FMA 4000. If Status
LED still constantly on with RED light, wait 6 minutes and start Auto Zero function (see
5.3.7 Zero Calibration). If after Zero Calibration the Fatal Error condition will be indicated again the unit has to be returned to the factory for repair.
9.3 Technical Assistance
OMEGA7 Engineering will provide technical assistance over the phone to quali­fied repair personnel. Please call our Flow Department at 800-872-9436 Ext.
2298. Please have your Serial Number and Model Number ready when you call.
37
Q
O2
= Q
a
= Q
r
X K = 1000 X 0.9926 = 992.6 sccm
where K = relative K factor to reference gas (oxygen to nitrogen)
1
d X C
p
where d = gas density (gram/liter) C
p
= coefficient of specific heat (cal/gram)
Q
a
K
a
Q
r
K
r
where Qa= mass flow rate of an actual gas (sccm) Q
r
= mass flow rate of a reference gas (sccm)
K
a
= K factor of an actual gas
K
r
= K factor of a reference gas
=
10. CALIBRATION CONVERSIONS FROM REFERENCE GASES
The calibration conversion incorporates the K factor. The K factor is derived from gas density and coefficient of specific heat. For diatomic gases:
=
K
=
K
gas
Note in the above relationship that d and Cp are usually chosen at the same con­ditions (standard, normal or other).
If the flow range of a Mass Flow Meter remains unchanged, a relative K factor is used to relate the calibration of the actual gas to the reference gas.
For example, if we want to know the flow rate of oxygen and wish to calibrate with nitrogen at 1000 SCCM, the flow rate of oxygen is:
Note: If particular K factor is activated via digital interface, the user does not need to perform any conversion. All conversion computations will be performed internally by MCU.
38
INDEX NAME DATA TYPE NOTES
0
BlankEEPROM char[10] Do not modify. Table Revision [PROTECTED]
1
SerialNumber char[20] Serial Number [PROTECTED]
2
ModelNumber char[20] Model Number [PROTECTED]
3
SoftwareVer char[10] Firmware Version [PROTECTED]
4
TimeSinceCalHr float Time since last calibration in hours.
5
Options1 uint Misc. Options*
6
BackLight int Back Light Level [0-4095]
7
AddressRS485 char[4] Two character address for RS485 only
8
GasNumber int Current Gas Table Number [0-9]
9
FlowUnits int Current Units of Measure [0-22]
10
AlarmMode char Alarm Mode ['E’- Enabled, 'D’ - Disabled]
11
LowAlarmPFS float Low Flow Alarm Setting [%FS] 0-Disabled
12
HiAlarmPFS float High Flow Alarm Setting [%FS] 0-Disabled
13
AlmDelay uint Flow Alarm Action Delay [0-3600sec] 0-Disabled
14
RelaySetting char[4] Relays Assignment Setting (N, T, H, L, R, M)
15
TotalMode char Totalizer Mode ['E’- Enabled, 'D’ - Disabled]
16
Total float Totalizer Volume in %*s (updated every 6 min)
17
TotalFlowStart float Start Totalizer at flow [%FS] 0 - Disabled
18
TotalVolStop float Totalizer Action Limit Volume [%*s] 0-Disabled
19
KfactorMode char D-Disabled, I-Internal, U-User Defined
20
KfactorIndex int Internal K-Factor Index [0-35]**
21
UserDefKfactor float User Defined K-Factor
22
UDUnitKfactor float K-Factor for User Defined Units of Measure
23
UDUnitTimeBase int User Defined Unit Time Base [1, 60, 3600 sec]
24
UDUnitDensity char User Defined Unit Density Flag [Y, N]
25
AoutScaleV float DAC 0-5 Vdc Analog Output Scale
26
DRC_DP float H/W DRC DP settings [0-255]
27
AoutScale_mA float DAC 4-20mA Analog Output Scale
28
AoutOffset_mA float DAC 4-20mA Analog Output Offset
29
SensorZero uint DPW value for Sensor Zero [0-1023]
30
Klag [0] float DRC Lag Constant [Do Not Alter]
31
Klag [1] float DRC Lag Constant [Do Not Alter]
32
Klag [2] float DRC Lag Constant [Do Not Alter]
33
Klag [3] float DRC Lag Constant [Do Not Alter]
34
Klag [4] float DRC Lag Constant [Do Not Alter]
APPENDIX I
OMEGA7 FMA 4000 EEPROM Variables Rev.A0 [10/2/2007]
Gas Independent Variables
39
INDEX NAME DATA TYPE NOTES
35
Klag [5] float DRC Lag Constant [Do Not Alter]
36
Kgain[0] float Gain for DRC Lag Constant [Do Not Alter]
37
Kgain[1] float Gain for DRC Lag Constant [Do Not Alter]
38
Kgain[2] float Gain for DRC Lag Constant [Do Not Alter]
39
Kgain[3] float Gain for DRC Lag Constant [Do Not Alter]
40
Kgain[4] float Gain for DRC Lag Constant [Do Not Alter]
41
Kgain[5] float Gain for DRC Lag Constant [Do Not Alter]
42
Zero_T float Resistance when last AutoZero was done [0-4095 count]
43
Tcor_K float Resistance correction coefficient [PFS/count]
44
AlarmLatch uint Alarm Latch [0-3]
45
TotalWarmDisable char Sensor Warm Up period Totalizer [D/E]
46
Reserved1 uint Reserved
47
LCD_Diagnostic char LCD Diagnostic Mode: [E/D]**
48
Reserved2 uint
Flow Reading Averaging: [0,1,2] (100, 250, 1000 ms), Default -1
49
N2_RollBack
char
Back to N2conversion mode: [E, D]
50
Reserved3 uint Reserved for Troubleshooting (do not change)
40
INDEX NAME DATA TYPE NOTES
100
GasIdentifer char[20] Name of Gas [If not calibrated = “Uncalibrated”]
101
FullScaleFlow float Full Scale Range in l/min
102
StdTemp float Standard Temperature
103
StdPressure float Standard Pressure
104
StdDensity float Gas Standard Density
105
CalibrationGas char[20]
Name of Gas used for Calibration
[If not calibrated=[“Uncalibrated”]
106
CalibratedBy char[20] Name of person who performed actual calibration
107
CalibratedAt char[20] Name of Calibration Facility
108
DateCalibrated char[12] Calibration Date
109
DateCalibrationDue char[12] Date Calibration Due
110
K_N
2
float Gas Parameters: K-factor relative to N
2
111
K_F1 float Reserved
112
K_F1 float Reserved
113
SensorTbl[0][Sensor Value] uint Index 0: Must be 120 (zero value) Do not Alter!
114
SensorTbl[0][Flow] float Index 0: Must be 0.0 (zero PFS) Do not Alter!
115
SensorTbl[1][Sensor Value] uint 10.0%F.S. A/D value from sensor [counts].
116
SensorTbl[1][Flow] float Actual Flow in PFS [0.1].
117
SensorTbl[2][Sensor Value] uint 20.0%F.S. A/D value from sensor [counts].
118
SensorTbl[2][Flow] float Actual Flow in PFS [0.2].
119
SensorTbl[3][Sensor Value] uint 30.0%F.S. A/D value from sensor [counts].
120
SensorTbl[3][Flow] float Actual Flow in PFS [0.3].
121
SensorTbl[4][Sensor Value] uint 40.0%F.S. A/D value from sensor [counts].
122
SensorTbl[4][Flow] float Actual Flow in PFS [0.4].
123
SensorTbl[5][Sensor Value] uint 50.0%F.S. A/D value from sensor [counts].
124
SensorTbl[5][Flow] float Actual Flow in PFS [0.5].
125
SensorTbl[6][Sensor Value] uint 60.0%F.S. A/D value from sensor [counts].
126
SensorTbl[6][Flow] float Actual Flow in PFS [0.6].
127
SensorTbl[7][Sensor Value] uint 70.0%F.S. A/D value from sensor [counts].
128
SensorTbl[7][Flow] float Actual Flow in PFS [0.7].
129
SensorTbl[8][Sensor Value] uint 80.0%F.S. A/D value from sensor [counts].
130
SensorTbl[8][Flow] float Actual Flow in PFS [0.8].
131
SensorTbl[9][Sensor Value] uint 90.0%F.S. A/D value from sensor [counts].
132
SensorTbl[9][Flow] float Actual Flow in PFS [0.9].
133
SensorTbl[10][Sensor Value] uint 100.0%F.S. A/D value from sensor [counts].
134
SensorTbl[10][Flow] float Flow in PFS. Should be 1.0 Do not Alter!
Calibration Table: Gas Dependent Variables.
Note: Values will be available for selected gas only.
41
INDEX ACTUAL GAS
K Factor Relative
to N
2
Cp
[Cal/g]
DENSITY
[g/I]
0 Acetylene C2H2 0.5829 .4036 1.162 1 Air 1.000 0.24 1.293 2 Allene (Propadiene) C3H4 0.4346 0.352 1.787
3 Ammonia NH
3
.7310 .492 .760
4 Argon Ar 1.4573 .1244 1.782
5 Arsine AsH3 0.6735 0.1167 3.478
6 Boron Trichloride BCl3 0.4089 0.1279 5.227
7 Boron Triflouride BF3 0.5082 0.1778 3.025
8 Bromine Br2 0.8083 0.0539 7.130
9 Boron Tribromide Br3 0.38 0.0647 11.18
10 Boromine Pentaflouride BrF5 0.26 0.1369 7.803
11 Boromine Triflouride BrF3 0.3855 0.1161 6.108
12 Bromotriflouromethane CBrF3 0.3697 0.1113 6.644
13 1,3-Butadiene C4H6 0.3224 0.3514 2.413
14 Butane C4H
10
.2631 .4007 2.593
15 1-Butane C4H8 0.2994 0.3648 2.503
16 2-Butane C4H8 CIS 0.324 0.336 2.503
17 2-Butane C4H8 TRANS 0.291 0.374 2.503
18 Carbon Dioxide CO
2
.7382 .2016 1.964
19 Carbon Disulfide CS
2
0.6026 0.1428 3.397
20
Carbon Monoxide C
O
1.00 .2488 1.250
21 Carbon Tetrachloride CCl4 0.31 0.1655 6.860
22 Carbon Tetrafluoride (Freon-14) CF4 0.42 0.1654 3.926
23 Carbonyl Fluoride COF2 0.5428 0.1710 2.945
24 Carbonyl Sulfide COS 0.6606 0.1651 2.680
25 Chlorine Cl
2
0.86 0.114 3.163
26 Chlorine Trifluoride ClF3 0.4016 0.1650 4.125 27 Chlorodifluoromethane (Freon-22) CHClF2 0.4589 0.1544 5.326
28 Chloroform CHCl
3
0.3912 0.1309 5.326
29 Chloropentafluoroethane (Freon-115) C2ClF5 0.2418 0.164 6.892
30 Chlorotrifluoromethane (Freon-13) CClF3 0.3834 0.153 4.660
31 Cyanogen C2N2 0.61 0.2613 3.322 32 Helium He 1.454 1.241 .1786 33 Hydrogen H
2
1.0106 3.419 .0899
34 Hydrogen H2(> 100 L/min) 1.92 3.419 0.0899
35 Oxygen O2 0.9926 0.2193 1.427
APPENDIX II INTERNAL “K” FACTORS
CAUTION: K-Factors at best are only an approximation. K factors should not
be used in applications that require accuracy better than +/- 5 to 10%.
42
APPENDIX III GAS FACTOR TABLE (“K FACTORS”)
CAUTION: K-Factors at best are only an approximation. K factors should not
be used in applications that require accuracy better than +/- 5 to 10%.
ACTUAL GAS
K FACTOR
Relative to N
2
Cp
[Cal/g]
Density
[g/I]
Acetylene C2H
2
.5829 .4036 1.162 Air 1.0000 .240 1.293 Allene (Propadiene) C3H
4
.4346 .352 1.787 Ammonia NH
3
.7310 .492 .760 *Argon Ar (<=10 L/min)
*Argon AR-1 (>=10 L/min)
1.4573
1.205
.1244 .1244
1.782
1.782
Arsine AsH
3
.6735 .1167 3.478 Boron Trichloride BCl
3
.4089 .1279 5.227 Boron Trifluoride BF
3
.5082 .1778 3.025 Bromine Br
2
.8083 .0539 7.130 Boron Tribromide Br
3
.38 .0647 11.18 Bromine PentaTrifluoride BrF
5
.26 .1369 7.803 Bromine Trifluoride BrF
3
.3855 .1161 6.108 Bromotrifluoromethane (Freon-13 B1) CBrF
3
.3697 .1113 6.644
1,3-Butadiene C4H
6
.3224 .3514 2.413
Butane C4H
10
.2631 .4007 2.593
1-Butene C4H
8
.2994 .3648 2.503
2-Butene C4H8 CIS
.324 .336 2.503
2-Butene C4H8TRANS
.291 .374 2.503
*Carbon Dioxide CO
2
(<10 L/min)
*Carbon Dioxide CO
2
-1 (>10 L/min)
.7382 .658
.2016 .2016
1.964
1.964
Carbon Disulfide CS
2
.6026 .1428 3.397
Carbon Monoxide C0
1.00 .2488 1.250
Carbon Tetrachloride CCl
4
.31 .1655 6.860
Carbon Tetrafluoride (Freon-14)CF
4
.42 .1654 3.926
Carbonyl Fluoride COF
2
.5428 .1710 2.945
Carbonyl Sulfide COS
.6606 .1651 2.680
Chlorine Cl
2
.86 .114 3.163
Chlorine Trifluoride ClF
3
.4016 .1650 4.125
Chlorodifluoromethane (Freon-22)CHClF
2
.4589 .1544 3.858
Chloroform CHCl
3
.3912 .1309 5.326
Chloropentafluoroethane(Freon-115)C2ClF
5
.2418 .164 6.892
Chlorotrifluromethane (Freon-13) CClF
3
.3834 .153 4.660
CyanogenC2N
2
.61 .2613 2.322
CyanogenChloride CICN
.6130 .1739 2.742
Cyclopropane C3H
5
.4584 .3177 1.877
* Flow rates indicated ( ) is the maximum flow range of the Mass Flow meter being used.
43
ACTUAL GAS
K FACTOR
Relative to N
2
Cp
[Cal/g]
Density
[g/I]
Deuterium D
2
1.00 1.722 1.799
Diborane B2H
6
.4357 .508 1.235
Dibromodifluoromethane CBr2F
2
.1947 .15 9.362
Dichlorodifluoromethane (Freon-12) CCl2F
2
.3538 .1432 5.395
Dichlofluoromethane (Freon-21) CHCl2F
.4252 .140 4.592
Dichloromethylsilane (CH3)2SiCl
2
.2522 .1882 5.758
Dichlorosilane SiH2Cl
2
.4044 .150 4.506
Dichlorotetrafluoroethane (Freon-114) C2Cl2F
4
.2235 .1604 7.626
1,1-Difluoroethylene (Freon-1132A) C2H2F
2
.4271 .224 2.857
Dimethylamine (CH3)2NH
.3714 .366 2.011
Dimethyl Ether (CH3)2O
.3896 .3414 2.055
2,2-Dimethylpropane C3H
12
.2170 .3914 3.219
Ethane C2H
6
.50 .420 1.342
Ethanol C2H6O
.3918 .3395 2.055
Ethyl Acetylene C4H
6
.3225 .3513 2.413
Ethyl Chloride C2H5Cl
.3891 .244 2.879
Ethylene C2H
4
.60 .365 1.251
Ethylene Oxide C2H4O
.5191 .268 1.965
Fluorine F
2
.9784 .1873 1.695
Fluoroform (Freon-23) CHF
3
.4967 .176 3.127
Freon-11 CCl3F
.3287 .1357 6.129
Freon-12 CCl2F
2
.3538 .1432 5.395
Freon-13 CClF
3
.3834 .153 4.660
Freon-13B1 CBrF
3
.3697 .1113 6.644
Freon-14 CF
4
.4210 .1654 3.926
Freon-21 CHCl2F
.4252 .140 4.592
Freon-22 CHClF
2
.4589 .1544 3.858
Freon-113 CCl2FCClF
2
.2031 .161 8.360
Freon-114 C2Cl2F
4
.2240 .160 7.626
Freon-115 C2ClF
5
.2418 .164 6.892
Freon-C318 C4F
8
.1760 .185 8.397
Germane GeH
4
.5696 .1404 3.418
Germanium Tetrachloride GeCl
4
.2668 .1071 9.565
*Helium He (<50 L/min) *Helium He-1 (>50 L/min) *Helium He-2 (>10-50 L/min)
1.454
2.43
2.05
1.241
1.241
1.241
.1786 .1786 .1786
Hexafluoroethane C2F6(Freon-116)
.2421 .1834 6.157
Hexane C6H
14
.1792 .3968 3.845
*Hydrogen H2-1 (<10-100 L) *Hydrogen H
2
-2 (>10-100 L)
*Hydrogen H
2
-3 (>100 L)
1.0106
1.35
1.9
3.419
3.419
3.419
.0899 .0899 .0899
* Flow rates indicated ( ) is the maximum flow range of the Mass Flow meter being used.
44
ACTUAL GAS
K FACTOR
Relative to N
2
Cp
[Cal/g]
Density
[g/I]
Hydrogen Bromide HBr
1.000 .0861 3.610
Hydrogen Chloride HCl
1.000 .1912 1.627
Hydrogen Cyanide HCN
.764 .3171 1.206
Hydrogen Fluoride HF
.9998 .3479 .893
Hydrogen Iodide HI
.9987 .0545 5.707
Hydrogen Selenide H2Se
.7893 .1025 3.613
Hydrogen Sulfide H2S
.80 .2397 1.520
Iodine Pentafluoride IF
5
.2492 .1108 9.90
Isobutane CH(CH3)
3
.27 .3872 3.593
Isobutylene C4H
6
.2951 .3701 2.503
Krypton Kr
1.453 .0593 3.739
*Methane CH4(<=10 L/min) *Methane CH
4
-1 (>=10 L/min)
.7175
.75
.5328 .5328
.7175 .7175
Methanol CH
3
.5843 .3274 1.429
Methyl Acetylene C3H
4
.4313 .3547 1.787
Methyl Bromide CH2Br
.5835 .1106 4.236
Methyl Chloride CH3Cl
.6299 .1926 2.253
Methyl Fluoride CH3F
.68 .3221 1.518
Methyl Mercaptan CH3SH
.5180 .2459 2.146
Methyl Trichlorosilane (CH3)SiCl
3
.2499 .164 6.669
Molybdenum Hexafluoride MoF
6
.2126 .1373 9.366
Monoethylamine C2H5NH
2
.3512 .387 2.011
Monomethylamine CH3NH
2
.51 .4343 1.386
Neon NE
1.46 .246 .900
Nitric Oxide NO
.990 .2328 1.339
Nitrogen N
2
1.000 .2485 1.25
Nitrogen Dioxide NO
2
.737 .1933 2.052
Nitrogen Trifluoride NF
3
.4802 .1797 3.168
Nitrosyl Chloride NOCl
.6134 .1632 2.920
Nitrous Oxide N2O
.7128 .2088 1.964
Octafluorocyclobutane (Freon-C318) C4F
8
.176 .185 8.397
Oxygen O
2
.9926 .2193 1.427
Oxygen Difluoride OF
2
.6337 .1917 2.406
Ozone
.446 .195 2.144
Pentaborane B5H
9
.2554 .38 2.816
Pentane C5H
12
.2134 .398 3.219
Perchloryl Fluoride ClO3F
.3950 .1514 4.571
Perfluoropropane C3F
8
.174 .197 8.388
Phosgene COCl
2
.4438 .1394 4.418
Phosphine PH
3
.759 .2374 1.517
* Flow rates indicated ( ) is the maximum flow range of the Mass Flow meter being used.
45
ACTUAL GAS
K FACTOR
Relative to N
2
Cp
[Cal/g]
Density
[g/I]
Phosphorous Oxychloride POCl
3
.36 .1324 6.843
Phosphorous Pentafluoride PH
5
.3021 .1610 5.620
Phosphorous Trichloride PCl
3
.30 .1250 6.127
Propane C3H
8
.35 .399 1.967
Propylene C3H
6
.40 .366 1.877
Silane SiH
4
.5982 .3189 1.433
Silicon Tetrachloride SiCl
4
.284 .1270 7.580
Silicon Tetrafluoride SiF
4
.3482 .1691 4.643
Sulfur Dioxide SO
2
.69 .1488 2.858
Sulfur Hexafluoride SF
6
.2635 .1592 6.516
Sulfuryl Fluoride SO2F
2
.3883 .1543 4.562
Tetrafluoroethane (Forane 134A) CF3CH2F
.5096 .127 4.224
Tetrafluorohydrazine N2F
4
.3237 .182 4.64
Trichlorofluoromethane (Freon-11) CCl3F
.3287 .1357 6.129
Trichlorosilane SiHCl
3
.3278 .1380 6.043
1,1,2-Trichloro-1,2,2 Trifluoroethane (Freon-113) CCl
2
FCClF
2
.2031 .161 8.36
Triisobutyl Aluminum (C4H9)AL
.0608 .508 8.848
Titanium Tetrachloride TiCl
4
.2691 .120 8.465
Trichloro Ethylene C2HCl
3
.32 .163 5.95
Trimethylamine (CH3)3N
.2792 .3710 2.639
Tungsten Hexafluoride WF
6
.2541 .0810 13.28
Uranium Hexafluoride UF
6
.1961 .0888 15.70
Vinyl Bromide CH2CHBr
.4616 .1241 4.772
Vinyl Chloride CH2CHCl
.48 .12054 2.788
Xenon Xe
1.44 .0378 5.858
46
APPENDIX IV COMPONENT DIAGRAM
TOP COMPONENT SIDE
Aug. 09, 2007
47
BOTTOM COMPONENT SIDE
Aug 09, 2007
APPENDIX V
DIMENSIONAL DRAWINGS
FMA 4000 WITHOUT READOUT
48
FMA 4000 WITH READOUT OPTION
49
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to con­tact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA nei­ther assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PUR­POSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a
“Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the
COST of the repair,
2. Model and serial number of the
product, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2001 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photo-
copied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
50
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA… Of Course!
Shop online at www.omega.com
TEMPERATURE
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
Wire: Thermocouple, RTD & Thermistor
Calibrators & Ice Point References
Recorders, Controllers & Process Monitors
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
Transducers & Strain Gages
Load Cells & Pressure Gages
Displacement Transducers
Instrumentation & Accessories
FLOW/LEVEL
Rotameters, Gas Mass Flow Meter & Flow Computers
Air Velocity Indicators
Turbine/Paddlewheel Systems
Totalizers & Batch Controllers
pH/CONDUCTIVITY
pH Electrodes, Testers & Accessories
Benchtop/Laboratory Meters
Controllers, Calibrators, Simulators & Pumps
Industrial pH & Conductivity Equipment
DATA ACQUISITION
Data Acquisition & Engineering Software
Communications-Based Acquisition Systems
Plug-in Cards for Apple, IBM & Compatibles
Datalogging Systems
Recorders, Printers & Plotters
HEATERS
Heating Cable
Cartridge & Strip Heaters
Immersion & Band Heaters
Flexible Heaters
Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
Metering & Control Instrumentation
Refractometers
Pumps & Tubing
Air, Soil & Water Monitors
Industrial Water & Wastewater Treatment
pH, Conductivity & Dissolved Oxygen Instruments
M-4651/0508
Loading...