OPERATIONAL FLOW RANGE: From 0-10 SCCM to 0-50 SLM
FMA-1900 SERIES DE is found to operate within the differential standard of 5-50 PSI (0.35-3.5 kg/cm²) with the optimum performance being at a at optimum of 30 PSI (2 kg/cm²).
TYPES of GASES: A wide range of gases including air, nitrogen, methane, carbon dioxide, argon, helium and hydrogen are reproducible with this model.
OUTPUT CONNECTORS: The devices come with linear output of 0-5 volts direct current into 2000 ohms minimum load resistance standard or linear 4-20 mA into 1000 split load adjusted maximum for 24 VDC supply voltage
CONTROL RANGE: Calibrated between 10 to 100 % of total
INPUT VOLTAGE REQUIREMENTS: 24 VDC regulated power supply 50mV RMS ripple with capability of delivering power not less than 6 watts (this is at 250mA) at a voltage of 24 VDC +10 percent optionally
ACCURACY: +1.5 percentage of full scale from 10 to 100 percent of full scale.
REPEATABILITY: +0.25 percent of full scale
TEMPERATURE COEFFICIENT: Maximum of 0.08 percent of full scale per degree centigrade.
PRESSURE COEFFICIENT: Maximum of 0.02 percent of full scale per PSI (0.07 kg/cm²).
REACTION TIME: One second to 63% of final value
GAS PRESSURE: Maximum of 10 kg/cm² gauge of 150 PSI is quite sufficient
LEAK INTEGRITY: This model covers a wide range of performance defect tolerances of membrane gas leak, up to 1x 10 -6 .
GAS AND AMBIENT TEMPERATURE: Operational temperature range of 32° to 122°F (0 to 50°C).
Q: What is the Maximum limit of flow which can the FMA-1900 pass?
A: The FMA-1900 is manufactured to pass a maximum flow rate measuring 0-50 standard litres of liquid in one minute.
Q: Which output signals does FMA-1900 give?
A: The signals are linear 0-5 VDC and 4-20 mA output signals.
Q: What is the power supply that is needed to run the FMA-1900?
A: The power supply needed for the FMA-1900 is 24 VDC and is regulated with least 6 watts (250mA) @ 24 VDC +10%.
Q: With which gases can FMA-1900 be used?
A: Many gases can be used with FMA-1900 such as air, nitrogen, methane, carbon dioxide, argon, helium, and hydrogen.
Q: What to do in case FMA-1900 goes to overrange condition ?
A: An overrange condition, is when the display or output reaches the maximum limit of its full scale range. The FMA-1900 will take time extending to up to several minutes to get back to normal functionality after the overrange condition is removed.
Q: Is FMA-1900 a loop powered device?
A: No, FMA-1900 do not fall in the category of loop powered devices.
Avoid connecting power on the input or output of the 4-20 mA section.
Q: What is the accuracy percentage of FMA-1900?
A: It could be seen that the accuracy percentage of FMA-1900: ±1.5% of full scale between 10 – 100% of full scale.
Q: What is the FMA-1900 installation procedure?
A: Ensure that the FMA-1900 instrument is clean and has the right plumbing fittings before it is fixed. Also ensure that the arrow on the transducer is in the correct position so that it indicates the flow direction.
User Manual
.bemxa~eahnk@m_
Mtg:/~.umega.com
e-mail:
info@omega.com
J
FMA-1900
Mass
Flow Meter
OMEGAneP
http://www.omega.com
On-Line Service
Internet e-mail
info@omega.com
Servicing North America:
USA:
IS0
9001
Canada:
Crrtifiod
Omega Drive,
Stamford, CT
(203)
Tel: e-mail: info@omega.com
976 Tel: (514) 856-6928
e-mail:
359-1660
Bergar
Lava1 (Quebec)
canada@omega.com
Box 4047
06907-0047
5Al
H7L
359-7700
(203)
FAX:
FAX: (514) 856-6886
One
For immediate technical or application assistance:
It is the policy of OMEGA to comply with all worldwide safety and apply. OMEGA is constantly pursuing certification of its products to the European New ApproachDirectives. OMEGA
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. acceptsno liability for any errors it contains, and reserves the right to alter specifications without notice.WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
Omega Engineering is not liable for any damages orpersonal injury, whatsoever, resulting from the use of Omega’s massflow meters or controllers with oxygen gas. Although the mass flowmeters and controllers are cleaned prior to shipment, we make noclaim or warranty that their cleanliness renders them safe for oxygenservice. The customer must clean Omega’s mass flow meters orcontrollers to the degree that they require for their oxygen flowapplications. This statement does not replace product warranty.
m
Over-tightening the pipe connection may crack the
fittings or shift the calibration of the FMA-1900 controller.
B
The maximum pressure and temperature in the flow line in
which your FMA-1900 is to be installed should not exceed 150 psig
(5O”C),
122°F
kg/cm*
gauge) or
B
The FMA-1900 is not a loop powered device! Do NOT
apply power to the 4-20
B
power supply (50 watts (250mA)
The FMA-1900 controller requires a 24 VDC regulated
mV
@
24 VDC
mA
output or input section.
RMS ripple) with the ability to provide at least 6
510%.
respectively.
If you are providing your ownpower source, refer to Section 2.3, paragraph 8, for specific powersupply requirements and jumper settings.
B
Do not use liquid leak detectors
search for leaks
to
inside or outside the FMA-1900. Instead, monitor pressure decay.
Omega’s FMA-1900 Series products can be calibrated to
measure and control the mass flow rate of gases in several
ranges from O-l 0 standard cubic centimeters per minute(SCCM) to +1.5%
range, and time response is 5 seconds to within 2% of setpoint.
The FMA-1900 is ideal for a complete range of gas flow
applications including general process control, laboratories,instrument OEM’s, gas panels, and flow calibration.
O-50
standard liters per minute (SLM). Accuracy is
of full scale over a wide temperature and pressure
Figure
I-
1 The FMA- 1900 Flow Controller
ZERO
SET POINT
The versatile FMA-1900 product digitally displays the massflow rate directly in engineering units or percent of full scale.
The FMA-1900 is a transducer requiring a regulated 24 VDCexternal power source. The O-5 VDC and 4-20 signals, which are linearly proportional to gas mass flow rate,are provided for recording, data-logging, or control. A“D”
connector is provided for power input, output signal and setpoint control. FMA-1900 is available in several basic configu-rations with either NPT (female) or compression inlet/outletfittings, and with or without the optional power supply.
output
mA
15-pin
l&Specifications
FLOW RATES: O-10 SCCM to O-50 SLM; flow ranges speci-
fied are for an equivalent flow of nitrogen at 760 mm Hg and
21
.l
“C (70°F). Other ranges in other units are available (e.g.,
SCFH or FMA-1900 SERIES MINIMUM DP REQUIREMENTS: 5-50
PSI optimum.
GASES: Most gases (e.g., air, nitrogen, methane, carbondioxide, argon, helium, hydrogen); check compatibility withwetted materials; specify when ordering.
OUTPUT SIGNALS: Linear O-5 VDC into 2000 Ohm minimumload resistance standard or linear 4-20maximum load resistance for 24 VDC supply (500 ohm/l5VDC supply).
CONTROL RANGE: Calibrated for 10 to 100% of full scale.INPUT POWER: 0 to 15
VDC
nm3/h).
kg/cm*)
flO%
differential standard; 30 PSI (2(0.35-3.5
mA
flO%;
into 1000 ohm
0 to 50 SLM/15 VDC
kg/cm*)
SLM/24
O-100%
fl.5%
of full scale from 1 ACCURACY
scale.
of
REPEATABILIWTEMPERATURE
PRESSURE COEFFICIENT: 0.02% of full scale per PSI (0.07
optional) into 22000 Ohms. O-20Ohms input impedance.
or better:
valve.)
MATERIALS:
&0.25%
COEFFICIENTI
kg/cm*)
COMMAND SIGNALS: O-5 VDC (O-10 VDC
full
0.08% of full scale per
kg/cm*)
gauge optimum.
lo-‘ATM
(Not recommended for use as a positive
10% glass-filled
cc/set
32” to 122°F (0 to 50°C)
mA
of full scale.
“C,
or better.
gauge maximum;
of helium maximum
66; 316 stainless steel,
Nyton
“C”-rings.
mA
and 4-20
into 250
-2-
INSTALLATION
2.1
Receipt of Your FMA-1900
After receiving your FMA-1900, carefully check the outside ofthe packing carton for damage incurred in shipment. If thepacking carton is damaged, notify the carrier at once regardingtheir liability. A report with the serial number and part numbershould be submitted to Customer Service. Call 800-622-2378for detailed instructions.
Remove the packing slip from its envelope and check that thecarton contains all parts listed. Make sure spare parts oraccessories are not discarded with the packing material. Incase of shortages, contact 800-872-9436Have your purchase order and model number available.
2.2
Return Shipment
Do not return any equipment without a Return Authorization,obtainable from the Customer Service Department at thenumber shown above.
Information describing the problem, the corrective action or work tobe accomplished at the factory, the purchase order number underwhich the work is to be done, and the name of person to contactmust be included with the returned equipment.
NOTE:Equipment returned for repair that is found to be com-pletely operational will be subject to the current “no problemfound” billing rate.identify possible problems with the installation or application.
In a case such as this, Omega will attempt to
(800~USA-WHEN).
2.3
Before Beginning the Installation
Read the following notes in their entirety before beginning
actual installation of your FMA-1900 flow controller.
1.
Using the flow direction arrow on the FMA-1900 to prop-erly orient the controller, install the FMA-1900 into the gasflow line. If you are utilizing quality paste pipe thread sealant. First tighten the fittings
by hand, then tighten no more than one and a half turns toavoid cracked fittings or creating a calibration shift.
The line pressure and temperature should not exceed 150psig (10 operating pressure differential is 50 psig.
2.
Apply power to your FMA-1900. If you are using the
Omega power supply, connect it to thetor on the side of your FMA-1900, then plug the powersupply into line power. If you are providing your ownpower source, refer to Section 2.3, paragraph 8, forspecific power supply requirements and jumper settings.
3.Upon application of power, the output signal will be at a highlevel for the first 10 to 20 seconds, after which (assuming zeroflow) it will drop to 0 VDC (or ration). Allow at least 15 minutes for complete warmup.
CAUTION:
Do NOT apply power to the 4-20
4.
After the warm-up period, your FMA-1900 will begin
monitoring the gas mass flow rate.
5.
Output Signals: The FMA-1900 has either O-5 VDC (O-l 0VDC optional) or 4-20 control range of the unit is 10% to 100% of the calibratedflow range. The output is
mass flow rate. The full scale range and the gas for whichthe unit was calibrated are shown on the FMA-1900 datatag. Section the electrical output signal hookup. For example, if you aremonitoring the O-5 VDC output signal, 5.00 VDC is theoutput signal for the full scale listed on the FMA-1900 datatag; 2.50 VDC is for one-half of full scale; and 0.00 VDC isfor zero flow.
signal, 20.00
mA
kg/cm*
gauge) or 150°F (66°C). Maximum
The FMA-1900 is not a loop powered device!
2.6,
ELECTRICAL CONNECTIONS, describes
If you are monitoring the 4-20
mA
is the output signal for the full scale; 12.00
is for one-half of full scale; and 4.00
l/4-inch pipe, use a good
15-pin
4mA,
depending on output configu-
mA
output or input section.
mA
output signals. The effective
linearty
proportional to the gas
mA
is for zero flow.
connec-
“D”
mA
output
l/2
Integral Display: The 3
6
engineering units or optional percent of full scale. The fullscale range and gas are shown on the instrument data tag.The decimal point for the flow rate is set at the factory andwill show automatically (e.g., “5.54” SLM or “76.4”
Overrange conditions are indicated by the display and/or
7.
output going to a high level, above the full scale range.After the over-range condition has been removed, it maytake several minutes for the FMA-1900 to recover andresume normal operation.
The FMA-1900 has more stringent power supply require-
8.
ments due to the presence of the valve.is operated in a control loop, power supply variationscannot be tolerated. This means the power supply must bea regulated 24 VDC with ripple not to exceed 50mV peak topeak, and capable of producing at least 250mA (6 watts).
The standard power supply for both the FMA-1900 is 24VDC. It is possible to operate the FMA-1900 on 15 VDC atreduced performance levels. There is a direct relationshipbetween the amount of power the valve requires and theflow rate. Due to this relationship, 15 VDC poweredcontrollers are limited to a flow rate of 15 SLM. To achieveflow rates above
The minimum power requirements for all FMA-1900controllers are as follows:
15 VDC installed) [If less than 15 SLM].
24 VDC removed)
A regulated power supply is required. Ripple contentshould not exceed 50mV peak to peak.
Refer to the component location diagram (Figure 3-2) forthe location of Jumper J3.
160mA,
@
250mA,
@
[
15 to 250 SLM].
digit LCD display reads directly in
%).
Because the valve
15
SLM, use a 24 VDC power supply.
Jumper J3 closed (shorting block
Jumper J3 open (shorting block
Loading...
+ 25 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.