Omega FMA-1900 User Manual

Specifications and Main Features

  • OPERATIONAL FLOW RANGE: From 0-10 SCCM to 0-50 SLM
  • FMA-1900 SERIES DE is found to operate within the differential standard of 5-50 PSI (0.35-3.5 kg/cm²) with the optimum performance being at a at optimum of 30 PSI (2 kg/cm²).
  • TYPES of GASES: A wide range of gases including air, nitrogen, methane, carbon dioxide, argon, helium and hydrogen are reproducible with this model.
  • OUTPUT CONNECTORS: The devices come with linear output of 0-5 volts direct current into 2000 ohms minimum load resistance standard or linear 4-20 mA into 1000 split load adjusted maximum for 24 VDC supply voltage
  • CONTROL RANGE: Calibrated between 10 to 100 % of total
  • INPUT VOLTAGE REQUIREMENTS: 24 VDC regulated power supply 50mV RMS ripple with capability of delivering power not less than 6 watts (this is at 250mA) at a voltage of 24 VDC +10 percent optionally
  • ACCURACY: +1.5 percentage of full scale from 10 to 100 percent of full scale.
  • REPEATABILITY: +0.25 percent of full scale
  • TEMPERATURE COEFFICIENT: Maximum of 0.08 percent of full scale per degree centigrade.
  • PRESSURE COEFFICIENT: Maximum of 0.02 percent of full scale per PSI (0.07 kg/cm²).
  • REACTION TIME: One second to 63% of final value
  • GAS PRESSURE: Maximum of 10 kg/cm² gauge of 150 PSI is quite sufficient
  • LEAK INTEGRITY: This model covers a wide range of performance defect tolerances of membrane gas leak, up to 1x 10 -6 .
  • GAS AND AMBIENT TEMPERATURE: Operational temperature range of 32° to 122°F (0 to 50°C).
  • WETTED MATERIALS: Viton O-ring, nickel plating, 316 stainless steel, 430F stainless steel, and 10% glass-filled Nylon 6/6

Frequently Asked Questions

Q: What is the Maximum limit of flow which can the FMA-1900 pass?

A: The FMA-1900 is manufactured to pass a maximum flow rate measuring 0-50 standard litres of liquid in one minute.

Q: Which output signals does FMA-1900 give?

A: The signals are linear 0-5 VDC and 4-20 mA output signals.

Q: What is the power supply that is needed to run the FMA-1900?

A: The power supply needed for the FMA-1900 is 24 VDC and is regulated with least 6 watts (250mA) @ 24 VDC +10%.

Q: With which gases can FMA-1900 be used?

A: Many gases can be used with FMA-1900 such as air, nitrogen, methane, carbon dioxide, argon, helium, and hydrogen.

Q: What to do in case FMA-1900 goes to overrange condition ?

A: An overrange condition, is when the display or output reaches the maximum limit of its full scale range. The FMA-1900 will take time extending to up to several minutes to get back to normal functionality after the overrange condition is removed.

Q: Is FMA-1900 a loop powered device?

A: No, FMA-1900 do not fall in the category of loop powered devices.

Avoid connecting power on the input or output of the 4-20 mA section.

Q: What is the accuracy percentage of FMA-1900?

A: It could be seen that the accuracy percentage of FMA-1900: ±1.5% of full scale between 10 – 100% of full scale.

Q: What is the FMA-1900 installation procedure?

A: Ensure that the FMA-1900 instrument is clean and has the right plumbing fittings before it is fixed. Also ensure that the arrow on the transducer is in the correct position so that it indicates the flow direction.

User Manual

.bemxa~eahnk@m_
Mtg:/~.umega.com
e-mail:
info@omega.com
J
FMA-1900
Mass
Flow Meter
OMEGAneP
http://www.omega.com
On-Line Service
Internet e-mail
info@omega.com
Servicing North America:
USA:
IS0
9001
Canada:
Crrtifiod
Omega Drive,
Stamford, CT
(203)
Tel: e-mail: info@omega.com
976 Tel: (514) 856-6928
e-mail:
359-1660
Bergar
Lava1 (Quebec)
canada@omega.com
Box 4047
06907-0047
5Al
H7L
359-7700
(203)
FAX :
FAX: (514) 856-6886
One
For i mmed iate technical or application assistance:
l-SOO-TC-OMEGASM
/
USA
and Canada:
Sales Service: l-800-826-6342 Customer Service: l-800-622-2378 Engineering Service: l-800-872-9436 TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
l-800-622-BE!?P
/
l-SOO-USA-WHENSM
/
Mexico and Latin America:
Tel:
Espaiiol:
En
(95)
800-TC-OMEGAsM
(203)
359-1660 ext: 2203
FAX: (95) 203-359-7807
espanol@omega.com
e- m ail:
Servicing Europe:
8034,118O
Postbus
Benelux:
Czech Republic:
France:
Germany/Austria:
United Kingdom:
Is0
9002 certified
It is the policy of OMEGA to comply with all worldwide safety and apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
will add
Tel:
(31)
20 6418405 Toll Free in Benelux: 06 0993344 e-mail:
nl@omega.com Ostravska 767,733 01 Karvina
(69)
Tel: 42 e-mail: czech@omega.com
Denis
9, rue
Tel: 33 0130-621-400
Toll
Free in France;
france@omega.com
e-mail: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: 49
(07056) Toll Free in Germany: 0130 112166 e-mail:
germany@omega.com
25 Swannington Road, Broughton Astley, Leicestershire, LE9
6TU, Tel: 44 (1455) 285520 FAX: 44
(1455)
the CE mark to every appropriate device upon certification.
LA Amstelveen, The Netherlands
6311899
Papin,
78190 Trappes
05-4-06342
3017
England
283912
Toll Free
in England: 0800-488-488
uk@omega.com
e-mail:
(31)
(07056)
7
Box
5EX,
(161)
20 6434643
England
8540
Omega Drive,
777-6611
regulations that
FAX:
FAX: 42 (69) 6311114
FAX: 33 0130-699-120
FAX: 49
P.O. Irlam, Manchester, M44 Tel: 44 FAX: 44 (161) 777-6622
B
Omega Engineering is not liable for any damages or personal injury, whatsoever, resulting from the use of Omegas mass flow meters or controllers with oxygen gas. Although the mass flow meters and controllers are cleaned prior to shipment, we make no claim or warranty that their cleanliness renders them safe for oxygen service. The customer must clean Omegas mass flow meters or controllers to the degree that they require for their oxygen flow applications. This statement does not replace product warranty.
m
Over-tightening the pipe connection may crack the
fittings or shift the calibration of the FMA-1900 controller.
B
The maximum pressure and temperature in the flow line in
which your FMA-1900 is to be installed should not exceed 150 psig
(5O”C),
122°F
kg/cm*
gauge) or
B
The FMA-1900 is not a loop powered device! Do NOT
apply power to the 4-20
B
power supply (50 watts (250mA)
The FMA-1900 controller requires a 24 VDC regulated
mV
@
24 VDC
mA
output or input section.
RMS ripple) with the ability to provide at least 6
510%.
respectively.
If you are providing your own power source, refer to Section 2.3, paragraph 8, for specific power supply requirements and jumper settings.
B
Do not use liquid leak detectors
search for leaks
to
inside or outside the FMA-1900. Instead, monitor pressure decay.
Operators Manual
M248310898
FMA-1900 SERIES FLOW CONTROLLERS
TABLE O F
.
1
Introduction
Description.. .....................................................................
1.1 Specifications
1.2
2. Installation Receipt of Your
2.1 Return
2.2 Before Beginning the Installation
2.3 Mechanical Installation
2.4 Plumbing
2.5 Electrical
2.6
3.
Operation
Referencing the Flow Rate to Other Temperature and
3.1 Pressure Accuracy
3.2 Overranging
3.3 The 4-20
3.4 Zero and
3.5 Mounting
3.6 Output Options
3.7 Setpoint Input Signal
3.8 Setpoint Configuration
3.9 Cold Sensor Lockout Circuit
3.10
3.11 Auto Shut-Off
............................................................................
..................................................................
..................................................................................
FMA-1900
Shipment..............................................................
Connections
Connections
...................................................................................
Conditions
.....................................
...................................................................
mA
Output Signal
Local
Setpoint
Position..........................................................
...............................................................
.................................................................
CONTENT S
1 1
2 4
...............................................
.....................................
....................................................
.....................................................
.....................................................
.......................................................
_.
..................................
...........................................
Adjustments
......................................................
...................................................
..........................................
............................
4 4 5 7 8
9
11
......
11 11 12 12 12 13 13 13 15 16 16
FMA-1900 SERIES FLOW CONTROLLERS
TABLE O F
CONTENTS CONTINUE D
3. Operation (continued)
3.12 On/Off Control
3.13Purging of Mass-Trak Products
3.13.1 Purging of Non-reactive
Appendix A Pin Connections
Appendix B Purge and Valve Off Connections
Appendix C K-factors and Gas Tables
...............................................................
.....................................
Gases
.....................................
..........................................................
.,............................. 19
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..*..........
16 17 17
18
20
-I-
INTRODUCTIO N
1.1
Description
Omegas FMA-1900 Series products can be calibrated to
measure and control the mass flow rate of gases in several
ranges from O-l 0 standard cubic centimeters per minute (SCCM) to +1.5%
range, and time response is 5 seconds to within 2% of set point.
The FMA-1900 is ideal for a complete range of gas flow
applications including general process control, laboratories, instrument OEMs, gas panels, and flow calibration.
O-50
standard liters per minute (SLM). Accuracy is
of full scale over a wide temperature and pressure
Figure
I-
1 The FMA- 1900 Flow Controller
ZERO
SET POINT
The versatile FMA-1900 product digitally displays the mass flow rate directly in engineering units or percent of full scale.
The FMA-1900 is a transducer requiring a regulated 24 VDC external power source. The O-5 VDC and 4-20 signals, which are linearly proportional to gas mass flow rate, are provided for recording, data-logging, or control. A D
connector is provided for power input, output signal and set point control. FMA-1900 is available in several basic configu- rations with either NPT (female) or compression inlet/outlet fittings, and with or without the optional power supply.
output
mA
15-pin
l&Specifications
FLOW RATES: O-10 SCCM to O-50 SLM; flow ranges speci-
fied are for an equivalent flow of nitrogen at 760 mm Hg and
21
.l
C (70°F). Other ranges in other units are available (e.g.,
SCFH or FMA-1900 SERIES MINIMUM DP REQUIREMENTS: 5-50
PSI optimum.
GASES: Most gases (e.g., air, nitrogen, methane, carbon dioxide, argon, helium, hydrogen); check compatibility with wetted materials; specify when ordering.
OUTPUT SIGNALS: Linear O-5 VDC into 2000 Ohm minimum load resistance standard or linear 4-20 maximum load resistance for 24 VDC supply (500 ohm/l5 VDC supply).
CONTROL RANGE: Calibrated for 10 to 100% of full scale. INPUT POWER: 0 to 15
VDC
nm3/h).
kg/cm*)
flO%
differential standard; 30 PSI (2 (0.35-3.5
mA
flO%;
into 1000 ohm
0 to 50 SLM/15 VDC
kg/cm*)
SLM/24
O-100%
fl.5%
of full scale from 1 ACCURACY
scale.
of
REPEATABILIW TEMPERATURE
PRESSURE COEFFICIENT: 0.02% of full scale per PSI (0.07
kg/cm*),
RESPONSE TIME: One second to 63% of final value.
GAS PRESSURE: 150 PSI (10 20 PSI (1.4
LEAK INTEGRITY: 1 x
to outside environment.
shutoff GASANDAMBIENTTEMPERATURE:
WEllED
430F stainless steel; nickel plating; Viton SETPOINT
optional) into 22000 Ohms. O-20 Ohms input impedance.
or better:
valve.)
MATERIALS:
&0.25%
COEFFICIENTI
kg/cm*)
COMMAND SIGNALS: O-5 VDC (O-10 VDC
full
0.08% of full scale per
kg/cm*)
gauge optimum.
lo-ATM
(Not recommended for use as a positive
10% glass-filled
cc/set
32to 122°F (0 to 50°C)
mA
of full scale.
C,
or better.
gauge maximum;
of helium maximum
66; 316 stainless steel,
Nyton
C-rings.
mA
and 4-20
into 250
-2-
INSTALLATIO N
2.1
Receipt of Your FMA-1900
After receiving your FMA-1900, carefully check the outside of the packing carton for damage incurred in shipment. If the packing carton is damaged, notify the carrier at once regarding their liability. A report with the serial number and part number should be submitted to Customer Service. Call 800-622-2378 for detailed instructions.
Remove the packing slip from its envelope and check that the carton contains all parts listed. Make sure spare parts or accessories are not discarded with the packing material. In case of shortages, contact 800-872-9436 Have your purchase order and model number available.
2.2
Return Shipment
Do not return any equipment without a Return Authorization, obtainable from the Customer Service Department at the number shown above.
Information describing the problem, the corrective action or work to be accomplished at the factory, the purchase order number under which the work is to be done, and the name of person to contact must be included with the returned equipment.
NOTE:Equipment returned for repair that is found to be com- pletely operational will be subject to the current no problem foundbilling rate. identify possible problems with the installation or application.
In a case such as this, Omega will attempt to
(800~USA-WHEN).
2.3
Before Beginning the Installation
Read the following notes in their entirety before beginning
actual installation of your FMA-1900 flow controller.
1.
Using the flow direction arrow on the FMA-1900 to prop- erly orient the controller, install the FMA-1900 into the gas flow line. If you are utilizing quality paste pipe thread sealant. First tighten the fittings
by hand, then tighten no more than one and a half turns to avoid cracked fittings or creating a calibration shift.
The line pressure and temperature should not exceed 150 psig (10 operating pressure differential is 50 psig.
2.
Apply power to your FMA-1900. If you are using the
Omega power supply, connect it to the tor on the side of your FMA-1900, then plug the power supply into line power. If you are providing your own power source, refer to Section 2.3, paragraph 8, for specific power supply requirements and jumper settings.
3.Upon application of power, the output signal will be at a high level for the first 10 to 20 seconds, after which (assuming zero flow) it will drop to 0 VDC (or ration). Allow at least 15 minutes for complete warmup.
CAUTION:
Do NOT apply power to the 4-20
4.
After the warm-up period, your FMA-1900 will begin
monitoring the gas mass flow rate.
5.
Output Signals: The FMA-1900 has either O-5 VDC (O-l 0 VDC optional) or 4-20 control range of the unit is 10% to 100% of the calibrated flow range. The output is
mass flow rate. The full scale range and the gas for which the unit was calibrated are shown on the FMA-1900 data tag. Section the electrical output signal hookup. For example, if you are monitoring the O-5 VDC output signal, 5.00 VDC is the output signal for the full scale listed on the FMA-1900 data tag; 2.50 VDC is for one-half of full scale; and 0.00 VDC is for zero flow.
signal, 20.00
mA
kg/cm*
gauge) or 150°F (66°C). Maximum
The FMA-1900 is not a loop powered device!
2.6,
ELECTRICAL CONNECTIONS, describes
If you are monitoring the 4-20
mA
is the output signal for the full scale; 12.00
is for one-half of full scale; and 4.00
l/4-inch pipe, use a good
15-pin
4mA,
depending on output configu-
mA
output or input section.
mA
output signals. The effective
linearty
proportional to the gas
mA
is for zero flow.
connec-
D
mA
output
l/2
Integral Display: The 3
6
engineering units or optional percent of full scale. The full scale range and gas are shown on the instrument data tag. The decimal point for the flow rate is set at the factory and will show automatically (e.g., 5.54SLM or 76.4
Overrange conditions are indicated by the display and/or
7.
output going to a high level, above the full scale range. After the over-range condition has been removed, it may take several minutes for the FMA-1900 to recover and resume normal operation.
The FMA-1900 has more stringent power supply require-
8.
ments due to the presence of the valve. is operated in a control loop, power supply variations cannot be tolerated. This means the power supply must be a regulated 24 VDC with ripple not to exceed 50mV peak to peak, and capable of producing at least 250mA (6 watts).
The standard power supply for both the FMA-1900 is 24 VDC. It is possible to operate the FMA-1900 on 15 VDC at reduced performance levels. There is a direct relationship between the amount of power the valve requires and the flow rate. Due to this relationship, 15 VDC powered controllers are limited to a flow rate of 15 SLM. To achieve flow rates above
The minimum power requirements for all FMA-1900 controllers are as follows:
15 VDC installed) [If less than 15 SLM].
24 VDC removed)
A regulated power supply is required. Ripple content should not exceed 50mV peak to peak.
Refer to the component location diagram (Figure 3-2) for the location of Jumper J3.
160mA,
@
250mA,
@
[
15 to 250 SLM].
digit LCD display reads directly in
%).
Because the valve
15
SLM, use a 24 VDC power supply.
Jumper J3 closed (shorting block
Jumper J3 open (shorting block
2.4
Mechanical Installation
CAUTION: The maximum pressure and temperature in the flow line in which your FMA-1900 is to be installed should not exceed
(5OC),
kg/cm*
150 psig (10
gauge) or 122°F
respectively.
In order to ensure a successful installation, inlet and outlet tubing or piping should be in a clean state prior to plumbing your FMA-1900 to the system. FMA-1900 is applicable to clean gas only because patticulates and other foreign matter may clog the sensor tube and laminar flow element over a period of time. If the gas contains particulate matter install a high-efficiency, 50 to 100 micron, in-line filter upstream of the FMA-1900.
Do not locate the FMA-1900 in areas subject to sudden temperature changes, moisture, or near equipment radiating significant amounts of heat. Allow adequate space for cable connectors and wiring. Be sure the arrow on the side of the transducer points in the direction of flow. You can obtain best results if you operate the FMA-1900 in the plane in which it was calibrated. If you mount the unit in a position other than its calibrated position, you may have mild to severe perfor- mance problems. (See Section 3.6.)
CAUTION: Do not use liquid leak detectors to search for leaks inside or outside the FMA-1900.Instead, monitor pressure decay
self-
B
#6,
Mount the FMA-1900 to a chassis with two
type
tapping screws.
CAUTION: These screws should extend into the flow body no
furtherthan
.15(4mm).
If screws extendfurtherthan
.15(4mm),
the flow body may be damaged. See Figure 2-1.
2- 1 Mounting the FMA- 7900 Series
Figure
t
2.75
BSCREWx.15DP2
6.%?UNCx.t3LG
PLBSELFTAF$INGTYPE
$5_ Plumbing Connections
Your FMA-1900 transducer is supplied with either female NPT
outlet
(standard) or compression inlet and should not be removed unless your FMA-1900 is being cleaned or calibrated for a new flow range. requires a good quality, paste, pipe thread sealant which should be used in the inlet and outlet fittings. lighten fitting only one and a half turns past hand tight.
CAUTION: Over-tightening the pipe connection may crack the fittings
or shiicalibratfon.
For installation of fittings, simply insert the tubing into the fitting. Make sure that the tubing rests firmly on the shoulder of the fitting and that the nut is finger-tight. While holding the fitting body steady with a back-up wrench, tighten the nut one and a quarter turns, watching the scribe mark make one complete revolution and continue to the nine oclock position. After this, the fitting can be reconnected by snugging with a wrench. Use a back-up wrench to avoid
damaging the inlet fitting.
Finally, check the system s entire flow path thoroughly for leaks before proceeding to Section 3, OPERATION.
CAUTION: All instruments are leak-tested prior to shipping. To
check your installation, test the fittings only.
detectors to search for leaks inside or outside the FMA-1900. Instead, monitor pressure decay.
IMPORTANT Install a section of straight pipe at least five pipe diameters in
length upstream of the transducer. DO NOT use reducers. If
the gas contains any particulate matter, an in-line filter is
recommended. There can be no restrictions (such as valves,
tubing or piping internal diameters, reducers, etc.) upstream or
downstream of the MFC less than the valve orifice diameter. Failure to comply with this requirement will result in severely impaired performance and possible oscillations in flow control- lers. Refer to Table 2-1 for minimum restriction diameters upstream or downstream of the flow controller.
l/4-inch
(outside diameter) compression
Scribe the nut at the six o clock position.
fittings. These fittings
l/4-inch pipe
Do not use liquid leak
TABLE 2-1 TYPICAL MINIMUM RESTRICTION DIAMETERS
(Under Standard
AP
Conditions of 39
Flow Ranges Relative to N2
O-10 to O-500 SCCM O-500 to O-l 000 SCCM O-2 to O-5 SLM O-10 SLM O-l 5 SLM O-30 to O-50
SLM
PSIG
Inlet and Ambient Outlet)
Valve Orifice Diameter (Typical-in inches)
0.02
0.03
0.05
0.05
0.065
0.083
2.6 Electrical Connections
CAUTION: The FMA-1900 is not a loop powered device! Do
mA
NOT apply power to the 4-20 FMA-1900 is provided with a
the side of D
connector are shown in Figure 2-2, and the pin assign-
theFMA-1900 enclosure. The pin numbers of this
ments are given in Table 2-2. Operating power input and output signals are supplied via the
output or input section.
D
15-pin
connector located on
D
connector.
D
Figure 2-2
Connector Location and Pin Number Assignments
TABLE 2-2 “D” CONNECTOR PIN ASSIGNMENTS
PIN #
FUNCTION SignalCommon
O-5 VDC Flow Signal* valve
Return
PowerCommon No Connection +24
VDC (15 VDC) Supply
Setpoint
Input
+15
NOTE:
VDC operation uses same supply pins.
See page 6,
#8.
PIN # FUNCTION
9
4-20
Signal Common
10
+5
11
Valveoff
12
+24
13 14
4-20
15
ChassisGrand
(Common)
mA
VDC Reference
VDC)
VDC (15
mA
Output
Suppty
*O-l 0 VDC optional
OPERATIO N
3.1 Referencing the Flow Rate to Other
m
Temperature and Pressure Conditions
The gas flow rate output of your FMA-1900 is referenced to
C
standardconditions of 21.1 (one atmosphere), unless you have specified otherwise in your order. FMA-1900, because it may make a difference if you are comparing the output of FMA-1900 with another type of flow meter. For example, the output reading of FMA-1900 will be approximately 7% lower if you compare it to a device that uses a standardtemperature of
Be sure you know the reference conditions of your
(70°F) and 760 mm of mercury
0°C
rather than 21
.lC.
3.2
Accuracy
The accuracy of FMA-1900 is effective control range of the device is 10% to 100%. The
&1.5%
of full scale accuracy means the O-5 VDC output signal is accurate to within VDC) and the 4-20 This means, for example, that the output signal for zero flow can be as high as get an output signal at zero flow that is within either of these two ranges, your FMA-1900 is functioning properly. With respect to the FMA-1900 digital readout, the accuracy is simply
1.5% times the full scale flow rate listed on the instrument data tag. For example, if full scale is 10 SLM, the digital readout will be accurate to fo.15 SLM, and the reading at zero flow may be as high as specification.
+0.15
kO.075
mA
output is accurate to within
kO.075
SLM and still be within the stated accuracy
51.5%
of full scale, and the
*
VDC (O-l 0 VDC accuracy
VDC or
It0.24mA. Please note if you
0.150
f0.24mA.
Overranging
3.3
If the flow rate exceeds the full scale range listed on the 1900 data tag, the output signal and digital display will read a higher value. The FMA-1900 has not been calibrated for overranged flows and will be both non-linear and inaccurate if an overrange condition exists. The O-5 VDC (O-l 0 optional) and 4-20 or more. On the digital display, the display cannot exceed the four digits 1999. If the flow rate exceeds 1999, the right-most digits will blank and only the left-hand display.
Overrange going to a condition has been removed, it may take several minutes for the FMA-1900 to recover and resume normal operation. An overrange condition will not harm the instrument.
mA
outputs can exceed full scale by as much as 50%
1
will appear on the
by
conditions are indicated
hii
level, above the full scale range. After the overrange
the display
andlor
FMA-
output
3.4
The 4-20
The 4-20 pin on the mum) to ground (see Section 2.6, Electrical Connections). Figure 3-l illustrates an installation with a current loop output.
Figure
mA Output Signal
output
mA
mAoutput
D
1
Hookup Wiring for a Single 4-20
3-
signal current flows from the 4-20
connector through the load (1000 Ohms maxi-
mA
Unit
Zero and Local
The zero and local ports marked on the front of your FMA-1900. If the zero output is more than ometer when you are absolutely certain that the system has reached its normal operating temperature and there is zero flow at the desired pressure and orientation.
Setpoint Adjustments
setpoint
potentiometers are accessed through
fl.5%
of scale, you may adjust the zero potenti-
The local using an external setpoint. Turning the potentiometer clock-
wise increases the setpoint.
setpoint potentiometer is used when you are not
Mounting Position
Unless specified otherwise, your FMA-1900 has been cali- brated for installation with the flow section in a horizontal plane (k15)
with the enclosure standing up. If your actual installation
position is different, you will have to make a zero adjustment. NOTE: The zero value may shift and be more pronounced
when under pressure and when mounted in different positions.
Output Options,, Meters and Controllers
The standard output for all FMA-1900 controllers is either a O-5
mA
VDC (O-l 0 VDC optional) or 4-20 linearly to the VDC (and O-10 VDC optional) output requires a minimum load resistance of 2000 Ohms, while the 4-20 dates a load resistance of up to 1000 Ohms at 24 VDC and 500 Ohms at 15 VDC.
Setpoint Input Signal
3.8
The
setpoint input signal is a direct linear representation of O-100% causes a condition of 0% flow to occur and a 5.00 VDC causes a flow condition equivalent to 100% of flow to occur.)
When the command (setpoint) signal is applied, the flow controller will respond to changes in the second to 63% of final value, and within five seconds to full scale.
setpoint
O-100%
of the mass flow full scale value. (A 0 VDC
Either O-5 VDC (O-10 VDC optional) or 4-20
commands are available.
mass flow full scale range. The O-5
signal that corresponds
mA
output accommo-
setpoint
setpoint
setpoint within one
f2%
of
mA
Figure 3-2 Component Location and
r-7
L-j
r--
I
I
___I
i--i
L
__I
___.
D/P
Switch Set Up
Setpoint
m
The DIP switches on the FMA-1900 flow controller are used to
configure The default configuration is for an internal O-5 VDC
on-board
Configuration3.9
-
setpoint
operation and the Positive Shut-Off feature.
setpoint
using the
setpoint
command potentiometer (accessible through a hole
in the case) and the positive shut-off option enabled. Other options include internally or externally sourced O-5 VDC
setpoint
with positive shut-off enabled or disabled and externally
setpoint
sourced 4-20
mA
with positive shut-off enabled or disabled.
A factory installed option of O-1 0 VDC in and out is also available.*
1
Move DIP switch
setpoint
potentiometer. Move this switch to the left (OFF) to select an
to the right (ON) to select the internal on-board
external setpoint, which you must supply. This is also the position
required if you choose to use a local Move DIP switches 2 and 3 to the right to select 4-20
setpoint
potentiometer.
setpoint
mA
input. Move these switches to the left to select O-5 VDC setpoint. If
setpoint
your FMA-1900 was originally set up for O-5 VDC
setpoint
decide to switch to a 4-20 VR12
to ensure that the
mA
setpoint
signal you may need to adjust
and output match at full scale.
and you
Move DIP switch 5 to the right to have the controller valve forced shut
setpoint
(auto shut-off) whenever the
is less than 2% of full scale.
Move this switch to the left to disable auto shut-off. A 5.1 VDC reference is provided for internal and external
setpoint
command pots. The reference provides approximately 0.125 volts
headroom to allow for external cabling and ensures the ability to always reach full scale when using these inputs. The intended to be provided externally by the user.
0-1OV
to be used with a
output, DIP switch 1 is moved to the right-hand
setpoint
O-1OV If the internal
input is
setpoint
is
position and DIP switch 4 must be in the left-hand (open) position. This allows the controller to operate from the internal (or external local) setpoint
command pot even though the output is
setpoint
NOTE: If the
input is not connected to some type of com-
0-1OV.
0-W
mand control device, the valve-off function must be activated or DIP
switch
#l must be in the internal sourceposition. If no
setpoint
command is present on a controller when powered-up and the valve
is not switched off, the valve will drift wide open.
setpoint,
The internal
S1-4
switch
I
opened. External set must be disconnected.
when used with the O-10 V in/out setup, requires DIP
3.10 Cold Sensor Lockout Circuit
FMA-1900 controllers incorporate a safety circuit that closes
the valve when a fault condition is detected that could result in
uncontrolled flow (valve wide open). The circuit operates by
monitoring the temperature of the sensor elements and forcing the output high if the temperature falls below a preset limit. There are several conditions under which this could occur:
a)
Operation at a temperature below that for which the instrument is rated.
b) Power failure while running at or near full scale. Upon
resumption of power, the valve will remain shut until minimum
operating temperature is again reached. c) Sensorfailure. The operation of this circuit may be checked by observing the
output upon power-up. first 10 to 20 seconds. After that, assuming zero flow, the output will drop to 0 VDC or observing.)
(The output signal should be high for the
you
4mA,
depending on which
output
are
3.11
Auto Shut-Off
All FMA-1900 controllers are provided with an Auto Shut-Off
feature that, when enabled, closes the valve at a command
signal level of 2% or less of full scale. To enable this feature on your FMA-1900, move DIP switch 5
to the right. The valve will then be forced shut whenever the setpoint is less than 2% of full scale. To disable the feature, move DIP switch 5 to the left. The valve will then remain open even when the The default is enabled.
setpoint falls below 2% of full scale.
On/Off Control 3.12
For all FMA-1900 controllers, on/off control is provided via a mechanical or open collector switch. You can implement this option manually by connecting an on/off switch between pins 3 and 12 of the operation resumes when pin 12 is left floating.
D
15-pin
connector (See Appendix B). Normal
3.13
Purging of FMA-1900 Products
The purge function opens the controller valve completely for the purpose of purging the meter and for quickly flushing unwanted gas from the flow path. When the valve is opened for purging, it allows flows far in excess of the rated full scale of the controller. Because of the
uncontrolled nature of the purge function, two conditions
must be met before a controller can be purged.
1)
The Valve OFF switch, if it is used, must be in the ON or open state.
2)
The Auto-Shutoff function cannot be active (the setpoint command signal must be above 2% of full scale or DIP switch 5 must be in the OFF position).
The activation of either the Valve OFF switch or the Auto-Shutoff function will override the purge command.
D
To activate purge, connect pin 4 of the tor to common through either a mechanical switch or an open-collector transistor or logic output capable of sinking at least 4mA (See Appendix B). The maximum voltage appearing on this pin is 5.0 VDC.
15-pin
connec-
3.13.1 Purging Non-reactive Gases
Purge non-reactive gases from the FMA-1900 with clean,
dry nitrogen or argon for a minimum of two hours.
3.13.2 Purging Reactive Gases
-
Purge reactive gases from the FMA-1900 using one of the
following methods:
Cycle purge. This is done by alternately evacuating
1)
and purging the device for two to four hours with clean, dry nitrogen or argon.
Purge the device with clean, dry nitrogen or argon for
2)
eight to twenty-four hours. Evacuate the device for eight to twenty-four hours.
3)
Appendix A. Pin Connections
Appendix C. K-factors and Gas Tables
The following tables provide K-factors and thermodynamic
properties of gases commonly used with mass flow controllers
and meters. The purpose of these tables is two-fold:
1.
Calibrating an actualgas with a reference gas. This is
latly
useful if the actual gas is not a common gas or if it is a
called nastygas (i.e., toxic, flammable, corrosive, etc.).
Interpreting the reading of a flow controller which has been
2. calibrated with a gas other than the actual gas.
the
In applying
where:
=
The volumetric flow rate of the gas referenced to standard
Q
conditions of
=
The
K
actual gas, and
The K-factor is derived from the first law of thermodynamics applied to the sensor tube.
where:
=
The constant amount of heat applied to the sensor tube,
H
=
The mass flow rate of the gas
m
=
The coefficient of specific heat of the gas
C,
given in the Tables (at
=
The temperature difference between the downstream and
AT
upstream coils.
N
=
A correction factor for the molecular structure of the gas given by the following table:
tables, the following fundamental relationship is used:
Q,/Q,=K,/K,
0°C
and 760 mmHg(SCCM or SLM),
=
refers to the
),
(6),
K
factor defined in equation
refers to the
=
)*
(
mC
AT
P
H=
N
(gm/min),
OC).
(
?eference
(Cal/gm);
patticu-
so-
(1)
gas.
(2)
C,
is
Number of Atoms in the Gas Molecule
Monatomic
Diatomic
Triatomic
Polyatomic
N
1.040 1 .ooo
0.941
The mass flow rate, m, can also be written as:
pQ
ri-r=
The gas mass density at standard conditions
p
where: p
is given in the tables (at
temperature difference,
780 mm Hg). Furthermore, the
0°C
AT,
is proportional to the output
(3)
(gil);=
voltage, E, of the mass flow meter, or
AT=aE where: a If we combine Equations (3) and
(2)
and solve for
where: b =
A
contstant
=
(4)
insert them into Equation
Q,
we get:
PC,)
Q
H/aE
= A constant if the output voltage is constant.
=
(bN/
For our purposes, we want the ratio of the flow rate,
Q,,
actual gas to the flow rate of a reference gas,
to produce
(4)
(5)
Q,,
for an
the same output voltage in a particular mass flow controller. We get this by
comgining Equations (1) and (5):
Q,/Q,=K,/K,=@J,/P,
Q,,)
(6)
Please note that the constant b cancels out. Equation (6) is
the fundamental relationship used in the accompanying tables.
For convenience, the tables give relative k-factors, which are
the ratios
K,/K2,
instead of the K-factors themselves.
In the third column of the tables, the relative K-factor is
Kreference
Ka*d
equivalent to the actual gas. In the fourth column, the relative
K-factor is used gas, nitrogen& enabling you to calculate some instances,
where the reference gas is a gas molecularly
K,&dKN
,
where the reference gas is the commonly
).
The remaining columns give
K,/K2
directly using Equation (6). In
C, and
p,
K,/K, from the tables may be different from
that which you calculate directly. The value from the tables is
preferred because in many cases it was obtained by experiment. Omega calibrates every FMA-1900 Series mass flow controller
with primary standards using the actual gas or a molecularly equivalent reference gas. The calibration certificate accompa- nying your flow controller will cite the reference gas used. When a reference gas is used, the actual flow rate will be within 2 to 4% of the calculated flow rate.
EXAMPLE 1: A flow controller is calibrated for nitrogen
(N,) and the flow rate is 1000 SCCM for a 5.000 VDC output signal. The flow rate for carbon dioxide at a 5.000 VDC output is:
KN*,
or
/
Kco,
=
QN*
/
Qco,
=
/
Qco,
= (0.74
1 .OOO) 1000
EXAMPLE 2: A flow controller is calibrated for hydrogen
740 SCCM
(H,)
and the flow rate is 100 SCCM for a 5.000 VDC output signal.
The flow rate for nitrous oxide
KN*O
QH*
=
/
QN*O
/
QN*O
=
1 .Ol) 100 = 70.3 SCCM
(0.71
(N,O)
KH*,
/
found.as follows:
is
or
Please note that the K-factors relative to nitrogen must be used in each case.
EXAMPLE 3: We want a flow controller to be calibrated for use with dichlorosilane
flow. We wish to use the
(CF,).
What flow of
QSIH~C~
QCF
=
Equation (6) is used for gas mixtures, but we must calculate
(SiH
CL,) at a 100 SCCM full scale
preterred
CF,
must we generate to do the calibration?
QCF
=
/
QCF
/
100
reference gas Freon-14
KcF,,
/
= 0.869
KsiH2Ch
100/0.869=115SCCM
N/
and N for a
p,
pC, for the mixture. The equivalent values of
C,
dual gas mixture are given as follows: The equivalent gas density is:
=
Total mass flow rate
rh,
61,
=
a
where:
(
rir,
=
refers to gas
),
+
,
and (
#l
=
p
=
refers to gas
)*
The equivalent specific heat is:
where: F, The equivalent value of N is: N
=
(rf+
1%
P
) and
/
)
P,
=
The equivalency relationships for p, C more than two gases have a form
61~)
/
(6,
Cp,
=
F, C, (fi2
F,
cm,
simr
=
/
,
and N for mixtures of
*P
ar to the dual-gas
(I$/
p,
+
(grn/min),
#2
C,
F,
+
(fi,
P
/
P,)
(br2
p,
t?~,)
(7)
)
m,)N,
/
in,)N, +
relationship given above.
Achul
on
Ch.“dul
Syttbol
u-tat.
K-k
Ret.
Rd.Ws
to
z
Cp
Rel.
to
N,
(CwQ)
BensKy
(g/l)
80%
o-ring
EIeakUhef
valv e
seat
Acetylene Air
Aflene
(Propadiie)
Ammonia
Argon
Arsine
Trichlodda
Boron
Trtftuoride
Boron Bromine
Tribomide
Boron
Pentafluoride
Bromine Bromine
Trfftuoride Bronmtdfloromethane (Fecwt3Bt)
1.3~B”tadiana Butane
-Butane
1 *-Butane 2-Butane
Modde
Carbon
Carbon
Disulfide
Cakm
Monoxide
Tetrachlortde
Carbon Carbon
Tetralluoride
(Freon-t)
Fiuotde
Cerbonyi
Sulide
Carbonyi Chlorine ChtonneTrifluodde ChforodiffuoronwJ-rane (Freon-22) Chloroform
(Freon-115) Chlorotriflurometha (Fecn-13) Cwogen Cyanogen
Chloride
cychlopropane
CJ,
C,“,
NH,
A
AsH,
BC, BF,
B, B,
BF, BF,
CBF,
C,H,
C.Hw
C,“,
C,H,
TRANS
C.H.
..
cq
Cs, c o
CCI,
CF.
COF,
co s
Cl,
CIF,
CHCIF,
CHCI,
C,CIF,Chkxopentafluoroethe
CCIF,
C,N,
CICN
C.H.
Cl.6
N,
N, N N,
A N, N,
N, N,
N, N,
N,
N, N N,
N N.
4
N; N,
N N,
N, NZ N, N, N,
N,
N,
N, N*
N,
se
1.00 .43
t.ooO 1.45
.67 .41 51N, .61 36
36 .37
.32
.26 .30
324
.74 56
.3t .42
54 66 .66 46 .46
.39 .24N*
3s
.61 .61
.4036
.246 ,352 ,492.73
.1244
.1167 .1279 .1776 .0539 .o647 .1369.26 .1161 .tt13
,351 .4007 364S
.336
,374,291 .x)16 .t426 .24661.00 .,655 .1664
.1710 .1651
,114 .(65l .1544
.1339
,164
,153
.2613 .t739 .3177.46
1.162
1.293
1.767 ,766
1.762
3.476
5.227
3.025
7.130
11.16
7.633
6.166
6.644
2.413
2.593
2.503
2.503
2.503
1.964
3.397 t.253
6.660
3.926
2.945
2.666
3.163
4.125
3.656
5.326
6.692
4.660
2.322
2.742
1.077 KR
NE0
NE0 NE0 NE0
KR
NE0
KR KRKR KR
KFt KR KR
KR KR KR
KR KR
KR KR KR
KR KR
KR
KR
Acluol
GOO
chornlcel
Symbol Gas
Ref .
K-fttG.
Ret. to
K&c.
N,Ret.Gas
to
Cp
Detac
(Cat/g)Rel.
Elastomer
(g6) O-ring
80%
Vatw
Seat
Deulerium Diborane Dibromodifluoromethane Dibromethane Dichiorodifluommethane (Freon-12) Dichlorolluoromethane (Freon-21) Dichlommethylsilane Dichlo osilane Dichlorotetralluoroethane (Freon-1, )
-Difluorcelhytene
.l 1
132A)
(Freon-t Dimethylamine Dimeyi
Ether
2.2-Dimethylp opana ElhaIW Ethanol EthylAcetylene Ethyl Chloride Ethylene Ethylene Oxide Fluorine Fluorolorm
(Freon-23)
Freon-l
1 Freon-l 2 Freon-13 Freon-t 3 Freon-14 Freon-21 Freon-22 Freon-t 13 Freon-114 Freon-115
8Freon-C31 Germane Germanium
Tetrachloride
D,
49
CBr F
CCI F,
CHCI,F
(CH,),SiCI,
sIH&I, t&t&F,
C,H,Fz
(CHJ NH
(CHJ O
C,H,
C&O
CP,
C,H,CI
C,“,
C,H,O)
F,
CHF,
CCI,F
CCl,F,
CCIF,
CF F,Bt
CF. CHCI F CHCIF.
CCI,FCCIF,
CFt,F,
C,CIF,
GaCL,
100
%
.&I
Y
.I9
Nz
.47
N,
.35
N
.;2
N*
.25
N,
.40
NZ
.22
N,
.43
N,
.37N, .39
N,
.22N, xl
N*CsH,
.39
N,
.32
NZ NZ
.39 .60N, .52N*
,980
%
66
N*
33
N,
.35
N*
36
N,
.37
N,
.42
N,
.42
N2
.46
N_
.20
N,
.22N* .24
N,
.17
N,C,F,
.57
N,GeH,
.27
N*
.1722
so6
.I5
.075
.I432
,140
.tS62
.t50
.1604
,224
,366 3414 ,391 .4097 .3395 .3513
,244 .1365
,266 .1673
,176 .1357 .1432
,153 .ttt3 .I654
,140 .1544
.t61
,166
,164
,165 .14o4 .to71
1.799
7.235
9.362
7.76
5.395
4.952
5.756
4.506
7.626
2.657
2.011 KR
2.055
3.219
1.342
2.055
2.413
2.679
1.251
1.965 KR
1.695
3.127
6.129
5.395
4.660
6.644
3.926
4.952
3.666
6.360
7.626
6.692
6.397
3.416
9.565
KR
KR KR KR
KR
KR KFI KR
KFI
KR KR
KR KR KR
KR KR KR KR KR
KR
KR KR KR KR KR KR
KFi
Actual
G8o
ChOltliCSf
svmbol
K-fee. C p
K-M.
Ref.
to
ReL
R&Gas
N,
to
(CalIo)
DSllSity
(grl)
00%
Elutottle
O-ring
Valve
Seat
Helium HexaRuoroalhane
16)
(Freon-1 Hexane Hydrogen Hydrogen Bromide Hydrogen Chloride Hydrogen Cyanide Hydnqen
fluoride
Hydrogen
Iodide HI
Selenide
Hydrogen Hydrogen Sulfide
Pentafluorida
Iodine lsobutane Isobutytene Krypton Methane Methand Methyl
Aoetytene Methyl Bromide Methyl Chloride Methyl Fluoride
Mercaptan
Methyl
Trichlomeilane
Methyl
Odde
Trifluortde
Chloride
Oxide
Hexaffuortde
Molybderwm Monoethykmine Monomethylamine Neon NE Nitric Nitrcgen Nitrogen Dioxide Nitrogen Nilrosyi Nitrous
(CH,)=f,
He
CF,
C,H,,
“*
HBr
HCI
HCN
HF
YSe
YS
IF,
CH(C”,),
C,H,
K
CH,
CH,OH
C,H. CH,B CH,CI
CH,F
CH,SH
MoF,
CH,NH,
CYN”,
NO
N,
NO,
NF,
NOCI
NO
He
1.636
N,
.16
N,
t.006 N, 1.066 N,
1.006 ,070t
N,
t.ooO
N,
l.ooo
N,
.79
N,
60
N, N*
N,
.29
N,
1.062
A
.72
N
56
N*
.43
N,
56
N,
63
N,
66
N,
.52
N, N,
.21
N N,
.5tN,
A
996
N,
1.006
N,
.74
N,
46
N, N,
.71
N
1.464 1.241 .1834 6.157
.3966
1.0, 3.419H, .0661 .1912 1.627
.317f 3479 .0545 5.707 .1025 3.613 .2397 1.520
.1106.25
3672
.0593 3.739
1.453 5328
3274 .3547 1.767 KR
.llx .1926 2.253 .3221 1.516 .2459 2.146
,164
.1373
.4343 1.366
,2451.006 1.46 .2326 1.339 .2465 1.25 .1933 2.052 .I797 .1632 .2066 1.964
.17&T
3.645 6699
3.610
1.206 603
wu
3.593
2.503,370,
,715
1.429
4.236
6.669
9.366
2.0112.67.35
.9CQ
3.166
2.920.6t
KR KR
KFI
KR.24
KR
KR KR KR
KR KR KR KR KR.27 UR
KR KR KFI KR.25 KR KFI KR
KR KR
Gas
Actual
(Freon-Csr 6)
Oxygen
Difluodde Oxygen Ozone Pentaborane Pentane Perchloryi
Fluoride Pefl”o opopane Phosgene Phosphine Phosphorous Phosphorous
Propane Propylene
Silica Sulfur Dioxide Sulfur Hexafluoride Sulfuryi TeoS Tetrafluorahydrazine
(Freon-ll) Trichlorisilane 1,1.2-Ttichloro-lv2.2 Ttifluorethane Trisobutyl Titanium Trichloro Trimethylamine Tungsten Uranium Hexafluoride Vinyl Bromide Vinyl Chloride Xenon
Oxychloride
Pentalluoride
TrichlorideFhosphomus
Tetrachlofids
Telrafluoride
Fluoride
(Freon-l
Aluminum
Tetrachloride
Ethylene
Hexasfuoride
Chemical
Symbol Gas
13) (C,H#
CH,CHBr CH&HCl
C,F,Octafluorqclobutane
OF,
0, 0,
B,H,
4Hz
CIOaF
C,F,
COCI,
PH,
POCI,
PH,
PCI, C,H, C,“,
SiHSilane SiCLSilicon
SiF
so: SF,
SO,F,
NSF,
CCI,FTtichlo ofluormelhans
SiHCI,
TZI,
(CH,),N
WF,
“Fe
X0
Raf.
N,
N, Y
N,
K-faC.
Rel. to
Ref.
Y
.63N,
Y
1.000 ,446N,
.26
N1
21
N, N,
,174
.44N,
1.070 .36
.36
NZ
41
NZ N* N,
N,
,090
NZ
.32
N,
.33
N*
.20
N,CCl,FCClF,
N, N,
.27
N,C,HCI, N,
.25
N1 N,
.46
N,
.46
N, A
Gas
Ral.
Cp
to
N,
DenSityK-fsc.
(Calfg)
,165
.1917 2.406 .2193 1.427
.3
,396
,151
,197 6.366
,237 ,132 .1810 .1250 3665
3169 .1270 7.560 .1691 .1466 .(592 6.516 .1543
.,a2
.1357
.1360
,161
,506 .120 6.465
,163 5.95.32 .3710 0610 .0668 .I241 4.772
,120s
,037s,993 1.44
Efaatornu
O-ring Valv e
A(&
6.397
2.144
2.616
3.219
4.571 KR.39
4.416
1.517
6.643
5.620
6.127
1.967 KR .a77 KR
1
.3541
1.433 KR.60
4.643
2.656
4.562
KR KR
4.64
6.129
6.043
6.360
6.648
2.639
KR
13.26
15.70 KR.20
2.766
5.856
Seat
KR.17
KR.36 KR
KR KR.1394 KR KR KR.30N, KR.30
KR.26 KR.35N, KR.69N, KR.26 KR.39N,
KFI KR.33N,
KR KR
KR.061 KR KFI KR.26
Teflon
KR
KR
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and
workmanship for a period of 13
additional one (1) month grace period to the normal
handling and shipping time. This ensures
coverage on each product. If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA s
Customer Service Department will issue an Authorized Return phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will
operation outside of design limits, improper repair, or unauthorized modification. This
WARRANTY Is VOID If the unit shows evidence of having been tampered with or shows evidence
being
of OMEGAs control. Components which wear are not warranted, including but not limited to
contact points, fuses, and O$Nlg
foranYdmnag9sthatresultftvmtheuseofRsproductsinaccwdsncewRhi&onna&n provided by OMEGA, either verbal or written. OMEGA warrants only that the manufactured WARRANTIES OR
WARRANTY OF HEREBY DISCLAIMED. herein based on contract, notexceedthepurchasepriceofthe shall OMEGA be liable for consequential, incidental or
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any
OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, or damage whatsoever arising out of the use of the Product(s) in such a manner.
damaged as a result of excessive corrosion; or current, heat, moisture or vibration;
improper specification; misapplication: misuse or other operating conditions outside of
is&--at4
to offer
assumes
H
will be as
by
REPRESENTADONS OF ANY KIND WHATSOEVER, EXPRESSED OR
are
exclusive and the total
warranty,
nuclear installation or activity, medical application, used on humans, or misused in any way, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability
months
triacs.
suBBestions
responsrbilii
spacified
TITLE,
UMlTAllON
negfigence,
from date of purchase. OMEGA Warranty adds an
one (1) year product warranty
that OMEGAs customers receive maximum
(AR)
number immediately upon
on the use of its various products. However,
for any omissions or errors ncr assumes
and free of defects. OMEGA
AND ALL IMPUED, EXCEPT THAT OF
OF
liability
lMPUED WARRANTIES INCLUDING ANY
FllNESS FOR A MERCHANTABBll-Y AND
UABIUTY:
The remedies of purchaser set forth
PARllCULAR PURPOSE ARE
of OMEGA with respect to thii order, whether
indemnification, strict liiilii or
MAKES
othewfise.
component upon which liability is based.
damages.
rq3ecial
to
cover
be repaired or replaced at no charge. OMEGAs WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing,
liibilitY
patis
OTHER
NO
shall
ln
no event
(1)
as
/
RETURN REQUESTS
INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN
AUTHORIZED RETURN (AR) NUMBER FROM OMEGA S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING
DELAYS). The assigned AR number should then be
marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to
prevent breakage in transit.
FOR
WARRANTY
the followina information available BEFORE
contacting
1. P.O. number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
OMEGA s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. (D
Coovriaht 1996 OMEGA ENGINEERING. INC. All riahts reserved. This document mav not be reproduced. translated, or reduced to any electronic medium or machine-readable prior written consent of OMEGA ENGINEERING, INC.
GMEGA:
RETURNS, please have
in technology and engineering.
FOR NON-WARRANTY REPAIRS, consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. P.O. number to cover the COST of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems relative to the product.
ohotocooied,
cooied.
form,
whole.or
in
in
hart,
without
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA...Of
Course!
TEMPERATURE
&
Ii??
Thermocouple, RTD
@
Wire: Thermocouple, RTD 0 Calibrators 0 Recorders, Controllers 0 Infrared Pyrometers
&
Ice Point References
PRESSURE, STRAIN AND FORCE
@
Transducers 0 Load Cells 0 Displacement Transducers 0 Instrumentation
&
&
Pressure Gauges
Thermistor Probes, Connectors,
&
Thermistor
&
Process Monitors
Strain Gauges
&
Accessories
.
FLOW/LEVEL
Flow
Computers
&
0
Rotameters, Gas Mass 0 Air Velocity Indicators 0
Turbine/I addlewheel
0 Totalizers
&
Batch Controllers
Flowmeters
Systems
pH/CONDUCTlVITY
&
pH
Electrodes, Testers
k?
Benchtop/Laboratory
0 0 Controllers, Calibrators, Simulators 0 Industrial
&
Conductivity Equipment
pH
Accessories
Meters
&
Pumps
DATA ACQUISITION
&
I@
Data Acquisition 0 Communications-Based Acquisition Systems 0 Plug-in Cards for Apple, IBM 0 Datalogging Systems @
Recorders, Printers
Engineering Software
&
Compatibles
& Plotters
Panels
&
Assemblies
HEATERS
0
Heating Cable 0 Cartridge &Ship Heaters
0 Immersion 0 Flexible Heaters 0 Laboratory Heaters
&
Band Heaters
ENVIRONMENTAL MONITORING AND CONTROL
&
0 Metering 0 Refractometers
0 Pumps 0 Air, Soil 0 Industrial Water
pH,
0
Control Instrumentation
&
Tubing
&
Water Monitors
Conductivity
&
Wastewater Treatment
&
Dissolved Oxygen
Instruments
M2483/0898
Loading...