Omega CNiTH-16D-AL, DPiTH-i16D User Manual

Step 17. Enter Alarm 1 Enable/Disable Submenu
Press dto display flashing
DSBL/ ENBL
.
If flashing
ENBL
is displayed, press a, if
DSBL
is displayed,
press buntil
ENBL
is displayed, then press dto store and
go to the next menu item.
Step 19. Select the Deviation Control Type Submenu
Press d. If flashing
_DEV
Deviation is displayed press a,
otherwise press buntil flashing
_DEV
is shown. Now
press dto store and go to next menu item.
Step 20. Select the Latched Type Submenu
Press d. If flashing
UNLT
Unlatched is displayed press a,
otherwise press buntil
UNLT
is displayed.
Press dto store and advance to next menu item.
Step 21. Select the Normally Open Type of Contact Closure Submenu
Press d. If flashing
N.o.
Normally Open is displayed,
press a, otherwise press buntil
N.o.
is displayed. Press
d
to store and advance to next menu item.
Step 22. Select the Above Type of Active Submenu
Press d. If flashing
ABoV
Above is displayed, press a,
otherwise press b until
ABoV
is displayed. Press d to store
and advance to next menu item.
Step 23. Enable Alarm 1 at Power On
(
A.P.oN
)
Press d. If flashing
ENBL
is displayed, press a, otherwise
press buntil
ENBL
is displayed. Press dto store and
advance to next menu item.
Step 24. Enter Alarm 1 High Submenu
Press atwice to skip
ALR.L
Alarm 1 Low value.
ALR.L
is
for below &
ALR.H
for above.
Step 25. Set the Alarm 1 High value
(
ALR.H
)
Press d. Press bor cuntil value to set the display to
002.0
. Press dto save.
Step 26. Enter the Alarm 2 Menu
The display will show
ALR2
the top menu for Alarm 2. Repeat steps from 17 to 25 to set for Alarm 2 the same conditions as for Alarm 1.
Step 29. Configuration of Display Color Selection
Press auntil the
COLR
Display Color Selection Menu
appears on the Display. Configure
COLR
as
N.CLR/GRN
(green),
1.CLR/ RED
(red),
2.CLR/AMBR
(amber). Please
refer to the operator’s manual if needed.
For color change on Setpoints refer to Owners Manual Section 2.
DPiTH-i16D Monitor
CNiTH-16D-AL Limit Alarm
Humidity + Temperature
OPERATION Step 1. Apply Power to the Instrument
When your device is first powered up it will display the ambient temperature (assume 75°F).
Step 2. Enter Setpoint 1 Menu
Press aone time from run mode to get to
SP1
Setpoint 1.
Step 3. Enter the Setpoint 1 Value Submenu
Press d. Display shows the previous selection of Setpoint 1.
Step 4. Change the Setpoint 1 Value
Press bor cuntil desired value is displayed.
Step 5. Store the Setpoint 1 Value
Set the Setpoint 1 to 10 degree higher than Process value (SP1 = 85) and press dto store, display flashes
STRD
message and advances to
SP2
Setpoint 2 Menu.
Step 6. Store the Setpoint 2 Value
Repeat steps 3 and 4. Set the Setpoint 2 to 5 degree higher than Process value (SP2 = 80) and press dto store, display flashes
STRD
message and advances to
CNFG
Configuration Menu.
Step 7. Enter the Reading Config Menu
Press dto enter
RDG
Reading Config Menu.
Step 8. Enter the submenu items of Rdg Config Menu
Press dto display
SNSR
Sensor submenu: Sensor selection
for Autotune, Loop, or Ramp and Soak
°F.°C
is for
temperature and
DRRh
is for Humidity
Step 9. Enter the submenu items of Rdg Config Menu
Press dto display Temp Unit submenu:
Step 10 Scroll thru selection for Temp Unit submenu
Press bto Scroll though the available selections of the Temperature Unit of your choice: °For °C.
Step 11. Store the Temperature Unit
Press d, display momentarily shows
STRD
the Unit has been stored and the instrument will go automatically to the next menu item.
Step 12. Enter the Filter Constant Submenu
Display shows
FLTR
Filter Constant Submenu.
Step 13. Display the Filter Constant Value Submenu
Press dto display the flashing, previously selected Filter Constant.
Step 14. Scroll through available Filter Constants
Press bto sequence thru Filter Constants
0001, 0002
,
0004, 0008, 0016, 0032, 0064
and
0128
.
Step 15. Store the Filter Constant
Press dmomentarily to store
0004
Filter Constant and the
instrument will automatically go to the next menu item.
Step 16. Enter Alarm 1 Menu
The display will show
ALR1
the top menu for Alarm 1. In the following steps we are going to enable Alarm 1, Deviation, Unlatch, Normally Open, Active Above, Enable at power-on and +2°F High Alarm i.e. Process Value > Setpoint 1 Value +2°F will activate Alarm 1.
• If Analog Output Option is installed and enabled, the controller will skip Alarm 1 Menu item to Analog Output.
• Alarm must be DISABLED if Ramp is ENABLED.
• Alarm1 will only work for Humidity, not Temperature.
MQS4005/1105
SPECIFICATION
SENSOR SPECIFICATIONS Relative Humidity Accuracy/Range:
±2% for 10 to 90% RH ±3% for 0 to 10%RH and 90 to 100%RH
Non-linearity: ±3% Hysteresis: ±1% RH Response Time:
4 sec (63% slowly moving air)
Repeatability: ±0.1% Resolution: 0.03%, 12bit
Temperature Accuracy/Range*:
±1°C (±2°F) for -40 to 0°C and 80 to
123.8°C (-40 to 32°F and 176 to 254°F) ±0.5°C (±1°F) for 0 to 80°C (32 to 176°F)
*NOTE: extended temp range is for Probe only, the Controller’s operating temp is 0-50°C
Response Time: 5 sec (63% slowly moving air)
Repeatability: ±0.1°C Resolution: 0.01°C, 14 bit
METER SPECIFICATIONS Display:
4-digit, 9-segment LED,
• 10.2 mm (0.40") Red, green, and amber programmable colors for setpoint and temperature units.
Output 1
:
Relay 250 Vac @ 3 A Resistive Load, SSR, Pulse
Output 2†:
Relay 250 Vac @ 3 A Resistive Load, SSR, Pulse
Only with -AL Limit Alarm option
Options: Communication
RS-232 / RS-485 or 10BaseT or
Excitation:
24 Vdc
@ 25 mA
Exc. not available for Low Power Option
Line Voltage/Power:
90 - 240 Vac ±10%,
50 - 400 Hz*,
or 110 - 375 Vdc, 5 W
* No CE compliance above 60 Hz
Low Voltage Power Option:
12 - 36 Vdc, 3 W**
** Units can be powered safely with 24 Vac
but No Certification for CE/UL are claimed.
Dimensions:
48H x 48W x 127D mm (1.89 x 1.89 x 5")
Weight:
159 g (0.35 lb)
Approvals:
UL, C-UL, CE per EN61010-1:2001
Alarm 1 is designed to monitor the humidity value around Setpoint 1 and Alarm 2 is designed to monitor the temperature value around Setpoint 2.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of one (1) year from the date of purchase. In addition to OMEGA’s standard warranty period, OMEGA Engineering will extend the warranty period for four (4) additional years if the warranty card enclosed with each instrument is returned to OMEGA.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGAand hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR W
ARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product
was PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
FOR NON-W
ARRANTY REPAIRS, consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the
repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2005 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
PATENT AND TRADEMARK NOTICE: This product is covered by one or more of the following patents: U.S. Pat. No. Des. 336,895; 5,274,577; 6,243,021 / CANADA 2052599; 2052600 / ITALY 1249456; 1250938 / GERMANY DE 41 34398 C2 / SPAIN 2039150; 2048066 / UK Patent No. GB2 249 837; GB2 248 954 / FRANCE BREVET NO. 91
12756. The “Meter Bezel Design” is a trademark of Newport Electronics, Inc. USED UNDER LICENSE. Other U.S. and International Patents pending or applied for.
This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning this device, as the guide contains important information relating to safety and EMC.
WARNING: These products are not designed for use in, and should not be used for, patient­connected applications.
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OEMGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
TRADEMARK NOTICE:
®
, omega.com
®
,
,
and ®are Trademarks of
OMEGA ENGINEERING, INC.
®
MADE IN
USA
®
®
OMEGAnet® On-Line Service
www.omega.com
Servicing North America:
USA: One Omega Drive, P.O. Box 4047
ISO 9001 Certified Stamford CT 06907-0047
TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA Customer Service: 1-800-622-2378 / 1-800-622-BEST Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
Mexico and TEL: (001)800-TC-OMEGA® FAX: (001) 203-359-7807 Latin American: En Espa
e-mail: espanol@omega.com
ñol: (001) 203-359-7803
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands TEL: +31 20 3472121 FAX: +31 20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: sales@omegaeng.nl
Czech Republic: Frystatska 184, 733 01 Karviná TEL: +420 59 6311899 FAX: +420 59 6311114 e-mail: info@omegashop.cz
France: 11, rue Jacques Cartier, 78280 Guyancourt TEL: +33 1 61 37 29 00 FAX: +33 1 30 57 54 27 Toll Free in France: 0800 466 342
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany TEL: +49 7056 9398-0 FAX: +49 7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
United Kingdom: One Omega Drive
ISO 9002 Certified River Bend Technology Centre
Northbank, Irlam Manchester M44 5BD United Kingdom TEL: +44 161 777 6611 FAX: +44 161 777 6622 Toll Free in England: 0800 488 488 e-mail: sales@omega.co.uk
e-mail: sales@omega.fr
Internet e-mail
info@omega.com
®
®
®
FUSE Connector Output Type For 115Vac For 230Vac DC
FUSE 1 Output 1 Relay 3 A(T) 3 A(T) ­FUSE 2 Output 2 Relay 3 A(T) 3 A(T) ­FUSE 3 Power N/A 100 mA(T) 100 mA(T) 100 mA(T) FUSE 4 Power N/A N/A N/A 400 mA(T)
1 Output 1/Setpoint 1/ Alarm 1 indicator
2 Output 2/Setpoint 2/ Alarm 2 indicator °C °C unit indicator for Temperature or Dewpoint °F °F unit indicator for Temperature or Dewpoint
%RH Display shows the Percent Relative Humidity
D Display shows the Dewpoint
a
Changes display to Configuration Mode and advances through menu items*
b
Used in Program Mode:
c
Used in Program Mode:
d
Accesses submenus in Configuration Mode
and stores
selected values*
CONFIGURATION
FLOW CHART
DESCRIPTION OF FRONT PANEL
The upper display may be RH, Temperature or Dewpoint readings depending on your Reading Configuration selections. Factory defaults are shown.
The Dual Display allows the user to observe the Relative Humidity or Dewpoint (upper display) and Temperature Value (lower display), at the same time.
Disassembly Instruction: If necessary, the unit may be removed from the panel and opened.
Warning: Disconnect all ac power from the unit before proceeding.
1. Make sure the AC power is disconnected.
2. Remove all wiring connections from the rear of the
meter. To remove power and input connectors squeeze top and bottom of the case near the connector site for release, then pull connectors from the case.
3. To remove meter from the case, squeeze top and bottom
of the bezel to release, then pull from case.
WIRING
Wire the instrument according to the figure shown below.
Warning: Do not connect ac power to your device until you have completed all input and output connections. This device must only be installed by a specially trained electrician with corresponding qualifications. Failure to follow all instructions and warnings may result in injury!
Connect the main power connections as shown in the figure below
This Quick Start Reference provides information on setting up your instrument for basic operation. The latest complete Communication and Operational Manual as well as free Software and ActiveX Controls are available at
www.omega.com/specs/iseries
or
on the CD-ROM enclosed with your shipment.
SAFETY CONSIDERATION
This device is marked with the international Caution symbol.
The instrument is a panel mount device protected in accordance with EN61010-1:2001. Remember that the unit has no power-on switch. Building installation should include a switch or circuit-breaker that must be compliant to IEC 947-1 and 947-3.
SAFETY:
• Do not exceed voltage rating on the label located on the top of the instrument housing.
• Always disconnect power before changing signal and power connections.
• Do not use this instrument on a work bench without its case for safety reasons.
• Do not operate this instrument in flammable or explosive atmospheres.
• Do not expose this instrument to rain or moisture.
EMC:
• Whenever EMC is an issue, always use shielded cables.
• Never run signal and power wires in the same conduit.
• Use signal wire connections with twisted-pair cables.
• Install Ferrite Bead(s) on signal wire close to the instrument if EMC problems persist.
MOUNTING Panel Mounting Instruction:
1. Using the dimensions from the panel cutout shown in
exploded views, cut an opening in the panel. 45mm +.61/-.00 square with R 1.5, 4 places (1.772”+.024/-.000 square with R 0.06”, 4 places) Panel thickness: 6.4mm (0.25”) max / 0.8mm (0.03”) min.
2. Insert the unit into the opening from the front of the panel,
so the gasket seals between the bezel and the front of the panel.
3. Slide the retainer over the rear of the case and tighten
against the backside of the mounting panel.
_____ Underline denotes factory default setup
START HERE
Button Function in Configuration Mode
• To enter the Menu, the user must first press
a
button.
• Use this button to advance/navigate to the next menu item. The user can navigate through all the top level menus by pressing a.
• While a parameter is being modified, press ato escape without saving the parameter.
• Press the up bbutton to scroll through “flashing” selections. When a numerical value is displayed press this key to increase value of a parameter that is currently being modified.
• Pressing the bbutton for approximately 3 seconds will speed up the rate at which the set point value increments.
• In the Run Mode, pressing the b button changes display from RH readings to Temperature readings.
• Press the down c button to go back to a previous Top Level Menu item.
• Press this button twice to reset the controller to the Run Mode.
• When a numerical value is flashing (except set point value) press cto scroll digits from left to right allowing the user to select the desired digit to modify.
• When a setpoint value is displayed press c to decrease value of a setpoint that is currently being modified. Pressing the c button for approximately 3 seconds will speed up the rate at which the setpoint value is decremented.
• In the Run Mode, pressing the c button changes from RH readings to Dewpoint readings.
• Press the enter d button to access the submenus from a Top Level Menu item.
• Press d to store a submenu selection or after entering a value - the display will flash a
STRD
message to confirm your selection.
• In the Run Mode, press d twice to enable Standby Mode with flashing
STBY
.
CONFIGURATION
The instrument has two different modes of operation. Run Mode: used to display Temperature and Relative Humidity. Menu Configuration Mode: used to navigate through the menu options and configure the controller.
a
MENU
b
(UP)
c
(DOWN)
d
ENTER
Output 1 and 2 are for -AL Limit Alarm Option only.
Rear View
Relative Humidity
Upper Display
Temperature
Lower Display
%RH
D
2
1
1
Refer to Operator’s Manual for important Input Probe Shield wiring notes
4 5 6
1/4 2/5 3/6
1/4 2/5 3/6
1/4 2/5 3/6
N/C
RED
WHITE
BLACK
GREEN
SHIELD
HUMd
YES/NO
tEMP
YES/NO
dEU
YES/NO
8
8
UNIt
YES/NO