OI\^EGA warrants this unit to be free of defects in materials and workmanship
and
to
give satisfactory service for a period of
If the unit sfiould malfunction, it must be returned to ttie factory for evaluation.
Our Customer Service Department will issue
immediately upon phone or written request. Upon examination by OMEGA, it
the unit is found to be defective it will be repaired or replaced at no charge.
However, this WARRANTY is VOID if the unit shows evidence of having been
tampered with or shows evidence of being damaged as a result of excessive
current, heat, moisture, vibration, or misuse, (Components which wear or which
are damaged by misuse are not
and triacs.
warranted.
one (1)
year
from
date of purchase.
an
Authorized Return (AR) number
These include contact
points,
fuses,
THERE ARE
OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT
LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND OF
FITNESS FOR A PARTICUUR PURPOSE. IN NO EVENT SHALL OMEGA
ENGINEERING, INC. BE LIABLE FOR CONSEQUENTIAL, INCIDENTAL OR
SPECIAL
AGREEMENT BY OMEGA ENGINEERING, INC. OR ANY BREACH OF ANY
WARRANTY BY OMEGA ENGINEERING, INC. SHALL NOT EXCEED THE
PURCHASE
FOR THE UNIT OR UNITS OR EOUIPMENT DIRECTLY AFFECTED BY SUCH
BREACH.
NO
WARRANTIES EXCEPT AS STATED
DAMAGES.
THE BUYER'S
PRICE PAID BY THE
SOLE
PURCHASER
REMEDY
TO OMEGA
HEREIN.
FOR
ANY
THERE ARE NO
BREACH OF THIS
ENGINEERING,
INC.
Return Requests/Inquiries
Direct all warranty and repair requests/inquiries to OMEGA Customer Service
Department, telephone number
contact the OMEGA Customer Service Department to obtain an authorized return (AR)
numtier. The designated AR number
package.
To avoid processing delays, also please be sure to Include:
1.
Returnee's name, address, and phone number.
2.
Model and Serial numbers.
3. Repair instructions.
(203)
359-1660.
should
then be marked
Before returning any instrument, please
on
the outside of
the
return
i^OMEGA
A. .^H
One Omega Drive, Box 4047
Stamford,
(203)359-1660 Telex: 996404 Cable: OMEGA
FAX: (203)359-7700
ENGiniEERING,
Connecticut 06907-(X)47
IIMC
OMEGA'-' is a registered trademark of OMEGA ENGINEERING, INC.
-, Copyright 1984 OMEGA ENGINEERING, INC. Alt rights reserved including
illustrations. Nothing in this manual may be reproduced in any manner, either wholly
or in part for any purpose whatsoever without written permission from
OMEGA ENGINEERING, INC.
Printed in
U.S.A.
M230/025
TABLE OF CONTENTS
GENERAL INFORMATION 1
Unpacking and Inspection 1
Optional Accessories 2
Preparation for Use 3
Safety Symbols
Environmental Conditions 4
Probe Matching 15
Troubleshooting ...: 15
Special Handling of Static Sensitive Devices 18
Component Layout 21
Schematic Diagram 23
and
Terms 4
GENERAL INFORMATION
The OI^EGA® IVIodels 873C and 873F are Hand Held Digital Thermometers. Both instruments have large easy to read liquid crystal displays
(LCD).
The input connection consists of a subminiature, type K thermocouple, female connector (SMP-K-F). This input accommodates any
OMEGA type K thermocouple probe fitted with an SMP-K-M connector. Cold junction electronic circuitry
compensates for ambient temperature changes. These instruments
also provide piece-wise linearization conformity for thermocouple nonlinearity. The instruments are powered by a 9V battery. Continuous
operation of 350 hours is typical with an alkaline battery or 200 hours
typical with a carbon-zinc battery.
The Model 873C measures temperature from - 55°C to 1000°C on
two ranges. The 200°C range measures temperature from -55.0°C to
199.9°C with a resolution of 0.1°C. The 1000°C range measures
temperature from -55°C to 1000°C with a resolution of 1°C. The accuracy is ± (0.25% rdg
of rdg -l-2°C) for -40°C to -55°C.
-i-
1.0°C)
in
the 873F/873C automatically
from -40°C to 1000°C and ± (0.25%
The Model 873F measures temperature from - 60°F to 200°F
ranges. The 200°F range measures temperature from -60.0°F to
199.9°F with a resolution of 0.1°F. The 2000°F range measures
temperature from -60°F to 1999°F with a resolution of 1°F. The accuracy of the Model 873°F is ±(0.25% of rdg -i-2.0°F).
UNPACKING AND INSPECTION
Remove the Packing List and verify that all equipment has been
received.
OMEGA Customer Service Department at (203) 359-1660.
Upon receipt of shipment, inspect the container and equipment for
any signs of damage. Take particular note of any evidence of rough
handling in transit. Immediately report any damage to the shipping
agent.
If there are any questions about the shipment, please call
NOTE
The carrier will not honor any claims unless all shipping
material is saved for their
removing contents, save packing material and carton in the
event reshipment is necessary.
examination.
After examining and
on
two
OPTIONAL ACCESSORIES
The following optional accessories can
and 873F to enhance the performance of the instrument. These
accessories are available:
Part Number
KS871
SC800
880A2
MN1604
SMP-K-M
SMP-K-F
OT-201-1
PREPARATION FOR USE
Description
Tilt Stand/Belt Clip/Probe Holder; can
be used as a tilt stand or clipped to
your belt. It will hold one compact
probe.
Soft Carrying Case
Deluxe Carrying Case
9V Replacement Battery
Subminiature Thermocouple Connec-
tor, Male
Subminiature Thermocouple Connec-
tor, Female
1 oz jar of High Thermal Conductivity
Paste for fast surface measurements.
be
used with the Models 873C
The Models 873F and 873C are supplied ready for use with battery
installed.
tor located on the top of the instrument (just above the display).
The symbol /!\ on the instrument denotes that the user should refer
to the opjerating instructions.
The WARNING used in this manual explains dangers that could result
in personal injury or death.
The CAUTION used in this manual explains hazards that could
damage the instrument.
Connect the appropriate probe to the miniature TC connec-
SAFETY SYMBOLS AND TERMS
ENVIRONMENTAL CONDITIONS
Operation of the Models 873C and 873F should take place at
bient temperature of -10°C to 50°C (14°F to 122°F), less than 80%
relative humidity up to 35°C, lineariy derate 3%
60°C.
R.H./C
an
am-
from 35°C to
OPERATION
The 873C and 873F Hand Held Digital Thermometers are easy to use
There is only one control (located on the side of the instrument) that
turns the instrument on/off and selects the range. (See Figures 1 and
2).
Use the following procedure to operate the instrument:
1.
Turn on the power to the instrument by selecting the desired range.
Model 873C:
A. 200°C range is the center position of the ON/OFF range switch.
B. 1000°C range is the upper position of the ON/OFF range switch.
Model 873F:
A. 200°F range is the center position of the ON/OFF range switch.
B. 2000°F range is the upper position of the ON/OFF range switch.
2.
Select the appropriate temperature probe and insert it into the input
jack.
3. Make the measurement:
A. Touch the probe tip to the material to be measured.
B. Read the display. The reading is in Celsius for the Model 873C.
The reading is in Fahrenheit for the Model 873F.
NOTE
An open sensor or overrange condition is indicated by:
CAUTION
Do not attempt to measure temperatures beyond
the range of the probe being used. Probe damage
may occur.
WARNING
A shock hazard exists on the Models 8730 and 873F
Input jacks when probes
or
sensors are exposed
to
voltage levels greater than 42V peak to earth ground.
SAFETY PRECAUTIONS
1.
Do not touch the probe tip when measuring excessively high or low
temperature.
2.
Do not exceed 42V peak from the probe tip or input jack to earth
ground.
Miniature TC Connector
OMEGA 873C THERMOMETER
Low Battery Indicator
LO
Minus Sign Displayed
Plus Sign Implied
•
BAT
-loao
3V4 Digit Display—
On/Off Range Switch
Figure 1. Model 873C Control Layout
200
TYPE K THEHMOCOUPLE
NiCr-N<AI
RANOE:
-IB'C
to
^ 42V MAX
4
lOOO'C
-^
Miniature TC Connector
Low Battery Indicator^
Minus Sign Displayed-
Plus Sign Implied
VA Digit Display-
On/Off Range Switch
OMEGA 873F THEHMOMETER
r DM I
-\QO.U
^=
TYPE K THEIIMOCOUPI.E
NiCr-NiAl
RANOE.
-tO'E
to
A 42V MAX
2000*F
+
^
Figure 2. Model 873F Control Layout
SERVICING INFORMATION
This section contains information needed to maintain the Models
873C and 873R The following information is included: functional
stallation/replacement, calibration, probe matching, troubleshooting,
parts list, schematic diagram and component layout.
WARNING
The information presented in this section is intended for use by qualified personnel only. Turn off
the power and remove all probes from the instru-
ment before replacing the battery.
FUNCTIONALITY CHECK
At room temperature hold the tip of a probe sensor between your
thumb and index finger. A reading between 25°C and 40°C (77°F and
104°F) should be observed to confirm functionality of the instrument.
PERFORMANCE VERIFICATION
Accuracy Check
1.
Prepare a distilled water ice bath or use an OMEGA* TRC III Ice
Point Reference Chamber.
2.
Connect a TC probe to the input jack of the instrument.
3. Immerse the TC probe into the TRC III or distilled water ice bath
and allow 10 minutes for thermal stabilization,
4.
Turn on the instrument and use Table 1 to verify that the readings
on each range are within specifications.
Table 1. Performance Verification
Range
200
1000
2000
873°C
0.0 ± 1°C
0 ± 1°C
873°F
32,0 ± 2°F
32 ± 2°F
ENVIRONMENTAL CONDITIONS
The Models 873C and 873F should be at an ambient temperature of
23°C ±1°C and a relative humidity of less than 80%.
DISASSEMBLY INSTRUCTIONS
Turn off the instrument and disconnect the
temperature probe before removing the bottom
cover.
When disassembling the instrument, handle the PC
board by the edges. Body
performance. Keep the PC board clean and free of
taminants.
1.
Place the unit face down on a bench or other similar surface and
remove the screws from the bottom cover.
2.
Separate the bottom cover from the rest of the instrument by
grasping the top of the case (just above the display) and carefully
lifting it away from the display.
3. Remove the battery.
4.
Remove the standoff (ST-166-1) that secures the PC board to the
case.
The standoff is located in the center of the PC board.
5. Remove the switch cap.
6. Lift the PC board out of the top cover (front panel)
7. Remove the shield.
8. The LCD is held in place on the PC board by a shock resistant
mounting and is assembled as shown in Figure 3.
9. To reassemble the unit, reverse the above procedure.
WARNING
NOTE
oil,
dirt, etc can degrade circuit
CAUTION
con-
When removing the
not to spread the four retaining hooks
bezel out too far. The plastic bezel hooks may fracture.
LCD
from the PC board be careful
on
the clear plastic
WARNING
When reassembling the instrument, be sure to
reinstall the switch cover. Common mode voltage
may be present on the switch creating a possible
hazard if the cover Is not installed.
fUa K 7/V RLUSTtn
a HEQUIRED)
BOTTOM COVER
Figure 3. Model 873 Exploded View
BATTERY INSTALLATION/REPLACEMENT
To install or replace the battery:
1.
Remove the bottom cover of the instrument (refer to the
disassembly instructions).
WARNING
Turn the instrument off and disconnect the input
probe before removing the bottom cover. Reinstall
the cover before resuming usa of the instrument.
2.
Remove the old battery.
3. Place the new battery in the battery compartment. Be sure to
observe the proper polarity. Refer to Figure 4.
4.
Reinstall the bottom cover before resuming use of the instrument.
BArrenv
TERMINALS
Rgure 4. Battery Installation
CALIBRATION
Required Test Equipment
1.
DC voltage calibrator wrth
2.
100: 1 divider- 1.98k and 20n wirewound resistors, 0.02% ratio
±0.01%
of setting ±20/iV accuracy
accuracy or better.
3. OMEGA TRC III Ice Point Reference Chamber or distilled water ice
bath.
4.
OMEGA type K TRP Reference Probe or make a thermocouple
Transition junction (see Calibration Set-Up).
Calibration Set-Up
1.
To make a thermocouple transition
of copper wire to each (type
K)
junction,
silver solder a piece
thermocouple wire and insulate the
junction with a waterproof jacket.
2.
Connect the thermocouple wires from the type K transition junc-
tion to an OMEGA Miniature Thermocouple Connector SMP-K-M.
3. Turn on the voltage calibrator.
4.
Plug in the miniature connector (see steps 1 and 2) into the 873
input.
5. Connect the copper wires to the 100:1 divider as shown in Figure
5. Connect the copper/chromel wire to the positive sida
6. Connect the 100:1 divider to the DC voltage calibrator as shown.
7. Use an OMEGA TRC III or prepare the distilled water ice bath as
follows:
A. Drill a hole in the cap of a Dewar flask or Thermos™ to
accommodate the thermocouple,
B. Firmly pack the flask with pea-size ice cubes made from
distilled water and then fill the flask with distilled water.
10
C. Replace the melted ice with more ice while removing the excess
water.
D. Place the cap on the flask. Immerse the thermocouples into the
flask and allow 20 minutes for temperature stabilization.
DC CALIBRATOR
100:1 VOLTAGE DIVIDER
*The
ICE BATH
20(1
resistor must be at a uniform temperature during calibration.
873
Figure 5. Set-Up Procedure
Calibration Procedure
1.
Remove the bottom cover as described in the disassembly instruc-
tions.
WARNING
To prevent a shock hazard, turn the instrument off
and remove the Input probe from the instrument
before removing the bottom cover.
2.
Turn on the Model 873 and perform the calibration adjustments
listed in Table 2 (873C) or Table 3 (873F). For access to calibration adjustments refer to Figure 4.
11
Table 2. Modal 8730 Calibration Adjustments
Step
1
2
3
4
Adjustment
Zero
-i-Gain (xl)
+ Gain(x10)
-Gain
Range
200°C
1000°C
200°C
200°C
Calibrator
Setting
0.00000
+ 41.269mV
-H7.139mV
-1.527mV
Trimmer
Adjust
R112
R115
R104
R105
Desired
Reading
00.0°C
1000°C
175.0°C
-40.0°C
Table 3. Model 873F Calibration Adjustment
Step
1
2
3
4
Adjustment
Zero
-i-Gain(xl)
+ Gain(x10)
-Gain
Range
200°F
2000°F
200°F
200°F
Calibrator
Setting
-0.692mV
-i-44.645mV
3.589mV
-1.929mV
Trimmer
Adjust
R112
R115
R104
R105
Desired
Reading
00.0°F
1990°F
190.0°F
-60.0°F
PROBE MATCHING
To improve the accuracy of the Model 873 while using a particular
proJTe, calibration can be adjusted so that probe errors in the region
of zero (0.0°C or 32°F) are compensated. Place the probe in an ice
bath and allow the reading to stabilize. Then reset the zero adjustment
(R112) to a
0.0°C
or 32°F reading.
TROUBLESHOOTING
The following troubleshooting information is intended to be used by
qualified electronic maintenance personnel who are familiar with the
proper use of standard electronic test equipment.
To gain access to the PC board, refer to the disassembly instructions.
Use the parts list, schematic diagram and component layout to
iden-
tify parts and check point locations.
13
DC Voltage Checks
1.
Battery:
2.
-HV
VBAT>7.5V
to (Analog Ground) = -1-2.9 ±0.3V
3. -fV to (Digital Ground) = +5V ±1V
4.
Reference Diode
CR101 (Anode) to (Analog Ground) = -1.23V ±0.03V.
5. Negative reference divider (R114) referenced to analog ground. *
CMOS devices are designed to function at high impedance levels.
Normal static charge can destroy these devices. Table 4 lists all the
static sensitive devices for the Models 873C and 873F. Steps 1
through 7 provide instructions on how to avoid damaging these
devices.
Table 4. Static Sensitive Devices
Schematic
Designation
U104
U105
U107
U108
U109
1.
Devices should be handled and transported in protective
tainers, antistatic tubes or conductive foam.
2.
Use a properiy grounded work bench and a grounding wriststrap.
3. Handle devices by the body only.
4.
Printed circuit boards must be grounded to bench while inserting
devices.
5. Use antistatic solder removers.
6. Use grounded tip soldering irons.
7. After devices are soldered or inserted into sockets they are protected and normal handling can resume.
con-
15
SPECIFICATIONS
MODEL 873F SPECIFICATIONS
TEMPERATURE SENSOR: Type K INICr-NiAl) Thermocouple,
RANGE SPAN RESOLUTION IBS to SyP; 2 years)
200°F -60.0° to 199.9° 0.1° ±(0,25% rdg -f-2.0°F)
2000°F -60° to 1999° 1^ ± (0.25% rdg -t- 2°F)
•Includes NBS confofmity,
82<'FI and
rors around 32°F may be compensated for by an intemal zero adjustment,
REPEATABILITY:
temperature.
TEMPERATURE COEFFICIENT: 65° to 82°F; included in accuracy
specification. From 14° to 65° and 82° to 122°F: less than ±(0,025%
of rdg -(-0.2°F)/°F,
INPUT CURRENT: 25nA typical, 50nA maximum.
TEMPERATURE ACCURACY*
coW
repeatability
junction compensation, temperature coefficient
errors.
Excludes
therrtiocouple
0.3°F
typical for 1 week at constant ambient
errors;
however thermocouple
165°
to
er-
MODEL 8730 SPECIFICATIONS
TEMPERATURE SENSOR: Type K (Ni&NiAl) Thermocouple,
RANGE SPAN RESOLUTION (18 to 28°C: 2 yeara)
200°C -40.0° to 199.9° 0.1° ± (0.25% rdg -t-
1000°C -40° to 1000° 1° ±(0.25% rdg -I- 1°C)
"Includes NBS conformity, cold junction compensation, temperature coefficient
(18° to 28°C) and repeatability errors. Excludes thermocouple errors; however tfiermocouple errors around 0°C may be compensated for by
REPEATABILrrV:
temperature.
TEMPERATURE COEFFICIENT: 18° to 28°C; included in accuracy
specification. From -10° to 18° and 28° to 50°C: less than ±(0.025%
of rdg -(-0.1°C)/°C.
INPUT CURRENT: 25nA typical, 50nA maximum.
TEMPERATURE ACCURACY*
-55.0° to -40.0° 0.1° ±(0.25% rdg -t- 2.0°C)
-55° to -40° If ± (0.25% rdg + 2°C)
an
intemal zero adjustment,
0.2°C
typical for 1 week at constant ambient
1.0°C)
Specifications subject to change without notice.
16
MODEL 873F and 873C GENERAL SPECIFICATIONS
(unless specified):
DISPLAY: 314 digit LCD, 13mnfi (0.5") height. Polarity and decimal point
indication.
CONVERSION RATE: 2.5 readings per second.
OVERRANGE AND OPEN SENSOR INDICATION: 3 least significant
digits blanked,
MAXIMUM COMMON MODE VOLTAGE: 42V peak to earth.
NORMAL MODE REJECTION RATIO: Greater than 40dB at 50 & 60Hz.
COMMON MODE REJECTION RATIO: Greater than 120dB at DC, 50 &
60Hz,
ENVIRONMENTAL LIMITS FOR OPERATING (Model 873F): 14° to
122°F,
less than 80% relative humidity up to 95°F; lineariy derate 1.5%
R.H./°F
ENVIRONMENTAL LIMITS FOR OPERATING (Model 873C): -10° to
ENVIRONMENTAL LIMITS FOR STORAGE (Model 873F): -30° to
ENVIRONMENTAL LIMITS FOR STORAGE (Model 873C): -35° to
POWER: 9V alkaline or carbon-zinc (NEDA 1604) battery.
BATTERY LIFE, CONTINUOUS: 350 hours typical with alkaline; 200
BATTERY INDICATOR: Display indicates "LO BAT" when less than 10%
DIMENSIONS: 160mm long x 69mm vinde x 31mm thick (6,3" x
CONSTRUCTION: Heavy duty ABS plastic housing.
ACCESSORIES SUPPLIED: Battery, Instruction Manual and Model
from 95° to 122°F.
50°C,
less than 80% relative humidity up to 35°C; lineariy derate 3%
R,H./°C
from 35° to 50°C.
140°F,
less than 90% relative humidity up to 95°F; linearly derate 1.5%
R,H,/°F
from 95° to 140°F,
60°C,
less than 90% relative humidity up to 35°C, linearly derate 3%
Resistor, 10k, 5%, 1/4W, Composition
Thick Film Resistor Networic
Thick Rim Resistor Network
Resistor, 10M, 5%, 1/4W, Composition
Potentiometer, Ik
Potentiometer, 50012
Potentiometer, 2k
Thick Film Resistor Network
Resistor, 191k, 1%, 1/8W, Metal Film
Resistor, 390k, 5%, 1/4W, Composition
Thick Film Resistor Network
Factory Selected Resistor (Part of
A31419)
Factory Selected Resistor (Part of
A31410)
Potentiometer, 2k
Factory Selected Resistor (part of
A31419)
.^^lf,
50V, Ceramic
.l/iF,
50V, Ceramic
4.7/JF,
25V, Aluminum
.33MF,
63V, Metalized
Sch
Location
Fl
Bl
C2
82
G1
G1
E2
F2
E2
G5
Dl
E3
A2
Fl
Fl
B2
82
E3
A2
Sev
Sev
Dl
Dl
Dl
D2
Sev
F2
F2
Sev
83
83
83
C3
18
Modei 873C and 873F, Parts Ust (Cont.)
Sch
Desig.
R114*
RIM""
R115
R116
R117
R118
S101
U101
U102
U103
U104
U105
U106
U107
U108
U109
VR101
W101
•Model 873F
"Model 8730
NOTE:
Description
Thick Film Resistor Network
Thick Film Resistor Networi<
Potentiometer, 2k
Resistor, 100k, 5%, 1/4W, Composition
Resistor, 100k, 5%, 1/4W, Composition
Resistor, 121k, 1%, 1/8W, Metal Film
Switch,
4P3T, On/Off Range Switch
Op Amp, LM308 (Factory selected; part
of A31419)
Current Source, LM334 (Factory selected;
part of A31410)
Low Power JFET Op Amp, TL062Cp
3-1/2 Digit A/D Converter, ICL7136
Quad 2 Input NOR Gate, 4001
Dual Synchronous Up Counter, 4518
12 Stage Binary Counter, 4020
Analog Multiplexer, 4051
Analog Multiplexer, 4053
Reference Diode, 1.23V
Jumper Wire
Mechanical Parts can be found in the expk>ded view (Figure 3).
Location
D2/G5
Sch
Sev
Sev
F4
G4
86
Dl
Sev
C2
03
F2
Sev
05
05
D5
G5
85
19
•
+T] SCHEMATIC DIAGRAM
! «f;r
a
=
»
a
il
i
^
^
'^H-
15 ^» i-
5 » I
ir^
*i
^""
I 11
8
tiM
RII3(2fe7K)_
RII3
•o
JIOOI
OR
(I2IK)-
JIOOI
"^
(ZDCI02
SIDE
^
CIO!
ED
2
SIOI
C109' 5 H-y >
C106
Q
iftiilj—
Rmz>
RIQ&
RIGS
RII5
n
n
1
JIOO2 JI003
n
o
o
TJ
o
z
m
00
O
O
WIOl
UI06
5
<
O
c
ut
OiVIEGA^...
YOUR SOURCE FOR
SENSING AND CONTROL
This is the FREE
complete reference
book you will want to
keep within reach at
all times. Send for your
LATEST COPY now!
One Omega Drive, Box 4047
Stamford,
(203)359-1660 Telex: 996404
FAX: (203)359-7700
CT 06907-0047
HMDilOOKAiio
BIEYiaOFEIllA
Phone: (203)359-1660
/^OMEGA
lA^
.AH
An OIWEGA Group Company
ENGiiUEERiNG.
IIMC.
"The WORLD'S Leader in the Field of
Sensing and Control"
n Thermocouple Probes and Assemblies
D Industrial Thermocouple Assemblies
n Infrared and Hand Held Instruments
D Hand Held and Surface Thermocouple Probes
n RTD Probes and Assemblies
D Thermistor Probes and Assemblies
D Connector Systems and Panels
a Wire: Thermocouple, RTD and Thermistor
D Cements, Coatings and Epoxles
n Calibrators and Ice Point References
D Analog Meters and Amplifiers
D Digital Thermometers
D Temperature and mV Recorders
n Solid State Controllers and Relays
D Digital Process Monitors
D Data Acquisition Modules and Data Loggers
D Computer Sensor Interface
D Technical and Reference Books
Return Requests/Inquiries
Direct all warranty and repair requests/inquiries to OMEGA Customer
Service Department, telephone number (203)359-1660, Before
returning any instrument, please contact the OMEGA Customer
Service Department to obtain an authorized return (AR) number. The
designated AR number should then be marked on the outside of the
return package.
To avoid processing delays, also please be sure to include: