United Kingdom:25 Swannington Road,P.O. Box 7, Omega Drive,
ISO 9002 CertifiedBroughton Astley, Leicestershire,Irlam, Manchester,
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is
constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the
CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any
errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
6.1Parts List for Models 868 and 869 ..............................................................................24
Page 4
jumper
Page 5
2.2BATTERY INSTALLATION
A nine volt battery is supplied with the instrument but is not installed, to
avoid possible damage due to leakage during storage or shipment.
Install the battery as follows:
1. Remove the instrument back cover.
2. Insert the battery in place; make sure correct polarity is observed
and battery terminals are contacting the battery clips.
3. Install back cover.
SECTION 3 OPERATION
3.1SAFETY PRECAUTION AND NOTES
WARNING
Do not subject the probe to a voltage more than 30 V
RMS, 42.4 V peak above earth ground, or a shock
hazard may result.
NOTE
Use only 100 ohm platinum RTD sensors that conform to the DIN
43760 standard (alpha = .00385). Other type sensors will give
inaccurate results. Maximum allowable lead resistance for
rated accuracy is 50 ohms per lead (four wire) or 10 ohms per
lead (three wire).
For best accuracy, it is recommended that the instruments be
used in the four wire configuration with a suitable four wire
probe. If three wire probes are used with the instrument in the
four wire mode, noisy readings will result (the displayed reading
will jump around).
3.2CONTROL AND DISPLAY
Figure 3-1 shows the control layout and nomenclature. The following
paragraphs contain information on probe connection, three and four
wire selection, range selection, and basic temperature measuring
procedures.
NOTE
2
Page 6
Page 7
figure 3-2.
jumper
Page 8
Page 9
SECTION 4 THEORY OF OPERATION
This section contains a brief description of operation that should
help a technician in understanding instrument operation during a
measurement, to aid in isolating possible malfunctions. Detailed
schematics of each model are provided at the end of the manual.
4.1THREE WIRE SIGNAL CONDITIONING
Two phases are necessary to condition the signal for digitization.
Each of these phases has a period of one-half the A/D converter back
plane (BP) period. These two phases are called the zero phase and the
signal phase, respectively. During each phase, appropriate FETs are
switched on to configure the circuit as required.
4.1.1Zero Phase
During the zero phase, certain FETs are switched on to configure the
circuit shown in Figure 4-1. This curcuit operates similarly to a sampleand-hold curcuit, in which voltage levels are capacitively stored for
later use. The following summarizes this function.
1. CRis connected across RR. The voltage charged on CRis
equal to IRR.
2. CZis connected across RZ+ L through LZ. The voltage charge
on CZis equal to I[RZ+ L].
Figure 4-1. Three Wire Zero Phase
6
Page 10
3. The input of A1is connected to ground. The output of A1is
equal to [A1Vos].
4. CC is connected between the output of A1and common. The
voltage charged on CCis equal to the output voltage of A
or [A1Vos].
5. CAis connected between the output of the A2network and
common. The voltage charged on CAis equal to the output
voltage of A1attentuated by A2or [A1A2Vos].
4.1.2Signal Phase
During the signal phase, the FET switching configuration changes so
that the voltages developed during the zero phase are connected to the
A/D converter. The configuration during the signal phase is shown in
Figure 4-2. The following discussion assumes that the circuit has gone
through more than one charge transformation:
1. CRis connected across CEand after a sufficient number of
zero and signal phases the voltage on CEapproaches that
stored on CR. That voltage is equal to IRR.
2. CZis connected between L2and the input of A1in a way that
opposes the voltage drop across RTand L4. Since the voltage
charged on CZis I(RZ+ L1) the voltage at the output of A1is
RZis made to equal the value of RTat 0°. This action eliminates the offset of RT(100Ω at 0°C).
3. CCis connected between the output of A1and the top of CDin
such a way that opposes the output of A1. The voltage
charged on CDis
If L4exactly equals L1then this reduces to A1I(RT-RZ).
1
[A1(I(RT+ L4) - I(RZ+ L2) + Vos]
or
[A1(I(RT- RZ+ L4- L1) + Vos]
A1(I(RT- RZ + L4 - L1) + Vos) -A1Vos
or
A1I(RT- RZ + L4 - L1)
4. CAis connected between the output of the A2network and the
top of CBin such a way that opposes the output of A2. The
voltage charged on CBis
A1A2(I(RT- RZ + L4 - L1) + Vos) -A1A2Vos
or
A1A2I(RT- RZ + L4 - L1)
If L4exactly equals L1then this reduces to A1A2I(RT- RZ)
7
.
Page 11
Page 12
4.2FOUR WIRE SIGNAL CONDITIONING
There are two phases necessary to condition the signal for
digitization. Each phase lasts for a period equal to one-half of the
back plane period of the A/D converter. These are called the zero
and signal phase.
4.2.1Zero Phase
During the zero phase, FETs are switched to configure the circuit
shown in Figure 4-3. The fourth wire adds an additional attenuation
that slightly increases the offset voltage at the input to amplifier A1.
1. CRis connected across RR. The voltage charged on CRis
equal to IRR.
2. CZis connected across RZ. The voltage charge in C
is equal to IRZ.
3. The input of A1is connected through L3. Since, ideally, zero
current flows into A1the voltage at the output of A1is equal
to [A1(IL4+ Vos)].
4. CC is connected between the output of A1and common. The
voltage charged on CCis equal to the output voltage of A
or [A1(IL4+ Vos)].
5. CAis connected between the output of the A2network and
common. The voltage charged on CAis equal to the output
voltage of A1attentuated by A2or [A1A2(IL4 + Vos)].
Z
1
Figure 4-3. Four Wire Zero Phase
9
Page 13
4.2.2Signal Phase
During the signal phase the 4 charged values of voltage are
transfered to other parts of the circuit (refer to simplified Figure 4-4).
The following explanations assume more than one charge transfer has
taken place.
1. CRis connected across CEand after a sufficient number of
zero and signal phases the voltage on CEapproaches that
stored on CR. That voltage is equal to IRR.
2. CZis connected between L2and the input of A1in a way that
opposes the voltage drop across RTand L4. Since the voltage
charged on CZis IRZthe voltage at the output of A1is
RZis made to equal the value of RTat 0°C. This action
eliminates the offset of RT(100Ω at 0°C).
3. CDis connected between the output of A1and the top of C
in such a way that opposes the output of A1. The voltage
charge on CDis
4. CAis connected between the output of the A2network and
the top of CBin such a way that opposes the output of A2.
The voltage charged on CBis
A1A2(I(RT - RZ + L4) + Vos) - A1A2(IL4+ Vos)
Due to this two phase measurement Vos and lead resistance
effects are eliminated.
[A1(I(RT + L4) - IRZ + Vos)]
or
A1(I(RT - RZ + L4) + IRZ + Vos)
D
A1(I(RT - RZ + L4) + Vos) - A1(IL4+ Vos)
or
A1I(RT - RZ)
or
A1A2I(RT - RZ)
Figure 4-4. Four Wire Signal Phase
10
Page 14
Page 15
4.6LOW-BATTERY DETECTOR
Low-battery detection is accomplished by comparing the
regulated voltage between V+ and common to the output of the
voltage divider (R107 and R110), which is connected across the
battery. When the battery voltage decreases, the output of the
voltage divider rises above analog common, causing
comparator U103A to change state. This action enables the LO
BAT annunciator on the display.
SECTION 5 SERVICE INFORMATION
5.1DISASSEMBLY
The instrument must be opened to replace the battery, to select
three or four wire operation, or for calibration. Troubleshooting
or parts replacement may require more complete disassembly
(see Figure 5-1). The sequence below describes the basic
procedure for disassembling the instrument.
1. Remove the back cover.
2. At this point, you obtain access to replace the battery, select
three wire or four wire operation, or perform calibration.
Replace cover.
3. To remove the PC board, unscrew the standoff securing the
board to the front case. The board may now be pulled free,
although the wires to the probe jack will still be attached.
When the board becomes free of the case, the switch cover
will pull free as well. If necessary, the board can be pulled
completely free by detaching the wires at the board end. The
probe jack may then be removed, if necessary, by removing
the nut securing it to the case and pushing the jack free from
the inside.
CAUTION
Handle the PC board only at the edges, whenever
possible, to avoid possible contamination, which could
degrade instrument performance.
4. The LCD assembly may be removed from the PC board by
carefully spreading the clips that secure the display to the
board. Once the assembly is free of the board, the various
parts will be loose, so handle the LCD with care.
12
Page 16
CAUTION
Do not touch the elastomer contact strips or mating surfaces
on the PC board. Also, use care when spreading the clips to
avoid breaking them.
5. The instrument may be reassembled by reversing the above
prodcedure, using Figure 5-1 as a guide. When assembling
the instrument, take special note of the following points:
A. If the LCD was removed, be sure it is aligned properly
on the board, with the clips fitted properly into the
notch on each side of the board.
B. If the probe jack was disconnected, make certain the
wires are connected properly. The wire colors are
marked on the PC board.
C. Make sure the PC board is properly secured to the front
case with the standoff. The switch cover should be placed
on the switch when installing the PC board.
D. Once the rear cover is in place, secure it with the
attachment screws.
Figure 5-1. Exploded View
13
Page 17
jumper
Page 18
Figure 5-2.
Connections for Performance Verification and Calibration
The procedure outlined in the last paragraph provides accurate
absolute instrument calibration, but it cannot compensate for
probe inaccuracy. Probe errors near 32°F (0°C) can be minimized
by using the following procedure.
1. Make up an ice water bath by firmly packing a dewar flask
or Thermos with pea-size ice cubes made of distilled water
and then filling the container with distilled water. Replace
melted ice with new ice while removing excess water during
the calibration procedure.
2. Connect the probe to be compensated to the instrument.
3. Drill a hole in the flask or Thermos cap just large enough to
accommodate the probe. Place the cap on the bath
container and pass the probe through the hole until the
probe tip rests at the center of the ice water bath.
4. Allow 20 minutes for the test fixture temperature to stabilize.
With the model 868 on the 200°F range, adjust R104 for a
reading of 32.0 on the display. For the Model 869, R104
should be adjusted for a reading of 00.0 with the instrument
on the 200°C range.
Using this method of probe calibration will uncalibrate the
instrument slightly when used with other probes.
5.4TROUBLESHOOTING
The troubleshooting information, along with appropriate
schematics and parts lists, are included to serve as a guide to
enable equipment repair. The schematics and parts lists may vary
slightly from actual production units and are to be used as a guide
only. Likewise, the troubleshooting section is a guide only and
cannot cover all possible contingencies that may occur.
NOTE
Each Model 868 and 869 is covered by a 13 month warranty as
described on the inside front cover of this manual. Warranty will
be void if the unit shows evidence of having been tampered with.
To gain access to the PC board for troubleshooting, the rear
cover and internal shield must be removed, as described in the
disassembly instructions. When troubleshooting or replacing
components, handle the PC board only by the edges to avoid
possible contamination.
Recommended Equipment: Digital Multimeter (DMM) with 10
megohm input resistance, ±0.05% basic accuracy; triggered
sweep oscilloscope with dc to 10 MHz bandwidth.
16
Page 20
5.4.1Voltage Checks
Several voltage checks can be made simply by connecting a
DMM to various A/D converter IC pins. Table 5-2 summarizes
these voltage readings.
STEPITEM/COMPONENTCONDITIONCOMMENTS
1Connect DMM HI to +VLeave connected for
2Connect DMM LO to>7.2 V dcBattery voltage
U102, pin 26 (-V)
3Connect DMM LO to5V ± 1 V dvDigital Common
U102, pin 37 (TEST)
4Connect DMM LO to3 VIN LO (COM)
U102, pin 30
5.4.2A/D Converter Checks
TABLE 5-2
VOLTAGE CHECKS
REQUIRED
all voltage checks.
A/D converter operation can be checked by measuring the input
and reference voltages and then comparing the displayed reading
to a value calculated from these input voltages.
1. Connect a temperature probe to the instrument; make sure
the temperature remains stable while making voltage
measurements.
2. Connect the DMM LO terminal to analog common (pin 32 of
U102).
3. Connect the DMM HI terminal to pin 31 (IN HI) of U102 and
record the reading.
4. Connect the DMM HI terminal to pin 30 (IN LO) of U102
and record the reading.
5. Connect the DMM HI terminal to pin 35 (REF HI) of U102
and record the voltage.
6. Connect the DMM HI terminal to pin 36 (REF LO) and record
the reading
Using the measured values above, calculate the displayed
reading (neglecting the decimal point) as follows:
Display = 1000(IN HI-IN LO)
(REF HI-REF LO)
If the displayed value does not agree with the calculated value to
within a few digits, the A/D converter is not functioning properly.
17
Page 21
5.4.3Waveform Checks
Several A/D converter waveform checks can be made using an
oscilloscope along with the information in Table 5-3.
1. Connect the oscilloscope LO input to analog common U102,
pin 32.
2. Connect the oscilloscope HI input to the A/D converter pin
indicated in the table.
3. Select an appropriate time base, input attenuator setting,
and trigger mode to stabilize the display.
4. Compare the displayed waveform with the corresponding
drawing in Table 5.3.
WAVEFORM CHECKS
TABLE 5-3
18
Page 22
5.5STATIC-SENSITIVE PARTS
MOS devices are designed to operate at very high impedance
levels. As a result, any normal static charge that builds up on
your person or clothing may be sufficient to destroy these
devices if they are not handled properly. Table 5-4 lists those
parts used in the Models 868 and 869 that might be destroyed
by static charge. When handling these devices, use the following
precautions:
1. Transport and handle these parts only in containers
designed to prevent static build-up. Typically, these parts will
be received in static-protected containers of plastic or foam.
Keep these devices in their original containers until ready for
installation.
2. Remove the devices from their protective containers only at a
properly grounded work station. Also ground yourself with a
suitable wrist strap.
3. Handle the devices only by the body; do not touch the
terminals or pins.
4. Any PC board into which the device is to be installed must
also be properly grounded.
5. Use only anti-static type solder suckers.
6. Use only grounded soldering irons.
7. Once the device is installed on the PC board, it is normally
adequately protected, and normal handling may resume.
CAUTION
This assembly contains electrostatic sensitive devices which
can be damaged by static discharge when touched. Observe
precaution when handling.
This performance verification procedure should be accomplished
after any parts replacement or circuit repair, or to check instrument
operation at any time incorrect operation is indicated.
19
Page 23
jumper
jumper
Page 24
22
SPECIFICATIONS continued
MODEL 869
TEMPERATURE SENSOR TYPE:
Three wire or four wire 100 Ωplatinum RTD (alpha = .00385)
4WIRE
ACCURACY*
SETTINGRANGE RESOLUTION(18 °C TO28 °C: 1 Year)
200 °C-100.0 °C to 199.9 °C 0.1
± 0.3 °C
-199.9 °C to -100.1 °C 0.1
± 1.5 °C
630 °C-100 °C to 630 °C 1
± 1 °C
-220 °C to -101 °C
1
±
2 °C
*ACCURACY:
Threewire accuracyisthe same if contactresistance errors areremoved by calibrationof instrument plusprobe at 0°C.Includes DIN43760 (ITS-90) conformity, repeatability,temperature coecient (18° to 28°C), time stability (one year) and errors with up to 50 Ω of lead resistance(each lead). Excludes probeerrors; however, probe errors around 0°C may be compensated by an internal adjustment.
REPEATABILITY:
0.1 °Cambient temperature.
TEMPERATURE COEFFICIENT:
18° to 28°C; included inaccuracy specication. From
-10°to 18°C, and28° to50°C: less than±0.015°C/°C.
MAXI
MUM LEAD RESISTANCE:
(each lead):
Four wire: 50 ΩThree wire: 10 Ω
SENSOR CURRENT:500 µAmax.
MODELS 868 AND 869GENERAL SPECIFICATIONS:
DISPLAY:
31/2
digit LCD, 0.5" (13 mm) height. Polarity and deci-
mal point indication
.
CONVERSION RATE:
1.5 readings per second.
OVERRANGE AND OPENSENSOR INDICATION:
Typical for one week at constant
PERFORMANCE VERIFICATION
RANGE °CVALUE (Ω)(18° to 28°C)
630°C9.9 Ω-221 to -219
200°C60.26 Ω-100.3 to -99.7
200°C100.00 Ω-0.3 to +0.3
200°C138.51 Ω99.7 to 100.3
200°C174.01 Ω194.7 to 195.3
630°C313.71 Ω599 to 601
SECTION 6 SPECIFICATIONS
RESISTANCEALLOWABLE READING
TEMPERATURE SENSOR TYPE:
SETTINGRANGE RESOLUTION(18 °C TO 28 °C: 1 Year)
200 °F-100.0 °F to 199.9 °F 0.1
1100 °F -100 °F to 1100 °F 1
*ACCURACY:
-199.9 °F to -100.1°F 0.1
-360 °F to -101 °F 1
Three wire accuracy is the same if contact
resistance errors are removed by calibration
of instrument plus probe at 32°F. Includes DIN
43760 (ITS-90) conformity, repeatability,
temperature coecient (65° to 82°F), time
stability (one year) and errors with up to 50 Ω
of lead resistance (each lead). Excludes probe
errors; however, probe errors around 32°F may
be compensated by an internal adjustment.
TABLE 5-5
(continued)
MODEL 869
MODEL 868
Three wire or four wire 100 Ω
platinum RTD (alpha = .00385)
4 WIRE
ACCURACY*
± 0.4 °F
± 1 °F
2 °F
±
±
4 °F
REPEATABILITY:0.2 °F Typical for one week at constant
ambient temperature.
TEMPERATURE COEFFICIENT: 65 °F to 82 °F; included in
MAXIMUM LEAD RESISTANCE:
(each lead):Four wire: 50 Ω
SENSOR CURRENT: 500 µ A max.
Three wire: 10 Ω
accuracy specication. From
14 °F to 65 °F, and 82 °F
to 122 °F: less than ±0.015 °F/°F.
21
Page 25
SPECIFICATIONS continued
TEMPERATURE SENSOR TYPE:
SETTINGRANGE RESOLUTION(18 °C TO 28 °C: 1 Year)
200 °C-100.0 °C to 199.9 °C 0.1
-199.9 °C to -100.1 °C 0.1
630 °C-100 °C to 630 °C 1
-220 °C to -101 °C
*ACCURACY:
Three wire accuracy is the same if contact
resistance errors are removed by calibration
of instrument plus probe at 0°C. Includes DIN
43760 (ITS-90) conformity, repeatability,
temperature coecient (18° to 28°C), time
stability (one year) and errors with up to 50 Ω
of lead resistance (each lead). Excludes probe
errors; however, probe errors around 0°C may
be compensated by an internal adjustment.
MODEL 869
Three wire or four wire 100 Ω
platinum RTD (alpha = .00385)
1
4 WIRE
ACCURACY*
± 0.3 °C
± 1.5 °C
± 1 °C
±
2 °C
REPEATABILITY:
TEMPERATURE COEFFICIENT:
MAXI
MUM LEAD RESISTANCE:
(each lead):
Four wire: 50 Ω
Three wire: 10 Ω
SENSOR CURRENT:500 µ A max.
MODELS 868 AND 869
GENERAL SPECIFICATIONS:
DISPLAY:
CONVERSION RATE:
OVERRANGE AND OPEN
SENSOR INDICATION:
0.1 °C
Typical for one week at constant
ambient temperature.
18° to 28°C; included in
accuracy specication. From
-10° to 18°C, and 28° to
50°C: less than ±0.015°C/°C.
digit LCD, 0.5" (13 mm) height. Polarity and deci-
31/ 2
mal point indication
.
1.5 readings per second.
22
Page 26
SPECIFICATIONS continued
MAXIMUM COMMON MODE
VOLTAGE:
COMMON MODE REJECTION
(Model 868):
42 V peak to earth.
Less than 0.001°F/volt at dc, 50 and 60 Hz
(100 Ω unbalance, LO driven).
COMMON MODE REJECTION
(Model 869):
Less than 0.001°C/volt at dc, 50 and 60 Hz
(100 Ω unbalance, LO driven).
ENVIRONMENTAL LIMITS FOR
OPERATING (Model 868):
14° to 122°F, less than 80% relative humidity
up to 95°F; linearly derate 1.5% RH/°F from
95° to 122°F.
ENVIRONMENTAL LIMITS FOR
OPERATING (Model 869):
-10° to 50°C, less than 80% relative humidity up to
35°C; linearly derates 3% RH/°C from 35° to 50°C.
ENVIRONMENTAL LIMITS FOR
STORAGE (Model 868):
-30° to 140°F, less than 90% relative humidity
up to 95°F; linearly derate 1.5% RH/°F from 95°
to 140°F.
ENVIRONMENTAL LIMITS FOR
STORAGE (Model 869):
-35° to 60°C, less than 90% relative humidity up to
35°C; linearly derate 3% RH/°C from 35° to 60°C.
RTD LINEARIZATION:Ratiometric dual-slope A/D with continuous
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials andworkmanship for a period of 13 monthsfrom date of purchase. OMEGA Warranty adds anadditional one (1) month grace period to the normal one (1) year product warrantyto coverhandling and shipping time. This ensures that OMEGA’s customers receive maximumcoverage on each product.
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’sCustomer Service Department will issue an Authorized Return (AR) number immediately uponphone or written request. Upon examination by OMEGA, ifthe unit is found to be defective it willbe repaired or replaced at no charge. OMEGA’sWARRANTY does not apply to defects resultingfrom any action of the purchaser, including but not limited to mishandling, improper interfacing,operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidenceof being damaged as a result of excessive corrosion; or current, heat, moisture or vibration;improper specification; misapplication; misuse or other operating conditions outside ofOMEGA’scontrol. Components which wear are not warranted, including but not limited tocontact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liabilityfor any damages that result from the use of its products in accordance with informationprovided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED ORIMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANYWARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AREHEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forthherein are exclusive and the total liability of OMEGA with respect to this order, whetherbased on contract, warranty, negligence, indemnification, strict liability or otherwise, shallnot exceed the purchase price of the component upon which liability is based. In no eventshall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGAisnot intended to be used, nor shall it be used: (1) asa “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity;or (2) in medical applications or used on humans. Should any Product(s) be used in or with anynuclear installation or activity, medical application, used on humans, or misused in any way,OMEGAassumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language,and additionally, purchaser will indemnify OMEGAand hold OMEGAharmless from any liabilityor damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN ANAUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT(IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then bemarked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging toprevent breakage in transit.
FOR WARRANTYRETURNS, please havethe following information available BEFORE contacting OMEGA:
1.P.O. number under which the product was PURCHASED,
2.Model and serial number of the productunder warranty, and
3.Repair instructions and/or specific problems relative to the product.
FOR NON-WARRANTYREPAIRS,
consultOMEGA for current repair charges. Have thefollowing information available BEFORE contacting OMEGA:
1. P.O. number to cover the COST
of the repair,
2.Model and serial number of product, and
3.Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, withoutprior written consent of OMEGA ENGINEERING, INC.
USA
MADE
IN
NOTES
31
Page 35
MADE
IN
USA
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and
workmanship for a period of 13 months from date of purchase. OMEGA Warranty adds an
additional one (1) month grace period to the normal one (1) year product warranty to cover
handling and shipping time. This ensures that OMEGA’s customers receive maximum
coverage on each product.
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s
Customer Service Department will issue an Authorized Return (AR) number immediately upon
phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will
be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting
from any action of the purchaser, including but not limited to mishandling, improper interfacing,
operation outside of design limits, improper repair, or unauthorized modification. This
WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence
of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration;
improper specification; misapplication; misuse or other operating conditions outside of
OMEGA’s control. Components which wear are not warranted, including but not limited to
contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability
for any damages that result from the use of its products in accordance with information
provided by OMEGA, either verbal or written. OMEGA warrants only that the parts
manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER
WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR
IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY
WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE
HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth
herein are exclusive and the total liability of OMEGA with respect to this order, whether
based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall
not exceed the purchase price of the component upon which liability is based. In no event
shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as
a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity;
or (2) in medical applications or used on humans. Should any Product(s) be used in or with any
nuclear installation or activity, medical application, used on humans, or misused in any way,
OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language,
and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability
or damage whatsoever arising out of the use of the Product(s) in such a manner.
WARRANTY/DISCLAIMER
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN
AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT
(IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be
marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to
prevent breakage in transit.
FOR WARRANTY
the following information available BEFORE
contacting OMEGA:
1. P.O. number under which the product was
PURCHASED,
2. Model and serial number of the product
under warranty, and
3. Repair instructions and/or specific
problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible.
This affords our customers the latest in technology and engineering.
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without
prior written consent of OMEGA ENGINEERING, INC.
RETURNS, please have
FOR NON-WARRANTY REPAIRS,
consult
OMEGA for current repair charges. Have the
following information available BEFORE
contacting OMEGA:
1. P.O. number to cover the COST
of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems
relative to the product.
Page 36
0315
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.