Omega 865, 866 User Manual

Page 1
®
User’s Guide
http://www.omega.com
e-mail: info@omega.com
Digital Thermometers
Model 865 and 866
Page 2
omega.com
TM
OMEGA
®
OMEGAnetSMOn-Line Service Internet e-mail
http://www.omega.com info@omega.com
USA: One Omega Drive, Box 4047
ISO 9001 Certified Stamford, CT 06907-0047
Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: canada@omega.com
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
Customer Service: 1-800-622-2378 / 1-800-622-BEST Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
SM
SM
SM
Mexico and Latin America:
Tel: (95) 800-TC-OMEGA En Espan˜ol: (203) 359-1660 ext: 2203 e-mail: espanol@omega.com
SM
FAX: (95) 203-359-7807
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: (31) 20 6418405 FAX: (31) 20 6434643 Toll Free in Benelux: 06 0993344 e-mail: nl@omega.com
Czech Republic: Ostravska 767, 733 01 Karvina
Tel: 42 (69) 6311899 FAX: 42 (69) 6311114 e-mail: czech@omega.com
France: 9, rue Denis Papin, 78190 Trappes
Tel: (33) 130-621-400 FAX: (33) 130-699-120 Toll Free in France: 0800-4-06342 e-mail: france@omega.com
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: 49 (07056) 3017 FAX: 49 (07056) 8540 Toll Free in Germany: 0130 11 21 66 e-mail: germany@omega.com
United Kingdom: 25 Swannington Road, P.O. Box 7, Omega Drive,
ISO 9002 Certified Broughton Astley, Leicestershire, Irlam, Manchester,
LE9 6TU, England M44 5EX, England Tel: 44 (1455) 285520 Tel: 44 (161) 777-6611 FAX: 44 (1455) 283912 FAX: 44 (161) 777-6622
Toll Free in England: 0800-488-488
e-mail: uk@omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts
no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
Page 3
TABLE OF CONTENTS
MODEL 865 AND 866 DIGITAL THERMOMETERS
SECTION PAGE
SECTION 1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
SECTION 2 INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.1 Unpacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.2 Preparation for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.3 Verification Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
SECTION 3 OPERATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
3.1 Safety Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
3.2 Operating Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
3.2.1 Model 865. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
3.2.2 Model 866. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
SECTION 4 THEORY OF OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
SECTION 5 SERVICE INFORMATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
5.1 Performance Accuracy Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
5.2 Disassembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
5.2.1 Special Handling of Static-Sensitive Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
5.2.2 Disassembly Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
5.2.3 Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.3 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.4 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.4.1 DC Voltage Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.4.2 A/D Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.4.3 Waveform Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
SECTION 6 SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.1 PARTS LIST and Component Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
i
Page 4
SECTION 1 INTRODUCTION
1.1 GENERAL DESCRIPTION
The OMEGA large, easy to read, liquid crystal displays (LCD). The input connection for each instrument consists of a the available accessory thermistor based probes and sensors. The resistance value of the thermistor at 25°C (77°F) is 2252 ohms. The instruments are pow­ered by a 9 V battery. Continuous operation of 350 hours is typical with an alkaline battery or 200 hours typical with a carbon-zinc battery.
The Model 865 measures temperatures on two ranges. The 300°F range measures temperature from –70°F to 300°F with a resolution of 1°F. The 200° F range measures temperature from –70°F to 199.9°F with a resolution of 0.1°F.
The Model 866 measures temperature from –55° to 150°C, on one range, with a resolution of 0.1°C.
1.2 FEATURES
®
Models 865 and 866 are Hand Held Digital Thermometers with
1
4 inch phone jack, which accommodates any one of
• High Accuracy/Resolution/Repeatability—to 0.1°
• Highly Versatile—Many applications with different probe configurations (probes not supplied)
• Continuous Automatic Linearization—Dual slope A/D
• Warning Indication—Overrange/Open Sensor/Low Battery
• Long-Life Battery—9 V Alkaline (MN-1604)
• Rugged Construction—Heavy-Duty ABS Plastic Housing
SECTION 2 BEFORE USING
2.1 UNPACKING
Remove the packing list and verify that all the equipment has been received. If there are any questions about the shipment, please call OMEGA Customer Service Department at (203) 359-1660.
Upon receipt of the shipment, inspect the container and equipment for any signs of damage. Take particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event reshipping is necessary.
NOTE
1
Page 5
2.2 PREPARATION FOR USE
The Models 865 and 866 are ready for use, after the battery is installed. To install the battery, refer to paragraph 5.2.3.
2.3 VERIFICATION CHECK
Install a probe and, at room temperature, hold the tip of the probe between your thumb and index finger. A reading between 25°C and 40°C should be obtained for the Model 866, and a reading of 77°F to 104°F should be obtained for the Model 865, to confirm the functioning of the instrument.
SECTION 3 OPERATION
3.1 SAFETY PRECAUTIONS
A shock hazard exists on the input jack when probes or sensors are exposed to voltage levels greater than 42 volts, peak to earth ground. Do not exceed 42 volts, peak to earth ground
WARNING
Do not attempt to measure temperatures beyond the range of the probe being used. Probe damage may occur. Maximum probe temperatures are given in the optional accessories section. Do not touch the probe tip when measuring excessively high or low temperatures.
3.2 OPERATING PROCEDURES
3.2.1 Model 865 (See Figure 3-1)
1. Turn power on by selecting the desired range.
2. Insert appropriate temperature probe into the input jack.
3. Touch the probe tip to the material to be measured and read the display in degrees Fahrenheit.
CAUTION
NOTE
An open sensor is indicated by a reading on the display of greater than -100.0°. Overrange is indicated by blanking of three least-significant digits. Low battery (10% life remaining) is indicated by LO BAT on the display.
2
Page 6
Figure 3-1 Model 865 Control Layout
3.2.2 Model 866 (See Figure 3-2)
1. Turn power on. Turning power on also selects the 150°C measurement range.
2. Insert appropriate temperature probe into the input jack.
3. Touch the probe tip to the material to be measured and read the display in degrees Celsius.
An open sensor is indicated by a reading on the display of greater than -100.0 °. Overrange is indicated by blanking of the three least-significant digits. Low battery (10% life remain­ing) is indicated by LO BAT on the display.
NOTE
3
Page 7
SECTION 4 THEORY OF OPERATION
The temperature sensor is thermistor based. The thermistor sensor is a negative temperature coefficient device. This means that as the temperature increases, the resistance of the thermistor decreases. This relationship is a logarithmic func­tion. The current that flows through the sensor is inversely proportional to the resistance of the sensor (Ohm’s Law).
U104B and associated resistors minimize the “on” resistance of the gates of U101A and B. The input signal must be conditioned by the Log Ratio amplifier, before being digitized by the A/D converter. The Log Ratio amplifier operates on two phases: the signal phase and the zero phase.
Each phase has a duration equal to converter (U102).
Figure 3-2 Model 866 Control Layout
1
2 of the backplane period of the A/D
4
Page 8
In the signal phase (see Figure 4-1), the source voltage VSis connected to the sen­sor R
. The current that flows through RTalso flows into the collector of Q101. The
T
output voltage of U104C is charged on capacitor C106. Also during this phase, the input to U104A is grounded and the voltage developed is charged on capacitor C105.
In the zero phase, the two charged values of voltage are transferred to other parts of the circuit, as depicted by Figure 4-2. The source voltage V from the sensor and connected to a zero reference resistor R flows through R
also flows in the collector of Q101. The output voltage of U104C
Z
is disconnected
S
. The current that
Z
is connected REF LO on U102 by way of C106 and U103B (see Schematic Diagram 866-106 or 865-106). The voltage that is applied to the input of U104A and across C111 is the difference between the output voltage of U104C and the voltage that was charged on C106 during the signal phase.
The voltage charged on C105 is now opposing the output of U104A and therefore transferred to C110. This is the voltage digitized by the A/D converter (U102). The fixed reference HI voltage for U102 is provided by current source U105 and its associated resistors.
Figure 4-1 Signal Phase
5
Page 9
Figure 4-2 Zero Phase
SECTION 5 SERVICE INFORMATION
5.1 PERFORMANCE ACCURACY CHECK
1. Set up the equipment as shown in Figure 5-1.
2. Follow Table 5-1 to verify instrument performance.
The LO of the resistance box must be connected to the LO (sleeve) of the
ed to the HI (tip) of the ¹/0 phone jack.
¹/0 phone jack. The HI of the resistance box must by connect-
NOTE
Figure 5-1 Calibration Setup
6
Page 10
Table 5-1
PERFORMANCE VERIFICATION
MODEL 865
Step Setting Decade Resistance Allowable Reading
1 200°F 75,750 ohms –39.4 to –40.6 2 200°F 255,700 ohms –68.8 to –71.2 3 200°F 19,220 ohms –00.5 to 00.5 4 200°F 220.8 ohms 188.9 to 191.1 5 300°F 43 ohms 298 to 3021
MODEL 866
Step Setting Decade Resistance Allowable Reading
1 150°C 75,790 ohms –39.6 to –40.4 2 150°C 217,100 ohms –54.2 to –55.8 3 150°C 7,355 ohms –00.3 to 00.3 4 150°C 152.8 ohms 99.4 to 100.6 5 150°C 41.9 ohms 149.2 to 150.7
5.2 DISASSEMBLY
5.2.1 Special Handling Of Static-Sensitive Devices
CMOS devices are designed to function at high impedance levels. Normal static charge can destroy these devices. U101, U102, and U103 are all static-sensitive devices, and steps 1 through 7 provide instructions on how to avoid damaging these devices.
1. Devices should be handled and transported in protective containers, anti-static tubes, or conductive foam.
2. Use a properly grounded workbench and a grounding wrist strap.
3. Handle device by the body only; do not touch terminals, pins, etc.
4. PC boards must be grounded to bench while inserting devices.
5. Use anti-static solder suckers.
6. Use grounded tip soldering irons.
7. After devices are soldered or inserted into sockets, they are protected and normal handling can resume.
7
Page 11
5.2.2 Disassembly Instructions
Turn the instrument off and disconnect the temperature probe before removing the bottom cover.
When disassembling the instrument, handle the PC board by the edges. Body oil, dirt, and moisture can degrade the circuit performance. Keep the PC board clean and free of contaminants.
1. Place the unit face down on a bench or other similar surface and remove the screws from the bottom cover.
2. Separate the bottom cover form the rest of the instrument by grasp­ing the top of the case (above the display) and gently lifting it away from the display.
3. Remove the battery.
WARNING
NOTE
4. Remove the standoff that secures the PC board to the case. (It is located in the center of the PC board.)
5. Remove the switch cover for the Model 866 and for the Model 865.
6. Lift the PC board out of the top cover (front panel).
CAUTION
When removing the LCD from the PC board, be careful not to spread the four retaining hooks on the clear plastic bezel too far. Plastic bezel hooks may fracture.
7. The LCD is held in place on the PC board by a shock resistant­mounting and is assembled as shown in Figure 5-2.
8. To reassemble the unit, reverse the above procedure.
8
Page 12
Figure 5-2 Models 865 and 866 Exploded View
When reassembling the instrument, be sure to reinstall the switch cover. Common-mode voltage may be present on the switch, creating a hazard if the cover is not reinstalled.
WARNING
9
Page 13
5.2.3 Battery Replacement
Turn the instrument off and disconnect the input probe before removing the bottom cover.
1. Remove the bottom cover of the instrument (see Figure 5-2).
2. Place the battery (MN-1604) in the battery compartment. Observe proper polarity (see Figure 5-3).
3. Reinstall the bottom cover.
WARNING
5.3 CALIBRATION
Calibration should be performed yearly or whenever the performance verification procedure indicates that the instrument is out of specification under laboratory conditions, having an ambient temperature of 23°C63°C and a relative humidity of less than 80%.
Use the following procedure to calibrate either Model 865 or Model 866. To locate the adjustment points, remove the bottom cover.
See Figure 5-3 above.
To prevent a shock hazard, turn the instrument off and remove the input probe from the instrument before remov­ing the bottom cover.
Figure 5-3 Battery Installation
WARNING
10
Page 14
1. Set up the equipment as shown in Figure 5-1.
2. Turn on the instrument.
3. Perform calibration. refer to Table 5-2 or Table 5-3, as applicable.
4. Verify correct calibration. Refer to paragraph 5.1, performance accuracy check.
5. Reinstall the bottom cover.
TABLE 5-2
MODEL 865 CALIBRATION
Step Range Input Adjustment Reading
A 200°F 19220 ohms R111 00.0 6 0 counts B 200°F 220.8 ohms R110 190.0 6 ¹/0 counts C 300°F 43 ohms R112 300 6 0 counts D R110 and R112 are interactive. Repeat steps B and C until the results are
repeatable.
Step Range Input Adjustment Reading
A 150°C 7355 ohms R111 00.0 6 0 counts B 150°C 152.8 ohms R110 100.0 6 ¹/0 counts C 150°C 41.9 ohms R112 150 6 0 counts D R110 and R112 are interactive. Repeat steps B and C until the results are
repeatable.
5.4 TROUBLESHOOTING
The troubleshooting information, along with appropriate schematics and parts lists, are included to serve as a guide to enable equipment repair. The schematics and parts lists may vary slightly from actual production units and are to be used as a guide only. Likewise, the troubleshooting section is a guide only and cannot cover all possible con­tingencies that may occur.
TABLE 5-3
MODEL 866 CALIBRATION
Each Model 865 and 866 is covered by a 12-month warranty as described on the inside front cover of this manual. Warranty will be void if the unit shows evidence of having been tampered with.
11
Page 15
5.4.1 DC VOLTAGE CHECKS
Using a precision digital voltmeter (DVM), make the following DC voltage checks in Table 5-4:
Step Item/Component Required Remarks
1 U102 pin 1 referenced >7.2V Battery Voltage
to pin26
2 U102 pin 1 referenced +5V 6 1V 5V Power Supply
to pin 37
3 U102 pin 1 referenced À3V
to pin 30
TABLE 5-4
DC VOLTAGE CHECKS
Condition
5.4.2 A/D MEASUREMENTS
To verify that the A/D and associated circuitry are operating properly, measure the four voltage points. Use the following formula to calculate the display reading.
1. Place a stable input to the instrument under test. For example, 190°F for the Model
865 and 150°C for the Model 866.
All the measurements in steps 2 through 5 are referenced to analog common.
2. Measure the voltage at U102 pin 31 (IN HI) and record the value.
3. Measure the voltage at U102 pin 30 (IN LO) and record the value.
4. Measure the voltage at U102 pin 35 (REF LO) and record the value.
5. Measure the voltage at U102 pin 36 (REF HI) and record the value.
NOTE
6. Calculate the display reading using the measured voltages and the following formula;
Display Reading (disregarding decimal point) =
(Pin 31 voltage)-(Pin 30 voltage)
1000 x
(Pin 35 voltage)-(Pin 36 voltage)
12
Page 16
5.4.3 WAVEFORM CHECKS
Using a precision oscilloscope, make the following waveform checks in Table 5-5:
TABLE 5-5
WAVEFORM CHECKS
Step Item/Component Waveform
1. (Referenced to analog ground)
U102 pin 40 (OSC 1)
2. U102 pin 39 (OSC 2)
3. U102 pin 38 (OSC 3) (Same as step 2.)
4. U102 pin 27 (INTEGRATOR)
(with negative full-scale input of approximately 1900 counts)
5. U102 pin 21 (Backplane)
13
Page 17
SECTION 6 SPECIFICATIONS
MODEL 865
TEMPERATURE SENSOR TYPE: Thermistor (2252 Ohm @ 77°F)
RANGE TEMPERATURE ACCURACY*
SPAN RESOLUTION 65° TO 82°F; 1 year
200°F – 40.0° to 199.9° 0.1° 6 (0.3% rdg + 0.5°F)
– 70.0° to – 40.1° 0.1° 6 (0.3% rdg + 1.0°F)
300°F – 70° to 300° 6 (0.3% rdg + 1°F)
*ACCURACY: Includes repeatability, temperature
coefficient (65° to 82°F), time stability (1 year), and linearization conformity errors. Excludes probe errors; how­ever, probe errors around 32°F may be compensated by an internal adjustment.
REPEATABILITY: 0.2°F typical for 1 week at constant
ambient temperature.
TEMPERATURE COEFFIECIENT: 65° to 82°F; included in accuracy
specification. From 14° to 65° and 82° to 122°F: less than 6 (0.06% rdg +0.01°F)/°F.
SENSOR VOLTAGE: 40 mV RMS maximum.
MODEL 866
TEMPERATURE SENSOR TYPE: Thermistor (2252 Ohm @ 25°C)
RANGE TEMPERATURE ACCURACY*
SPAN RESOLUTION 18˚ TO 28˚F; 1 year
150°F – 40.0° to 150.0° 0.1° 6 (0.3% rdg + 0.3°C)
– 55.0° to – 40.1° 0.1° 6 (0.3% rdg + 0.6°C)
14
Page 18
*ACCURACY: Includes repeatability, temperature
coefficient (18° to 28 °C), time stability (1 year), and linearization conformity errors. Excludes probe errors; however, probe errors around 0°C may be compen­sated by an internal adjustment.
REPEATABILITY: 0.1°C typical for 1 week at constant ambi-
ent temperature.
TEMPERATURE COEFFICIENT: 18° to 28°C; included in accuracy specifica-
tion. From –10° to 18° and 28° to 50° C: less than 6 (0.06% rdg + 0.01°C)/°C.
SENSOR VOLTAGE: 40 mV RMS maximum.
MODEL 865 AND 866 GENERAL SPECIFICATIONS (unless specified): DISPLAY: 3
1
2-digit LCD O.5" (13 mm) height. Polarity
and decimal point indication.
CONVERSION RATE: 1.5 readings per second.
OPEN SENSOR INDICATION: Display reads between –100° and –199°. OVERRANGE INDICATION: 3 least significant digits blanked. MAXIMUM COMMON
MODE VOLTAGE: 42 V peak to earth. COMMON MODE REJECTION
(Model 865): Less than 0.01°F/volt at dc, 50 and 60 Hz.
(2kV unbalance, LO driven)
COMMON MODE REJECTION (Model 866): Less than 0.01°C/volt at dc, 50 and 60 Hz.
(2kV unbalance, LO driven)
ENVIRONMENTAL LIMITS FOR OPERATING (Model 865): 14° to 122°F, less than 80% relative
ENVIRONMENTAL LIMITS FOR OPERATING (Model 866): 10° to 50°C, less than 80% relative
humidity up to 95°F; linearly derate 1.5% R.H./°F from 95° to 122°F.
humidity up to 35°C; linearly derate 3% R.H./°C from 35° to 50°C.
15
Page 19
ENVIRONMENTAL LIMITS FOR STORAGE (Model 865): –30° to 140°F, less than 90% relative
humidity up to 95° F; linearly derate 1.5 R.H./°F from 95° to 140°F.
ENVIRONMENTAL LIMITS FOR STORAGE (Model 866): –35° to 60°C, less than 90% relative
humidity up to 35° C; linearly derate 3% R.H./°C from 35° to 60°C.
THERMISTOR LINEARIZATION: Ratiometric dual-slope A/D with continu-
ous linearization.
INPUT CONNECTION:
1
40 (6 mm) phone jack.
POWER: 9 V alkaline or carbon-zinc (NEDA 1604)
battery.
BATTERY LIFE, CONTINUOUS: 350 hours typical with alkaline battery;
200 hours typical with carbon-zinc battery.
BATTERY INDICATOR: Display indicates “LO BAT” when less
than 10% of life remains.
DIMENSIONS, WEIGHT: H: 6.30 (160 mm) x W: 2.70 (69 mm) D:
1.20 (31 mm) Net weight 7.5 oz. (210 gm)
CONSTRUCTION: Heavy duty ABS plastic housing.
OPTIONAL ACCESSORIES
Any OMEGA thermistors with 2252 ohms are suitable. See the ON series or the THX series in the latest OMEGA Temperature Measurement Handbook.
Model 8660 Tilt/Stand/Belt Clip/Probe Holder—for the Model 865 or 866. Model 8668 Soft Carrying Case is a padded vinyl carrying case that holds one hand held meter, a Model 8660 and a probe.
16
Page 20
6.1 PARTS LIST FOR MODELS 865 AND 866
SCHEMATIC SCHEMATIC
DESIG. DESCRIPTION LOCATION
BA101 9V Alkaline Battery, NEDA 1604 F1 C101 Capacitor, 0.047, 50V Metalized Polypropylene E2 C102 Capacitor, 0.1
C103 Capacitor, 0.1 C104 Capacitor, 0.33 C105 Capacitor, 0.33 C106 Capacitor, 0.33
µ
F, 50V Ceramic F1
µ
F, 50V Ceramic G1
µ
F, 63V Metalized Polyester E2
µ
F, 63V Metalized Polyester D3
µ
F, 63V Metalized Polyester D1 C107 Capacitor, 1000p, 1000V Ceramic Disc. C2 C108 Capacitor, 0.1 C109 Capacitor, 0.33 C110 Capacitor, 0.33 C111 Capacitor, 0.33
µ
F, 50V Ceramic F1
µ
F, 63V Metalized Polyester E3
µ
F, 63V Metalized Polyester E3
µ
F, 63V Metalized Polyester E1 C112 Capacitor, 47pF, 500V Silver Mica or Ceramic E2 DS101 3 J1001
1
2" Digit Liquid Crystal Display G4
1
4" Phone Jack A1 J1002 Battery Clip E1 J1003 Battery Clip F1 Q101 NPN Transistor, MPS8099 C2 R101 Resistor, 910 kilohms, 5%, R102* Resistor, 5.4 kilohms, 0.1%,
1
4W, Composition E2
1
8W, Metal Film D4 R102** Jumper R103 Factory selected, part of 865-600 B-6 R104* Resistor, 4.6 kilohms, 0.1%, R104** Resistor, 10 kilohms, 0.1%, R105* Resistor, 44 kilohms, 0.1%, R105** Resistor, 21.77 kilohms, 0.1%, R106* Resistor, 16.5 kilohms, 1%, R106** Resistor, 6.34 kilohms, 1%,
1
8W, Metal Film D3
1
8W, Metal Film D3
1
8W, Metal Film D3
1
8W, Metal Film D3
1
8W, Metal Film C3
1
8W, Metal Film C3 R107 Thick Film Resistor Network SEV R108 Factory selected, Part of 31410 E4 R109* Resistor, 3.92 kilohms, 1%, R109** Resistor, 3.65 kilohms, 1%,
1
8W, Metal Film F5
1
8W, Metal Film F5 R110 Pot, 200 ohms E4 R111* Pot, 5 kilohms C3 R111** Pot, 2 kilohms C3 R112 Pot, 10 ohms D1 R113 Resistor, 392 kilohms, 1%,
1
8W, Metal Film E2
S101 493T, Slide Switch SEV
17
Page 21
MODEL 865 AND 866 PARTS LIST (continued)
U101 Triple 2 Channel Analog Multiplexer, 4053 SEV
1
U102 3
2 Digit Low Power A/D converter, ICL7136CPL F2 U103 Triple 2 Channel Analog Multiplexer, 4053 SEV U104 Quad Low Power J-FET, LF44CN SEV U105 Current Source, LM334 E4 VR101 6.2V, Zener Diode, 1N753 C1
* Model 865
** Model 866
Figure 6-1 Component Layout
18
Page 22
Page 23
Page 24
WARRANTY/DISCLAIMER
USA
MADE
IN
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, includ­ing but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negli­gence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY
RETURNS, please have the following information available BEFORE contacting OMEGA:
1. P.O. number under which the product was
PURCHASED,
2. Model and serial number of the product
under warranty, and
3. Repair instructions and/or specific
problems relative to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. P.O. number to cover the COST
of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.
Page 25
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
TEMPERATURE
MU
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
MU
Wire: Thermocouple, RTD & Thermistor
MU
Calibrators & Ice Point References
MU
Recorders, Controllers & Process Monitors
MU
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
MU
Transducers & Strain Gauges
MU
Load Cells & Pressure Gauges
MU
Displacement Transducers
MU
Instrumentation & Accessories
FLOW/LEVEL
MU
Rotameters, Gas Mass Flowmeters & Flow Computers
MU
Air Velocity Indicators
MU
Turbine/Paddlewheel Systems
MU
Totalizers & Batch Controllers
pH/CONDUCTIVITY
MU
pH Electrodes, Testers & Accessories
MU
Benchtop/Laboratory Meters
MU
Controllers, Calibrators, Simulators & Pumps
MU
Industrial pH & Conductivity Equipment
DATA ACQUISITION
MU
Data Acquisition & Engineering Software
MU
Communications-Based Acquisition Systems
MU
Plug-in Cards for Apple, IBM & Compatibles
MU
Datalogging Systems
MU
Recorders, Printers & Plotters
HEATERS
MU
Heating Cable
MU
Cartridge & Strip Heaters
MU
Immersion & Band Heaters
MU
Flexible Heaters
MU
Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
MU
Metering & Control Instrumentation
MU
Refractometers
MU
Pumps & Tubing
MU
Air, Soil & Water Monitors
MU
Industrial Water & Wastewater Treatment
MU
pH, Conductivity & Dissolved Oxygen Instruments
M0179/0896
Loading...