BRONCHOSCOPE DISINFECTION GUIDE
EXERA™ BF-160 Series and EVIS™ BF-240 Series Bronchovideoscopes
Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information
WARNING:
This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the
detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that
HIGH-LEVEL DISINFECTION
1
Suction Cleaning Adapter
Fill basin with disinfectant. Attach a suction cleaning adapter (MAJ-222) to the previously cleaned broncho­scope. Immerse completely in disinfectant solution.
2
Attach a syringe to the adapter. Pull the plunger of the syringe to fill the instrument/suction channels and channel cleaning adapter with disinfectant solution.
RINSING
was included with your
3
While immersed, detach the adapter from the bronchoscope. Use a clean, lint-free cloth to wipe away any bub­bles clinging to external surfaces.
Tube to suction
bronchoscope
4
Soak all equipment according to the time and temperature recommended by the disinfectant manufacturer. Use a timer to measure the exact contact time.
when purchased.
5
Connect the suction cleaning adapter and syringe. Inject air to remove disinfectant from the instrument/suction channels, then remove the bronchoscope from the disinfectant.
DRYING
9
6
Disconnect the suction cleaning adapter and immerse the bronchoscope and adapter in sterile water. Using a sterile, lint-free cloth, thoroughly rinse and wipe all surfaces.
10
7
Connect suction cleaning adapter to bronchoscope and to a tube from a suction pump. Aspirate water through channels for 30 seconds. Remove from water and aspirate air through channels for 60 seconds.
To Suction Pump
8
Immerse the bronchoscope's distal end in 70% alcohol and aspirate alcohol for 5 seconds. Remove the distal end from the alcohol and aspirate air for 20 seconds.
Suction Cleaning
Adapter (MAJ-222)
To Suction Pump
Alcohol
TRANSPORTATION
Turn the suction pump off. Hold the control section with the openings pointing down and disconnect the suction cleaning adapter.
NOMENCLATURE
Single-Use Suction Valve (MAJ-209)
Suction Cylinder
Remote Switches
Single-Use Biopsy Valve (MAJ-210)
Instrument/Suction Channel Port
Control Section
UP/DOWN Angulation Control Lever
Boot
Dry the bronchoscope and adapter with an alcohol-moistened cloth. Make sure that the equipment is completely dry prior to storage.
Universal Cord
Insertion Tube
Electrical Connector
Venting Connector
Bending Section
Light Guide
Water-Resistant Cap
(MH-553)
Distal Tip
Hold all parts of the endoscope when carrying. Do not let the ends of the bronchoscope swing freely.
IMPORTANT
•Wear appropriate personal protective   equipment during reprocessing.
•Autoclave all autoclavable accessories.
•Store the bronchoscope hanging vertically   with all valves and caps removed.
•Use a compatible disinfectant and test  it for proper concentration as recom-  mended by the manufacturer.
•Do not use the carrying case to store   "patient-ready" endoscopes. Any broncho-  scope placed in the carrying case must be   reprocessed prior to patient use.
Clean the bronchoscope
meticulously prior to disinfection.
© 2004 Olympus America Inc. / Printed in USA 01/04 – PWC-100016