Olympus EXERA BF-160 Bronchovideoscope, EVIS BF-240 User Manual

To use this chart as a wallchart, cut off the top of the chart as illustrated.
To insert this chart into a binder, punch out the holes and fold the chart as illustrated.
Suction Cleaning Adapter
Tube to suction
To Suction Pump
Suction Cleaning
Adapter (MAJ-222)
To Suction Pump
Clean the bronchoscope
meticulously prior to disinfection.
Single-Use Suction Valve (MAJ-209)
Suction Cylinder
Single-Use Biopsy Valve (MAJ-210)
Remote Switches
Universal Cord
Instrument/Suction Channel Port
Electrical Connector
Water-Resistant Cap
(MH-553)
Venting Connector
Distal Tip
Bending Section
Light Guide
Boot
UP/DOWN Angulation Control Lever
Control Section
Insertion Tube
Alcohol
HIGH-LEVEL DISINFECTION
RINSING
TRANSPORTATION
DRYING
NOMENCLATURE
IMPORTANT
1
2
3
4
5
6
9
10
7
8
© 2004 Olympus America Inc. / Printed in USA 01/04 – PWC-100016
EXERA™ BF-160 Series and EVIS™ BF-240 Series Bronchovideoscopes
BRONCHOSCOPE DISINFECTION GUIDE
Refer to the companion BRONCHO SCOPE CLEANING G UIDE for detailed cleaning info rmation
WARNING:
This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the
detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that
was included with your
bronchoscope
when purchased.
Attach a syringe to the adapter. Pull the plunger of the syringe to fill the instrument/suction channels and channel cleaning adapter with disinfectant solution.
Fill basin with disinfectant. Attach a suction cleaning adapter (MAJ-222) to the previously cleaned broncho­scope. Immerse completely in disinfectant solution.
While immersed, detach the adapter from the bronchoscope. Use a clean, lint-free cloth to wipe away any bub­bles clinging to external surfaces.
Soak all equipment according to the time and temperature recommended by the disinfectant manufacturer. Use a timer to measure the exact contact time.
Connect the suction cleaning adapter and syringe. Inject air to remove disinfectant from the instrument/suction channels, then remove the bronchoscope from the disinfectant.
Disconnect the suction cleaning adapter and immerse the bronchoscope and adapter in sterile water. Using a sterile, lint-free cloth, thoroughly rinse and wipe all surfaces.
Connect suction cleaning adapter to bronchoscope and to a tube from a suction pump. Aspirate water through channels for 30 seconds. Remove from water and aspirate air through channels for 60 seconds.
Turn the suction pump off. Hold the control section with the openings pointing down and disconnect the suction cleaning adapter.
Dry the bronchoscope and adapter with an alcohol-moistened cloth. Make sure that the equipment is completely dry prior to storage.
Hold all parts of the endoscope when carrying. Do not let the ends of the bronchoscope swing freely.
•Wear appropriate personal protective   equipment during reprocessing.
•Autoclave all autoclavable accessories.
•Store the bronchoscope hanging vertically   with all valves and caps removed.
•Use a compatible disinfectant and test  it for proper concentration as recom-  mended by the manufacturer.
•Do not use the carrying case to store   "patient-ready" endoscopes. Any broncho-  scope placed in the carrying case must be   reprocessed prior to patient use.
Immerse the bronchoscope's distal end in 70% alcohol and aspirate alcohol for 5 seconds. Remove the distal end from the alcohol and aspirate air for 20 seconds.