Analyzer: Olympus AU 400 - 600 - 640
Application: ACID PHOSPHATASE - Codes AC 0120 TC
Preparation: POWDER SINGLE REAGENT. FOLLOW PREPARATION PROGRAM INDICATED IN INSERT
SHEET. INSTALL AS R1
Storage: REFRIGERATE AT 2-8°C
TEST NAME: ACP
Sample: Volume 20 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 0 ml Dilution 0 ml
Wavelenght: Pri. 410 Sec. 800 Min OD Max OD
Method: RATE L - 0.1 H 2.5
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.3
Last L - 0.1 First H 0.3
Measuring point 1: First 12 Last 27
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: NO L 1 H 60
Correlation factor:
A 1 B 0
On-board stability period: 30
Unit: U/L
Factor: 853
IUS.422.0.2 Directions for use Edition 2010-08-01 page 4 of 41
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: ALBUMIN - Codes BC 0100 / 0500 / 1000 / 1500 CH
Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT
Storage: REFRIGERATE (2-8°C)
Stability: AS INDICATED IN THE LABEL
TEST NAME: ALB
Sample: Volume 2 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 295 ml Dilution 0 ml R2 volume 0 ml Dilution 0 ml
Wavelenght: Pri. 600 Sec. 800 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.5
Last L - 0.1 First H 0.5
Measuring point 1: First 0 Last 4
Measuring point 2: First Last
Linearity: Dynamic range:
No Lag Time: YES L 0.1 H 6.0
Correlation factor:
A 1 B 0
On-board stability period: 90
Unit: g/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 5 of 41
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: ALP
Sample: Volume 5 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 410 Sec. 480 Min OD Max OD
Method: RATE L -0.1 H 2.5
Reaction slope: + Reagent OD limit:
First L -0.1 First H 1.5
Last L -0.1 First H 1.5
Measuring point 1: First 13 Last 19
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: NO L 5 H 3000
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: U/L
Factor: 2757
IUS.422.0.2 Directions for use Edition 2010-08-01 page 6 of 41
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: ALP
Sample: Volume 5 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 410 Sec. 480 Min OD Max OD
Method: RATE L - 0.1 H 2.0
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 1.2
Last L - 0.1 First H 1.2
Measuring point 1: First 20 Last 24
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: NO L 7 H 3000
Correlation factor:
A 1 B 0
On-board stability period: 30
Unit: U/L
Factor: 2757
IUS.422.0.2 Directions for use Edition 2010-08-01 page 7 of 41
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: AMYLASE FL - Codes AM F060 / F120 / F245 CHPreparation: R1 - LIQUID READY TO USE SINGLE REAGENT
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: AMY
Sample: Volume 6 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 240 ml Dilution 0 ml R2 volume 0 ml Dilution 0 ml
Wavelenght: Pri. 410 Sec. 480 Min OD Max OD
Method: RATE L - 0.1 H 2.0
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.3
Last L - 0.1 First H 0.3
Measuring point 1: First 5 Last 10
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: NO L 10 H 3000
Correlation factor:
A 1 B 0
On-board stability period: 30
Unit: U/L
Factor: 3178
IUS.422.0.2 Directions for use Edition 2010-08-01 page 8 of 41
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: AMY
Sample: Volume 8 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 192 ml Dilution 0 ml R2 volume 48 ml Dilution 0 ml
Wavelenght: Pri. 410 Sec. 480 Min OD Max OD
Method: RATE L - 0.1 H 2.0
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.3
Last L - 0.1 First H 0.3
Measuring point 1: First 20 Last 27
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: NO L 6 H 1500
Correlation factor:
A 1 B 0
On-board stability period: 30
Unit: U/L
Factor: 3480
IUS.422.0.2 Directions for use Edition 2010-08-01 page 9 of 41
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: AMYP
Sample: Volume 5 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 410 Sec. 480 Min OD Max OD
Method: RATE L - 0.1 H 2.0
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.3
Last L - 0.1 First H 0.3
Measuring point 1: First 20 Last 27
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: NO L 2 H 2500
Correlation factor:
A 1 B 0
On-board stability period: 30
Unit: U/L
Factor: 3480
IUS.422.0.2 Directions for use Edition 2010-08-01 page 10 of 4110 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: BILT
Sample: Volume 10 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 520 Sec. 800 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.5
Last L - 0.1 First H 0.5
Measuring point 1: First Last 27
Measuring point 2: First Last 10
Linearity: Dynamic range:
No Lag Time: L 0.05 H 20
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 12 of 4112 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: BILD
Sample: Volume 10 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 540 Sec. 660 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.5
Last L - 0.1 First H 0.5
Measuring point 1: First Last 27
Measuring point 2: First Last 10
Linearity: Dynamic range:
No Lag Time: L 0.05 H 20
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 13 of 4113 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: CALCIUM - Code CA 0305 / 0505 CH
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT B
R2 - INSTALL LIQUID READY TO USE REAGENT A
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: CA
Sample: Volume 5 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 145 ml Dilution 0 ml R2 volume 145 ml Dilution 0 ml
Wavelenght: Pri. 570 Sec. 660 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.5
Last L - 0.1 First H 1.2
Measuring point 1: First Last 27
Measuring point 2: First Last 10
Linearity: Dynamic range:
No Lag Time: L 0.1 H 20
Correlation factor:
A 1 B 0
On-board stability period: 30
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 14 of 4114 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: CALCIUM ASX - Code CA 0100 / 0500 CH
Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT
Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C)
Stability: AS INDICATED IN THE LABEL
TEST NAME: CA
Sample: Volume 3 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 295 ml Dilution 0 ml R2 volume 0 ml Dilution 0 ml
Wavelenght: Pri. 660 Sec. 700 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 2.0
Last L - 0.1 First H 2.0
Measuring point 1: First Last 10
Measuring point 2: First Last
Linearity: Dynamic range:
No Lag Time: L 0.2 H 20
Correlation factor:
A 1 B 0
On-board stability period: 90
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 15 of 4115 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: CHOL
Sample: Volume 3 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 295 ml Dilution 0 ml R2 volume 0 ml Dilution 0 ml
Wavelenght: Pri. 520 Sec. 700 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.5
Last L - 0.1 First H 0.5
Measuring point 1: First Last 27
Measuring point 2: First Last
Linearity: Dynamic range:
No Lag Time: L 1 H 700
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 16 of 4116 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: HDL direct FL- Code HD F080 / F245 / F400 CH
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: 60 DAYS OPEN BOTTLES
TEST NAME: HDLC
Sample: Volume 2 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 180 ml Dilution 0 ml R2 volume 60 ml Dilution 0 ml
Wavelenght: Pri. 600 Sec. 700 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.3
Last L - 0.1 First H 0.3
Measuring point 1: First 0 Last 27
Measuring point 2: First 0 Last 10
Linearity: Dynamic range:
No Lag Time: L 1 H 220
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 17 of 4117 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: LDL direct FL- Code DL F080 CH
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: 60 DAYS OPEN BOTTLES
TEST NAME: LDLC
Sample: Volume 2 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 180 ml Dilution 0 ml R2 volume 60 ml Dilution 0 ml
Wavelenght: Pri. 600 Sec. 700 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.3
Last L - 0.1 First H 0.3
Measuring point 1: First 0 Last 27
Measuring point 2: First 0 Last 10
Linearity: Dynamic range:
No Lag Time: L 1 H 400
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 18 of 4118 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: CHE
Sample: Volume 4 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 40 ml Dilution 0 ml
Wavelenght: Pri. 410 Sec. 480 Min OD Max OD
Method: RATE L 0.4 H 2.5
Reaction slope: - Reagent OD limit:
First L 1.1 First H 2.2
Last L 1.1 First H 2.2
Measuring point 1: First 16 Last 20
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: NO L 0.4 H 25*
Correlation factor:
A 1 B 0
On-board stability period: 30
Unit: U/L
Factor: 66*
* NOTE: results in kU/L
IUS.422.0.2 Directions for use Edition 2010-08-01 page 19 of 4119 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: CK
Sample: Volume 10 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 340 Sec. 660 Min OD Max OD
Method: RATE L -0.1 H 2.0
Reaction slope: + Reagent OD limit:
First L -0.1 First H 0.5
Last L -0.1 First H 0.5
Measuring point 1: First 21 Last 27
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: NO L 1 H 1000
Correlation factor:
A 1 B 0
On-board stability period: 30
Unit: U/L
Factor: 4127
IUS.422.0.2 Directions for use Edition 2010-08-01 page 20 of 4120 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: CK-MB FL - Codes MB F060 / F120 CH
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: CKMB
Sample: Volume 10 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 340 Sec. 660 Min OD Max OD
Method: RATE L -0.1 H 2.0
Reaction slope: + Reagent OD limit:
First L -0.1 First H 0.5
Last L -0.1 First H 0.5
Measuring point 1: First 21 Last 27
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: NO L 2 H 1000
Correlation factor:
A 1 B 0
On-board stability period: 30
Unit: U/L
Factor: 8254
IUS.422.0.2 Directions for use Edition 2010-08-01 page 21 of 4121 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: COPPER - Code CU 0100 CH
Preparation: R1 - MIX REAGENT A + REAGENT B AS INDICATED IN INSERT SHEET
Storage: ROOM TEMPERATURE - DO NOT REFRIGERATE IF POSSIBLE
Stability: 2 WEEKS
TEST NAME: CU
Sample: Volume 20 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 295 ml Dilution 0 ml R2 volume 0 ml Dilution 0 ml
Wavelenght: Pri. 600 Sec. 700 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.5
Last L - 0.1 First H 0.5
Measuring point 1: First Last 27
Measuring point 2: First Last
Linearity: Dynamic range:
No Lag Time: L 4 H 500
Correlation factor:
A 1 B 0
On-board stability period: 10
Unit: microg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 22 of 4122 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT B
R2 - INSTALL LIQUID READY TO USE REAGENT A
Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C)
Stability: AS INDICATED IN THE LABEL
TEST NAME: CREA
Sample: Volume 20 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 145 ml Dilution 0 ml R2 volume 145 ml Dilution 0 ml
Wavelenght: Pri. 520 Sec. 800 Min OD Max OD
Method: FIXED L - 0.1 H 2.5
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.5
Last L - 0.1 First H 0.5
Measuring point 1: First 20 Last 24
Measuring point 2: First Last
Linearity: Dynamic range:
No Lag Time: L 0.2 H 20
Correlation factor:
A 1 B 0
On-board stability period: 14
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 23 of 4123 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: GGT
Sample: Volume 25 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 410 Sec. 480 Min OD Max OD
Method: RATE L -0.1 H 2.0
Reaction slope: + Reagent OD limit:
First L -0.1 First H 1.5
Last L -0.1 First H 1.5
Measuring point 1: First 14 Last 23
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: NO L 2 H 800
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: U/L
Factor: 1571
IUS.422.0.2 Directions for use Edition 2010-08-01 page 24 of 4124 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: GLUC
Sample: Volume 3 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 295 ml Dilution 0 ml R2 volume 0 ml Dilution 0 ml
Wavelenght: Pri. 520 Sec. 700 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.5
Last L - 0.1 First H 0.5
Measuring point 1: First Last 27
Measuring point 2: First Last
Linearity: Dynamic range:
No Lag Time: L 1 H 500
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 25 of 4125 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: GLUCOSE UV FL - Code GL F601 CH
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: GLUC
Sample: Volume 2.5 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 340 Sec. 380 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.5
Last L - 0.1 First H 0.5
Measuring point 1: First Last 27
Measuring point 2: First Last 10
Linearity: Dynamic range:
No Lag Time: L 1 H 700
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 26 of 4126 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: GOT
Sample: Volume 25 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 340 Sec. 660 Min OD Max OD
Method: RATE L 0.7 H 2.5
Reaction slope: - Reagent OD limit:
First L 1.0 First H 2.5
Last L 1.0 First H 2.5
Measuring point 1: First 14 Last 27
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: YES L 2 H 440
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: U/L
Factor: 1746
IUS.422.0.2 Directions for use Edition 2010-08-01 page 27 of 4127 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: GPT
Sample: Volume 25 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 340 Sec. 660 Min OD Max OD
Method: RATE L 0.7 H 2.5
Reaction slope: - Reagent OD limit:
First L 1.0 First H 2.5
Last L 1.0 First H 2.5
Measuring point 1: First 14 Last 27
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: YES L 2 H 440
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: U/L
Factor: 1746
IUS.422.0.2 Directions for use Edition 2010-08-01 page 28 of 4128 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: IRON FZ - Codes FE F245 / F400 CH
Preparation: R1 - INSTALL REAGENT A AFTER PREPARATION
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: FE
Sample: Volume 50 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 570 Sec. 800 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.3
Last L - 0.1 First H 0.3
Measuring point 1: First 0 Last 12
Measuring point 2: First 0 Last 10
Linearity: Dynamic range:
No Lag Time: L 25 H 1000
Correlation factor:
A 1 B 0
On-board stability period: 30
Unit: microg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 29 of 4129 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: LDH
Sample: Volume 4 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 340 Sec. 380 Min OD Max OD
Method: RATE L 0.7 H 2.5
Reaction slope: - Reagent OD limit:
First L 1.0 First H 2.5
Last L 1.0 First H 2.5
Measuring point 1: First 12 Last 27
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: YES L 30 H 2000
Correlation factor:
A 1 B 0
On-board stability period: 30
Unit: U/L
Factor: 8015
IUS.422.0.2 Directions for use Edition 2010-08-01 page 30 of 4130 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: LIPASE FL - Code LP F080 CH
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: ALP
Sample: Volume 4 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 570 Sec. 800 Min OD Max OD
Method: RATE L -0.1 H 2.5
Reaction slope: + Reagent OD limit:
First L -0.1 First H 0.8
Last L -0.1 First H 0.8
Measuring point 1: First 13 Last 19
Measuring point 2: First Last
Linearity: 20% Dynamic range:
No Lag Time: NO L 3 H 300
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: U/L
IUS.422.0.2 Directions for use Edition 2010-08-01 page 31 of 4131 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: MAGNESIUM - Codes MG 0200 / 0500 CH
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT B
R2 - INSTALL LIQUID READY TO USE REAGENT A
Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C)
Stability: AS INDICATED IN THE LABEL
TEST NAME: MG
Sample: Volume 3 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 145 ml Dilution 0 ml R2 volume 145 ml Dilution 0 ml
Wavelenght: Pri. 520 Sec. 800 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L 0.3 First H 2.0
Last L 0.3 First H 2.0
Measuring point 1: First Last 27
Measuring point 2: First Last 10
Linearity: Dynamic range:
No Lag Time: L 0.1 H 8
Correlation factor:
A 1 B 0
On-board stability period: 14
Unit: mEq/l
IUS.422.0.2 Directions for use Edition 2010-08-01 page 32 of 4132 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: UREA
Sample: Volume 3 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 340 Sec. 660 Min OD Max OD
Method: RATE L 1.0 H 2.5
Reaction slope: - Reagent OD limit:
First L 1.2 First H 2.5
Last L 1.2 First H 2.5
Measuring point 1: First 12 Last 18
Measuring point 2: First Last
Linearity: 25% Dynamic range:
No Lag Time: YES L 1 H 300
Correlation factor:
A 1 B 0
On-board stability period: 60
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 38 of 4138 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: URIC ACID T FL - Code AU F402 CH
Preparation: R1 - MIX REAGENT A + REAGENT B AS INDICATED IN INSERT SHEET
Storage: REFRIGERATE AT 2-8°C
Stability: 90 DAYS
TEST NAME: UA
Sample: Volume 10 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 295 ml Dilution 0 ml R2 volume 0 ml Dilution 0 ml
Wavelenght: Pri. 540 Sec. 700 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.5
Last L - 0.1 First H 0.5
Measuring point 1: First Last 27
Measuring point 2: First Last
Linearity: Dynamic range:
No Lag Time: L 0.2 H 25
Correlation factor:
A 1 B 0
On-board stability period: 90
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 39 of 4139 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
Storage: REFRIGERATE AT 2-8°C
Stability: AS INDICATED IN THE LABEL
TEST NAME: UA
Sample: Volume 10 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 200 ml Dilution 0 ml R2 volume 50 ml Dilution 0 ml
Wavelenght: Pri. 540 Sec. 700 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.5
Last L - 0.1 First H 0.5
Measuring point 1: First Last 27
Measuring point 2: First Last 10
Linearity: Dynamic range:
No Lag Time: L 0.1 H 35
Correlation factor:
A 1 B 0
On-board stability period: 90
Unit: mg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 40 of 4140 of 4141
Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640
Analyzer: Olympus AU 400 - 600 - 640
Application: ZINC - Code ZN 0125 CH
Preparation: R1 - MIX REAGENT A + REAGENT B AS INDICATED IN INSERT SHEET
Storage: REFRIGERATE AT 2-8°C
Stability: 2 WEEKS
TEST NAME: ZN
Sample: Volume 15 ml Dilution 0 ml Pre-dilution Rate: 1
Reagents: R1 volume 295 ml Dilution 0 ml R2 volume 0 ml Dilution 0 ml
Wavelenght: Pri. 570 Sec. 800 Min OD Max OD
Method: END L H
Reaction slope: + Reagent OD limit:
First L - 0.1 First H 0.8
Last L - 0.1 First H 0.8
Measuring point 1: First Last 27
Measuring point 2: First Last
Linearity: Dynamic range:
No Lag Time: L 5 H 1000
Correlation factor:
A 1 B 0
On-board stability period: 10
Unit: microg/dl
IUS.422.0.2 Directions for use Edition 2010-08-01 page 41 of 4141 of 4141