THIS MANUAL MUST BE READ CAREFULLY PRIOR TO THE OPERATION OF
THIS DEVICE. THIS DEVICE WILL PERFORM AS DESIGNED ONLY IF USED IN
ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONS. IMPROPER USE
MAY CAUSE THE DEVICE TO FAIL TO PERFORM AS DESIGNED AND MAY CAUSE
INJURY TO THE PATIENT AND/OR HEALTHCARE PROFESSIONAL.
W WARNING
®
The warranties made by Ohio Medical Corporation
not installed, used and serviced in accordance with the instructions in this manual. Please protect yourself and
your patients by following them. We encourage our customers to write or call regarding this equipment prior to
use or for any additional information relative to use or repairs.
This product is not intended as a life-sustaining or life-supporting device.
FEDERAL (USA) LAW RESTRICTS THIS DEVICE TO SELL BY OR ON THE ORDER OF A PHYSICIAN
Ohio Medical Corporation® | 1111 Lakeside Drive | Gurnee, IL 60031 USA
with respect to these products are voided if the products are
P/N 814358 [Rev.1] 09/2010
P/N 814358 [Rev.1] 09/2010
Ohio Medical Corporation
®
Medical Instrument Warranty
This product is sold by Ohio MedicalTM under the warranties set forth in the following paragraphs. Such warranties
are extended only with respect to the purchase of this product directly from Ohio Medical or Ohio Medical’s
Authorized Dealers as new merchandise and are extended to the rst Buyer thereof, other than for purpose of
resale.
For a period of twelve (12) months from the date of original delivery to Buyer, to Buyer’s order, or to an Ohio
Medical Authorized Dealer, this product, other than its expendable parts, is warranted to be free from functional
defects in materials and workmanship and to conform to the description of the product contained in the operation
manual and accompanying labels and/or inserts, provided that the same is properly operated under conditions of
normal use, that regular periodic maintenance and service is performed and that replacements and repairs are
made in accordance with the instructions provided. This same warranty is made for a period of sixty (60) days
with respect to the expendable parts. The foregoing warranties shall not apply if the product has been repaired
other than by Ohio Medical or in accordance with written instructions provided by Ohio Medical, or altered by
anyone other than Ohio Medical, or if the product has been subject to abuse, misuse, negligence, or accident.
Ohio Medical’s sole and exclusive obligation and Buyer’s sole and exclusive remedy under the above warranties
is limited to repairing or replacing, free of charge, at Ohio Medical’s option, a product, which is telephonically
reported to the nearest Ohio Medical Regional Service Ofce and which, if so advised by Ohio Medical, is
thereafter returned with a statement of the observed deciency, not later than seven (7) days after the expiration
date of the applicable warranty,to the designated Ohio Medical Service Ofce during normal business hours,
transportation charges prepaid, and which, upon Ohio Medical’s examination, is found not to conform with
the above warranties. Ohio Medical shall not be otherwise liable for any damages including, but not limited to
incidental damages, consequential damages, or special damages.
There are no express or implied warranties which extend beyond the warranties herein above set forth. Ohio
Medical makes no warranty of merchantability or tness for a particular purpose with respect to the product or
parts thereof.
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor General WARNINGS and CAUTIONS
General Warnings and Cautions
W WARNING
1. The MiniOX® 3000 Oxygen Monitor will perform
to specications only if it is used and serviced in
accordance with the manufacturer’s instructions.
This instrument is to be used only by qualied,
trained personnel who have carefully read the
operating manual and labels and who have observed
the information set forth. If this instrument does not
perform as described in this manual, the instrument
must not be used until the condition is rectied.
2. The MiniOX
®
3000 Oxygen Monitor must be calibrated
prior to each use. A two point calibration check must
be performed weekly. See Section 3, Operation. If
the instrument cannot be calibrated, the sensor must
be replaced. If the instrument is still unable to be
calibrated, the instrument must be serviced.
3. The oxygen sensor has a minimal response to
certain gases other than oxygen. Be aware of these
gases and their interference levels. See Appendix C,
Interferent Gases and Vapors.
4. The oxygen sensor is affected by changes in
barometric pressure. See Appendix B, Effects of
Pressure, Humidity and Temperature.
5. The sensor is a sealed unit containing a potassium
hydroxide electrolyte, which is caustic. If the sensor
should develop a leak and the contents come
in contact with skin or clothing, rinse area with
large quantities of water. In case of eye contact,
immediately ush eyes for at least 15 minutes,
holding eyes open. Call a physician. Dispose of a
leaking sensor immediately; it must be disposed of
in accordance with all applicable federal, state, and
local regulations.
6. Ensure a tight t exists between the sensor and the tee
adapter. Ohio Medical tee adapters are engineered
to t securely with Ohio Medical sensors. However,
the sensor retaining strap must be used to prevent
accidental separation of sensor and tee adapter.
7. Never operate the MiniOX
®
3000 Oxygen Monitor if it
is suspected that water or other liquids have entered
into the case. If this occurs, immediately turn the unit
OFF and contact your nearest Ohio Medical Service
Center for additional information.
8. Use of devices generating or emitting electromagnetic
radiation near the MiniOX
®
3000 Oxygen Monitor may
interfere with the proper operation of the product,
causing it to fail to perform as designed. Particularly,
the electromagnetic radiation from the interfering
device may cause the product to display incorrect/
erratic values or to stop operating. Special attention
should be paid to the patient if this occurs.
9. Never use the MiniOX
®
3000 Oxygen Monitor in
combustible atmosphere, such as occurs with
ammable anesthetics. Such use could result in
ignition of atmosphere.
10. Never allow an excess length of cable near the
patient’s head or neck, as such could result in
strangulation. Secure excess cable to the bed rail or
other suitable object.
FAILURE TO COMPLY WITH THESE WARNINGS
COULD JEOPARDIZE THE WELL-BEING OF THE
PATIENT AND/OR HEALTH CARE PROFESSIONAL.
W CAUTION
1. Remove internal battery if unit is to be non-operational
for extended periods of time.
2. Observe polarity when inserting a new battery.
Incorrect connection may cause damage to the
instrument.
3. The MiniOX
®
3000 Oxygen Monitor must never be
immersed in any cleaning solution, autoclaved, or
exposed to temperatures greater than 70oC.
4. Use only genuine Ohio Medical Corporation
accessories and replacement parts. Failure to do
so may seriously impair the monitor’s performance.
Repair or alteration of the MiniOX
®
3000 Oxygen
Monitor beyond the scope of the maintenance
instructions or by anyone other than an authorized
Ohio Medical Corporation service person could
WC-1
P/N 814358 [Rev.1] 09/2010
cause the product to fail to perform as designed.
5. Improper mounting of the sensor in a breathing circuit
may result in inaccurate readings. The sensor MUST
be mounted with the deector pointing downward to
prevent moisture collection on the sensor membrane
(see Figures 3-3 and 3-4 in Section 3, Operation).
Install the sensor upstream from the humidier to
minimize its exposure to moisture.
6. Never use a MiniOX
®
3000 Oxygen Monitor with a
cable that appears worn, cracked, or has damaged
insulation.
FAILURE TO COMPLY WITH THESE CAUTIONS
COULD RESULT IN INSTRUMENT DAMAGE AND/OR
FAILURE OF UNIT TO PERFORM TO SPECIFICATIONS.
MiniOX® 3000 Oxygen Monitor General WARNINGS and CAUTIONS
Denition of Warnings and Cautions
W WARNING
W CAUTION
NOTE:
Statement citing a potential safety hazard and
possible injury to yourself or others.
Statement citing a possibility of damage to the
instrument or other property.
The monitor is for use by trained health care
professionals:
• under the supervision, or on the order, of a
physician
• in a hospital or clinical setting
• during emergency transport
General Description
• fall below the preset (or default) low alarm setting
• rise above the preset (or default) high alarm
setting
®
The MiniOX
3000 Oxygen Monitor:
• detects low and depleted battery conditions
• activates audible and visual alarms
• alarms for:
• sensor disconnection or malfunction
• various internal operating errors
®
The MiniOX
3000 Oxygen Monitor conducts self-
checks:
• at power-up (battery installation)
• at turn on
• during operation
Additionally, the monitor has two operator-initiated
test functions:
Battery-operated and microprocessor-controlled,
®
the MiniOX
3000 Oxygen Monitor measures
oxygen concentrations in the 0% to 100% range.
The monitor’s performance features ensure reliable
and accurate oxygen measurement. These features
include:
• calibration function
• high and low oxygen concentration alarms
• low and depleted battery alarms
• oxygen sensor indicator
• automatic error detection
• battery test
• oxygen alarm test
The calibration function allows calibration of
the device against room air (dened as oxygen
concentration of 20.8%) or 100% O
. Audible and
2
visual alarms alert the operator when monitor
calibration is required.
• The Alarm Test veries the operation of the high
and low oxygen level alarms
• The Battery Test assesses the relative remaining
battery life
®
The MiniOX
3000 Oxygen Monitor consists of two
components: the instrument and the oxygen sensor.
The front of the hand-held instrument features:
• a touch-sensitive keypad
• a liquid crystal display (LCD) that shows:
• monitor status
• continuous oxygen concentrations
• preset alarm levels
• two red light-emitting diodes (LEDs) which serve
as visual alarms
The back of the instrument case features:
High and low oxygen concentration alarms may be
set in the ranges of:
• 16% to 100% (high alarm) and
• 15% to 99% (low alarm)
or the default high/low settings may be used (50%
and 18%, respectively). Audible and visual alarms
activate when oxygen concentrations:
1-1
P/N 814358 [Rev.1] 09/2010
• a bail bar to allow the instrument to “stand” on a
horizontal surface during monitoring operations
• a plastic wedge that slides into an optional
bracket for mounting the instrument on a
horizontal or vertical pole
The galvanic oxygen sensor consists of a deector
assembly and a plastic housing containing two
electrodes. A coiled cable connects the sensor to the
instrument. Plugs at each end of the cable snap into
jacks (one located in the sensor housing and one
located in the instrument) and are held securely in
place by twist collars.
The oxygen sensor is introduced into a breathing
circuit through an Ohio Medical Corporation tee
adapter connecting two lengths of tubing. The
sensor/tee adapter assembly is positioned with the
sensor deector pointing downward to ensure that
moisture does not collect on the sensor membrane.
The retaining strap ensures that the sensor remains
securely in place in the tee adapter.
See the following appendices:
• Appendix A, Specications
• Appendix F, Accessories/Replacement Parts
Operating Principles
MiniOX® 3000 Oxygen Sensor
The oxygen sensor includes two electrodes:
• provides the operator interface
• controls internal functions
• monitors data and system status
Operator Interface
When the operator enters commands using the
keypad, the microprocessor interprets these
commands and responds by:
• displaying text in the LCD
• activating the LEDs
• emitting audible tones
Internal Functions
During operation, the microprocessor coordinates all
internal functions including:
• self-testing
• response to keypad commands
• data collection
• display update
• a gold cathode exposed to the atmosphere
through a uoropolymer membrane
• a lead anode submersed in a potassium
hydroxide solution
When oxygen diffuses through the membrane, the
electrochemical reduction of oxygen on the cathode
and the corresponding oxidation on the anode
generate an electrical current proportional to the
partial pressure of oxygen in the sample atmosphere.
The instrument temperature compensates, amplies,
and converts the electrical current, displaying O
2
values.
The Ohio Medical Corporation oxygen sensor is selfzeroing; when no oxygen is present:
• no current is produced
• 0% oxygen is displayed
The sensor has a minimal response to gases other
than oxygen (see Appendix C, Interferent Gases and
Vapors).
• conrmation that a programmed sequence is
properly followed
Data and System Status
The signal from the oxygen sensor is amplied
and converted to a digital value through electronic
circuitry. Following software compensation, the
microprocessor:
• compares current data to preselected values
• displays the updated O
concentration
2
• if appropriate, activates audible and visual alarms
The microprocessor also monitors internal systems,
indicating such operating conditions as:
• calibration required
• low battery voltage
• sensor disconnect
• system error
®
MiniOX
3000 Instrument
The MiniOX® 3000 Oxygen Monitor is batterypowered and microprocessor-controlled. The
microprocessor:
Performance Features
Calibration
The MiniOX® 3000 Oxygen Monitor must be
calibrated:
• when environmental conditions (pressure,
humidity and temperature) change
When a MiniOX
®
3000 Oxygen Monitor with a sensor
connected is turned ON, the monitor performs a selftest and a calibration requirement occurs:
• “CAL” ashes on the display, telling the operator
to carry out the calibration procedure
The calibration procedure is initiated by exposing the
sensor to a calibration gas [either room air (dened
as oxygen concentration of 20.8%) or 100% oxygen]
and pressing the corresponding key on the keypad.
The following appears on the display:
• “CAL LOCKED”
• the selected calibration gas concentration.
When the unlock key is pressed the monitor
displays:
• “CAL”
• a 10-segment bar graph which “counts down” two
seconds per bar for 20 seconds
The monitor automatically calibrates to the selected
concentration. At the end of the 20 seconds, the
monitor:
• an audible alarm activates
• the measured concentration appears in the
display
The operator can silence the audible alarm for
three 30-second intervals for a total of 90 seconds;
however, the visual alarm continues to ash. At
the end of the silence period, the audible alarm
reactivates if the alarm condition is not corrected. See
Section 3, Silencing an Alarm
Low Battery Alarms
®
The MiniOX
3000 Oxygen Monitor features a twostage battery alarm that warns of depleted and
expired battery voltage:
• The rst alarm alerts the operator that the monitor
has approximately six hours of operating time
remaining:
• a warning message appears in the display
• an audible alarm sounds at 30-second
intervals
• If the operator does not replace the battery after
this alarm, a second low battery alarm activates
when the battery is no longer able to support
monitoring. The monitor:
• displays a warning message
• activates an audible and visual alarm (See
Section 3, Low Battery Alarms)
• emits a beep indicating that calibration is
complete
• enters the operating mode
See Section 3, Calibration
Low/High Oxygen Concentration Alarms
®
The MiniOX
3000 Oxygen Monitor has audible
and visual alarms that activate when oxygen
concentrations exceed preset low or high alarm
settings. Default settings are 18% and 50%,
respectively; however, the operator may select alarm
levels between 15% and 100%. See Section 3,
Setting the Alarms
When the MiniOX® 3000 Oxygen Monitor detects an
oxygen concentration that exceeds the preset alarm
limit:
• the red LED for that alarm ashes
1-3
P/N 814358 [Rev.1] 09/2010
Sensor Indicator
®
During monitoring, the MiniOX
3000 Oxygen Monitor
displays a warning message; audible and visual
alarms activate if:
• the oxygen sensor becomes disconnected
• the cable fails or disconnects from the sensor or
If a calibration is performed using a calibration
gas other than room air or 100% oxygen, or if the
microprocessor detects a calibration error during
normal operation:
• the display ashes a warning message: “CAL”
and “ERR” ve times; then, “CAL” ashes
• audible and visual alarms activate, alerting
the operator that recalibration is required. See
Section 3, Calibration
Error Handling
®
The MiniOX
3000 Oxygen Monitor performs a self-
test
• when it is turned ON
• after a battery is installed (power-up)
• during routine monitoring
If the microprocessor detects an error during these
routine self-tests or during monitoring,
Test Functions
In addition to alarms that alert the operator to low and
®
expired battery voltage, the MiniOX
3000 Oxygen
Monitor features:
• a keypad function that allows the operator to
check the relative battery life at any time. See
Section 3, Test Functions
• a test function that allows the operator to ensure
that the high and low alarms activate at the preset
alarm levels. See Section 3, Test Functions
Radio Frequency/Electromagnetic Insensitivity
®
The MiniOX
3000 Oxygen Monitor is designed to be
insensitive to radio frequency interference (RFI) and
electromagnetic interference (EMI). However, if the
monitor comes in close contact with an RF transmitter
or local electrical disturbance, an erratic reading may
occur.
• the display ashes:
• “ERR”
• an error code
• audible and visual alarms activate
• the monitor ceases to operate until the
appropriate service is performed. See Appendix
E, Error Codes
To reset the monitor, press SET or turn the monitor
OFF, then ON (press I/0 twice). For further details,
see Appendix E, Error Codes
P/N 814358 [Rev.1] 09/2010
1-4
MiniOX® 3000 Oxygen Monitor Section 2, Setting Up the Monitor
Section 2
®
Setting Up the MiniOX
3000 Oxygen
Monitor
The MiniOX® 3000 Oxygen Monitor is packaged in
a shipping carton suitable for instrument storage. If
there is obvious damage to the shipping carton or its
contents, contact Ohio Medical Corporation Repair at:
866-549-6446, Option #1
To set up the monitor:
1. Verify that you have the following (see Figure
2-1):
®
• MiniOX
• Oxygen sensor and deector in sealed
package
3000 Oxygen Monitor
• 10-foot coiled cable with twist collar
• Tee adapter
• Sensor retaining strap
• 9-volt alkaline battery
®
• MiniOX
3000 Operation Manual CD
2. Check the sensor’s manufacturing date on the
sensor package. If the sensor is put into service
within six months after this date, the sensor will
meet performance specications.
3. Remove the sensor from the sealed package and
attach it to the coiled cable.
a. Firmly press the connector until it snaps into
place; tighten the twist collar.
b. Insert the opposite end of the coiled cable into
the jack on the side panel of the instrument;
tighten the twist collar.
4. Remove the deector from the package.
SENSOR
RETAINING STRAP
TEE
ADAPTER
MINIOX® 3000
OXYGEN MONITOR
OXYGEN SENSOR AND DEFLECTOR
IN SEALED CONTAINER
9-VOLT
ALKALINE
BATTERY
TWIST
COLLAR
10 FT
COILED
CABLE
2-1
Figure 2-1
MiniOX® 3000 Oxygen Monitor
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Section 2, Setting Up the Monitor
a. Insert the gasket into the open end of the
deector, ensuring that the gasket is properly
seated within the deector
b. Gently screw the deector onto the sensor
(For monitoring in a breathing circuit, see
Section 3, Installing the Sensor in a Breathing
Circuit)
5. Unscrew the two battery cover screws on the back
of the instrument.
a. Remove the cover
b. Install the battery (See Section 4, Battery
Replacement for full battery installation
instructions)
After battery installation:
• the display ashes
• the instrument performs a self-test
• then turns OFF
6. Proceed to Section 3, Operation
W CAUTION
Do not handle the sensor unnecessarily
during calibration or use. Body heat can
cause the sensor’s thermistor to change
disproportionately to the change in
gas sample temperature at the sensing
electrode. This can produce some error
until thermal equilibrium is restored.
See Appendix F, Accessories/Replacement Parts
®
for ordering MiniOX
3000 Monitor accessories and
replacement parts (Figure 2-2).
AUDIBLE ALARMS
LED ALARMS
DISPLAYED OXYGEN READING
HIGH ALARM POINT SETTING
LOW ALARM POINT SETTING
ON/OFF KEY
BATTERY TEST AND ALARM
TEST KEY
21% CALIBRATION KEY
UNLOCK KEY
(ENABLES CALIBRATION SEQUENCE)
DECREASE ALARM SET POINT KEY
SET KEY
(ENABLES ADJUSTMENTS OF
ALARM SET POINTS)
100% CALIBRATION KEY
DEFLECTOR
SENSOR
CABLE ASSEMBLY
MONITOR
STAND
SILENT KEY
INCREASE ALARM SET POINT KEY
Figure 2-2
Assembled MiniOX® 3000 Oxygen Monitor
P/N 814358 [Rev.1] 09/2010
2-2
MiniOX® 3000 Oxygen Monitor Section 3, Operation
Section 3
Operation
This section describes the following operational
procedures and functions of the MiniOX® 3000
Oxygen Monitor:
• Calibration
• Two-Point Linearity Check
• Setting the Alarms
• Alarm Conditions
• Silencing an Alarm
• Test Functions
• Installing the Sensor in a Breathing Circuit
• Mounting the Instrument with a Bracket
Calibration
®
The MiniOX
calibration:
3000 Oxygen Monitor requires
• when the operating environment changes
• after the monitor is turned ON
• if the sensor is disconnected and reconnected to
the instrument
Recommended practice of calibration is against
room air (dened as oxygen concentration of 20.8%).
When a more precise measurement is desired,
repeat calibration using 100% oxygen concentration.
Changes in pressure, humidity or temperature may
affect calibration accuracy (see Appendix B, Effects of Pressure, Humidity, and Temperature).
W CAUTION
Do not handle the sensor unnecessarily
during calibration or use. Body heat can
cause the sensor’s thermistor to change
disproportionately to the change in
gas sample temperature at the sensing
electrode. This can produce some error
until thermal equilibrium is restored.
To Calibrate In Room Air (Figure 3-1):
• daily, while in operation
EXPOSE SENSOR
1
2
3
TO ROOM AIR
PRESS I/O
PRESS 21%
CALIBRATE
1. Expose sensor to room air.
PRESS UNLOCK
AFTER 20 SECOND
CALIBRATION COMPLETE
AND
20.8% +2% DISPLAYED
4
5
3-1
Figure 3-1
Calibrating the MiniOX® 3000 Oxygen Monitor in Room Air
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Section 3, Operation
2. Press I/0 to turn ON the instrument.
• “CAL” ashes in the display
3. Press 21%. The following appears on the display:
• “CAL”
• “LOCKED”
• “21% CAL”
4. Press UNLOCK. The following appears:
• “CAL”
• “21% CAL”
• a 10-segment bar graph that “counts down”
two seconds per bar for 20 seconds
5. After 20 seconds, the calibration process is
complete the device:
• displays 20.8% ± 2% O
(18.8% to 22.8%)
2
• proceeds to the monitoring mode
• displays the current oxygen concentration as
%O
2
To Calibrate at 100% O
(Figure 3-2)
2
Note: Prior to calibrating at 100% O2, the MiniOX®
3000 Oxygen Monitor must rst be calibrated
in room air.
1. Calibrate in room air (see Section 3, To Calibrate
in Room Air)
2. Expose the sensor to 100% oxygen and allow
the readings to stabilize prior to initiating the
calibration.
• “CAL” ashes in the display.
3. Press 100%. The following appears on the
display:
• “CAL”
• “LOCKED”
• “100% CAL”
4. Press UNLOCK. The following appears on the
display:
• “CAL”
• “100% CAL”
• a 10-segment bar graph that “counts down”
two seconds per bar for 20 seconds.
5. After 20 seconds, the calibration process is
complete; the device:
• displays 100.0% +0/-2% (98% to 100%)
• proceeds to the monitoring mode
• displays the current oxygen concentration as
%O
2
CALIBRATE IN
1
2
3
ROOM AIR
EXPOSE SENSOR
TO 100% OXYGEN
PRESS 100%
CALIBRATE
PRESS UNLOCK
AFTER 20 SECOND
CALIBRATION COMPLETE
AND
100% +0/-2% DISPLAYED
4
5
Figure 3-2
Calibrating the MiniOX® 3000 Oxygen Monitor at 100% O
P/N 814358 [Rev.1] 09/2010
2
3-2
MiniOX® 3000 Oxygen Monitor Section 3, Operation
NOTE: The MiniOX® 3000 Oxygen Monitor has a
ve second “time out” following keypad
functions. If you do not press UNLOCK
within ve seconds, the instrument
returns to the ashing “CAL” mode.
NOTE: During operation if “CAL ERR”
ashes in the display, visual, audible
alarms activate, and then “CAL”
ashes, turn OFF the instrument
and repeat calibration procedure.
When recalibrating, be sure to select
the calibration value and use the
corresponding calibration gas. If “CAL
ERR” reoccurs, it may be necessary
to replace the sensor. (See Section 4,
“Sensor Replacement”)
NOTE: During operation if “CAL” displays, you
must recalibrate the monitor. If “CAL”
displays following proper recalibration, it
may be necessary to replace the sensor.
(See Section 4, Sensor Replacement)
Two-Point Linearity Check
To evaluate sensor performance, conduct a two-point
linearity check on the monitor every week or any time
you suspect that the sensor is not accurately reading
oxygen concentrations. Acceptable range for room air
measurement during a linearity check is:
two-point linearity check
• If the variation is still greater than ±2%,
replace the sensor. (See Section 4, Sensor
Replacement)
Setting the Alarms
The default Alarm Set points are:
• 18% for Low Alarm (appears in the lower left
corner of the display)
• 50% for High Alarm (appears in the lower right
corner of the display)
These alarm points can be reset between 15%
and100%.
®
NOTE: The MiniOX
3000 Oxygen Monitor
retains current alarm settings when
turned OFF, if the alarm settings are not
below 18%. However, if the alarm settings
are lower than 18% when the MiniOX®
3000 Oxygen Monitor is turned OFF or
following battery replacement, alarm set
points return to the default settings of
18% and 50%.
To Set the Low Alarm
1. Press SET once. The following appears on the
display:
• 20.8% ±2% (18.8% to 22.8%). This variation
allows for differences in:
• sampling methods
• accuracy of gas concentration
• precision of initial setting
If linearity deviates more than ±2%, the sensor is
nearing end-of-life and should be replaced.
To perform a two-point linearity check:
1. Calibrate in room air.
2. Place the sensor in a stream of oxygen with a
known concentration of 100% until the reading
stabilizes.
3. Recalibrate the monitor.
4. After calibration, measure room air. This reading
should be 20.8% ±2% (18.8% or 22.8%) within
ve minutes.
• If the variation is greater than ±2%, repeat the
• “AL”
• up/down arrows
2. Using the arrow keys, scroll up or down to the
desired Low Alarm set point (15% to 99%).
®
3. The MiniOX
3000 Oxygen Monitor “locks” this
value. After ve seconds, the monitor:
• beeps once
• automatically proceeds to the Monitoring
Mode
NOTE: The Low Alarm CANNOT be disabled or
set:
• below 15%
• above 99%
• higher than or equal to the High Alarm setting
To Set the High Alarm
1. Press SET twice. The following appears on the
display:
3-3
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Section 3, Operation
• “AL”
• up/down arrows
2. Using the arrow keys, scroll up or down to the
desired High Alarm set point (16% to 100%).
®
3. The MiniOX
3000 Oxygen Monitor “locks” this
value. After ve seconds, the monitor:
• beeps once
• automatically proceeds to Monitoring Mode
(Press SET once after Step 2 to manually
proceed to Monitoring Mode)
NOTE: The HIGH Alarm value:
• CANNOT be set equal to, or less than, the
Low Alarm value
• CAN be disabled by increasing the alarm set
point beyond 100% until “--” displays
Alarm Conditions
High and Low Oxygen Concentration Alarms
®
When the MiniOX
3000 Oxygen Monitor detects an
oxygen concentration that exceeds the preset alarm
limit:
• the red LED ashes for that alarm (High or Low)
• an audible alarm activates
• the measured concentration displays
Silencing an Alarm
If the alarm condition is not corrected within the
preset silence interval (30,60, or 90 seconds):
• the audible alarm reactivates
If the alarm condition is corrected and then reoccurs
within this interval:
• both audible and visual alarms activate
2. To manually terminate the silence mode, press
the down arrow (labeled with a ▼ icon).
Low Battery Alarms
®
The MiniOX
3000 Oxygen Monitor features a twostage battery alarm that warns of depleted and
expired battery voltage:
• The rst alarm alerts the operator that the monitor
has approximately six hours of operating time
remaining:
• “LOW BAT” appears on the display
• an audible alarm sounds at 30-second
intervals
• If the operator does not replace the battery after
this alarm, a second low battery alarm activates
when the battery is no longer able to support
monitoring.
• the monitor displays “LOW BAT” and “---”
• audible and visual alarms are activated
The MiniOX® 3000 Oxygen Monitor features a silence
function that allows the operator to temporarily mute
the O2 concentration audible alarm.
1. When an audible alarm activates, press the silent
key (labeled with a horn icon) to mute the alarm
for up to three 30-second intervals, press:
• once for a 30-second interval
• twice (within two seconds) for a 60-second
interval
• three times (within four seconds) for a
90-second interval
The following appears on the display:
• a horn icon
• a three-bar graph showing a “countdown”
of 10 seconds per bar for each 30 second
interval
NOTE: When the audible alarm is silenced, the
appropriate visual alarm continues to
ash.
P/N 814358 [Rev.1] 09/2010
For periodic checks on battery status, see Section 3,
Test Functions”
Sensor Indicator
®
During monitoring, the MiniOX
3000 Oxygen Monitor
displays “SENSOR” and “OFF”; audible and visual
alarms activate if:
• the oxygen sensor becomes disconnected
• the cable fails or disconnects from the sensor or
instrument
• the sensor membrane is perforated
• the thermistor circuit opens
3-4
MiniOX® 3000 Oxygen Monitor Section 3, Operation
Error Handling
The MiniOX® 3000 Oxygen Monitor performs a selftest
• when it is turned ON
• after a battery is installed
If the microprocessor detects an error during these
routine self-tests or during monitoring:
• the display ashes “ERR” and an error code
• audible and visual alarms activate
• the monitor ceases to operate until the
appropriate service is performed; to reset the
Table 3-1 Alarms and Status Messages
ALARMCAUSECORRECTIVE ACTION
alarm:
Low O
2
O2 concentration appears,
visual alarm ashes, and
audible alarm activates
High O
O2 concentration appears,
visual alarm ashes, and
audible alarm activates
“SENSOR OFF” displays;
visual and audible alarms
activate
During calibration, “CAL ERR”
ashes, visual and audible
alarms activate; then “CAL”
ashes
During calibration, “CAL”
ashes
“LOW BAT” displays and monitor beeps every 30 seconds
“LOW BAT” displays “---”
ashes, both visual alarms
activate, and monitor emits a
three-tone beep for four minutes; monitor shuts OFF
“ERR” and error code (01 to
08) display, visual and audible
alarms activate and monitor
locks
Alarm:
2
Oxygen concentration is below the
preset Low Alarm setting
Oxygen concentration is above the
present High Alarm setting
Cable is disconnected or malfunctioning
Sensor is disconnectedVisually check cable for damage. If necessary, replace cable
Sensor membrane is perforatedIf the alert persists, replace sensor
Thermistor circuit is open on the
sensor circuit board
Calibration performed using a
gas other than room air or 100%
oxygen or wrong calibration key
pressed (100%/21%)
Sensor signal is out of monitoring
range
Monitor has approximately six
hours of operating time remaining
Battery is expired and monitor is
not operating
Microprocessor detects internal
error
monitor, press SET or turn the monitor OFF,
then ON (press I/0 twice). For further details see
Appendix E, Error Codes
NOTE: During normal operation, if a key is pressed
and held longer than nine seconds, audible
and visual alarms activate.
1. To reset the monitor, press SET or turn the
monitor OFF, then ON (Press I/0 twice). For
further details see Appendix E, Error Codes
®
Table 3-1 is a summary of MiniOX
3000 Oxygen
Monitor alarms and status messages.
Check patient and delivered oxygen concentration
Verify that low alarm setting is appropriate
Check patient and delivered oxygen concentration
Verify that high alarm setting is appropriate
Check cable and sensor connections
If alert persists following sensor replacement, call Ohio Medical
Corporation service
Recalibrate the monitor, ensure proper calibration key pressed
for calibration gas of room air or 100% oxygen
If the alert persists, replace sensor
If alert persists following sensor replacement, call Ohio Medical
Corporation service
Recalibrate the monitor, ensure proper calibration key pressed
for calibration gas of room air or 100% oxygen
If the alert persists, replace the sensor
If alert persists following sensor replacement, call Ohio Medical
Corporation service
Replace the battery as soon as possible; then, recalibrate the
monitor and reset the high and low alarm values
Replace the battery as soon as possible; then recalibrate the
monitor and reset the high and low alarm values
Note error code and see Appendix E, Error Codes
Disconnect Battery
Call Ohio Medical Corporation Service
3-5
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Section 3, Operation
Test Functions
Battery Test
In addition to alarms alerting the operator to low and
expired battery voltage, the MiniOX® 3000 Oxygen
Monitor features a keypad function allowing the
operator to check the relative battery life at any time.
To test battery status:
1. Press TEST once. The following appears on the
display:
• “BATTERY TEST”
• a 10-bar graph showing relative battery life.
Ten bars indicate a fully charged battery; one
bar indicates a battery near depletion
®
2. After ve seconds, the MiniOX
Monitor proceeds to the Monitoring mode,
displaying oxygen concentrations as %O2.
High and Low Alarms Test
3000 Oxygen
Installing the Sensor in a Breathing Circuit
To use the sensor in a breathing circuit, you will need
the following:
• Sensor (with deector)
• Tee adapter
• Retaining Strap
To install the sensor in a breathing circuit:
1. Install the tee adapter into the breathing circuit
upstream from the humidier. Make sure that side
port of the tee adapter is facing upward.
2. Remove the coiled cable from the sensor.
3. Firmly insert the sensor (with deector) into the
tee adapter with the deector pointing downward
(Figure 3-3) to prevent moisture from condensing
onto the sensor membrane (See Appendix B, Effects of Pressure, Humidity, and Temperature).
Make sure that the sensor ts tightly into the tee
adapter.
®
The MiniOX
3000 Oxygen Monitor features a test
function to ensure that the high and low alarms
activate at the preset alarm levels. To test the alarms:
Press TEST twice.
• “ALARM TEST” displays
• the instrument automatically scrolls up to the high
alarm value and ashes this value on the display
• the audible alarm and visual high alarm activate
(If the high alarm value is disabled, the instrument
scrolls down from 100% to the low alarm value)
• the instrument then automatically scrolls down to
the low alarm value and ashes this value on the
display
• the audible alarm and visual low alarm activate
®
• the MiniOX
3000 Oxygen Monitor proceeds
to the Monitoring Mode, displaying oxygen
concentrations as %O
NOTE: The MiniOX
2
®
3000 Oxygen Monitor is not
monitoring during the Alarm Test which
may take up to 30 seconds to complete.
NOTE: The Alarm Test will not function if the
MiniOX® 3000 Oxygen Monitor is in an
alarm condition.
4. Install one end of the retaining strap over a post
on the side of the tee adapter.
5. Loop the strap over the sensor, inserting the strap
center hole over the sensor cable jack.
6. Install the other end of the strap over the other tee
adapter post.
7. Re-attach the coiled cable to the sensor. Tighten
the twist collar.
The sensor is securely placed for monitors (See
Figure 3-4).
Mounting the Instrument with a Pole Bracket
®
The MiniOX
3000 Oxygen Monitor (as shown in
Figure 3-5) can be attached to a vertical or horizontal
pole using the Ohio Medical Corporation Mounting
Bracket (See Appendix F, Accessories/Replacement Parts). To mount the MiniOX® 3000 Oxygen Monitor:
1. Congure the bracket for either horizontal or
vertical attachment by adjusting the dovetail
mounting plate on the bracket.
2. Slide the mounting bracket onto the pole; secure
by tightening the screw knob.
3. Slide the wedge on the back of the monitor into
the dovetail mounting plate.
P/N 814358 [Rev.1] 09/2010
3-6
MiniOX® 3000 Oxygen Monitor Section 3, Operation
SENSOR CABLE
RETAINING STRAP
SENSOR
GASKET
DEFLECTOR
O-RING
15mm TEE
3-7
Figure 3-3
Installing Sensor into Tee Adapter
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Section 3, Operation
SENSOR CABLE
RETAINING STRAP
SENSOR
DEFLECTOR
15mm TEE
ADAPTER
FASTEN RETAINING STRAP
TO POSTS ON TEE
Figure 3-4
Sensor in Breathing Circuit
INSTRUMENT
MOUNTING
DOVETAIL
SLIDE DOVETAIL
COMPLETELY INTO
BRACKET
MOUNTING
BRACKET
INSTALLED ON
POLE
Figure 3-5
MiniOX® 3000 Oxygen Monitor with Pole Mounting Bracket
P/N 814358 [Rev.1] 09/2010
3-8
MiniOX® 3000 Oxygen Monitor Section 4, Maintenance and Care
Section 4
Maintenance and Care
W WARNING
Use only genuine Ohio Medical
Corporation replacement parts when
performing any maintenance procedures
included in this manual. Failure to do
so may seriously impair the monitor’s
performance. Repair or alteration of the
MiniOX® 3000 Oxygen Monitor beyond the
scope of the maintenance instructions or
by anyone other than an authorized Ohio
Medical Corporation service person could
cause the product to fail to perform as
designed.
This section describes the following general
maintenance and care procedures for the MiniOX®
3000 Oxygen Monitor:
• Battery Replacement
• Sensor Replacement
• Deector Replacement
• Cable Replacement
• Cleaning
Battery & Battery Cover Replacement
®
The MiniOX
3000 Oxygen Monitor requires one
standard 9-volt alkaline battery. To replace the
battery:
1. Verify that the monitor is turned OFF. The display
should be blank.
2. Pull out the support stand from the back of the
case.
3. Unscrew the two screws on the battery cover in
the back of the instrument and remove cover. See
Figure 4-1.
4. Remove the battery from the case and unsnap
the battery from the battery holder.
NOTE: To ensure proper start-up, wait at least
45 seconds before connecting the fresh
battery to the battery connector.
5. Snap the terminal of the new battery into the
battery holder.
6. Install the battery cover and screw into place.
Make sure that the battery cover is properly
®
seated and at on the back of the MiniOX
3000
Oxygen Monitor case.
7. Recalibrate the monitor. Reset the low and high
alarms, if desired.
NOTE: To maximize battery life, press I/O to
®
turn OFF the MiniOX
3000 Oxygen
Monitor when not monitoring. In
order to retain alarm settings, do not
remove battery .
4-1
Figure 4-1
Battery Cover Removal
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Section 4, Maintenance and Care
Sensor Replacement
1. Verify that the monitor is turned OFF. The display
To ensure safe and effective use of your device, the
sensor must be replaced with a MiniOX® sensor as
should be blank.
2. Disconnect the old cable from the sensor.
this sensor is manufactured for this instrument. Use
of other types of sensors has not been tested and is
not endorsed by Ohio Medical Corporation. Use of
other sensor types will void your warranty.
3. Disconnect the old cable from the instrument.
4. Attach the new cable to the sensor. Firmly press
the connector until the sensor snaps into place.
Tighten the twist collar.
There are no serviceable parts in the sensor or cable
assemblies, the entire unit must be replaced.
Replace the sensor when:
5. Attach the new cable to the instrument. Firmly
press the connector until it snaps into place.
Tighten the twist collar.
6. Recalibrate the instrument.
• Room air reading is greater than 20.8% ±2%
(18.8% to 22.8%) in Two-Point Linearity Check
®
• The MiniOX
3000 Oxygen Monitor will not
Cleaning, Disinfection and Sterilization
W CAUTION
calibrate.
• “Sensor” and “OFF” display and audible and
visual alarms persist when sensor and cable
connections are correct and cable is viable.
Never autoclave, immerse, or expose the
®
MiniOX
3000 Oxygen Monitor (including
sensor) to high temperatures (>70oC). Never
expose the device to pressure, irradiation,
vacuum, steam, or chemicals (other than
To replace the sensor:
alcohol or mild cleaning agents).
1. Verify that the monitor is turned OFF. The display
should be blank.
2. Disconnect the expired sensor from the coiled
cable.
3. Attach a new sensor to the coiled cable and rmly
press the connector until the sensor snaps into
place. Tighten the twist collar.
4. Recalibrate the monitor.
Deector Replacement
To replace the deector:
1. Gently unscrew the old deector and remove the
gasket.
2. Insert a new gasket into the large end of the new
deector. Carefully screw the new deector onto
the sensor, making sure that the new gasket is
properly seated.
See Figure 3-3 for proper sensor and deector
assembly.
Cable Replacement
Clean the instrument and sensor by wiping with a
cloth lightly dampened with Isopropyl Alcohol or mild
detergent. Make sure that no moisture seeps into the
instrument case or cable jack port.
Instrument
When cleaning or disinfecting the instrument, care
must be taken to prevent entry of solutions into the
instrument case. If it is suspected that solutions or
moisture enters the case, verify performance by
conducting a self-test (see Possible physical damage to instrument, Appendix D).
Cleaning
The external surfaces of the unit may be cleaned
by wiping them with a cloth moistened with a mild
detergent solution.
Disinfection
The external surfaces of the unit may be disinfected
by wiping them with a cloth moistened with ethanol
or Cidex. The instrument is not designed to withstand
the conditions imposed by steam, ethylene oxide or
radiation sterilization.
To replace the coiled cable:
P/N 814358 [Rev.1] 09/2010
4-2
MiniOX® 3000 Oxygen Monitor Section 4, Maintenance and Care
Sensor and Cable
Cleaning
The external surfaces of the oxygen sensor and of
the cable may be cleaned by wiping them with a cloth
moistened with a mild detergent solution.
Disinfection
The external surfaces of the oxygen sensor housing
and of the cable may be disinfected by wiping them
with a cloth moistened with ethanol or Cidex. The
instrument is not designed to withstand the conditions
imposed by steam, ethylene oxide or radiation
sterilization.
Sensor Deector, Retaining Strap and Tee
Adapter
Cleaning
The sensor deector, retaining strap and tee
adapter may be cleaned by wiping them with a cloth
moistened with a mild detergent solution. The parts
must be thoroughly dry before they are used.
Sterilization
The sensor deector, retaining strap and tee adapter
may be sterilized using Cidex (per manufacturer’s
instructions), steam or ethylene oxide. Due to the
varying conditions imposed on materials during
sterilization, it is not possible to determine the exact
number of times sterilization processes can be
carried out. Therefore, the operator must carefully
examine the sensor deector, retaining strap and tee
adapter after sterilization and prior to use to verify
that the item is t for use. The operator must verify
that the items are free from tears and cracks and that
the items have not undergone any material changes
that may compromise their tness for use (e.g.,
brittleness and dimensional changes). The operator
must also examine the items to verify that the items
are free of chemical residuals resulting from the
sterilization process.
Because of the variability of cleaning, disinfection
and sterilization processes, Ohio Medical Corporation
cannot provide specic sterilization instructions, nor
can the sterility of an item be ensured.
Disinfection
The sensor deector, retaining strap and tee adapter
may be disinfected by washing them with ethanol or
Cidex (per manufacturer’s instructions). The parts
must be thoroughly dry before they are used.
4-3
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Appendix A, Specications
Appendix A
Specications
Table A-1. Specications
ALARM SYSTEM
O
2
LOW/HIGH O
ALARM
ACCURACY
LINEARITY
NOMINAL RESPONSE TIME
directed at sensor at 2 liters per
(O
2
minute at RTP)
SENSOR LIFEOver one year in normal medical conditions (nominal 750,000 O
SENSOR SHELF LIFESix months minimum (stored in sealed package)
BATTERY LIFEApproximately 1500 device hours, assuming 80% ON, 20% OFF
OPERATING TEMPERATURE RANGE0
STORAGE TEMPERATURE-40
HUMIDITY5% to 95% RH (non-condensing)
DIMENSIONS
WEIGHT
BATTERY ALARM
SYSTEM
INSTRUMENT5.98” X 3.26” X 1.31”
SENSOR1-1/5” OD x 1-3/4” long
INSTRUMENT260 g (9.2 oz)
SENSOR35 g (1.25 oz)
STAGE 1
STAGE 2
*RTP: Room temperature and pressure, e.g., 23oC + 3o and ambient barometric pressure
2
Pulsating alarm, ashing LEDs
+ 1% of full scale (RTP*) as calibrated in the environment of use (total accuracy, including
linearity, is 2%)
+1% of full scale (RTP*)
90% in 20 seconds
97% in 30 seconds
hours)
2
o
C to 40oC (32oF to 104oF)
o
C to 70oC (-40oF to 158oF)
Warning that approximately six hours of monitoring life remain; LOW BAT appears on display; beep sounds every 30 seconds until battery is replaced or Stage 2 occurs.
Alarm that proper operation is not assured; monitor shuts down, audible and visual alarms
are activated.
P/N 814358 [Rev.1] 09/2010
A-1
MiniOX® 3000 Oxygen Monitor Appendix B, Effects of Pressure, Humidity and Temperature
Appendix B
Effects of Pressure, Humidity and
Temperature
W CAUTION
To ensure accurate and reliable oxygen
monitoring, it is necessary to have a
thorough understanding of the effects of
pressure, humidity, and temperature on the
sensor.
Effects of Pressure
The sensor responds to partial pressure (not
percentage) of oxygen. Changes in barometric
pressure change the reading, even if the percent of
oxygen in the sample remains constant.
Partial pressure of oxygen (PO
of oxygen (%O
) times the pressure at which the
2
sample is measured (mmHg).
PO2 = (%O2 ) (mmHg)
For example: at sea level, the pressure equals 760
mmHg and dry air contains 21% O2. Therefore:
PO2 = (21%) (760 mmHg)
) equals the percent
2
W CAUTION
Do not expose the sensor to pressure
outside the range of 600 to 900 mmHg
(23.62 to 35.43 Inches Hg), as this may
cause inaccuracies.
Effects of Humidity
The presence of humidity in an oxygen sample
decreases the actual concentration of oxygen.
Humidity in a sample has the same effect as diluting
the sample with another gas. For example, if 100%
oxygen is saturated with 100% humidity, the actual
concentration of oxygen drops to 96% - 97%.
As with all oxygen gas sensors, condensation on
the sensor membrane blocks the ow of oxygen,
resulting in a lower oxygen concentration reading and
an increased response time. This is a typical problem
resulting from locating the sensor downstream from
the humidier in an oxygen delivery system; clearing
the sensor face and deector restores normal
operation. However, to avoid this problem when using
the sensor in a breathing circuit, position the sensor
upstream of the humidier and mount the sensor with
the deector pointing downward to prevent moisture
from draining onto the sensor membrane.
= 160 mmHg
PO
2
If you calibrate an instrument to read 21% at 760
mmHg partial pressure and then take the instrument
to an area above sea level, a lower reading occurs
due to a lower partial pressure. For a pressure of 700
mmHg:
PO
= (21%) (700 mmHg )
2
=147 mmHg
PO
2
The percent reading on the instrument is derived from
the following formula:
PO
PO
When PO
Actual = 147 mmHg
2
Sea level = 160 mmHg
2
sea level is 21%
2
(21 %) (147 mmHg)
Actual = (160 mmHg) = 19.3%
PO
2
Therefore, to eliminate error due to pressure
changes, the instrument must be calibrated at the
pressure in which it is used.
B-1
P/N 814358 [Rev.1] 09/2010
Effects of Temperature
Due to an internal thermistor (temperature variable
®
resistor), the MiniOX
3000 sensor is minimally
affected by temperature change. Variations in the
sensor reading from temperature change are less
than 3% when the instrument is calibrated and used
in a monitoring environment of 0oC to 40°C (32°F to
104°F).
Operating Temperature Range:
0o to 40oC (32o to 104oF)
W CAUTION
Do not use instrument outside operating
temperature range.
W CAUTION
Do not handle the sensor more than
necessary during calibration or use. Body
heat can cause the sensor’s thermistor to
change disproportional to the change in
gas sample temperature at the sensing
electrode. This can produce some error,
until thermal equilibrium is restored.
MiniOX® 3000 Oxygen Monitor Appendix C, Interferent Gases and Vapors
Appendix C
Interferent Gases and Vapors
Table C-1. Interferent Gases and Vapors
InterferentVOLUME % DRYINTERFERENCE EQUIVALENT OF % O
If problem persists, call Ohio Medical Corporation service
Press I/O to run ON monitor
If monitor does not respond, replace the battery
If problem persists, call Ohio Medical Corporation service
If alert persist following sensor replacement, call Ohio
Medical Corporation service
Recalibrate the monitor, ensure proper calibration key
pressed for calibration gas of room air or 100% oxygen
If alert persists, replace sensor
If alert persists following sensor replacement, call Ohio
Medical Corporation service
Recalibrate the monitor, ensure proper calibration key
pressed for calibration gas of room air or 100% oxygen
If alert persists, replace sensor
If alert persists following sensor replacement, call Ohio
Medical Corporation service
Replace the battery as soon as possible; then recalibrate the
monitor and reset the high and low alarm values
Replace the battery as soon as possible; then recalibrate the
monitor and reset the high and low alarm values
Note error code and see Appendix E, Error Codes
Disconnect battery
Call Ohio Medical Corporation service
Turn instrument OFF, then ON to initiate self-test. If
instrument is damaged, “ERR” and error code (01 to 08)
display. Disconnect battery. Call Ohio Medical Corporation
service
Display is blank
No Response to keypad
commands
“SENSOR OFF” displays:
visual and audible alarms
activate
During calibration, “CAL
ERR” ashes, visual and
audible alarms activate;
then “CAL” ashes
During operation, “CAL”
ashes
‘LOW BAT” displays and
monitor beeps every 30
seconds
“LOW BAT” displays
“---” ashes, both
visual alarms activate,
and monitor emits a
three-tone beep for four
minutes; monitor shuts
OFF
“ERR” and error code (01
to 08) display, visual and
audible alarms activate
and monitor locks
Possible physical damage to instrument
Monitor is turned OFF or Battery is expired.
Battery is expired
Cable is disconnected or malfunctioningCheck cable and sensor connections
Sensor is disconnectedVisually check cable for damage. If necessary, replace cable
Sensor membrane is perforatedIf alert persists, replace sensor
Thermistor circuit is open on the sensor
circuit board
Calibration performed using a gas other
than room air or 100% oxygen or wrong
calibration key pressed (100%/21%)
Sensor signal is out of monitoring range
Monitor has approximately six hours of
operating time remaining
Battery is expired and monitor is not operating
Microprocessor detects internal error
Instrument is dropped and/or exposed to
uids
D-1
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Appendix E, Error Codes
Appendix E
Error Codes
The MiniOX® 3000 Oxygen Monitor performs
diagnostic tests to detect errors that could cause
unreliable monitor operation. These tests are
performed:
• when a battery is installed (power-up)
• when the monitor is turned ON
• during routine monitoring
During self-test or any time the monitor detects an
operational error, the MiniOX
• ceases operation
• Display “ERR” and an error code
• activates visual and audible alarms
®
3000 Oxygen Monitor;
Error alarms continue until:
1. The monitor is shut off (press I/O)
2. The monitor is reset and error does not reoccur
(press set)
3. The battery is disconnected
If an error reoccurs, the monitor is inoperative and
must not be used; call Ohio Medical Corporation
Repair at:
866-549-6446, Option #1
The error code that appears on the display
corresponds to a specic failure. The audible alarm is
a beep that corresponds to this code (e.g., “ERR 3”
triggers a three-beep alarm).
Table E-1. Error Codes
ERROR DISPLAYCORRESPONDING ERROR
01System error
02Random Access Memory (RAM) error
03Read Only Memory (ROM) error
04Analog error
05Timing error
06Keypad error*
07Battery error **
08LCD error
*Press SET. If error reoccurs, call Ohio Medical Corporation Repair
**Replace battery. If error reoccurs, call Ohio Medical Corporation Repair
P/N 814358 [Rev.1] 09/2010
E-1
MiniOX® 3000 Oxygen Monitor Appendix F, Accessories and Replacement Parts
Appendix F
Accessories and Replacement Parts
Table F-1, Accessories and Replacement Parts
ITEMPART NO.
Oxygen Sensor406931
Mounting Bracket (Pole)474664
Mounting Bracket (Wall)10023945
Coiled Cable472045
Tee Adapter473021
Retaining Strap634249
Operation Manual on CD600700
Operation Manual, printed814358
®
MiniOX
Battery Cover Assembly710974
Battery Cover Screw, Knurled655579
Deector803229
Dovetail474606
Screw, #4x1/4” (set of 4) thread forming450001
Inspection and ServicingSVC-301
Cleaning, Calibration and TestingSVC-302
Enclosure ReplacementSVC-303
Circuit Board ReplacementSVC-304
Connector ReplacementSVC-305
3000 Oxygen Monitor814365
Servicing
F-1
P/N 814358 [Rev.1] 09/2010
NOTES
P/N 814358 [Rev.1] 09/2010
Ohio Medical Corporation®
1111 Lakeside Drive
Gurnee, IL 60031 USA
phone: 1-866-549-6446
fax: 1-847-855-6218
www.ohiomedical.com
P/N 814358 [Rev.1] 09/2010
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.