Ohio MiniOX 3000 User manual

MiniOX® 3000
Oxygen Monitor
Operation Manual
THIS MANUAL MUST BE READ CAREFULLY PRIOR TO THE OPERATION OF THIS DEVICE. THIS DEVICE WILL PERFORM AS DESIGNED ONLY IF USED IN ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONS. IMPROPER USE MAY CAUSE THE DEVICE TO FAIL TO PERFORM AS DESIGNED AND MAY CAUSE INJURY TO THE PATIENT AND/OR HEALTHCARE PROFESSIONAL.
W WARNING
®
The warranties made by Ohio Medical Corporation not installed, used and serviced in accordance with the instructions in this manual. Please protect yourself and your patients by following them. We encourage our customers to write or call regarding this equipment prior to use or for any additional information relative to use or repairs.
This product is not intended as a life-sustaining or life-supporting device.
FEDERAL (USA) LAW RESTRICTS THIS DEVICE TO SELL BY OR ON THE ORDER OF A PHYSICIAN
Ohio Medical Corporation® | 1111 Lakeside Drive | Gurnee, IL 60031 USA
1-866-549-6446 | fax: 1-847-855-6218 | www.ohiomedical.com
with respect to these products are voided if the products are
P/N 814358 [Rev.1] 09/2010
P/N 814358 [Rev.1] 09/2010
Ohio Medical Corporation
®
Medical Instrument Warranty
This product is sold by Ohio MedicalTM under the warranties set forth in the following paragraphs. Such warranties are extended only with respect to the purchase of this product directly from Ohio Medical or Ohio Medical’s Authorized Dealers as new merchandise and are extended to the rst Buyer thereof, other than for purpose of resale.
For a period of twelve (12) months from the date of original delivery to Buyer, to Buyer’s order, or to an Ohio Medical Authorized Dealer, this product, other than its expendable parts, is warranted to be free from functional defects in materials and workmanship and to conform to the description of the product contained in the operation manual and accompanying labels and/or inserts, provided that the same is properly operated under conditions of normal use, that regular periodic maintenance and service is performed and that replacements and repairs are made in accordance with the instructions provided. This same warranty is made for a period of sixty (60) days with respect to the expendable parts. The foregoing warranties shall not apply if the product has been repaired other than by Ohio Medical or in accordance with written instructions provided by Ohio Medical, or altered by anyone other than Ohio Medical, or if the product has been subject to abuse, misuse, negligence, or accident.
Ohio Medical’s sole and exclusive obligation and Buyer’s sole and exclusive remedy under the above warranties is limited to repairing or replacing, free of charge, at Ohio Medical’s option, a product, which is telephonically reported to the nearest Ohio Medical Regional Service Ofce and which, if so advised by Ohio Medical, is thereafter returned with a statement of the observed deciency, not later than seven (7) days after the expiration date of the applicable warranty,to the designated Ohio Medical Service Ofce during normal business hours, transportation charges prepaid, and which, upon Ohio Medical’s examination, is found not to conform with the above warranties. Ohio Medical shall not be otherwise liable for any damages including, but not limited to incidental damages, consequential damages, or special damages.
There are no express or implied warranties which extend beyond the warranties herein above set forth. Ohio Medical makes no warranty of merchantability or tness for a particular purpose with respect to the product or parts thereof.
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor General WARNINGS and CAUTIONS
General Warnings and Cautions
W WARNING
1. The MiniOX® 3000 Oxygen Monitor will perform to specications only if it is used and serviced in accordance with the manufacturer’s instructions. This instrument is to be used only by qualied, trained personnel who have carefully read the operating manual and labels and who have observed the information set forth. If this instrument does not perform as described in this manual, the instrument must not be used until the condition is rectied.
2. The MiniOX
®
3000 Oxygen Monitor must be calibrated prior to each use. A two point calibration check must be performed weekly. See Section 3, Operation. If the instrument cannot be calibrated, the sensor must be replaced. If the instrument is still unable to be calibrated, the instrument must be serviced.
3. The oxygen sensor has a minimal response to certain gases other than oxygen. Be aware of these gases and their interference levels. See Appendix C,
Interferent Gases and Vapors.
4. The oxygen sensor is affected by changes in barometric pressure. See Appendix B, Effects of
Pressure, Humidity and Temperature.
5. The sensor is a sealed unit containing a potassium hydroxide electrolyte, which is caustic. If the sensor should develop a leak and the contents come in contact with skin or clothing, rinse area with large quantities of water. In case of eye contact, immediately ush eyes for at least 15 minutes, holding eyes open. Call a physician. Dispose of a leaking sensor immediately; it must be disposed of in accordance with all applicable federal, state, and local regulations.
6. Ensure a tight t exists between the sensor and the tee adapter. Ohio Medical tee adapters are engineered to t securely with Ohio Medical sensors. However, the sensor retaining strap must be used to prevent accidental separation of sensor and tee adapter.
7. Never operate the MiniOX
®
3000 Oxygen Monitor if it is suspected that water or other liquids have entered into the case. If this occurs, immediately turn the unit OFF and contact your nearest Ohio Medical Service Center for additional information.
8. Use of devices generating or emitting electromagnetic radiation near the MiniOX
®
3000 Oxygen Monitor may interfere with the proper operation of the product, causing it to fail to perform as designed. Particularly, the electromagnetic radiation from the interfering device may cause the product to display incorrect/ erratic values or to stop operating. Special attention should be paid to the patient if this occurs.
9. Never use the MiniOX
®
3000 Oxygen Monitor in combustible atmosphere, such as occurs with ammable anesthetics. Such use could result in ignition of atmosphere.
10. Never allow an excess length of cable near the patient’s head or neck, as such could result in strangulation. Secure excess cable to the bed rail or other suitable object.
FAILURE TO COMPLY WITH THESE WARNINGS COULD JEOPARDIZE THE WELL-BEING OF THE PATIENT AND/OR HEALTH CARE PROFESSIONAL.
W CAUTION
1. Remove internal battery if unit is to be non-operational for extended periods of time.
2. Observe polarity when inserting a new battery. Incorrect connection may cause damage to the instrument.
3. The MiniOX
®
3000 Oxygen Monitor must never be immersed in any cleaning solution, autoclaved, or exposed to temperatures greater than 70oC.
4. Use only genuine Ohio Medical Corporation accessories and replacement parts. Failure to do so may seriously impair the monitor’s performance. Repair or alteration of the MiniOX
®
3000 Oxygen Monitor beyond the scope of the maintenance instructions or by anyone other than an authorized Ohio Medical Corporation service person could
WC-1
P/N 814358 [Rev.1] 09/2010
cause the product to fail to perform as designed.
5. Improper mounting of the sensor in a breathing circuit may result in inaccurate readings. The sensor MUST be mounted with the deector pointing downward to prevent moisture collection on the sensor membrane (see Figures 3-3 and 3-4 in Section 3, Operation). Install the sensor upstream from the humidier to minimize its exposure to moisture.
6. Never use a MiniOX
®
3000 Oxygen Monitor with a cable that appears worn, cracked, or has damaged insulation.
FAILURE TO COMPLY WITH THESE CAUTIONS COULD RESULT IN INSTRUMENT DAMAGE AND/OR FAILURE OF UNIT TO PERFORM TO SPECIFICATIONS.
MiniOX® 3000 Oxygen Monitor General WARNINGS and CAUTIONS
Denition of Warnings and Cautions
W WARNING
W CAUTION
NOTE:
Statement citing a potential safety hazard and possible injury to yourself or others.
Statement citing a possibility of damage to the instrument or other property.
Advisory on instrument function.
P/N 814358 [Rev.1] 09/2010
WC-2
MiniOX® 3000 Oxygen Monitor Table of Contents
Section 1
Introduction ..........................................................................................1-1
General Description.................................................................................................................. 1-1
Operating Principles. ................................................................................................................1-2
MiniOX® 3000 Oxygen Sensor ............................................................................................ 1-2
MiniOX® 3000 Instrument .................................................................................................... 1-2
Operator Interface .............................................................................................................. 1-2
Internal Functions .............................................................................................................. 1-2
Data and System Status .....................................................................................................1-2
Performance Features ............................................................................................................ 1-2
Calibration ............................................................................................................................ 1-2
Low/High Oxygen Concentration Alarms ............................................................................. 1-3
Low Battery Alarms ............................................................................................................. 1-3
Sensor Indicator .................................................................................................................. 1-3
Calibration Needed Indicator .............................................................................................. 1-3
Error Handling ..................................................................................................................... 1-4
Test Functions ...................................................................................................................... 1-4
Radio FrequencylElectromagnetic Insensitivity. ................................................................... 1-4
Section 2
Setting Up the MiniOX® 3000 Oxygen Monitor ................................ 2-1
Figure 2-1 MiniOX® 3000 Oxygen Monitor ......................................................................2-1
W CAUTION ............................................................................................................... 2-2
Figure 2-2 Assembled MiniOX® 3000 Oxygen Monitor ................................................... 2-2
Section 3
Operation ..............................................................................................3-1
Calibration ................................................................................................................................ 3-1
W CAUTION .................................................................................................................... 3-1
To Calibrate In Room Air, Figure 3-1 ................................................................................... 3-1
Figure 3-1 Calibrating the MiniOX® 3000 Oxygen Monitor in Room Air.......................... 3-1
To Calibrate at 100% O2 , Figure 3-2 .................................................................................. 3-2
Figure 3-2 Calibrating the MiniOX® 3000 Oxygen Monitor at 100% O2 ........................... 3-2
Two-Point Linearity Check....................................................................................................... 3-3
Setting the Alarms. .................................................................................................................. 3-3
To Set the Low Alarm ........................................................................................................... 3-3
To Set the High Alarm ......................................................................................................... .3-4
Alarm Conditions ..................................................................................................................... 3-4
High and Low Oxygen Concentration Alarms ..................................................................... 3-4
Silencing an Alarm ............................................................................................................... 3-4
Low Battery Alarms. ............................................................................................................ 3-4
Sensor Indicator ................................................................................................................... 3-4
Error Handling ...................................................................................................................... 3-5
Table 3-1 Alarms and Status Messages .......................................................................... 3-5
Test Functions ..................................................................................................................... 3-6
Battery Test .......................................................................................................................... 3-6
High and Low Alarms Test. .................................................................................................. 3-6
Installing the Sensor in a Breathing Circuit. .............................................................................3-6
TOC-1
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Table of Contents
Mounting the Instrument with a Pole Bracket ...........................................................................3-6
Figure 3-3 Installing Sensor into Tee Adapter ..................................................................... 3-7
Figure 3-4 Sensor in Breathing Circuit ................................................................................. 3-8
Figure 3-5 MiniOX® 3000 Oxygen Monitor with Pole Mounting Bracket .............................. 3-8
Section 4
Maintenance and Care ........................................................................4-1
W CAUTION ................................................................................................................... 4-1
Battery & Battery Cover Replacement. ....................................................................................4-1
Figure 4-1. Battery Cover Removal ................................................................................ 4-1
Sensor Replacement ................................................................................................................4-2
Deector Replacement .............................................................................................................4-2
Cable Replacement ..................................................................................................................4-2
Cleaning , Disinfection and Sterilization .................................................................................... 4-2
W CAUTION .................................................................................................................... 4-2
Instrument ........................................................................................................................... 4-2
Sensor and Cable ...............................................................................................................4-3
Sensor Deector, Retaining Strap and Tee Adapter ............................................................4-3
Appendix A
Specications ............................................................................................. A-1
T able A-1. Specications. .................................................................................................A-1
Appendix B
Effects of Pressure, Humidity and Temperature ............................ B-1
W CAUTION ...................................................................................................................B-1
Effects of Pressure ..................................................................................................................B-1
W CAUTION ...................................................................................................................B-1
Effects of Humidity ...................................................................................................................B-1
Effects of Temperature ............................................................................................................B-1
W CAUTION ...................................................................................................................B-1
W CAUTION ...................................................................................................................B-1
Appendix C
Interferent Gases and Vapors............................................................ C-1
Table C-1 Interferent Gases and Vapors ....................................................................... C-1
Appendix D
Troubleshooting ................................................................................ D-1
Table D-1 Troubleshooting ............................................................................................ D-1
Appendix E
Error Codes ......................................................................................... E-1
Table E-1 Error Codes ....................................................................................................E-1
Appendix F
Accessories and Replacement Parts ................................................F-1
Table F-1 Accessories and Replacement Parts .............................................................F-1
P/N 814358 [Rev.1] 09/2010
TOC-2
MiniOX® 3000 Oxygen Monitor Section 1, Introduction
Section 1 Introduction
The MiniOX® 3000 Oxygen Monitor provides continuous, direct monitoring of oxygen mixtures in a variety of applications, including:
Respiratory Therapy (e.g., respirators, ventilators, pediatric incubators)
Anesthesiology (e.g., anesthesia machines)
Oxygen Therapy (e.g., oxygen tents)
The monitor is for use by trained health care professionals:
under the supervision, or on the order, of a physician
in a hospital or clinical setting
during emergency transport
General Description
fall below the preset (or default) low alarm setting
rise above the preset (or default) high alarm setting
®
The MiniOX
3000 Oxygen Monitor:
detects low and depleted battery conditions
activates audible and visual alarms
alarms for:
sensor disconnection or malfunction
various internal operating errors
®
The MiniOX
3000 Oxygen Monitor conducts self-
checks:
at power-up (battery installation)
at turn on
during operation
Additionally, the monitor has two operator-initiated test functions:
Battery-operated and microprocessor-controlled,
®
the MiniOX
3000 Oxygen Monitor measures oxygen concentrations in the 0% to 100% range. The monitor’s performance features ensure reliable and accurate oxygen measurement. These features include:
calibration function
high and low oxygen concentration alarms
low and depleted battery alarms
oxygen sensor indicator
automatic error detection
battery test
oxygen alarm test
The calibration function allows calibration of the device against room air (dened as oxygen concentration of 20.8%) or 100% O
. Audible and
2
visual alarms alert the operator when monitor calibration is required.
The Alarm Test veries the operation of the high and low oxygen level alarms
The Battery Test assesses the relative remaining battery life
®
The MiniOX
3000 Oxygen Monitor consists of two
components: the instrument and the oxygen sensor.
The front of the hand-held instrument features:
a touch-sensitive keypad
a liquid crystal display (LCD) that shows:
monitor status
continuous oxygen concentrations
preset alarm levels
two red light-emitting diodes (LEDs) which serve as visual alarms
The back of the instrument case features:
High and low oxygen concentration alarms may be set in the ranges of:
16% to 100% (high alarm) and
15% to 99% (low alarm)
or the default high/low settings may be used (50% and 18%, respectively). Audible and visual alarms activate when oxygen concentrations:
1-1
P/N 814358 [Rev.1] 09/2010
a bail bar to allow the instrument to “stand” on a horizontal surface during monitoring operations
a plastic wedge that slides into an optional bracket for mounting the instrument on a horizontal or vertical pole
The galvanic oxygen sensor consists of a deector assembly and a plastic housing containing two electrodes. A coiled cable connects the sensor to the
MiniOX® 3000 Oxygen Monitor Section 1, Introduction
instrument. Plugs at each end of the cable snap into jacks (one located in the sensor housing and one located in the instrument) and are held securely in place by twist collars.
The oxygen sensor is introduced into a breathing circuit through an Ohio Medical Corporation tee adapter connecting two lengths of tubing. The sensor/tee adapter assembly is positioned with the sensor deector pointing downward to ensure that moisture does not collect on the sensor membrane. The retaining strap ensures that the sensor remains securely in place in the tee adapter.
See the following appendices:
Appendix A, Specications
Appendix F, Accessories/Replacement Parts
Operating Principles
MiniOX® 3000 Oxygen Sensor
The oxygen sensor includes two electrodes:
provides the operator interface
controls internal functions
monitors data and system status
Operator Interface
When the operator enters commands using the keypad, the microprocessor interprets these commands and responds by:
displaying text in the LCD
activating the LEDs
emitting audible tones
Internal Functions
During operation, the microprocessor coordinates all internal functions including:
self-testing
response to keypad commands
data collection
display update
a gold cathode exposed to the atmosphere through a uoropolymer membrane
a lead anode submersed in a potassium hydroxide solution
When oxygen diffuses through the membrane, the electrochemical reduction of oxygen on the cathode and the corresponding oxidation on the anode generate an electrical current proportional to the partial pressure of oxygen in the sample atmosphere. The instrument temperature compensates, amplies, and converts the electrical current, displaying O
2
values.
The Ohio Medical Corporation oxygen sensor is self­zeroing; when no oxygen is present:
no current is produced
0% oxygen is displayed
The sensor has a minimal response to gases other than oxygen (see Appendix C, Interferent Gases and
Vapors).
conrmation that a programmed sequence is properly followed
Data and System Status
The signal from the oxygen sensor is amplied and converted to a digital value through electronic circuitry. Following software compensation, the microprocessor:
compares current data to preselected values
displays the updated O
concentration
2
if appropriate, activates audible and visual alarms
The microprocessor also monitors internal systems, indicating such operating conditions as:
calibration required
low battery voltage
sensor disconnect
system error
®
MiniOX
3000 Instrument
The MiniOX® 3000 Oxygen Monitor is battery­powered and microprocessor-controlled. The microprocessor:
Performance Features
Calibration
The MiniOX® 3000 Oxygen Monitor must be calibrated:
P/N 814358 [Rev.1] 09/2010
1-2
MiniOX® 3000 Oxygen Monitor Section 1, Introduction
daily, while in operation
each time the monitor is turned ON
following sensor disconnection/reconnection
when environmental conditions (pressure, humidity and temperature) change
When a MiniOX
®
3000 Oxygen Monitor with a sensor connected is turned ON, the monitor performs a self­test and a calibration requirement occurs:
“CAL” ashes on the display, telling the operator
to carry out the calibration procedure
The calibration procedure is initiated by exposing the sensor to a calibration gas [either room air (dened as oxygen concentration of 20.8%) or 100% oxygen] and pressing the corresponding key on the keypad. The following appears on the display:
“CAL LOCKED”
the selected calibration gas concentration.
When the unlock key is pressed the monitor displays:
“CAL”
a 10-segment bar graph which “counts down” two
seconds per bar for 20 seconds
The monitor automatically calibrates to the selected concentration. At the end of the 20 seconds, the monitor:
an audible alarm activates
the measured concentration appears in the display
The operator can silence the audible alarm for three 30-second intervals for a total of 90 seconds; however, the visual alarm continues to ash. At the end of the silence period, the audible alarm reactivates if the alarm condition is not corrected. See
Section 3, Silencing an Alarm
Low Battery Alarms
®
The MiniOX
3000 Oxygen Monitor features a two­stage battery alarm that warns of depleted and expired battery voltage:
The rst alarm alerts the operator that the monitor
has approximately six hours of operating time remaining:
a warning message appears in the display
an audible alarm sounds at 30-second intervals
If the operator does not replace the battery after this alarm, a second low battery alarm activates when the battery is no longer able to support monitoring. The monitor:
displays a warning message
activates an audible and visual alarm (See
Section 3, Low Battery Alarms)
emits a beep indicating that calibration is complete
enters the operating mode
See Section 3, Calibration
Low/High Oxygen Concentration Alarms
®
The MiniOX
3000 Oxygen Monitor has audible and visual alarms that activate when oxygen concentrations exceed preset low or high alarm settings. Default settings are 18% and 50%, respectively; however, the operator may select alarm levels between 15% and 100%. See Section 3,
Setting the Alarms
When the MiniOX® 3000 Oxygen Monitor detects an oxygen concentration that exceeds the preset alarm limit:
the red LED for that alarm ashes
1-3
P/N 814358 [Rev.1] 09/2010
Sensor Indicator
®
During monitoring, the MiniOX
3000 Oxygen Monitor displays a warning message; audible and visual alarms activate if:
the oxygen sensor becomes disconnected
the cable fails or disconnects from the sensor or
instrument
the sensor membrane is perforated
the thermistor circuit opens
Calibration Needed Indicator
®
The MiniOX
3000 Oxygen Monitor ashes “CAL”:
when the monitor is turned ON
following sensor disconnection/reconnection
MiniOX® 3000 Oxygen Monitor Section 1, Introduction
If a calibration is performed using a calibration gas other than room air or 100% oxygen, or if the microprocessor detects a calibration error during normal operation:
the display ashes a warning message: “CAL” and “ERR” ve times; then, “CAL” ashes
audible and visual alarms activate, alerting the operator that recalibration is required. See
Section 3, Calibration
Error Handling
®
The MiniOX
3000 Oxygen Monitor performs a self-
test
when it is turned ON
after a battery is installed (power-up)
during routine monitoring
If the microprocessor detects an error during these routine self-tests or during monitoring,
Test Functions
In addition to alarms that alert the operator to low and
®
expired battery voltage, the MiniOX
3000 Oxygen
Monitor features:
a keypad function that allows the operator to check the relative battery life at any time. See
Section 3, Test Functions
a test function that allows the operator to ensure that the high and low alarms activate at the preset alarm levels. See Section 3, Test Functions
Radio Frequency/Electromagnetic Insensitivity
®
The MiniOX
3000 Oxygen Monitor is designed to be insensitive to radio frequency interference (RFI) and electromagnetic interference (EMI). However, if the monitor comes in close contact with an RF transmitter or local electrical disturbance, an erratic reading may occur.
the display ashes:
“ERR”
an error code
audible and visual alarms activate
the monitor ceases to operate until the appropriate service is performed. See Appendix
E, Error Codes
To reset the monitor, press SET or turn the monitor OFF, then ON (press I/0 twice). For further details,
see Appendix E, Error Codes
P/N 814358 [Rev.1] 09/2010
1-4
MiniOX® 3000 Oxygen Monitor Section 2, Setting Up the Monitor
Section 2
®
Setting Up the MiniOX
3000 Oxygen
Monitor
The MiniOX® 3000 Oxygen Monitor is packaged in a shipping carton suitable for instrument storage. If there is obvious damage to the shipping carton or its contents, contact Ohio Medical Corporation Repair at:
866-549-6446, Option #1
To set up the monitor:
1. Verify that you have the following (see Figure 2-1):
®
MiniOX
Oxygen sensor and deector in sealed package
3000 Oxygen Monitor
10-foot coiled cable with twist collar
Tee adapter
Sensor retaining strap
9-volt alkaline battery
®
MiniOX
3000 Operation Manual CD
2. Check the sensor’s manufacturing date on the sensor package. If the sensor is put into service within six months after this date, the sensor will meet performance specications.
3. Remove the sensor from the sealed package and attach it to the coiled cable.
a. Firmly press the connector until it snaps into
place; tighten the twist collar.
b. Insert the opposite end of the coiled cable into
the jack on the side panel of the instrument; tighten the twist collar.
4. Remove the deector from the package.
SENSOR
RETAINING STRAP
TEE
ADAPTER
MINIOX® 3000
OXYGEN MONITOR
OXYGEN SENSOR AND DEFLECTOR
IN SEALED CONTAINER
9-VOLT
ALKALINE
BATTERY
TWIST
COLLAR
10 FT
COILED
CABLE
2-1
Figure 2-1
MiniOX® 3000 Oxygen Monitor
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Section 2, Setting Up the Monitor
a. Insert the gasket into the open end of the
deector, ensuring that the gasket is properly seated within the deector
b. Gently screw the deector onto the sensor
(For monitoring in a breathing circuit, see
Section 3, Installing the Sensor in a Breathing Circuit)
5. Unscrew the two battery cover screws on the back of the instrument.
a. Remove the cover
b. Install the battery (See Section 4, Battery
Replacement for full battery installation instructions)
After battery installation:
the display ashes
the instrument performs a self-test
then turns OFF
6. Proceed to Section 3, Operation
W CAUTION
Do not handle the sensor unnecessarily during calibration or use. Body heat can cause the sensor’s thermistor to change disproportionately to the change in gas sample temperature at the sensing electrode. This can produce some error until thermal equilibrium is restored.
See Appendix F, Accessories/Replacement Parts
®
for ordering MiniOX
3000 Monitor accessories and
replacement parts (Figure 2-2).
AUDIBLE ALARMS
LED ALARMS
DISPLAYED OXYGEN READING
HIGH ALARM POINT SETTING
LOW ALARM POINT SETTING
ON/OFF KEY
BATTERY TEST AND ALARM
TEST KEY
21% CALIBRATION KEY
UNLOCK KEY (ENABLES CALIBRATION SEQUENCE)
DECREASE ALARM SET POINT KEY
SET KEY
(ENABLES ADJUSTMENTS OF
ALARM SET POINTS)
100% CALIBRATION KEY
DEFLECTOR
SENSOR
CABLE ASSEMBLY
MONITOR
STAND
SILENT KEY
INCREASE ALARM SET POINT KEY
Figure 2-2
Assembled MiniOX® 3000 Oxygen Monitor
P/N 814358 [Rev.1] 09/2010
2-2
MiniOX® 3000 Oxygen Monitor Section 3, Operation
Section 3 Operation
This section describes the following operational procedures and functions of the MiniOX® 3000 Oxygen Monitor:
Calibration
Two-Point Linearity Check
Setting the Alarms
Alarm Conditions
Silencing an Alarm
Test Functions
Installing the Sensor in a Breathing Circuit
Mounting the Instrument with a Bracket
Calibration
®
The MiniOX calibration:
3000 Oxygen Monitor requires
when the operating environment changes
after the monitor is turned ON
if the sensor is disconnected and reconnected to the instrument
Recommended practice of calibration is against room air (dened as oxygen concentration of 20.8%). When a more precise measurement is desired, repeat calibration using 100% oxygen concentration. Changes in pressure, humidity or temperature may affect calibration accuracy (see Appendix B, Effects of Pressure, Humidity, and Temperature).
W CAUTION
Do not handle the sensor unnecessarily during calibration or use. Body heat can cause the sensor’s thermistor to change disproportionately to the change in gas sample temperature at the sensing electrode. This can produce some error until thermal equilibrium is restored.
To Calibrate In Room Air (Figure 3-1):
daily, while in operation
EXPOSE SENSOR
1
2
3
TO ROOM AIR
PRESS I/O
PRESS 21%
CALIBRATE
1. Expose sensor to room air.
PRESS UNLOCK
AFTER 20 SECOND
CALIBRATION COMPLETE
AND
20.8% +2% DISPLAYED
4
5
3-1
Figure 3-1
Calibrating the MiniOX® 3000 Oxygen Monitor in Room Air
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Section 3, Operation
2. Press I/0 to turn ON the instrument.
“CAL” ashes in the display
3. Press 21%. The following appears on the display:
“CAL”
“LOCKED”
“21% CAL”
4. Press UNLOCK. The following appears:
“CAL”
“21% CAL”
a 10-segment bar graph that “counts down” two seconds per bar for 20 seconds
5. After 20 seconds, the calibration process is complete the device:
displays 20.8% ± 2% O
(18.8% to 22.8%)
2
proceeds to the monitoring mode
displays the current oxygen concentration as %O
2
To Calibrate at 100% O
(Figure 3-2)
2
Note: Prior to calibrating at 100% O2, the MiniOX®
3000 Oxygen Monitor must rst be calibrated in room air.
1. Calibrate in room air (see Section 3, To Calibrate
in Room Air)
2. Expose the sensor to 100% oxygen and allow the readings to stabilize prior to initiating the calibration.
“CAL” ashes in the display.
3. Press 100%. The following appears on the display:
“CAL”
“LOCKED”
“100% CAL”
4. Press UNLOCK. The following appears on the display:
“CAL”
“100% CAL”
a 10-segment bar graph that “counts down” two seconds per bar for 20 seconds.
5. After 20 seconds, the calibration process is complete; the device:
displays 100.0% +0/-2% (98% to 100%)
proceeds to the monitoring mode
displays the current oxygen concentration as %O
2
CALIBRATE IN
1
2
3
ROOM AIR
EXPOSE SENSOR
TO 100% OXYGEN
PRESS 100%
CALIBRATE
PRESS UNLOCK
AFTER 20 SECOND
CALIBRATION COMPLETE
AND
100% +0/-2% DISPLAYED
4
5
Figure 3-2
Calibrating the MiniOX® 3000 Oxygen Monitor at 100% O
P/N 814358 [Rev.1] 09/2010
2
3-2
MiniOX® 3000 Oxygen Monitor Section 3, Operation
NOTE: The MiniOX® 3000 Oxygen Monitor has a
ve second “time out” following keypad functions. If you do not press UNLOCK within ve seconds, the instrument returns to the ashing “CAL” mode.
NOTE: During operation if “CAL ERR”
ashes in the display, visual, audible alarms activate, and then “CAL” ashes, turn OFF the instrument and repeat calibration procedure. When recalibrating, be sure to select the calibration value and use the corresponding calibration gas. If “CAL ERR” reoccurs, it may be necessary to replace the sensor. (See Section 4, “Sensor Replacement”)
NOTE: During operation if “CAL” displays, you
must recalibrate the monitor. If “CAL” displays following proper recalibration, it may be necessary to replace the sensor. (See Section 4, Sensor Replacement)
Two-Point Linearity Check
To evaluate sensor performance, conduct a two-point linearity check on the monitor every week or any time you suspect that the sensor is not accurately reading oxygen concentrations. Acceptable range for room air measurement during a linearity check is:
two-point linearity check
If the variation is still greater than ±2%, replace the sensor. (See Section 4, Sensor
Replacement)
Setting the Alarms
The default Alarm Set points are:
18% for Low Alarm (appears in the lower left corner of the display)
50% for High Alarm (appears in the lower right corner of the display)
These alarm points can be reset between 15% and100%.
®
NOTE: The MiniOX
3000 Oxygen Monitor retains current alarm settings when turned OFF, if the alarm settings are not below 18%. However, if the alarm settings are lower than 18% when the MiniOX® 3000 Oxygen Monitor is turned OFF or following battery replacement, alarm set points return to the default settings of 18% and 50%.
To Set the Low Alarm
1. Press SET once. The following appears on the display:
20.8% ±2% (18.8% to 22.8%). This variation allows for differences in:
sampling methods
accuracy of gas concentration
precision of initial setting
If linearity deviates more than ±2%, the sensor is nearing end-of-life and should be replaced.
To perform a two-point linearity check:
1. Calibrate in room air.
2. Place the sensor in a stream of oxygen with a known concentration of 100% until the reading stabilizes.
3. Recalibrate the monitor.
4. After calibration, measure room air. This reading should be 20.8% ±2% (18.8% or 22.8%) within ve minutes.
If the variation is greater than ±2%, repeat the
“AL”
up/down arrows
2. Using the arrow keys, scroll up or down to the desired Low Alarm set point (15% to 99%).
®
3. The MiniOX
3000 Oxygen Monitor “locks” this
value. After ve seconds, the monitor:
beeps once
automatically proceeds to the Monitoring Mode
NOTE: The Low Alarm CANNOT be disabled or
set:
below 15%
above 99%
higher than or equal to the High Alarm setting
To Set the High Alarm
1. Press SET twice. The following appears on the display:
3-3
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Section 3, Operation
“AL”
up/down arrows
2. Using the arrow keys, scroll up or down to the desired High Alarm set point (16% to 100%).
®
3. The MiniOX
3000 Oxygen Monitor “locks” this
value. After ve seconds, the monitor:
beeps once
automatically proceeds to Monitoring Mode (Press SET once after Step 2 to manually proceed to Monitoring Mode)
NOTE: The HIGH Alarm value:
CANNOT be set equal to, or less than, the Low Alarm value
CAN be disabled by increasing the alarm set point beyond 100% until “--” displays
Alarm Conditions
High and Low Oxygen Concentration Alarms
®
When the MiniOX
3000 Oxygen Monitor detects an oxygen concentration that exceeds the preset alarm limit:
the red LED ashes for that alarm (High or Low)
an audible alarm activates
the measured concentration displays
Silencing an Alarm
If the alarm condition is not corrected within the preset silence interval (30,60, or 90 seconds):
the audible alarm reactivates
If the alarm condition is corrected and then reoccurs within this interval:
both audible and visual alarms activate
2. To manually terminate the silence mode, press the down arrow (labeled with a ▼ icon).
Low Battery Alarms
®
The MiniOX
3000 Oxygen Monitor features a two­stage battery alarm that warns of depleted and expired battery voltage:
The rst alarm alerts the operator that the monitor
has approximately six hours of operating time remaining:
“LOW BAT” appears on the display
an audible alarm sounds at 30-second intervals
If the operator does not replace the battery after this alarm, a second low battery alarm activates when the battery is no longer able to support monitoring.
the monitor displays “LOW BAT” and “---”
audible and visual alarms are activated
The MiniOX® 3000 Oxygen Monitor features a silence function that allows the operator to temporarily mute the O2 concentration audible alarm.
1. When an audible alarm activates, press the silent key (labeled with a horn icon) to mute the alarm for up to three 30-second intervals, press:
once for a 30-second interval
twice (within two seconds) for a 60-second interval
three times (within four seconds) for a 90-second interval
The following appears on the display:
a horn icon
a three-bar graph showing a “countdown” of 10 seconds per bar for each 30 second interval
NOTE: When the audible alarm is silenced, the
appropriate visual alarm continues to ash.
P/N 814358 [Rev.1] 09/2010
For periodic checks on battery status, see Section 3,
Test Functions”
Sensor Indicator
®
During monitoring, the MiniOX
3000 Oxygen Monitor displays “SENSOR” and “OFF”; audible and visual alarms activate if:
the oxygen sensor becomes disconnected
the cable fails or disconnects from the sensor or
instrument
the sensor membrane is perforated
the thermistor circuit opens
3-4
MiniOX® 3000 Oxygen Monitor Section 3, Operation
Error Handling
The MiniOX® 3000 Oxygen Monitor performs a self­test
when it is turned ON
after a battery is installed
If the microprocessor detects an error during these routine self-tests or during monitoring:
the display ashes “ERR” and an error code
audible and visual alarms activate
the monitor ceases to operate until the appropriate service is performed; to reset the
Table 3-1 Alarms and Status Messages
ALARM CAUSE CORRECTIVE ACTION
alarm:
Low O
2
O2 concentration appears, visual alarm ashes, and audible alarm activates
High O O2 concentration appears, visual alarm ashes, and audible alarm activates
“SENSOR OFF” displays; visual and audible alarms activate
During calibration, “CAL ERR” ashes, visual and audible alarms activate; then “CAL” ashes
During calibration, “CAL” ashes
“LOW BAT” displays and mon­itor beeps every 30 seconds
“LOW BAT” displays “---” ashes, both visual alarms activate, and monitor emits a three-tone beep for four min­utes; monitor shuts OFF
“ERR” and error code (01 to
08) display, visual and audible alarms activate and monitor locks
Alarm:
2
Oxygen concentration is below the preset Low Alarm setting
Oxygen concentration is above the present High Alarm setting
Cable is disconnected or malfunc­tioning
Sensor is disconnected Visually check cable for damage. If necessary, replace cable
Sensor membrane is perforated If the alert persists, replace sensor
Thermistor circuit is open on the sensor circuit board
Calibration performed using a gas other than room air or 100% oxygen or wrong calibration key pressed (100%/21%)
Sensor signal is out of monitoring range
Monitor has approximately six hours of operating time remaining
Battery is expired and monitor is not operating
Microprocessor detects internal error
monitor, press SET or turn the monitor OFF, then ON (press I/0 twice). For further details see Appendix E, Error Codes
NOTE: During normal operation, if a key is pressed
and held longer than nine seconds, audible and visual alarms activate.
1. To reset the monitor, press SET or turn the monitor OFF, then ON (Press I/0 twice). For further details see Appendix E, Error Codes
®
Table 3-1 is a summary of MiniOX
3000 Oxygen
Monitor alarms and status messages.
Check patient and delivered oxygen concentration
Verify that low alarm setting is appropriate
Check patient and delivered oxygen concentration
Verify that high alarm setting is appropriate
Check cable and sensor connections
If alert persists following sensor replacement, call Ohio Medical Corporation service
Recalibrate the monitor, ensure proper calibration key pressed for calibration gas of room air or 100% oxygen
If the alert persists, replace sensor
If alert persists following sensor replacement, call Ohio Medical Corporation service
Recalibrate the monitor, ensure proper calibration key pressed for calibration gas of room air or 100% oxygen
If the alert persists, replace the sensor
If alert persists following sensor replacement, call Ohio Medical Corporation service
Replace the battery as soon as possible; then, recalibrate the monitor and reset the high and low alarm values
Replace the battery as soon as possible; then recalibrate the monitor and reset the high and low alarm values
Note error code and see Appendix E, Error Codes
Disconnect Battery
Call Ohio Medical Corporation Service
3-5
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Section 3, Operation
Test Functions
Battery Test
In addition to alarms alerting the operator to low and expired battery voltage, the MiniOX® 3000 Oxygen Monitor features a keypad function allowing the operator to check the relative battery life at any time. To test battery status:
1. Press TEST once. The following appears on the display:
“BATTERY TEST”
a 10-bar graph showing relative battery life. Ten bars indicate a fully charged battery; one bar indicates a battery near depletion
®
2. After ve seconds, the MiniOX Monitor proceeds to the Monitoring mode, displaying oxygen concentrations as %O2.
High and Low Alarms Test
3000 Oxygen
Installing the Sensor in a Breathing Circuit
To use the sensor in a breathing circuit, you will need the following:
Sensor (with deector)
Tee adapter
Retaining Strap
To install the sensor in a breathing circuit:
1. Install the tee adapter into the breathing circuit upstream from the humidier. Make sure that side port of the tee adapter is facing upward.
2. Remove the coiled cable from the sensor.
3. Firmly insert the sensor (with deector) into the tee adapter with the deector pointing downward (Figure 3-3) to prevent moisture from condensing onto the sensor membrane (See Appendix B, Effects of Pressure, Humidity, and Temperature). Make sure that the sensor ts tightly into the tee adapter.
®
The MiniOX
3000 Oxygen Monitor features a test function to ensure that the high and low alarms activate at the preset alarm levels. To test the alarms:
Press TEST twice.
“ALARM TEST” displays
the instrument automatically scrolls up to the high
alarm value and ashes this value on the display
the audible alarm and visual high alarm activate
(If the high alarm value is disabled, the instrument scrolls down from 100% to the low alarm value)
the instrument then automatically scrolls down to
the low alarm value and ashes this value on the display
the audible alarm and visual low alarm activate
®
the MiniOX
3000 Oxygen Monitor proceeds to the Monitoring Mode, displaying oxygen concentrations as %O
NOTE: The MiniOX
2
®
3000 Oxygen Monitor is not monitoring during the Alarm Test which may take up to 30 seconds to complete.
NOTE: The Alarm Test will not function if the
MiniOX® 3000 Oxygen Monitor is in an alarm condition.
4. Install one end of the retaining strap over a post on the side of the tee adapter.
5. Loop the strap over the sensor, inserting the strap center hole over the sensor cable jack.
6. Install the other end of the strap over the other tee adapter post.
7. Re-attach the coiled cable to the sensor. Tighten the twist collar.
The sensor is securely placed for monitors (See Figure 3-4).
Mounting the Instrument with a Pole Bracket
®
The MiniOX
3000 Oxygen Monitor (as shown in Figure 3-5) can be attached to a vertical or horizontal pole using the Ohio Medical Corporation Mounting Bracket (See Appendix F, Accessories/Replacement Parts). To mount the MiniOX® 3000 Oxygen Monitor:
1. Congure the bracket for either horizontal or
vertical attachment by adjusting the dovetail mounting plate on the bracket.
2. Slide the mounting bracket onto the pole; secure
by tightening the screw knob.
3. Slide the wedge on the back of the monitor into
the dovetail mounting plate.
P/N 814358 [Rev.1] 09/2010
3-6
MiniOX® 3000 Oxygen Monitor Section 3, Operation
SENSOR CABLE
RETAINING STRAP
SENSOR
GASKET
DEFLECTOR
O-RING
15mm TEE
3-7
Figure 3-3
Installing Sensor into Tee Adapter
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Section 3, Operation
SENSOR CABLE
RETAINING STRAP
SENSOR
DEFLECTOR
15mm TEE
ADAPTER
FASTEN RETAINING STRAP
TO POSTS ON TEE
Figure 3-4
Sensor in Breathing Circuit
INSTRUMENT
MOUNTING
DOVETAIL
SLIDE DOVETAIL
COMPLETELY INTO
BRACKET
MOUNTING
BRACKET
INSTALLED ON
POLE
Figure 3-5
MiniOX® 3000 Oxygen Monitor with Pole Mounting Bracket
P/N 814358 [Rev.1] 09/2010
3-8
MiniOX® 3000 Oxygen Monitor Section 4, Maintenance and Care
Section 4 Maintenance and Care
W WARNING
Use only genuine Ohio Medical Corporation replacement parts when performing any maintenance procedures included in this manual. Failure to do so may seriously impair the monitor’s performance. Repair or alteration of the MiniOX® 3000 Oxygen Monitor beyond the scope of the maintenance instructions or by anyone other than an authorized Ohio Medical Corporation service person could cause the product to fail to perform as designed.
This section describes the following general maintenance and care procedures for the MiniOX® 3000 Oxygen Monitor:
Battery Replacement
Sensor Replacement
Deector Replacement
Cable Replacement
Cleaning
Battery & Battery Cover Replacement
®
The MiniOX
3000 Oxygen Monitor requires one
standard 9-volt alkaline battery. To replace the battery:
1. Verify that the monitor is turned OFF. The display
should be blank.
2. Pull out the support stand from the back of the
case.
3. Unscrew the two screws on the battery cover in
the back of the instrument and remove cover. See Figure 4-1.
4. Remove the battery from the case and unsnap
the battery from the battery holder.
NOTE: To ensure proper start-up, wait at least
45 seconds before connecting the fresh battery to the battery connector.
5. Snap the terminal of the new battery into the
battery holder.
6. Install the battery cover and screw into place.
Make sure that the battery cover is properly
®
seated and at on the back of the MiniOX
3000
Oxygen Monitor case.
7. Recalibrate the monitor. Reset the low and high
alarms, if desired.
NOTE: To maximize battery life, press I/O to
®
turn OFF the MiniOX
3000 Oxygen Monitor when not monitoring. In order to retain alarm settings, do not remove battery .
4-1
Figure 4-1
Battery Cover Removal
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Section 4, Maintenance and Care
Sensor Replacement
1. Verify that the monitor is turned OFF. The display To ensure safe and effective use of your device, the sensor must be replaced with a MiniOX® sensor as
should be blank.
2. Disconnect the old cable from the sensor. this sensor is manufactured for this instrument. Use
of other types of sensors has not been tested and is not endorsed by Ohio Medical Corporation. Use of other sensor types will void your warranty.
3. Disconnect the old cable from the instrument.
4. Attach the new cable to the sensor. Firmly press
the connector until the sensor snaps into place. Tighten the twist collar.
There are no serviceable parts in the sensor or cable assemblies, the entire unit must be replaced.
Replace the sensor when:
5. Attach the new cable to the instrument. Firmly
press the connector until it snaps into place. Tighten the twist collar.
6. Recalibrate the instrument.
Room air reading is greater than 20.8% ±2%
(18.8% to 22.8%) in Two-Point Linearity Check
®
The MiniOX
3000 Oxygen Monitor will not
Cleaning, Disinfection and Sterilization
W CAUTION
calibrate.
“Sensor” and “OFF” display and audible and
visual alarms persist when sensor and cable connections are correct and cable is viable.
Never autoclave, immerse, or expose the
®
MiniOX
3000 Oxygen Monitor (including sensor) to high temperatures (>70oC). Never expose the device to pressure, irradiation, vacuum, steam, or chemicals (other than
To replace the sensor:
alcohol or mild cleaning agents).
1. Verify that the monitor is turned OFF. The display should be blank.
2. Disconnect the expired sensor from the coiled cable.
3. Attach a new sensor to the coiled cable and rmly press the connector until the sensor snaps into place. Tighten the twist collar.
4. Recalibrate the monitor.
Deector Replacement
To replace the deector:
1. Gently unscrew the old deector and remove the gasket.
2. Insert a new gasket into the large end of the new deector. Carefully screw the new deector onto the sensor, making sure that the new gasket is properly seated.
See Figure 3-3 for proper sensor and deector assembly.
Cable Replacement
Clean the instrument and sensor by wiping with a cloth lightly dampened with Isopropyl Alcohol or mild detergent. Make sure that no moisture seeps into the instrument case or cable jack port.
Instrument
When cleaning or disinfecting the instrument, care must be taken to prevent entry of solutions into the instrument case. If it is suspected that solutions or moisture enters the case, verify performance by conducting a self-test (see Possible physical damage to instrument, Appendix D).
Cleaning
The external surfaces of the unit may be cleaned by wiping them with a cloth moistened with a mild detergent solution.
Disinfection
The external surfaces of the unit may be disinfected by wiping them with a cloth moistened with ethanol or Cidex. The instrument is not designed to withstand the conditions imposed by steam, ethylene oxide or radiation sterilization.
To replace the coiled cable:
P/N 814358 [Rev.1] 09/2010
4-2
MiniOX® 3000 Oxygen Monitor Section 4, Maintenance and Care
Sensor and Cable
Cleaning
The external surfaces of the oxygen sensor and of the cable may be cleaned by wiping them with a cloth moistened with a mild detergent solution.
Disinfection
The external surfaces of the oxygen sensor housing and of the cable may be disinfected by wiping them with a cloth moistened with ethanol or Cidex. The instrument is not designed to withstand the conditions imposed by steam, ethylene oxide or radiation sterilization.
Sensor Deector, Retaining Strap and Tee Adapter
Cleaning
The sensor deector, retaining strap and tee adapter may be cleaned by wiping them with a cloth moistened with a mild detergent solution. The parts must be thoroughly dry before they are used.
Sterilization
The sensor deector, retaining strap and tee adapter may be sterilized using Cidex (per manufacturer’s instructions), steam or ethylene oxide. Due to the varying conditions imposed on materials during sterilization, it is not possible to determine the exact number of times sterilization processes can be carried out. Therefore, the operator must carefully examine the sensor deector, retaining strap and tee adapter after sterilization and prior to use to verify that the item is t for use. The operator must verify that the items are free from tears and cracks and that the items have not undergone any material changes that may compromise their tness for use (e.g., brittleness and dimensional changes). The operator must also examine the items to verify that the items are free of chemical residuals resulting from the sterilization process.
Because of the variability of cleaning, disinfection and sterilization processes, Ohio Medical Corporation cannot provide specic sterilization instructions, nor can the sterility of an item be ensured.
Disinfection
The sensor deector, retaining strap and tee adapter may be disinfected by washing them with ethanol or Cidex (per manufacturer’s instructions). The parts must be thoroughly dry before they are used.
4-3
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Appendix A, Specications
Appendix A Specications
Table A-1. Specications
ALARM SYSTEM
O
2
LOW/HIGH O ALARM
ACCURACY
LINEARITY
NOMINAL RESPONSE TIME
directed at sensor at 2 liters per
(O
2
minute at RTP)
SENSOR LIFE Over one year in normal medical conditions (nominal 750,000 O
SENSOR SHELF LIFE Six months minimum (stored in sealed package)
BATTERY LIFE Approximately 1500 device hours, assuming 80% ON, 20% OFF
OPERATING TEMPERATURE RANGE 0
STORAGE TEMPERATURE -40
HUMIDITY 5% to 95% RH (non-condensing)
DIMENSIONS
WEIGHT
BATTERY ALARM SYSTEM
INSTRUMENT 5.98” X 3.26” X 1.31”
SENSOR 1-1/5” OD x 1-3/4” long
INSTRUMENT 260 g (9.2 oz)
SENSOR 35 g (1.25 oz)
STAGE 1
STAGE 2
*RTP: Room temperature and pressure, e.g., 23oC + 3o and ambient barometric pressure
2
Pulsating alarm, ashing LEDs
+ 1% of full scale (RTP*) as calibrated in the environment of use (total accuracy, including linearity, is 2%)
+1% of full scale (RTP*)
90% in 20 seconds
97% in 30 seconds
hours)
2
o
C to 40oC (32oF to 104oF)
o
C to 70oC (-40oF to 158oF)
Warning that approximately six hours of monitoring life remain; LOW BAT appears on dis­play; beep sounds every 30 seconds until battery is replaced or Stage 2 occurs.
Alarm that proper operation is not assured; monitor shuts down, audible and visual alarms are activated.
P/N 814358 [Rev.1] 09/2010
A-1
MiniOX® 3000 Oxygen Monitor Appendix B, Effects of Pressure, Humidity and Temperature
Appendix B Effects of Pressure, Humidity and Temperature
W CAUTION
To ensure accurate and reliable oxygen monitoring, it is necessary to have a thorough understanding of the effects of pressure, humidity, and temperature on the sensor.
Effects of Pressure
The sensor responds to partial pressure (not percentage) of oxygen. Changes in barometric pressure change the reading, even if the percent of oxygen in the sample remains constant.
Partial pressure of oxygen (PO of oxygen (%O
) times the pressure at which the
2
sample is measured (mmHg).
PO2 = (%O2 ) (mmHg)
For example: at sea level, the pressure equals 760 mmHg and dry air contains 21% O2. Therefore:
PO2 = (21%) (760 mmHg)
) equals the percent
2
W CAUTION
Do not expose the sensor to pressure outside the range of 600 to 900 mmHg (23.62 to 35.43 Inches Hg), as this may cause inaccuracies.
Effects of Humidity
The presence of humidity in an oxygen sample decreases the actual concentration of oxygen. Humidity in a sample has the same effect as diluting the sample with another gas. For example, if 100% oxygen is saturated with 100% humidity, the actual concentration of oxygen drops to 96% - 97%.
As with all oxygen gas sensors, condensation on the sensor membrane blocks the ow of oxygen, resulting in a lower oxygen concentration reading and an increased response time. This is a typical problem resulting from locating the sensor downstream from the humidier in an oxygen delivery system; clearing the sensor face and deector restores normal operation. However, to avoid this problem when using the sensor in a breathing circuit, position the sensor upstream of the humidier and mount the sensor with the deector pointing downward to prevent moisture from draining onto the sensor membrane.
= 160 mmHg
PO
2
If you calibrate an instrument to read 21% at 760 mmHg partial pressure and then take the instrument to an area above sea level, a lower reading occurs due to a lower partial pressure. For a pressure of 700 mmHg:
PO
= (21%) (700 mmHg )
2
=147 mmHg
PO
2
The percent reading on the instrument is derived from the following formula:
PO
PO
When PO
Actual = 147 mmHg
2
Sea level = 160 mmHg
2
sea level is 21%
2
(21 %) (147 mmHg)
Actual = (160 mmHg) = 19.3%
PO
2
Therefore, to eliminate error due to pressure changes, the instrument must be calibrated at the pressure in which it is used.
B-1
P/N 814358 [Rev.1] 09/2010
Effects of Temperature
Due to an internal thermistor (temperature variable
®
resistor), the MiniOX
3000 sensor is minimally affected by temperature change. Variations in the sensor reading from temperature change are less than 3% when the instrument is calibrated and used in a monitoring environment of 0oC to 40°C (32°F to 104°F).
Operating Temperature Range:
0o to 40oC (32o to 104oF)
W CAUTION
Do not use instrument outside operating temperature range.
W CAUTION
Do not handle the sensor more than necessary during calibration or use. Body heat can cause the sensor’s thermistor to change disproportional to the change in gas sample temperature at the sensing electrode. This can produce some error, until thermal equilibrium is restored.
MiniOX® 3000 Oxygen Monitor Appendix C, Interferent Gases and Vapors
Appendix C Interferent Gases and Vapors
Table C-1. Interferent Gases and Vapors
Interferent VOLUME % DRY INTERFERENCE EQUIVALENT OF % O
Carbon Dioxide 10% 0.1%
Cyclopropane 50% 0.2%
Desurane 7.5% 0.1%
Diethyl Ether 20% 0.3%
Enurane 4% 0.1%
Halothane 5% 0.2%
Helium 80% 0.2%
Isourane 3% 0.5%
Methoxyurane 4% 0.1%
Nitrogen 80% 0.2%
Nitrous Oxide 80% 0.1%
2
P/N 814358 [Rev.1] 09/2010
C-1
MiniOX® 3000 Oxygen Monitor Appendix D, Troubleshooting
Appendix D Troubleshooting
Table D-1. Troubleshooting
PROBLEM CAUSE CORRECTIVE ACTION
Press I/O to turn ON monitor
If monitor does not respond, replace the battery
If problem persists, call Ohio Medical Corporation service
Press I/O to run ON monitor
If monitor does not respond, replace the battery
If problem persists, call Ohio Medical Corporation service
If alert persist following sensor replacement, call Ohio Medical Corporation service
Recalibrate the monitor, ensure proper calibration key pressed for calibration gas of room air or 100% oxygen
If alert persists, replace sensor
If alert persists following sensor replacement, call Ohio Medical Corporation service
Recalibrate the monitor, ensure proper calibration key pressed for calibration gas of room air or 100% oxygen
If alert persists, replace sensor
If alert persists following sensor replacement, call Ohio Medical Corporation service
Replace the battery as soon as possible; then recalibrate the monitor and reset the high and low alarm values
Replace the battery as soon as possible; then recalibrate the monitor and reset the high and low alarm values
Note error code and see Appendix E, Error Codes
Disconnect battery
Call Ohio Medical Corporation service
Turn instrument OFF, then ON to initiate self-test. If instrument is damaged, “ERR” and error code (01 to 08) display. Disconnect battery. Call Ohio Medical Corporation service
Display is blank
No Response to keypad commands
“SENSOR OFF” displays: visual and audible alarms activate
During calibration, “CAL ERR” ashes, visual and audible alarms activate; then “CAL” ashes
During operation, “CAL” ashes
‘LOW BAT” displays and monitor beeps every 30 seconds
“LOW BAT” displays “---” ashes, both visual alarms activate, and monitor emits a three-tone beep for four minutes; monitor shuts OFF
“ERR” and error code (01 to 08) display, visual and audible alarms activate and monitor locks
Possible physical dam­age to instrument
Monitor is turned OFF or Battery is ex­pired.
Battery is expired
Cable is disconnected or malfunctioning Check cable and sensor connections
Sensor is disconnected Visually check cable for damage. If necessary, replace cable
Sensor membrane is perforated If alert persists, replace sensor
Thermistor circuit is open on the sensor circuit board
Calibration performed using a gas other than room air or 100% oxygen or wrong calibration key pressed (100%/21%)
Sensor signal is out of monitoring range
Monitor has approximately six hours of operating time remaining
Battery is expired and monitor is not oper­ating
Microprocessor detects internal error
Instrument is dropped and/or exposed to uids
D-1
P/N 814358 [Rev.1] 09/2010
MiniOX® 3000 Oxygen Monitor Appendix E, Error Codes
Appendix E Error Codes
The MiniOX® 3000 Oxygen Monitor performs diagnostic tests to detect errors that could cause unreliable monitor operation. These tests are performed:
when a battery is installed (power-up)
when the monitor is turned ON
during routine monitoring
During self-test or any time the monitor detects an operational error, the MiniOX
ceases operation
Display “ERR” and an error code
activates visual and audible alarms
®
3000 Oxygen Monitor;
Error alarms continue until:
1. The monitor is shut off (press I/O)
2. The monitor is reset and error does not reoccur
(press set)
3. The battery is disconnected
If an error reoccurs, the monitor is inoperative and must not be used; call Ohio Medical Corporation Repair at:
866-549-6446, Option #1
The error code that appears on the display corresponds to a specic failure. The audible alarm is a beep that corresponds to this code (e.g., “ERR 3” triggers a three-beep alarm).
Table E-1. Error Codes
ERROR DISPLAY CORRESPONDING ERROR
01 System error
02 Random Access Memory (RAM) error
03 Read Only Memory (ROM) error
04 Analog error
05 Timing error
06 Keypad error*
07 Battery error **
08 LCD error
*Press SET. If error reoccurs, call Ohio Medical Corporation Repair
**Replace battery. If error reoccurs, call Ohio Medical Corporation Repair
P/N 814358 [Rev.1] 09/2010
E-1
MiniOX® 3000 Oxygen Monitor Appendix F, Accessories and Replacement Parts
Appendix F Accessories and Replacement Parts
Table F-1, Accessories and Replacement Parts
ITEM PART NO.
Oxygen Sensor 406931
Mounting Bracket (Pole) 474664
Mounting Bracket (Wall) 10023945
Coiled Cable 472045
Tee Adapter 473021
Retaining Strap 634249
Operation Manual on CD 600700
Operation Manual, printed 814358
®
MiniOX
Battery Cover Assembly 710974
Battery Cover Screw, Knurled 655579
Deector 803229
Dovetail 474606
Screw, #4x1/4” (set of 4) thread forming 450001
Inspection and Servicing SVC-301
Cleaning, Calibration and Testing SVC-302
Enclosure Replacement SVC-303
Circuit Board Replacement SVC-304
Connector Replacement SVC-305
3000 Oxygen Monitor 814365
Servicing
F-1
P/N 814358 [Rev.1] 09/2010
NOTES
P/N 814358 [Rev.1] 09/2010
Ohio Medical Corporation® 1111 Lakeside Drive Gurnee, IL 60031 USA phone: 1-866-549-6446 fax: 1-847-855-6218 www.ohiomedical.com
P/N 814358 [Rev.1] 09/2010
Loading...