Ohio 7300 Service manual

Schaerer Mayfield USA, Inc.
Go To Table Of Contents
Modular Surgery
Table
Service and
Parts Manual
SMI
7300
SF-1586 Part No. 504-0023-00 Rev. E (5/04)
CA802000
FOR USE BY SMI TRAINED TECHNICIANS ONLY
Go To Table Of Contents
TABLE OF CONTENTS
Section/Paragraph Page Section/Paragraph Page
IMPORTANT INSTRUCTIONS
General Safety Instructions........................................ iv
Warnings.................................................................... iv
Warranty Instructions ................................................. iv
SECTION I GENERAL INFORMATION
1.1 Scope of Manual ......................................... 1-1
1.2 How to Use Manual ..................................... 1-1
1.3 Description of 7300 Modular Surgery
Table ....................................................... 1-1
1.4 Specifications ............................................. 1-7
1.5 Parts Replacement Ordering ....................... 1-8
1.6 Special Tools .............................................. 1-9
SECTION II TESTING AND TROUBLESHOOTING
2.1 Operational Test.......................................... 2-1
2.2 Troubleshooting Procedures ........................ 2-6
SECTION III SCHEDULED MAINTENANCE
3.1 Scheduled Maintenance .............................. 3-1
SECTION IV MAINTENANCE/SERVICE INSTRUCTIONS
4.1 Introduction ................................................. 4-1
4.2 Checking Oil Level / Adding Oil To
Reservoir .................................................. 4- 1
4.3 Calibration Of Return To Level / Neutral
Position .................................................... 4- 2
4.4 Hand Control Button Board
Removal / Installation ............................... 4-4
4.5 Hand Control Board
Removal / Installation ............................... 4-4
4.6 Seat Cylinders Synchronization
Procedure ................................................. 4- 5
4.7 Column Cylinder Slave Line Volume
Adjustment ............................................... 4- 7
4.8 Trendelenburg Cylinder
Removal / Installation ............................... 4-9
4.9 Trendelenburg Pilot Operated Check Valve
Removal / Installation ............................. 4-11
4.10 Trendelenburg Cylinder Orifice(s)
Removal / Installation ............................. 4-13
4.11 Trendelenburg Position Sensor
Removal / Installation ............................. 4-14
4.12 Lateral Tilt Cylinder
Removal / Installation ............................. 4-15
4.13 Lateral Tilt Orifice(s)
Removal / Installation ............................. 4-16
4.14 Lateral Tilt Pilot Operated Check Valve
Removal / Installation ............................. 4-18
4.15 Lateral Tilt Position Sensor
Removal / Installation ............................. 4-20
4.16 Seat Cylinder
Removal / Installation ............................. 4-21
4.17 Seat Cylinder Orifice(s)
Removal / Installation ............................. 4-26
4.18 Seat Cylinder Pilot Operated Check Valve
Removal / Installation ............................. 4-27
4.19 Seat Position Sensor
Removal / Installation ............................. 4-28
4.20 Column Cylinder
Removal / Installation ............................. 4-31
4.21 Column Pilot Operated Check Valve
Removal / Installation ............................. 4-34
4.22 Typical L.H. and R.H. Control Valve Solenoids
Removal / Installation ............................. 4-35
4.23 Typical Base Mounted Control Valve Solenoids
Removal / Installation ............................. 4-37
4.24 Discharge Filter Replacement .................... 4-39
4.25 Main Floor Lock or Outrigger
Floor Lock Cylinder
Removal / Installation ............................. 4-41
4.26 Motor Or Motor Pump
Removal / Installation ............................. 4-42
4.27 Pump Unit
Removal / Installation ............................. 4-44
4.28 Motor Pump Output Check Valve
Removal / Installation ............................. 4-47
4.29 Motor Pump Pressure Relief Valve
Removal / Installation ............................. 4-48
4.30 Outrigger Pressure Relief Valve
Removal / Installation ............................. 4-50
4.31 Main Controller Board
Removal / Installation ............................. 4-53
4.32 Charging / Power Driver Board
Removal / Installation ............................. 4-55
4.33 Emergency Override Panel
Removal / Installation ............................. 4-57
4.34 Distribution Board
Removal / Installation ............................. 4-58
4.35 RFI Filter
Removal / Installation ............................. 4-59
(*) Indicates that there has been a serial number break for the illustration
and that there are additional point page(s) following the original page.
© Schaerer Mayfield USA, Inc. 2004 Page iPrinted in U.S.A.
TABLE OF CONTENTS - CONTINUED
Return To Table Of Contents
Section/Paragraph Page Section/Paragraph Page
4.36 Transformer
Removal / Installation ............................. 4-61
4.37 Line Power Pilot Lamp
Removal / Installation ............................. 4-63
4.38 Plug Receptacle
Removal / Installation ............................. 4-63
4.39 Batteries
Removal / Installation ............................. 4-64
4.40 Primary Thermostat
Removal / Installation ............................. 4-65
4.41 Secondary Thermostat
Removal / Installation ............................. 4-66
4.42 Floor Lock Status Switch
Removal / Installation ............................. 4-66
4.43 Head Section Gas Spring
Removal / Installation / Adjustment ........ 4-68
4.44 Foot Control Interface Board
Removal / Installation ............................. 4-69
4.45 Typical Latch Mechanism
Removal / Installation ............................. 4-70
4.46 Typical Floor Lock Check Valve
Removal / Installation ............................. 4-71
4.47 Articulating Leg Transfer Board
Adjustments............................................ 4-73
4.48 Bell Footpiece Traction Crank Assembly
Disassembly / Assembly ........................ 4-75
SECTION V SCHEMATICS AND DIAGRAMS
5.1 Electrical Schematics / Wiring Diagrams..... 5-1
5.2 Hydraulic Flow Diagram................................ 5-4
5.3 Hand Control Messages ............................... 5- 5
5.4 Error Code Chart .......................................... 5- 6
SECTION VI PARTS LIST
6.1 Introduction ................................................. 6-1
6.2 Description of Columns ............................... 6-1
6.3 Torque Specifications And Important
Assembly Notes ....................................... 6-1
Pictorial Index ............................................. 6-2
Accessories ................................................. 6- 3
Labels And Decals ....................................... 6-4
Cushions and Table Tops............................. 6- 5
Covers and Shields ...................................... 6-6
Head Section Assembly............................... 6-7
Bridge Components...................................... 6-8
Lateral Tilt Cylinder Assembly...................... 6- 9
Upper Hose Connections............................ 6-10
L.H. Side Components ............................... 6-11
L.H. Control Valve Assembly ..................... 6-12
L.H. Seat Cylinder Assembly ..................... 6-13
R.H. Side Components............................... 6-14
Emergency Override Panel Assembly ........ 6-15
R.H. Control Valve Assembly..................... 6-16
R.H. Seat Cylinder Assembly..................... 6-17
Column Assembly Components ................. 6-18
Column Assembly ...................................... 6-19
Trendelenburg Cylinder Assembly ............. 6-20
Column Cylinder Assembly ........................ 6-21
Hand Control Assembly.............................. 6-22
Base Cover Components ........................... 6-23
Base Hydraulic / Electrical
Components............................................ 6-24
Electrical Control Box Components ............ 6-25
Base Hydraulic Plumbing ........................... 6-26
Base Control Valve Assembly.................... 6-27
Hydraulic Pump / Motor Components ......... 6-28
Hydraulic Motor Components ..................... 6-29
Hydraulic Pump Components ..................... 6-30
Floor Lock Components ............................. 6-31
Main Floor Lock Cylinder Components ....... 6-32
Outrigger Floor Lock Cylinder
Components............................................ 6-33
Caster Components ................................... 6-34
Power Driver Board Components................ 6-35
Battery Components .................................. 6-36
10210 Perineal Post Assembly .................. 6-37
10322 Cross Arm Support Assembly ......... 6-38
10360 Safety Strap .................................... 6-39
10395 Lottes Anklet ................................... 6-40
10508 Well Leg Support Assembly............. 6-41
10509 Equipment Cart................................ 6-42
10620 Drape Stand Assembly .................... 6-43
10821 Removable Siderail.......................... 6-44
10823 Leg Transfer Board .......................... 6-45
10824 Horizontal Perneal Post ................... 6-46
10825 Tibia Extender ................................. 6-47
10830 Well Leg Support (Siderail Mount).... 6-48
10836 Offset Adapter Mount ...................... 6-49
13100 / 13200 Montreal Lateral
Position Device....................................... 6-50
(*) Indicates that there has been a serial number break for the illustration
and that there are additional point page(s) following the original page.
© Schaerer Mayfield USA, Inc. 2004
Page iiPrinted in U.S.A.
TABLE OF CONTENTS - CONTINUED
Return To Table Of Contents
Section/Paragraph PageSection/Paragraph Page
22111 Pin and Wire Holder ......................... 6-51
22119 Traction Bow Assembly ................... 6-52
22314 Traction Cuff Assembly ................... 6-53
70542 Bierhoff Leg Holder .......................... 6-54
70556 Siderail Extension Set ..................... 6-55
71070 Siderail Socket ................................ 6-56
71071 I.V. Arm Board Assembly ................ 6-57
71073 Siderail Clamp ................................. 6-58
71074 Conventional Ether Screen .............. 6-59
71075 Foot Rest......................................... 6-60
71077 Siderail Extender Set ....................... 6-61
71078 A/P Cassette Holder ........................ 6-62
71079 Lithotomy Leg Holder ....................... 6-63
71080 Universal Ether Screen .................... 6-64
71082 Lateral Cassette Holder ................... 6-65
71083 Cross Arm Support .......................... 6-66
71084 / 71085 / 71093 I.V. Arm Board
Cushion .................................................. 6-67
71088 Neuro Headrest Adapter................... 6-68
71090 Lateral Support Set.......................... 6-69
71092 Infant Arm & Hand Table ................. 6-70
71094 Table Side Extenders ...................... 6-71
71097 Tri Clamp 2 ...................................... 6-72
71098 Flexguard Stirrup Set ....................... 6-73
71099 Guardian Stirrup Set ........................ 6-74
71100 Leg Prep Stand................................ 6-75
71101 Snaplock Arm Board........................ 6-76
71102 / 71103 Snaplock Arm
Board Cushion ........................................ 6-77
71104 Shoulder Brace Set.......................... 6-78
71105 Universal Patient Brace /
71113 Circular Head Cushion ..................... 6-79
71106 / 71107 Uro Catch Bags................... 6-80
71108 Winged Anesthesia Screen.............. 6-81
71110 Footswitch ....................................... 6-82
71111 Split Leg Set - 71118 / 71119 Split
Leg Pad Set............................................ 6-83
71112 Foot Extension Assembly /
71114 Foot Extension Cushion................... 6-84
71116 Conductive Safety Strap /
20715 Nonconductive Safety Strap ............ 6-85
71120 Side Extender Shelf......................... 6-86
71125 Image Amplification Extension ....... 6-87
71129 X-Ray Arm Board............................. 6-88
71130 I.A. Board (36-inch).......................... 6-89
73001 Pelvic Base w/ Cushion ................... 6-90
73002 Leg Transfer Board w/ Cushion ........ 6-91
73003 Drain Pan......................................... 6-92
73004 I.A. Extention w/ Cushion (55-inch).. 6-93
73005 X-Ray Top Set................................. 6-94
73006 Back Section ................................... 6-95
73007 Ortho Pelvic Base /
73023 / 73024 Board w/ Cushion............ 6-96
73008 Abductor Bar ................................... 6-97
73009 Abductor Bar Clamp ........................ 6-98
73010 Traction Unit .................................... 6-99
73011 Traction Unit Extention .................. 6-100
73012 Popliteal Support ........................... 6-101
73013 Lateral Well Leg Holder .................. 6-102
73014 Shoulder Arthroscopic Attachment. 6-103 Shoulder Arthroscopic Lumbar
Support Assembly ................................ 6-104
Shoulder Arthroscopic Head Rest............. 6-105
73015 Articulating Bracket ....................... 6-106
73016 Drapery Kit .................................... 6-107
73019 Back Section X-Ray Top ............... 6-108
73020 Leg Transfer Frame ....................... 6-109
73021 Yellofin Leg Holder System.... Not Shown
73022 Clear Vision Drapes ............... Not Shown
10531 Bell Footpiece Traction Assy......... 6-110
73025 Orthopedic Cart ............................. 6-111
73026 General Surgery Cart ..................... 6-112
COMMENTS ............................................................ 7-1
(*) Indicates that there has been a serial number break for the illustration
and that there are additional point page(s) following the original page.
Page iiiPrinted in U.S.A.© Schaerer Mayfield USA, Inc. 2004
IMPORTANT INSTRUCTIONS
Return To Table Of Contents
General Safety Instructions
Safety First: The primary concern of Schaerer Mayfield USA is that this surgery table is
maintained with the safety of the patient and staff in mind. Toassure that services and repairs are completed safely and correctly, proceed as follows:
(1 ) Read this entire manual before performing any
services or repairs on this surgery table.
(2 ) Be sure you understand the instructions con-
tained in this manual before attempting to service or repair this surgery table.
Warnings
Throughout this manual are Note, Caution, and Danger paragraphs that call attention to particular procedures. These items are used as follows:
NOTE
A note is used to amplify an operating procedure, practice or condition.
Warranty Instructions
Refer to the SMI “Limited Warranty” printed on the back cover of the Installation and Operation Manual for warranty information. Failure to follow the guidelines listed below will void the warranty and / or render the 7300 Modular Surgery Table unsafe for operation.
• In the event of a malfunction, do not attempt to operate the surgery table until necessary repairs have been made.
• Do not attempt to disassemble surgery table, replace malfunctioning or damaged components, or perform adjustments unless you are one of SMI’s authorized service technicians.
• Do not substitute parts of another manufacturer when replacing inoperative or damaged components. Use only SMI replacement parts.
CAUTION
A CAUTION is used for an operating
procedure, practice, or condition which, if not correctly followed, could result in equipment damage.
DANGER
A DANGER is used for an operating
procedure, practice, or condition which, if not correctly followed, could result in loss of life or serious personal injury.
© Schaerer Mayfield USA, Inc. 2004
Page ivPrinted in U.S.A.
SECTION I
Return To Table Of Contents
GENERAL INFORMATION
SECTION I
GENERAL INFORMATION
1.1 Scope of Manual
This manual contains detailed troubleshooting, sched­uled maintenance, maintenance, and service instruc­tions for 7300 Modular Surgery Table. This manual is intended to be used by SMI’s authorized service technicians.
1.2 How to Use Manual
A. Manual Use When Performing Scheduled Mainte-
nance. (1) Perform inspections and services listed in
Scheduled Maintenance Chart (Refer to para 3.1).
(2) If a component is discovered to be faulty or out
of adjustment, replace or adjust component in accordance with maintenance/service instruc­tions (Refer to para 4.1).
B. Manual Use When Unit Is Malfunctioning And
Cause Is Unknown. (1) Perform an operational test on unit (Refer to
para 2.1).
(2) Perform troubleshooting procedures listed in
Troubleshooting Guide (Refer to para 2.2).
(3) If a component is discovered to be faulty or out
of adjustment, replace or adjust component in accordance with maintenance/service instruc­tions (Refer to para 4.1).
C. Manual Use When Damaged Component Is Known.
(1) Replace or adjust component in accordance
with maintenance/service instructions (Refer to para 4.1).
1.3 Description Of 7300 Modular Surgery Table
A. General Description (See Figure 1-1, Sheets 1
and 2).
The 7300 Modular Surgery Table is designed for use as a general surgery table on which virtually any surgical procedure can be accomplished. 7300 was designed for maximum C-arm access and X-ray visibility.
Major serviceable components of the surgery table are:
1. Motor pump which includes two motors, two pump
units, two motor pump pressure relief valves, two primary thermostats, two secondary thermostats, and a motor pump output check valve, and a discharge filter,
2. Trendelenburg (UP and DOWN) solenoid valves,
Trendelenburg cylinder pilot operated check valves, and Trendelenburg cylinder, trendelenburg position sensor.
3. Seat (UP and DOWN) solenoid valves, seat cylinder
pilot operated check valves, seat cylinders and, seat position sensor.
4. Lateral tilt (LEFT and RIGHT) solenoid valves,
lateral tilt cylinder pilot operated check valves, lateral tilt cylinder, and lateral tilt position sensor.
5. Column (UP and DOWN) solenoid valves, column
cylinder pilot operated check valve, column cylinder, and column assembly which includes slides.
6. Main and Outrigger Floor Lock solenoid valves,
three main floor lock cylinders, two outrigger floor lock cylinders, outrigger pressure relief valve, main and outrigger floor lock status switches.
7. Head section with two gas cylinders.
8. Main Controller Board which includes a 10 amp
fuse.
9. Charging / Power Driver Board which includes a 40
amp fuse.
10. RFI filter.
11. Transformer.
12. Four batteries.
13. Hand control which includes button board and
hand control board.
14. Foot control which includes foot switch pad and
foot control interface board.
15. Emergency override board.
16. Distribution board.
17. Plug Receptacle which includes two 5 amp fuses,
and line power pilot lamp.
© Schaerer Mayfield USA, Inc. 2004
Page 1-1Printed in U.S.A.
SECTION I
Return To Table Of Contents
GENERAL INFORMATION
Figure 1-1. Major Components (Sheet 1 of 2)
Page 1-2Printed in U.S.A.© Schaerer Mayfield USA, Inc. 2004
SECTION I
Return To Table Of Contents
GENERAL INFORMATION
Figure 1-1. Major Components (Sheet 2 of 2)
Page 1-3Printed in U.S.A.© Schaerer Mayfield USA, Inc. 2004
SECTION I
Return To Table Of Contents
GENERAL INFORMATION
B. Theory of Operation. See Figures 5-1, Sheets 1
and 2, and Figure 5-2 for electrical schematic / wiring diagram. See Figure 5-3 for hydraulic schematic.
Electrical Theory Of Operation
When power cord is plugged into power cord recep­tacle, line voltage (approximately 115 VAC) is supplied to table. Line voltage is applied across an RFI filter. RFI filter reduces line conducted RFI / EMI that is present on incoming power line. Across RFI filter output is a line power pilot lamp which illuminates to indicate power is present at output of RFI filter. Transformer primary lines are also connected across RFI filter output. Transformer steps line voltage down to 32.5 VAC ± 2 VAC. The 32.5 VAC from output of transformer is applied to charging / power driver board. Charging / power driver board uses a DC rectifying circuit to convert 32.5 VAC to approximately 27.8 VDC. Charging / power driver board regulates charging rate of four batteries by regulating rate at which 27.8 VDC is applied to batteries. Charging / power driver board also supplies 27.5 VDC to main controller board to power it. Transformer has thermal fuses connected in each primary winding and they are embedded between the primary and secondary windings. If transformer over­heats, normally closed (N.C.) thermal fuses open, disconnecting power to transformer. Transformer is primarily used to supply power to charging / power driver board, so board can charge batteries and supply power to rest of table. Charging / power driver board regulates charge rate of batteries when batteries are being charged. This means that a full load is not continuously being placed on transformer, which is important because transformer is not sized to draw a full current load continuously. If batteries are low and a function is selected, power to drive motor pump(s) is drawn from batteries first and then from transformer as necessary. If batteries are too low or a function is selected for too long, continuous current draw thru transformer will overheat transformer very quickly, causing thermal fuses to blow.
CAUTION
It is important that a table with a low bat­tery charge be operated only in case of
emergency and for a very short time period (less
than a minute).
Logic Theory Of Operation
When ENABLE button is pressed, a 5 VDC signal is sent to main controller board, activating it. Main controller board performs a self diagnostic check on itself. If self diagnostic check fails, error code E11 (Internal RAM / Register Failure) is displayed on hand control. Main controller board and charging / power driver board uses status circuits to check functionality of following electrical components: motors, charging / power driver board, valve spool solenoids, main controller board, position sensors, foot control switch pad, and hand control button board. A status circuit, on charging / power driver board, monitors amount of current draw thru motor pump windings, when a function has been selected. If current draw exceeds a predetermined value, main controller board stops all functions and displays error code E02 (Overcurrent - Motor Pump #1) or E03 (Overcurrent ­Motor Pump #2), which indicates a failure of a motor pump. If current draw falls below a predetermined value, main controller board stops all functions and displays error code E04, which indicates failure of charging / power driver board. A status circuit, on main controller board, monitors enable circuitry for each valve spool solenoid and detects if a valve spool solenoid or enable circuitry of main controller board is functioning correctly. If not, main controller board stops all functions and displays error code E05 (Valve Drive Failure), which indicates failure of either a valve spool solenoid or main controller board. Another status circuit, on main controller board, moni­tors voltage input from position sensors. If voltage value exceeds or falls below a predetermined range, main controller board stops all functions and displays appropriate error code: E07 (Trendelenburg Position Sensor Failure), E08 (Tilt Position Sensor Failure), E09 (Seat Position Sensor Failure). An error code indicates failure of either a position sensor, main controller board, or wiring. Main controller board also monitors main floor lock status switch, outrigger floor lock status switch, motor pump #1 primary thermostat, and motor pump # 2 primary thermostat. When hand control is ENABLED, main controller board checks if normally open (N.C.) main floor lock status switch and outrigger floor lock status switch is untripped. If main controller does not detect that energizes main floor lock valve spool and motor pump, causing main floor locks to extend. After eight seconds, main controller de-energizes main floor lock valve spool
both
status switches are untripped, it
Page 1-4Printed in U.S.A.© Schaerer Mayfield USA, Inc. 2004
SECTION I
Return To Table Of Contents
GENERAL INFORMATION
and motor pump, and then energizes outrigger floor lock valve spool and motor pump, causing outrigger floor locks to extend. After six seconds, main controller de-energizes outrigger floor lock valve spool and motor pump. At this time main controller checks status of status switches again. If main controller does not detect that all functions and displays E01 (Floor Lock Status Switch Is Not Responding). Also, when hand control is ENABLED, main controller board checks if normally closed (N.C.) motor pump primary thermostats are closed. If main controller board detects an open circuit, it stops all functions and displays error code E13 (Overheat - Motor Pump #1) or E14 (Overheat - Motor Pump #2), which indicates that motor pump has overheated, opening its primary thermostat. Also, when hand control is ENABLED, main controller board checks status of each set of switches in hand control; there are two normally open (N.O.) switches for each function, one for solenoid valve and one for motor. If only one of the set of two switches is detected as being pressed, main controller board detects this and displays error code E15 (Hand Control Switch Failure). This indicates that one switch is stuck in closed posi­tion. Also, if a function button is pressed and then error code E15 displays, this indicates that only one of two switches is working properly. This safety feature prevents unintended table movement due to a switch malfunction. Also, when a function is selected using foot control, main controller board checks status of each set of switches in the foot control pad; there are two normally open (N.O.) switches for each function. If there is a set of switches for which only one of two switches is detected as being pressed, the main controller board detects this and displays error code E06 (Foot Switch Failure) on hand control. This indicates that one of two switches is stuck in closed position or not working properly. This safety feature prevents unintended table movement due to a switch malfunction. Motor pumps also have secondary thermostats which provide backup protection if primary thermostats fail. If a motor pump overheats and its primary thermostat does not shut off motor pump, normally open (N.O.) secondary thermostat closes, directly shorting current past motor windings to ground, causing 40 amp fuse on charging / power driver board to blow, which stops motor pump.
both
status switches are untripped, it stops
Hydraulic System Theory Of Operation
When a function is selected using hand control or foot control, motor pump(s) begins to run and selected function's valve spool shifts. Hydraulic oil flows thru a strainer in reservoir and into pump chamber. Pump pumps oil thru an internal check valve, and a discharge filter. Oil is now in supply line, ready to power selected function's cylinder(s). Motor pump internal check valve allows oil to flow thru to the supply line, but closes when motor pump stops running, preventing oil from back flowing into motor pump, which prevents gravity drain out of supply line and back into motor pump. Discharge filter removes any contami­nants from oil before oil reaches any hydraulic compo­nents. This prevents reduced life expectancy of hydraulic components such as seals and o-rings and prevents clogging of valves and check valves. There is a 250 BAR (3625 PSI) pressure relief valve on each motor pump. If pressure reaches or exceeds this pressure, valve opens dumping oil back into reservoir.
Oil is now in supply line, ready to power selected function's cylinder(s).
Trendelenburg and Lateral Tilt
If
Trendelenburg
control was selected (in normal table orientation), motor pump is energized and valve spool for that function moves to an up function position (oil will now flow in new oil flow path that has been created; this path is shown by straight flow symbols on hydraulic schematic) and oil flows thru a check valve (A) and into the base of cylinder, extending cylinder. Check valve (A) prevents oil from escaping from cylinder after valve spool is de­energized, keeping cylinder from drifting. Oil also flows thru a pilot line (represented by a dashed line on hydraulic schematic) and extends a pilot piston. Pilot piston forces open check valve (B) on retracting side of cylinder, allowing oil to escape from top side of cylin­der, thru valve spool and into the oil return line. If
Reverse Trendelenburg
hand control was selected (in normal table orientation), motor pump is energized and valve spool for that function moves to a down function position (oil will now flow in new oil flow path that has been created; this path is shown by crossed flow symbols on the hydraulic schematic) and oil flows thru a check valve (B) and into top of cylinder, retracting cylinder. Check valve (B) prevents oil from escaping from cylinder after valve spool is de-energized, keeping cylinder from drifting. Oil also flows thru a pilot line (represented by a dashed line on hydraulic schematic) and extends a pilot piston.
or
Lateral Tilt Left
or
button on hand
Lateral Tilt Right
button on
Page 1-5Printed in U.S.A.© Schaerer Mayfield USA, Inc. 2004
SECTION I
Return To Table Of Contents
GENERAL INFORMATION
Pilot piston forces open check valve (A) on base of cylinder, allowing oil to flow from base of cylinder, thru valve spool, and into oil return line.
Seat Functions
If
SEAT UP
normal table orientation), motor pump is energized and valve spool for that function moves to an up function position (oil will now flow in new oil flow path that has been created; this path is shown by straight flow symbols on hydraulic schematic) and oil flows thru a check valve (A) and into base of left cylinder (master cylinder), extending cylinder. Check valve (A) prevents oil from escaping from cylinder after valve spool is de­energized, keeping cylinder from drifting. Oil also flows thru two pilot lines (represented by dashed lines on hydraulic schematic) and extends two pilot pistons (B and D). One pilot piston forces open a check valve (B) which allows oil to flow out of top side of left cylinder thru another check valve (C) and into base of right cylinder (slave cylinder), extending cylinder. Other pilot piston forces open check valve (D) on top side of right cylinder (slave cylinder), allowing oil to flow from top side of cylinder, flow thru valve spool and into oil return line. If
SEAT DOWN
normal table orientation), motor pump is energized and valve spool for that function moves to a down function position (oil will now flow in new oil flow path that has been created; this path is shown by crossed flow symbols on hydraulic schematic) and oil flows thru a check valve (D) and into top of right cylinder (slave cylinder), retracting cylinder. Check valve (D) prevents oil from escaping from right cylinder after valve spool is de-energized, keeping cylinder from drifting. Oil also flows thru two pilot lines (represented by dashed lines on hydraulic schematic) and extends two pilot pistons. One pilot piston forces open a check valve (C) which allows oil to flow out of base of right cylinder thru another check valve (B) and into top side of left cylinder (master cylinder), retracting cylinder. Other pilot piston forces open check valve (A) on base of left cylinder (master cylinder), allowing oil to flow out of base of cylinder, thru valve spool, and into oil return line.
Column Functions
If
TABLE UP
pump energizes and valve spool for that function moves to an up function position (oil will now flow in new oil flow path that has been created; this path is shown by straight flow symbols on hydraulic schematic) and oil flows thru a check valve (A) and into base of column
button on hand control is selected (in
button on hand control is selected (in
button on hand control is selected, motor
cylinder, extending cylinder. Check valve (A) prevents oil from escaping from column cylinder after valve spool is de-energized, keeping cylinder from drifting. Oil flows out of top side of column cylinder as necessary, thru valve spool, and into return line. If
TABLE DOWN
motor pump energizes and valve spool for that function moves to a down function position (oil will now flow in new oil flow path that has been created; this path is shown by crossed flow symbols on hydraulic sche­matic) and oil flows to top side of base cylinder, retract­ing cylinder. Oil also flows thru a pilot line (represented by a dashed line on hydraulic schematic) and extends a pilot piston. Pilot piston forces open check valve (A) on base of cylinder, allowing oil to flow out of base of cylinder, thru valve spool, and into oil return line.
Floor Lock Functions
If
ENABLE / LOCK
motor pump energizes and valve spool for main floor lock cylinders moves to an up function position (oil will now flow in new oil flow path that has been created; this path is shown by straight flow symbols on hydraulic schematic) and oil flows thru a check valve (A) and into base of three main floor lock cylinders, extending cylinders. Check valve (A) prevents oil from escaping from three main floor lock cylinders after valve spool is de-energized, keeping cylinders from drifting. After approximately 10 seconds, main controller board de-energizes motor pump and main floor lock spool and then reenergizes motor pump and energizes outrigger floor lock valve spool to an up function position (oil will now flow in new oil flow path that has been created; this path is shown by straight flow symbols on hydraulic schematic) and oil flows thru two check valves (B and C) and into the base of two outrigger floor lock cylin­ders, extending cylinders. As soon as pressure in outrigger floor lock cylinders reaches 10 BARS (145 PSI), outrigger pressure relief valve opens, dumping oil back to return line. Outrigger cylinders were designed to add stability to table on an uneven floor. Three main floor lock cylinders lift table off of casters; then the two outrigger floor lock cylinders extend with enough power to stabilize table further. Check valves (B and C) prevent oil from escaping from two outrigger floor lock cylinders after valve spool is de­energized, keeping cylinders from drifting. If
UNLOCK & ENABLE
selected for three seconds, motor pump energizes and main floor lock valve spool moves to a down function position (oil will now flow in new oil flow path that has been created; this path is shown by crossed flow
button on hand control is selected,
button on hand control is selected,
without lifting
buttons on hand control is
table
Page 1-6Printed in U.S.A.© Schaerer Mayfield USA, Inc. 2004
SECTION I
Return To Table Of Contents
GENERAL INFORMATION
symbols on hydraulic schematic). Oil flows thru three pilot lines (represented by dashed lines on hydraulic schematic) and extends three pilot pistons, which forces open three check valves; check valves A, B, and C. With all three check valves open, oil flows out of base of three main floor lock cylinders and two outrigger floor lock cylinders, causing them to retract. There is a strong spring in each floor lock cylinder which helps retract cylinder.
Seat Slave Line Valve
On the seat functions, there is a manually operated two way-valve (slave line valve) which is used to add or remove oil from closed loop (slave line) to synchronize stroke of left (master) and right (slave) cylinders.
Column Slave Line Valve
There is a manually operated two-way valve (slave line valve) on column function, which is used to add or remove oil from middle stage of column cylinder (if there is too much oil in second stage, column cylinder will "jump" downward and make a noise when column down function is selected. If there is not enough oil in second stage, full height potential of column cylinder will not be reached).
Orifices
Orifices are internal to Trendelenburg, Lateral Tilt, and Seat, cylinders to meter flow of oil into cylinders, thereby regulating speed of cylinders.
1.4 SPECIFICATIONS
Factual data for the 7300 Modular Surgery Table is provided in Table 1-1.
Table 1-1. Specifications
Description Data
Weight
Normal (w/ head & pelvic sections & upholstery) ...........
........................................................... 660 lb. (299 kg)
With Shipping Carton.......................... 764 lb. (347 kg)
Shipping Carton........ 58 in. "L" x 30 in. "W" x 40 in. "H"
(147 cm x 76 cm x 102 cm)
Dimensions: (w/ Head, Back, Pelvic & Leg Transfer Board.)
Table Top Length ............................. 75.6 in. (192 cm)
Table Top Width ...................................... 20.5 (52 cm)
Overall Width......................................... 24 in. (61 cm)
Overall Length..................................... 78 in. (198 cm)
Table Positioning:
Table Top Height (Adjustable) .......... 27.7 (± 0.2) in. to
43.9 (± 0.2) in.
(70 to 112 cm)
Trendelenburg ......................................0° to 28° (±1°)
Reverse Trendelenburg....................... 0° to -28° (±1°)
Lateral Tilt................ 0° to -18° (±1°) in either direction
Seat Section .......... + 75° (±1°) to -40° (±1°) in relation
to back section
Coordinating Flex (Normal Position):
Trendelenburg ................+20° (±5°) above horizontal
Seat Section ......-40° (±5°) in relation to back section
Coordinating Reflex (Normal Position):
Trendelenburg ................. -28° (±5°) below horizontal
Seat Section ..... +60° (±5°) in relation to back section
Head Section:
Mounted on Head End...........+30° (±3°) to -30° (±3°)
Mounted on Foot End ............+30° (±3°) to -75° (±5°)
Page 1-7Printed in U.S.A.© Schaerer Mayfield USA, Inc. 2004
SECTION I
MODEL
NUMBER
SERIAL
NUMBER
SMI
7100-001
115 VAC
2 AMP 60 HZ TDWXXXXX
MODEL
INPUT
RATING
SERIAL NO.
Return To Table Of Contents
GENERAL INFORMATION
Table Speed (Table in normal orientation, full battery
charge, ambient oil temp.):
Extend Column Fully .................. 16.0 to 21.0 seconds
Retract Column Fully.................. 16.0 to 21.0 seconds
Reverse Trend. to Trend. ........... 20.5 to 24.5 seconds
Trend. to Reverse Trend. ........... 16.0 to 20.0 seconds
Lat. Tilt R to Lat. Tilt L .................. 9.5 to 12.5 seconds
Lat. Tilt L to Lat. Tilt R ................ 10.0 to 13.0 seconds
Seat Down to Seat Up ................ 18.0 to 21.0 seconds
Seat Up to Seat Down ................ 18.5 to 21.5 seconds
Weight Capacity (maximum):
Normal Table Orientation ..................500 lbs. (227 kg)
Reverse Table Orientation..............400 lbs. (181.4 kg)
Head Section.........................................40 lbs. (18 kg)
Overweight Operation ................ Patient Positioning is
restricted. To learn of limita-
tions, contact the Surgical
Table Product Manager at
1-800-643-6275
Hydraulic System Oil ................ ISO Viscosity Grade 32
Premium Hydraulic Fluid
Reservoir Capacity ..................... 2.1 Quarts (2.0 Liters)
(2) Refer to the Parts List to determine the item
numbers of the parts, part numbers of the parts, descriptions of the parts, and quantities of parts needed and record this data (Refer to para 6.1).
NOTE
Ask the Purchasing Department of the company that owns the unit for this information. Otherwise, this information may be obtained from the dealer that sold the unit.
(3) Determine the installation date of the unit and
record this data.
Serial Number when contacting SMI
(4)Call SMI with recorded information and ask
for Technical Services Department. Use phone number 1-800 643-6275
Please have the Model and
System Capacity ........................ 3.7 Quarts (3.5 Liters)
Battery Charging Time.............................. 8 to 12 hours
Motor Pump Pressure
Relief Valve ...............Opens At 250 BARS (3625 PSI)
Outrigger Pressure
Relief Valve ...................Opens At 10 BARS (145 PSI)
Electrical Requirements:
115 VAC Unit........................... 110 - 120 VAC, 60 HZ,
2 amp, single phase
Battery Power Output............ 24 VDC, 18 amp-hours
1.5 Parts Replacement Ordering
If a part replacement is required, order the part directly from the factory as follows:
(1) Refer to Figure 1-2 to determine the location of
the model number and serial number of the unit and record this data.
SMI
7100
CA7681
Figure 1-2. Model Number / Serial
Number Location
© Schaerer Mayfield USA, Inc. 2004
Page 1-8Printed in U.S.A.
1.6 Special Tools
Return To Table Of Contents
Table 1-2 lists all of the special tools needed to repair the unit, how to obtain the special tools, and the pur­pose of each special tool.
Table 1-2 is on the following page.
Table 1-2. Special Tools List
SECTION I
GENERAL INFORMATION
Description of Special Tool
Multimeter Commercially Available Any Type Used to perform continuity and voltage checks. Pin Extractor (has
interchangable bits - M3x0.5, M4x0.7, M5x0.8, M6x1.0, 1/4-20, 10-24, and 5/16-18)
Spanner Wrench - locking nut Schaerer Mayfield USA
Spanner Wrench - brass pivot puck
Retaining Bolt Schaerer Mayfield USA
Spanner Wrench - retainer cap Schaerer Mayfield USA
Pressure Relief Valve Test Harness
Blanking Plug Kit Schaerer Mayfield USA
Soldering Iron Commercially Available Any Type Used to connect wires to terminals with solder. 60 / 40 Solder Commercially Available Any Type Type of solder which should be used when
Internal retaining ring pliers Commercially Available Any Type Used to remove snap rings which secure bearings
Torque Wrench Commercially Available Any Type Used to tighten hardware to specified torque
Transport Tool Schaerer Mayfield USA
Tipping Tool Schaerer Mayfield USA
Manufacturer's
Name / Address / Phone
Schaerer Mayfield USA 4900 Charlemar Drive Cincinnati, Ohio 45227 (800) 755-6381
4900 Charlemar Drive Cincinnati, Ohio 45227 (800) 755-6381
Schaerer Mayfield USA 4900 Charlemar Drive Cincinnati, Ohio 45227 (800) 755-6381
4900 Charlemar Drive Cincinnati, Ohio 45227 (800) 755-6381
4900 Charlemar Drive Cincinnati, Ohio 45227 (800) 755-6381
Schaerer Mayfield USA 4900 Charlemar Drive Cincinnati, Ohio 45227 (800) 755-6381
4900 Charlemar Drive Cincinnati, Ohio 45227 (800) 755-6381
4900 Charlemar Drive Cincinnati, Ohio 45227 (800) 755-6381
4900 Charlemar Drive Cincinnati, Ohio 45227 (800) 755-6381
Manufacturer's
Part Number
M05390 Used to pull pins which secure cylinders, handles,
etc. in place. Can also be used to remove caster fork.
M07235 Used to remove locking nut which secures top of
column cylinder in column.
M07236 Used to remove brass pivot puck which secures
top of column cylinder in column.
M07407 Used to hold rod of column cylinder stationary,
while special tools, M07235 and M07236 are used.
M07161 Used to remove retainer cap which secures
latching mechansim in place.
M07443 Used to adjust crack settings of motor pump
pressure relief valves and outrigger pressure relief valve.
502-0176-00 Used to cap the end of hydraulic lines to allow
pressure checks to be made.
soldering wires on surgery table.
in base of cylinders.
values.
502-0169-00 A dolly type tool which is used to transport Model
7100 or 7300 surgical tables around.
502-0168-00 Used to tip over Model 7100 or 7300 surgical
tables to access base of table for servicing.
Purpose of Special Tool
© Schaerer Mayfield USA, Inc. 2004
Page 1-9Printed in U.S.A.
SECTION I
Return To Table Of Contents
GENERAL INFORMATION
© Schaerer Mayfield USA, Inc. 2004
Page 1-10Printed in U.S.A.
TESTING AND TROUBLESHOOTING
Return To Table Of Contents
SECTION II
TESTING AND TROUBLESHOO TING
SECTION II
2.1 Operational Test
In order to effectively diagnose malfunctions of the 7300 Modular Surgical Table, it may be necessary to perform an operational test as follows:
DANGER
Refer to the Operator Manual for
complete instructions on operating the surgical table. Failure to do so could result in personal injury. Power must be supplied to the unit to measure the voltage at the Test Points. Do not touch any bare wires or terminals while making these voltage checks. Failure to do so may result in electrical shock which could result in serious personal injury or death.
NOTE
The Operational Test, for the most part, only de­scribes what should happen when the surgical table is operated. If the surgical table does something other than described, a problem has been discovered. Refer to the Troubleshooting Guide to determine the cause of the problem and its correction.
(1 ) Plug power cord into table's plug receptacle.
See Figure 2-1, Sheets 1 and 2.
(8 ) Move the surgery table around. (9) Observe. The casters should swivel freely and
the table should roll easily on the casters. The
table should not wobble. (10) Press ENABLE button on hand control. (11) Observe. The motor pump should run and the
main floor lock cylinders should extend. After
eight seconds, the motor pump should stop
running. Then the motor pump should begin
running again and the outrigger floor lock
cylinders should extend. After six seconds, the
motor pump should stop running. The hand
control display should read LOCKING FLOOR
while the floor lock cylinders are extending. (12) Insert pins of Head Section into sockets on
column end of table. Rotate latch handles to
locked position (to move latch handles to locked
position, rotate them away from you) and then
pull and push on head sections. (13) Observe. When the latch handles are rotated to
the locked position, the section being installed
should now be held securely in place. No side-
to-side or in-and-out movement should be
possible.
(2) Observe. The line power pilot lamp should
illuminate. (3 ) Unplug the power cord. (4 ) Press ENABLE button on hand control. (5) Observe. The hand control display should
illuminate and display the current table position
and battery charge level (XX POSITION and
BATT LEVEL XX. After 10 seconds, the hand
control display should extinguish. (6 ) Press and hold UNLOCK & ENABLE buttons on
hand control for three seconds. (7) Observe. The motor pump should run and the
table should lower (floor lock cylinders should
retract) until table is on its casters. The hand
control display should read UNLOCKING
FLOOR and then ON WHEELS.
© Schaerer Mayfield USA, Inc. 2004
(14) Press the ENABLE and REVERSE POSITION
buttons on hand control.
(15) Observe. The hand control display should read
REVERSE POSITION.
(16) Press the FLEX button and then REFLEX
button.
(17) Observe. Hand control display should indicate
these positions are "disabled" and cannot be used when table is in REVERSE POSITION.
(18) Press the ENABLE and NORMAL POSITION
buttons on hand control.
(19) Observe. The hand control display should read
NORMAL POSITION.
Page 2-1 Printed in U.S.A.
SECTION II
Return To Table Of Contents
TESTING AND TROUBLESHOOTING
C
LINE POWER
PILOT LAMP
HEAD
SECTION
HAND
CONTROL
LATCH
HANDLE
EMERGENCY
OVERRIDE
PANEL
FOOT
CONTROL
RELEASE
HANDLE
ENABLE LOCK
LEVEL
SMI
7300
SMI
PLUG
B
DISABLE
UNLOCK
7100
SMI
7300
BACK
SECTION
RECEPTACLE
A
SEAT SECTION (PELVIC BASE)
POWER
CORD
TRENDEL-
ENBURG
REVERSE
TRENDEL-
ENBURG
TABLE
UP
TABLE DOWN
FLOOR
LOCK
CYLINDERS
LATERAL
TILT
LEFT
LATERAL
TILT
RIGHT
Figure 2-1. Operational Test (Sheet 1 of 2)
LOCKED
POSITION
AUTO CAPTURE POSITION
UNLOCKED
POSITION
LATCH HANDLE
POSITIONS
(TYP. 2 PLACES)
CA819700
© Schaerer Mayfield USA, Inc. 2004
Page 2-2Printed in U.S.A.
SECTION II
Return To Table Of Contents
TESTING AND TROUBLESHOOTING
BUTTONS
SMI
DISABLE
LATERAL
TILT LEFT
SYSTEM
OVERRIDE
SYSTEM
OVERRIDE
7300
DISABLE
TABLE DOWN
TABLE
UP
TRENDELENBURG
REVERSE
TRENDELENBURG
UNLOCKLOCK
SEAT
UP
SEAT
DOWN
HAND
CONTROL
DISPLAY
ENABLE
NORMAL
POSITION
TRENDELENBURG
TABLE
UP
LATERAL
TILT LEFT
SEAT
UP
FLEX
ENABLE
LOCK
DISABLE
NORMAL
REVERSE
POSITION
TREND
POSITION
REV TREND
HEIGHT UP
HEIGHT DOWN
TILT L
TILT R
SEAT UP SEAT DOWN
FLEX
UNLOCKLEVEL
SMI
LEVEL
(RETURN TO)
DISABLE
REVERSE POSITION
REVERSE
TRENDELENBURG
TABLE DOWN
LATERAL
REFLEX
TILT RIGHT
SEAT
DOWN
7300
REFLEX
UNLOCK
LATERAL
TILT RIGHT
LEVEL
UNLOCK
(RETURN TO)
Figure 2-1. Operational Test (Sheet 2 of 2)
(20) Press the ENABLE button and then press the
TABLE UP, TABLE DOWN, TRENDELENBURG, REVERSE TRENDELENBURG, LATERAL TILT LEFT, LATERAL TILT RIGHT, SEAT UP, SEAT DOWN, FLEX, and REFLEX buttons, running each function thru its full range of motion, and checking its speed and range of motion perfor­mance.
(21) Observe. The table performance should meet
the following specifications:
Range Of Motion:
Table Top Height (Adjustable) ........... 27.7 (± 0.2) in. to
43.9 (± 0.2) in. (70 to 112 cm)
Trendelenburg ....................................... 0° to 28° (±1°)
Reverse Trendelenburg ......................... 0° to -28° (±1°)
Lateral Tilt................. 0° to -18° (±1°) in either direction
CA819800
Seat Section ........... + 25° (±1°) to -40° (±1°) in relation
to back section
Coordinating Flex (Normal Position):
Trendelenburg .................. +20° (±5°) above horizontal
Coordinating Reflex (Normal Position):
Trendelenburg ................... -12° (±5°) below horizontal
Table Speed (Table in normal orientation, full battery
charge, ambient oil temp.):
Extend Column Fully................... 16.0 to 21.0 seconds
Retract Column Fully .................. 16.0 to 21.0 seconds
Reverse Trend. to Trend. ............ 20.5 to 24.5 seconds
Trend. to Reverse Trend. ............ 16.0 to 20.0 seconds
Lat. Tilt R to Lat. Tilt L .................. 9.5 to 12.5 seconds
Lat. Tilt L to Lat. Tilt R ................ 10.0 to 13.0 seconds
Seat Down to Seat Up ................ 18.0 to 21.0 seconds
Seat Up to Seat Down ................ 18.5 to 21.5 seconds
© Schaerer Mayfield USA, Inc. 2004
Page 2-3Printed in U.S.A.
SECTION II
Return To Table Of Contents
TESTING AND TROUBLESHOOTING
(22) Press the ENABLE button and then press the
LEVEL button until the table reaches its level position. Using a protractor, check the table top to ensure it is level within ±2°.
(23) Observe. The table top should move to a level
position within ±2° as follows: The lateral tilt function should level out first; then a sequence of the Seat, and Trendelenburg functions should move, with each function moving no more than 10° at a time.
(24) Squeeze the release handle on the head section
and move the head section thru its full range of motion.
(25) The head section should move smoothly and
should not take excessive force. The head section should meet the following range of motion specifications:
Head Section:
Mounted on Head End ........... +30° (±3°) to -30° (±3°)
Mounted on Foot End ............ +30° (±3°) to -75° (±5°)
(26) Move head section to level position and then
place approximately 40 lbs. (18 kgs) at Point C (head section). Observe head section for two to
three minutes. (27) Observe. The head section should not drift. (28) Place approximately 400 lbs. (181.4 kg) of
weight on table top as follows: 28 lbs. (12.7 kg)
at Point C (head section), 152 lbs. (69 kg) at
Point B (back section), and 168 lbs. (76.2 kg) at
Point A (seat section).
(33) Observe. Each function should move as
depicted by the illustration on its button.
(34) Press each function button on the foot control.
Then place the table top in REVERSE position. Then press each function button on the foot control again.
(35) Each function should move as depicted when
table top is in NORMAL position. The table should not operate, using the foot control, when the table top is in REVERSE position.
(36) Plug power cord into table's power receptacle.
DANGER
Electrical power must be connected to
perform the following steps. Do not touch any bare wires or terminals. Failure to do so could cause an electrical shock which could result in serious personal injury or death.
(37) See Figure 2-2 for this step. Check for 32.5
VAC ± 2 VAC across Test Points C and D and across Test Points D and E.
(38) Observe. There should be 32.5 VAC ±2 VAC
across
(39) See Figure 2-2 for this step. Check for 27.7
VAC ± 2 VAC across Test Points F and G and across Test Points H and J. Check for 27.8 VAC ±2 VAC across Test Points K and L.
(40) Observe. The proper voltages should be
present across
both
sets of test points.
each
set of test points.
(29) Press ENABLE button and then press TABLE
UP and TABLE DOWN buttons to run column function up and down several times.
(30) Observe. Column function should move up and
down steadily and smoothly. No binding condition should be evident, especially when column is lowering. There should be no exces-
sive side-to-side play in column assembly. (31) Remove weight from table top. (32) Press and hold the SYSTEM OVERRIDE button
while pressing the function buttons, one at a
time, on the emergency override panel.
© Schaerer Mayfield USA, Inc. 2004
Page 2-4Printed in U.S.A.
SECTION II
Return To Table Of Contents
TESTING AND TROUBLESHOOTING
POWER CORD RECEPTACLE
5 amp. FUSE
L
115
N
VAC
5 amp. FUSE
Note
Wire colors may change without notification
ST5b
3
1
11
9
15
5
7 8
6
10
2
12
141316
4
RED
RED
RED
RED
RED
BLACK
BLACK
BROWN
BLUE
BLACK
BLACK
BLACK
BLACK
WHITE
VIOLET
GREEN
YELLOW
ORANGE
RED
4
5
FOOT
CONTROL
6
PORT
1
NOTE: FOOT AND HAND CONTROL
PLUGS ARE POSITIONED TO VIEW
TERMINAL (BACK) END.
5
6
HAND
CONTROL
7
PORT
1
2
GREEN
3
2
4
3
= DENOTES "TEST POINTS"
!!
LINE POWER PILOT LAMP
A
BROWN
BROWN
RFI
FILTER
BLUE
YELLOW / GREEN
YELLOW / GREEN
ST6
12
GREY / PINK
GREY / WHITE
BROWN / GREY
WHITE / GREEN
BROWN / GREEN
YELLOW / WHITE
BROWN / YELLOW
BLUE
GREEN
YELLOW
RED
BROWN
GREY
GREY BLUE
GREEN
WHITE
YELLOW
RED
BROWN
9111413161518172019 10
RED / BLUE
B
7
8
WHITE
BLUE
YELLOW / GREEN
3
5 6
4
GREY
VIOLET
WHITE YELLOW
1 2
BLUE
M
N.C.
80° c
(176° F)
FUSE
OUTRIGGER
FLOOR LOCK
STATUS SWITCH
BLACK
2
WHITE
GREY
WHITE
N
100° c
MOTOR
(212° F)
PUMP
SECONDARY
L2
#2
THERMAL
RED
1
BROWN
N.O.
N.O.
N.C.C.
(SWITCHES SHOWN IN UNTRIPPED POSITION)
BROWN
FUSE
MAIN FLOOR
LOCK STATUS
SWITCH
BLACK
2
WHITE
Q
RED
1
BROWN
N.C.
P2
80° c
(176° F)
L1
PRIMARY
THERMAL
FUSE
N.O.
N.C.C.
4
BROWN
X
ST3
W
8
6
10
12
7
9 11
PINK
WHITE
GREEN
YELLOW
RED / BLUE
GREY / PINK
RED
R
2 1 3 5
BLUE
GREY
VIOLET
BLACK
P
ST10
8
10 12 14
6
4
2
7
1 3 5
13
9
11
RED
PINK
BLUE
BLACK
BROWN
VIOLET
GREEN
YELLOW
RED / BLUE
GREY / PINK
WHITE / GREEN
BROWN / GREEN
RED
PINK
BLUE
GREY
RED / BLUE
BLACK
GREEN
VIOLET
BROWN
YELLOW
GREY / PINK
3
911
5
7
1
2
12
10
6
8
4
St23
WHITE
K
ST3a
ST2a
TRANSFORMER
THERMAL
FUSE
BLACK
RED
THERMAL
FUSE
YELLOW
BLACK
RED
2
2
ST1
ST1
3
3
4
4
ST8 (P2)
1
1
CHARGING / POWER
ST8
BLACK
RED
BLACK
C5
ST3a
(BATT. -)
C D E
9 10
V
DRIVER BOARD
T
9 10
RED
YELLOW
BLUE
BLACK
BROWN
2 1 3 5
7 8
7 8
WHITE
VIOLET
4
S
ST5a
ST5a
ST2a
(BATT. +)
5 6
5 6
GREY
GREY
6
3 4
3 4
VIOLET
WHITE
8 7
ST2b
2
3
1
9111413161518172019
5
7
1 3 5
8
6
10
2
4
MAIN CONTROLLER BOARD
1
2
3
WHITE
GREEN
BROWN
2 4
2
1 3
4
St35
BLACK
GREEN
BROWN
YELLOW
GREEN
BROWN
YELLOW
H
10 amp.
FUSE
1
1
St1
2 2
RED
PINK
BROWN
YELLOW
RED
RED
2
1 3
4
BLUE
St37
5
6
7
8
GREY
RED
2
1 3
4
BLUE
St36
5
6
WHITE
7
8
GREY
J
SEAT POSITION
YELLOW
SENSOR
16 14 13
12 10 8 6 4 2
RED
12
ST6
St1
St4
15
11 9 7
St5b
5 3 1
LATERAL TILT
POSITION SENSOR
GREEN
2
1
1
3
3
4
5
6
St33
St32
U
ST7a
4
(PUMPS +)
1 2
1 2
BLUE
RED
10 9
6
ST2b
ST9 a & b
ST7 (P1)
BLACK
YELLOW
8 7
WHITE
FUSE
40 amp.
ST4
2 1
BLACK
BROWN
10 9
GREEN
BROWN
L
- +- +- +-
+
BATTERY BATTERY BATTERY BATTERY
ST9a
RED
ST9b
RED
F
2 1
G
RED
GREY
WHITE
BLACK
BROWN
8
6
4
2
7
1 3 5
ST7
8
6
4
2
7
1 3 5
ST7
TRENDELENBURG POSITION SENSOR
2
3
100° c
MOTOR
(212° F)
PUMP
SECONDARY
L2
#1
THERMAL
N.O.
ST4
FUSE
PI
L1
PRIMARY THERMAL
WHITE
BLACK
BLACK
GREEN
10
9
10
9
YELLOW
8
10
6
4
2
12
7
9 11
1 3 5
ST3
ON ON
12
8
10 12 14
6
4
2
7
1 3 5
13
9
11
ST10
1
Figure 2-2. Voltage Checks
© Schaerer Mayfield USA, Inc. 2004
CA819900
Page 2-5Printed in U.S.A.
SECTION II
Return To Table Of Contents
TESTING AND TROUBLESHOOTING
2.2 Troubleshooting Procedures
Table 2-1 is a Troubleshooting Guide which is used to determine the cause of the malfunction.
Table 2-1. Troubleshooting Guide
Problem Symptom Probable Cause Check Correction
Table seems powerless even when plugged into a wall outlet.
Line Power Pilot Lamp does not illuminate (with power cord plugged in).
Line Power Pilot Lamp illuminates, but batteries do not recharge and table is powerless (with power cord plugged in).
Line Power Pilot Lamp illuminates, but table is powerless.
Facility circuit breaker providing power to table is tripped.
One of two 5 amp fuses in the power cord receptacle is blown.
RFI filter is blown. See Figure 5-1 for this
Line power pilot lamp is burned out.
Wires or wire connections between power cord receptacle and transformer are broken, loose or dirty.
Thermal fuse(s) in transformer are blown.
Transformer is blown. See Figure 5-1 for this
Charging / power driver board is malfunctioning.
Wires or wire connections between transformer and charging / power driver board are broken, loose, or dirty.
10 amp fuse on main controller board is blown.
Check to see if facility circuit breaker is tripped.
Remove both 5 amp fuses from power cord receptacle and perform a continuity check on the fuses.
check. Check for 115 VAC ± 5.0 VAC across Test Points A and B.
Replace suspect line power pilot lamp with known working line power pilot lamp.
Perform a continuity check on all suspect wires or connections.
See Figure 5-1 for this check. Check for 32.5 ±
2.0 VAC across Test Points C and D and across Test Points D and E.
check. Check for 32.5 ±
2.0 VAC across Test Points C and D and across Test Points D and E.
See Figure 5-1 for this check. Check for 27.7 +/- 0.3 VDC across Test Points F and G. Check for 27.7 +/- 0.3 VDC across Test Points H and J. Check for 27.8 +/- 0.3 VDC across Test Points K and L.
Perform a continuity check on all suspect wires or connections.
Perform continuity check on 10 amp fuse.
If circuit breaker is tripped, determine what caused the circuit breaker to trip, correct the problem, and then reset / replace circuit breaker.
Replace any blown 5 amp fuses.
If 115 VAC ± 5.0 VAC is not present across Test Points A and B, replace RFI filter. Refer to para 4.35.
Replace line power pilot lamp. Refer to para 4.37.
Clean any dirty connections. Tighten any loose connections. Replace any broken connectors or wires.
If 32.5 VAC ± 2.0 VAC is not present across one or both of the Test Point groups, replace the transformer. Refer to para 4.36.
If 32.5 VAC ± 2.0 VAC is not present across one or both of the Test Point groups, replace the transformer. Refer to para 4.36.
If proper voltage is not present at Test Points, replace charging / power driver board. Refer to para
4.38.
Clean any dirty connections. Tighten any loose connections. Replace any broken connectors or wires.
Replace 10 amp fuse. See Figure 5-1 for fuse location.
© Schaerer Mayfield USA, Inc. 2004
Page 2-6Printed in U.S.A.
Problem Symptom Probable Cause Check Correction
Return To Table Of Contents
Table operates when power cord is plugged in, but will not operate when power cord is unplugged.
Table 2-1. Troubleshooting Guide
When ENABLE button on hand control is pressed, its display does not illuminate (with power cord unplugged).
TESTING AND TROUBLESHOOTING
40 amp fuse on charging / power driver board is blown.
10 amp fuse on main controller board is blown.
Batteries are dead and are not supplying charging / power driver board with enough power to operate.
See Figure 5-1 for this check. Check fuse visually or perform continuity check on fuse (may need to check secondary thermal fuse of motor pump. If blown, thermal fuse will cause 40 amp fuse to blow).
Perform continuity check on 10 amp fuse.
See Figure 5-1 for this check. Check individual batteries by checking for
13.2 +/- 0.4 VDC across terminals of each battery (Test Points M and N).
SECTION II
Replace 40 amp fuse. See Figure 5-1 for fuse location.
Replace 10 amp fuse. See Figure 5-1 for fuse location.
If 13.2 +/- 0.4 VDC is not present across Test Points, replace batteries. Refer to para 4.39.
Batteries do not recharge in 6 to 8 hours from low charge.
Hand control does not work.
Batteries do not recharge very quickly.
When ENABLE button on hand control is pressed, its display does not illuminate.
Charging / power driver board is malfunctioning.
One of thermal fuses in transformer is blown.
Batteries are dead. See Figure 5-1 for this
10 amp fuse on main controller board is blown.
Main controller board or hand control board is malfunctioning.
Wires or wire connections between main controller board, distribution board, and hand control port are broken, loose or dirty.
Motor pump secondary thermostat is closed creating a direct short to ground, causing 40 amp fuse on charging / power driver board to blow.
See Figure 5-1 for this check. Check for 27.7 +/- 0.3 VDC across Test Points F and G.
See Figure 5-1 for this check. Check for 32.5 +/- 2.0 VAC across Test Points C and D, and across Test Points D and E.
check. Check individual batteries by checking for
13.2 +/- 0.4 VDC across terminals of each battery (Test Points M and N).
Perform continuity check on 10 amp fuse.
See Figure 5-1 for this check. Check for 5.0 VDC across Test Points P and Q. If 5 VDC is present, check for 5.0 VDC across Test Points Q and R within 10 seconds of pressing ENABLE button on hand control.
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
See Figure 5-1 for this check. Check for a resistance value of 0 ohms between Test Points S and T for motor # 1 or Test Points U and V for motor # 2.
If proper voltage is not present at Test Points, replace charging / power driver board. Refer to para 4.32.
If 32.5 +/- 2.0 VAC is not present across one or both of the Test Point groups, replace the transformer. Refer to para 4.36.
If 13.2 +/- 0.4 VDC is not present across Test Points, replace batteries. Refer to para 4.39.
Replace 10 amp fuse. See Figure 5-1 for fuse location.
If 5.0 VDC is not present across Test Points P and Q, the main controller board is malfunctioning. If 5.0 VDC is not present across Test Points Q and R during check, then main controller board or hand control board is malfunctioning. Refer to para
4.31 or 4.5. Clean any dirty connections.
Tighten any loose connections. Replace any broken connectors or wires.
Replace malfunctioning motor, secondary thermostat, and 40 amp fuse. Refer to para 4.26 and 4.41.
© Schaerer Mayfield USA, Inc. 2004
Page 2-7Printed in U.S.A.
SECTION II
Return To Table Of Contents
TESTING AND TROUBLESHOOTING
Table 2-1. Troubleshooting Guide
Problem Symptom Probable Cause Check Correction
Hand control does not work - Continued.
When ENABLE button on hand control is pressed, its display does not illuminate
- Continued.
Display of hand control board is malfunctioning.
Replace suspect hand control board with known working hand control board.
Replace hand control board. Refer to para 4.5.
Foot control does not work.
When a button on hand control is pressed, error code E15 displays.
When a function button on hand control is pressed, the motor pump does not run and no error codes display.
Foot control operates fine sometimes, but not every time.
When foot switch on foot control is pressed, nothing happens (hand control display should illuminate when foot switch is depressed).
Button board is malfunctioning.
Button board is malfunctioning. One of two button switches for a function is stuck in closed position or will not close.
Hand control board is malfunctioning.
Button board is malfunctioning.
Main controller board is malfunctioning.
Wires or wire connections between main controller board, distribution board, and hand control port are broken, loose or dirty.
Table is in reverse orientation.
The hand control or emergency override panel is being used (hand control and / or emergency override panel takes priority over foot control).
One of two foot switches for a function is stuck in closed position or will not close.
Replace suspect button board with known working button board.
Perform a continuity check on both button switches of suspect function (when a button switch is pressed, there should be continuity between its two terminals).
Replace suspect hand control board with known working hand control board.
Replace suspect button board with known working button board.
Replace suspect main controller board with known working main controller board.
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
Foot control is locked out from being used when table top is in reverse orientation - foot control may only be used when table top is in normal orientation.
Check if problem occurrs only when hand control or emergency override panel is being used.
See Figure 5-1 for this check. Perform a continuity check on both foot switches of suspect function (when foot switch is pressed, there should be continuity between its two terminals).
Replace button board. Refer to para 4.4.
If check fails, replace button board. Refer to para 4.4.
Replace hand control board. Refer to para 4.5.
Replace button board. Refer to para 4.4.
Replace main controller board. Refer to para 4.31.
Clean any dirty connections. Tighten any loose connections. Replace any broken connectors or wires.
Use hand control or set up table top in normal orientation. Inform operator that Foot Control cannot be used when table is in reverse orientation.
Do not use foot control simultaneously with hand control or emergency override panel. Inform operator of Emergency Override Panel, Hand Control, and Foot Control priorities.
If check fails, replace foot switch pad.
© Schaerer Mayfield USA, Inc. 2004
Page 2-8Printed in U.S.A.
Problem Symptom Probable Cause Check Correction
Return To Table Of Contents
Foot control does not work
-Continued
Error code E01 (Floor Lock Status Switch Is Not Responding) displays on hand control.
Table 2-1. Troubleshooting Guide
When foot switch on foot control is pressed, nothing happens (hand control display should illuminate when foot switch is depressed) -Continued.
Floor lock cylinders do not extend or retract properly and error code E01 appears on hand control.
Floor lock cylinders extend properly, but error code E01 appears on hand control.
TESTING AND TROUBLESHOOTING
Foot control interface board is malfunctioning.
10 amp fuse on main controller board is blown.
Main controller board or foot control interface board is malfunctioning.
Wires or wire connections between main controller board, distribution board, and hand control port are broken, loose or dirty.
Check valve(s) for main floor lock cylinder(s) or outrigger floor lock cylinder(s) are stuck open, allowing floor lock cylinders to retract after extending.
Foreign matter in main floor lock or outrigger floor lock valve spool.
Outrigger pressure relief valve is malfunctioning, preventing outrigger cylinders from extending fully.
Inside surfaces of main floor lock cylinder(s) or outrigger floor lock cylinder(s) are dirty / covered with foreign matter, which prevents the cylinder(s) from extending or retracting fully.
Main floor lock status switch remains tripped after main floor locks have extended (out of adjustment).
Main floor lock status switch is malfunctioning - stuck open.
Replace suspect foot control interface board with known working foot control interface board.
Perform continuity check on 10 amp fuse.
See Figure 5-1 for this check. Check for 5.0 VDC across Test Points P and W. If 5.0 VDC is present, check for 5.0 VDC across Test Points W and X within 10 seconds of pressing a foot button on foot control.
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
Remove check valve(s) and check for foreign matter.
Remove valve spool(s) and check for foreign matter. Flush out valve spool manifold with oil.
Replace suspect pressure relief valve with known working pressure relief valve.
Check for foreign matter on inside surfaces of floor lock cylinders.
Check if status switch is tripped.
See Figure 5-1 for this check. Perform continuity check on normally closed status switch (switch untripped = continuity).
SECTION II
Replace foot control interface board. Refer to para 4.44.
Replace 10 amp fuse. See Figure 5-1 for fuse location.
If 5.0 VDC is not present across Test Points P and W, the main controller board is malfunctioning. If 5.0 VDC is not present across Test Points W and X during check, then main controller board or foot control interface board is malfunctioning. Refer to para
4.31 or 4.44. Clean any dirty connections.
Tighten any loose connections. Replace any broken connectors or wires.
Remove and clean check valve(s). Replace any damaged pilot pistons or check valves.
Push on manual release pin of valve spool. If still stuck, remove valve spool(s) and check for foreign matter. Flush out valve spool manifold with oil.
Replace outrigger pressure relief valve and spring. Refer to para 4.30.
Disassemble and clean floor lock cylinder(s).
If necessary, adjust main floor lock status switch. Refer to para 4.42.
If malfunctioning, replace main floor lock status switch. Refer to para 4.42.
© Schaerer Mayfield USA, Inc. 2004
Page 2-9Printed in U.S.A.
SECTION II
Return To Table Of Contents
TESTING AND TROUBLESHOOTING
Table 2-1. Troubleshooting Guide
Problem Symptom Probable Cause Check Correction
Error code E01 (Floor Lock Status Switch Is Not Responding) displays on hand control - Continued.
Error Code E02 (Overcurrent - Motor #1) displays on hand control.
Error code E03 (Overcurrent - Motor #2) displays on hand control.
Error code E04 (Motor Driver Failure) displays on hand control.
Floor lock cylinders do not extend or retract properly and error code E01 appears on hand control - Continued.
Floor lock cylinders do not retract properly and Error code E01 appears on hand control.
When function is selected, error code E02 appears on hand control.
When function is selected, error code E03 appears on hand control.
When function is selected, error code E04 appears on hand control.
Outrigger floor lock status switch remains tripped after outrigger floor locks have extended (out of adjustment).
Outrigger floor lock status switch is malfunctioning - stuck open.
Wires or wire connections between floor lock status switches and main controller board are broken, loose or dirty.
One of the floor lock status switches is malfunctioning.
Retaining ring which holds ouside floor lock sleeve onto floor lock cylinder piston is missing.
Motor #1 is malfunctioning. Replace suspect motor #1
Charging / power driver board is malfunctioning.
Motor #2 is malfunctioning. Replace suspect motor #2
Charging / power driver board is malfunctioning.
Wires or wire connections between charging / power driver board and pumps are broken, loose, or dirty.
Charging / power driver board is malfunctioning.
Motor is burned out (most likely motor #1).
Check if status switch is tripped.
See Figure 5-1 for this check. Perform continuity check on normally closed status switch (switch untripped = continuity).
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
Perform continuity check on N.C. floor lock status switch (switch untripped = continuity).
Check to see if retaining ring is missing.
with known working motor pump.
Replace suspect charging / power driver board with known working board.
with known working motor pump.
Replace suspect charging / power driver board with known working board.
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
Replace suspect charging / power driver board with known working board.
Replace suspect motor with known working motor. (If only COLUMN UP function causes error code E04, motor #2 could be malfunctioning - motor #2 only runs for this function during normal operation).
If necessary, adjust outrigger floor lock status switch. Refer to para 4.42.
If malfunctioning, replace outrigger floor lock status switch. Refer to para 4.42.
Clean any dirty connections. Tighten any loose connections. Replace any broken connectors or wires.
If necessary, adjust or replace floor lock status switch. Refer to para 4.42.
Replace retaining ring.
Replace motor #1. Refer to para 4.26.
Replace charging / power driver board. Refer to para
4.32. Replace motor #2. Refer to
para 4.26.
Replace charging / power driver board. Refer to para
4.32. Clean any dirty connections.
Tighten any loose connections. Replace any broken connectors or wires.
Replace charging / power driver board. Refer to para
4.32. Replace motor. Refer to
para 4.26.
© Schaerer Mayfield USA, Inc. 2004
Page 2-10Printed in U.S.A.
Problem Symptom Probable Cause Check Correction
Return To Table Of Contents
Error code E05 (Valve Driver Failure) displays on hand control.
Error code E06 (Footswitch Failure) displays on hand control.
Error code E07 (Trendelenburg Position Sensor Failure) displays on hand control.
Table 2-1. Troubleshooting Guide
When ENABLE button is pressed, or a function button is selected, error code E05 appears on hand control.
When a footswitch button is depressed, error code E06 appears on hand control.
When ENABLE button is pressed, or a function button is selected, error code E07 appears on hand control.
TESTING AND TROUBLESHOOTING
A valve spool solenoid is malfunctioning.
Wires or wire connections between valve spool solenoid and main controller board are broken, loose or dirty.
Main controller board is malfunctioning - valve driver circuitry is malfunctioning.
One of two foot switches for a function is stuck in closed position or will not complete circuit when depressed.
Foot control interface board is malfunctioning.
Wires or wire connections between foot switches and foot control interface board are broken, loose or dirty.
Trendelenburg position sensor is malfunctioning.
Wires or wire connections between Trendelenburg position sensor, distribution board, and main controller board are broken, loose, or dirty.
Main controller board is malfunctioning.
Press and hold the SYSTEM OVERRIDE button on the emergency panel, while pressing, one at a time, all the remaining function buttons on the emergency panel (except for DISABLE button). All functions should operate except for function with malfunctioning valve spool solenoid.
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
Replace suspect main controller board with known working main controller board.
See Figure 5-1 for this check. Perform a continuity check on both foot switches of suspect function (switch not depressed = open / switch depressed = continuity.
Replace suspect foot control interface board with known working foot control interface board.
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
Replace suspect Trendelenburg position sensor with known working position sensor.
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
Replace suspect main controller board with known working main controller board.
SECTION II
Replace malfunctioning valve spool solenoid. Refer to para
4.22 or 4.23.
Clean any dirty connections. Tighten any loose connections. Replace any broken connectors or wires.
Replace main controller board. Refer to para 4.31.
If check fails, replace foot switch pad.
Replace foot control interface board. Refer to para 4.44.
Clean any dirty connections. Tighten any loose connections. Replace any broken connectors or wires.
Replace Trendelenburg position sensor. Refer to para
4.11.
Clean any dirty connections. Tighten any loose connections. Replace any broken connectors or wires.
Replace main controller board. Refer to para 4.31.
© Schaerer Mayfield USA, Inc. 2004
Page 2-11Printed in U.S.A.
SECTION II
Return To Table Of Contents
TESTING AND TROUBLESHOOTING
Table 2-1. Troubleshooting Guide
Problem Symptom Probable Cause Check Correction
Error code E08 (Tilt Position Sensor Failure) displays on hand control.
Error code E09 (Seat Position Sensor Failure) displays on hand control.
Error code E11 (Internal RAM / Register Failure) displays on hand control.
Error code E12 (Calibrating data Failure) displays on hand control.
When ENABLE button is pressed, or a function button is selected, error code E08 appears on hand control.
When ENABLE button is pressed, or a function button is selected, error code E09 appears on hand control.
When ENABLE button is pressed, error code E11 appears on hand control.
When ENABLE button is pressed, error code E12 appears on hand control.
Tilt position sensor is malfunctioning.
Wires or wire connections between tilt position sensor, distribution board, and main controller board are broken, loose or dirty.
Main controller board is malfunctioning.
Seat position sensor is malfunctioning.
Wires or wire connections between seat position sensor, distribution board, and main controller board are broken, loose or dirty.
Main controller board is malfunctioning.
Battery power level is too low to properly power table.
Main controller board is malfunctioning.
Replace suspect tilt position sensor with known working position sensor.
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
Replace suspect main controller board with known working main controller board.
Replace suspect seat position sensor with known working position sensor.
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
Replace suspect main controller board with known working main controller board.
Plug power cord into table and press ENABLE button on hand control to see if error code E11 displays now.
Replace suspect main controller board with known working main controller board.
Replace tilt position sensor. Refer to para 4.15.
Clean any dirty connections. Tighten any loose connections. Replace any broken connectors or wires.
Replace main controller board. Refer to para 4.31.
Replace seat position sensor. Refer to para 4.19.
Clean any dirty connections. Tighten any loose connections. Replace any broken connectors or wires.
Replace main controller board. Refer to para 4.31.
If error code still appears, recharge batteries.
Replace main controller board. Refer to para 4.31.
Error code E13 (Overheat
- Motor #1) displays on hand control.
© Schaerer Mayfield USA, Inc. 2004
When ENABLE button is pressed, error code E13 appears on hand control.
Motor #1 is overheated and has tripped its primary thermostat.
Wait 60 to 90 minutes to allow motor to cool down and the primary thermostat to reset.
After table is operational again, check to see if motor #1 overheats continually.
Wait 60 to 90 minutes to allow motor #1 to cool down and the primary thermostat to reset. Replace primary thermostat. Refer to para
4.40. If motor #1 overheats
continually (E13), replace primary thermostat. Refer to para 4.40. If motor #1 still continues to overheat (E13), replace motor. Refer to para
4.26.
Page 2-12Printed in U.S.A.
Problem Symptom Probable Cause Check Correction
Return To Table Of Contents
Error code E13 (Overheat
- Motor #1) displays on hand control - Continued.
Error code E14 (Overheat
- Motor #2) displays on hand control.
Error code E15 (Hand Control Button Switch Failure) displays on hand control.
When ENABLE and UNLOCK buttons on hand control are pressed, floor lock cylinders do not retract.
Table 2-1. Troubleshooting Guide
When ENABLE button is pressed, error code E13 appears on hand control ­Continued.
When ENABLE button is pressed, error code E14 appears on hand control.
When ENABLE button is pressed, error code E15 appears on hand control.
Main floor lock cylinders do not retract properly.
TESTING AND TROUBLESHOOTING
Wires or wire connections between motor #1 primary thermostat and main controller board are broken, loose, or dirty.
Motor #2 is overheated and has tripped its primary thermostat.
Wires or wire connections between motor #2 primary thermostat and main controller board are broken, loose, or dirty.
One of two button switches for a function is stuck in closed position or will not complete circuit when pressed.
Hand control board is malfunctioning.
Wires or wire connections between hand control board, hand control port, distribution board, and main control board are broken, loose, or dirty.
All floor lock valve spool is clogged.
Inside surfaces of main floor lock cylinder is dirty / covered with foreign matter, which prevents the cylinder from retracting fully.
Pilot operated check valve is malfunctioning - not opening check valve or check valve is clogged.
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
Wait 60 to 90 minutes to allow motor to cool down and the primary thermostat to reset.
After table is operational again, check to see if motor #2 overheats continually.
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
See Figure 5-1 for this check. Perform a continuity check on both button switches of suspect function (button switch not depressed = open / button switch depressed = continuity.
Replace suspect hand control board with known working hand control board.
See Figure 5-1 for this check. Perform a continuity check on all suspect wires or connections.
Check for foreign matter in valve spool.
Check for foreign matter on inside surfaces of main floor lock cylinders.
Check for broken or bent components or foreign matter in pilot operated check valve.
SECTION II
Clean any dirty connections. Tighten any loose connections. Replace any broken connectors or wires.
Wait 60 to 90 minutes to allow motor to cool down and the primary thermostat to reset. Replace primary thermostat. Refer to para 4.40.
If motor #2 overheats continually (E14), replace primary thermostat. Refer to para 4.40. If motor #2 still continues to overheat (E14), replace motor. Refer to para
4.26. Clean any dirty connections.
Tighten any loose connections. Replace any broken connectors or wires.
If check fails, replace button board. Refer to para 4.4.
Replace hand control board. Refer to para 4.5.
Clean any dirty connections. Tighten any loose connections. Replace any broken connectors or wires.
Remove valve spool and clean valve spool ports out. Replace valve spool component if necessary.
Disassemble and clean main floor lock cylinders.
Remove pilot operated check valve components. Clean out any foreign matter. Replace any broken or bent components.
© Schaerer Mayfield USA, Inc. 2004
Page 2-13Printed in U.S.A.
SECTION II
Return To Table Of Contents
TESTING AND TROUBLESHOOTING
Table 2-1. Troubleshooting Guide
Problem Symptom Probable Cause Check Correction
When ENABLE and UNLOCK buttons on hand control are pressed, floor lock cylinders do not retract - Continued.
All functions run too slow. No error codes are
One function is too slow; the rest are operating properly.
One function is too fast; the rest are operating properly.
Outrigger floor lock cylinders do not retract properly.
displayed, but all functions are too slow or do not extend or retract fully.
No error codes are displayed, but function is too slow or does not extend or retract fully.
No error codes are displayed, but column function moves too slowly or does not extend or retract fully.
No error codes are displayed, but function is too fast.
All floor lock valve spool is clogged.
Inside surfaces of outrigger floor lock cylinder is dirty / covered with foreign matter, which prevents the cylinder from retracting fully.
Pilot operated check valve(s) is malfunctioning - not opening check valve or check valve is clogged.
Motor pump is worn or leaking.
Motor pump pressure relief valve is malfunctioning.
Discharge filter is clogged with foreign matter.
Reservoir strainer is clogged with foreign matter.
Valve spool for suspect function is clogged.
One of pilot operated check valves for suspect function is malfunctioning - not opening check valve or check valve is clogged.
One of orifices for suspect function is clogged with foreign matter (does not apply to column cylinder or floor lock cylinders).
Cylinder for suspect function is malfunctioning.
Column plates and wear pads are binding.
One of orifices for suspect function is missing ( NOTE: only applies to early units which have orifices external to cylinders.)
Check for foreign matter in valve spool.
Check for foreign matter on inside surfaces of outrigger floor lock cylinders.
Check for broken or bent components or foreign matter in pilot operated check valve.
Replace suspect motor pump with known working motor pump.
Replace suspect pressure relief valve with known working pressure relief valve.
_ Replace discharge filter.
Visually check reservoir strainer for excessive foreign matter.
Check for foreign matter in valve spool.
Check for broken or bent components or foreign matter in pilot operated check valve.
Remove orifice and check for foreign matter.
Replace suspect cylinder with known working cylinder.
Check for foreign matter, corrosion, nicks, scratches, or other damage to wear surfaces of column plates or wear pads.
Check for missing orifice. Replace orifice.
Remove valve spool and clean valve spool ports out. Replace valve spool component if necessary.
Disassemble and clean outrigger floor lock cylinders.
Remove pilot operated check valve components. Clean out any foreign matter. Replace any broken or bent components.
Replace motor pump. Refer to para 4.26 or 4.27.
Replace pressure relief valve and spring. Refer to para
4.29.
Refer to para 4.24. Replace reservoir strainer.
Refer to para 4.27.
Remove valve spool and clean valve spool ports out. Replace valve spool component if necessary.
Remove pilot operated check valve components. Clean out any foreign matter. Replace any broken or bent components.
Remove orifice and clean.
Replace malfunctioning cylinder.
Replace column plates or wear pads.
© Schaerer Mayfield USA, Inc. 2004
Page 2-14Printed in U.S.A.
Loading...
+ 201 hidden pages