Ohaus VX User Manual

3000 Series Indicators Instruction Manual
T32XW Indicator
i
ii
3000 Series Indicators
TABLE OF CONTENTS
2.4 Mounting Bracket ....................................................................................................................................EN-12
EN-1
EN-2
3000 Series Indicators
TABLE OF CONTENTS (Cont.)
3000 Series Indicators
TABLE OF CONTENTS (Cont.)
4.6.1 Weighing .....................................................................................................................................EN-28
6.3.1 Physical Seals .............................................................................................................................. EN-32
6.3.2 Audit Trail Seal .............................................................................................................................EN-33
EN-3
EN-4
3000 Series Indicators
1. INTRODUCTION
This manual contains installation, operation and maintenance instructions for the T32XW Indicator. Please read this manual completely before installation and operation.
1.1 Safety Precautions
For safe and dependable operation of this equipment, please comply with the following safety precautions:
• Verify that the input voltage range printed on the data label matches the local AC power to be used.
• Make sure that the power cord does not pose a potential obstacle or tripping hazard.
• Use only approved accessories and peripherals.
• Operate the equipment only under ambient conditions specified in these instructions.
• Disconnect the equipment from the power supply before cleaning.
• Do not operate the equipment in hazardous or unstable environments.
• Do not immerse the equipment in water or other liquids.
• Service should only be performed by authorized personnel.
• The Indicator is supplied with a grounded power cable. Use only with a compatible grounded power outlet.
3000 Series Indicators
1.2 Overview of Parts and Controls
EN-5
1
2
3
4
5
6
TABLE 1-1. T32XW PARTS.
Item Description
1 Data Label
2 Front Housing
3 Adjusting Knob (2)
4 Control Panel
5 Mounting Bracket
6 Screw (4)
7 Rear Housing
8 Safety Data Label
9 Strain Relief for RS232
10 Strain Relief for Load Cell
Cable
11 Power cord
7
8
9 10
Figure 1-1. Indicator.
11
EN-6
1
2
3
4
7
5
6
8
J8
J1
J5
J7
W1
W2
J2
GND
LFT OFF
RXD
-EXE
-SEN
-SIG
CGND
+SIG
LFT ON
+SEN
TXD
+EXE
TXD GND
RXD
+EXC +SIG -SIG -EXC +SENS GND -SENS
1.2 Overview of Parts and Controls (Cont.)
3000 Series Indicators
LOAD CELL WIRING
RS232 WIRING
Figure 1-2. Main PC Board.
TABLE 1-2. MAIN PC BOARD.
Item Description
1 Keypad Connector J8
2 Battery Connector
3 Line Power Input
4 Sense Jumper W2
5 Load Cell Terminal Block J5
6 Sense Jumper W1
7 RS232 Terminal Block J7
8 LFT On / Off Switch
3000 Series Indicators
1
2
3
4
56
78
910
11
12
1314
15
1.2 Overview of Parts and Controls (Cont.)
EN-7
Figure 1-3. Controls and Indicators.
TABLE 1-3. CONTROL PANEL.
No. Designation
1 Capacity Label Window
2 Pound symbol
3 Ounce symbol
4 Kilogram, gram symbols
5 TARE
6 TARE symbol
7 FUNCTION
8 NET symbol
9 PCS symbol
10 PRINT
11 Calibration Mode symbol
12 Battery symbol
13 ON/ZERO
14 Center of Zero symbol
15 Stable weight indicator
Menu
Units
Off
button
Mode
button
button
button
EN-8
1.3 Control Functions
Button
3000 Series Indicators
TABLE 1-4. CONTROL FUNCTIONS.
Primary Function (Short Press)
Secondary Function (Long Press)
Menu Function (Short Press)
ON/ZERO
If Indicator is On, sets zero.
Off
Turns the Indicator on or off.
Yes
Accepts the current setting on the display.
PRINT
Sends the current value to the COM port if AUTOPRINT is set to Off.
Units
Changes the weighing Unit.
No
Advances to the next menu or menu item.
Rejects the current setting on the display and advances to the next available setting.
Increments the value.
FUNCTION
Initiates an application mode.
Mode
Allows changing the application mode.
Press and hold allows scrolling through modes.
Back
Moves Back to previous menu item.
Decrements the value.
TARE
Performs a tare operation.
Menu
Enter the User menu.
View the Audit Trail event counters (extended press)
Exit
Exits the User menu.
Aborts the calibration in progress.
3000 Series Indicators
1
2
3
4
5
6
78
9
2. INSTALLATION
2.1 Unpacking
Unpack the following items:
• Indicator
• Mounting Bracket
• Knobs (2)
• Capacity Label Sheet
• Instruction Manual CD
• Warranty Card
• LFT sealing Kit
2.2 External Connections
2.2.1 RS232 interface Cable to the indicator
Connect the optional RS232 cable to the RS232 connector Figure 1-1, item 9).
EN-9
Pin Connection
1 N/C
2 TXD
3 RXD
4 N/C
5 GND
6 N/C
7 N/C
8 N/C
9 N/C
2.2.2 AC Power
Connect the AC plug to a properly grounded electrical outlet.
Figure 2-1. RS232 Pins.
EN-10
3000 Series Indicators
2.2.3 Battery Power
The indicator can be operated on the internal rechargeable battery when AC power is not available. The indicator will automatically switch to battery operation if there is a power failure or the power cord is removed.
Note: Before using the indicator for the first time, the internal rechargeable battery should be fully charged for up to 12 hours. The indicator can be operated during the charging process. The battery is protected against over charging and the indicator can remain connected to the AC power line.
Connect AC power to the indicator and allow it to charge. While the battery is charging, the triangle above the battery function symbol will light. When the battery is fully charged, this triangle will disappear.
The indicator can operate for up to 100 hours on a fully charged battery.
During battery operation, a flashing triangle above the battery function symbol indicates the battery is low and requires recharging. Approximately 60 minutes of operation will remain when the battery symbol starts to blink. The indicator will display Lo.BAT and automatically turn off when the battery is fully discharged.
CAUTION
BATTERY IS TO BE REPLACED ONLY BY AN AUTHORIZED OHAUS SERVICE DEALER.
RISK OF EXPLOSION CAN OCCUR IF REPLACED WITH THE WRONG TYPE OR CONNECTED IMPROPERLY.
Dispose of the lead acid battery according to local laws and regulations.
2.2.4 Mounting Bracket to the Indicator
Align the mounting bracket over the threaded holes in the side of the indicator and install the knobs. Adjust the indicator to the desired angle and tighten the knobs.
2.3 Internal Connections
Some connections require the housing to be opened.
2.3.1 Opening the Housing
CAUTION: ELECTRICAL SHOCK HAZARD. REMOVE ALL POWER CONNECTIONS TO THE INDICATOR BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS. THE HOUSING SHOULD ONLY BE OPENED BY AUTHORIZED AND qUALIFIED PERSONNEL, SUCH AS AN ELECTRICAL TECHNICIAN.
3000 Series Indicators
Remove the four hex head screws from the rear housing. Open the housing by carefully pulling the top of the front housing forward. Once all connections are made, reattach the front housing. The screws should be tightened fully to maintain a watertight seal.
2.3.2 Scale Base to the Indicator
Pass the load cell cable through the strain relief (Figure 1-1, item 10) and attach it to terminal block J5 (Figure 1-2, item 5). Re-tighten the strain relief to ensure a watertight seal.
Jumper Connections
For a 4-wire load cell with no sense wires: Jumpers W1 and W2 must be shorted. For a 6-wire load cell that includes sense wires, see Figure 2-2. Jumpers W1 and W2 must be opened. For load cells with an extra ground shield wire: Connect the shield to the center position (GND) of J5.
EN-11
Pin Connection
J5-1 +EXCITATION
J5-2 +SENSE
J5-3 +SIGNAL
J5-4 GND
J5-5 -SIGNAL
J5-6 -SENSE
J5-7 -EXCITATION
Figure 2-2. Jumper Connections.
After wiring is completed and jumpers are in place, replace the indicator housing screws. Make sure the strain relief is properly tightened.
2.3.3 RS232 Interface Cable to the indicator
Pass the optional RS232 cable through the strain relief (Figure 1-1, item 9) and attach it to terminal block J7 (Figure 1-2, item 7). Re-tighten the strain relief to ensure a water tight seal.
Pin Connection
J7-1 TXD
J7-2 RXD
J7-3 GND
EN-12
3000 Series Indicators
2.4 Mounting Bracket
Attach the bracket to a wall or table using fasteners (not supplied) that are appropriate for the type of mounting surface. The bracket will accommodate up to 6 mm (1/4”) diameter screws. Locate the mounting holes as shown in Figure 2-3.
Figure 2-3. Mounting Bracket Dimensions.
3000 Series Indicators
3 SETTINGS
3.1 Menu Structure
TABLE 3-1. MENU STRUCTURE.
CALIBRATIONSETUP
SPAN RESET RESET RESET RESET LINEARITY NO NO NO NO GEO YES YES YES YES 0…31 LEGAL FOR TRADE AVERAGING COUNT KILOGRAM END CAL OFF LOW OFF OFF
 ON  MEDIUM  ON  ON  CALIBRATION UNIT  HI  END MODE  POUND
 KILOGRAM  AUTO ZERO  OFF  POUND  OFF  ON  CAPACITY  0.5d  GRAM  5…20000  1d  OFF  GRADUATION  3d  ON  0.001…20  EXPAND MODE  OUNCE  POWER ON UNIT  OFF  OFF  AUTO  ON  ON  GRAM  BACKLIGHT  POUND OUNCE  KILOGRAM  AUTO  OFF  POUND  ON  ON  OUNCE  OFF  END UNIT  POUND OUNCE  AUTO OFF  ZERO RANGE  OFF  0%  SET 1  2%  SET 2  100%  SET 5
END SETUP END READOUT
READOUT
MODE
UNIT
PRINT RESET RESET
 NO  NO  YES  YES  BAUD  LOCK CAL  300, …19200  OFF  PARITY  ON  7 EVEN  LOCK SETUP  7 ODD  OFF  7 NONE  ON  8 NONE  LOCK READOUT  STOP  OFF  1  ON  2  LOCK MODE  HANDSHAKE  OFF  OFF  ON  XON-XOFF  LOCK UNIT  STABLE ONLY  OFF  OFF  ON  ON  LOCK PRINT  AUTO PRINT  OFF  OFF  ON  WHEN STABLE  END MENU LOCK  LOAD  LOAD AND ZERO  INTERVAL  1…3600  CONTINUOUS  CONTENT  GROSS  NET  TARE  UNIT
END PRINT
MENU LOCK
EN-13
END
EN-14 3000 Series Indicators
3.2 Menu Navigation
TO ENTER THE MENU MODE
Press and hold the Menu button until MENU appears on the display. The first upper level menu appears on the display. Summary of button navigation functions in menu mode:
--Yes Allows entry into the displayed menu.
- Accepts the displayed setting and advances to the next menu item.
--No Skips by the displayed menu.
- Rejects the displayed setting or menu item and advances to the next available item.
--Back Moves backwards through the upper and middle level menus.
- Backs out of a list of selectable items to the previous middle level menu.
--Exit Exits from menu directly to the active weighing mode.
3.3 Calibration Menu
Two calibration processes are available: Span Calibration and Linearity Calibration.
NOTES:
1. Make sure that appropriate calibration masses are available before beginning calibration.
2. Make sure that the scale base is level and stable during the entire calibration process.
3. Calibration is unavailable with LFT set to On.
4. Allow the Indicator to warm up for approximately 5 minutes after stabilizing to room temperature.
5. To abort calibration, press the Exit button anytime during the calibration process.
Span Perform Linearity Perform Geographic Adjustment Set 00…Set 19… Set 31 End Calibration Exit CALIBRATE menu
3.3.1 Span Calibration
Span Calibration uses two points to adjust the scale. The first point is the zero value where there is no weight on the scale. The second point is the Span value where a calibration mass is placed on the scale.
When SPAN is displayed, press the Yes button to access the Span Calibration menu item.
The display flashes 0.
With no weight on the scale, press the Yes button to establish the zero point.
The display shows --C-- while the zero point is established.
The display flashes the span calibration point. Place the specified weight on the scale and press the Yes button.
To choose a different span point, repeatedly press the No button to increment the selections or press the Back button to decrement the selections. Refer to Table 3-3 for available span points. When the desired value is displayed, place the specified weight on the scale and press the Yes button.
EN-153000 Series Indicators
The display shows --C-- while the span point is established.
If span calibration was successful, the scale exits to the active weighing mode and displays the actual weight value.
3.3.2 Linearity Calibration
Linearity calibration uses 3 calibration points. The first calibration point is established with no weight on the scale. The second calibration point is established at approximately half capacity. The third calibration point is established at capacity. The Linearity calibration points are fixed and cannot be altered by the user during the calibration procedure. Refer to Table 3-3 for the linearity points.
When LINEAr is displayed, press the Yes button to access the Linearity Calibration menu item.
The display flashes 0. With no weight on the scale, press the Yes button to establish the zero point.
The display shows --C-- while the zero point is established.
The display flashes the mid calibration point.
Place the specified weight on the scale and press the Yes button.
The display shows --C-- while the mid point is established.
The display flashes the full calibration point.
Place the specified weight on the scale and press the Yes button.
The display shows --C-- while the full point is established.
If linearity calibration was successful, the scale exits to the active weighing mode and displays the actual weight value.
EN-16 3000 Series Indicators
3.3.3 Geographical Adjustment Factor
The Geographcial Adjustment Factor (GEO) is used to compensate for variations in gravity.
Note: Changing the GEO Factor alters the calibration. The GEO value was set at the factory and should only be changed by an authorized manufacturer’s representative or certified verirication personnel.
Refer to table 3-2 to determine the GEO factor that corresponds to your location.
3.3.4 End Calibration
Advance to the next menu.
EN-173000 Series Indicators
TABLE 3-2. GEOGRAPHICAL ADJUSTMENT VALUES
Elevation in meters
0 325 650 975 1300 1625 1950 2275 2600 2925 3250
325 650 975 1300 1625 1950 2275 2600 2925 3250 3575
Elevation in feet
0 1060 2130 3200 4260 5330 6400 7460 8530 9600 10660
1060 2130 3200 4260 5330 6400 7460 8530 9600 10660 11730
Latitude GEO value 0°00’ 5°46’ 5 4 4 3 3 2 2 1 1 0 0 5°46’ 9°52’ 5 5 4 4 3 3 2 2 1 1 0 9°52’ 12°44’ 6 5 5 4 4 3 3 2 2 1 1 12°44’ 15°06’ 6 6 5 5 4 4 3 3 2 2 1 15°06’ 17°10’ 7 6 6 5 5 4 4 3 3 2 2 17°10’ 19°02’ 7 7 6 6 5 5 4 4 3 3 2 19°02’ 20°45’ 8 7 7 6 6 5 5 4 4 3 3 20°45’ 22°22’ 8 8 7 7 6 6 5 5 4 4 3 22°22’ 23°54’ 9 8 8 7 7 6 6 5 5 4 4 23°54’ 25°21’ 9 9 8 8 7 7 6 6 5 5 4 25°21’ 26°45’ 10 9 9 8 8 7 7 6 6 5 5 26°45’ 28°06’ 10 10 9 9 8 8 7 7 6 6 5 28°06’ 29°25’ 11 10 10 9 9 8 8 7 7 6 6 29°25’ 30°41’ 11 11 10 10 9 9 8 8 7 7 7 30°41’ 31°56’ 12 11 11 10 10 9 9 8 8 7 7 31°56’ 33°09’ 12 12 11 11 10 10 9 9 8 8 7 33°09’ 34°21’ 13 12 12 11 11 10 10 9 9 8 8 34°21’ 35°31’ 13 13 12 12 11 11 10 10 9 9 8 35°31’ 36°41’ 14 13 13 12 12 11 11 10 10 9 9 36°41’ 37°50’ 14 14 13 13 12 12 11 11 10 10 9 37°50’ 38°58’ 15 14 14 13 13 12 12 11 11 10 10 38°58’ 40°05’ 15 15 14 14 13 13 12 12 11 11 10 40°05’ 41°12’ 16 15 15 14 14 13 13 12 12 11 11 41°12’ 42°19’ 16 16 15 15 14 14 13 13 12 12 11 42°19’ 43°26’ 17 16 16 15 15 14 14 13 13 12 12 43°26’ 44°32’ 17 17 16 16 15 15 14 14 13 13 12 44°32’ 45°38’ 18 17 17 16 16 15 15 14 14 13 13 45°38’ 46°45’ 18 18 17 17 16 16 15 15 14 14 13 46°45’ 47°51’ 19 18 18 17 17 16 16 15 15 14 14 47°51’ 48°58’ 19 19 18 18 17 17 16 16 15 15 14 48°58’ 50°06’ 20 19 19 18 18 17 17 16 16 15 15 50°06’ 51°13’ 20 20 19 19 18 18 17 17 16 16 15 51°13’ 52°22’ 21 20 20 19 19 18 18 17 17 16 16 52°22’ 53°31’ 21 21 20 20 19 19 18 18 17 17 16 53°31’ 54°41’ 22 21 21 20 20 19 19 18 18 17 17 54°41’ 55°52’ 22 22 21 21 20 20 19 19 18 18 17 55°52’ 57°04’ 23 22 22 21 21 20 20 19 19 18 18 57°04’ 58°17’ 23 23 22 22 21 21 20 20 19 19 18 58°17’ 59°32’ 24 23 23 22 22 21 21 20 20 19 19 59°32’ 60°49’ 24 24 23 23 22 22 21 21 20 20 19 60°49’ 62°90’ 25 24 24 23 23 22 22 21 21 20 20 62°90’ 63°30’ 25 25 24 24 23 23 22 22 21 21 20 63°30’ 64°55’ 26 25 25 24 24 23 23 22 22 21 21 64°55’ 66°24’ 26 26 25 25 24 24 23 23 22 22 21 66°24’ 67°57’ 27 26 26 25 25 24 24 23 23 22 22 67°57’ 69°35’ 27 27 26 26 25 25 24 24 23 23 22 69°35’ 71°21’ 28 27 27 26 26 25 25 24 24 23 23 71°21’ 73°16’ 28 28 27 27 26 26 25 25 24 24 23 73°16’ 75°24’ 29 28 28 27 27 26 26 25 25 24 24 75°24’ 77°52’ 29 29 28 28 27 27 26 26 25 25 24 77°52’ 80°56’ 30 29 29 28 28 27 27 26 26 25 25 80°56’ 85°45’ 30 30 29 29 28 28 27 27 26 26 25 85°45’ 90°00’ 31 30 30 29 29 28 28 27 27 26 26
EN-18 3000 Series Indicators
3.4 Setup Menu
When the Indicator is used for the first time, enter this menu to set the Capacity and Graduation.
Reset No, Yes Legal for Trade Off, On Cal Unit kg, lb Capacity 5…20000 Graduation 0.001…20 Power On Unit g, kg, lb, oz, lb:oz, Auto Zero Range 0%, 2%, 100% End Setup Exit SETUP menu
3.4.1 Reset
Reset the Setup menu to the factory defaults. No = not reset. Yes = reset.
NOTE: If the Legal for Trade menu item is set to ON, the Capacity, Graduation, Zero Range and Legal For Trade settings are not reset.
3.4.2 Legal for Trade
Set the legal for trade status. OFF = off ON = on Turning on the “LFT” menu setting has the following effects:
• Zero-range is set and locked on “2”.
• Auto Zero Tracking is set and locked on 0.5d
• The lb:oz unit is not available as a power-on setting.
3.4.3 Calibration Unit
Set the unit during calibration. CAL UN kg = Calibrate using kg weights CAL UN lb = Calibrate using pound weights
3.4.4 Capacity
Set the scale capacity from 5 to 20000. Refer to the Setup Table 3.3 for available settings.
TABLE 3-3. SETUP AND CALIBRATION VALUES
EN-193000 Series Indicators
Capacity Graduation size
with LFT OFF
5 0.0005, 0.001, 0.002,
0.005
10 0.0005, 0.001, 0.002,
0.005, 0.01
15 0.001, 0.002, 0.005, 0.01 0.005, 0.01 5, 10, 15 5, 15 20 0.001, 0.002, 0.005, 0.01,
0.02
25 0.002, 0.005, 0.01, 0.02 0.005, 0.01, 0.02 5, 10, 15, 20, 25 10, 25 30 0.002, 0.005, 0.01, 0.02 0.005, 0.01, 0.02 5, 10, 15, 20, 25, 30 15, 30 40 0.002, 0.005, 0.01, 0.02 0.01, 0.02 5, 10, 15, 20, 25, 30, 40 20, 40 50 0.005, 0.01, 0.02, 0.05 0.01, 0.02, 0.05 5, 10, 15, 20, 25, 30, 40, 50 25, 50 60 0.005, 0.01, 0.02, 0.05 0.01, 0.02, 0.05 5, 10, 15, 20, 25, 30, 40, 50, 60 30, 60 75 0.005, 0.01, 0.02, 0.05 0.02, 0.05 5, 10, 15, 20, 25, 30, 40, 50, 60, 75 30, 75 100 0.005, 0.01, 0.02, 0.05,
0.1
120 0.01, 0.02, 0.05, 0.1 0.02, 0.05, 0.1 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120 60, 120 150 0.01, 0.02, 0.05, 0.1 0.05, 0.1 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150 75, 150 200 0.02, 0.01, 0.02, 0.05,
0.1, 0.2
250 0.05, 0.1, 0.2 0.05, 0.1, 0.2 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
300 0.02, 0.05, 0.1, 0.2 0.05, 0.1, 0.2 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
400 0.02, 0.05, 0.1, 0.2 0.1, 0.2 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
500 0.05, 0.1, 0.2, 0.5 0.1, 0.2, 0.5 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
600 0.05, 0.1, 0.2, 0.5 0.1, 0.2, 0.5 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
750 0.05, 0.1, 0.2, 0.5 0.2, 0.5 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
1000 0.05, 0.1, 0.2, 0.5, 1 0.2, 0.5, 1 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
1200 0.1, 0.2, 0.5, 1 0.2, 0.5, 1 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
1500 0.1, 0.2, 0.5, 1 0.5, 1 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
2000 0.1, 0.2, 0.5, 1, 2 0.5, 1, 2 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
2500 0.2, 0.5, 1, 2 0.5 ,1, 2 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
3000 0.2, 0.5, 1, 2 0.5 ,1 ,2 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
5000 0.5, 1, 2, 5 1, 2, 5 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
6000 0.5, 1, 2, 5 1, 2, 5 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
7500 0.5, 1, 2, 5 2, 5 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
10000 0.5, 1, 2, 5, 10 2, 5, 10 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
12000 1, 2, 5, 10, 20 2, 5, 10 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
15000 1, 2, 5, 10 5, 10 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
20000 1, 2, 5, 10, 20 5, 10, 20 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200,
Graduation size with LFT ON Span calibration points Linearity
calibration points
0.001, 0.002, 0.005 5 2, 5
0.002, 0.005, 0.01 5, 10 5, 10
0.005, 0.01, 0.02 5, 10, 15, 20 10, 20
0.02, 0.05, 0.1 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100 50, 100
0.05, 0.1, 0.2 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120, 150, 200 100, 200
120, 250
250
150, 300
250, 300
200, 400
250, 300, 400
250, 500
250, 300, 400, 500
300, 600
250, 300, 400, 500, 600
300, 750
250, 300, 400, 500, 600, 750
500, 1000
250, 300, 400, 500, 600, 750, 1000
600, 1200
250, 300, 400, 500, 600, 750, 1000, 1200
750, 1500
250, 300, 400, 500, 600, 750, 1000, 1200, 1500
1000, 2000
250, 300, 400, 500, 600, 750, 1000, 1200, 1500, 2000
1200, 2500 250, 300, 400, 500, 600, 750, 1000, 1200, 1500, 2000, 2500
1500, 3000 250, 300, 400, 500, 600, 750, 1000, 1200, 1500, 2000, 2500, 3000
2500,5000 250, 300, 400, 500, 600, 750, 1000, 1200, 1500, 2000, 2500, 3000, 5000
2500,5000 250, 300, 400, 500, 600, 750, 1000, 1200, 1500, 2000, 2500, 3000, 5000, 6000
3000,7500 250, 300, 400, 500, 600, 750, 1000, 1200, 1500, 2000, 2500, 3000, 5000, 6000, 7500
5000,10000 250, 300, 400, 500, 600, 750, 1000, 1200, 1500, 2000, 2500, 3000, 5000, 6000, 7500, 10000
6000,12000 250, 300, 400, 500, 600, 750, 1000, 1200, 1500, 2000, 2500, 3000, 5000, 6000, 7500, 10000, 12000
7500,15000 250, 300, 400, 500, 600, 750, 1000, 1200, 1500, 2000, 2500, 3000, 5000, 6000, 7500, 10000, 12000, 15000
10000,20000 250, 300, 400, 500, 600, 750, 1000, 1200, 1500, 2000, 2500, 3000, 5000, 6000, 7500, 10000, 20000
EN-20 3000 Series Indicators
3.4.5 Graduation
Set the scale readability.
0.001, 0.002, 0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20.
NOTE: Not all settings are available for each capacity. Refer to the Setup Table 3.3 for available settings.
3.4.6 Power On Unit
Set the unit that will be active at power on. oz, lb, g, kg, lb:oz or Auto (last unit in use when power was turned off.)
3.4.7 Zero Range
Set the percentage of scale capacity that may be zeroed. 0% = zeroing disabled
2% = zero up to 2 percent of capacity 100% = zero up to full capacity
3.4.8 End Setup
Advance to the next menu.
3.5 Readout Menu
Enter this menu to customize display functionality.
Reset: No, Yes Filter Level Lo, Med, Hi Auto Zero Tracking Off, 0.5d, 1d, 3d Backlight Off, On, Auto Auto Shut Off Off End Readout Exit READOUT menu
3.5.1 Reset
Set the Readout menu to factory default settings. No = not reset Yes = reset If the Legal for Trade menu item is set to ON, the Stable Range, Averaging Level, Auto Zero Tracking and Auto Off settings are not reset.
3.5.2 Filter
Set the amount of signal filtering. LO = less stability, faster stabilization time (<1 sec.) MEd = normal stability, stabilization time (<2 sec.) HI = greater stability, slower stabilization time (<3 sec.)
3.5.3 Auto-Zero Tracking
Set the automatic zero tracking functionality. OFF = disabled.
0.5 d = the display will maintain zero until a drift of 0.5 divisions per second has been exceeded. 1 d = the display will maintain zero until a drift of 1 division per second has been exceeded. 3 d = the display will maintain zero until a drift of 3 divisions per second has been exceeded.
NOTE: When the LFT menu item is set to ON, the selections are limited to 0.5d and 3d. The setting is locked when the hardware lock switch is set to the ON position.
EN-213000 Series Indicators
3.5.4 Backlight
Set the display backlight functionality. OFF = always off. ON = always on. AUtO = turns on when a button is pressed or the displayed weight changes. turns off after 5 seconds of no activity.
3.5.5 Auto Off Timer
Set the automatic shut off functionality. OFF = disabled SEt 1 = powers off after 1 minute of no activity. SEt 2 = powers off after 2 minutes of no activity. SEt 5 = powers off after 5 minutes of no activity.
3.5.6 End Readout
Advance to the next menu.
EN-22 3000 Series Indicators
3.6 Mode Menu
Enter this menu to activate the desired application modes.
3.6.1 Reset
Set the Mode menu to the factory defaults. No = not reset. Yes = reset.
NOTE: If the Legal for trade menu item is set ON, the settings are not reset.
3.6.2 Parts Counting Mode
Set the status. OFF = Disabled ON = Enabled
Reset: No, Yes Count: Off, On End Mode Exit MODE menu
3.6.3 End Mode
Advance to the next menu.
3.7 Unit Menu
kg
Enter this menu to activate the desired units. Default settings are bold.
3.7.1 Reset
Set the Unit menu to the factory defaults. Settings: NO = not reset. YES = reset If the Legal for Trade menu item is set ON, the settings are not reset.
EN-233000 Series Indicators
Reset: No, Yes Kilograms: Off, On Pounds: Off, On Grams: Off, On Ounces: Off, On Pounds:Ounces Off, On End Unit Exit UNIT menu
3.7.2 Kilogram Unit
Set the status. OFF = Disabled ON = Enabled
3.7.3 Pound Unit
Set the status. OFF = Disabled ON = Enabled
3.7.4 Gram Unit
Set the status. OFF = Disabled ON = Enabled
3.7.5 Ounce Unit
Set the status. OFF = Disabled ON = Enabled
3.7.6 Pound Ounce Unit
Set the status. OFF = Disabled ON = Enabled
EN-24
3.7.7 End Unit
Advance to the next menu.
3.8 Print Menu
3000 Series Indicators
Enter this menu to define printing parameters. Default settings are bold.
3.8.1 Reset
Set the Print menu to factory defaults. NO = not reset.
YES = reset.
NOTE: If the Legal for Trade menu item is set to ON, the following settings are not reset: Stable, Auto Print
Reset No, Yes Baud Rate: 300, 600, 1200, 2400, 4800, 9600, 19200 Parity: 7 Even, 7 Odd, 7 None, 8 None
Stop Bit 1 or 2 Handshake: Off, XON/XOFF Stable Only Off, On Auto Print Off, On Stable (-> Load, Load and Zero), Interval (-> 1…3600), Continuous
Content Gross (->Off, On) Net (->Off, On) Tare (->Off, On) Unit (->Off, On) End Print
Exit PRINT menu
3.8.2 Baud
Set the Baud rate. 300 = 300 bps 600 = 600 bps 1200 =1200 bps 2400 = 2400 bps 4800 = 4800 bps 9600 = 9600 bps 19200 = 19200 bps
3.8.3 Parity
Set the data bits and parity. 7 EVEN = 7 data bits, even parity. 7 Odd = 7 data bits, odd parity. 7 NONE = 7 data bits, no parity. 8 NONE = 8 data bits, no parity.
3000 Series Indicators
3.8.4 Stop Bit
Set the number of stop bits. 1 = 1 stop bit. 2 = 2 stop bits.
3.8.5 Handshake
Set the flow control method. NONE = no handshaking. ON-OFF = XON/XOFF software handshaking.
3.8.6 Print Stable Data Only
Set the print critera. OFF = values are printed immediately. ON = values are only printed when the stability criteria are met.
EN-25
3.8.7 Auto Print
Set the automatic printing functionality. OFF = disabled. ON.StAb = printing occurs each time the stability criteria are met. INtEr = printing occurs at the defined interval. CONt = printing occurs continuosly.
When INtEr is selected, set the Print Interval. 1 to 3600 (seconds)
3.8.8 Content
Select the additional content of the printout. GROSS OFF = Gross weight is not printed. ON = Gross weight is printed. NET OFF = Net weight is not printed. ON = Net weight is printed. TARE OFF = Tare weight is not printed. ON = Tare weight is printed. UNIT OFF = Unit is not printed. ON = Unit weight is printed.
3.8.9 End Print
Advance to the next menu.
EN-26
3.9 Menu Lock Menu
3000 Series Indicators
Enter this menu. Default settings are bold.
3.9.1 Reset
Set the menu Lock menu to factory defaults. NO = not reset. YES = reset.
NOTE: Settings for LFT controlled menu items are not reset.
3.9.2 Lock Calibration
Set the status. OFF = Calibration menu is not locked. ON = Calibration menu is locked and hidden.
Reset: No, Yes Lock Calibration Menu Off, On Lock Setup Menu Off, On Lock Readout Menu Off, On Lock Mode Menu Off, On Lock Unit Menu Off, On Lock Print Menu Off, On End Lock Menu
3.9.3 Lock Setup
Set the status. OFF = Setup menu is not locked. ON = Setup menu is locked and hidden.
3.9.4 Lock Readout
Set the status. OFF = Readout menu is not locked. ON = Readout menu is locked and hidden.
3.9.5 Lock Mode
Set the status. OFF = Mode menu is not locked. ON = Mode menu is locked and hidden.
3.9.6 Lock Unit
Set the status. OFF = Unit menu is not locked. ON = Unit menu is locked and hidden.
3000 Series Indicators
EN-27
3.9.7 Lock Print
Set the status. OFF = Print menu is not locked. ON = Print menu is locked.
3.9.8 End Lock
Advance to the next menu.
3.10 Security Switch
A security switch is located on the Main PCB board. When the switch is set to the on position, user menu settings that were locked in the Menu Lock can not be changed.
Open the housing as explained in Section 2.3.1. Set the position of security switch to ON as shown in Figure 1-3.
4 OPERATION
4.1 Turning Indicator On/Off
To turn the Indicator on, press the and hold the ON/ZERO performs a display test, momentarily displays the software version, and then enters the active weighing mode.
Off
button for 2 seconds. The Indicator
To turn the Indicator off, press and hold the ON/ZERO
Off
button until OFF is displayed.
4.2 Zero Operation
Zero can be set under the following conditions:
• Automatically at Power On (initial zero).
• Semi-automatically (manually) by pressing the ON/ZERO
• Semi-automatically by sending the Zero command (Z or alternate zero command).
Press the ON/ZERO operation.
Off
button to zero the weight display. The scale must be stable to accept zero
Off
button.
4.3 Manual Tare
When weighing an item that must be held in a container, taring stores the container weight in memory. Place the empty container on the scale (example 0.5 kg) and press the TARE button. The display will show the net weight.
To clear the Tare value, empty the scale and press the TARE button. The display will show the gross weight.
EN-28
3000 Series Indicators
4.4 Changing Units of Measure
Press and hold the PRINT will be displayed (refer to Section 3.7).
Units
button until the desired measuring unit appears. Only measuring units enabled in the Unit Menu
4.5 Printing Data
Printing the displayed data to a printer or sending the data to a computer requires that the communication parameters in the Print Menu are set (refer to Section 3.8).
Press the PRINT must be Off).
Units
button to send the displayed data to the communication port (the Auto-Print Mode in Section 3.8 function
4.6 Application Modes
Only modes enabled in the mode menu will be displayed (refer to Section 3-6).
4.6.1 Weighing
Place the item to be weighed on the scale. The illustration indicates a sample of 1.5 kg, Gross weight.
Note: To return to the Weighing mode from the Parts Counting mode, press and hold the button until WEIGH is displayed.
Mode
4.6.2 Parts Counting
Use this mode to count parts of uniform weight. The Indicator determines the quantity based on the average weight of a single part. All parts must be uniform in weight for accurate measurements.
To enter the Parts Counting mode, press and hold the
Mode
button until Count is displayed.
Average Piece Weight (APW)
When the
NOTE: If no APW has been previously stored, the CLr.PW display is skipped and the display shows PUt10Pcs.
Clearing a Stored APW Press the Yes button to clear the stored APW.
Mode
button is released, CLr.PW Pcs is displayed.
3000 Series Indicators
Recalling a Stored APW Press the No button to recall the existing APW.
EN-29
Press the FUNCTION
Mode
button to temporarily display the APW value.
Establishing the Average Piece Weight (APW)
The display shows Put10 Pcs.
Establishing a New APW
Press the No button to increment the sample size. Choices are 5, 10, 20, 50, 100 and 200.
To establish the APW, place the specified quantity of samples on the scale and press the Yes button to capture the weight.
Begin Counting
Place the parts on the scale and read the count. If a container is used, be sure to tare the empty container first.
EN-30 3000 Series Indicators
5 SERIAL COMMUNICATION
The Indicator include an RS232 serial communication interface.
The setup of RS232 operating parameters are more fully explained in Section 3.8. The physical hardware connection is explained in in Section 2.2.
The interface enables display data to be sent to a computer or printer. A computer can be used to control some functions of the indicator using the commands listed in Table 5-1.
5.1 Interface Commands
Communicate to the indicator using the command characters listed in Table 5-1.
TABLE 5-1. SERIAL INTERFACE COMMAND TABLE.
Command Character
IP Immediate Print of displayed weight (stable or unstable).
P Print stable displayed weight (according to stability setting).
CP Continuous Print.
SP Print when stable.
xP Interval Print x = Print Interval (1-3600 sec)
Z Same as pressing Zero button
T Same as pressing Tare button
xT Download Tare value in grams (positive values only). Sending 0T clears tare (if allowed)
PU Print current unit: g, kg, lb, oz, lb:oz
xU Set scale to unit x: 1=g, 2=kg, 3=lb, 4=oz, 5=lb:oz
PV Version: print name, software revision and LFT ON (if LFT is set ON).
Esc R Global reset to reset all menu settings to the original factory defaults
NOTES:
• Commands sent to the Indicator must be terminated with a carriage return (CR) or carriage return-line feed (CRLF).
• Data output by the Indicator is always terminated with a carriage return-line feed (CRLF).
• The xT (preset tare) command is not available when LFT is set to ON.
Function
5.2 Output Format
The default serial output format is shown below.
Field: Polarity Space Weight Space Unit Stability Legend CR LF Length: 1 1 7 1 5 1 3 1 1
Definitions: Polarity, “-” sign if negative, blank if positive. Weight, up to 6 numbers and 1 decimal, right justified, leading zero blanking. Units, up to 5 characters. Stability, “?” character is printed if not stable, blank if stable. Legend, up to 3 characters: G = gross weight, NET = net weight, T = tare
EN-313000 Series Indicators
EN-32 3000 Series Indicators
6. LEGAL FOR TRADE
When the indicator is used in trade or a legally controlled application it must be set up, verified and sealed in accordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that all pertinent legal requirements are met.
6.1 Settings
Before verification and sealing, perform the following steps:
1. Verify that the menu settings meet the local weights and measures regulations.
2. Perform a calibration.
3. Set Legal for Trade to ON in the Setup menu.
4. Exit the menu.
5. Disconnect power from the indicator and open the housing as explained in Section 2.3.1.
6. Set the position of the security switch to ON as shown in Section 1.2, Figure 1-2, Item 8.
7. Close the housing.
8. Reconnect power and turn the indicator on.
NOTE: For installations that employ the audit trail sealing method, steps 5 to 8 are not required. However, the security switch may be set to ON to safeguard against unintentional changes to configuration and calibration settings.
NOTE: When Legal for Trade is set to ON and the security switch is set to ON, the following menu settings cannot be changed: Span Calibration, Linearity Calibration, GEO, LFT, Calibration Unit, Capacity, Graduation, Power On Unit, Zero Range, Auto Zero Tracking, Expanded Mode, Count Mode, Kilogram Unit, Pound Unit, Gram Unit, Ounce Unit, Pound Ounce Unit, Stable Only. To enable editing of these menu settings, return the security switch to the off position and set LFT menu item to off.
6.2 Verification
The local weights and measures official or authorized service agent must perform the verification procedure. Please contact your local weights and measures office for further details.
6.3 Sealing
6.3.1 Physical Seals
For jurisdictions that use the physical sealing method, the local weights and measures official or authorized service agent must apply a security seal to prevent tampering with the settings. Refer to the illustrations below for sealing methods.
EN-333000 Series Indicators
SEAL
Seal
CALIBRATION
MODE
CALIBRATION
MODE
Figure 6-1. Wire Seal Figure 6-2. Paper Seal
6.3.2 Audit Trail Seal
For jurisdictions that use the audit trail sealing method, the local weights and measures official or authorized service agent must record the current configuration and calibration event counter values at the time of sealing. These values will be compared to values found during a future inspection.
NOTE: A change to an event counter value is equivalent to breaking a physical seal.
The audit trail uses two event counters to record changes to configuration and calibration settings.
The configuration event counter (CFG) will index by 1 when exiting the menu if one or more of the following settings
• are changed Legal for Trade, Calibration Unit, Capacity, Graduation, Power On Unit, Zero Range, Auto Zero Tracking, Expanded Mode, Count Mode, Kilogram Unit, Pound Unit, Gram Unit, Ounce Unit, Pound Ounce Unit, Stable Only. Note that the counter only indexes once, even if several settings are changed. The configuration event counter values range from CFG000 to CFG999. When the value reaches CFG999, the count starts over at CFG000. The calibration event counter (CAL) will index by 1 when exiting the menu if a Span Calibration, Linearity Calibration
• or GEO setting change is made. Note that the counter only indexes once, even if several settings are changed. The calibration event counter values range from CAL000 to CAL999. When the value reaches CAL999, the count starts over at CAL000.
The event counters can be viewed by pressing and holding the MENU button. While the button is held, the display will show MENU followed by Audit.
Release the button when Audit is displayed to view the audit trail information.
EN-34 3000 Series Indicators
CALIBRATION
MODE
CALIBRATION
MODE
CALIBRATION
MODE
The audit trail information is displayed in the format CFGxxx and CALxxx.
Then the indicator returns to normal operation.
7 MAINTENANCE
CAUTION: DISCONNECT THE UNIT FROM THE POWER SUPPLY BEFORE CLEANING.
7.1 Indicator Cleaning
• Use approved cleaning solutions for the stainless-steel Indicator housing and rinse with water. Dry thoroughly.
• Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the control panel.
7.2 Troubleshooting
TABLE 7-1. TROUBLESHOOTING.
SYMPTOM PROBABLE CAUSE(s) REMEDY
Unit will not turn on. Power cord not plugged in or properly
connected.
Check power cord connections. Make sure power cord is plugged in properly into the power outlet.
EN-353000 Series Indicators
Power outlet not supplying electricity.
Battery power used up.
Other failure.
Cannot zero the Scale, or will not zero when turned on.
Unable to calibrate. Lock Calibration Menu set to On.
Cannot display weight in desired weighing unit.
Cannot change menu settings. Menu has been locked.
Battery indicator is flashing. Battery discharged. Connect indicator to power and charge
Battery fails to charge fully. Battery is defective. Have the battery replaced by an authorized
Error 7.0 Unstable weight reading when defining
Load on Scale exceeds allowable limits.
Load on Scale is not stable.
Load Cell damage.
Lock switch is “on”.
LFT menu set to On.
Incorrect value for calibration mass.
Unit not set to On. Enable unit in the Units Menu.
Lock switch set on.
reference weight.
Check power source.
Reconnect AC power to charge the battery.
Service required.
Remove load on Scale.
Wait for load to become stable.
Service required.
Set Lock Calibration Menu to Off. Refer to Section 3.9 Menu Lock.
Set the Lock switch to Off.
Set LFT menu to Off.
Use correct calibration mass.
Refer to Section 3.7 in the Unit Menu.
Set selected menu to Off in the Lock Menu. Lock Switch on the circuit board may need to be set to the Off position.
Set the Lock switch to off.
battery.
Ohaus service dealer.
Unstable Error, check platform location.
EN-36 3000 Series Indicators
TABLE 7-1. TROUBLESHOOTING (Cont.).
SYMPTOM PROBABLE CAUSE(s) REMEDY
Error 8.1 Weight reading exceeds Power On Zero limit. Remove load from scale. Recalibrate scale.
Error 8.2 Weight reading below Power On Zero limit. Add load to scale. Recalibrate scale.
Error 8.3 Weight reading exceeds Overload limit. Reduce load on scale.
Error 8.4 Weight reading below Underload limit. Add load to scale. Recalibrate scale.
Err 9.0 Internal fault Service required.
Err 9.5 Calibration data not present. Calibrate scale.
Err 53 EEPROM data incorrect. Service required.
CAL E Calibration Error. Calibration value outside
allowable limits.
LOW.rEF The average piece weight of the parts is small
(warning).
REF.WT Err The average piece weight of the parts is too
small.
Repeat calibration using correct calibration weights.
Use parts with average piece weight greater than or equal to 1 division.
Use parts with a average piece weight greater than or equal to 0.1 division.
7.3 Service Information
If the troubleshooting section does not resolve your problem, contact an authorized Ohaus Service Agent. For Service assistance in the United States, call toll-free 1-800-526-0659 between 8:00 AM and 5:00 PM Eastern Standard Time. An Ohaus Product Service Specialist will be available to assist you. Outside the USA, please visit our website www.ohaus.com to locate the Ohaus office nearest you.
8. TECHNICAL DATA
8.1 Specifications
Materials
Housing: stainless steel Keypad: polyester Display Window: polycarbonate
Ambient conditions
The technical data is valid under the following ambient conditions: Ambient temperature: -10°C to 40°C / 14°F to104°F Relative humidity: Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C. Altitude: up to 2000m Operability is assured at ambient temperatures between -10°C. and 40°C.
TABLE 8-1. SPECIFICATIONS
Capacity Range 5 to 20000 kg or lb
Maximum Displayed Resolution 1:20,000
Type Approved Resolution 1:6,000
Minimum Average Piece Weight (APW)
Weighing Units kg, lb, g, oz, lb:oz
Functions Weighing, Parts Counting
Display 1 in./2.5 cm digit height, 6-digit, 7-segment
1.5 in./3.8 cm high x 4.9 in./12.5 cm wide backlit LCD
Backlight White LED
Keypad 4-button mechanical switches
Ingress Protection IP65 for PCBA Chamber
Load Cell Excitation Voltage 5V DC
Load Cell Drive Up to 4 x 350 ohm Load Cells
Load Cell Input Sensitivity Up to 3 mV/V
Stabilization Time Within 2 Seconds
Auto-zero Tracking Off, 0.5, 1 or 3 Divisions
Zeroing Range 0%, 2% or 100% of Capacity
Span Calibration 5 kg or 5 lb to 100% Capacity
Interface RS232
Overall Dimensions (W x D x H) (in/mm)
Net Weight (lb/kg) 7.1 / 3.2
Shipping Weight (lb/kg) 9.7 / 4.4
Operating Temperature Range -10°C to 40°C/14°F to 104°F
Power Internal rechargeable, Sealed Lead-Acid Battery (58-hour typical life)
100-240 VAC~0.5A MAX / 50-60 Hz, Internal Power Supply
8.3 x 2.8 x 5.8 / 212 x 71 x 149
1d
EN-373000 Series Indicators
EN-38 3000 Series Indicators
8.2 Accessories
TABLE 8-2. ACCESSORIES.
DESCRIPTION PART NUMBER
Column Mount Kit, 35 cm painted steel 80251743 Column Mount Kit, 70 cm painted steel 80251744 Column Mount Kit, 35 cm stainless steel 80251745 Column Mount Kit, 70 cm stainless steel 80251746 Interface Cable/PC 9-pin 80500552 Interface Cable/PC 25-pin 80500553
8.3 Drawings and Dimensions
239 mm
212 mm
9 in
.
8.3 in
.
149mm
5.8 in
.
71 mm
2.8 in
.
201 mm
8.1 in
.
EN-393000 Series Indicators
Figure 8-1. Indicator Overall Dimensions with Mounting Bracket.
EN-40 3000 Series Indicators
8.4 Compliance
Compliance
Compliance to the following standards is indicated by the corresponding mark on the product.
Marking Standard
This product conforms to the EMC directive 2004/108/EC, the Low Voltage Directive 2006/95/EC and the Non-automatic Weighing Instruments Directive 90/384/EEC. The complete Declaration of Conformity is available from Ohaus Corporation.
AS/NZS4251.1 Emission, AS/NZS4252.1 Immunity
UL60950-1: 2003
EC Emissions Note
This device complies with EN55011/CISPR 11 Class B Group 1.
Important notice for verified weighing instruments
Weighing Instruments verified at the place of manufacture bear one of the preceding mark on the packing label and the green ‘M’ (metrology) sticker on the descriptive plate. They may be put into service immediately.
Weighing Instruments to be verified in two stages have no green ‘M’ (metrology) on the descriptive plate and bear one of the preceding identification mark on the packing label. The second stage of the initial verification must be carried out by the approved service organization of the authorized representative within the EC or by the national weight & measures (W+M) authorities.
The first stage of the initial verification has been carried out at the manufacturers work. It comprises all tests according to the adopted European standard EN 45501:1992, paragraph 8.2.2.
If national regulations limit the validity period of the verification, the user of the weighing instrument must strictly observe the re-verification period and inform the respective W+M authorities.
EN-413000 Series Indicators
Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection.
For disposal instructions in Europe, refer to www.ohaus.com/weee.
FCC Note
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Industry Canada Note
This Class A digital apparatus complies with Canadian ICES-003. Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada.
ISO 9001 Registration
In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the Ohaus quality management system is compliant with the ISO 9001 standard’s requirements. On May 15, 2003, Ohaus Corporation, USA, was re-registered to the ISO 9001:2000 standard.
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus.
This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for any consequential damages.
As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for further details.
Ohaus Corporation 19A Chapin Road P.O. Box 2033 Pine Brook, NJ 07058, USA Tel: (973) 377-9000 Fax: (973) 593-0359 www.ohaus.com
*80252869*
P/N 80252869 © 2009 Ohaus Corporation, all rights reserved.
Printed in China
Loading...