Ohaus Corporation, 29 Hanover Road, Florham Park, New Jersey, 07932, USA
Declaration of Conformity
marked with “CE” - are in conformity with the directives and standards mentioned.
Konformitätserkärung
gekennzeichnet mit “CE” - mit den genannten Richtlinien und Norman übereinstimmen.
Déclaration de conformité
dessous cité - munis de la mention «CE» - sont conformes aux directives et aux normes mentionnées ci-après.
Declaración de Conformidad
balanzas indicados a continuación - con el distintivo ,CE’ - están conformes con las directivas y normas citadas.
Dichiarazione di conformità
specificati di seguito - contrassegnati con la marcatura “CE” - sono conformi alle direttive e norme citate.
Balance Type/Waagentyp/Type de balance/Modelo de balanza/Tipo di biliancia Explorer and Voyager
Marked with:DirectiveStandard
gekennzeichnet mit:RichtlinieNorm
munis de la mention:DirectiveNorme
con el distintivo:DirectivaNorma
contrassegnati con laDirettivaNorma
marcatura:
Year of attachment ofElectromagnetic compatibilityEN45501:1992, EN50082-1:1992 Immunity
the CE markEU 89/336, 92/31, 93/68EN55022:1987 Funkstörungen
Jahr der erstenelektromagnetischeEN45501:1992, EN50082-1:1992 Immunität
Eichung EU 89/336, 92/31, 93/68EN55022:1987 Emissions parasites
Année de la premèreCompatibilité électromagnétiqueEN45501:1992, EN50082-1:1992 Immunité
vérificationEU 89/336, 92/31, 93/68EN55022:1987 Radiointerferencias
Año de la primeraCompatibilidad electromagnéticaEN45501:1992, EN50082-1:1992 Inmunidad
verificaciónEU 89/336, 92/31, 93/68EN55022:Verträglichkeit 1987 Radiointerferenze
annodella primaCompatibilità elettromagneticaEN45501:1992, EN50082-1:1992 Immunità
verifica
We, Ohaus Corporation, declare under our sole responsibility that the balance models listed below
Wir, Ohaus Corporation, erklären in alleiniger Verantwortung, dass die untenstehenden Waagentypen
Nous, Ohaus Corporation, déclarons sous notre seule responsabilité, que les types de balance ci-
Nostras, Ohaus Corporation, declaramos bajo responsabilidad exclusiva que los modelos de
Noi, Ohaus Corporation, U.S.A, dichiariamo sotto nostra unica responsabilità, che i tipi di bilance
EU 73/23 Low VoltageIEC1010-1 & EN60950:1992 Safety Regulations
EU 73/23 NiederspannungIEC1010-1 & EN60950:1992 Sicherheitsbestimmungen
EU 73/23 Basse tensionIEC1010-1 & EN60950:1992 Consignes de sécurité
EU 73/23 Baja tensiónIEC1010-1 & EN60950:1992 Disposiciones sobre seguridad
EU 73/23 Bassa tensioneIEC1010-1 & EN60950:1992 Prescrizioni . di sicurezza
EU 89/336, 92/31, 93/68EN55022:1987 Emissions
96
M
122
ISO 9001 Certificate for Ohaus Corporation.
Quality International, BVQI, and was awarded the ISO 9001 certificate. This certifies that Ohaus Corporation, USA, has a quality system
that conforms with the international standards for quality management and quality assurance (ISO 9000 series). Repeat audits are carried
out by BVQI at intervals to check that the quality system is operated in the proper manner.
ISO 9001-Zertifikat für Ohaus Corporation.
International BVQI geprüft, und erhielt das ISO 9001 Zertifikat. Dieses bescheinigt, dass Ohaus Corporation, USA über ein Qualitätssystem
verfügt, welches den internationalen Normen für Qualitätsmanagement und Qualitátssicherung (ISO 9000er-Reihe) entspricht. Anlässlich
von Wiederhol-Audits durch die BVQI wird periodisch überprüft, ob das Qualitätssystem zweckmässig gehandhabt.
Certificat ISO 9000 pour Ohaus Corporation.
Quality International BVQI et a obtenu le certificat, degré ISO 9001. Celui-ci atteste que Ohaus Corporation, USA, dispose d’un système
qualité correspondant aux normes internationales pour la gestion de la qualité et pour I’assurance qualité (degré ISO 9000). Des audits
réguliers effectués par la BVQI vérifient si le système qualité est appliqué de facon appropriée.
Certificado ISO 9001 para Ohaus Corporation.
Quality International (BVQI) y ha obtenido el certificado ISO 9001. Esto acredita que Ohaus Corporation, USA, dispone de un sistema de
calidad que cumple las normas internacionales para gestión y garantfa de calidad (ISO serie 9000). Con ocasión de las inspecciones de
repetibilidad por parte de la BVQI, se comprueba periódicamente si el sistema de calidad se manipula de forma correcta.
Certificato ISO 9001 per la Ohaus Corporation.
ISO 9001 sin dal 1994 dall Bureau Veritas Quality International BVQI, e così fomice la dimostrazione che il suo sistema die Garanzia Qualità
soddisfa i massimi requisite. ll sistema della garanzia della qualità Ohaus Corporation viene verificato periodicamente dall BVQI, dando cosi
evidenza di.
T2914
EU 90/384 NAWIEN45501:1992 Non Automatic Weighing Instruments
EU 90/384 FNSWEN45501:1992 für nicht selbsttätige Waagen
EU 90/384 BFNAEN45501:1992 balances à fonctionnement non automatique
EU 90/384 PBNAEN45501:1992 para balanzas no automátäcas
EU 90.384 BFNAEN45501:1992 per bilance a funzionamento non automatics
Ohaus Corporation, USA, was examined and evaluated in 1994 b y the Bureau V eritas
Die Firma Ohaus Corporation, USA, wurde 1994 durch die Bureau Veritas Quality
La société Ohaus Corporation, USA, a été contrôlée en 1994 par Bureau Veritas
La firma Ohaus Corporation, USA, ha sido inspeccionada por la Bureau Veritas
ll sistema di garanzia della qualità della Società Ohaus Corporation, USA è certificato
James Ohaus
President
John Enterline
Marketing Manager
Notice
Certified scales,scales used for legal applications have the general type designation E...5 / V...5 and EU type Approval (T2914). The year
of the initial verification is shown next to the CE mark. Such scales are verified in the factory and carry the "M" mark on the actual scale and
the packaging. The year of the initial verification is shown next to the CE mark. If the letter M is shown against a solid bac kground, the scale
may be put into operation immediately. Should the background be partitioned and hatched, the scale must be verified at its place of use by
the certified Ohaus service. If national regulations limit the duration of the validity of the v erification certificate in individual countries, the end
user of such a scale is personally responsible for arranging the repeat verification in good time.
Hinweise
Geeichte/eichpflichtige Waagen tragen die allgemeine Typenbezeichnung E... 5 / V...5. Für sie liegt eine EU Bauartzulassung vor (T2914).
Das Jahr der ersten ist neben dem CE Zeichen aufgeführt. Solche W aagen sind ab Werk geeich und tragen die K ennz eichnung "M" au f dem
Gerät selbst und auf der Verpackung. Erscheint der Buchstab M auf v ollem Grund, darf die Waage sofort in Betrieb genommen werden. Ist
der Grund geteilt und schraffiert, muss die Waage am Verwendungsort durch den zertifizierten Ohaus Service ortsgeeicht werden. Sofern
gemäss den nationalen V orschriften in den einz elnen Staaten die Güitigkeitsdauer der Eichung beschränkt ist, ist der Betreiber einer soichen
Waage Für die rechtzeitige Nacheichung selbst verantwortlich.
Remarques
Les balances vérifiées/admissibles à la vérification portent la désignation de modèle générale E...5 / V ... 5. Elles font l’objet d’une
approbation de modèle UE (T2914). L’année de la vérification primitive est indiquée à côté de la marque CE. Ces balances sont vérifidées
d’origine et portent la marque "M" sur I’appareil lui-même et sur l’emballage, Si la lettre M apparaît sur un fond totalement vert, la balance
peut être mise en service immédiatement. Si le fond est divisé et hachuré, la balance doit être vérifiée sur le lieu d’ustilisation par le service
après-vente Ohaus certifié. Dans les pays où la durée de validité de la vérification est limitée par des prescriptions nationales, l’utilisateur
est lui-même responsable de la vérification ultérieure d’une telle balance en temps voulu.
Notas
Las balanzas verificadas/verificables llevan la designatión general E...5 / V ...5 y cuentan con una aprobación de modelo UE (T2914). EL
año de la primera verficación está indicado al lado del distintivo CE. Estas balanzas están verificadas en fábrica y Ilevan la designatión "M"
sobre el propio aparato y sobre el embalaje. Cuando la letra M aparece sobre fondo sólido, la balanza se puede poner inmediatamente en
funcionamiento. Si el f ondo está dividido y ra yado , la balanza ha de ser verificada en el lugar de uso por el se vicio técnico Ohaus certificado.
Si la duración de la validez de la verificación está limitada de acuerdo con las normas de los distintos países, el propio usuario de tal balanza
es responsable de la verificación posterior a su debido tiempo.
Avvertenza
Le bilance approvate hanno la denominazione del modello E... 5 / V ...5. Per esse esiste un’appprovazione CE del tipo. L’anno delia prima
verifica è indicato a fianco della marcatura CE. I tipi marcati con un contrassegno "M" su sfondo verde pieno possono essere impiegati da
subito. I tipi marcati con ii contrassegno "M" su sfondo nero/barrato diagonalmente dovranno essere verificati sul luogo d’installazione da
parte d’un tecnico autorizzato dal Servizio Assistenza Ohaus o ispettore dell’Ufficio Metrico. Queste bilance sono state verificate in f abbrica
e recano il contrassegno "M" sull’apparecchio stesso, e sull’imballo . É obbligo dell’untente denunciare la detenzione dello strumento all’ufficio
metrico competente per territorio e sottoporio alia prescritta verifica periodica come da disposizioni ministeriali.
NOTE: THIS EQUIPMENT HAS BEEN TESTED AND FOUND TO COMPLY WITH THE LIMITS FOR A CLASS A
DIGITAL DEVICE, PURSU ANT TO PART 15 OF THE FCC RULES .
THESE LIMITS ARE DESIGNED T O PRO VIDE REASONABLE PRO TECTION AGAINST HARMFUL INTERFERENCE
WHEN THE EQUIPMENT IS OPERATED IN A COMMERCIAL ENVIRONMENT. THIS EQUIPMENT GENERATES,
USES, AND CAN RADIATE RADIO FREQUENCY ENERGY AND, IF NO T INSTALLED AND USED IN A CCORD ANCE
WITH THE INSTRUCTION MANUAL, MAY CAUSE HARMFUL INTERFERENCE TO RADIO COMMUNICATIONS.
OPERATION OF THIS EQUIPMENT IN A RESIDENTIAL AREA IS LIKELY TO CA USE HARMFUL INTERFERENCE IN
WHICH CASE THE USER WILL BE REQUIRED TO CORRECT THE INTERFERENCE AT HIS OWN EXPENSE.
THIS DIGITAL APPARATUS DOES NOT EXCEED THE CLASS A LIMITS FOR RADIO NOISE EMISSIONS FROM
DIGITAL APPARA TUS AS SET OUT IN THE INTERFERENCE-CA USING EQUIPMENT STANDARD ENTITLED “DIGITAL APPARA TUS”, ICES-003 OF THE DEPARTMENT OF COMMUNICATIONS.
CET APPAREIL NUMERIQUE RESPECTE LES LIMITES DE BRUITS RADIOELECTRIQUES APPLICABLES AUX
APPAREILS NUMERIQUES DE CLASSE A PRESCRITES DANS LA NORME SUR LE MATERIEL BROUILLEUR :
“APPAREILS NUMERIQUES”, NMB-003 EDICTEE PAR LE MINISTRE DES COMMUNICATIONS .
Unauthorized changes or modifications to this equipment are not permitted.
TABLE OF CONTENTSTABLE OF CONTENTS
TABLE OF CONTENTS
TABLE OF CONTENTSTABLE OF CONTENTS
OVERVIEW OF CONTROLS .......................................................................................................................1
1.GETTING TO KNOW YOUR BALANCE ...................................................................................................... 2
2.1Unpacking and Checking the Standard Equipment ...................................................................................... 2
2.2Selecting the Location .................................................................................................................................. 3
2.3Setting Up and Leveling the Balance............................................................................................................ 3
2.4Installing Cover Plate and Pan ..................................................................................................................... 4
2.6Connecting Power ........................................................................................................................................4
3.OPERATING YOUR BALANCE ................................................................................................................... 5
3.1Main Menu ...................................................................................................................................................5
3.2Turning on the Balance ................................................................................................................................ 6
3.3.4 User Calibration ...................................................................................................................................... 10
3.3.5 Calibration Test ...................................................................................................................................... 11
3.4Setting Measuring Unit ............................................................................................................................... 12
3.14 Density ..................................................................................................................................................... 26
3.14.1 Balance Preparation with Density Kit .................................................................................................... 26
i
TABLE OF CONTENTS (Cont.)TABLE OF CONTENTS (Cont.)
TABLE OF CONTENTS (Cont.)
TABLE OF CONTENTS (Cont.)TABLE OF CONTENTS (Cont.)
3.14.2 Solid Density Determinations................................................................................................................... 27
3.14.3 Improving the accuracy of the Result of Solid Density Determinations .....................................................30
3.14.4 Liquid Density Determinations .................................................................................................................. 30
3.16 Printing Data .............................................................................................................................................. 32
4.SETTING UP YOUR BALANCE................................................................................................................. 34
4.7Print current Settings..................................................................................................................................42
4.11 Menu Lock-Out Protection.......................................................................................................................... 46
4.12 Legal for Trade (LFT) ................................................................................................................................. 47
5.CARE AND MAINTENANCE......................................................................................................... .............48
5.3Error Codes List ......................................................................................................................................... 52
5.5Service Information ....................................................................................................................................53
1LCD display on off button.
2Go Back buttonPermits going back in men us.
3Help buttonProvides additional information on subject displayed in the current menu.
11
4Print buttonWhen pressed, prints data either on an external printer or computer.
5
5
>O/T< buttonWhen pressed, sets balance to zero.
6b uttonWhen pressed, travels up through menu options and selects alpha n umeric characters.
7 button
When pressed, travels down through menu options and selects alphanumeric characters.
8 buttonWhen pressed, travels to the left through displays.
9 buttonWhen pressed, travels to the right through displa ys .
10Enter buttonWhen pressed, accepts item on display.
11Leveling feetUsed to level the balance .
12Leveling indicatorIndicates lev eling position of the balance.
1
1. GETTING T O KNO W Y OUR BALANCE
Please read through this section carefully, as it contains important information for safe and economical operation of
your Vo yager Balance .
1.1 Introduction
Thank you for deciding to purchase a Voyager Balance from Ohaus. Thanks to a new modular design, your Voyager
Balance lets you adapt the balance to your changing needs. Remote displays, upgraded displays which can be
table, wall or to w er mounted are a v ailab le as accessories. It offers a high le vel of operating convenience and useful
functions to make accurate measurements. A new , large, graphic LCD panel with a 240 x 128 pixel resolution is a
Back lighted, Cold Cathode Fluorescent (CCFL) type. Pop up displays makes operation of the balance extremely
simple. The use of the up/down, left/right arrow panel buttons enable selections from the men us. The Enter button
on the front panel when pressed permits any highlighted menu item to be enabled. A Go Back button permits going
back up to three lev els in any menu. A Help b utton provides assistance when required. Panel controls are clearly
marked as to their function with large Tare buttons on either side of the front panel. Oper ation and setup of the
balance is straightforward and easy. The Voyager Balance is a v ailab le in a variety of full scale capacities ranging
from 62 grams to 8,100 grams with dual range models av ailab le. Legal for Trade versions are a v ailable.
Behind your instrument stands OHAUS , a leading manuf acturer of scales , balances and analytical measuring
instruments. Our Aftermarket Department is staffed with trained instrument technicians and is dedicated to provide
you, the customer , with the fastest service possible in the event your instrument requires servicing. OHAUS also has
a Customer Service Department to answer any inquiries regarding applications and accessories.
To ensure you mak e full use of the possibilities offered by your Vo y ager balance , we advise you to read through
these operating instructions very carefully .
2. INST ALLATION
2.1 Unpacking and Checking the Standard Equipment
Open the package and remove the instrument and the accessories. Check the completeness of the delivery. The
following accessories are part of the standard equipment of your new Voy ager balance.
• Check the instrument for transport damage. Immediately inform your Ohaus dealer if y ou ha v e complaints or
parts are missing.
✓✓
✓
✓✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓✓
✓
✓✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓✓
✓
✓✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
✓ ✓
✓
✓ ✓
• Store all parts of the packaging. This packaging guarantees the best possible protection f or the tr anport of your
instrument.
2
2.2 Selecting the Location
The balance should always be used in an en vironment which is free from e xcessive air currents, corrosives, vibration,
and temperature or humidity extremes. These f actors will affect displayed weight readings .
DO NOT install the balance:
• Next to open windows or doors causing drafts or
rapid temperature changes.
Go Back
Heip
0%
Print
WEIGHT
>O/T<>O/T<
CONTRAST
CONTRAST
100%
0.00
MOVE HIGHLIGHT
GRAMS
STABLE
MAIN MENU
Enter
OHAUS
>O/T<>O/T<
®
• Near air conditioning or heat vents.
• Near vibrating, rotating or reciprocating equipment.
• Near magnetic fields or equipment that generates
magnetic fields.
Go Back
Heip
0%
Print
W
EIG
HT
>O/T<>O/T<
CO
CO
NTRAST
NTRAST
100%
0.00
M
O
VE HIG
G
RAM
S STABLE
HLIG
H
T
M
AIN M
ENU
Enter
OHAUS
>O/T<>O/T<
Go Back
Heip
0%
Print
W
EIGHT
>O/T<
CONTRAST
CONTRAST
100%
0.00
MOVE HIGHLIGHT
GRAMS
STABLE
MAIN M
ENU
Enter
OHAUS
>O/T<
®
®
• Allow sufficient space around the instrument for
ease of operation and keep awa y from radiating heat
sources.
2.3 Setting Up and Leveling the Balance
Exact horizontal positioning and stable installation are prerequisites for repeatab le results . To compensate small
irregularites or inclinations at the location, the instrument can be leveled.
For exact horizontal positioning, the balance is equipped
with a level indicator located at the front on the control
panel and two leveling f eet located at the rear of the
balance.
Position the balance in the intended operating location.
Adjust the leveling f eet at the rear of the balance until the
• On an unlevel work surface.
Leveling
Indicator
Go Back
Help
0%
Print
>O/T<
WEIGHT
CONTRAST
CONTRAST
0.000
100%
MOVE HIGHLIGHT
GRAMS
STABLE
MAIN MENU
Enter
OHAUS
>O/T<
®
Leveling
Foot
air bubble in the indicator is centered.
NOTE: The instrument should be leveled each time its
location is changed.
3
2.4 Installing Cover Plate and Pan
Go Back
Help
0%
Print
WEIGHT
CONTRAST
>O/T<
0.000
100%
MOVE HIGHLIGHT
GRAMS
STABLE
MAIN MENU
Enter
OHAUS
®
>O/T<
Pan
Cover Plate
2.5 Installing Wind Shield
Wind Shield
Go Back
Heip
0%
Print
>O/T<
CONTRAST
CONTRAST
MOVE HIGHLIGHT
GRAMS
WEIGHT
0.00
MAIN MENU
OHAUS
®
100%
STABLE
Enter
>O/T<
Balances in the range of 62g to 410g are shipped with
the pan and the protective cov er plate not installed. On
balances equipped with a draft shield, slide open the side
door and place the cover plate into position. Then, insert
the pan into the center hole which is the measuring
transducer. Higher capacity balances with 6" or 8" pans
do not have a cover plate.
On 610g to 4100g balances with 0.01g resolution, a wind
shield is required to reduce the possibility of air currents
from disturbing the pan. When the wind shield is in
place, air currents are deflected up over the pan.
Make sure the wind shield is firmly snapped into place.
2.6 Connecting Power
AC Adapter Connection
Rear of Balance
Connect the AC Adapter supplied to the three pin connector located at the rear of the balance.
The balance is now ready for operation.
4
MAIN MENU
CHANGE UNITS
SET BALANCE
CALIBRATION
CUSTOM MENU
CHANGE MODE
LIBRARY
LIBRARY 0.00% USED
EXIT TO WEIGH
EXIT TO WEIGH
MAIN MENU
;LIBRARY
CALIBRATION
CUSTOM MENU
CHANGE MODE
CHANGE UNITS
SET BALANCE
SETUP BALANCE
READOUT
INTERFACE
PRINT OPTION
SETUP GLP
SET TIME/DATE
AUTOCAL ENABLE
PRINT CURRENT SETTINGS
RESET
LOCK OUT
SOFTWARE VERSION
3. OPERA TING YOUR BALANCE
MAIN MENU
;LIBRARY
CALIBRATION
CUSTOM MENU
CHANGE MODE
CHANGE UNITS
SET BALANCE
SETUP BALANCE
INTERFACE
PRINT OPTION
SETUP GLP
SET TIME/DATE
AUTOCAL ENABLE
PRINT CURRENT SETTINGS
RESET
LOCK OUT
SOFTWARE VERSION
EXIT TO WEIGH
3.1 Main Menu
The Vo y ager balance has one main menu from which all selections are made . Shown below is the normal weighing
display screen and the main menu screen.
WEIGHT
Go Back
Help
PrintEnter
0%
CONTRAST
CONTRAST
0.00
GRAMSSTABLE
MOVE HIGHLIGHT
100%
MAIN MENU
Weighing Display Screen
Go Back
Help
PrintEnter
CHANGE MODE
CUSTOM MENU
MAIN MENU
CHANGE UNITS
CALIBRATION
SET BALANCE
LIBRARY
Main Menu Screen
MENU SELECTION
The menus shown below are selected from the main menu screen by using the arrow buttons and pressing ENTER.
The screens shown are all of the primary selections that can be made. By using the arrow buttons, further selections
can be made in each menu to set parameters.
The Vo y ager balance is ready to oper ate after the the installation procedures are performed. When the balance is
first turned on and it completes its checks, and is calibrated, it can be used to weigh or tare materials without setting
the menus. The balance is preset at the factory to weigh in grams
It is recommended that you read this manual carefully and set the balance to operate for y our specific applications
using the procedures in Chapter 4 Setting up Your Balance and calibrate the balance before using.
LCD Display On/Off
To turn the balance LCD display ON, press the ON/OFF button (circled b utton with an I inside) located at the upper
left-hand corner of the panel once. To turn OFF, press button again.
Stabilization
Before initially using the balance, allow time for it to adjust to its new environment. The balance only requires to be
plugged in to warm up. Recommended warm up period is twenty (20) minutes. The internal circuits of the balance
are powered whenev er it is plugged into a po w er source .
3.3 Calibration
V oyager balances offer a choice of fiv e calibration methods: Automatic Calibration (AutoCalTM), Span Calibration, User
Calibration, Linearity Calibration, and Calibration Test.
•
Span
-Span calibration ensures that the balance reads correctly within specifications using
two weight values: zero and a weight value at 100% of the balance’s full capacity.
•Linearity
•User
•Calibration Test
•
AutoCal
-Linearity calibration minimizes deviation between actual and displayed weights within
the balance’s w eighing r ange. Three weight values are used: zero, a weight v alue at
midpoint of the balances weighing range, and a weight value at or near the balance’s
specified capacity .
-User calibration is a method where the balance can be calibrated using a mass of
known value by entering that value into the balance.
-Calibration test allows the stored calibration data to be tested against the current
mass being used for the test.
TM
Automatic calibration (AutoCalTM) of the balance is accomplished by an internal
mass.
Calibration Menu Protection
NOTES:
• Calibration may be lock ed out to prevent unauthorized
personnel from changing calibration. If calibration has
been locked out, you can only access Internal W eight
Calibration and Calibration Test.
• To lock out calibration menu, after calibration, refer to
the section titled Menu Lock-Out Protection.
• Linearity , Span and User calibr ation are disab led f or
Type Approv ed/LFT balances.
Calibration Masses
Before beginning calibration, make sure masses are
availab le. If you begin calibration and realize calibration
masses are not availab le, exit the menu. The balance
will retain previously stored calibration data. Calibration
should be performed as necessary to ensure accurate
weighing. Masses required to perform the procedures
are listed in the adjacent table.
Masses must meet or exceed ASTM Class 1 Tolerance.
Calibration masses are available as accessories.
6
3.3.1 Auto Calibration (AutoCalTM)
On Vo yager balances equipped with the A utoCal
tion mass. Auto calibration can be performed at any time providing the balance has warmed up to operating temperature.
TM
feature, calibration can be accomplished using an internal calibra-
PROCEDURE
WEIGHT
Go Back
Help
PrintEnter
0%
CONTRAST
0.00
GRAMS STABLE
MOVE HIGHLIGHT
100%
MAIN MENU
• Press arrow b utton to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CALIBRATION.
• Press Enter button.
• Press or arrow button and select AUTOCAL.
CHANGE MODE
CUSTOM MENU
CALIBRATE
SPAN
LINPLEASEERARITY
USER
CHANGE MODE
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
INTERNAL CALIBRATION
PLEASE CLEAR THE PAN
AND PRESS < ENTER > . . .
CHANGE MODE
CUSTOM MENU
MAIN MENU
CHANGE UNITS
SPAN
CALIBRATION
LINEARITY
USER
CHANGE MODE
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
MAIN MENU
CHANGE UNITS
SPAN
CALIBRATION
LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
INTERNAL CALIBRATION
PLEASE WAIT . . .
SET BALANCE
MAIN MENU
CALIBRATE
RETURN TO WEIGH
CHANGE UNITS
CALIBRATION
SET BALANCE
MAIN MENU
CALIBRATE
RETURN TO WEIGH
CHANGE UNITS
CALIBRATION
SET BALANCE
SET BALANCE
LIBRARY
LIBRARY
• Clear the pan.
• Press Enter button, screen displays PLEASE WAIT,
the balance is automatically calibrated and then returns to a weighing mode.
NOTE:
AutoCalTM calibration uses an internal mass in the
balance for calibration and is done automatically when
selected.
WEIGHT
Go Back
Help
PrintEnter
0%
CONTRAST
0.00
GRAMS STABLE
MOVE HIGHLIGHT
100%
MAIN MENU
7
3.3.2 Span Calibration
Span calibration normally requires that calibration be made using a mass equal to the full capacity of the balance,
howev er, the Voyager balance can be calibrated using other lesser values as specified on the display.
PROCEDURE
WEIGHT
Go Back
Help
PrintEnter
CHANGE MODE
CUSTOM MENU
SPAN
LINPLEASEERARITY
PLEASE CLEAR THE PAN
USER
CHANGE MODE
AND PRESS < ENTER > . . .
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
0%
CONTRAST
MAIN MENU
CHANGE UNITS
SPAN
LINEARITY
USER
CHANGE MODE
CALIBRATION
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
MAIN MENU
CALIBRATE
INTERNAL CALIBRATION
CHANGE UNITS
SPAN
LINPLEASEERARITY
USER
CHANGE MODE
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
INTERNAL CALIBRATION
PLEASE WAIT . . .
CALIBRATION
GRAMS STABLE
MOVE HIGHLIGHT
SET BALANCE
MAIN MENU
CALIBRATE
CHANGE UNITS
RETURN TO WEIGH
CALIBRATION
SET BALANCE
MAIN MENU
CALIBRATE
RETURN TO
CHANGE UNITS
WEIGH
SPAN
CALIBRATION
LINPLEASEERARITY
PLEASE PUT: 4000.00 G ON PAN.
USER
AND PRESS < ENTER > . . .
CHANGE MODE
CALIBRATION TEST
AUTOCAL
OTHER WEIGHTS MAY BE USED:
EXIT TO WEIGH
3000 2000 1000
CUSTOM MENU
100%
0.00
MAIN MENU
SET BALANCE
SET BALANCE
MAIN MENU
CALIBRATE
SPAN CALIBRATION
LIBRARY
CHANGE UNITS
CALIBRATION
LIBRARY
• Press arrow b utton to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CALIBRATION.
• Press Enter button.
• Press or arrow button and select SPAN.
• Press Enter button.
• Clear the pan and press Enter button.
• Place indicated mass value on pan and press Enter
button.
• Display indicates if calibration was successful and the
difference between the last calibration.
• Press any button, balance returns to weighing mode.
• Remove masses from the pan.
NOTE: The samples shown on the displa ys w ere for an
4.1kg balance.
SET BALANCE
LIBRARY
SPAN
LINPLEASEERARITY
PLEASE WAIT . . .
USER
CHANGE MODE
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
MAIN MENU
CALIBRATE
INTERNAL CALIBRATION
CHANGE UNITS
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION
CHANGE MODE
CALIBRATION TEST
AUTOCALEXIT TO WEIGH
ENTER USER CAL VALUE
SPAN CALIBRATION
0
SPAN CAL SUCCESSFULLY DONE!
UP/DOWN TO SCROLL CHARACTERS
THE LAST DIFFERENCE BETWEEN THIS CAL
LEFT/RIGHT TO MOVE CURSOR
AND LAST CAL IS:
ENTER TO ACCEPT
0.01 GRAMS
BACK TO QUIT
PRESS ANY KEY TO EXIT . . .
SET BALANCE
MAIN MENU
LIBRARY
CHANGE UNITS
CALIBRATION
SET BALANCE
LIBRARYCUSTOM MENU
WEIGHT
Go Back
Help
PrintEnter
0%
CONTRAST
0.00
GRAMS STABLE
MOVE HIGHLIGHT
100%
MAIN MENU
8
3.3.3 Linearity Calibration
Linearity calibration utilizes three calibration points, one at zero , center span and full span. This method minimizes
deviation between actual and displa y ed weights within the balance's weighing range . Three w eight v alues are used;
zero, a w eight v alue at midpoint of the balance's weighing range and a w eight v alue at or near the specified capacity.
PROCEDURE
WEIGHT
Go Back
Help
PrintEnter
CHANGE MODE
CUSTOM MENU
SPAN
LINEARITY
PLEASE CLEAR THE PAN
USER
CHANGE MODE
AND PRESS ENTER KEY . . .
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
0%
CONTRAST
MAIN MENU
CHANGE UNITS
CALIBRATION
MAIN MENU
CALIBRATE
LINEARITY CALIBRATION
CHANGE UNITS
SPAN
LINEARITY
PLEASE WAIT . . .
USER
CHANGE MODE
CALIBRATION
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
CHANGE MODE
CUSTOM MENU
MAIN MENU
CALIBRATE
LINEARITY CALIBRATION
CHANGE UNITS
SPAN
LINEARITY
CALIBRATION
USER
CHANGE MODE
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
0.00
GRAMS STABLE
MOVE HIGHLIGHT
SET BALANCE
SPAN
LINEARITY
USER
CALIBRATION TEST
RETURN TO
AUTOCAL
WEIGH
EXIT TO WEIGH
SET BALANCE
RETURN TO WEIGH
PLEASE PUT 2000.00 G ON PAN
AND PRESS < ENTER > . . .
MAIN MENU
CALIBRATE
CHANGE UNITS
CALIBRATION
SET BALANCE
MAIN MENU
CALIBRATE
LINEARITY CALIBRATION
RETURN TO
CHANGE UNITS
WEIGH
CALIBRATION
100%
MAIN MENU
SET BALANCE
SET BALANCE
LIBRARY
• Press arrow b utton to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CALIBRATION.
• Press Enter button.
• Press or arrow button and select LINEARITY.
• Press Enter button, LINEARITY CALIBRA TION is displayed.
• Clear pan and press Enter button.
LIBRARY
• Place indicated mass value on pan. This will be one
half of the total capacity of the balance, then press
Enter button.
• Place indicated mass value on pan. This will be the
total capacity of the balance, then press Enter button.
PLEASE W AIT is displa y ed f ollo wed b y the actual weight
value placed on the pan which is the maximum capacity
of the balance.
• Remove the masses from the pan. The balance is
now calibrated.
NOTE: The samples shown on the displa ys w ere for an
4.1kg balance.
SPAN
LINEARITY
PLEASE WAIT . . .
USER
CHANGE MODE
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
MAIN MENU
CALIBRATE
LINEARITY CALIBRATION
CHANGE UNITS
SPAN
LINEARITY
USER
CHANGE MODE
CALIBRATION
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
SET BALANCE
MAIN MENU
CALIBRATE
LINEARITY CALIBRATION
PLEASE PUT 4000.00 G ON PAN
CHANGE UNITS
AND PRESS < ENTER > . . .
RETURN TO WEIGH
CALIBRATION
SET BALANCE
LIBRARY
WEIGHT
Go Back
Help
PrintEnter
0%
CONTRAST
100%
4000.00
GRAMS STABLE
MOVE HIGHLIGHT
MAIN MENU
9
3.3.4 User Calibration
User calibration is used when it is desired to calibrate the balance using a mass of known value. To use this calibration feature, proceed as f ollows:
PROCEDURE
WEIGHT
Go Back
Help
PrintEnter
CHANGE MODE
CUSTOM MENU
SPAN
2000
LINEARITY
USER
CHANGE MODE
UP/DOWN TO SCROLL CHARACTERS
CALIBRATION TEST
LEFT/RIGHT TO MOVE CURSOR
AUTOCAL
ENTER TO ACCEPT
EXIT TO WEIGH
BACK TO QUIT
CUSTOM MENU
0%
CONTRAST
MAIN MENU
CHANGE UNITS
CALIBRATION
MAIN MENU
CALIBRATE
ENTER CAL VALUE
CHANGE UNITS
SPAN
0
LINEARITY
USER
CHANGE MODE
PLEASE CLEAR THE PAN
UP/DOWN TO SCROLL CHARACTERS
CALIBRATION TEST
CALIBRATION
AND PRESS ENTER KEY . . .
LEFT/RIGHT TO MOVE CURSOR
AUTOCAL
ENTER TO ACCEPT
EXIT TO WEIGH
BACK TO QUIT
CUSTOM MENU
MOVE HIGHLIGHT
SPAN
LINEARITY
USER
CHANGE MODE
CALIBRATION TEST
RETURN TO
AUTOCAL
WEIGH
EXIT TO WEIGH
CUSTOM MENU
SET BALANCE
MAIN MENU
CALIBRATE
ENTER CAL VALUE
USER CALIBRATION
CHANGE UNITS
RETURN TO WEIGH
SPAN
0
LINEARITY
USER
CHANGE MODE
CALIBRATION
UP/DOWN TO SCROLL CHARACTERS
CALIBRATION TEST
LEFT/RIGHT TO MOVE CURSOR
AUTOCAL
ENTER TO ACCEPT
EXIT TO WEIGH
BACK TO QUIT
CUSTOM MENU
100%
0.00
GRAMS STABLE
SET BALANCE
CALIBRATE
ENTER CAL VALUE
USER CALIBRATION
PLEASE PUT: 2000 G ON PAN.
AND PRESS < ENTER > . . .
MAIN MENU
MAIN MENU
CALIBRATE
CHANGE UNITS
CALIBRATION
SET BALANCE
MAIN MENU
CHANGE UNITS
RETURN TO WEIGH
CALIBRATION
SET BALANCE
LIBRARY
SET BALANCE
LIBRARY
• Press arrow button to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CALIBRATION.
• Press Enter button.
• Press or arrow button and select USER.
• Press Enter button, USER CALIBRATION is displayed.
• Using the arrow buttons, enter a v alue which is at
least 25% of the full span value, press Enter b utton.
The sample displays shown used a 2kg mass on an
4.1kg balance.
• Clear the pan and press Enter button.
• Place the selected mass value on the pan and press
Enter button.
• Display indicates if calibration was successful and the
difference between the last calibration.
• Press any button, balance returns to weighing mode.
MAIN MENU
CALIBRATE
SPAN
LINEARITY
USER
CHANGE MODE
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
ENTER CAL VALUE
USER CALIBRATION
0
PLEASE WAIT . . .
UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
ENTER TO ACCEPT
BACK TO QUIT
CHANGE UNITS
SPAN
LINEARITY
USER
CHANGE MODE
CALIBRATION
UP/DOWN TO SCROLL CHARACTERS
CALIBRATION TEST
LEFT/RIGHT TO MOVE CURSOR
AUTOCAL
ENTER TO ACCEPT
EXIT TO WEIGH
BACK TO QUIT
CUSTOM MENU
SET BALANCE
MAIN MENU
CALIBRATE
ENTER CAL VALUE
USER CALIBRATION
0
USER CAL SUCCESSFULLY DONE!
CHANGE UNITS
RETURN TO WEIGH
THE LAST DIFFERENCE BETWEEN THIS CAL
AND LAST CAL IS:
0.01 GRAMS
PRESS ANY KEY TO EXIT . . .
CALIBRATION
SET BALANCE
LIBRARY
10
WEIGHT
Go Back
Help
PrintEnter
0%
CONTRAST
0.00
GRAMS STABLE
MOVE HIGHLIGHT
100%
MAIN MENU
3.3.5 Calibration test
Calibration test feature allows a chec k of a known calibr ation mass against the last stored calibr ation inf ormation in
the balance. The sample displays shown are for a 4.1 kilogram balance.
PROCEDURE
WEIGHT
Go Back
Help
PrintEnter
CHANGE MODE
CUSTOM MENU
SPAN
LINEARITY
PLEASE CLEAR THE PAN
USER
CHANGE MODE
AND PRESS < ENTER > . . .
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
0%
CONTRAST
MAIN MENU
CHANGE UNITS
CALIBRATION
MAIN MENU
CALIBRATE
CALIBRATION TEST
CHANGE UNITS
SPAN
LINEARITY
PLEASE WAIT . . .
CALIBRATION
USER
CHANGE MODE
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
GRAMS STABLE
MOVE HIGHLIGHT
SET BALANCE
SPAN
LINEARITY
USER
CHANGE MODE
CALIBRATION TEST
RETURN TO
AUTOCAL
WEIGH
EXIT TO WEIGH
CUSTOM MENU
SET BALANCE
MAIN MENU
CALIBRATE
CALIBRATION TEST
LIBRARY
CHANGE UNITS
CALIBRATION
100%
0.00
MAIN MENU
MAIN MENU
CALIBRATE
CHANGE UNITS
CALIBRATION
SET BALANCE
LIBRARY
SET BALANCE
LIBRARY
• Press arrow button to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CALIBRATION.
• Press Enter button.
• Press or arrow button and select CALIBRATION TEST .
• Press Enter button.
• Clear the pan and press Enter button.
• Place the indicated mass value on the pan and press
Enter button.
• Display indicates if calibration was successful and the
difference between the last calibration.
• Remove masses from the pan.
• Press any button, balance returns to weighing mode.
MAIN MENU
CALIBRATE
CALIBRATION TEST
SPAN
LINEARITY
PLEASE PUT: 4000.00 G ON PAN.
USER
CHANGE MODE
AND PRESS < ENTER > . . .
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
CHANGE UNITS
CALIBRATION
SPAN
LINEARITY
PLEASE WAIT . . .
USER
CHANGE MODE
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CUSTOM MENU
SET BALANCE
MAIN MENU
CALIBRATE
CALIBRATION TEST
LIBRARY
CHANGE UNITS
ENTER USER CAL VALUE
CALIBRATION TEST
SPAN
0
SPAN CAL SUCCESSFULLY DONE!
LINEARITY
CALIBRATION
THE LAST DIFFERENCE BETWEEN THIS CAL
USER
CHANGE MODE
AND LAST CAL IS:
CALIBRATION TEST
0.01 GRAMS
AUTOCAL
PRESS ANY KEY TO EXIT . . .
EXIT TO WEIGH
SET BALANCE
MAIN MENU
CALIBRATE
LIBRARY
CHANGE UNITS
CALIBRATION
SET BALANCE
LIBRARYCUSTOM MENU
11
WEIGHT
Go Back
Help
PrintEnter
0%
CONTRAST
0.00
GRAMS STABLE
MOVE HIGHLIGHT
100%
MAIN MENU
3.4 Setting Measuring Unit
Before using the balance f or the first time , the desired measuring unit should be set first. The f ollowing measuring
units are availab le: MILLIGRAMS, GRAMS , KILOGRAMS, PENNY WEIGHTS , CARATS, OUNCES, TR OY OUNCES,
GRAINS, HONG KONG TAELS, SINGAPORE TAELS , ROC T AELS, POUNDS, NEWTONS, TICALS and CUSTOM
UNITS.
Procedure
WEIGHT
Go Back
Help
PrintEnter
0%
CONTRAST
0.00
GRAMS STABLE
MOVE HIGHLIGHT
MAIN MENU
100%
MAIN MENU
To select a measuring unit, proceed as follows:
• Press arrow button to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CHANGE UNITS.
• Press Enter button.
• Press or arrow button and select the
desired measuring unit. (Grams is shown).
• Press Enter button to save setting. The balance will
CHANGE MODE
CUSTOM MENU
CHANGE UNITS
MILLIGRAMS
GRAMS
KILOGRAMS
CHANGE MODE
PENNY WEIGHTS
CARATS
OUNCES
TROY OUNCES
GRAINS
CUSTOM MENU
HONG KONG TAELS
SINGAPORE TAELSEIGH
CHANGE UNITS
GRAINS
HONG KONG TAELS
SINGAPORE TAELS
CHANGE MODE
ROC TAELS
MOMMES
POUNDS
NEWTONS
TICALS
CUSTOM MENU
CUSTOM UNITS
EXIT TO WEIGH
CHANGE UNITS
CALIBRATION
MAIN MENU
CHANGE UNITS
CALIBRATION
MAIN MENU
CHANGE UNITS
CALIBRATION
SET BALANCE
LIBRARY
SET BALANCE
LIBRARY
SET BALANCE
LIBRARY
now weigh in the selected measuring unit.
NOTE: Depending upon the balance model/capacity,
some measuring units may not be availab le when
selected.
When CUSTOM UNITS is selected, up to three diff erent entries may be made. You would select SETUP
and follow the prompts and enter UNIT NAME, SET
MANTISSA, SELECT DP, SELECT LSD and then
SA VE & EXIT.
WEIGHT
Go Back
Help
PrintEnter
0%
CONTRAST
CONTRAST
0.00
GRAMSSTABLE
MOVE HIGHLIGHT
100%
MAIN MENU
12
>O/T<
>O/T<
1620.00
WEIGHT
0%
100%
CONTRAST
GRAMS
STABLE
MOVE HIGHLIGHT
CONTRAST
MAIN MENU
Enter
Go Back
Heip
Print
OHAUS
®
>O/T<
>O/T<
200.00
WEIGHT
0%
100%
CONTRAST
GRAMS
STABLE
MOVE HIGHLIGHT
CONTRAST
MAIN MENU
Enter
Go Back
Heip
Print
OHAUS
®
3.5 Basic Weighing
V oyager balances are shipped with grams enabled. When the balance is to be used with other Type Approved/Legal
for Trade units of measure, the desired unit must be enabled. F or w eigh below applications , refer to section on
density measurements.
WEIGHT
Go Back
Help
PrintEnter
0%
CONTRAST
0.00
GRAMS STABLE
MOVE HIGHLIGHT
100%
MAIN MENU
Procedure
• Press arrow b utton to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CHANGE MODE.
• Press Enter button.
• Press or arrow button and select BASIC
WEIGHING.
Zero/Tare
When weighing material or objects that must be held in a
container, taring stores the container weight in the
balance’s memory, separate from the weight of the
material in the container.
• Place an empty container on the pan. Its weight is
displayed. (200 gram container is shown).
• Press
the display blanks until stable w eight
O/T
readings are received, then indicates zero . The
container’s weight is stored in memory.
• Add material to the container . As material is added,
its net weight is displayed. (Example is 1620 grams).
• Removing the container and material from the platform will cause the balance to display the container’s
weight as a negative number. The tared weight will
remain in memory until
the balance is turned off.
• Pressing
O/T
resets the balance to zero.
is pressed again or
O/T
13
(
Example
(
Example
Container 200g)
Material 1620g)
3.6 Parts Counting
EASY COUNT
0%
100%
PIECESUNSTABLE
OPTIMIZE
MAIN
A.P.W. =
SIZE =
TARE WT =
WEIGHT =
CONTRAST
NEW
SET A.P.W.
PLEASE PUT:
10
SAMPLES
AND PRESS < ENTER > . . .
The Voyager balance can be set to either an Easy Count or Advanced Count parts counting method.
Easy Count setting enables a simplified method for counting parts. When selected, Easy Count displays a screen
which requires that the number of sample parts be entered. After this entry , putting a quantity of samples on the pan,
the balance will display the actual number of samples. Since the balance determines the quantity based on the
average weight of a single part, all parts must be reasonably uniform in weight.
Advanced Count setting contains a number of entry screens which include assigning a library name, filling and
sorting applications and statisical information which can be printed.
EXIT TO WEIGH
DIFF WEIGHING
FORMULATION
QUICK CHECK
CUSTOM MENU
STATISTICS
DENSITY
100%
MAIN MENU
SET BALANCE
MAIN MENU
CHANGE MODE
PARTS COUNTING
RETURN TO WEIGH
CHANGE UNITSPART
CALIBRATION
0%
A.P.W. =
SIZE =
<UP>/<DOWN> TO SCROLL CHARACTERS
<LEFT>/<RIGHT> TO MOVE CURSOR
<ENTER> TO ACCEPT
<BACK> TO ABORT
CONTRAST
10
Procedure
SET BALANCE
EASY COUNT
RETURN TO WEIGH
ENTER SAMPLE SIZE
NEW
TARE WT =WEIGHT =
PIECESUNSTABLE
OPTIMIZE
• Press arrow b utton to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CHANGE MODE.
• Press Enter button.
• Press or arrow button and select PARTS
COUNTING.
• Press Enter button.
• Press or arrow button and select EASY
COUNT.
• Press Enter button.
• Enter the numeric sample size as indicated on the
screen using the arrow buttons to set the value.
• Press Enter button.
• Place sample size on pan and press Enter button.
• Remove samples from pan and place samples to be
measured on the pan. The display indicates thenumber of pieces based on the sample size. Repeated
batches of samples may be placed on the pan and
counted. Sample shown belo w indicates 10 pieces .
AARE WT 0
TC ERR OFF
AUTO OPTIMIZATION OFF
FUNCTION LINK NONE
SAVE TO LIBRARY
EAVE AND START
LIBRARY NAME
<UP>/<DOWN> TO SCROLL CHARACTERS
<LEFT>/<RIGHT> TO MOVE CURSOR
<ENTER> TO ACCEPT
<BACK> TO ABORT
Advanced Count setting contains a number of entry screens which include assigning a library name, filling and sorting
applications and statisical information which can be printed. Ref er to screen 4 belo w which contains the following
entries:
LIBRARY NAME - A name up to 10 characters can be entered and stored to identify the item to be counted.
A.P.W. - This average piece weight, when selected, either a sample siz e or actual piece weight can be entered.
T ARE WEIGHT - This is the tare weight of the container holding the samples.
AUTO OPTIMIZATION - An ON or OFF function. When set ON, optimizes the accuracy based on sample size.
FUNCTION LINK - Various pop-up screens permit the entry of four options, NONE, FILLING, CHECK WEIGHING
and STATISTIC. When FILLING is selected, a target weight is entered which is sho wn as 100% on the bar gr aph on
the display. When material is added to the balance pan, it is display ed as a percentage and weight. When CHECK
WEIGHING is selected, a separate pop-up display has entries for nominal pieces , o v er pieces , under pieces, display
type, library name and save and exit. This type of function permits checking of individual pieces against the stored
information in the balance. When STATISTICS is selected, provides display of Standard Deviation, either population
or sample, Mean, Sum, High, Low and Difference readings. Each can be individually set ON or OFF.
SA VE TO LIBRARY - When selected, saves all settings to the library .
RUN - When selected, starts program.
Procedure
WEIGHT
Go Back
Help
PrintEnter
0%
CONTRAST
MOVE HIGHLIGHT
MAIN MENU
100%
0.00
GRAMS STABLE
MAIN MENU
• Press arrow b utton to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CHANGE MODE.
• Press Enter button.
• Press or arrow button and select PARTS
COUNTING.
• Press Enter button.
• Press or arrow button and select ADV ANCED COUNT.
• Press Enter button.
• Continue through all menus and make the required
settings.
Filling or Fill weighing permits you to enter a target weight, then view other loads as a percentage of the reference
which has been set in the balance parameters. The load you place on the platform is displa y ed as a percentage of
what was entered into the balance. A twin bar display indicates up to 89% on the first bar and up to 110% on the
second bar with a large numeric display.
<UP>/<DOWN> TO SCROLL CHARACTERS
<LEFT>/<RIGHT> TO MOVE CURSOR
<ENTER> TO ACCEPT
<BACK> TO ABORT
CHANGE UNITSPART
CALIBRATION
• Press arrow b utton to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CHANGE MODE.
• Press Enter button.
• Press or arrow button and select FILLING.
• Press Enter button.
• Using the arrow buttons, enter the target weight.
• Press Enter button.
• Place the load on the balance pan, the display indicates on the bar graph as a percentage and displays
the actual load weight numerically .
SET BALANCE
LIBRARY
16
FILLING
Go Back
Help
PrintEnter
0%
90%110%
0.00
0.00
TARGET = 0.00
CONTRAST
DIFF WT =
GRAMS STABLE
NEW STOP
89%
0.00
MAIN
3.8 Animal Weighing
Animal weighing permits you to weigh small animals directly on the balance. To compensate f or activ e subjects, a
setup menu permits you to enter a smoothing filter labeled Good, Better and Best which averages the subjects
movements and displays an accurate weight. A single bar display indicates up to 100% of the capacity of the balance. The large numeric display indicates the weight of the subject.
• Press or arrow button and select ANIMAL
WEIGHING.
• Press Enter button.
• Press or buttons to set the AW Filter, then
press Enter button. PUT ANIMAL ON PAN . . . is
displayed.
• Place the subject on the balance pan, a countdown
appears on the display which allows the balance to
accurately indicate the weight of the subject. The bar
graph indicates the percentage of weight relative to
the capacity of the balance.
100%
RESET ( ) = 5 PREVIOUS ( ) = 5
0%
CONTRAST
CONTRAST
ANIMAL WEIGHING
GRAMSSTABLE
STOP
PUT ANIMAL ON PAN . . .
FILTER
GRAMSSTABLE
FILTER
STOP
MENU
MENU
100%
17
ANIMAL WEIGHING
Go Back
Help
PrintEnter
0%
CONTRAST
GRAMSSTABLE
FILTER
STOP
0.00
100%
MENU
3.9 Check Weighing
Check weighing mode permits you to weigh an item, set balance parmeters such as the nominal weight, over weight,
under weight, type of display: normal shown with a bar graph and a numeric display or sign which spells out under ,
accept, over in large letters and assign a library name. This can then be recalled later eliminating the need to enter
weighing parameters again. This type of weighing is where individual items must be checked against preset parameters. Since man y displays are repetitive, not all will be shown in the following procedure
Procedure
WEIGHT
Go Back
Help
PrintEnter
CHANGE MODE
CUSTOM MENU
CHANGE MODE
CUSTOM MENU
CHANGE MODE
CUSTOM MENU
0%
CONTRAST
BASIC WEIGHING
PARTS COUNTING
FILLING
CHANGE MODE
ANIMAL WEIGHING
CHECK WEIGHING
DIFF WEIGHING
FORMULATION
QUICK CHECK
CUSTOM MENU
STATISTIC
DENSITY
BASIC WEIGHING
NOMINAL WT 0.00
PARTS COUNTING
OVER WT 0.00
FILingL
UNDER WT 0.00
ANIMAL WEIGHING
LIBRARY NAME
CHECK WEIGHING
SAVE & EXIT
DIFF WEIGHING
R
FOR
EXIT TO WEIGH
QUICK CHECK
STATISTIC
DENSITY
BASIC WEIGHING
NOMINAL WT 0.00
PARTS COUNTING
OVER WT 0.00
FILLING
UNDER WT 0.00
ANIMAL WEIGHING
LIBRARY NAME
CHECK WEIGHING
SAVE & EXIT
DIFF WEIGHING
R
FORMULATION
EXIT TO WEIGH
QUICK CHECK
STATISTICS
DENSITY
SET NOMINAL WEIGHT
0.00
BASIC WEIGHING
UP/DOWN TO SCROLL CHARACTERS
NOMINAL WT 0.00
PARTS COUNTING
LEFT/RIGHT TO MOVE CURSOR
OVER WT 0.00
FILLING
CHANGE MODE
ENTER TO ACCEPT
UNDER WT 0.00
ANIMAL WEIGHING
BACK TO QUIT
LIBRARY NAME
CHECK WEIGHING
<UP>/<DOWN> TO SCROLL CHARACTERS
SAVE & EXIT
DIFF WEIGHING
<LEFT>/<RIGHT> TO MOVE CURSOR
R
FORMULATION
<ENTER> TO ACCEPT
EXIT TO WEIGH
QUICK CHECK
CUSTOM MENU
<BACK> TO ABORT
STATISTIC
DENSITY
NAME
<UP>/<DOWN> TO SCROLL CHARACTERS
<LEFT>/<RIGHT> TO MOVE CURSOR
<ENTER> TO ACCEPT
<BACK> TO ABORT
MOVE HIGHLIGHT
MAIN MENU
CHANGE UNITS
MAIN MENU
CHANGE MODE
CALIBRATION
CHANGE UNITSPART
BASIC WEIGHING
CALIBRATION
NOMINAL WT 300.0 0
PARTS COUNTING
OVER WT 0.00
FILL
CHANGE MODE
UNDER WT 0.00
ANIMAL WEIGHING
LIBRARY NAME
CHECK WEIGHING
SAVE TO LIBRARY
DIFF WEIGHING
RUN
FORMULATION
EXIT TO WEIGH
QUICK CHECK
CUSTOM MENU
STATISTICS
DENSITY
MAIN MENU
CHANGE MODE
SETUP
CHANGE UNITSPART
MAIN MENU
CHANGE MODE
SETUP
SET OVER WEIGHT
CHANGE UNITSPART
0.00
CALIBRATION
BASIC WEIGHING
NOMINAL WT 0.00
PARTS COUNTING
OVER WT 0.00
FILLING
CHANGE MODE
UNDER WT 0.00
ANIMAL WEIGHING
CALIBRATION
LIBRARY NAME
CHECK WEIGHING
<UP>/<DOWN> TO SCROLL CHARACTERS
SAVE & EXIT
DIFFIAL WEIGHING
<LEFT>/<RIGHT> TO MOVE CURSOR
R
FORMULATION
<ENTER> TO ACCEPT
EXIT TO WEIGH
QUICK CHECK
CUSTOM MENU
<BACK> TO ABORT
STATISTICS
DENSITY
MAIN MENU
CHANGE MODE
SETUP
LIBRARY NAME
CHANGE UNITSPART
CALIBRATION
100%
0.00
GRAMS STABLE
CHANGE MODE
RETURN TO WEIGH
SET UNDER WEIGHT
295.00
SET BALANCE
MAIN MENU
SET BALANCE
LIBRARY
SET BALANCE
MAIN MENU
SETUP
RETURN TO WEIGH
CHANGE UNITSPART
CALIBRATION
SET BALANCE
SET BALANCE
MAIN MENU
CHANGE MODE
SETUP
CHANGE UNITSPART
RETURN TO WEIGH
CALIBRATION
LIBRARY
SET BALANCE
LIBRARY
SET BALANCE
LIBRARY
• Press arrow button to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CHANGE MODE.
• Press Enter button.
• Press or arrow button and select CHECK
WEIGHING.
• Press Enter button.
• Press or arrow button and start with NOMINAL WT .
• Press Enter button.
• Using the arrow buttons, enter the NOMINAL WT
value, then, press Enter button.
• Repeat steps and enter parameters for OVER WT,
and UNDER WT.
• Enter a library name for the object of measurement
and press Enter button.
• Press arrow button and select SAVE TO
LIBRARY and press Enter button.
• Press arrow button and select RUN.
• Press Enter button.
• Place item to be checked on the balance pan, the
normal display indicates the weight of the item. The
bar graph indicates whether the item is under ,
accepable or over in weight based on the settings
made in the balance.
Display below illustrates a 300 gram sample weight.
CHECK WEIGHING
Go Back
Help
PrintEnter
UNDER ACCEPT
OVER = 305.00
UNDER = 295.00
299.96
GRAMS STABLE
CONTRAST
NEW
NOMINAL = 300.00
DIFF = -0.04
STOP
OVER
MAIN
18
3.10Differential W eighing
Differential weighing stores tare and weight values so a sample can be dried and the diff erence calculated at a later
time. Up to 100 samples can be stored. samples can be added to the applications library. Batch and individual
samples can be accommodated.The features in the Voy ager balance f or diff erential weighing include:
RECALL - When selected, brings up previously stored library names associated with each sample.
LIBRARY NAME - A name up to 10 characters can be entered and stored to identify the sample item.
T ARE WEIGHT - This is the tare weight of the container holding the samples, can be set to NO or YES.
AUT O TARE - Used when it is desired to automatically tare the storing container's weight, can be set to NO or YES.
AUTO SAMPLE DETECT - This f eature is used for repetitive sample weighing, can be set NO or YES.
MODE - Can be set to INDIVIDUAL or BATCH.
FINAL RESUL T - Can be set to show results in PERCENTAGE or WEIGHT.
SAMPLE NUMBERS - Desired number of samples can be entered.
SA VE TO LIBRARY - P ermits all entries to be sav ed to the Library.
EXIT TO WEIGH - When selected, will exit to standard weighing mode.
Procedure
• Press arrow button to select MAIN MENU.
WEIGHT
Go Back
Help
PrintEnter
BASIC WEIGHING
PARTS COUNTING
FILLING
CHANGE MODE
ANIMAL WEIGHING
CHECK WEIGHING
DIFF WEIGHING
FORMULATION
QUICK CHECK
CUSTOM MENU
STATISTICS
DENSITY
0%
CONTRAST
MAIN MENU
CHANGE MODE
CHANGE UNITSPART
BASIC WEIGHING
PARTS COUNTING
SETUP
FILL
CHANGE MODE
RECALL
CALIBRATION
ANIMAL WEIGHING
RESUME
CHECK WEIGHING
EXIT TO WEIGH
DIFF WEIGHING
FORMULATION
QUICK CHECK
CUSTOM MENU
STATISTICS
DENSITY
GRAMS STABLE
MOVE HIGHLIGHT
SET BALANCE
MAIN MENU
CHANGE MODE
DIFF WEIGHING
CHANGE UNITSPART
RETURN TO WEIGH
CALIBRATION
100%
0.00
MAIN MENU
SET BALANCE
LIBRARY
or
MAIN MENU
CHANGE MODE
BASIC WEIGHING
PARTS COUNTING
SETUP
FILL
CHANGE MODE
RECALL
ANIMAL WEIGHING
RESUME
CHECK WEIGHING
EXIT TO WEIGH
DIFF WEIGHING
FORMULATION
QUICK CHECK
CUSTOM MENU
STATISTICS
DENSITY
DIFF WEIGHING
CHANGE UNITSPART
CALIBRATION
BASIC WEIGHING
PARTS COUNTING
S
FILL
CHANGE MODE
RECALL
TARE WEIGHT NO
ANIMAL WEIGHING
RSETUP
AUTO TARE NO
CHECK WEIGHING
EXIT TO WEIGH
AUTO SAMPLE DETECT NO
DIFFERENTIAL WEIGHING
MODE INDIVIDUAL
FORMULATE
FINAL RESULT WEIGHT
QUICK CHECK
SAMPLE NUMBERS 3
CUSTOM MENU
STATISTIC
LIBRARY NAME NAME
DENSITY
SAVE TO LIBRARY
SET BALANCE
MAIN MENU
CHANGE MODE
RETURN TO WEIGH
DIFF WEIGHING
CHANGE UNITSPART
CALIBRATION
SETUP
SET BALANCE
RETURN TO
WEIGH
• Press Enter button.
• Using the arrow buttons, select CHANGE MODE.
• Press Enter button.
• Press or arrow button and select DIFFWEIGHING.
• Press Enter button.
• Press or arrow button and select RECALL,
SETUP or EXIT TO WEIGH.
• Press Enter button. If RECALL was selected, you
can scan the library and select the desired item.
• If SETUP is selected, continue by pressing Enter
button.
• On the SETUP MENU, using the arrow b uttons and
Enter button, select YES or NO for each of the
following items on the menu: TARE WEIGHT, AUTO
TARE, A UTO SAMPLE DETECT; for MODE, select
either INDIVIDUAL or BATCH; for FINAL RESULT,
select either WEIGHT or PERCENT ; for SAMPLE
NUMBERS, enter the desired number of samples to
be weighed; enter a name for LIBRARY NAME; scroll
down using the arrow button and select SAVE TO
LIBRARY by pressing Enter b utton.
• Scroll to RUN, press Enter button. DIFF WEIGHING-
SAMPLE 1 is displayed as sho wn belo w.
DIFF WEIGHING -- NAME-- SAMPLE 1
Go Back
Help
PrintEnter
0%
TARE WT=
INIT WT= 0.00
GRAMSSTABLE
CONTRAST
ENTER
0.00
EDIT
100%
MAIN
19
3.10Differential W eighing (Cont.)
Initial Weighing
DIFF WEIGHING - - NAME -- SAMPLE 1
0%
TARE WT= 19.09
INIT WT= 0.00
100%
19.09
CONTRAST
GRAMSSTABLE
ENTER
EDIT
DIFF WEIGHING -- NAME -- SAMPLE 1
0%
TARE WT= 19.09
INIT WT= 306.17
MAIN
100%
306.17
CONTRAST
GRAMSSTABLE
ENTER
EDIT
MAIN
Procedure (Cont.)
First, the container is tared and stored, then the product
is intially weighed in the container and stored for each
sample. A name is giv en to the samples which is stored
in the balance. After all samples have been weighed and
entered, a summary table indicates the weight of the
containers and each sample.
After the samples have undergone a process such as
heating or cooling, the entire procedure is repeated
starting with sample number 1 and continued until all
samples have been completed. The balance then
displays a new table which indicates the tare weight of
the container, the initial weight of the product, the final
weight of the product and the difference weight. To enter
samples, proceed as follows:
• Place container on pan, wait for stab le indication,
then press Enter button. The display indicates the
container weight.
• Place the first sample in the container on the pan.
The display indicates the initial weight of the first
sample.
• Press Enter button, display changes to sample 2.
• Remove first sample and container from pan.
• Place container on pan for sample two, press Enter
button. The displa y indicates the container weight.
• Place the second sample in the container on the pan.
The display indicates the initial weight of the second
sample.
• Press Enter button, display changes to sample 3.
• Remove second sample and container from pan.
• Repeat the above procedure for all samples. The
example shown at left is f or one sample.
When the last sample is placed on the balance, a summary display indicates the TARE WT and INITIAL WT.
• Using the arrow buttons, select CONTINUE or SAVE,
then enter Data File Name.
100%
Final Weighing
When all of the samples have been remov ed and processed externally , reenter Differential W eighing and
select RESUME. The final samples are weighed with
their containers one after another. When the last sample
is weighed on the balance, a final summary display is
shown. See example at bottom of the page. The final
summary indicates the TARE WT, INIT WT, FINAL WT
and DIFF (difference weight). This can be printed. The
RE-SAMPLE and EDIT selections on the displays allow
corrections to be made.
CONTRAST
RETURN
RE-SAMPLE
DELETE
20
3.11 Formulation
The Vo y ager balance can store betw een 200 minimum and 500 formulations limited only by the memory capacity of
the resident library . Each formulation can be named and hav e up to 10 components specified and identified b y name .
Once the formulations are stored in the balance library, they may be recalled and used at any time. Each component
of a given formulation can be specified as to its weight or percentage. The balance will display each element of a
formulation on a dual bar graph as a percentage and also displays the desired weight. Thus , each product may be
placed on the pan until 100% is indicated. Names are limited to 10 characters. The illustrations in this procedure are
samples for one formulation using a higher resolution balance.
SET REFERENCE
NUMEM NAUMBERS 0
SETUP
SAVE TO LIBRARY
RUN
EXIT TO WEIGH
MAIN MENU
FORMULATION
CHANGE UNITS
CALIBRATION
LIBRARY NAME
WEIGH TYPE WEIGHT
SET REFERENCE
NUMBER 0
SETUP
SAVE TO LIBRARY
RUN
EXIT TO WEIGH
CHANGE UNITS
CALIBRATION
BASIC WEIGHING
PARTS COUNTING
SETUP
FILL
CHANGE MODE
RECALL
ANIMAL WEIGHING
RESUME
CHECK WEIGHING
EXIT TO WEIGH
DIFFERENTIAL WEIGHING
FORMULATION
QUICK CHECK
CUSTOM MENU
STATISTICS
DENSITY
CHANGE MODE
FORMULATION
FORMULA SETUP
CHANGE UNITS
FORMULA 1
<UP>/<DOWN> TO SCROLL CHARACTERS
<LEFT>/<RIGHT> TO MOVE CURSOR
<ENTER> TO ACCEPT
<BACK> TO ABORT
MAIN MENU
CHANGE MODE
FORMULATION
FORMULA SETUP
SELECT WEIGH TYPE
CHANGE UNITS
WEIGHT
PERCENT
CALIBRATION
0.00
GRAMS STABLE
SET BALANCE
LIBRARY
SET BALANCE
MAIN MENU
RETURN TO
LIBRARY NAME
WEIGH
CALIBRATION
SET BALANCE
RETURN TO WEIGH
MAIN MENU
SET BALANCE
MAIN MENU
CHANGE MODE
FORMULATION
RETURN TO WEIGH
CHANGE UNITS
CALIBRATION
SET BALANCE
RETURN TO WEIGH
100%
SET BALANCE
RETURN TO WEIGH
SET BALANCE
LIBRARY
• Press arrow button to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CHANGE MODE.
• Press Enter button.
• Press or arrow button and select FORMULATION.
• Press Enter button.
• Press or arrow button and select SETUP.
• Press Enter button, FORMULA SETUP is displayed.
• Press or arrow button and select LIBRARY
NAME.
• Press Enter button.
R
• Enter a library name for the 1st formula using the
arrow buttons.
• Press Enter button when name is completed, FORMULA SETUP is displayed again.
• Press arrow button and select WEIGH TYPE.
• Press Enter button, a new screen with WEIGHT and
PERCENT is displayed. WEIGHT allo ws components
of the formula to be specified by weight. PERCENT
allows components to be specified by percentage.
FORMULA TIONS BY WEIGHT
• Press or arrow button and select WEIGHT.
• Press Enter button.
• Using arrow button, scroll to ITEM NUMBER.
•Press Enter button. A new screen, SET ITEM
NUMBER appears.
• Press or arrow button and enter the number of components in the formula.
• Press Enter button, FORMULA SETUP isdisplayed.
• Press or arrow button and select SETUP.
• Press Enter button, NAME menu is displayed with the
number of items you entered f or the first f ormula.
• Press or arrow button and select item 1.
• Press Enter button, COMPONENT NAME display
appears.
• Using arrow buttons, enter the component name for
item 1 in the formula.
21
3.11 Formulation (Cont.)
MAIN MENU
CHANGE MODE
BASIC WEIGHING
PARTS COUNTING
SETUP
FILL
CHANGE MODE
R
LIBRARY NAME
ANIMAL WEIGHING
RECALL
WEIGH TYPE WEIGHT
CHECK WEIGHING
EXIT TO WEIGH
SET REFERENCE
DIFFERENTIAL WEIGHING
NUMBER OF ITEMS 0
FORMULATE
SETUP
QUICK CHECK
SAVE TO LIBRARY
CUSTOM MENU
STATISTIC
RUN
DENSITY
EXIT TO WEIGH
BASIC WEIGHING
PARTS COUNTING
SETUP
FILL
CHANGE MODE
R
ANIMAL WEIGHING
RECALL
CHECK WEIGHING
EXIT TO WEIGH
DIFFERENTIAL WEIGHING
FORMULATE
QUICK CHECK
CUSTOM MENU
STATISTIC
DENSITY
BASIC WEIGHING
PARTS COUNTING
SETUP
FILL
CHANGE MODE
R
ANIMAL WEIGHING
RECALL
CHECK WEIGHING
EXIT TO WEIGH
DIFFERENTIAL WEIGHING
FORMULATE
QUICK CHECK
CUSTOM MENU
STATISTIC
DENSITY
FORMULATION
FORMULA SETUP
CHANGE UNITS
BASIC WEIGHING
PARTS COUNTING
SETUP
FILL
CHANGE MODE
R
LIBRARYNAME
ANIMAL WEIGHING
RECALL
CALIBRATION
WEIGH TYPE
CHECK WEIGHING
EXIT TO WEIGH
SET REFERENCE
DIFFRENTIAL WEIGHING
NUMBER NUMBER
FORMULATE
SETUP
QUICK CHECK
SAVE TO LIBRARY
CUSTOM MENU
STATISTIC
RUN
DENSITY
EXIT TO WEIGH
MAIN MENU
CHANGE MODE
FORMULATION
FORMULA SETUP
LIBRARY NAME FORMULA1
WEIGH TYPE WEIGHT
SET REFERENCE
NUMBER OF ITEMS 3
SETUP
SAVE TO LIBRARY
RUN
EXIT TO WEIGH
LIBRARY NAME FORMULA1
WEIGH TYPE WEIGHT
SET REFERENCE
NUMBER OF ITEMS 3
SETUP
SAVE TO LIBRARY
RUN
EXIT TO WEIGH
CHANGE UNITS
BASIC WEIGHING
PARTS COUNTING
SETUP
FILL
CHANGE MODE
R
LIBRARYNAME
ANIMAL WEIGHING
CALIBRATION
RECALL
WEIGHT TYPE
CHECK WEIGHING
EXIT TO WEIGH
SET REFERENCE
DIFFERENTIAL WEIGHING
NUMBEM NUMBER 0
FORMULATE
SETUP
QUICK CHECK
SAVE TO LIBRARY
CUSTOM MENU
STATISTIC
RUN
DENSITY
EXIT TO WEIGH
MAIN MENU
CHANGE MODE
FORMULATION
FORMULA SETUP
CHANGE UNITS
CALIBRATION
SET BALANCE
MAIN MENU
CHANGE MODE
FORMULATION
FORMULA SETUP
SET ITEM NUMBER
CHANGE UNITS
RETURN TO
3
WEIGH
<UP>/<DOWN> TO SCROLL CHARACTERS
<LEFT>/<RIGHT> TO MOVE CURSOR
CALIBRATION
<ENTER> TO ACCEPT
<BACK> TO ABORT
SET BALANCE
MAIN MENU
CHANGE MODE
FORMULATION
FORMULATION SETUP
# NAME VALUE
CHANGE UNITS
COMPONENT NAME
RETURN TO
1 A1 0.00
A1
WEIGH
23
UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
BASIC WEIGHING
CALIBRATION
ENTER TO ACCEPT
PARTS COUNTING
SETUP
BACK TO QUIT
FILL
CHANGE MODE
R
LIBRARYNAME
ANIMAL WEIGHING
RECALL
WEIGHT TYPE
CHECK WEIGHING
EXIT TO WEIGH
SET REFERENCE
DIFFERENTIAL WEIGHING
NUMBEM NUMBER 0
FORMULATE
SETUP
QUICK CHECK
SAVE TO LIBRARY
CUSTOM MENU
STATISTIC
RUN
DENSITY
EXIT TO WEIGH
SET BALANCE
BASIC WEIGHING
PARTS COUNTING
SETUP
FILL
CHANGE MODE
R
LIBRARY NAME FORMULA1
ANIMAL WEIGHING
RECALL
WEIGH TYPE WEIGHT
RETURN TO
CHECK WEIGHING
EXIT TO WEIGH
WEIGH
SET REFERENCE
DIFFERENTIAL WEIGHING
NUMBER OF ITEMS 3
FORMULATE
SETUP
QUICK CHECK
SAVE TO LIBRARY
CUSTOM MENU
STATISTIC
RUN
DENSITY
EXIT TO WEIGH
MAIN MENU
CHANGE MODE
FORMULATION
FORMULATION SETUP
# NAME VALUE
CHANGE UNITS
1 A1 20.00
2 A2 30.00
3 A3 50.00
CALIBRATION
MAIN MENU
CHANGE MODE
FORMULATION
FORMULA SETUP
CHANGE UNITS
CALIBRATION
SET BALANCE
RETURN TO WEIGH
SET BALANCE
RETURN TO WEIGH
SET BALANCE
RETURN TO WEIGH
SET BALANCE
RETURN TO WEIGH
Procedure (Cont.)
FORMULA TIONS BY WEIGHT (cont.)
• Press Enter button, a NAME menu appears with the
value shaded.
• Press Enter button, ENTER COMPONENT WEIGHT
menu is displayed.
• Using the arrow buttons, enter the component weight
for item 1 in the formula.
• Press Enter button when desired weight is entered.
The NAME menu appears again. Repeat steps
(identified with arrow R on previous page) and enter
the component names and weight values f or the first
formula. Samples shown at left.
• When all of the entries have been made, select SAVE
TO LIBRARY in the FORMULATION SETUP menu,
and then select RUN, press Enter button. The
balance displays the first component of the formula
and the required weight.
• Place a container on the pan and tare by pressing
>O/T< button.
• Add the required amount to the container until thebalance indicates 100% on the bar graph and the proper
weight.
• Using the arrow buttons, select NEXT sho wn at the
bottom of the screen and press Enter button.The
second component of the formula is displayed. Add
the required weight. Select NEXT and repeat the
procedure for all items in the formula.
FORMULATION -- FORMULA1
0%
90% 110%
89%
20.00
A1 = 20.00 DIFF WT = 0.00
CONTRAST
GRAMS STABLE
FORMULATION -- FORMULA1
NEXT
0%
90% 110%
PREVIOUS
MAIN
89%
30.00
A2 = 30.00 DIFF WT = 0.00
CONTRAST
GRAMS STABLE
FORMULATION -- FORMULA1
NEXT
0%
90% 110%
PREVIOUS
MAIN
89%
50.00
A3 = 50.00 DIFF WT = 0.00
CONTRAST
Example using a total of 100 grams in the formulation.
GRAMS STABLE
NEXT
PREVIOUS
MAIN
• When you hav e completed weighing all items in the
formula, select NEXT again and a display appears
which indicates the Target Weight, Result and Difference W eight f or the formula. See displa y below.
• To select pre viously stored f ormulations, select
FORMULATION menu and select RECALL. This will
bring up the library of formulations.
FORMULA1 TARGET RESULT DIFF WT
A1 20.00 20.05 0.05
Formulations can be done by using percentages instead of weight values . See page 21 and repeat procedure up to
where screen choice for WEIGH TYPE is displa y ed, Then continue as follows:
• Press arrow button and select WEIGH TYPE.
• Press Enter button, a new screen with WEIGHT and
PERCENT is displayed. PERCENT allows components to be specified by percentage.
• Press or arrow button and select PERCENT.
• Press Enter button.
• Using arrow button, scroll to SET REFERENCE.
• Press Enter button. A new screen, ENTER TARGET
WEIGHT appears.
• Using arrow buttons, enter target weight.
• Press Enter button, FORMULA SETUP isdisplayed.
• Using arrow button, scroll to NUMBER OF
ITEMS .
• Press Enter button. A new screen, SET ITEM
NUMBER appears.
• Press or arrow button and enter the number of components in the formula.
• Press Enter button, FORMULA SETUP isdisplayed.
• Press or arrow button and select SETUP.
• Press Enter button, NAME menu is displayed with the
number of items you entered f or the first f ormula.
• Press or arrow button and select item 1.
• Press Enter button, COMPONENT NAME display
appears.
• Using arrow buttons, enter the component name for
item 1 in the formula.
• Press Enter button, a NAME menu appears with the
value shaded.
• Press Enter button, ENTER COMPONENT %
menu is displayed.
• Using the arrow buttons, enter the component percentage for item 1 in the formula.
• Press Enter button when desired weight is entered.
The NAME menu appears again. Repeat steps
and enter the component names and percent values
for the first formula.
• When all of the entries have been made , select SAVE
TO LIBRAR Y in the FORMULATION SETUP menu,
and then select RUN, press Enter button. The
balance displays the first component of the formula
and the required weight.
• Place a container on the pan and tare by pressing
>O/T< button.
• Add the required amount to the container until the
balance indicates 100% on the bar graph and the
proper percentage.
• Using the arrow buttons and select NEXT shown at
the bottom of the screen and press Enter button.The
second component of the formula is displayed. Add
the required weight. Select NEXT and repeat the
procedure for all items in the formula.
• When you have completed weighing all items in the
formula, select NEXT again and a display appears
which indicates the Target Weight, Result and Difference Weight for the formula. See display below.
FORMULA2 TARGET RESULT DIFF WT
B1 10.00 10.03 0.03
Quick check weighing permits you to place a reference sample or a sample weight on the balance pan which is used
as a reference weight to measure against similar samples. A single bar display indicates up to 100% of the capacity
of the balance, the difference in weight between the original sample and suceeding samples is displayed. The
difference in percentage is also shown along with the reference weight. The large n umeric displa y indicates the
weight of the subject.
PLEASE PUT THE REFERENCE WEIGHT
ON PAN AND PRESS < ENTER > . . .
WAIT FOR STABLE . . .
SET BALANCE
QUICK CHECK
RETURN TO WEIGH
SET REFERENCE
100%
• Press arrow button to select MAIN MENU.
• Press Enter button.
• Using the arrow buttons, select CHANGE MODE.
• Press Enter button.
• Press or arrow button and select QUICK
CHECK.
• Press Enter button.
• Place the reference weight on the pan and press the
Enter button.
• Remove reference weight from the pan and place
sample to be compared against the reference weight
on the pan.
• The balance displays the difference of the sample
weight against the reference weight in a measuring
unit and percentage. The bar graph indicates the
percentage of weight of the capacity of the balance.
• To enter a ne w reference weight, select NEW at the
bottom of the display using the arrow button and
press the Enter button and repeat procedure.
CONTRAST
GRAMSUNSTABLE
NEW
STOP
MAIN
24
QUICK CHECK
Go Back
Help
PrintEnter
0%
DIFF = 0.00 REF WT = 300.44
DIFF % = 0.00
300.44
GRAMS STABLE
CONTRAST
NEW
STOP
100%
MAIN
Loading...
+ 299 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.