OHAUS ST5000-F, ST5000-B User manual

Page 1
STARTER 5000
Bench pH Meter Instruction Manual
STARTER 5000
Medidor de pH de banco
STARTER 5000
Manuel d’instructions
PH-mètre de laboratoire
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
TABLE OF CONTENTS
INTRODUCTION ................................................................ 1
1
1.1 Definition of Signal Warnings and Symbols ................................. 1
1.2 Safety Precautions ........................................................................... 2
2 INST ALLATION ................................................................. 3
2.1 Package contents ............................................................................. 3
2.2 Installing the stand-alone electrode holder ................................... 4
2.3 Installing the power adapter ........................................................... 4
2.4 Connect the pH electrodes ............................................................. 5
3 DISPLAY A ND ICONS ....................................................... 5
3.1 Turn on the meter ............................................................................. 5
3.2 Turn off the meter............................................................................. 6
3.3 Icons .................................................................................................. 6
3.4 Menu directory ................................................................................. 9
4 CALIBRATION, MEASUREMENT AND PRINT .............. 10
4.1 Calibration ...................................................................................... 10
 Buffer group ..................................................................................... 10
 Performing calibration ....................................................................... 11
4.2 Sample measurement .................................................................... 12
 pH measurement ............................................................................. 12
 mV and Rel.mV measurement ......................................................... 12
4.3 Temperature measurement ........................................................... 12
4.4 Printing ............................................................................................ 13
5 SETUP ............................................................................. 14
5.1 System Setting ............................................................................... 14
 System setting Save Mode .......................................................... 14
 System Setting Output ................................................................. 14
 System Setting Restore Factory Setup ....................................... 14
 System Setting Service Password .............................................. 14
5.2 Measurement Setting ..................................................................... 15
 Measurement Setting Endpoint Mode .......................................... 15
 Measurement Setting GLP Measurement Mode .......................... 15
 Measurement Setting Continuous Measure ................................. 15
5.3 Calibration Setting ......................................................................... 16
 Calibration Setting Buffer Group .................................................. 16
 Calibration Setting Calibration Reminder ..................................... 16
5.4 Data Log .......................................................................................... 16
6 MAINTENANCE ............................................................... 17
6.1 Error message ................................................................................ 17
6.2 Meter maintenance ........................................................................ 17
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
6.3 Electrode maintenance .................................................................. 17
6.4 Information ..................................................................................... 17
7 TECHNICAL DATA .......................................................... 18
7.1 Specifications ................................................................................. 18
8 COMPLIANCE ................................................................. 19
9 BUFFER GROUPS .......................................................... 20
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
STARTER 5000 Bench pH Meter EN-1
1 INTRODUCTION
Thank you for choosing this OHAUS product. Please read the manual completely before using the STARTER 5000 bench pH meter to avoid incorrect operation.
STARTER 5000 is a 0.001pH bench meter with a color touch-screen LCD. It’s a new generation of intuitive and easy-to-use bench meters for reliable measurements. A few of the many existing features are:
x Bi-lingual graphical user interface x Auto, manual time endpoint which can assist users to freeze the stable reading value to
meet different customer requirements
x 8 buffer groups (US, EU, NIST and JJG) with 1 self-defined buffer x IP54 water and dust protection
1.1 Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results.
Signal Words
WARNING For a hazardous situation with medium risk, possibly resulting in
injuries or death if not avoided.
CAUTION For a hazardous situation with low risk, resulting in damage to the
device or the property or in loss of data, or injuries if not avoided.
Attention For important information about the product
Note For useful information about the product
Warning Symbols
General Hazard Electrical Shock Hazard
Explosion Hazard Corrosive Hazard
Alternating Current Direct Current
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
EN-2 STARTER 5000 Bench pH Meter
1.2 Safety Precautions
CAUTION Read all safety warnings before installing, making connections, or
servicing this equipment. Failure to comply with these warnings could result in
x Verify that the AC adapter's input voltage range and the plug type are compatible with the
local AC power supply.
x Make sure that the power cord does not pose a potential obstacle or tripping hazard. x Use the equipment only in dry locations. x Dry off any liquid spills immediately. The instrument is not watertight. x Use only approved accessories and peripherals. x Operate the equipment only under ambient conditions specified in these instructions. x Disconnect the equipment from the power supply when cleaning. x Do not operate the equipment in hazardous or unstable environments. x Service should only be performed by authorized personnel.
x I f the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
personal injury and/or property damage. Retain all instructions for future reference.
WARNING Do not use this instrument in an environment subject to explosion hazards. The housing of the instrument is not gas tight (explosion hazard due to spark formation, corrosion caused by the ingress of gases).
CAUTION When using chemicals and solvent s, comply with the instructions of the producer and the general lab safety rules.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
ST ARTER 5000 Bench pH Meter EN-3
Model
Description
P/N
ST350
3-in-1 glass refillable pH Electrode
30129354
ST310
3-in-1 plastic refillable pH Electrode
83033965
ST210
2-in-1 plastic refillable pH Electrode
83033966
ST320
3-in-1 plastic gel pH Electrode(no need to fill)
83033967
ST260
2-in 1 glass double salt bridge pH Electrode
30129357
ST230
2-in-1 glass muddy sample pH Electrode
83033968
STSURF
2-in-1 flat surface pH Electrode
30129470
STMICRO5
2-in-1 glass micro sample pH Electrode
30087566
STMICRO8
2-in-1 glass micro sample pH Electrode
30087569
STTEMP30
Temperature Electrode
83033970
STORP1
Gel plastic ORP electrode
30038555
STORP2
Refillable glass ORP electrode
30038553
ST5000-B
Units
ST5000 meter
1
ST5000 In use cover
1
Stand-alone electrode holder
1 set
4GB USB flash drive
1
Buffer powder set (4.01; 7.00; 10.00)
1 set
Power supply
1 set
ST350 3-in-1 glass refillable pH electrode
1
Buffer powder set (4.01; 7.00; 10.0)
83033971
Buffer pH1.68 250ml
30100424
Buffer pH4.01 250ml
30100425
Buffer pH7.00 250ml
30100427
Buffer pH10.01 250ml
30100429
Buffer pH12.45 250ml
30100440
pH electrode Reference Electrolyte
30059255
pH electrode protection solution
30059256
Electrode holder stand alone
30058733
ST5000 In use cover
30129897
SP Cable Adapter RS232 F-M SF40A (for ST5000)
30059316
2 INSTALLATION
Carefully unpack the box.
2.1 Package contents
The model ST5000-B (basic package) has the following items:
In addition to ST5000-B content, the ST5000-F package also includes the following:
Each pH buffer powder should be dissolved in 250ml pure water or deionized water in a volumetric flask. You also can order the bottled buffer solution from Ohaus. More electrodes, solutions, accessories and spare parts, please contact Ohaus for more details.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
EN-4 STARTER 5000 Bench pH Meter
2.2 Installing the stand-alone electrode holder
Install the electrode arm on the base:
After adjusting the tension knob to some extent, t he upper arm can be move up and down. When the pH electrode is placed on the arm and calibration or measurement is being performed, the storage bottle of pH electrode can be placed into the base hole for easy storage:
2.3 Installing the power adapter
Insert the adapter plug into the power adapter slot.
CAUTION Only use an adapter plug that is compatible with the mains power receptacle. Make sure that the adapter plug is fully seated in the power adapter slot before plugging it into the mains power receptacle.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
STARTER 5000 Bench pH Meter EN-5
2.4 Connect the pH electrodes
There are 2 sockets for the electrode. “pH” socket (BNC); “TEMP.” Socket (Cinch). For ST350 3-in-1 electrode, connect both the BNC and Cinch socket. For 2 -in-1 pH electrodes (e.g. ST230), only connect the BNC.
3 DISPLAY AND ICONS
3.1 Turn on the meter
STARTER 5000 is a color touch screen 0.001pH bench meter. It does not have any buttons. After connecting with power, the meter will turn on, if meter in screen protection, you can tap the screen to turn on the screen. Note: If customer did not operate the meter for more than 6 hours, the meter will completely shut off automatically. You cannot turn on the meter by tap ping the screen any more. You need to reconnect power to turn on the meter.
Select the language and then press to enter the main interface. Main Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
EN-6 STARTER 5000 Bench pH Meter
User ID
Sample ID
Information
Date
Time
pH measurement mode. Shift between mV and Rel.mV mode or
Signal icon, when it
is blinking that means the meter is receiving signal (in
measurement or calibration).
 
Endpoint icons, auto When in manual-endpoint, if the signal is stable, the meter will show the stable signal
icon. Tap the stop icon
to reach manual endpoint.
Signal stable icon
Tempera
3.2 Turn off the meter
To turn off the meter, please tap the turn-off button in the top-left of main interface to enter
the turn -off interface. Then tap main interface again.
Turn-off interface
3.3 Icons
icon
to turn-off the meter or tap to go back to

-endpoint, manual-endpoint and time-endpoint.
ture icon.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
ST ARTER 5000 Bench pH Meter EN-7
Endpoint:"
Tap this area to enter the interface to set the endpoint mode. Available modes are Auto, Manual and Time end
is
displayed.
Sensor ID:"
Tap this area to enter the interface to select, set or change the sensor ID. The current sensor ID is displayed.
Cond:"
Tap this area to sensor performance icon (judged by calibration data) is displayed.
and offset: ± (0 Electrode
or offset: ± (
) mV Electrode condition is acceptable
%
or offset: ± ( Electrode
not
good or needs cleaning
Calibration reminder icon calibration from last calibration. When the time is reached, this icon will be displayed and blink to remind that it is time to perform a calibration.
Buffers:”
This area displays the current buffer group, tap this area to enter the interface to select the buffer group.
Start icon. Tap this icon and the meter wil
icon will blink. The start icon area will show the stop icon
Calibration icon. Tap this icon to start calibration.
Save icon. After saving the current measurement result (endpoint reached), this icon color will change
, indicating that the reading has
already been saved and cannot be saved twice.
Output icon. Output
If “Shut off connection” is chosen in
icon
therwise the output icon will be displayed in the
same position.
Information, measure and calibrate tap area The information area is also the shortcut to enter the setting interface:
point. The current endpoint mode
Sensor
There are 3 sensor condition icons:
Slope: more than 95%
-15) mV
condition is good
enter the sensor calibration data interface. The current
Slope: 90-95%
15-35
. Set the time you want the meter to alert you to do
l begin measurement, the signal
Slope: less than 90
35-60) mV
condition is
from black to light gray
data to external device such as a printer.
System Setting Output, the save
will be displayed .O
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 11
EN-8 STARTER 5000 Bench pH Meter
Return to main interface
Return to previous menu
Enter, confirm or apply current setup, finish setting
Edit
Page up
Page down
Delete
Increase (light)
Decrease (light)
Graph display (for the measurement or calibration data)
Save
Output to external device (USB flash drive, printer)
Full keyboard
Next calibration point
Current Calibration point again
End of calibration
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
ST ARTER 5000 Bench pH Meter EN-9
System Setting
Date
Time
Save Mode
Output
Screen Brightness
Baud Rate
Screen Protection
Restore Factory Setup
Instrument Information
Service Passwo r d
Measurem Setting
Resolution
MTC Temp. Input
pH/mV/Rel.mV Mode
Continuous Measure
Endpoint Mode
ID setting
GLP Measurement Mode
Calibration Setting
Buffer Group
Calibration Reminder
Data Measurement Data
Calibration Data
3.4 Menu directory
The menus of first and secondary directory for STARTER 5000:
ent
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 13
EN-10 STARTER 5000 Bench pH Meter
4 CALIBRATION, MEASUREMENT AND PRINT
The meter display will turn off automatically according to the time set in the menu System Setting
Screen Protection LCD Auto-off. To turn on the screen again, tap the screen.(See 3.1)
A complete pH measurement process should follow the procedure:
a) Rinse the pH electrode and connect to meter b) Prepare the buffer and calibrate the pH electrode c) Prepare the sample d) Conduct pH measurement of the sample e) Record measurement data
f) Rinse the pH electrode and store Prepare the pH electrode: the pH electrode should be rinsed with pure water before and after usage. Check if the electrode is physically damaged. (Be careful with the glass bulb)
The pH electrode should be stored in the storage bottle. After the pH electrode has been put into the sample or buffer solution, stir several seconds then wait 30 to 60 seconds for the signal to be stable, and then press the button to operate (Calibration or measurement).
WARNING Do not operate the equipment in hazardous environments. The equipment is not explosion protected.
CAUTION When using chemicals and solvents, comply with the instructions of the chemical producer and the general lab safety rules.
4.1 Calibration
 Buffer group
STARTER 5000 can perform 1-5 point calibrations for predefined buffer groups. For self-defined buffer group, it can do 1-9 point calibration.
Before calibration, make sure the current buffer group is the group wanted. When using a new pH electrode, a calibration should be performed before any m easurement. If a pH electrode is not used for a few days, a calibration should be performed before next usage.
Calibration is to display the right pH value when the meter receive the mV value signal from the pH electrode.
Slope: the linear coefficient between mV and pH acco rding to theoretical value (e.g. -
59.16mV/pH @ 25°C means 100% slope);
Offset: the mV value when the pH value is 7.00. (Theoretical value is 0 mV)
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 14
ST ARTER 5000 Bench pH Meter EN-11
5°C
1.67
4.01
7.09
10.25
10°C
1.67
4.00
7.06
10.18
15°C
1.67
4.00
7.04
10.12
20°C
1.68
4.00
7.02
10.06
25°C
1.68
4.01
7.00
10.01
30°C
1.68
4.01
6.99
9.97
35°C
1.69
4.02
6.98
9.93
40°C
1.69
4.03
6.97
9.89
45°C
1.70
4.05
6.97
9.86
50°C
1.71
4.06
6.96
9.83
STARTER 5000 will give an electrode condition icon to show the performance of the pH electrode according to the calibration data. Normally, if we get sad face icon after a proper calibration, you may need to replace the pH electrode with a new one.
STARTER 5000 automatically corrects for the t emperature dependence of the buffer pH values. E.g. for US standard buffer group "1.68, 4.01, 7.00, 10.01" reference the following table:
For example, with calibration buffer 10.01 at 15°C, the calibration result should be 10.12 pH.
 Performing calibration
After rinsing the pH electrode, put the electrode into the first buffer, stir for 10s then press to start the calibration. When in calibration, the signal icon blinks. When the endpoint has been
reached, are now 3 options:
Quit the calibration and go back to main interface
Finish current point calibration, go to calibration result interface
Do the next point calibration
Note: During calibrating process, the ST5000 is always in Auto Endpoint Mode. Note: The use of 3-in-1 pH electrode which include the temperature sensor or a separate
temperature electrode is recommended. When using the MTC mode, enter the correct temperature value and keep all buffer and sample solutions at the set temperature.
appears while disappears and current point calibration is finished, There
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 15
EN-12 STARTER 5000 Bench pH Meter
4.2 Sample measurement
 pH measurement
Be sure the electrode has been calibrated before measurement. After rinsing,
Place the pH electrode in the sample.
Press
measurement is finished.
Note: If the signal is not stable all the time, it is recommended to use manual endpoint mode to avoid bad repeatability.
to start the pH measurement, when the endpoint has been reached, the current
 mV and Rel.mV measurement
Press to shift to or mode; follow the same procedure as for pH measurement to perform mV or Rel.mV measurement. If the meter is connected t o an ORP electrode, the mV result is the value of Oxidation-Reduction Potential value; also called Redox value.
4.3 Temperature measurement
For better accuracy, we recommend to use either a built-in or a separate temperature electrode.
If a temperature electrode is used, ATC and the sample temperature are displayed. If the meter does not detect a temperature electrode, it automatically switches to the manual
temperature compensation mode and MTC appears. MTC temperature should be set.
Note: STARTER 5000 accepts NTC 30 kΩ temperature sensor. For ATC (Auto Temperature
Compensation) or MTC (Manual Temperature Compensation), the Temperature Compensation only corrects for the change in the output of the electrode, not for the change in the actual solution. That means the meter corrects the signal (mV) from the pH electrodes t o get a more accurate pH value according to the real temperature.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 16
ST ARTER 5000 Bench pH Meter EN-13
4.4 Printing
STARTER 5000 can print the measurement or calibration data through RS232. After the printer (e.g. OHAUS SF-40A) is connected to the meter properly, make the following settings(See 5.1):
please order the connect cable for ST5000 (P/N 30059316) to connect SF40A and ST5000.
System Setting Output , select Printer System Setting Baud Rate, select 9600.
The format for the print-out measurement is:
Date Time User SampleID
6-13-2015 14:20:22 User01 Sample02
SensorID Endpoint mode MTC/ATC
Sensor02 Manual ATC
Temperature Result
25.3°C 6.999pH The format for the print-out calibration data is: Calibration Data
SensorID Serial Number Sensor02 14060012 Cal. Date Cal. Time ATC/MTC 01-01-2015 14:20:33 ATC Buffers mV Offset Slope
4.00pH 176.94mV
7.00pH 0.05mV 0.05mV 98.22% When performing continuous measurement, the format for the print-out calibration data is: Timing Measure Record
SensorID Serial Number
Sensor03 14060012
Date Time Temperature Measurement
01-01-2015 14:20:25 23.5°C 6.998pH
Date Time Temperature Measurement
01-01-2015 14:20:30 23.5°C 6.999pH
Date Time Temperature Measurement
01-01-2015 14:20:35 23.5°C 7.000pH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 17
EN-14 STARTER 5000 Bench pH Meter
5 SETUP
5.1 System Setting
System setting is used to set meter parameters, the meter information could also be f ound in it. Please see section 3.4 for reference.
 System setting ►Save Mode
Save mode has 2 options, one is Auto Save Endpoint Reading, another is Manual Save Endpoint Reading. Auto Save Endpoint Reading means that when the measurement reac hes the endpoint (auto, manual or time endpoint), the meter will save the endpoint reading automatically.
The Save icon saved only once.
will change from black to light grey
; An endpoint reading can be
 System Setting ►Output
When selecting shut off connection, the main interface will display the save icon , or
display the Output Icon

 System Setting Restore Factor y Setup
Restoring the factory setup includes the following:
1. Shut off GLP mode
2. Shut off A uto Save m ode
3. Shut off co ntinuous measure
4. Shut off (output) connection
5. Shut off calibration Reminder
6. The resolution is: 0.001pH, 0.01mV
7. Measurement mode: pH
8. Self buffer group will be deleted
9. Baud rate: 9600
10. Brightness of screen: 70%
11. Shut off Screen Protection
12. MTC: 25°C
13. Endpoint: Auto
14. Time endpoint: 10 Second
15. Delete all the calibration data
 System Setting ► Service Pass word
Not used.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 18
ST ARTER 5000 Bench pH Meter EN-15
5.2 Measurement Setting
 Measurement Setting E ndpoint Mode
Auto Endpoint In this mode, the meter will reach the endpoint and lock the reading according to the following stability criteria automatically:
0.1pH resolution: the signal of the sensor input may not change by more than 0.3mV in 6 seconds.
0.01pH resolution: the signal of the sensor input may not change by more than 0.1mV in 6 seconds.
0.001pH resolution: the signal of the sensor input may not change by more than 0.03mV in 6 seconds.
Time Endpoint In this mode, the user needs to input the endpoint time, e.g. 15 seconds. This means after starting the measurement for 15s, the meter will reach the endpoint.
Manual Endpoint In this mode, the user needs to decide when the signal reading is stable, then press the stop icon
to reach manual endpoint.
 Measurement Setting ►GLP Measurement Mode
GLP mode requires each measurement to have the User ID and Sam ple ID. If there is no sample ID or user ID, the meter will show the interface to let user input the required data.
 Measurement Setting ►Co ntinuous Measure
During continuous measure, the user needs to input the time interval to get each reading result. Below is the Graph of continuous measure.
The Y axis is pH value. The x axis is the Number of measurement result.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 19
EN-16 STARTER 5000 Bench pH Meter
5.3 Calibration Setting
 Calibration Setting ►Buffer Group
Before calibration, please make sure the current buffer group is correct for your buffer solutions. User can select from the 8 predefined buffer groups. User can also set self-defined buffer groups. The adjacent 2 calibration point difference should be larger than 0.99pH.
 Calibration Setting ►Calibration Reminder
Enter this setting to set the time, from the last calibration, after how many hours the meter should
remind to perform calibration again. The meter will show the calibration reminder icon blink, to remind user to perform calibration.
and
5.4 Data Log
Tap Data to get into the Data log interface, including measurement data and calibration data. You can delete all the measurement data here. The only way to delete calibration data is to restore factory setup. The data can be output to a USB Flash Drive (≤4GB).
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 20
ST ARTER 5000 Bench pH Meter EN-17
6 MAINTENANCE
6.1 Error message
When the input value/measurement result is out of range or calibration data is not good enough, the meter will display error information. If the error cannot be solved by the error message displayed, please contact Ohaus sales representative for further assistance.
6.2 Meter maintenance
The STARTER 5000 series instruments do not require any maintenance other than occasional wiping with a damp cloth. Attention: Follow these safety precautions when cleaning the instrument
x Disconnect the instrument from the power supply x Do not open the instrument x Do not allow liquids to enter the instrument
Attention: The housing is made of acrylonitrile butadiene styrene (ABS). This material is attacked by some organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK). Any spillage should be immediately wiped off.
6.3 Electrode maintenance
Make sure the electrode is filled with electrolyte solution. Always store the electrode according to the electrode instruction manual and do not allow it to dry out. If the electrode response becomes sluggish or the slope is not good enough, try the following:
o Soak the electrode in 0.1M HCl for more than 8 hours.
o For fat or oil contaminant, degrease the membrane with cotton wool soaked in either
acetone or a soap solution. After electrode treatment, a new calibration should be performed. If the electrode slope is still not good, the electrode might need to be replaced.
6.4 Information
Tap the Information icon to enter the information interface. Here the user can input related information by tapping the area with "
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
" in the bottom left such as sensor purchase date, etc.
Page 21
EN-18 STARTER 5000 Bench pH Meter
Model
SSTARTER 5000
Measuring range
-2.00020.000 pH
-30
Resolution
0.1/0.01/0.001 pH 1
0.1
Error limits
± 0.002 pH ± ± 0.1
Calibration
1-9 points 8 predefined buffer group+1 self-defined buffer group
Memory
1000 measurements 10*10 calibration data
Power supply
100-240V~0.3A input, 9V 1.12A output AC Adapter
Size/weight
Approximately 210 x 170 x 80 mm / 0.65 kg
Ship dimension/weight
Approximately 372 x 310 x 185 mm / 2.30 kg Display
Touch screen liquid crystal with backlight
Input
BNC, impedance > 3 x 10e+12 Ω Cinch, NTC 30 kΩ
Temperature-compensation
ATC & MTC
Housing
ABS
7 TECHNICAL DATA
7.1 Specifications
The technical data is valid under the following ambient conditions:
x Indoor use only x Altitude: Up to 2000 m x Specified Temperature range: 5°C to 40°C x Humidity: maximum relative humidity 80 % for temperatures up t o 30°C decreasing linearly to
50% relative humidity at 40°C
x Mains supply voltage fluctuations: up to ±10% of the nominal voltage x Installation category II x Pollution degree: 2 x Operability is assured at ambient temperatures between 5°C to 40°C
2000.002000.00 mV °C …130°C
/0.1/0.01 mV
°C
0.03% FS mV °C
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 22
STARTER 5000 Bench pH Meter EN-19
8 COMPLIANCE
This product conforms to the EMC Directive 2004/108/EC and the Low Voltage Directive 2006/95/EC.
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements. The Batteries Directive 2006/66/EC introduces new requirements from September 2008 on removability of batteries from waste equipment in EU member States. To comply with this Directive, this device has been designed for safe removal of the batteries at end-of-life by a waste treatment facility. Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.Thank you for your contribution to environmental protection.
FCC Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Industry Canada Note
This Class A digital apparatus complies with Canadian ICES-001
ISO 9001 Registration
In 1994, OHAUS Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the OHAUS quality management system is compliant with the ISO 9001 standards requirements. On June 21, 2012, OHAUS Corporation, USA, was re-registered to the ISO 9001:2008 standard.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 23
EN-20 STARTER 5000 Bench pH Meter
e
9 BUFFER GROUPS
STARTER 5000 automatically correct for the temperature dependence of the buffer group pH value given in the following tables. Please choose the buffer group you want before calibration.
JJG119 ((Ref. 25°C, 0.01pH) Temp °C 1.68 4.00 6.86 9.18 12.46
5 1.67 4.00 6.95 9.39 13.21 10 1.67 4.00 6.92 9.33 13.01 15 1.67 4.00 6.90 9.28 12.82 20 1.68 4.00 6.88 9.23 12.64
25 1.68 4.00 6.86 9.18 12.46
30 1.68 4.01 6.85 9.14 12.29 35 1.69 4.02 6.84 9.11 12.13 40 1.69 4.03 6.84 9.07 11.98 45 1.70 4.04 6.83 9.04 11.83 50 1.71 4.06 6.83 9.02 11.70
US standard (Ref. 25°C) Temp °C 1.68 4.01 7.00 10.01
5 1.67 4.01 7.09 10.25 10 1.67 4.00 7.06 10.18 15 1.67 4.00 7.04 10.12 20 1.68 4.00 7.02 10.06
25 1.68 4.01 7.00 10.01
30 1.68 4.01 6.99 9.97 35 1.69 4.02 6.98 9.93 40 1.69 4.03 6.97 9.89 45 1.70 4.05 6.97 9.86 50 1.71 4.06 6.96 9.83
Europe standard (Ref. 25°C) Tempe
C 2.00 4.01 7.00 9.21 11.00
5 2.02 4.01 7.09 9.45 11.72 10 2.01 4.00 7.06 9.38 11.54 15 2.00 4.00 7.04 9.32 11.36 20 2.00 4.00 7.02 9.26 11.18
25 2.00 4.01 7.00 9.21 11.00
30 1.99 4.01 6.99 9.16 10.82 35 1.99 4.02 6.98 9.11 10.64 40 1.98 4.03 6.97 9.06 10.46 45 1.98 4.04 6.97 9.03 10.28 50 1.98 4.06 6.97 8.99 10.10
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 24
STARTER 5000 Bench pH Meter EN-21
MERCK (Ref. 20°C) Temp °C 2.00 4.00 7.00 9.00 12.00
5 2.01 4.04 7.07 9.16 12.41 10 2.01 4.02 7.05 9.11 12.26 15 2.00 4.01 7.02 9.05 12.10
20 2.00 4.00 7.00 9.00 12.00
25 2.00 4.01 6.98 8.95 11.88 30 2.00 4.01 6.98 8.91 11.72 35 2.00 4.01 6.96 8.88 11.67 40 2.00 4.01 6.95 8.85 11.54 45 2.00 4.01 6.95 8.82 11.44 50 2.00 4.00 6.95 8.79 11.33
DIN (19267) (Ref. 25°C) Temp °C 1.09 4.65 6.79 9.23 12.75
5 1.08 4.67 6.87 9.43 13.63 10 1.09 4.66 6.84 9.37 13.37 15 1.09 4.66 6.82 9.32 13.16 20 1.09 4.65 6.80 9.27 12.96 25 1.09 4.65 6.79 9.23 12.75 30 1.10 4.65 6.78 9.18 12.61 35 1.10 4.65 6.77 9.13 12.45 40 1.10 4.66 6.76 9.09 12.29 45 1.10 4.67 6.76 9.04 12.09 50 1.11 4.68 6.76 9.00 11.98
JIS Z 8802 (Ref. 25°C) Temp °C 1.679 4.008 6.865 9.180
5 1.668 3.999 6.951 9.395 10 1.670 3.998 6.923 9.332 15 1.672 3.999 6.900 9.276 20 1.675 4.002 6.881 9.225
25 1.679 4.008 6.865 9.180
30 1.683 4.015 6.853 9.139 35 1.688 4.024 6.844 9.102 40 1.694 4.035 6.838 9.068 45 1.700 4.047 6.834 9.038 50 1.707 4.060 6.833 9.011
DIN(19266) / NIST (Ref. 25°C) Temp °C 1.680 4.008 6.865 9.183 12.454
5 1.668 4.004 6.950 9.392 13.207 10 1. 670 4.001 6.922 9.331 13.003 15 1. 672 4.001 6.900 9.277 12.810 20 1. 676 4.003 6.880 9.228 12.627
25 1. 680 4.008 6.865 9.183 12.454
30 1. 685 4.015 6.853 9.144 12.289 35 1. 691 4.026 6.845 9.110 12.133 40 1. 697 4.036 6.837 9.076 11.984 45 1. 704 4.049 6.834 9.046 11.841 50 1. 712 4.064 6.833 9.018 11.705
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 25
EN-22 STARTER 5000 Bench pH Meter
JJG119 (Ref. 25°C, 0.001pH) Temp °C 1.680 4.003 6.864 9.182 12460
5 1.669 3.999 6.949 9.391 13.210 10 1.671 3.996 6.921 9.330 13.011 15 1.673 3.996 6.898 9.276 12.820 20 1. 676 3.998 6.879 9.226 12.637
25 1. 680 4.003 6.864 9.182 12.460
30 1. 684 4.010 6.852 9.142 12.292 35 1.688 4.019 6.844 9.105 12.130 40 1. 694 4.029 6.838 9.072 11.975 45 1.700 4.042 6.834 9.042 11.828 50 1. 706 4.055 6.833 9.015 11.697
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 26
ST ARTER 5000 Bench pH Meter EN-23
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warrant y period Ohaus will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus. This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the dat e of shipment t o the authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for any consequential damages. As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for further details.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...