OHAUS ST3100-F, ST3100-B User manual

Page 1
STARTER 3100 Bench pH Meter Instruction Manual
Manual de instrucciones del pHmetro de mesa STARTER 3100
STARTER 3100 PH-mètre de laboratoire
Manuel d’instructions
STARTER 3100 Medidor de pH de Bancada Manual de Instruções
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
TABLE OF CONTENTS
1 INTRODUCTION ................................................................... 1
1.1 Definition of Signal Warnings and Symbols ...................................... 1
1.2 Safety Precautions ............................................................................ 2
1.3 Display and controls .......................................................................... 3
2 INSTALLATION ..................................................................... 5
2.1 Package contents .............................................................................. 5
2.2 Installing the stand-alone electrode holder ....................................... 6
2.3 Installing the power adapter .............................................................. 6
2.4 Connect the pH electrodes ................................................................ 7
2.5 Meter stand for adjusting view angle ................................................. 7
2.6 Attached quick guide ......................................................................... 7
3 SETUP ................................................................................... 8
3.1 Set temperature unit and MTC value ................................................ 8
3.2 Selecting a predefined buffer group .................................................. 8
4 STARTER 3100 OPERATION ............................................... 9
4.1 Calibration ......................................................................................... 9
4.1.1 Buffer group ................................................................................... 9
4.1.2 Performing 1-point calibration ..................................................... 10
4.1.3 Performing 2-point calibration ...................................................... 11
4.2 Sample measurement ..................................................................... 12
4.2.1 pH measurement ......................................................................... 12
4.2.2 mV measurement ........................................................................ 12
4.3 Temperature measurement ............................................................. 12
4.4 Using the memory ........................................................................... 12
4.4.1 Storing a reading ......................................................................... 12
4.4.2 Recalling from memory ............................................................... 12
4.4.3 Clearing the memory ................................................................... 13
4.5 Printing ............................................................................................ 13
5 MAINTENANCE .................................................................. 15
5.1 Error message ................................................................................. 15
5.2 Meter maintenance.......................................................................... 15
5.3 Electrode maintenance ................................................................... 16
5.4 Self-diagnosis .................................................................................. 16
5.5 Recover factory settings .................................................................. 16
6 TECHNICAL DATA .............................................................. 17
6.1 Specifications .................................................................................. 17
6.2 Compliance ..................................................................................... 18
7 BUFFER GROUP ................................................................ 19
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
STARTER 3100 Bench pH Meter EN-1
General hazard
Explosion hazard
Corrosive hazard
Alternating current
Direc
1 INTRODUCTION
Thank you for choosing OHAUS.
Please read the manual completely before using the STARTER 3100 bench pH meter to ensure proper setup, operation and maintenance.
STARTER 3100 has an excellent performance/price ratio and is designed with many useful features including a stand-alone electrode holder and a beeper for error alert. Other accessories such as pH electrodes and buffer solutions are also available.
STARTER 3100 offers many practical features and functions such as:
x Large backlit LCD with Quick Guide attached under the meter to assist with user operation x Stand-alone electrode holder for maximum flexibility x Electrode condition icon automatically showing you the health of your pH electrode x RS232 port and memory to store up to 99 measurements
1.1 Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results.
Signal Words WARNING For a hazardous situation with medium risk, possibly resulting in
CAUTION For a hazardous situation with low risk, resulting in damage to the
Attention For important information about the product. Note For useful information about the product Warning Symbols
t current
injuries or death if not avoided.
device or the property or in loss of data, or injuries if not avoided.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
EN-2 STARTER 3100 Bench pH Meter
1.2 Safety Precautions
CAUTION: Read all safety warnings before installing, making connections, or servicing this
equipment. Failure to comply with these warnings could result in personal injury and/or property damage. Retain all instructions for future reference.
x Verify that the input voltage range printed on the data label and the plug type matches the
local AC power to be used.
x Make sure that the power cord does not pose a potential obstacle or tripping hazard. x Use the equipment only in dry locations. x Dry off any liquid spills immediately. The instrument is not watertight. x When using chemicals and solvents, comply with the instructions of the chemical producer
and the general lab safety rules.
x Use only approved accessories and peripherals. x Operate the equipment only under ambient conditions specified in these instructions. x Disconnect the equipment from the power supply when cleaning. x Do not operate the equipment in hazardous or unstable environments. x Service should only be performed by authorized personnel.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
STARTER 3100 Bench pH Meter EN-3
1
Electrode condition
and offset: ± (0 Electrode
good
or offset: ± ( Electrode acceptable
or offset: ± ( Electrode
not
good or needs cleaning
2
Endpoint stability icon
; Auto endpoint icon
3
Measurement icon - ; means measurement or calibration is running when blinks
4
Calibration icon - ; means calibration in progress when display
5
Setup icon
group
etc.
6
pH/mV reading or slope in calibration process
7
Calibration point / Buffer group /Memory number / Error index
8
Auto temperature compensation - ATC ; Manual temperature compensation - MTC
9
Temperature during measurement or offset (mV) in calibration process
2 3 4 1 5
8
9
1
2 3 4
5
6
1.3 Display and controls
Displays
Slope: more than 95%
-15) mV
condition is
- ; instrument is in the setup mode, can set temperature(MTC), buffer
Slope: 90-95%
15-35) mV
condition is
Slope: less than 90%
35-60) mV
condition is
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
EN-4 STARTER 3100 Bench pH Meter
Button
Press & release
Press & hold
- Start or finish measurement
- Confirm setting, store entered
- Turn auto endpoint on / off
- Start calibration
- Recall the latest calibration
- Meter turn on
-
to measurement
- Meter turn off
- Store current reading to memory
-
-
- Recall stored data
-
- Switch between pH and mV
-
- Scroll down through the memory
- Enter setup mode
-
Turn on/turn off the backlight of the LCD
Controls
for 3 seconds
value
Exit and return screen
Increase value during setting Scroll up through the memory
Decrease value during setting
Start self-diagnosis
,
data : slope and offset
Print current memory data
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
STARTER 3100 Bench pH Meter EN-5
Model
Description
P/N
ST310
3-in-1 plastic refillable pH Electrode
83033965
ST210
2-in-1 plastic refillable pH Electrode
83033966
ST320
3-in-1 plastic gel pH Electrode(no need to fill)
83033967
ST230
2-in 1 glass muddy sample pH Electrode
83033968
STORP1
Gel plastic ORP electrode
30038555
STORP2
Refillable glass ORP electrode
30038553
STTEMP30
Temperature Electrode
83033970
ST
U
1
Stand-alone electrode holder
1 set
In use cover
1
12V Power supply
1 set
pH Buffer Powder Set (4.01, 7.00, 10.01)
1 set
ST310 3-in-1 refillable pH electrode
1
pH Buffer Powder Set (4.01, 7.00, 10.01)
1 set
ST320 3-in-1 gel pH electrode
1
2 INSTALLATION
Carefully unpack the box.
2.1 Package contents
The model ST3100-B (basic package) has the following items:
3100-B
STARTER 3100
In addition to ST3100-B content, the ST3100-F package also includes the following:
In addition to ST3100-B content, the ST3100-H package also includes the following:
Each pH buffer powder should be dissolved in 250ml pure water or deionized water in a volumetric flask.
Additional electrodes available include the following:
nits
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
EN-6 STARTER 3100 Bench pH Meter
Buffer powder set (4.01; 7.00; 10.00)
83033971
Buffer pH1.68 250ml
30100424
Buffer pH4.01 250ml
30100425
Buffer pH7.00 250ml
30100427
Buffer pH10.01 250ml
30100429
Buffer pH12.45 250ml
30100430
Electrode holder stand alone (3100/3100C)
30058733
In use cover
30058734
Buffers and Accessories:
2.2 Installing the stand-alone electrode holder
Install the electrode arm on the base,
After adjusting the tension knob to some extent, you can move the upper arm up and down. When the pH electrode is installed on the arm, the storage bottle of pH electrode fits into the base hole.
2.3 Installing the power adapter
Insert the right adapter clip into the power adapter slot. The Ohaus balance power supply is also 12V which could also be used for the pH meter.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
STARTER 3100 Bench pH Meter EN-7
2.4 Connect the pH electrodes
There are 3 sockets for electrode. “pH” socket (BNC); “TEMP.” Socket (Cinch) and “REF” Socket (2mm banana).
For ST310 electrode connect to the BNC and Cinch socket. For 2-in-1 pH electrode (e.g. S T210) you only need to connect the BNC.
REF socket is for separate reference electrode. (e.g.STREF1)
2.5 Meter stand for adjusting view angle

One unique design of the STARTER 3100 is the meter stand, which can be used to adjust the view angle of the display in case of working on high table.
2.6 Attached quick guide
Another unique design of the STARTER 3100 is the attached quick guide, the quick guide is attached into the bottom housing of the meter.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
EN-8 STARTER 3100 Bench pH Meter
b1
1.68
4.01
7.00
10.01
3 SETUP
3.1 Set temperature unit and MTC value
Please note: If a temperature electrode is used, Automatic Temperature Compensation (ATC) and the sample temperature are displayed on the screen. You may then choose to skip MTC setup (below).
If the meter does not detect a temperature electrode or one is not used, the meter automatically switches to Manual Temperature Compensation (MTC) mode and MTC appears on the screen.
MTC can be set as follows:
Power the meter on by pressing
Press and hold
temperature unit blinks (°C or °F).
Press
Press
Then
Continue with MTC temperature setting by using
compensation accordingly
Press
Press
The default MTC temperature value setting is 25°C (77°F). Note: °C = 5/9 (°F - 32)
or to switch between °C and °F.
until the setup icon appears on the display and the current
to confirm your selection.
to confirm the setting
to return to the measurement screen.
.
or to adjust temperature
3.2 Selecting a predefined buffer group
After confirming the MTC compensation temperature value, make the buffer group selection. Use
or to select a buffer group among 3 buffer groups (4.1.1). Press to
confirm the setting or press
The default buffer group is b1.
to leave to return to the measurement screen.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
STARTER 3100 Bench pH Meter EN-9
b1
1.68
4.01
7.00
10.01b22.00
4.01
7.00
9.21
11.00
b3
1.68
4.00
6.86
9.18
12.46
4 STARTER 3100 OPERATION
Standard procedure of pH measurement is as follows:
a) pH electrode preparation b) buffer preparation and pH electrode calibration c) sample preparation d) pH measurement e) Record measurement results or print
f) Rinse the pH electrode and properly store pH electrode preparation: pH electrode should be rinsed with pure water befor e and after using. Check if the electrode is physically damaged. (Be careful with the glass bulb) The pH electrode should be stored in the storage bottle; the solution in the bottle is 3M KCl solution. After placing the pH electrode into the sample or buffer solution, user should stir several seconds then wait 30 to 60 seconds for the signal to stabilize, and then press the button to operate (Calibration or measurement).
WARNING Do not operate the equipment in hazardous environments. The equipment is not explosion protected.
WARNING When using chemicals and solvents, comply with the instructions of the chemical producer and the general lab safety rules.
4.1 Calibration
4.1.1 Buffer group
STARTER 3100 can perform 1-, 2- or 3- point calibrations. There are 3 buffer groups in the meter, you can select the buffer group you prefer (see 4.3), default buffer is b1 US standards; the buffer value will be automatically recognized during calibration. The 3 predefined buffer groups are (at 25°C):
STARTER 3100 automatically corrects for the temperature dependence of the buffer pH values given in the following table - buffer group b1. This means, if the buffer solutions’ temperature is 15°C, the calibration value you will get should be pH1.67 (buffer 1.68), pH4.00 (buffer 4.01), pH7.04(buffer 7.00) and pH10.12 (buffer 10.01).
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
EN-10 STARTER 3100 Bench pH Meter
5 °C
1.67
4.01
7.09
10.25
10 °C
1.67
4.00
7.06
10.18
15 °C
1.67
4.00
7.04
10.12
20 °C
1.68
4.00
7.02
10.06
25 °C
1.68
4.01
7.00
10.01
30 °C
1.68
4.01
6.99
9.97
35 °C
1.69
4.02
6.98
9.93
40 °C
1.69
4.03
6.97
9.89
45 °C
1.70
4.05
6.97
9.86
50 °C
1.71
4.06
6.96
9.83
Calibration: pH electrodes need to be calibrated with pH standard buffer solution before a proper pH measurement can be made. Calibration is to display the right pH value when meter receive the mV value signal from the pH electrode.
Slope: the linear coefficient between mV and pH according to theoretical value (e.g. -
59.16mV/pH @ 25 °C which means 100% slope); Offset: the mV value when pH value is 7.00. (Theoretical value is 0 mV);
4.1.2 Performing 1-point calibration
When performing calibration, Ohaus recommends using Auto End Point Mode. After powering the meter on, be sure the top of the screen shows
Mode.
Auto or Manual End point Mode:
to ensure the meter is in Auto End point
o Press and hold to change the End Point Mode. o When in Manual Mode, to manually reach a pH measurement or calibration value,
you need to press button­then the sample reading or calibration value freezes, freezes on the display.
o When in Auto End Point Mode, the meter determines when the reading is stable
then displays and locks the reading or calibration value automatically, the reading freezes and on the display.
Note: With the 1-point calibration only the offset is adjusted. If the sensor was previously calibrated with multi-point calibration the previously stored slope will remain. Otherwise theoretical 100% slope (-59.16 mV / pH) will be used.
blinks 3 times then disappears; blinks 3 times and freezes
when reading is stable and displays :
blinks 3 times and
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 13
STARTER 3100 Bench pH Meter EN-11
When STARTER 3100 is in pH measurement mode (see 4.2); place the pH electrode in a calibration buffer, stir 5 seconds, wait for 30 seconds, then:
Press
and appear on the top of the screen, is blinking during calibration.
The meter reaches endpoint automatically according to the preselected auto-endpoint
mode, the calibration point pH value (e.g. 7.00) with the temperature display on the screen.
The 1-point calibration is finished; There are now 3 options (OHAUS recommends conducting at least a 2 point calibration);
1. Press
2. Press shown on the display for 3 seconds then return to the measurement screen.
3. Press
Cal 1” displays on the bottom left of the screen and “Cal ” is blinking.
to do the 2-point calibration.
to store the 1-point calibration and exit, the offset and the slope are
to reject the calibration, return to the measurement screen.
4.1.3 Performing 2-point calibration
Perform 1-point calibration as described above. Rinse the pH electrode with pure water and wipe off with tissue.
Place the electrode in the next calibration buffer, stir and wait, then press
displays on the bottom left of the screen and and appear, is blinking during calibration.
The mete r reaches endpoint according to the endpoint mode, the calibration point pH value
(e.g. 4.01) with the temperature display on the screen.
Cal ” is blinking. On the top o f the screen
, Cal 2
The 2-point calibration is finished. There are 3 options:
1. Press
2. Press shown on the display for 3 seconds then return to the measurement screen.
3. Press
Note: To perform a 3 point calibration, follow the instructions for performing a 2 point calibration using a third buffer.
Find Quality Products Online at: sales@GlobalTestSupply.com
to do the 3-point calibration.
to store the 2-point calibration and exit, the offset and slope are
to reject the calibration, return to the measurement screen.
www.GlobalTestSupply.com
Page 14
EN-12 STARTER 3100 Bench pH Meter
4.2 Sample measurement
4.2.1 pH measurement
Place the pH electrode in the sample. Stir for 5 seconds; wait for 30 seconds.
Press
measurement.
When meter reaches endpoint, the pH value with the temperature display on the screen.
to start the pH measurement, appears on the display. is blinking during
4.2.2 mV measurement
 Press to switch between pH measurement mode and mV measurement mode.  Follow the same procedure as for pH measurement to perform mV measurement.
4.3 Temperature measurement
For better accuracy, we recommend to use either a built-in or a separate temperature electrode.
If a temperature electrode is used, ATC and the sample temperature are displayed. If the meter does not detect a temperature electrode, it automatically switches to the manual
temperature compensation mode and MTC appears. MTC temperature should be set.
Note: STARTER 3100 accepts NTC 30 kΩ temperature sensor. ATC (Auto Temperature
Compensation) or MTC (Manual Temperature Compensation), the Temperature Compensation only corrects for the change in the output of the electrode, not for the change in the actual solution. This means the meter corrects the signal(mV) from the electrode to get a more accurate pH value according to the real temperature.
4.4 Using the memory
4.4.1 Storing a reading
The STARTER 3100 can store up to 99 endpoint results.
Press button
has been stored.
If you press further data you will have to clear the memory. (See 4.4.3)
when the measurement reaches endpoint. M01 indicates that one result
when M99 is displayed, FUL displays to indicate the memory is full. To store
4.4.2 Recalling from memory
Press and hold to recall the stored values from memory when the current
measurement reaches endpoint.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 15
STARTER 3100 Bench pH Meter EN-13
Press button or to scroll through the stored results. R01 to R99 indicates which
result is being displayed.
Press
to exit.
4.4.3 Clearing the memory
Pressing or to scroll through the stored results until MRCL appears.
Press
There are now 2 options:
Press
Press
, CLr blinks;
to confirm the deletion of all the stored data.
to return to the measurement mode without deleting the memory.
4.5 Printing
If the printer is connected to the STARTER 3100(e.g. SF-F40A, green light means connection success; printer baud rate 9600bps; 8 data bit; none parity bit; 1 stop bit), a print-out is automatically generated after each end pointed measurement or calibration. The format for the print-out following a pH measurement is: End Point, Value, Temp., ATC/MTC Auto EP, 4.01pH, 25.0 °C, MTC
The details for the second line are: Auto EP, 4.01pH, 25.0 °C , MTC | | | |---- Manual Temperature Compensation | | |---------------- Temperature value and unit ºC | |------------------------- pH value |---------------------------------------- Auto End Point
The format for the print-out following a mV measurement is: End Point, Value, Temp., ATC/MTC Manual EP, 182mV, 23.2 °C, ATC
The print-out in case of an error message is: End Point, Value, Temp., ATC/MTC Error3
The print-out for a 2-point calibration is: Buffer1: 4.01pH mV1: 178mV Temp.1: 25.0 °C Buffer2: 7.00pH mV2: 3mV Temp.2: 25.0 °C
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 16
EN-14 STARTER 3100 Bench pH Meter
Slope: 99% Offset: 5mV ATC/MTC: MTC
The print-out for a 3-point calibration is: Buffer1: 4.01pH mV1: 178mV Temp.1: 25.0 °C Buffer2: 7.00pH mV2: 0mV Temp.2: 25.0 °C Buffer3: 9.21pH mV3: -130mV Temp.3: 25.0 °C Slope: 100% Offset: 0mV ATC/MTC: ATC
Printing from memory: When scrolling through the memory you can print the ent ry that is currently
viewed by pressing and holding (3 seconds) the Recall 01: End Point, Value, Temp., ATC/MTC Manual EP, 4.01pH, 35.6 °C, ATC
. The printout format is as follows:
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 17
STARTER 3100 Bench pH Meter EN-15
Error 0
Memory access error
Reset to factory settings
Error 1
Self-diagnosis failed
Repeat the self-diagnosis procedure and make sure that you finish pressing all five keys within two minutes.
Error 2
Measured values out of range
Check if the electrode is properly connected and placed in the sample solution.
Error 3
Measured buffer temperature out of range (
C)
Keep the pH buffer temperature within the range for calibration
Error 4
Offset out of range offset > 60mV or <
Make sure the pH buffer is correct and fresh; Clean or replace the pH electrode.
Error 5
Slope out of range
Make sure the buffer is correct and fresh; Clean or replace the
Error 6
Meter cannot recognize the buffer
Make sure the buffer is correct and fresh; check if the buffer has not been used more than once.
Error 9
The current data set has already been stored once
An endpoint reading can only be stored once. Perform a new measurement to store.
Error 10
The sample temperature out of range
Check the sample temperature, the temperature sensor.
5 MAINTENANCE
5.1 Error message
<5 or >40 °
- 60 mV
If an error happens, the meter will also beep 3 times to alert. For further technical support please contact Ohaus. (US please contact 1-800-672-7722).
pH electrode.
5.2 Meter maintenance
Never unscrew the two halves of the housing! The STARTER 3100 series instruments do not require any maintenance other than occasional wipe with a damp cloth and the replacement of depleted batteries. The housing is made of acrylonitrile butadiene styrene (ABS). This material is attacked by some organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK). Any spillage should be immediately wiped off.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 18
EN-16 STARTER 3100 Bench pH Meter
b
5.3 Electrode maintenance
Make sure the electrode is filled with electrolyte solution. Always store the electrode according to the electrode instruction manuals and do not allow it to dry out.
If the electrode response becomes sluggish or the slope is not acceptable, try the following:
o Soak the electrode in 0.1M HCl for at least 8 hours. o For fat or oil contaminant, degrease the membrane with cotton wool soaked in either
acetone or a soap solution.
After electrode treatment, a new calibration should be performed. If the electrode slope is still not acceptable, the electrode might need to be replaced.
5.4 Self-diagnosis
When the meter is on, press and hold and simultaneously until the
meter displays the full screen. Each icon blinks one after the other. This way you may check whether all icons are correctly shown. The next step is to check that the keys are functioning correctly. This requires user interaction.
When b
blinks, five icons are displayed.
Press the five keys in any order. Each time you press a key an icon disappears from the
screen, continue to press the other keys until all the icons have disappeared.
When the self-diagnosis has been completed successfully, PAS appears. If self-diagnosis fails, error message Err 1 appears.
Note: You have to finish pressing all five keys within 2 minutes, otherwise Err 1 appears and you will have to repeat the procedure.
5.5 Recover factory settings
When the meter is off, press and hold & & together for 3
seconds, the screen displays
Press
then restart the meter.
Or press
to reset factory settings (MTC, slope and offset, etc.), displa y
to quit the setting, display then turn off the meter.
and blinks, this means “Reset”. There are 2 options:
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 19
STARTER 3100 Bench pH Meter EN-17
Model
Measuring range
-2.00…16.00 pH
-5
Resolution
0.01 pH 1 mV
0.1
Error limits
± 0.01 pH ± 1 mV ± 0.5
Calibration
3 points 3 predefined buffer groups
Memory
99 measurements The last calibration data
Power supply
AC Adapter Input: 100-240V ~ X.XA50/60 Hz AC Adapter Output: 12V
X.XA
Size/weight
220 W x 175 D x 78 H mm / 0.75 kg Display
Liquid crystal with backlight
Input
BNC, impedance > 10e+12 Ω Cinch, NTC 30 kΩ
Reference input
2mm banana socket
Temperature-compensation
ATC & MTC
Housing
ABS
6 TECHNICAL DATA
6.1 Specifications
Ambient conditions
x Indoor use only x Altitude: Up to 2000 m x Specified Temperature range: 5 °C to 40 °C x Humidity: maximum relative humidity 80 % for temperatures up to 30 °C decreasing linearly to
50% relative humidity at 40 °C
x Mains supply voltage fluctuations: up to ±10% of the nominal voltage x Installation category II x Pollution degree: 2 x Operability is assured at ambient temperatures between 5 °C to 40 °C
ST3100
1999…1999 mV
°C…110°C
°C
°C
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 20
EN-18 STARTER 3100 Bench pH Meter
6.2 Compliance
This product conforms to the EMC Directive 2004/108/EC and the Low Voltage Directive 2006/95/EC.
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements. Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related. Thank you for your contribution to environmental protection.
FCC Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
ISO 9001 Registration
In 1994, OHAUS Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the OHAUS quality management system is compliant with the ISO 9001 standards requirements. On June 21, 2012, OHAUS Corporation, USA, was re-registered to the ISO 9001:2008 standard.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 21
STARTER 3100 Bench pH Meter EN-19
7 BUFFER GROUP
STARTER 3100 automatically correct for the temperature dependence of the buffer group pH value given in the following tables (b2, b3), you can find b1. (see 3.1.1)
Buffer group b2 Europe standard Temp °C pH2.00 pH4.01 pH7.00 pH9.21 pH11.00 5 2.02 4.01 7.09 9.45 11.72 10 2.01 4.00 7.06 9.38 11.54 15 2.00 4.00 7.04 9.32 11.36 20 2.00 4.00 7.02 9.26 11.18
25 2.00 4.01 7.00 9.21 11.00
30 1.99 4.01 6.99 9.16 10.82 35 1.99 4.02 6.98 9.11 10.64 40 1.98 4.03 6.97 9.06 10.46 45 1.98 4.04 6.97 9.03 10.28 50 1.98 4.06 6.97 8.99 10.10
Buffer group b3 JJG119 Temp °C pH1.68 pH4.00 pH6.86 pH9.18 pH12.46 5 1.67 4.00 6.95 9.39 13.21 10 1.67 4.00 6.92 9.33 13.01 15 1.67 4.00 6.90 9.28 12.82 20 1.68 4.00 6.88 9.23 12.64
25 1.68 4.00 6.86 9.18 12.46
30 1.68 4.01 6.85 9.14 12.29 35 1.69 4.02 6.84 9.11 12.13 40 1.69 4.03 6.84 9.07 11.98 45 1.70 4.04 6.83 9.04 11.83 50 1.71 4.06 6.83 9.02 11.70
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 22
EN-20 STARTER 3100 Bench pH Meter
Ohaus products are warranted against defects in materials and workmanship from the date of
LIMITED WARRANTY
delivery through the duration of the warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) th at proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus. This warranty does not apply if the product has been damaged b y accident or misuse, exposed to radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shi pment to the authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for any consequential damages.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com