STDO11 EN-1
Model
Item NO
Description
STDO11
30031639
Galvanic DO electrode
Specifications:
Connection: BNC Cable length: 1.1m Shaft length: 120mm Shaft diameter: 12mm Shaft material: Plastic Temperature range: 0-50 Measurement range: 0-200% Storage Solution: 10% NaCl
DO electrode is to measure the Dissolved Oxygen in the water. Galvanic DO electrode
does not need to warm up before measurement. It can reach 95% of the final reading after 1 minute measurement.
If you want to measure the temperature at the same time, you can purchase temperature electrode STTEMP30 (83033970).
www.GlobalTestSupply.com
Page 3
EN-2 STDO11
DO electrode operation:
1. Unpacking
Carefully remove the protective bottle from the tip of the electrode by unscrewing the lid then removing the bottle. Remove the shorting plug from the connector and store in a safe place. Be careful because the protective bottle lid is tightly fit on the electrode.
2. Calibration
Connect the electrode to the instrument, rinse with deionized water and carefully dry the tip of the electrode with a tissue. Then follow your instrument instruction manual for detailed calibration.
After calibration and measurement, you should rinse the DO electrode then storage in the protective bottle.
www.GlobalTestSupply.com
Page 4
STDO11 EN-3
FAQ:
1. DO electrode condition
Electrode condition icon give you an intuitive information about the DO electrode performance.
Slope: 80-125%
Electrode condition is good
Slope: 60-80%
Electrode condition is not good, may need cleaning or replace
2. Cleaning
This galvanic oxygen electrode has no customer serviceable parts. Cleaning is important however, and the electrode should be rinsed thoroughly in pure water between and after measurements. NEVER allow the sample to dry on the electrode.
If dirt or sample do dry onto the membrane, the tip of the sensor can be immersed in pure water and ultrasonically cleaned for 30 seconds.
3. Storage
After rinsing the electrode, refit the storage bottle and shorting plug and store in a cool dry place.