Ohaus SPX621, SPX622, SPX6201, SPX8200, SPX223 Instruction Manual

...
TM
Scout
Instruction Manual
Balanzas Scout
Balances Scout
Mode d’emploi
Scout
Bedienungsanleitung
Series Balances - SPX
TM
Series - SPX
TM
Séries - SPX
TM
Waagen-Serien - SPX
TM
Bilance Serie Scout
Manuale d’Istruzioni
- SPX
EN-2 SCOUTTM SERIES BALANCES - SPX
CAUTION: Read all safety warnings before installing, making connections, or
servicing this equipment. Failure to comply with these warnings could result in personal injury and/or property damage. Retain all instructions for future reference.
SCOUTTM SERIES BALANCES - SPX EN-1
1. INTRODUCTION
This manual contains installation, operation and maintenance instructions for the Scout SPX
Series Balances. Please read the manual completely before using the balance.
1.1 Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results.
Signal Words
WARNING For a hazardous situation with medium risk, possibly resulting in injuries or death if not avoided.
CAUTION For a hazardous situation with low risk, resulting in damage to the device or
the property or in loss of data, or injuries if not avoided. Attention For important information about the product. Note For useful information about the product
Warning Symbols
Attention Symbol Electric Shock Hazard
1.2 Safety Precautions
Verify that the local AC power supply is within the input voltage range printed on the AC
adapter’s data label.
Only connect the AC adapter to a compatible grounded electrical outlet. Do not position the scale such that it is difficult to disconnect the AC adapter from the
power receptacle.
Make sure that the power cord does not pose a potential obstacle or tripping hazard. This equipment is intended for indoor use and should only be operated in dry locations. Operate the equipment only under ambient conditions specified in the user instructions. Do not operate the equipment in hazardous or unstable environments. Do not drop loads on the pan. Only use approved accessories and peripherals. Disconnect power from the equipment before cleaning or servicing.
Service should only be performed by authorized personnel.
2. INSTALLATION
2.1 Installing Components
Refer to the illustrations and instructions below to identify and assemble your Scout balance with its components. All components must be assembled before using the balance.
Release the red Transportation Lock on subplatform of the balance by turning the red pointer 90 counter-clockwise.
Figure 2-1. Transportation Lock
Subplatform
Transportation
Unlocked
Position
Locked
Position
EN-2 SCOUTTM SERIES BALANCES - SPX
2.1.1 Releasing the transportation Lock
2.1.2 Installing the Weighing Pan
Balances with a rectangular pan are placed into the sub-platform as shown and rotated counter-clockwise until it locks. Round pans are placed straight down on sub-platform.
Figure 2-2. Pan Installation
2.1.3 Security Slot
A security slot is provided at the rear of the balance allowing the balance to be secured by an optional cable and lock accessory.
2.2 Selecting the Location
For best performance, the Scout SPX balance should be used in a clean, stable environment. Do not use the balance in environments with excessive drafts, with rapid temperature changes, near magnetic fields or near equipment that generates magnetic fields, or vibrations.
2.3 Leveling the Balance
The Scout has an illuminated level indicator as a reminder that the balance should be leveled for accurate weighing. There is a level bubble in a small round window on the front of the balance. To level the balance, adjust the feet at each corner until the bubble is centered in the circle. Be sure the equipment is level each time its location is changed.
Figure 2-3. Level indicator
AC Adapter Input Jack
Battery Compartment Cover
Interface Connection
Security Switch
Weigh Below Hole
Security slot (Anti-theft lock slot)
SCOUTTM SERIES BALANCES - SPX EN-3
2.4 Connecting Power
AC Adapter Installation
AC power is used to power the scale when battery power is not needed. First, connect the AC Adapter (supplied) to the AC Adapter Input Jack at the rear of the balance then connect the AC plug to an electrical outlet.
Figure 2-4. Rear and bottom view of balance
Battery Installation
Install the four “AA” batteries with polarity as shown in the battery compartment.
Note:
After power on, it is recommended to warm up the balance for at least 5 minutes before using it.
2.5 Initial Calibration
When the Balance is first installed, and when it is moved to another location, it must be calibrated to ensure accurate weighing results. Have the appropriate calibration masses available before beginning calibration. Refer to the Calibration Section for masses and calibration procedure.
Button
Primary Function (Short Press)
Zero/On
Turns the balance on
If balance is On, sets Zero
Print
Sends the current value to the selected COM ports if AUTOPRINT is set to Off.
Function
Initiates an application mode.
Tare
Enter/clear a tare value.
Secondary Function (Long Press)
Zero/Off
Turns the balance Off.
Units
Changes the weighing unit.
Mode
Allows changing the application mode.
Menu
Enter the User menu.
Menu Function (Short Press)
Yes
Accepts the current setting on the display.
No
Advances to the next menu or menu item.
Rejects the current setting on the display and advances to the next available setting.
Back
Moves Back to previous menu item.
Exit
Exits the User menu.
Aborts the calibration in progress.
EN-4 SCOUTTM SERIES BALANCES - SPX
3. OPERATION
3.1 Controls
Figure 3-1. Scout Control Panel
TABLE 3-1. Button Functions
Notes: 1 Short Press: Press less than 1 second.
2
Long Press: Press and hold for more than 2 seconds.
Item
Description
Item
Description
1
Battery charge symbol
8
Kilogram, grain symbols
2
Stable weight symbol
9
Check Weighing symbols
3
Negative symbol
10
Pound, Ounce, Pound:Ounce symbols
4
Center of Zero symbol
11
Percent, dwt, g/mol symbols
5
NET symbol
12
Gravity, Pieces, t hk, t sg, t tw, Hold symbols
6
Preset Tare symbol
13
Newton, Totalization, Dynamic, Bluetooth symbols
7
Pointer symbols
2
3
4
1
5
6 8
9
10
11
12
13
7
SCOUTTM SERIES BALANCES - SPX EN-5
Figure 3-1. Scout Display
TABLE 3-2. Display Symbols*
Note: * Some symbols might not be available depending on different models.
3.2 Turning Balance On/Off
To turn the balance on, press and hold the On/Zero Off button for 1 second. The balance performs a display test, momentarily displays the software version, and then enters the active weighing mode. To turn the balance off, press and hold the On/Zero Off button until OFF is displayed.
3.3 Calibration Operation
When the balance is operated for the first time, a span calibration is recommended to ensure accurate weighing results. Before performing the calibration, be sure to have the appropriate calibration weights. Ensure that the Security switch is set to unlocked position.
Press and hold Menu until [mMeNU] (Menu) is displayed. When the button is released, the display will show [C.A.L]. Press Yes to accept. [SpaN] will then be shown. Press Yes to begin the span
calibration. [--C--] will be displayed while zero reading is stored. Next, the display shows the calibration
weight value (press No to toggle value). Place the specified calibration mass on the pan. [--C--] will be displayed while the reading is stored. The display will show [done] if the calibration was
successful. The balance returns to the previous application mode and is ready for use.
EN-6 SCOUTTM SERIES BALANCES - SPX
3.4 Weighing Mode
This mode is the factory default setting.
1. If needed, press and hold Mode until [wWeIGH] (Weigh) is displayed.
2. If required, place an empty container on the pan and press Tare.
3. Add sample to the pan or container. The display shows the weight of the sample.
3.5 Counting Mode
This mode counts large numbers of items based on the weight of a reference count.
1. Place an empty container on the pan and press Tare.
2. Press and hold Mode until [Count] (Count) is displayed. [CLr.PwW] (Clear Average Piece Weight, APW) will then display.
If no APW exists, the balance will display [Put.10], proceed to step 5.
3. Press No to use the stored APW. Proceed to step 7.
4. Press Yes to establish an APW.
5. The balance will then display the stored sample size, i.e. [Put 10]. Press No or Back to toggle the choices (5, 10, 20, 50 or 100).
6. Put the indicated number of pieces on the pan then press Yes to calculate the APW. The display shows the piece count.
7. Add additional pieces until the desired count is reached.
8. To clear the stored APW press and hold Mode until [Count] is displayed. Press Yes when
[Clr.pwW] is displayed. Note: Press Function to view the current APW.
3.6 Percent Mode
This mode measures the weight of a sample as a percentage of a reference weight.
1. If required place an empty container on the pan and press Tare.
2. Press and hold Mode until [Percnt] is displayed. [Clr.ref] (clear reference) will then display. If no reference weight exists, the balance will display [Put.ref], proceed to step 5.
3. Press No to use the stored reference weight and proceed to step 6.
4. Press Yes to establish a new reference. Balance will now display [Put.ref].
5. Add the desired reference material to the pan or container. Press Yes to store the reference weight. The display shows 100%.
6. Replace the reference material with the sample material. The display shows the percentage of the sample compared to reference weight.
7. To clear the stored reference press and hold Mode until [Percnt] is displayed. Press Yes when [Clr.ref] is displayed.
Note: Press Function to view the current reference weight.
SCOUTTM SERIES BALANCES - SPX EN-7
3.7 Check Mode
Use this mode to compare the Weight to a target weight range. The balance supports positive, negative and zero check weighing.
3.7.1 Check Weighing
Use this mode to compare the weight of items to a target weight range.
1. Press and hold Mode until [CHeCk] (Check) is displayed. [CLr.ref] (clear check limits) will then display.
2. Press No to use the stored check limits and proceed to step 5.
3. Press Yes to establish new check limits. The balance will then display [Set. Lo]. Press Yes to view the “Low” limit value. Press Yes to accept or No to edit the “Low” limit value. The
stored value then displays with the first digit highlighted [000.000 kg]. Repeatedly press No
until the desired number appears. Press Yes to accept and highlight the next digit. Repeat until all the digits are correct. Press Yes to accept the “low” limit value, [Set. Hi] will be
displayed.
4. Repeat the same procedure to accept or edit the “high” value.
5. If required, place an empty container on the pan and press Tare.
6. Place sample material on the pan or in the container. If the sample weight is under the
target weight range, the under icon will light. If the sample is within the target weight range, the accept symbol will light. If the sample
is over the target weight range, over icon will light.
Note: Press Function to view the low and high check limits.
Positive Check
Positive check is used to determine when the material added to the balance is within the target range. In this case the UNDER and OVER limits must be positive values. (The OVER limit must be greater than the UNDER limit.) Add material to the balance until it is within the ACCEPT ( ) range.
Negative Check
Negative check is used to determine when the material removed from the balance is within the target range. In this case the UNDER and OVER limits are both negative values. (The UNDER limit must be greater than the OVER limit.) Place the item to be weighed on the balance and press TARE. Remove a portion of the item until it is within the ACCEPT range.
Zero Check
Zero check is used when comparing subsequent samples to an initial reference sample. In this case, the UNDER limit must be a negative value and the OVER limit must be a positive value. Place the reference item on the balance and press TARE. Remove the reference sample and place the item to be compared on the balance to determine if it is within the ACCEPT range.
EN-8 SCOUTTM SERIES BALANCES - SPX
3.8 Totalization Mode
This mode allows the user to store a series of weight measurements. Totalize mode has been initiated when the symbol is displayed and the current unit is displayed.
Notes: Only positive numbers are totalized.
1. Press and hold Mode until [total] (Totalization) is displayed. [CLr.tot] will then be displayed.
2. Press Yes or No key to clear the current totalized data or not. When a weight is added to the scale the value is displayed.
3. If required, place an empty container on the pan and press Tare. Add the first item, its
weight is displayed. Press Function to store the weight, the “symbol will flash and the display will show the total weight.
4. Press Tare (or remove the weight in previous operation) and add the next item. The
scale will display its weight. Press Function to store its weight. The symbol will flash and the new total weight will be displayed.
5. Repeat step 4 for all of the items to be accumulated.
6. To clear the stored total press and hold Mode until [total] is displayed. When [Clr. tot] is displayed, press Yes.
3.9 Hold Mode
There are two modes for the display hold:
- Peak Hold: allows the user to capture and store the highest stable weight value (>=5d).
- Display Hold (default): allows the user to capture and store the first stable weight value (>=5d).
Start If no weight value is held on the display, press Function key to begin. The [rEAdY] (Ready) will
be displayed until a weight is added on the pan. When the stable value is being held on the display, the Hold icon (H) will blink and the displayed weight will not change.
Reset If the pan is empty and a weight value was held on the display, a single short press of the Function key will clear the held value and show the new weight on the pan.
1. Press and hold Mode until [HoLd] is displayed.
2. If required, place an empty container on the pan and press Tare. Zero value will then display.
3. Press Function key to begin. The [rEAdY] (Ready) will be displayed.
4. Place samples to be weighed on the pan.
5. The stable value will be held on the display, the Hold icon (H) will blink.
Sub-Menus
C.a.l
S.e.t.u.p
M.o.d.e
U.n.i.t
E.n.d
User Menu
Items:
Span Lin
Reset
Yes/No
Filter
Low/Med/High
AZT
0.5/1/3/Off
StableRange
0.5/1/2//5
Backlight
Off/On/Auto
Auto Tare
Off/On/On-acc
Auto Off
Off/1/5/10
Reset
Yes/No
Weighing
On/Off
Counting
On/Off
Percent
On/Off
Check
On/Off
Totalize
On/Off
Hold
Disp/Peak/Off
Reset g kg ct N oz ozt dwt lb lb:oz grn hkt* sgt* twt* t*
tola/tical
C
End
End
End
End
Span: Perform Linearity: Perform End Calibration: Exit menu
No/Back
Long press -
Enter menu
Short press -
Accept
Short press – Go to Next/Prev. menu
Yes
Exit
Short press –
Exit menu
SCOUTTM SERIES BALANCES - SPX EN-9
4. MENU SETTINGS
The User Menu allows the customizing of balance settings. Note: Additional Sub-Menus may be available if Interface Options are installed. See Interface
User Manual for the additional setting information.
4.1 Menu Navigation
User Menu:
Notes:
* Only available in certain regions. When the Security Switch is in locked position, the Calibration function is hidden, Units are
locked to the current setting.
Security switch:
4.2 Calibration Menu
Enter this menu to perform calibrations.
Span [SpaN]
Initiates a span calibration procedure (zero and span)
Lin [LIN] Initiates a linearity calibration procedure (zero, mid-point and span).
End Cal [End]
Advance to the next menu or return to the top of the current menu.
Reset: no, yes Filter: Low, Med, High Auto Zero Tracking: off, 0.5d, 1d, 3d Stable: 0.5d, 1d, 2d, 5d Backlight: off, on, auto Auto Tare: off, on, on-acc Auto Off: off, 1, 5, 10 End Setup: Exit menu
EN-10 SCOUTTM SERIES BALANCES - SPX
4.3 Setup Menu
Enter this menu to set balance parameters.
Note: Bold always represents factory default value Reset [reset]
Reset the Setup menu to factory defaults. NO = not reset YES = reset
Filter [FILtEr] Set the amount of signal filtering. LOW = less stability, faster stabilization time MED = normal stability, stabilization time HI = greater stability, slower stabilization time
AZT [AZt] Set the automatic zero tracking functionality. OFF = disabled
0.5d = the display will maintain zero until a change of 0.5 divisions per second has been exceeded. 1d = the display will maintain zero until a change of 1 divisions per second has been exceeded. 3d = the display will maintain zero until a change of 3 divisions per second has been exceeded.
Stable Range [Stable] Set the amount the reading can vary while the stability symbol remains on.
0.5d = 0.5 balance division 1d = 1 balance division 2d = 2 balance division 5d = 5 balance division
Back Light [LIGHt] Sets backlight functionality. OFF = always off ON = always on AUTO = turns on when a button is pressed or the displayed weight changes. Note: When connected with power pack, the backlight is always on.
Auto Tare [A.tArE] Set the automatic tare functionality. OFF = Automatic Tare is disabled ON = the first stable gross weight is tared ON-ACC = stable gross loads within the accept limits are tared (in Check weighing mode)
Reset: no, yes Weigh: off, on Count: off, on Percent: off, on Check: off, on Totalize: off, on Hold: Disp, Peak, Off End Mode: Exit menu
SCOUTTM SERIES BALANCES - SPX EN-11
Auto off [A.OFF]
Set the automatic shut off functionality. OFF = disabled 1 = powers off after 1 minute of no activity 5 = powers off after 5 minutes of no activity 10 = powers off after 10 minutes of no activity
End Setup [End] Advance to the next menu or return to the top of the current menu.
4.4 Mode Menu
This menu activates modes so they will be available for use with the Mode button.
Reset [reset] Reset the Mode menu to factory defaults. NO = not reset YES = reset
Weigh [wWEIGH] Set the status. OFF = disabled ON = enabled
Count [COUNt] Set the status. OFF = disabled ON = enabled
Percent [PerCNt] Set the status. OFF = disabled ON = enabled
Check [CHECk] Set the sub-mode OFF = disabled ON = enabled
Totalize [total] Set the sub-mode OFF = disabled ON = enabled
g: off, on kg: off, on ct: off, on N: off, on oz: off, on ozt: off, on dwt: off, on lb: off, on lb:oz: off, on grn: off, on hkt: off, on sgt: off, on twt: off, on t: off, tola, tical C: off, on End Unit: Exit menu
EN-12 SCOUTTM SERIES BALANCES - SPX
Hold [Hold]
Set the sub-mode. OFF = disabled Peak Hold = allows the user to capture and store the highest stable weight value (>=5d). Display Hold = allows the user to capture and store the first stable weight value (>=5d).
When the stable value is being held on the display, the “Hold” icon will blink and the displayed
weight will not change.
Advance to the next menu or return to the top of the current menu.
End Mode [End]
4.5 Unit Menu
This menu activates units so they will be accessible with the Units button. The units in the menu must be turned “on” to be active. Note: Available units vary by model and local regulations.
Custom Unit (C)
Use the Custom Unit to display weight in an alternative unit of measure. The custom unit is defined using a conversion factor based on gram unit, where the conversion factor is the number of custom units per gram expressed in scientific notation (Factor x 10^Exponent).
Factor
Set the conversion factor (0.1 to 1.99) using the numeric keypad. The default setting is 1.0.
SCOUTTM SERIES BALANCES - SPX EN-13
Exponent
Set the factor multiplier.
-3 = divide the Factor by 1000 (1x10-3)
-2 = divide the Factor by 100 (1x10-2)
-1 = divide the Factor by 10 (1x10-1) 0 = multiply the Factor by 1 (1x100) 1 = multiply the Factor by 10 (1x101) 2 = multiply the Factor by 100 (1x102) 3 = multiply the Factor by 1000 (1x103)
Least Significant Digit
Set the graduation. Settings of 0.5, 1, 2, 5, 10, 100 are available. Note: Custom Unit is locked at Off position when the Security Switch is set to the locked
position.
Follow below instructions below to enter the unit menu to set the Custom Unit. Press and hold Menu until [mMeNU] is displayed. When the button is released, the display will show [C.A.L]. Press No through the succeeding menu items: [C.A.L], [SEtUP], [mModE] until [Unit] is
displayed.
1. When [Unit] is displayed, press Yes.
2. [rESEt] will be displayed. Press No, [Unit g] will be displayed.
3. Press No to toggle through active units until [Unit c] is displayed.
4. Press Yes, [On c] will be displayed, then press Yes.
5. [Factor] is displayed. Press Yes to view the default Factor. Press Yes to accept or No to edit the “factor multiplier”. The stored value is then displayed with the first digit highlighted
[1.0000000]. Repeatedly press No until the desired number appears. Press Yes to accept and highlight the next digit. Repeat until all the digits are correct. Press Yes to accept the “Factor value.
6. [EPt] is displayed. Repeat the same procedure as in step 5 to accept or edit the Exponent value. Press Yes to accept the Exponent value.
7. [LSd] is displayed. Repeat the same procedure as in step 5 to accept or edit the Least Significant Digit value. Press the Yes key to accept the graduation value, [End] is displayed.
8. When [End] is displayed, press Exit to exit to weighing mode.
Use an optional interface connectivity kit to connect the balance either to a computer or a printer.
Below Interface kit accessories are available: RS232, USB Host, USB Device, Ethernet,
Bluetooth®*.
Figure 4-4. Rear of the balance
* Interface kits may vary according to local regulations
The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by OHAUS is under license.
Interface connection
AC Adapter jack
Security slot
Weigh Below Hook
Weigh Below Hook installed
Weigh below opening
EN-14 SCOUTTM SERIES BALANCES - SPX
4.6 Additional Features
Weigh Below Hook
The Scout Balance is equipped with a weigh below hook for weighing below the balance. The weigh below hook is located at the reverse side of the battery cover as shown below. To use this feature, remove the red protective cover underneath for the weigh below opening. Attention: Before turning the balance over, remove the Pan and Pan Support (if present), and turn the transportation lock to “locked” position to prevent damage.
Figure 4-2. Weigh below
The balance can be supported using lab jacks or any other convenient method. Ensure the balance is level and secure and that the transportation lock has been released. Power on the balance, then use a string or wire to attach items to be weighed.
Figure 4-3. Weigh below application
Connecting the Interface
WARNING: Electric Shock Hazard. Disconnect the equipment from the power
supply before cleaning. Electric Shock Hazard.
Symptom
Possible Cause
Cannot turn on
No power to balance
Poor accuracy
Improper calibration; Unstable environment
Cannot calibrate
Security switch in locked position
Err 8.1
Weight exceeds power on zero range.
Err 8.2
Weight below power on zero range.
Err 8.3
Over load (weight exceeds rated capacity)
Err 8.4
Under load (pan removed)
Err 8.5
Tare out of range
Err 8.6
Displayed value exceeds 999999 (possible in Totalization mode)
ref.err
Parts counting or percentage error – sample weight <1d. Balance shows error then exits parts counting or goes to [Clr.APU].
Lo.ref
Percent Reference weight or APW is too low for accurate results
CAL E
Fail to do calibration.
USB.Err
Cannot find menu or app file in U-disk.
DESCRIPTION
Item Number
DESCRIPTION
Item Number
RS232 kit
30268982
Stacking Kit, x1
30268988
USB Host kit
30268983
Specific Gravity kit
30269020
USB Device Kit
30268984
Auxiliary Display Kit
30269019
Bluetooth Kit*
30268985
Carrying Case
30269021
Ethernet Kit
30268986
In-Use Cover
30269022
Stacking Kit, x6
30268987
Printers and Cables
Contact OHAUS
SCOUTTM SERIES BALANCES - SPX EN-15
5. MAINTENANCE
5.1 Cleaning
The housing may be cleaned with a cloth dampened with a mild detergent if necessary. Attention: Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the housing
or control panel.
5.2 Troubleshooting
The following table lists common problems and possible causes and remedies. If the problem persists, contact OHAUS or your authorized dealer.
TABLE 5-1
5.3 Service Information
If the troubleshooting section does not resolve or describe your problem, contact your authorized OHAUS service agent. For service assistance or technical support in the United States call toll-free 1-800-672-7722 ext. 7852 between 8:00 AM and 5:00 PM EST. An OHAUS product service specialist will be available to provide assistance. Outside the USA, please visit
our web site, www.ohaus.com to locate the OHAUS office nearest you.
5.4 Accessories
TABLE 5-2. ACCESSORIES
Note: * Bluetooth kit is only available in certain regions according to the local regulations.
Model
SPX123
SPX223
SPX222
SPX422
SPX622
SPX1202
SPX2202
Capacity (g)
120
220
220
420
620
1200
2200
Readability (g)
0.001
0.001
0.01
0.01
0.01
0.01
0.01
Repeatability (Std. Dev.) (g)
0.002
0.002
0.01
0.01
0.01
0.02
0.02
Linearity (g)
0.003
0.003
0.01
0.01
0.02
0.03
0.03
Span Calibration Mass*
100 g
200 g
200 g
200 g
300 g
1 kg
2 kg
Linearity Calibration Mass
50, 100 g
100, 200 g
100, 200 g
200, 400 g
300, 600 g
500 g,
1 kg
1 kg, 2 kg
Stabilization Time (s)
1.5 1 1.5
Construction
ABS plastic housing with 304 stainless steel (SST) pan
Draftshield
Yes
No
Calibration
User-selectable external span or linearity calibration / Digital with external weight
Tare Range
Full Capacity by subtraction
Weighing Units**
g, kg, ct, N, oz, ozt, dwt, lb, lb:oz, grn,Tael (HongKong),Tael (Singapore),Tael (Taiwan), tical,
tola, C (Custom unit)
Application Modes
Weighing, Parts Counting, Percent Weighing, Check Weighing, Totalization, Display Hold
Typical Battery Life
80 hours
120 hours
80 hours
Specified Temperature Range
10° C (50°F) to 40° C (104° F) at 10% to 80% relative humidity, non-condensing
Storage Conditions
-20° C (-4°F) to 55° C (131° F) at 10% to 90% relative humidity, non-condensing
Communication
RS232, USB Host, USB Device, Ethernet or Bluetooth*** (all available as accessory)
Display Type
Backlit LCD: 6-digit 7-segment with white LED backlight
Display Size
0.78 in / 20 mm digits
Pan Size (W x D)
Ø 93 mm / 3.7 in
Ø 120 mm / 4.7 in
170 x 140 mm /
6.7 x 5.5 in
Balance Dimensions (W x D x H)
202 x 222 x 103 mm /
8 x 8.7 x 4.1 in
202 x 224 x 54 mm / 8 x 8.8 x 2.1 in
Shipping Dimensions (W x D x H)
300 x 250 x 129mm /
11.8 x 9.8 x 5.1 in
300 x 250 x 86 mm / 11.8 x 9.8 x 3.4 in
Net Weight
1 kg / 2.2 lb
Shipping Weight
1.5 kg / 3.3 lb
EN-16 SCOUTTM SERIES BALANCES - SPX
6. TECHNICAL DATA
The technical data is valid under the following ambient conditions: Indoor use only Operating temperature range: +5°C to +40°C Relative humidity: 10% to 80% at 31°C, decreasing linearly to 50% at 40°C, non-condensing Altitude: Up to 2000 m Power: AC power adaptor input 100-240V 50/60 Hz and output 5 V DC 1 A, or 4 AA batteries Pollution degree: 2 Installation category: II Main supply voltage fluctuations: up to ± 10% of the nominal voltage
6.1 Specifications
TABLE 6-1. SPECIFICATIONS
Notes:
* Calibration weights are included with models up to 620g capacity. ** Availability is dependent on model and region. *** Bluetooth kit is only available in certain regions according to the local regulations
Model
SPX123
SPX223
SPX222
SPX422
SPX622
SPX1202
SPX2202
Gram (g)
120 x 0.001
220 x 0.001
220 x 0.01
420 x 0.01
620 x 0.01
1200 x 0.01
2200 x 0.01
Kilogram (kg)
/ / / / /
1.2 x 0.00001
2.2 x 0.00001
Carat (ct)
600 x 0.005
1100 x 0.005
1100 x 0.05
2100 x 0.05
3100 x 0.05
6000 x 0.05
11000 x 0.05
Newton (N)
1.17679 x
0.00001
2.15744 x
0.00001
2.1574 x
0.0001
4.1188 x
0.0001
6.0801 x
0.0001
11.7679 x
0.0001
21.5744 x
0.0001
Ounce (oz)
4.23290 x
0.00005
7.76030 x
0.00005
7.7600 x
0.0005
14.8150 x
0.0005
21.8700 x
0.0005
42.3290 x
0.0005
77.6030 x
0.0005
Ounce Troy (ozt)
3.85810 x
0.00005
7.07320 x
0.00005
7.0730 x
0.0005
13.5030 x
0.0005
19.9335 x
0.0005
38.5810 x
0.0005
70.7320 x
0.0005
Pennyweight (dwt)
77.162 x
0.001
141.463 x
0.001
141.46 x 0.01
270.07 x 0.01
398.67 x 0.01
771.62 x 0.01
1414.63 x
0.01
Pound (lb) / / / /
1.36690 x
0.00005
2.64555 x
0.00005
4.85020 x
0.00005
Pound:Ounce (lb:oz)
/ / /
/
1lb:5.8700oz
x 0.0005oz
2lb:10.3290oz
x 0.0005oz
4lb:13.6030oz
x 0.0005oz
Grain (grn)
1851.88 x
0.02
3395.12 x
0.02
3395.0 x
0.2
6481.6 x
0.2
9568.0 x
0.2
18518.8 x 0.2
33951.2 x 0.2
Model
SPX421
SPX621
SPX2201
SPX6201
SPX8200
Capacity (g)
420
620
2200
6200
8200
Readability (g)
0.1
0.1
0.1
0.1
1
Repeatability (Std. Dev.) (g)
0.1
0.1
0.1
0.1
1
Linearity (g)
0.1
0.1
0.1
0.2
1
Span Calibration Mass*
200 g
300 g
2 kg
5 kg
8 kg
Linearity Calibration Mass
200, 400 g
300, 600 g
1 kg, 2 kg
3 kg, 6 kg
4 kg, 8 kg
Stabilization Time (s)
1
Construction
ABS plastic housing with 304 stainless steel (SST) pan
Draftshield
No
Calibration
User-selectable external span or linearity calibration / Digital with external weight
Tare Range
Full Capacity by subtraction
Weighing Units**
g, kg, ct, N, oz, ozt, dwt, lb, lb:oz, grn,Tael (HongKong),Tael (Singapore),Tael (Taiwan),
tical, tola, C (Custom unit)
Application Modes
Weighing, Parts Counting, Percent Weighing, Check Weighing, Totalization, Display Hold
Typical Battery Life
120 hours
80 hours
120 hours
Specified Temperature Range
10° C (50°F) to 40° C (104° F) at 10% to 80% relative humidity, non-condensing
Storage Conditions
-20° C (-4°F) to 55° C (131° F) at 10% to 90% relative humidity, non-condensing
Communication
RS232, USB Host, USB Device, Ethernet or Bluetooth*** (all available as accessory)
Display Type
Backlit LCD: 6-digit 7-segment with white LED backlight
Display Size
0.78 in / 20 mm digits
Pan Size (W x D)
Ø 120 mm /
4.7 in
170 x 140 mm / 6.7 x 5.5 in
Balance Dimensions (W x D x H)
202 x 222 x 103 mm /
8 x 8.7 x 4.1 in
202 x 224 x 54 mm / 8 x 8.8 x 2.1 in
Shipping Dimensions (W x D x H)
300 x 250 x 129mm /
11.8 x 9.8 x 5.1 in
300 x 250 x 86 mm / 11.8 x 9.8 x 3.4 in
Net Weight
1 kg / 2.2 lb
Shipping Weight
1.5 kg / 3.3 lb
SCOUTTM SERIES BALANCES - SPX EN-17
TABLE 6-2. SPECIFICATIONS cont.
Capacity x Readability:
TABLE 6-3. SPECIFICATIONS cont.
Notes: * Calibration weights are included with models up to 620g capacity. ** Availability is dependent on model and region. *** Bluetooth kit is only available in certain regions according to the local regulations
Capacity x Readability:
Model
SPX421
SPX621
SPX2201
SPX6201
SPX8200
Gram (g)
420 x 0.1
620 x 0.1
2200 x 0.1
6200 x 0.1
8200 x 1
Kilogram (kg)
/
/
2.2 x 0.0001
6.2 x 0.0001
8.2 x 0.001
Carat (ct)
2100 x 0.5
3100 x 0.5
11000 x 0.5
31000 x 0.5
41000 x 5
Newton (N)
4.119 x 0.001
6.080 x 0.001
21.574 x 0.001
60.801 x 0.001
80.41 x 0.01
Ounce (oz)
14.815 x 0.005
21.870 x 0.005
77.600 x 0.005
218.700 x 0.005
289.25 x 0.05
Ounce Troy (ozt)
13.500 x 0.005
19.930 x 0.005
70.730 x 0.005
199.335 x 0.005
263.60 x 0.05
Pennyweight (dwt)
270.1 x 0.1
398.7 x 0.1
1414.6 x 0.1
3986.7 x 0.1
5270 x 1
Pound (lb)
/
1.3670 x 0.0005
4.8500 x 0.0005
13.6685 x
0.0005
18.080 x 0.005
Pound:Ounce (lb:oz)
/
1lb:5.870oz x
0.005oz
4lb:13.600oz x
0.005oz
13lb:10.700oz x
0.005oz
18lb:1.25oz x
0.05oz
Grain (grn)
6480 x 2
9570 x 2
33950 x 2
95680 x 2
126540 x 20
Model A B C
with draftshield
202 mm / 8.0 in.
222 mm / 8.7 in.
103 mm / 4.1 in.
w/o draftshield
202 mm / 8.0 in.
224 mm / 8.8 in.
54 mm / 2.1 in.
Mark
Standard
This product complies with the applicable harmonized standards of EU Directives 2011/65/EU (RoHS), 2014/30/EU (EMC), 2014/35/EU (LVD) and 2014/31/EU (NAWI). The EU Declaration of Conformity is available online at www.ohaus.com/ce.
EN 61326-1
CAN/CSA-C22.2 No. 61010-1 UL Std. No. 61010-1
EN-18 SCOUTTM SERIES BALANCES - SPX
TABLE 6-4. SPECIFICATIONS cont.
6.2 Drawings
Figure 6.1 Dimensions
6.3 Compliance
Compliance to the following standards is indicated by the corresponding mark on the product.
Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related. Disposal instructions in Europe are available online at www.ohaus.com/weee. Thank you for your contribution to environmental protection.
SCOUTTM SERIES BALANCES - SPX EN-19
Important notice for verified weighing instruments in the EU
When the instrument is used in trade or a legally controlled application it must be set up, verified and sealed in accordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that all pertinent legal requirements are met. Weighing Instruments verified at the place of manufacture bear the following supplementary metrology marking on the descriptive plate.
Weighing Instruments to be verified in two stages have no supplementary metrology marking on the descriptive plate. The second stage of conformity assessment must be carried out by the applicable weights and measures authorities. If national regulations limit the validity period of the verification, the user of the weighing instrument must strictly observe the re-verification period and inform the weights and measures authorities As verification requirements vary by jurisdiction, the purchaser should contact their local weights and measures office if they are not familiar with the requirements.
FCC Note
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Industry Canada Note
This Class B digital apparatus complies with Canadian ICES-003.
EN-20 SCOUTTM SERIES BALANCES - SPX
ISO 9001 Registration
In 1994, OHAUS Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the OHAUS quality management system is compliant with the ISO 9001 standard’s requirements. On June 21, 2012, OHAUS Corporation, USA, was re-registered to the ISO 9001:2008 standard.
LIMITED WARRANTY
OHAUS products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period OHAUS will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to OHAUS. This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than OHAUS. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied warranty is given by OHAUS Corporation. OHAUS Corporation shall not be liable for any consequential damages.
As warranty legislation differs from state to state and country to country, please contact OHAUS or your local OHAUS dealer for further details.
Balances ScoutTM Séries – SPX ES 1
PRECAUCIÓN : Lea atentamente todas las advertencias en material de
seguridad antes de proceder a la instalación, de llevar a cabo conexiones u operaciones de mantenimiento en estos equipos. Del incumplimiento del contenido de estas advertencias podrían resultar daños personales y/o materiales. Conserve estas instrucciones para poder consultarlas en el futuro.
1. INTRODUCCIÓN
Este manual contiene las instrucciones para la instalación, el funcionamiento y el mantenimiento de las Balanzas de la gama Scout SPX Series. Por favor, lea este manual en su
integridad antes de proceder a utilizar la balanza.
1.1 Definición de los símbolos y señales de advertencia
Las notas en materia de seguridad se hallan señalizadas con términos y símbolos de advertencia. Los mismos pueden transmitir informaciones sobre cuestiones y alertas de seguridad. Señalan cuestiones y advertencias de seguridad. Ignorar las indicaciones en materia de seguridad puede ser causa de daños tanto personales como al dispositivo, averías y lecturas erróneas.
Términos de advertencia
ALERTA Indica una situación de un nivel de riesgo o peligro medio, de la que podrían resultar daños o incluso la muerte en el caso de no evitarse. PRECAUCIÓN Indica una situación de un nivel de riesgo o peligro bajo, de la que podrían resultar daños materiales, daños al dispositivo o las pérdida de datos, o daños físicos, en el caso de no evitarse. Atención Para informaciones de relevancia sobre el producto. Nota Para informaciones de utilidad sobre el producto.
Símbolos de advertencia
Riesgos general Riesgos de descarga eléctrica
1.2 Precauciones de seguridad
Compruebe que la alimentación eléctrica de corriente AC está en el rango del voltaje de
entrada impreso en la etiqueta del adaptador.
Conecte el adaptador AC únicamente a un enchufe compatible que disponga de toma de
tierra.
No colocar la balanza de tal modo que resulte difícil de desconectar el adaptador AC de la
toma de corriente.
Asegúrese de que el cable de corriente no represente un obstáculo potencial o pueda
provocar caídas y/o tropiezos.
Este equipo ha sido concebido y fabricado para ser utilizado en espacios interiores, y
debería hacerse uso del mismo en emplazamientos secos.
Haga funcionar el equipo únicamente en las condiciones ambientales señalas las
instrucciones del usuario.
No hacer funcionar el equipo en entornos peligrosos o inestables. No dejar caer objetos o cargas sobre la bandeja. Hacer uso únicamente de los accesorios y periféricos aprobados por el fabricante. Desconectar el equipo de la red eléctrica antes de proceder a su limpieza o efectuar
operaciones de mantenimiento.
Las operaciones de mantenimiento deben de ser únicamente llevadas a cabo por personal
autorizado.
ES 2 Balances ScoutTM Séries – SPX
Desbloquee el cierre / bloqueo de transporte girando el puntero rojo 90 grados en el sentido contrario a las agujas del reloj (de DchaaIzda) para desbloquearlo.
Gráfico 2-1. Bloqueo de transporte
Sub-platforma
Indicador del
bloqueo de
transporte
Posición de
bloqueo
Posición de desbloqueo
2. INSTALACIÓN
2.1 Instalación de componentes
Consulte las instrucciones y los gráficos que aparecen a continuación para identificar y efectuar el montaje de los componentes de su balanza Scout. Todas los elementos y componentes deben de ser montados antes de hacer uso de la balanza.
2.1.1 Desbloqueo del cierre de seguridad de transporte
2.1.2 Instalar la bandeja de pesaje
Las balanzas provistas de una bandeja rectangular se deben colocar encajándolas en la sub­plataforma como se muestra a continuación y girándolas en el sentido contrario a las agujas del reloj (de Dcha a Izda) hasta que se produzca el encaje. Las bandejas redondas se colocan directamente sobre el soporte de la bandeja.
Gráfico 2-2. Instalación de la bandeja de pesaje
2.1.3 Ranura de seguridad
Existe una ranura de seguridad en la parte trasera de la balanza que permite el aseguramiento de la balanza mediante un cable opcional y un accesorio de cierre de bloqueo.
2.2 Seleccionar el emplazamiento idóneo
Para mejorar sus condiciones de funcionamiento, la balanza Scout debería ser colocada en un entorno limpio y estable. No hacer uso de la balanza en entornos con excesivas corrientes de aire, sometidos a cambios bruscos de temperatura, a proximidad de campos magnéticos o cerca de equipos que generen vibraciones o campos magnéticos.
2.3 Equilibrado de la balanza
La balanza Scout dispone de un indicador iluminado que sirve para recordar que la balanza debe hallarse equilibrada para un pesaje preciso. Hay una burbuja de equilibrado en una pequeña ventana en la parte frontal de la balanza. Para equilibrar la balanza, ajustar los soportes en cada una de las esquinas hasta que la burbuja quede centrada en el círculo. Asegúrese de que el equipo sea equilibrado cada vez que se cambia el equipo de emplazamiento.
Conexión del jack del adaptador AC
Cubierta del compartimento de las
pilas
Conexión de Interfaz
Interruptor de seguridad
Orificio de pesaje inferior
Ranura de seguridad (cierre de
bloqueo anti-robo)
Balances ScoutTM Séries – SPX ES 3
Gráfico 2-3. Indicador de equilibrio
2.4 Conexión a la corriente eléctrica
Instalación del adaptador AC
Se puede utilizar la corriente AC para hacer funcionar la balanza cuando no resulte necesario la energía de las pilas / baterías. Primero, conectar el adaptador AC (suministrado con el equipo) al jack de la conexión de entrada del adaptador AC situado en la partes posterior de la balanza y luego conecte el enchufe AC a una toma de corriente eléctrica.
Figure 2-4. Partes posterior e inferior de la balanza
Instalación de pilas / baterías
Instalar cuatro pilas “AA” con la polaridad tal y como se señala en el comportamiento de las pilas.
Nota: Después de la alimentación, se recomienda dejar que el equilibrio caliente durante al menos 5 minutos antes de usarla.
2.5 Calibración inicial
Cuando se instala la balanza por primera vez, y cuando el equipo es desplazado a otro emplazamiento, debe de procederse a su calibrado para garantizar la precisión de los resultados de pesaje. De un modo alternativo, las balanzas pueden ser calibradas manualmente mediante el uso de masas externas. Debe de disponer de las masas de calibrado adecuadas antes de iniciar el proceso de calibrado. Por favor, consulte el apartado de calibrado en lo referente a las masas y al proceso de calibrado.
ES 4 Balances ScoutTM Séries – SPX
Botón
Función primaria (Pulsación corta)
Cero/On -
encendido
Enciende la balanza
Si la balanza está encendida, la configura a cero
Imprimir
Envía el valor actual al Puerto de comunicación seleccionado si el modo AUTOPRINT (impresión automática) está en Off .apagado.
Función
Inicia un modo de aplicación.
Tara
Introducir / borrar un valor de tara. Función secundaria
(Pulsación larga)
Zero/Off -
apagado
Apaga la balanza.
Unidades
Modifica la unidad de pesaje.
Modo
Permite la modificación del modo de aplicación.
Menú
Acceder al menú del usuario.
Función Menú (Pulsación corta)
Yes - Sí
Acepta la configuración actual que aparece en pantalla.
No
Permite avanzar hasta el siguiente menú o el siguiente elemento del menú.
Anula la configuración que aparece en la pantalla y avanza hasta la siguiente configuración disponible.
Atrás
Permite volver hacia atrás al elemento anterior del menú.
Salir
Permite salir del Menú del usuario.
Interrumpe el proceso de calibrado en curso.
3. FUNCIONAMIENTO
3.1 Controles
Gráfico 3-1.Panel de Control Scout
TABLA 3-1. Botones de funciones
Notas:
2
1
Pulsación corta : Pulsar durante menos de 1 segundo. Pulsación larga : Pulsar y mantener pulsado durante más de 2 segundos.
Elemento
Descripción
Elemento
Descripción
1
Símbolo de carga de la batería
8
Símbolos de Kilogramos, grano 2
Símbolo de peso estable
9
Símbolos de verificación de pesaje
3
Símbolo “Negativo”
10
Libras, Onzas, Símbolos de Libras:Onzas
4
Símbolo Cero centrado
11
Símbolos de porcentaje, dwt, g/mol
5
Símbolo NETO
12
Símbolos G (Gravedad), Piezas, t hk, t sg, t tw, H (Mantener en pantalla)
6
Símbolo Tara configurada
13
Símbolos Newtons, Totalización, Dinámico, Bluetooth
7
Símbolos del cursor
2
3
4
1
5
6 8
9
10
11
12
13
7
Balances ScoutTM Séries – SPX ES 5
Figure 3-2. Pantalla de la Balanza Scout
TABLE 3-2.Símbolos mostrados en pantalla
Nota : * Algunos símbolos pueden no hallarse disponibles en función de los distintos modelos.
3.2 Encendido / apagado de la balanza
Encender la balanza, pulsar y mantener pulsado el botón [On/Zero Off]– Encendido / Cero – Apagado durante 1 segundo. La balanza efectuará una comprobación automática de la pantalla y mostrará momentáneamente la versión del software, y luego iniciará el modo de pesaje activo. Para apagar la balanza, pulsar y mantener pulsado el botón [On/Zero Off]- Encendido / Cero – Apagado hasta que se muestra OFF (apagado) en pantalla.
3.3 Operación de calibración
Cuando se pone en marcha la balanza primera vez, se recomienda llevar a cabo una calibración de rango para garantizar la precisión de los resultados de pesaje. Antes de efectuar el calibrado, asegúrese de disponer de los pesos de calibrado adecuados. Asegúrese de que el interruptor de seguridad está en la posición de desbloqueo.
Pulsar y mantener pulsado Menu hasta que [mMeNU] (Menú) se muestre en pantalla. Cuando se suelte el botón, la pantalla mostrará el mensaje [C.A.L]. Pulsar Yes - sí para aceptar. Entonces se mostrará el mensaje [SpaN] en pantalla. Pulsar Yes - sí para iniciar el calibrado de rango. Se mostrará el mensaje [--C--] mientras que se almacena la lectura del “cero”. Seguidamente,
se mostrará en pantalla el valor del peso de calibrado. Pulsar la tecla [No] para modificar dicho valor. Colocar la masa de calibrado especificada en la bandeja. Se mostrará el mensaje [--C--] mientras que se almacena la lectura en la memoria. Se mostrará el mensaje [done] si el
calibrado se ha llevado a cabo con éxito. La balanza volverá al modo de la aplicación anterior y se encuentra preparada para ser utilizada.
ES 6 Balances ScoutTM Séries – SPX
3.4 Modo de pesaje
Este es el modo de la configuración por defecto de salida de fábrica.
1. Si fuese necesario, pulsar y mantener pulsado el botón [Mode] hasta que [wWeIGH] (Weigh – peso) aparezca en pantalla.
2. Si fuese necesario, coloque un recipiente vacío sobre la bandeja y pulse Tare - Tara.
3. Coloque una muestra sobre la bandeja o en el recipiente. El peso de la muestra se mostrará en la pantalla.
3.5 Modo de recuento
Este modo permite contar grandes cantidad y cifras de elementos sobre la base de un peso de referencia para el recuento.
1. Si fuese necesario, coloque un recipiente vacío sobre la bandeja y pulse Tare - Tara.
2. Pulsar y mantener pulsado el botón [Mode] hasta que [Count] (Recuento) aparezca en pantalla. Entonces aparecerá en pantalla : [CLr.PwW] (Borrar el peso medio del elemento:
AveragePieceWeight, APW). En el caso de no existir un peso medio (APW) del elemento a pesar, la balanza mostrará
el mensaje [PUt.10], proceder al paso 5.
3. Pulsar No para hacer uso del APW (peso medio del elemento a pesar) almacenado en la memoria. Proceder al paso 7.
4. Pulsar Yes – sí para configurar un APW (peso medio del elemento a pesar).
5. Entonces la balanza mostrará en pantalla el tamaño de la muestra, es decir [PUt 10]. Pulsar No o Back – vuelta atrás para modificar la selección efectuada (5, 10, 20, 50 o
100).
6. Introducir el número de elementos indicado sobre la bandeja y luego Yes - sí para calcularel APW (peso medio del elemento a pesar). La pantalla mostrará el recuento de elementos.
7. Añadir elementos adicionales hasta alcanzar la cifra deseada.
8. Para borrar de la memoria el APW (peso medio del elemento a pesar) almacenado, pulsar y mantener pulsado el botón [Mode] hasta que [Count] aparezca en pantalla. Pulsar Yes –
cuando [Clr.pwW] aparezca en pantalla. Nota : Pulsar Function (función) para ver el APW (peso medio del elemento a pesar)
actualmente almacenado en la memoria.
3.6 Modo Porcentaje
Este modo permite medir el peso de un elementode muestra como un porcentaje de un peso de referencia.
1. Si fuese necesario, coloque un recipiente vacío sobre la bandeja y pulse Tare - Tara.
2. Pulsar y mantener pulsado el botón [Mode] hasta que [Percnt] aparezca en pantalla. Entonces aparecerá en pantalla [Clr.ref] (borrar referencia).En el caso de no
existir
un peso de referencia, la balanza mostrará el mensaje [Put.ref], proceder al paso 5.
3. Pulsar No para hacer uso del peso de referencia almacenado en la memoria y proceder al paso 6.
4. Pulsar Yes - sí para configurar un nuevo peso de referencia. Entonces aparecerá en pantalla en mensaje [Put.ref].
5. Coloque el material de referencia deseado sobre la bandeja o en el interior del recipiente Pulsar Yes– sí para almacenar en la memoria el peso de referencia. La pantalla mostrará
:100%.
6. Remplazar el material de referencia con el material de muestra. La pantalla mostrará el
porcentaje del peso de muestra en comparación con el peso de referencia.
Balances ScoutTM Séries – SPX ES 7
7. Para borrar de la memoria el peso de referencia almacenado en la memoria, pulsar y
mantener pulsado [Mode] hasta que [Percnt] aparezca en pantalla. Pulsar Yes – sí
cuando [Clr.ref] aparezca en pantalla. Nota : Pulsar Function (función) para ver el peso de referencia actualmente almacenado en la memoria.
3.7 Modo de Comprobación
Se utiliza este modo para comparar el Peso con un rango de peso a alcanzar. La balanza soporta un pesaje de comprobación positivo, negativo y cero.
3.7.1 Pesaje de comprobación
Utilizar este modo para comparar el Peso de los elementos con el rango de peso a alcanzar.
1. Pulsar y mantener pulsado el botón [CHeCk] (Check - comprobar) aparezca en pantalla.
Entonces aparecerá en pantalla [CLr.ref] (borrar límite de comprobación).
2. Pulsar Nopara hacer uso del límite de comprobación almacenado en la memoria y proceder
al paso 5.
3. Pulsar Yes - sí para configurar un nuevo límite de comprobación de referencia. Entonces aparecerá en pantalla en mensaje[Set. Lo]. PulsarYes - sí para ver el valor del límite
“inferior”. PulsarYes - sí para aceptar oNopara editar el valor del límite inferior”. Entonces se mostrará en pantalla el valor almacenado en la memoria, con el primer dígito señalado
[000.000kg]. Pulsar No repetidamente hasta que aparezca la cifra deseada. Pulsar Yes- para aceptar y señalar el siguiente dígito. Repetir hasta que todos los dígitos sean correctos. PulsarYes- sí para aceptar el valor del límite “inferior, aparecerá en pantalla el
mensaje :[Set. Hi].
4. Repetir el mismo procedimiento para aceptar o editar el valor del límite superior”.
5. Si fuese necesario, coloque un recipiente vacío sobre la bandeja y pulse Tare - Tara.
6. Coloque el material de muestra sobre la bandeja o en el recipiente. Si el peso de la muestra está por debajo del rango del peso establecido, se encenderá el icono “inferior”, .
Si la muestra está dentro del rango del peso definido, se encenderá el icono de “aceptado”
. Si el peso de la muestra está por encima del rango del peso establecido, se encenderá el icono .
Nota : Pulsar Function (función) para ver los límites de comprobación inferior y superior.
Comprobación Positiva
La comprobación positiva se utiliza para determinar si el material añadido colocado sobre la balanza se halla dentro del margen del peso establecido. En este caso los límites INFERIORESy SUPERIORES deben de ser valores positivos. (El límite SUPERIOR debe de ser mayor que el límite INFERIOR.) Añadir material adicional a la bascule hasta que alcance el rango ACEPTADO ( ).
Comprobación negativa
La comprobación negativa se utiliza para determinar si el material sustraído del removed añadido colocado sobre la balanza se halla dentro del margen del peso establecido. En este caso los límites INFERIORES y SUPERIORES son en ambos casis valores negativos. (El límite INFERIOR debe de ser mayor que el límiteSUPERIOR.) Colocar el elemento que debe de ser pesado sobre la balanza y pulsar TARE - TARA. Sustraer una porción o parte del elemento hasta alcanzar el rango ACEPTADO.
Zero Check
Zero check is used when comparing subsequent samples to an initial reference sample. In this case, the UNDER limit must be a negative value and the OVER limit must be a positive value. Place the reference item on the balance and press TARE. Remove the reference sample and place the item to be compared on the balance to determine if it is within the ACCEPT range.
ES 8 Balances ScoutTM Séries – SPX
3.8 Modo Totalización
Este modo permite al usuario almacenar en la memoria una serie de mediciones de pesaje. El modo de totalización se inicia cuando el símbolo ” aparece en pantalla y se muestra la unidad actual. Nota :Únicamente se totalizan cifras positivas.
1. Pulsar y mantener pulsado el botón [Mode] hasta que [total] (Totalización) aparezca en pantalla. Entonces aparecerá en pantalla[CLr.tot].
2. Pulsar la tecla Yes - sí o Nopara borrar o no el dato de totalización actualmente almacenado en la memoria. Cuando se añade / coloca un peso en la balanza el valor aparecerá en
pantalla.
3. Si fuese necesario, coloque un recipiente vacío sobre la bandeja y pulse Tare - Tara. Añada
el primer elemento, su peso aparecerá en pantalla.
Pulsar Function (function) para almacenar su peso en la memoria, el símbolo
parpadeará en la pantalla en la que aparecerá el peso total.
4. Pulsar Tare - Tara (o quitar el peso de la operación anterior) y añada el siguiente elemento
La balanza mostrará su peso en pantalla.
Pulsar Function (function) para almacenar su peso en la memoria, el símbolo
parpadeará en la pantalla en la que aparecerá el nuevo peso total.
5. Repetir el paso 4 para todos los elementos cuyo peso haya que acumular.
6. Para borrar de la memoria el total almacenado, pulsar y mantener pulsado Mode hasta que
[total] aparezca en pantalla. Cuando aparezca en pantalla [Clr.tot] , pulsar Yes - sí.
3.9 Modo de mantenimiento de los valores mostrados en pantalla “Hold”
Hay dos modos que permiten mantener los valores mostrados en pantalla “hold” :
- PeakHold – Mantener el valor más elevado “pico” :permite al usuario capturar y almacenar en la memoria el valor de pesaje estable más elevado (>=5d).
- DisplayHold – Mantener los valores mostrados en pantalla (configuración por defecto):permite al usuario capturar y almacenar en la memoria el primer valor de pesaje estable(>=5d).
Inicio Si no se mantiene en pantalla ningún valor de pesaje, pulsar la tecla [Function] para iniciar. Se
mostrará en pantalla el mensaje [rEAdY] (Ready - Preparado) hasta que se coloque un peso en la balanza. Cuando la balanza muestre en pantalla un valor estable, el icono Hold– mantener en pantalla ­(H)parpadeará y el peso mostrado en pantalla no cambiará.
Reset - Reconfigurar Si no hay ningún peso colocado en la bandeja y se mantiene un valor mostrado en la pantalla, basta con una pulsión corta de la tecla [Function]– función para eliminar el valor mostrado en pantalla y que se muestre el valor del nuevo pesaje en la pantalla.
1. Pulsar y mantener pulsado el botón Mode hasta que [HoLd] aparezca en pantalla.
2. Si fuese necesario, coloque un recipiente vacío sobre la bandeja y pulse Tare - Tara. Entonces se mostrará en pantalla el valor Cero.
3. Pulsar la tecla [Function] para iniciar. Se mostrará en pantalla el mensaje [rEAdY] (Ready - Preparado).
4. Colocar las muestras que deban ser pesadas sobre la bandeja.
5. La pantalla mostrará el valor estable, el icono “Hold” (H) – mantener en pantalla ­pestañeará.