Marked with: EC Directive (Including applicable amendments) Standard
73/23/EC IEC 1010 -1:1990 + A1: 92 + A2: 95
Electrical equipment for use within specified Safety requirements for Electrical Equipment for Measurement,
voltage limits Control Laboratory Use, Part 1: General Requirements
CE marking affixed in 97
89/336/EC EN61326: -1:1997 (class B) + A1: 1998
Electromagnetic compatibility EMC Emissions, residential, commercial and light industry.
EN61000-3-2:1995 + A1:1998 + A2: 1998; EN61000-3-3:1995
EMC Part 3 (for equipment rated input current < or=16A)
Limits- Section 2: Limits for harmonic current emissions
Limits- section 3: Limitation of voltage fluctuations and flicker in low
voltage supply systems
xx
ISO 9001 Registration for Ohaus Corporation.
tional, BVQI, and was awarded ISO 9001 registration. This certifies that Ohaus Corporation, USA, has a quality system that conforms with the international
standards for quality management and quality assurance (ISO 9000 series). Repeat audits are carried out by BVQI at intervals to check that the quality
system is operated in the proper manner.
Additional Standards
CAN/CSA-C22.2 No. 1010.1-92; UL Std. No. 3101-1
Safety requirements for Electrical Equip. for measurement, Control and Laboratory Use, Part 1; General Requirements
FCC
FCC, Part 15, class A Emission
EU 90/384 NAWI EN45501:1992 Non Automatic Weighing Instruments
T2914
xx= year CE affixed
Ohaus Corporation, USA, was examined and evaluated in 1994 by the Bureau Veritas Quality Interna-
Ted Xia
President
Ohaus Corporation
Pine Brook, NJ USA
Date: March 6, 2003
Johan Dierbach
General Manager
Ohaus Europe
Greifensee, Switzerland
Date: March 6, 2003
AS/NZS4251.1 AS/NZS4252.1 Emission and Immunity
315
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the
interference at his own expense.
This Class A digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada
3.1Overview of Controls .............................................................................................................................. EN-8
3.2Overview of Display Indicator ................................................................................................................ EN-10
3.5.5 Units ........................................................................................................................................ EN-26
3.5.13 Sealing the Balance ................................................................................................................. EN-30
3.6Printing Data ....................................................................................................................................... EN-30
4. CARE AND MAINTENANCE ........................................................................................................................ EN-31
4.3Error Codes List ................................................................................................................................... EN-32
4.4Service Information .............................................................................................................................. EN-33
4.5Replacement Parts ............................................................................................................................... EN-33
5. TECHNICAL DATA ...................................................................................................................................... EN-34
* 4100 g, 6100 g and 8100 g balances with internal calibration
are equipped with a 6" Pan and Windshield.
The following items are supplied with all balances:
AC Power Adapter, Instruction Manual, Warranty Card, In Use
Cover
Remove packing material from the instrument.
Check the instrument for transport damage. Immediately
inform your Ohaus dealer if you have complaints or parts
are missing.
Store all parts of the packaging. This packaging
guarantees the best possible protection for the transport of your
instrument.
1.3 Safety Precautions
Please follow the safety precautions as listed.
CAUTION:
Do not operate the balance around corrosive fumes.
Only use the adapter provided with the balance.
Do not try to service the Explorer® Pro balance.
Page 6
EN-4 Explorer Pro
BOTTOM OF BALANCE
WEIGH BELOW
HOOK LOCATION
REMOVE
PROTECTIVE
COVER.
WEIGH BELOW
HOOK IS INSIDE OF
BALANCE.
2.2Installing Components
2.2.1Platform Installation
Balances are shipped with the pan not installed. On balances
equipped with a draft shield, slide open the side door and insert
the pan into the center hole.
2.2.2Windshield Installation
On 610 g to 6100 g balances with 0.01 g resolution, a wind-
shield is required to reduce the possibility of air currents from
disturbing the pan. When the windshield is in place, air currents
are deflected up over the pan. Make sure the windshield is firmly
snapped into place.
Wind Shield
NOTE: 4100 g, 6100 g and 8100 g balances with internal
calibration are equipped with a 6" Pan and Windshield.
2.2.3Weigh Below Preparation
The Explorer Pro balance is equipped with a weigh below hook
at the bottom of the balance. To use this feature, remove the
protective cover underneath the balance. See illustration for
location. The balance can be supported using lab jacks or any
other convenient method. Make sure the balance is level and
secure. Apply power and operate the balance. Attach items to be
weighed to the hook underneath the balance.
Page 7
Explorer Pro EN-5
2.3Selecting the Location
2.3.1Balance Location
The balance should always be used in an environment which is
free from excessive air currents, corrosives, vibration, and
temperature or humidity extremes. These factors will affect
displayed weight readings.
DO NOT install the balance:
Next to open windows or doors causing drafts or rapid
temperature changes.
Near air conditioning or heat vents.
Near vibrating, rotating or reciprocating equipment.
Near magnetic fields or equipment that generates magnetic
fields.
On an unlevel work surface.
Allow sufficient space around the instrument for ease of
operation and keep away from radiating heat sources.
2.3.2Leveling the Balance
Exact horizontal positioning and stable installation are
prerequisites for repeatable results. To compensate for small
irregularities or inclinations at the location, the instrument can
be leveled.
For exact horizontal positioning, the balance is equipped with a
level indicator located at the front on the control panel and two
leveling feet located at the rear of the balance.
Position the balance in the intended operating location. Adjust
the leveling feet at the rear of the balance until the air bubble in
the indicator is centered.
NOTE: The instrument should be leveled each time its location is
changed.
Correct Incorrect
Leveling Indicator
Leveling
Foot
Page 8
EN-6 Explorer Pro
12345
6789
2.4Connecting Power and Communications
2.4.1AC Adapter Installation
Connect the AC Adapter supplied to the three pin connector located
at the rear of the balance.
The balance is now ready for operation.
AC Adapter Connection
Rear of Balance
2.4.2RS232 Interface
Explorer Pro balances are equipped with a bidirectional RS232
compatible interface for communication with serial printers and
computers. When the balance is connected directly to a printer,
displayed data can be output at any time by simply pressing the
Print button, or by using the Auto Print feature. Connecting the
balance to a computer enables you to operate the balance from
the computer, as well as receive data such as displayed weight,
weighing mode, stability status, etc.
Male Connector
The following sections describe the hardware and software
provided with the balance
Hardware
On the rear of the balance, the right-hand, 9-pin male subminiature
D connector is provided for interfacing to other devices. The
pinout and pin connections are shown in the adjacent illustration.
Refer to paragraph 3.5.6 for setup.
The balance is equipped with hardware handshaking, it will not
output any data unless pin 5 (CTS) is held in a high state (+3 to
+15 V dc). Interfaces not utilizing the CTS handshake may tie pin
5 to pin 6 to defeat it.
Output Formats
Data output can be initiated in one of three ways: 1) By pressing
PRINT; 2) Using the Auto Print feature; 3) Sending a print
command (P) from a computer.
1 N/C
2 Data Out (TXD)
3 Data In (RXD)
4N/C
5 Clear To Send (CTS)
6 Data Terminal Ready (DTR)
7 Ground
8 Request To Send (RTS)
9N/C
Additional information is located in Section 5, Technical Data
which contains the RS232 Command Table.
Page 9
Explorer Pro EN-7
2.4.3 Initial Calibration
Calibration Masses
Calibration masses are required if the balance is not equipped
with internal calibration. Before beginning calibration, make sure
masses are available. If you begin calibration and realize
calibration masses are not available, exit the menu. The balance
will retain previously stored calibration data. Calibration should
be performed as necessary to ensure accurate weighing. Masses
required to perform the procedures are listed in the following table.
NOTE: Any of the calibration modes can be terminated at any
time by pressing the Menu button.
CALIBRATION MASSES
LINEARITY SPAN ONLY
CAPACITYMASSESMASSES
62 g 20g/50 g 50 g
162 g 50g/150 g150 g
110 g 50g/100 g100 g
210 g 100g/200 g200 g
410 g 200g/400 g400 g
510g/610 g 200g/500 g500 g
1550 g 500g/1500 g 1500 g
2100 g1000g/2000 g 2000 g
4100 g2000g/4000 g 4000 g
6100 g2000g/5000 g 5000 g
8100 g4000g/8000 g 8000 g
It is recommended that masses must meet or exceed ASTM
Class 1 Tolerance. Calibration masses are available as
accessories.
Calibrating from the Weighing Screen
When the balance is first turned on, three soft keys appear on
the weighing screen. The CALIBRATE soft key is highlighted.
This permits calibration immediately without entering the
CALIBRATION menu. On balances equipped with internal
calibration, the balance automatically calibrates. On balances
that do not have internal calibration, an external mass is required.
WEIGHING 01/01/03 03:05:45: PM
UNITS
*
0g
Press the Enter button with the CALIBRATE soft key highlighted.
On balances that do not have internal calibration, a message
appears on screen indicating the mass value to be placed on the
pan and also indicates other values of masses that can be used.
Place the required mass on the pan and press the Enter button.
The balance performs a span calibration.
2nd UNITS
CALIBRATE
0.00
0.00g
g
6100g
Page 10
EN-8 Explorer Pro
3OPERATION
3.1Overview of Controls
1
WEIGHING 01/01/03 03:05:45: PM
13
UNITS
*
2nd UNITS
0.00
12
11
10
9
0g
8
No.DesignationFunction
1 IPower On/ Off button.
CALIBRATE
0.00g
g
6100g
7
10
2
3
4
5
9
6
7
14
2 buttonWhen pressed in Menu mode, single press moves menu selection bar in an up direction and
highlights the activated field or changes the settings of a selected field in increasing order.
When at an alphanumeric field, then the number or letter is incremented.
3 buttonWhen pressed in Menu mode, single press moves the cursor within a alphanumeric field
to the left (example date 07/02/2003). Selection does not wrap at end.
In application mode, single press moves to the next soft key selection to the left.
4 buttonWhen pressed in Menu mode, single press moves the cursor within a alphanumeric field
to the right (example date 07/02/2003). Selection does not wrap at end.
In application mode, single press moves to the next soft key selection to the right.
Page 11
Explorer Pro EN-9
No.Designation Function
5
6Enter buttonWhen pressed in application modes, functions as an Enter button to accept Soft-key selection.
7 & 9>O/T< buttonsWhen pressed, performs tare function or zero function.
8Spirit LevelProvides a leveling indication for the balance.
10LCD DisplayBacklit LCD display provides all necessary indications for operation.
11Mode buttonWhen pressed, causes the balance to cycle through all active application modes.
12Print buttonWhen pressed in any application mode, causes a print function to occur.
buttonWhen pressed In Menu mode, single press moves menu selection bar in a down direction
and highlights the selected field or changes the settings of a selected field in decreasing
order.
When at an alphanumeric field, then the number or letter is decremented.
When pressed in menu, functions as an Accept button for accepting and saving
parameter of selected items.
13Menu buttonWhen pressed in any application mode, will bring up the Menu screen. Pressing the button
within a Menu screen will return to the application mode.
14FeetTwo adjustable feet used to level the balance.
Page 12
EN-10 Explorer Pro
3.2Overview of Display Indicator
12
WEIGHING 01/01/03 03:05:45: PM
UNITS
Message Line
13
12
11
No.DesignationFunction
1ApplicationIndicates the active application by function.
2DateIndicates current date when properly set.
*
-
Result Line Result Line
Result Line Result Line
Result Line Result Line
0g
200.0000
0
2nd UNITSCALIBRATE
10
3
dwt
6100g
4
5
6
7
8
9
3TimeIndicates current time when properly set.
4Soft KeysThree Soft-keys are displayed in the upper display area. Depending on the application mode
selected, the functionality changes. A Soft-key can be selected by using the left and right
arrow buttons. Pressing the Enter button on a highlighted soft-key will activate the
function.
5Message LineContains instructional messages.
6Numerical DisplayIndicates primary weight/load.
7Unit of measureIndicates active unit of measure.
8Result Lines 1-6Six lines that contain balance results.
9CapacityIndicates full balance capacity.
10Bar GraphShaded area indicates capacity used for, under, accept, over for check weighing application,
or target values for filling application.
11Indicates center of zero (only active in legal for trade).
0
12Indicates negative value.
13Stability indicator.
*
Page 13
Explorer Pro EN-11
3.3Menu
3.3.1Menu Structure
The Explorer Pro balance utilizes a menu structure that permits
entering various menus by using a dedicated Menu button.
Pressing the Menu button allows access to additional sub menus.
This feature permits establishing a set quantity of pieces that can
be shown as a percentage when pieces are added. A split bar
display indicates 0% to 90% on the first bar and 90% to 110%
on the second bar.
Refer back to Parts Counting, enter the PARTS COUNTING SETUP
menu and change Mode to Fill.
The balance is now in the PC FILLING mode.
SETUP
Select SETUP soft key and press the Enter button. The PARTS
COUNTING SETUP is shown.
PARTS COUNTING SETUP
Mode: Fill
Unit: g
Sample Size: 10
APW: Off
Tare: 0.00
Auto Optimize: Off
Target:1000
Display Settings: Default
Size/APW Key: Unlocked
New Count Key: Unlocked
Exit
Page 19
Explorer Pro EN-17
g
g
One additional entry has been added to the PARTS COUNTING
SETUP, that is Target. Refer to the Adjustments section of parts
counting for a description of all other settings.
Target: 0-9999 (default is 1000)
Defines the target pieces for the 100% limit.
Display Settings: Custom, Default (default is Default)
Selecting Display Settings to Custom will bring up the DISPLAY
SETTINGS screen that will allow individual display items to be
turned on or off. APW, Sample Size, Tare, Weight are described
under parts counting Display Settings.
Two new entries Target and Difference appear on this screen.
Selecting Exit will return balance to previous screen.
Target: On, Off (default is On)
Can be turned on or off.
Difference: On, Off (default is On)
Can be turned on or off.
3.4.3Percent Weighing
Percent weighing allows placing a reference load on the balance,
then view other loads as a percentage of the reference. The load
placed on the pan as a reference is displayed as 100%.
Subsequent loads are displayed as a percentage of the reference
are limited. The maximum limit is the capacity of the balance.
The minimum limit is 100d. The display data includes reference
weight in unit of measurement, difference value in percent,
difference value in measuring unit and a bar graph indicating
present used capacity of the balance.
PERCENT WEIGHING 01/01/03 03:05:45 PM
REFERENCE
Add Reference Weight, Press Enter
SETUP
NEW REF
*
Ref: 0.00g
Diff: 0.00 %
Diff: 0.00g
0g
0.00 g
6100
SAMPLE SIZE/APW
Select SAMPLE SIZE soft key and enter the sample size desired.
APW
With the APW soft key highlighted, press the Enter button. The
APW is highlighted at the lower left of the screen. Using the
arrow buttons and the Enter button, the APW value can be
changed. Follow screen instructions and place sample on pan.
Remove the sample and add items on pan until 100% is
highlighted at the bottom of the screen as shown. In the following
example, the sample size was 10, the APW was 2.00g, the Target
Unit: g, kg, etc...(default is g)
Cycles through all units turned On in the Units menu.
Over: 0-99999999 (default is 15)
Defines over limit value.
Target: 0-99999999 (default is 10).
Defines target value.
Under: 0-99999999 (default is 5)
Defines under limit value.
Display Settings: Custom, Default (default is Default)
Setting the display to Custom allows individual items Over, Target,
Under, Difference, Difference % to be turned On or Off. In the
default setting, all items are turned on. Additionally, the item
reading can be set to weight or message. Default is Weight.
Reading allows a choice of either weight readings to be displayed
in large numeral or message that displays OVER, ACCEPT and
UNDER in large letters. Display below indicates message format.
CHECK WEIGHING 01/01/03 03:05:45 P
UNITS
*
UNDER
Over: 15.00 g
Target: 10.00 gDiff: -10.00 g
Under: 5.00 gDiff: -100.00 %
SETUP
CALIBRATE
0.00g
3.4.6Gross / Net / Tare Weighing
Gross/Net/Tare (G/N/T) application allows to display Gross
(sample plus container weight), NET (sample weight) and TARE
(container weight) simultaneously.
G/N/T WEIGHING 01/01/03 03:05:45 PM
UNITS
2nd UNITS
TARE
*
0.00
Gross: 0.00 g
Net: 0.00 g
Tare: 0.00 g
0g
USE
G/N/T WEIGHING
Press the O/T button to zero the balance. In this application, the
O/T button functions as a Zero and not as a Tare.
Place a container on the pan. With the TARE soft key highlighted,
press the Enter button. The containers weight is stored in memory
in the balance. Place the material in the container. The balance
immediately displays the Gross, Net and Tare weights. The net
weight is displayed as large numerals. Example shown represents
a container weight of 50g and material of 200g. The gross
weight is displayed as 250g. Before a new container is used,
zero the balance by using the O/T button.
G/N/T WEIGHING 01/01/03 03:05:45 PM
UNITS
2nd UNITS
TARE
g
0.00 g
6100
Exit:
When selected, returns to Check Weighing mode.
UNITS
When the UNITS soft key is highlighted, each press of the Enter
button cycles through the measuring units that are On.
*
Gross: 250.00 g
Net: 200.00 g
Tare: 50.00 g
0g
200.00
200.00 g
g
6100
Page 24
EN-22 Explorer Pro
ADJUSTMENTS
The G/N/T WEIGHING display contains three soft keys at the top
of the screen UNITS, 2nd UNITS and TARE.
TARE
When the TARE soft key is selected, pressing the Enter button will
store the current display reading as the tare value.
2nd UNITS
When 2nd UNITS soft key is highlighted, each press of the Enter
button changes the measuring units that are on and will appear
under the large numerals. The last option when cycling through
the units is Off.
UNITS
When the UNITS soft key is highlighted, each press of the Enter
button cycles through all active measuring units on the Gross,
Net and Tare displays.
3.4.7Filling
Filling permits entering a target reference weight. Material can be
placed on the pan and by monitoring the Difference weight and
the bar display, an accurate fill can be achieved. A split bar
display indicates 0% to 90% on the first bar and 90% to 110%
on the second bar.
ADJUSTMENTS
The FILLING display contains three soft keys at the top of the
screen TARGET, SETUP and NEW TARGET.
NEW TARGET
With the NEW TARGET soft key highlighted, press the Enter button.
Follow display instructions to Add Target Weight, Press Enter.
This establishes a new target weight and is displayed on the
screen.
SETUP
When SETUP soft Key is highlighted, press the Enter button. The
FILLING SETUP display is shown.
FILLING SETUP
Unit: g
2 nd unit: g
Target Weight: 0000
Display Settings: Custom
Target Key: Unlocked
New Target Key: Unlocked
Exit
Units: g, kg, etc...(default is g)
Cycles through all units turned On in the Units menu.
FILLING 01/01/03 03:05:45 PM
TARGET
*
Target: 50.00 g
Diff: -50.00 g
0%
SETUP
0.00
50%
NEW TARGET
90%
100%
0.00 g
g
110%
USE
FILLING
A sample (target weight) is placed on the pan and the Enter
button is pressed. The balance stores this weight as shown on
the display as Target. The sample is removed from the pan and
the material is added to the pan. The display provides Target
Weight, Difference weight and the split bar at the bottom of the
screen displays the percentage of the sample compared against
the target weight. When a container is used, zero the balance
before placing the actual sample on the pan. See Adjustments to
set other balance parameters.
2nd UNITS: g, kg, etc...(default is g)
Cycles through all units turned On in the Units menu..
Target Weight: 0-99999999 (default is 0)
Allows entering a specific target filling weight.
Display Settings: Custom, Default (default is Default)
Setting the display to Custom allows individual items Target
Weight, Difference, 2nd weight to be turned On or Off. In the
default setting, all items are turned on.
Target Key: Locked, Unlocked (default is Unlocked)
This feature allows to lock or unlock the soft key function.
New Target Key: Locked, Unlocked (default is Unlocked)
This feature allows to lock or unlock the soft key function.
Exit:
When selected, balance returns to the Filling Mode.
Page 25
Explorer Pro EN-23
TARGET
With the TARGET soft key highlighted, press the Enter button.
The FILLING display is shown with the value of the Target
highlighted. Using the arrow buttons and the Enter button, the
Target value can be changed.
FILLING 01/01/03 03:05:45 PM
TARGET
*
Target: 0.00 g
Diff: -0.00 g
0%
SETUP
0.00
50%
NEW TARGET
90%
100%
g
0.00 g
110%
3.5Balance Settings
The Explorer Pro balance contains ten submenus that are
accessible from the Menu. The submenus are Calibration, Balance
Each of the submenus contain settings that will affect the operation
of the balance. Please review all submenus settings to obtain
the best performance from the balance. Make the necessary
settings to suit specific needs.
3.5.1Calibration
Press the Menu button and select CALIBRATION. Press the enter
button, the CALIBRATION screen is displayed. Explorer Pro
balances offers a choice of five calibration methods: Internal
Calibration, Autocal Calibration, Span Calibration, Linearity
Calibration and User Calibration,
IMPORTANT !
DO NOT DISTURB THE BALANCE
DURING CALIBRATION.
Internal Calibration
On Explorer Pro balances equipped with internal calibration,
calibration is accomplished using the internal calibration mass.
Internal calibration can be performed at any time providing the
balance has warmed up to operating temperature and is level.
AutoCal: On, Off (default is On)
Explorer Pro balances equipped with internal calibration also
contain AutoCal. When AutoCal is set ON, the balance performs
a self calibration when a measured predefined delta temperature
change occurs. After calibration, the display returns to the last
application mode. Environmental conditions must be met for the
AutoCal to be successfully completed.
AutoCal Delta: -100 - +100 (default is 0)
Explorer Pro balances equipped with internal calibration also
contain AutoCal Delta. AutoCal Delta allows the internal calibration
mass value to be adjusted. This permits calibrating the balance
using an external mass which is traceable to a certified standard.
Perform the internal calibration procedure. After the calibration is
completed, zero the balance.
Place a certified mass equal to the span calibration value of the
balance.
Linearity, Span and User calibration are disabled for Type
Approved/LFT balances Class II and Class III.
CALIBRATION
Internal
AutoCal : Off
AutoCal Delta (d): 0
Span
Linearity
User
User Cal Weight (g): 500.00
Cal Test
Exit
Page 26
EN-24 Explorer Pro
Compare the reading on the balance to the expected weight being
used. If the reading is above the expected weight, the delta entered
in digits is negative. If the reading is below the expected weight,
the delta entered in digits is positive.
See following example:
Actual Weight Reading:200.0014
Expected Weight Reading: 200.0000
Delta Weight (d): 0.0014
Delta weight in Digits:-14
Recalibrate using internal calibration. After calibration, place the
certified mass on the pan and see if the mass agrees with the
displayed value. If not, repeat procedure until internal calibration
reading agrees with the certified mass.
Span Calibration:
Span calibration utilizes two calibration points, one at zero load
and the other at specified full span. See following table.
CALIBRATION MASSES
LINEARITY SPAN ONLY
CAPACITYMASSESMASSES
62 g 20g/50 g 50 g
162 g 50g/150 g 150 g
110 g 50g/100 g 100 g
210 g 100g/200 g 200 g
410 g 200g/400 g 400 g
510g/610 g 200g/500 g 500 g
1550 g 500g/1500 g1500 g
2100 g 1000g/2000 g2000 g
4100 g 2000g/4000 g4000 g
6100 g 2000g/5000 g5000 g
8100 g 4000g/8000 g8000 g
It is recommended that masses must meet or exceed ASTM
Class 1 Tolerance. Calibration masses are available as
accessories.
Additional calibration values to be used are shown on the display
screen. The best accuracy is achieved using the mass closest
to the full span value. After calibration, the display returns to the
last application mode.
Linearity Calibration:
Linearity calibration utilizes three calibration points, one at zero
load and two at specified calibration masses. This method
minimizes deviation between actual and displayed weights within
the balance's weighing range. After calibration, the display returns
to the last application mode.
User Calibration:
User calibration is used when it is desired to calibrate the balance
using a user defined mass. The user defined mass value has to
be entered in the User Cal Weight. After calibration, the display
returns to the last application mode.
User Calibration Weight: 25%-100% of capacity (default is Span)
Defines the mass value for User calaibration. The user calibration
may now be done with the mass selected.
Cal Test:
Calibration test feature allows a check of a known calibration
mass against the last stored calibration information in the balance.
The display indicates the difference in weight between calibration
mass placed on the pan and the previous weight value which
was stored in the balance. After Cal Test, the display returns to to
the last application mode.
3.5.2Balance Options
Press the Menu button, and select Balance Options.
Press the Enter button. The BALANCE OPTIONS screen is
displayed.
BALANCE OPTIONS
Language: English
Display Date/Time: Date/Time
Date Format: DD/MM/YY
Date: 00/00/00
Time Format: 12 Hour
Time: 00:00:00
Contrast:
Brightness:
Audible Signal: Off
Exit
Page 27
Explorer Pro EN-25
Languages: English, Spanish, etc...(default is English)
Allows the selection of English, Spanish, French, German or Italian
as the operating language to be displayed.
Display Date/Time: Time, date, etc...(default is Date/Time)
Allows display selection of Time, Date, Date/Time or Off in the
upper right-hand corner of the Application Mode screen.
Date Format: MM/DD/YY, YY/MM/DD, ETC...(default is MM/DD/
YY)
Allows setting of one of six date formats: MM/DD/YY,
YY/MM/DD , DD/MM/YY, DD/YY/MM, MM/YY/DD or YY/DD/MM.
Date: OO/00/00 (default is none)
Allows setting the present date.
Time Format: 12 Hour, 24 hour(default is 12 Hour)
Allows setting the time format of either 12 hours or 24 hours.
Time: 00:00:00 (default is none)
Allows setting the present time of hours, minutes and AM/PM.
Valid numbers change with time format from 1->12 to 0->23.
When in 24 hour mode, no AM/PM flag appears.
Contrast:
Allows adjusting the contrast level of the display.
READOUT
Filter Level: Medium
Stability Level (d): 1.0
Auto Zero (d): 0.5
Auto Tare: Off
Legal for Trade: Off
Exit
Filter Level: Low, Medium, High (default is Medium)
Sets the balance reading averaging level to a value of Low,
Medium or High. A higher filter level will provide a more repeatable
measurement.
Stability Level: 0.5, 1.0, etc... (default is 1.0)
Sets the balance stability level for the stability indicator to either
0.5, 1.0, 2.0 or 5.0. A 0.5 setting is equivalent to end result of
.5 display digit. A 5.0 setting is equivalent to end result of 5
display digits.
Auto Zero: Off, 0.5, etc... (default is 0.5)
Sets the balance auto zero level to either Off, 0.5, 1.0, 2.0 or
5.0. Auto Zero tracking compensates for the drift at zero load.
0.5d means 0.5 display digit drift compensation per second.
5.0d means 5 display digit drift compensation per second.
Brightness:
Allows adjusting the brightness level of the display.
Audible Signal: On, Off (default is Off)
With audible signal set On, each button press will generate a
tone.
Exit:
When selected, returns to previous screen.
3.5.3Readout
Press the Menu button, and select Readout.
Press the Enter button. The READOUT screen is displayed.
Auto Tare: On, Off (default is Off)
Sets the Auto Tare function to On or Off. This feature provides
auto tare in all application modes. When Auto Tare is On, the
balance waits for the container/load to be placed on the pan and
automatically tares it. This function is repeated for each new
container/load.
Legal for Trade: On, Off (default is Off)
Sets the Legal for Trade (LFT) function either On or Off. Refer to
paragraph paragraph 3.5.11 for additional information.
Page 28
EN-26 Explorer Pro
3.5.4Application Modes
Press the Menu button, and select Application Modes.
Press the Enter button APPLICATION MODES screen is displayed.
APPLICATION MODES
Weighing: On
Parts Counting: Off
Animal Weighing: Off
Percent Weighing: Off
Check Weighing: Off
G/N/T: Off
Filling: Off
Exit
All application modes are displayed on this screen. Each mode
may be set on or off. These modes are cycled through when the
Mode button is pressed.
3.5.5Units
Press the Menu button, and select Units. The UNITS screen is
displayed and displays a list of the available weighing units.
Depending on the model, some units are not available.
In some models, the number of units contained in the balance
exceeds the screen capacity. In order to view all units, the down
arrow must be pressed repeatedly to view all available units.
UNITS
Carat (ct): Off
Penny weight (dwt): Off
Troy OZ (ozt): Off
Grain (GN): Off
HK Tael (Hkt): Off
SG Tael (SGt): Off
ROC Tael (ROt): Off
Newton (N): Off
Momme (m): Off
Tical (ti): Off
Custom (Cst): Off
Exit
Milligram (mg): On, Off (default is Off)
Unit (mg) = g x 1000, displayed readability by 1.
Gram (mg): On, Off (default is On)
Unit (g) = g x 1, displayed readability by 1.
Kilogram (kg): On, Off (default is Off)
Unit (kg) = g x .001, displayed readability by 1.
OZ avdp (oz): On, Off (default is Off)
Unit Ounces (oz) = g x .002204623, displayed readability by 5.
Pound avdp (lb): On, Off (default is Off)
Unit Pounds (lb) = g x .03527396, displayed readability by 5.
Carats (ct): On, Off (default is Off)
Unit (ct) = g x 5, displayed readability by 5.
UNITS
Milligram (mg): Off
Gram: On
Kilogram: Off
Oz avdp (oz): Off
Pound avdp (lb): Off
Carat (ct): Off
Penny weight (dwt): Off
Troy OZ (ozt): Off
Grain (GN): Off
HK Tael (Hkt): Off
SG Tael (SGt): Off
ROC Tael (ROt): Off
Scroll down using arrow button to view remaining units.
Pennyweight (dwt): On, Off (default is Off)
Unit (dwt) = g x .6430149, displayed readability by 1.
Troy OZ (ozt): On, Off (default is Off)
Unit (ozt) = g x .03215075, displayed readability by 5.
Grain (GN): On, Off (default is Off)
Unit (GN) = g x 15.43236, displayed readability by 2.
Hong Kong Tael (HKt): On, Off (default is Off)
Unit (HKt) = g x 0.02671725, displayed readability by 5.
Singapore Tael (SGt): On, Off (default is Off)
Unit (SGt) = g x 0.02645547, displayed readability by 5.
ROC Tael (ROt): On, Off (default is Off)
Unit (ROt) = g x 0.02666667, displayed readability by 5.
Page 29
Explorer Pro EN-27
Newton (N): On, Off (default is Off)
Unit (N) = g x 0.00980665, displayed readability by 1.
MOMME (m): On, Off (default is Off)
Unit (m) = g x 0.2666667, displayed readability by 5.
Tical (ti): On, Off (default is Off)
Unit (ti) = g x 0.0612395, displayed readability by 1.
Custom Unit (Cst): On, Off (default is Off)
When the Custom Unit is set On, the CUSTOM UNIT screen will
appear.
This feature can be used to create a custom weighing unit. It
permits entering a conversion factor which the balance will use
to convert grams to the desired unit of measure.
Conversion Factor x Grams = Custom Unit
Unit (Cst) = g x user set custom factor, displayed resolution can
not exceed resolution in gram.
EXPONENTS
E-3Moves decimal point 3
places to the left.
E-2Moves decimal point 2
places to the left.
E-1Moves decimal point 1
place to the left.
E0Leaves decimal point in
normal position.
E1Moves decimal point 1
place to the right.
E2Moves decimal point 2
places to the right.
E3Moves decimal point 3
places to the right.
LSDs
LSD .5Adds one decimal place
display counts by 5s.
LSD 1Display counts by 1s.
LSD 2Display counts by 2s.
LSD 5Display counts by 5s.
LSD 10Display counts by 10s.
LSD 100Display counts by 100s.
Conversion factors are expressed in scientific notation
and entered into the balance in three parts:
Mantissa (0.1 and 1.999999)
Exponent (10
E
)
Least Significant Digit (LSD)
CUSTOM UNIT
Factor: 1.0
Exponents: E1
LSD: 1
Exit
SCIENTIFIC NOTATION
Mantissa
Conv. 0.1 and Exponent Man-Exp.
Factor1.999999 tissa (10E)
123.4 = .1234 x 1000 =.1234 x 10
12.34= .1234 x 100 = .1234 x 10
1.234= .1234 x 10 = .1234 x 10
.1234= .1234 x1 = .1234 x 10
.01234 = .1234 x .1 = .1234 x 10
.001234 = .1234 x .01 = .1234 x 10
.000123 = .123 x .001 = .123 x 10
Between
3
2
1
0
-1
-2
-3
3.5.6 Interface
Press the Menu button, and select Interface.
Press the Enter button, The INTERFACE screen is displayed.
INTERFACE
Baud Rate: 2400
Parity: None
Data Bits: 7
Stop Bits: 2
Exit
Baud Rate: 300, 1200, etc...(default is 2400)
Baud Rate is selectable between 300, 1200, 2400, 4800 or
9600 BPS.
Parity Bit: None, Odd, Even (default is None)
Parity Bit is selectable between None, Odd or Even.
Data Bits: 7, 8 (default is 7)
Data Bits is selectable between 7 or 8 data bits.
Stop Bits: 1, 2 (default is 2)
Stop Bits is selectable between 1 or 2 stop bits.
Exit:
Will return to previous display.
Page 30
EN-28 Explorer Pro
3.5.7Print Options
Press the Menu button, and select Print Options.
Press the Enter button PRINT OPTIONS screen is displayed.
PRINT OPTIONS
Auto Print: Off
Print Interval: Off
Stable Data: Off
Numeric Data: Off
Exit
Auto Print: Off, interval, etc... (default is Off)
When enabled, the Auto Print feature causes the balance to
automatically output display data in one of three ways:
continuously, at user specified time intervals, or upon stability.
OFFTurns off the auto print feature.
IntervalProvides a user specified printing interval
StableProvides printed data each time a stable
reading is achieved.
ContinuousOutputs data continuously.
Print Interval: 1-3600 (default is 1)
When Auto Print is set to Interval, a specified printing interval
between 1 and 3600 seconds can be set.
Stable Data: Load, Load & Zero (default is Load)
When Auto Print is set to Stable, an option of Load or Load &
Zero can be set.
LoadWill print stable load data only.
Load & ZeroWill print stable load data and stable zero
data.
Numeric Data: On, Off (default is Off)
When set On, will print numeric data.
GLP PRINT OPTIONS
Project Name: OHAUS
User Name: OHAUS
Date & Time: Off
Balance ID: Off
Project Name: Off
User Name: Off
Calibration: Off
Reference: Off
Exit
Project Name: (default is OHAUS)
A project name of up to 8 characters may be entered.
User Name: (default is OHAUS)
A user name of up to 8 characters may be entered.
Date & Time: On, Off (default is Off)
When set On, will output date and time.
Balance ID: On, Off (default is Off)
When set On, will output Balance ID.
Project Name: On, Off (default is Off)
When set On, will output project name.
User Name: On, Off (default is Off)
When set On, will output User name.
Calibration: On, Off (default is Off)
When set On, prints out the calibration data after the completion
of every calibration process. See sample of span calibration.
SAMPLE PRINTOUT
---------------SPAN CAL-------------Cal: 400.000 g
The default setting for all items on the LOCKOUT screen are
Unlocked.
Balance Options:
Language: English
Display Date/Time: Date/Time
Date Format: MM/DD/YY
Time Format: 12 Hour
Readout:
Filter Level: Medium
Stability Level (d): 1.0
Auto Zero (d): 0.5
Auto Tare: Off
Legal for Trade: Off
Application Modes:
Weighing: On
All others Off
Units:
Grams On, all others Off.
3.5.10Factory Reset
The FACTORY RESET allows parameters to be set to the default
values. Each menu item can be reset by setting to Yes. After
accepting the changes, the balance will reset the selected menus
items. Global Reset will change all menus to a factory setting in
one step.
Press the Menu button, and select Factory Reset.
Press the Enter button FACTORY RESET screen is displayed.
FACTORY RESET
Calibration: No
Balance Options: No
Readout: No
Application Modes: No
Units: No
Interface: No
Print Options: No
GLP Print Options: No
Lockout: No
Global Reset: No
Accept Changes: No
The balance default parameters are listed as follows:
Legal for Trade (LFT) is a software controlled option which can
be set ON in the Readout menu. When LFT is set ON, certain
items in the Calibration, Readout and Print menus are
automatically preset and locked to permit the balance to operate
in a legal for trade application and works in conjunction with a
Lockswitch. Default setting is OFF. See default table.
LFT DEFAULT TABLE
LFT and LockswitchDefault Value
Balance Menu
Readout
Stability LevelLocked to 1
Auto ZeroLimited to OFF & 0.5
Auto CalLocked to ON
Print Current SettingsBalance will only output stable data.
When the balance is first turned ON and LFT has been previously
set ON, the initial display will indicate that LFT is ON.
The last digit on the weighing display will be highlighted in white
and is used to indicate the auxiliary digit.
Lockswitch shown in
Locked Position
Sealing Screw
Switch Plate
Lead Seal
or
Securing Sticker
Example of Sealing Method
Depending upon country regulations, additional settings will have
to be locked. Before sealing the balance, check with your local
Weights and Measure official.
3.5.12Hardware Lockswitch
Access to the various menus can be disabled setting the Lockswitch located on the PC board inside the balance to locked
position. The Lockswitch locks out all menus which have been
set to LOCKED. The default setting for the Lockswitch is UNLOCKED.
3.5.13Sealing the Balance
Certified balances have a securing sticker and additional labeling
applied at the factory. When subsequent verification is carried
out, they can be sealed either with a lead seal and wire, or with a
new securing sticker.
3.6Printing Data
Printing data requires that the Interface menu, Print Options and
GLP Print Options are set properly.
Pressing the Print button will initiate printing each time it is
pressed.
Sample printout is shown below with GLP Options turned on.
SAMPLE PRINTOUT
DATE: 19/02/03
09:56:16
9.989g
Ref:
Balance id:000001B9A925
Project Name:OHAUS
User Name: OHAUS
Page 33
Explorer Pro EN-31
4.CARE AND MAINTENANCE
4.1 Cleaning
To keep the balance operating properly, the housing and pan should be kept clean and free from foreign material. If necessary, a
cloth dampened with a mild detergent may be used. Keep calibration masses in a safe dry place.
4.2Troubleshooting
SYMPTOM
Unit will not turn on.
Incorrect weight reading.
Cannot display weight in desired unit.
Cannot access desired application.
Unable to change menu settings.
RS232 interface not working.
PROBABLE CAUSE(S)
Power cord not plugged in or properly
connected to balance.
Balance was not tared before weighing.
Balance out of level.
Balance not properly calibrated.
Desired unit not enabled.
Desired application mode is not enabled.
Menu locked.
Interface parameters not properly set up.
REMEDY
Check power cord connections.
Press >O/T< with no weight on the
pan, then weigh item.
Adjust leveling feet.
Recalibrate correctly.
Enable units in Units menu.
Enable desired application in
Application Modes menu.
Verify that Lockswitch is in the Off
position, unlock menu in the
Lockout menu.
Verify interface settings in RS232
menu correspond to those of
peripheral device.
Unstable readings.
Error message display.
Incorrect calibration
Incorrect cable being used.
Cable connections.
Vibration on table surface.
Balance not tared.
Internal calibration not adjusted properly.
Refer to Accessory list for proper
cable.
Check cable connections are
installed properly. Check correct
cable end is plugged into the
balance.
Check environmental conditions.
Close draft shield doors.
Change averaging level to a higher
setting or place balance on a stable
surface
See Error Codes list.
Tare the balance.
Perform calibration adjustments.
Page 34
EN-32 Explorer Pro
4.3Error Codes List
Error Codes List
The following list describes the various error codes that can appear on the display and the suggested remedy.
Data Errors
1.0 Transient error (hardware error, probably static discharge). If error persists, the balance must be
serviced.
1.1 Balance temperature transducer hardware error.
1.2 No data from main board.
Tare Errors
2.0 Balance is unable to stabilize within time limit after taring. Environment is too hostile or balance needs
recalibration.
Calibration Errors
3.0 Incorrect or no calibration mass used for calibration. Recalibrate with correct masses.
RS232 Errors
4.4 RS232 buffer is full.
User Errors
7.0 User entry out of bounds.
7.2 Number outside of display capacity.
Over-Under Load Errors
8.0 Hardware error causing an internal weight signal which is too low. Check if pan is off. If not, the balance
must be serviced.
8.1 Hardware error caused by an internal weight signal which is too high. Check load on the pan which may
be excessive. If error persists, the balance must be serviced.
8.2 Power-on load out of specification (LFT only)
8.3 Rated capacity exceeded. Remove excessive weight from pan.
8.4 Underload condition on balance. Check that the proper pan is installed.
8.5 Auto Cal weight internal sensor indicated its weight on the pan.
CheckSum Errors
9.1 Bad factory checksum. If error persists, have the balance serviced.
9.2 Bad factory checksum. If error persists, have the balance serviced.
9.3 Bad factory checksum. If error persists, have the balance serviced.
9.4 Auto Cal data failed checksum. This failure will disable access to the Auto Cal
9.5 Factory calibration data failed checksum.
9.6 Bad program checksum.
9.7 Bad CMOS checksum.
9.8 User calibration data failed checksum.
9.9 Factory temperature compensation data failed checksum.
feature (if installed).
Page 35
Explorer Pro EN-33
4.4Service Information
If the Troubleshooting section does not resolve or describe your problem, you will need to contact an authorized Ohaus Service
Agent. For Service assistance in the United States, please call Aftermarket, Ohaus Corporation toll-free at (800) 526-0659. An
Ohaus Product Service Specialist will be available to help you. Global contact addresses and phone numbers are accessible on
www.ohaus.com.
4.5Replacement Parts
DescriptionU.S. Part No. Global Part No.
Power Pack, 100/120 V ac US Plug (Cord set part of power pack)490202-01021202536
Power Pack, (Cord set required for UK, European and Australian)490203-01021202537
Cord Set, 230 V ac, UK Plug 76448-00 89405
Cord Set, 230 V ac, European Plug 76212-00 87925
Cord Set, 230 V ac, Australian plug 76199-01 88751
In-Use Display Cover Kit80850042
4.6Accessories
Description
Calibration Masses - ASTM Class 1 Tolerance:
20 g 49024-1180780022
50 g 49054-1180730028
100 g 49015-1180780020
200 g 49025-1180780023
500 g 49055-1180780029
1 kg 49016-1180780021
2 kg 49026-1180780024
4 kg 49046-1180780027
Security Device 470004-0180850043
Density Determination Kit 470007-0180850045
Information Settings, Selectable Environmental Settings, Selectable Auto-Print Settings, Integral Weigh
Below Hook, Contrast and Brightness Control, Protective In-Use cover
Tare range Full capacity by subtraction
Stabilization time (s)4
Calibration External / Internal
Display Type LCD Dot Matrix w/CCFL Backlight
Display Size (in/cm) 2.5 x 4.7 / 64 x 120
Pan size (in/mm) 3.5/9
Operating temperature range:Non LFT w/internal calibration 10°C to 40° C / 50°F to 104° F
w/o internal calibration 10°C to 30°C / 50°F to 86° F
Tare rangeFull capacity by subtraction
Stabilization time (s) 3
Operating temperature range:Non LFT w/internal calibration 10°C to 40°C / 50°F to 104° F
Information Settings, Selectable Environmental Settings, Selectable Auto-Print Settings, Integral Weigh
Below Hook, Contrast and Brightness Control, Protective In-Use cover
All others 10°C to 30°C / 50°F to 86°F
Plug configuration for US, Euro, UK, Japan & Australia
* Moveable FineRange
TM
** Balances with Auto Cal are equipped with a 6.8 in. x 6.8 in. / 17.2 cm x 17.2 cm Pan and Windshield.
*** Units availability is country dependent.
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the
warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be
defective at no charge, provided that the product is returned, freight prepaid, to Ohaus.
This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive
materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than
Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the
authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for
any consequential damages.
As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for
further details.
Page 39
Explorer Pro EN-37
Index
A
AC Adapter Installation 6
Accessories 33
Admissible Ambient Conditions 35
Admissible ambient conditions 35
Ambient Temperature rang 35
Analytical Balances 35
Animal Weighing 19
Application Modes 26
Applications 12
APW 14, 16
Atmospheric humidity 35
Audible Signal 25
Auto Print 28
Auto Tare 25
Auto Zero 25
AutoCal 23
AutoCal Delta 23
Average Piece Weight 13
Languages 25
LCD Display 9
Lead Seal 30
Legal for Trade 25
Leveling the Balance 5
LFT DEFAULT TABLE 30
LFT Legal for Trade 30
LIMITED WARRANTY 37
Linearity Calibration 24
Lockout 29
Page 40
EN-38 Explorer Pro
Index
M
Menu 11
Menu Structure 11
N
Navigation 11
NEW COUNT 13
New Count Key 14
NEW REF 18
Numeric Data 28
O
OPERATION 8
Output Formats 6
Overview of Controls 8
Overview of Display Indicator 10
P
Parity Bit 27
PARTS COUNTING 13
Parts Counting 13
PARTS COUNTING-CHECK 15
PARTS COUNTING-FILL 16
PC CHECK 15
PC FILLING 16
PERCENT WEIGHING 17, 18
Percent Weighing 17
Platform Installation 4
Pollution degree 35
Power On/Off 11
Power Pack 33
Power supply voltage 35
Precision Balances 36
Print Interval 28
Print Options 28
Printer 33
Printer Cable 33
Printing Data 30
Project Name 28
R
Rear of Balance 6
REFERENCE 18
Reference 28
Replacement Parts 33
Replacement parts 33
RS232 Commands 34
RS232 Interface 6
RS232 Interface Cable 33
RS232 interface not working 31
TECHNICAL DATA 34
Time 25
Time Format 25
Troubleshooting 31
Turning on the Balance 11
U
U.S. Part No. 33
Unable to change menu settings 31
Unit will not turn on 31
Units 26
Unpacking 3
Unstable readings 31
User Calibration 24
User Calibration Weight 24
User Name 28
V
Voltage fluctuations 35
W
Warranty 36
Weigh Below Preparation 4
Weighing 12
Weighing with a Container 12
Windshield Installation 4