Ohaus E1M110 User Manual

Ohaus CorporationOhaus Corporation
Ohaus Corporation
Ohaus CorporationOhaus Corporation
29 Hanover Road Florham Park, NJ 07932-0900
INSTRUCTION MANUAL
Balances
1
Declaration of Conformity
with the directives and standards mentioned.
Konformitätserkärung
genannten Richtlinien und Normen übereinstimmen.
Déclaration de conformité
«CE» - sont conformes aux directives et aux normes mentionnées ci-après.
Declaración de Conformidad
con el distintivo ,CE’ - están conformes con las directivas y normas citadas.
Dichiarazione di conformità
contrassegnati con la marcatura “CE” - sono conformi alle direttive e norme citate.
Balance Type/Waagentyp/Type de balance/Modelo de balanza/Tipo di biliancia Explorer and Voyager
Marked with: Directive Standard
gekennzeichnet mit: Richtlinie Norm munis de la mention: Directive Norme con el distintivo: Directiva Norma contrassegnati con la Direttiva Norma marcatura:
Year of attachment of Electromagnetic compatibility EN45501:1992, EN50082-1:1992 Immunity the CE mark EU 89/336, 92/31, 93/68 EN55022:1987 Funkstörungen Jahr der ersten elektromagnetische Verträglichkeit EN45501:1992, EN50082-1:1992 Immunität Eichung EU 89/336, 92/31, 93/68 EN55022:1987 Emissions parasites Année de la premère Compatibilité électromagnétique EN45501:1992, EN50082-1:1992 Immunité vérification EU 89/336, 92/31, 93/68 EN55022:1987 Radiointerferencias Año de la primera Compatibilidad electromagnética EN45501:1992, EN50082-1:1992 Inmunidad verificación EU 89/336, 92/31, 93/68 EN55022:Verträglichkeit 1987 Radiointerferenze annodella prima Compatibilità elettromagnetica EN45501:1992, EN50082-1:1992 Immunità verifica
We, Ohaus Corporation, declare under our sole responsibility that the balance models listed below marked with “CE” - are in conformity
Wir , die Ohaus Corporation, erklären in alleiniger Verantw ortung, dass die untenstehenden W aagentypen, gekennzeichnet mit “CE” - mit den
Nous, Ohaus Corporation, déclarons sous notre seule responsabilité, que les types de balance ci-dessous cité - munis de la mention
Nostras, Ohaus Corporation, declaramos bajo responsabilidad exclusiva que los modelos de balanzas indicados a continuación -
Noi, Ohaus Corporation, U.S.A, dichiariamo sotto nostra unica responsabilità, che i tipi di bilance specificati di seguito -
EU 73/23 Low Voltage IEC1010-1 & EN60950:1992 Safety Regulations EU 73/23 Niederspannung IEC1010-1 & EN60950:1992 Sicherheitsbestimmungen EU 73/23 Basse tension IEC1010-1 & EN60950:1992 Consignes de sécurité EU 73/23 Baja tensión IEC1010-1 & EN60950:1992 Disposiciones sobre seguridad EU 73/23 Bassa tensione IEC1010-1 & EN60950:1992 Prescrizioni . di sicurezza
EU 89/336, 92/31, 93/68 EN55022:1987 Emissions
EU 90/384 NAWI EN45501:1992 Non Automatic Weighing Instruments EU 90/384 FNSW EN45501:1992 für nicht selbsttätige Waagen EU 90/384 BFNA EN45501:1992 balances à fonctionnement non automatique EU 90/384 PBNA EN45501:1992 para balanzas no automátäcas EU 90.384 BFNA EN45501:1992 per bilance a funzionamento non automatics
ISO 9001 Certificate for Ohaus Corporation.
and was awarded the ISO 9001 certificate. This certifies that Ohaus Corporation, USA, has a quality system that conforms with the international standards for quality management and quality assurance (ISO 9000 series). Repeat audits are carried out by BVQI at intervals to check that the quality system is operated in the proper manner.
ISO 9001-Zertifikat für Ohaus Corporation.
erhielt das ISO 9001 Zertifikat. Dieses bescheinigt, dass Ohaus Corporation, USA über ein Qualitätssystem verfügt, welches den internationalen Normen für Qualitätsmanagement und Qualitátssicherung (ISO 9000er-Reihe) entspricht. Anlässlich von Wiederhol-Audits durch das BVQI wird periodisch überprüft, ob das Qualitätssystem zweckmässig gehandhabt wird.
Certificat ISO 9000 pour Ohaus Corporation.
obtenu le certificat, degré ISO 9001. Celui-ci atteste que Ohaus Corporation, USA, dispose d’un système qualité correspondant aux normes internationales pour la gestion de la qualité et pour I’assurance qualité (degré ISO 9000). Des audits réguliers effectués par la BVQI vérifient si le système qualité est appliqué de facon appropriée.
Certificado ISO 9001 para Ohaus Corporation.
obtenido el certificado ISO 9001. Esto acredita que Ohaus Corporation, USA, dispone de un sistema de calidad que cumple las normas internacionales para gestión y garantfa de calidad (ISO serie 9000). Con ocasión de las inspecciones de repetibilidad por parte de la BVQI, se comprueba periódicamente si el sistema de calidad se manipula de forma correcta.
Certificato ISO 9001 per la Ohaus Corporation.
Bureau V eritas Quality International BVQI, e così fomice la dimostrazione che il suo sistema die Garanzia Qualità soddisfa i ma ssimi requisite. ll sistema della garanzia della qualità Ohaus Corporation viene verificato periodicamente dall BVQI, dando cosi evidenza di.
Ohaus Corporation, USA, was examined and ev aluated in 1994 by the Bureau V eritas Quality International, BVQI,
Die Firma Ohaus Corporation, USA, wurde 1994 durch das Bureau V eritas Quality International BVQI geprüft, und
La société Ohaus Corporation, USA, a été contrôlée en 1994 par Bureau V eritas Quality International BVQI et a
La firma Ohaus Corporation, USA, ha sido inspeccionada por la Bureau V eritas Quality International (BVQI) y ha
ll sistema di garanzia della qualità della Società Ohaus Corporation, USA è certificato ISO 9001 sin dal 1994 dall
2
Notice
Certified scales,scales used for legal applications have the general type designation E...5 / V ...5 and EU type Approval (T2914). The year of the initial verification is shown next to the CE mark. Such scales are verified in the factory and carry the "M" mark on the actual scale and the packaging. The year of the initial verification is shown next to the CE mark. If the letter M is shown against a solid background, the scale may be put into operation immediately . Should the background be partitioned and hatched, the scale must be verified at its place of use by the certified Ohaus service. If national regulations limit the duration of the validity of the verification certificate in individual countries, the end user of such a scale is personally responsible for arranging the repeat verification in good time.
Hinweise
Geeichte/eichpflichtige Waagen tragen die allgemeine T ypenbezeichnung E... 5 / V ...5. Für sie liegt eine EU Bauartzulassung v or (T2914). Das Jahr der ersten Eichung ist neben dem CE Zeichen aufgeführt. Solche W aagen sind ab Werk geeicht und tragen die Kennzeichn ung "M" auf dem Gerät selbst und auf der V erpackung. Erscheint der Buchstabe M auf vollem Grund, darf die W aage sofort in Betrieb genommen werden. Ist der Grund geteilt und schraffiert, muss die W aage am V erwendungsort durch den zertifizierten Ohaus Service ortsgeeicht werden. Sof ern gemäss den nationalen Vorschriften in den einzelnen Staaten die Güitigkeitsdauer der Eichung beschränkt ist, ist der Betreiber einer soichen Waage für die rechtzeitige Nacheichung selbst verantwortlich.
Remarques
Les balances vérifiées/admissibles à la vérification portent la désignation de modèle générale E...5 / V ... 5. Elles font l’objet d’une approbation de modèle UE (T2914). L ’année de la vérification primitive est indiquée à côté de la marque CE. Ces balances sont vérifidées d’origine et portent la marque "M" sur I’appareil lui-même et sur l’emballage, Si la lettre M apparaît sur un fond totalement vert, la balance peut être mise en service immédiatement. Si le fond est divisé et hachuré, la balance doit être vérifiée sur le lieu d’ustilisation par le service après-vente Ohaus certifié. Dans les pays où la durée de validité de la vérification est limitée par des prescriptions nationales, l’utilisateur est lui-même responsable de la vérification ultérieure d’une telle balance en temps voulu.
Notas
Las balanzas verificadas/verificables llevan la designatión general E...5 / V ...5 y cuentan con una aprobación de modelo UE (T2914). EL año de la primera verficación está indicado al lado del distintivo CE. Estas balanzas están verificadas en fábrica y Ilevan la designatión "M" sobre el propio aparato y sobre el embalaje. Cuando la letra M aparece sobre fondo sólido, la balanza se puede poner inmediatamente en funcionamiento. Si el fondo está dividido y rayado, la balanza ha de ser verificada en el lugar de uso por el sevicio técnico Ohaus certificado. Si la duración de la validez de la verificación está limitada de acuerdo con las normas de los distintos países, el propio usuario de tal balanza es responsable de la verificación posterior a su debido tiempo.
Avvertenza
Le bilance approvate hanno la denominazione del modello E... 5 / V ...5. P er esse esiste un’appprovazione CE del tipo. L ’anno delia prima verifica è indicato a fianco della marcatura CE. I tipi marcati con un contrassegno "M" su sfondo verde pieno possono essere impiegati da subito. I tipi marcati con ii contrassegno "M" su sfondo nero/ barrato diagonalmente dovranno essere verificati sul luogo d’installazione da parte d’un tecnico autorizzato dal Servizio Assistenza Ohaus o ispettore dell’Ufficio Metrico. Queste bilance sono state verificate in fabbrica e recano il contrassegno "M" sull’apparecchio stesso, e sull’imballo. É obbligo dell’untente denunciare la detenzione dello strumento all’ufficio metrico competente per territorio e sottoporio alia prescritta verifica periodica come da disposizioni ministeriali.
3
NOTE: THIS EQUIPMENT HAS BEEN TESTED AND FOUND T O COMPL Y WITH THE LIMITS FOR A CLASS A DIGITAL
DEVICE, PURSUANT T O P AR T 15 OF THE FCC RULES . THESE LIMITS ARE DESIGNED T O PRO VIDE REASONABLE PRO TECTION AGAINST HARMFUL INTERFERENCE
WHEN THE EQUIPMENT IS OPERATED IN A COMMERCIAL ENVIRONMENT. THIS EQUIPMENT GENERATES, USES, AND CAN RADIATE RADIO FREQUENCY ENERGY AND, IF NO T INSTALLED AND USED IN ACCORDANCE WITH THE INSTR UCTION MANU AL, MA Y CA USE HARMFUL INTERFERENCE TO RADIO COMMUNICATIONS. OP­ERATION OF THIS EQUIPMENT IN A RESIDENTIAL AREA IS LIKELY TO CAUSE HARMFUL INTERFERENCE IN WHICH CASE THE USER WILL BE REQUIRED TO CORRECT THE INTERFERENCE AT HIS OWN EXPENSE.
THIS DIGITAL APPARATUS DOES NOT EXCEED THE CLASS A LIMITS FOR RADIO NOISE EMISSIONS FROM DIGITAL APPARATUS AS SET OUT IN THE INTERFERENCE-CA USING EQ UIPMENT STAND ARD ENTITLED “DIGI­T AL APPARA TUS”, ICES-003 OF THE DEPARTMENT OF COMMUNICATIONS.
CET APPAREIL NUMERIQUE RESPECTE LES LIMITES DE BRUITS RADIOELECTRIQUES APPLICABLES AUX APPAREILS NUMERIQUES DE CLASSE A PRESCRITES DANS LA NORME SUR LE MATERIEL BROUILLEUR : “APP AREILS NUMERIQUES”, NMB-003 EDICTEE PAR LE MINISTRE DES COMMUNICA TIONS.
Unauthorized changes or modifications to this equipment are not permitted.
4

TABLE OF CONTENTS

OV ERVIEW OF CONTR OLS ........................................................................................................................ 7
OVER VIEW OF DISPLA Y INDICATORS ......................................................................................................8
1. GETTING TO KNOW Y OUR BALANCE .............................................................................................................. 9
1.1 Introduction ....................................................................................................................................................... 9
2. INST ALLA TION ................................................................................................................................................... 9
2.1 Unpacking and Checking the Standard Equipment ............................................................................................. 9
2.2 Selecting the Location ...................................................................................................................................... 1 0
2.3 Setting Up and Leveling the Balance ................................................................................................................ 10
2.4 Connecting Power ............................................................................................................................................. 1 1
3. OPERA TING YOUR BALANCE ......................................................................................................................... 12
3.1 The Menu (Basic Settings of the Instrument) ................................................................................................... 12
3.2 T urning On the Balance..................................................................................................................................... 13
3.3 Calibr ation ..................................................................................................................................................... 13
3.3.1 Internal Calibration (InCALTM) .................................................................................................................. 1 5
3.3.2 Span Calibration ..................................................................................................................................... 1 6
3.3.3 User Calibration ...................................................................................................................................... 17
3.3.4 Linearity Calibration ................................................................................................................................ 18
3.3.5 Calibration Test ....................................................................................................................................... 1 9
3.3.6 Calibration GLP Printout ......................................................................................................................... 20
3.4 Weighing ..................................................................................................................................................... 21
3.5 Percent Weighing .............................................................................................................................................. 2 2
3.6 Parts Counting.................................................................................................................................................. 23
3.7 Animal Weighing ............................................................................................................................................... 24
3.8 Weigh Below..................................................................................................................................................... 2 5
3.9 Printing Data..................................................................................................................................................... 2 5
4. SETTING UP Y OUR BALANCE........................................................................................................................ 26
4.1 Setting Date and Time ...................................................................................................................................... 26
4.2 Readout ..................................................................................................................................................... 27
4.3 Good Laboratory Practices (GLP) Data............................................................................................................. 28
4.4 Good Laboratory Practices (GLP) Set .............................................................................................................. 28
5
TABLE OF CONTENTS (Cont.)
4.5 Print ..................................................................................................................................................... 29
4.6 RS 232 ..................................................................................................................................................... 31
4.7 Legal for T rade (LFT)......................................................................................................................................... 32
4.8 Mode ..................................................................................................................................................... 33
4.9 Units ..................................................................................................................................................... 3 4
4.10 Global ..................................................................................................................................................... 3 4
4.11 Custom Unit ............................................................................................................................................... 36
4.12 Menu Lock-Out Protection .......................................................................................................................... 38
5. CARE AND MAINTENANCE ............................................................................................................................ 39
5.1 Troubleshooting................................................................................................................................................. 39
5.2 RS232 Interface............................................................................................................. ................................... 40
5.3 Error Codes List................................................................................................................................................ 42
5.4 Information Messages ...................................................................................................................................... 42
5.5 Service Information........................................................................................................................................... 44
5.6 Replacement Parts ........................................................................................................................................... 44
5.7 Accessories ..................................................................................................................................................... 44
5.8 Specifications................................................................................................................................................... 45
6
OVERVIEW OF CONTROLS
12
1
2
3
4
5
No. Designation Function
1 I Po wer on off b utton. 2 Mode button Selects standard weighing, percent, parts counting and animal weighing modes. 3 Units button Selects weighing units.
6
8
9 7 10
11
4 Setup button Selects various submenus: calibration, date, time , readout, GLP data, GLP set, print,
RS232, LFT, function, units, global, custom.
5 >O/T< button When pressed, sets balance to zero.
6 button When pressed, tra v els up through submenus.
7 button When pressed, tra v els down through submenus.
8 button When pressed, tr a v els to the left through men us.
9 button When pressed, tr a v els to the right through menus.
10 Enter/Print b utton When in menus, selects item on display, otherwise prints data. 11 Leveling f eet Used to le v el the balance. 12 Leveling indicator Indicates leveling position of the balance (located at rear of balance).
7
OVERVIEW OF DISPLAY INDICATORS
6
5 4
3 2
1
No. Function
1 Use (Pointer Group) key to change - used to
prompt the user while navigating through the menu system.
2 Standard (7) segment numeric characters. Seven
characters are availab le and are used f or display ing weight values.
3 Center of Zero - Indicates Center of Zero in
Legal For Trade (LFT) mode. 4 P - This symbol is not used. 5 T - This symbol is not used. 6 Stable - Indicates that the measured value has
become stable. 7 B/G - This symbol is not used. 8 Change - Is displayed together with Mode, Units
or Setup signifying that a change to balance settings is being performed.
8
7
9
1718
No. Function
12 (British Flag) - Are (14) segment alpha numeric
13 Custom - The user can input a factor to meet
14 Symbols for weighing modes, include:
10
16
characters. Seven characters are used to present features and functions.
unique unit measure applications.
% - Percent weighing. PC - Parts counting. GN - Grain. N - Newtons. m - Mommes. g - Grams. kg - Kilograms. dwt - Penn yweight. lb - Pounds. oz - Ounces. ct - Carats. t - Taels. Taels are available in three types;
Hong Kong, Singapore, and Taiwan.
ti - Tical. oz t - Ounces troy.
11
12 13
14
15
9 Mode - Is displayed when the Mode button is
pressed. Allo ws the user to know what area of the balance menu is being addressed.
10 Units - Is displayed when the Units button is
pressed. Allows the user to know what area of the balance menu is being addressed.
11 Setup - Is displayed when the Setup button is
pressed. Allows the user to know what area of the balance menu is being addressed.
15
15 press >O/T< -This symbol is not used.
16 - Differentiated digit for LFT. 17 press Enter - Used as a prompt to the user to
press the Enter button. The menu item display ed is accepted/selected.
18 NET - This symbol is not used.
8
1. GETTING TO KNOW YOUR BALANCE
Please read through this section carefully, as it contains important information for safe and economical operation of your Explorer Balance.
1.1 Introduction
Thank you for deciding to purchase an Explorer Balance from Ohaus. Thanks to a new modular design, your Explorer Balance lets you adapt the balance to your changing needs. Remote displays , upgr aded displays which can be table, wall or tow er mounted are a vailable as accessories. It off ers a high level of operating convenience and useful functions to make accurate measurements. A unique LCD panel has a large 7 digit, 7 segment display which indicates the weight value of an item being measured and a 7 digit British Flag display (14 segments) which spells out items selected in the submenus. In addition, the display contains English words to indicate the status of the balance. Arro w indicators in the display prompt the user as to what panel keys are to be pressed to initiate a change . Panel controls are clearly marked as to their function with large Tare buttons on either side of the front panel. Opera­tion and setup of the balance is straightforward and easy. The Explorer Balance is a v ailab le in full scale capacities of 12,000 grams, 22,000 grams and 32,000 grams. Legal for Trade v ersions are available.
Behind your instrument stands OHAUS , a leading manufacturer of precision scales and balances. An Aftermarket Department with trained instrument technicians is dedicated to provide you with the fastest service possible in the event y our instrument requires servicing. OHAUS also has a Customer Service Department to answer any inquiries regarding applications and accessories.
To ensure you mak e full use of the possibilities off ered b y your Explorer balance, we advise you to read through these operating instructions very carefully .
2. INSTALLATION
2.1 Unpacking and Checking the Standard Equipment
Open the package and remove the instrument and the accessories. Check the completeness of the delivery. The following accessories are part of the standard equipment of your new Explorer balance.
Equipment 12,000g 22,000g 32,000g
••
AC Power Adapter
••
••
Instruction Manual
••
••
Warranty Card
••
••
Weigh Below Hook
••
• Remove packing material from the instrument.
• Check the instrument for transport damage. Immediately inform your Ohaus dealer if you have complaints or parts are missing.
• Store all parts of the packaging. This pac kaging guar antees the best possible protection for the transport of your instrument.
✓ ✓
✓ ✓
✓ ✓
✓ ✓
✓ ✓
✓ ✓
✓ ✓
✓ ✓
✓ ✓
9
2.2 Selecting the Location
The balance should always be used in an en vironment which is free from e xcessive air currents, corrosives, vibration, and temperature or humidity extremes. These factors will affect display ed w eight readings.
DO NOT install the balance:
Next to open windows or doors causing drafts or rapid temperature changes.
Near air conditioning or heat vents.
Near vibrating, rotating or reciprocating equipment.
Near magnetic fields or equipment that generates
magnetic fields.
On an unlevel work surface.
Allow sufficient space around the instrument for
ease of operation and keep awa y from radiating heat sources.
2.3 Setting Up and Leveling the Balance
Exact horizontal positioning and stable installation are prerequisites for repeatab le results . To compensate small irregularites or inclinations at the location, the instrument can be leveled.
Leveling Indicator In Rear
Leveling Feet
For exact horizontal positioning, the balance is equipped with a level indicator located at the rear of the balance and leveling f eet located at the front and rear of the balance.
Position the balance in the intended operating location. Adjust the leveling f eet
10
2.4 Connecting Power
Auxiliary Data Connection
RS232 Connection
AC Adapter Connection
Left Side of Balance
Connect the AC Adapter supplied to the three pin con­nector located at the side of the balance.
The balance is now ready for operation.
11
3 OPERATING Y OUR BALANCE
3.1 The Menu (Basic Settings of the Instrument)
The Explorer balance has three basic menus; each is selected b y front panel b uttons marked Mode, Units and Setup. Mode Button
The Mode button, when pressed, permits the selection of four weighing modes which are: weigh, percent, count and animal weighing. These modes are controlled by an on or off selection made in the Setup menu under the submenu
Mode as displayed. Units Button
The Units button, when pressed, allows the balance to display v alues in selected measuring unit.
Setup Button
The Setup button, when pressed, allows entry into thirteen submenus which allows you to set the balance for specific operating parameters. Each of the thirteen submenus contain settings which are user selectable. The table below illustrates the various submenus and the functions which are selectable . The items shown on the men u, which are bolded, are the factory default settings. In other words, if you did not enter the Setup men u, the balance would func­tion in the basic manner shown by the various settings which are bolded. The setup submenus shown below are arranged in the order as displayed in the balance .
START
CALIBRATION
InCALTM Calibration (No Lock)* Span Calibration User Calibration Linearity Calibration Calibration Test (No Lock) Lock ON/OFF Exit
* (If option is installed.)
READOUT
Filter -0-, -1-, -2-, -3­Stable .5d, 1d, 2d, 5d Auto 0 OFF, .5d, 1d, 5d Lock ON/OFF Exit
PRINT
Auto Print OFF, Cont, Inter, On Stb Interval Enter 1 seconds Stable ON/OFF Numeric ON/OFF GLP Cont ON/OFF GLP Tare ON/OFF Reference ON/OFF Lock ON/OFF
Exit
MODE
Weigh ON/OFF Percent ON/OFF Count ON/OFF Animal ON/OFF Lock ON/OFF Exit
SETUP SUBMENUS
DATE
Type m/d/y Set Date Exit
GLP DATA
User Number Enter 7 digits Project Number Enter 7 digits Lock ON/OFF Exit
RS232
Baud 300, 1200, 2400, 4800, 9600 Parity None, -E-, -Odd-, -0-, -1­Data 7,8 Stop 1, 2 Lock O N/ OFF Exit
UNITS
Units ON/OFF Lock ON/OFF Exit
12
TIME
Type 12 Hour/24 Hour Set Time Adjust Enter xx seconds Exit
GLP SET
Time ON/OFF Balance ID ON/OFF User Number ON/OFF Project Number ON/OFF Difference ON/OFF Name ON/OFF Lock ON/OFF Exit
LEGAL FOR TRADE
LFT LOCK ON/OFF
(Locked using switch)
GLOBAL
List NO/YES Reset NO/YES Version Software No. Lock ON/OFF
CUSTOM
Factor Enter 7 digits Exponent Enter +3 to -3 LSD Enter .5 to 100 Lock ON/OFF Exit
3.2 Turning On the Balance
The Explorer balance is ready to operate after the installation procedures are performed. When the balance is first turned on and it completes its checks, it can be used to weigh or tare materials without setting the menus.
It is recommended that you read this manual carefully and set the balance to operate for y our specific applications using the procedures in Chapter 4 Setting up Your Balance and calibrate the balance before using.
The balance is a high precision instrument and will give you years of service if kept clean and handled carefully. If you have an y prob lems oper ating the instrument or require additional information, please feel free to contact our Product Service Department at (800) 526-0659.
In this section, you will enter the menu for the first time. Do not worry if you are unfamiliar with the function of the buttons on the panel, the display pro vides the necessary coaching as you go along.
Power On/Off
To turn the balance ON, press the ON/OFF button (circled button with an I inside) located at the upper left­hand corner of the panel once. To turn OFF, press button again.
Stabilization
Before initially using the balance, allow time for it to adjust to its new environment. The balance only requires to be plugged in to warm up. Recommended warm up period is thirty (30) minutes. The internal circuits of the balance are powered whenev er it is plugged into a po w er source.

Calibration

Refer to paragraph 3.3 and calibrate the balance bef ore proceeding.
3.3 Calibration
Explorer balances offer a choice of five calibration methods: Internal Calibration (InCALTM), Span Calibration, User Calibration, Linearity Calibration, and CalTestTM.
User
Linearity
Cal Test
TM
InCal
Span
- Span calibration ensures that the balance reads correctly within specifications using two
- User calibration is a method where the balance can be calibr ated using a mass of kno wn
- Calibration test allows the stored calibration data to be tested against the current mass being
Internal calibration (InCALTM) of the balance is accomplished by an internal mass. (If option is installed.)
NOTE: When CAL NOW is display ed, the balance should be calibr ated bef ore weighing again.
weight values: zero and a weight value at incremental values of full capacity and or 100% of the balance’s full capacity.
value and by entering that numeric value into the balance.
- Linearity calibration minimizes deviation between actual and display ed weights within the balance’s weighing range . Three w eight v alues are used: zero, a weight value at midpoint of the balances weighing range, and a weight value at or near the balance’s specified capacity.
used for the test.
Lock
- Can be set on or off. When set on, Span, User and Linearity are locked out and cannot be used.
13
3.3 Calibration (Cont.) Calibration Menu Protection
NOTES:
• Calibration may be lock ed out to prevent unauthorized personnel from changing calibration. If calibration has been locked out, you can only access Internal Calibration (InCALTM) when installed and Cal Test.
• To lock out calibration menu, after calibration, refer to the section titled Menu Lock-Out Protection.
Calibration Masses
Before beginning calibration, make sure masses are availab le. If you begin calibration and realize calibration masses are not availab le, exit the menu. The balance will retain previously stored calibration data. Calibration should be performed as necessary to ensure accurate weighing. Masses required to perform the procedures are listed in the adjacent table.
NOTE:
Any of the calibration modes can be terminated
time
by pressing either the Mode, Units or Setup buttons .
at any
CALIBRA TION MASSES
LINEARITY SPAN ONLY CAPACITY MASSES MASSES
12,000g 5,000g/10,000g 10,000g 22,000g 10,000g/20,000g 20,000g 32,000g 15,000g/30,000g 30,000g
It is recommened that masses must meet or exceed ASTM Class 1 Tolerance. Calibration masses are available as accessories.
14
3.3.1 Internal Calibration (InCALTM)
On Explorer balances equipped with the InCalTM feature, calibration can be accomplished using an internal calibration mass. When the balance requires calibration, a screen prompt of CAL NOW appears. Internal calibration can be performed at any time providing the balance has warmed up to operating temperature.
Procedure
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYPE is displa yed.
NOTE: DO NOT DISTURB THE BALANCE DURING CALIBRA TION.
• Press Enter button, CAL TYPE InCAL is display ed.
• Press Enter button, INCAL is displayed. NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remov e the w eight from the pan). The balance automatically resumes calibration.
After a few seconds, CAL SET is displa yed (the internal weight is positioned and then removed), the display then returns to WEIGH mode.
15
3.3.2 Span Calibration
Span calibration utilizes two calibration points, one at zero and the other at full span or incremental v alues starting at approximately 25% of full capacity. As an example, an 12kg balance will accept either 4kg, 6kg, 8kg, or 10kg f or span calibration. Values which are below or in between will not be accepted and the balance will displa y its maxim um capacity. Sample display illustrates an 12kg balance
Procedure
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYPE is display ed.
• Press Enter button, CAL TYPE InCAL is display ed.
• Press button to select SPAN calibration, CAL TYPE SPAN is display ed.
• Press Enter button, WORKING is display ed.
NOTE: If a weight is left on the pan, the balance will display CLR PAN (remov e the w eight from the pan). The balance automatically resumes calibration.
• Display changes to PUT WT 10000g.
• Place specified calibration mass on pan. NOTE: For an 12kg balance, either 4kg, 6kg, 8kg, or
10kg can be used for span calibration. The PUT WEIGHT message indicates the calibration mass that is on the pan.
• Press Enter button, WORKING is display ed. After a few seconds CAL SET is displayed, the display then returns to WEIGH mode.
Span calibration is completed.
• Remove calibration mass from the pan.
16
3.3.3 User Calibration
User calibration is used when it is desired to calibrate the balance using a mass of known value. To use this calibra­tion feature, proceed as f ollows:
Procedure
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYPE is displa yed.
• Press Enter button, CAL TYPE InCAL is dis ­played.
• Press button twice to select USER calibration, CAL TYPE USer is displa y ed.
• Press Enter button, the display indicates the last calibration mass value which was entered with the
first digit flashing. (Sampleillustrates 6000g).
• Press and or and enter the desired mass value. This number must be at least
25% of the full span value.
(FLASHING)
• Press Enter button, WORKING is displayed.
NOTE: If a weight is left on the pan, the balance will display CLR PAN (remov e the w eight from the pan). The balance automatically resumes calibration.
• Display changes to PUT WT 6000g.
• Place specified calibration mass on pan.
• Press Enter button, WORKING is displayed. After a few seconds CAL SET is displayed, the display then returns to WEIGH mode.
User calibration is completed.
• Remove calibration mass from the pan.
17
3.3.4 Linearity Calibration
Linearity calibration utilizes three calibration points, one at zero , center span and full span. This method minimizes deviation between actual and displa yed weights within the balance's w eighing range . Three weight values are used; zero, a weight v alue at midpoint of the balance's weighing range and a w eight v alue at or near the specified capacity. Sample display illustrates an 12kg balance.
Procedure
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYPE is displa yed.
• Press Enter button, CAL TYPE InCAL is display ed.
• Press button three times to select LIN calibra­tion, CAL TYPE Lin is displa y ed.
• Press Enter button, WORKING is display ed.
NOTE: If a weight is left on the pan, the balance will display CLR PAN (remo v e the weight from the pan). The balance automatically resumes calibration.
• Display changes to PUT WT 5000 g. The displayed weight is half the capacity of the balance.
• Place specified calibration mass on pan.
• Press Enter button, WORKING is displayed. After a few seconds displa y changes to PUT WT 10000 g.
• Place specified calibration mass on pan.
• Press Enter button, WORKING is displayed. After a few seconds CAL SET is displayed, the display then returns to WEIGH mode.
Linearity calibration is completed.
• Remove calibration mass from the pan.
18
3.3.5 Calibration T est
Calibration test feature allows a check of a kno wn calibr ation mass against the last stored calibration information in the balance. Sample display illustrates an 12kg balance.
Procedure
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYPE is displa yed.
• Press Enter button, CAL TYPE InCAL is display ed.
• Press button to select CALTEST calibration, CALtESt is displayed.
• Press Enter button, WORKING is displa yed.
NOTE: If a weight is left on the pan, the balance will display CLR PAN (remo v e the weight from the pan). The balance automatically resumes calibration.
• Display changes to PUT WT 10000 g.
• Place specified calibration mass on pan.
• Press Enter button, WORKING is display ed. After a few seconds, WEIGH DIFF is displayed. The display now indicates the actual difference in weight between what value was just placed on the pan and the previous weight value which was stored in the bal­ance. After approximately 8 seconds,the display returns to the WEIGH mode.
• Remove calibration test mass from the pan.
19
3.3.6 Calibration GLP Printout
If any option in the GLP Set Menu is turned On, GLP automatically prints data after calibration is completed.
Span Calibration Printout
When performing Span calibration with all GLP options turned on, a printout is automatically made after the calibration is completed.
InCAL
When performing InCALTMcalibration with all GLP options turned on, a printout is automatically made after the calibration is completed.
TM
Calibration Printout
- - - - - SPAN CAL - - - - - -
7/01/97 1:00:00 PM Bal Id 1234 Cal: 10000.0g Old: 10000.0g Dif: 0.0g
Wt. Ref ......................................
USER NO 2056853 PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - - INCAL - - - - - ­7/01/97 1:00:00 PM Bal Id 1234 Cal: 10000.0g Old: 10000.0g Dif: 0.0g
Wt. Ref ......................................
USER NO 2056853 PROJ NO 100012
Name........................................
Linearity Calibration Printout
When performing a Linearity calibration with all GLP option turned on, a printout is automatically made after the calibration is completed.
Calibration T est Printout
When performing a Calibration Test with all GLP options turned on, a printout is automatically made after the calibration is completed.
- - - - - END - - - - -
- - - - - LIN CAL - - - - - ­7/01/97 1:00:00 PM Bal Id 1234 Cal: 10000.0g Old: 9999.8g Dif: 0.2g
Wt. Ref ......................................
USER NO 2056853 PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - - CAL TEST - - - - - ­7/01/97 1:00:00 PM Bal Id 1234 Cal: 10000.0g Act: 10000.2g Dif: 0.2g
Wt. Ref ......................................
USER NO 2056853 PROJ NO 100012
Name........................................
20
- - - - - END - - - - -
3.4 Weighing
NOTE: The Explorer balances are shipped with grams only enabled. When the balance is to be used with other
Type Approv ed/Legal for Trade units of measure, the desired unit must be enabled. Ref er to paragr aph 4.9 to enable other measuring units. For weigh below measurements, ref er to paragr aph 3.8.
Procedure
• Press to rezero the display.
• Press Units button to select measuring unit.
• Press or button for desired measuring unit.
•Press Enter button, balance is now ready for weigh ing.
• Place the object(s) or material to be weighed on the pan. Example illustrates a 200 gram weight.
• Wait for the stability indicator to appear before read­ing the w eight.
Zero/Tare
When weighing material or objects that must be held in a container, taring stores the container weight in the bal­ance’s memory, separate from the weight of the material in the container .
(
Example
Container 200g)
Procedure
• Press with no load on the pan to set the balance to zero.
• Place an empty container on the pan. Its weight is displayed.
• Press the display blanks until stable weight readings are received, then indicates zero . The container’s weight is stored in memory.
• Add material to the container. As material is added, its net weight is display ed.
• Removing the container and material from the plat­form will cause the balance to display the container’s weight as a negative number. The tared weight will
remain in memory until is pressed again or the balance is turned off.
• Pressing resets the balance to zero .
(
Example
Material 1620g)
21
3.5 Percent W eighing
Percent W eighing is load on the balance, then view other loads as a percentage of the reference. The load you place on the pan as a reference may be displayed as any percentage you select from 5% to 100% (in 1% increments). One hundred percent does not necessarily have to represent the reference load. Subsequent loads, displa y ed as a percentage of the reference are limited only by the capacity of the balance. The default setting is Reference 100%. Refer to paragraph
4.8 to enable percent weighing.
enabled only
when Percent is turned ON. Percent weighing permits you to place a reference
Procedure
• Press the Mode button.
• Press or button until PERCENT is dis­played.
• Press Enter button, PUT>PAN 100% is displayed. If a container is used, the balance can be tared at this point.
• Put the reference load on the pan.
• Press or button and select ref erence weight percentage (P ercent Range 5 to 100). Hold
button down f or f ast change .
• Press Enter button to save setting, W ORKING is displayed ... calculating reference w eight.
Balance displays reference weight for five seconds in selected measuring unit, then displays the percentage.
• Remove the reference weight from the pan and re­place it with another load. The second load is dis­played as a percentage of the ref erence .
NOTE: The PERCENT display (number of digits) is a function of the accuracy of the balance and the size of the reference weight. The displa y examples were with a 300g mass used with an 12 kilogram balance.
22
3.6 Parts Counting
Parts Counting is In the parts counting mode, the balance displays the quantity of parts you place on the pan. Since the balance determines the quantity based on the average weight of a single part, all parts must be reasonably uniform in weight.
enabled only
when Count is turned ON in the Mode submenu and selected with the Mode button.
Procedure
• Press the Mode button.
• Press or button until COUNT is displayed.
• Press Enter button to save setting, PUT>PAN 100 PC is displayed.
• Press >O/T< if taring is required.
• Press or button and select sample size . Sample size is 5 to 1000 pieces.
• Place sample size on the pan.
• Press Enter button to continue, display indicates WORKING.
Balance displays the reference w eight of an individual piece part for five seconds and then displays the total number pieces on the pan.
• Remove the sample and place parts to be counted on the pan. Balance displays number of pieces.
Update
Update is a function which permits placing additional samples which are greater than the value of the original sample but less than three times the value. This action increases the accuracy of the measurement.
• Place sample on the pan which is at least one but not more than three times the original sample size.
• Press Mode button, COUNT is displayed.
• Press Enter button, UPDATE is displayed.
• Press Enter button, WORKING is displayed then the reference weight f ollow ed b y the new sample siz e .
23
3.7 Animal Weighing
Animal Weighing is
enabled only
when Animal is turned ON in the Mode submenu.
Procedure
• Press the Mode button.
• Press or button until ANIMAL is displayed.
• Press Enter button to continue, LEVEL is displayed.
• Press or button to change animal weighing level, 0, 1, 2 or 3. 0 level represents an inactive
subject, 3 is used for a very active subject.
• Press Enter button to continue, AU T O is displayed.
• Press or button to select AUTO ON or OFF.
• Press Enter button to continue.
When the AUTO function is set ON, different subjects can be weighed one after another without pressing any buttons. When the balance displays READY, simply place subject on pan.
Start Animal Cycle
• Place animal container if used on pan.
• Press to tare the container .
• Place subject on pan.
• The animal cycle will automatically start if AUTO was set to ON.
• Press Enter button to start animal cycle if AU TO was set to OFF.
During Animal Cycle
• Display shows countdown to AW0.
Completed Animal Cycle
• Balance displays weight until specimen is removed from the pan.
24
3.8 Weigh Below
The Explorer balance is equipped with a weigh below hook which can be attached to the bottom of the balance. To install the hook, remove the protective cap from the bottom of the balance as sho wn in the figure below. Screw the weigh below hook into the bottom of the balance.
3.9 Printing Data
Printing data to an external computer or printer requires that the communications parameters in the Setup menu, Print options and communication parameters be set first. Refer to page 23 Print menu settings and page 25 for RS232 communication settings.
Procedure
• Press the Print button. Printing to an external printer or computer will occur each time the Print button is
pressed unless autoprint feature is turned on in which case printing can occur in a continuous fashion, at specified intervals or each time a stable reading is achieved.
Sample printout is shown below with time turned on.
SAMPLE PRINTOUT
1/01/97 12:01:37 AM
429.5 g
For a re vie w of of printing samples, refer to Section 4 Setting Up Your Balance. What is printed is controlled by the GLP Set Menu and the selection of GLP Cont. or GLP Tare in the Print Menu.
25
4. SETTING UP YOUR BALANCE
4.1 Setting Date and Time
Your Explorer balance provides date and time data which can be viewed on a computer or printed out on an external printer. When y ou put y our ne w instrument into operation f or the first time , y ou should enter the current date and the time. These settings are retained as long as the balance remains connected to a power source.
Date
Date is a feature which enables the balance to be set to a U.S.A. date standard or European date standard. U.S. standard has the month, date, followed b y the y ear, each separated by (/) in the printout. The European date standard has the day first, followed b y the month and then the year; each separated by a period. The def ault setting is U.S.A. Standard.
Procedure
• Press the Setup button, CAL is displayed.
• Press or button and select Date from the menu.
• Press Enter button, TYPE is display ed.
• Press Enter button, SET M d y, d M y, y M d, M y d, y d M, or d y M is displayed.
• Press or button and select type of date.
• Press Enter button, SAVED is displayed, then SET is displayed.
• Press Enter button, first digit of date is flashing.
• Using arrow buttons, enter the correct date.
• When the correct date is entered, press Enter button, SA VED displa ys momentarily and EXIT appears.
• Press Enter button, balance returns to a weighing mode.
Time
Time is a feature which enables the balance to be set to the current time in either U.S.A. standards (12 hour periods) or European/Military standards (24 hour periods). The default setting is U .S.A. Standard.
Procedure
• Press the Setup button, CAL is displayed.
• Press or button and select Time from the menu.
• Press Enter button, TYPE is display ed.
• Press Enter button, TYPE 12 hr is displayed.
• Press or button and select 12 hr or 24 hr.
• Press Enter button, SAVED is displa y ed momentarily then SET is displayed.
• Press Enter button,SET with time is flashing.
• Using arrow buttons, enter the correct time.
• When the correct time is entered, press Enter button, SA VED displa ys momentarily and EXIT appears.
• Press Enter button, balance returns to a weighing mode.
Adjustments up to +60 seconds a month can be made to the balance internal clock. Repeat the first seven steps, ADJUST is display ed. Using arrow buttons,
enter time correction and press Enter button.
26
4.2 Readout
The Readout menu is used to adapt the balance to environmental conditions. It contains four submenus: Stable, Auto 0, Filter, Loc k and Exit. Lock enables you to program balance parameters and to loc k the settings .
Procedure
To select any of the items in the Readout menu, proceed as follows:
• Press the Setup button, CAL is displayed.
• Press or button until READOUT is dis­played.
• Press Enter button to continue.
• Press or button until either STABLE, AU TO 0, FILTER, LOCK or EXIT is displayed.
• Press Enter button to continue.
• Press or button and select the desired menu setting.
• Press Enter button to save settings, SAVED is dis­played.
• Press or button to continue or EXIT.
Stability
The stability range specifies the weighing results and must be within a preset tolerance limit for a certain time to turn the stability indicator ON. When a displayed w eight changes be y ond the allowable range, the stability indicator turns OFF, indicating an unstable condition. Factory default setting is sho wn in bold type .
• Press Enter button to continue.
.5 d Smallest range: stability indicator is ON only
when displayed weight is within .5 divisions.
1 d Reduced range. 2 d Normal range. 5 d Largest range, stability indicator is ON even though displa y ed weight
changes slightly .
When the RS232 interface is configured to print stable data only, the stability range also gov erns data output. Dis­played data will only be output if it is within the selected stability r ange .
Auto-Zer o
Auto-Zero minimizes the eff ects of temperature changes and shift on the z ero reading. The balance maintains the zero display until the threshold is e xceeded. F actory default setting is sho wn in bold type.
OFFT urns Auto-Zero OFF.
.5 d Sets threshold to .5 divisions.
1 d Sets threshold to 1 division. 3 d Sets threshold to 3 divisions.
Filter
Filter compensates for vibration or excessiv e air currents. Def ault settings are shown bold.
-0- reduced stability, fastest stabilization time
-1- normal stability , normal stabilization time
-2- more stability, slow stabilization time.
Lock
Lock ON/OFF can only be changed when the hardware Lockswitch is set OFF/unlocked. A menu is locked when the menu lock is set ON and the Lockswitch is ON. Loc k when selected and turned on, locks all of the entries made under the Readout menu. In the locked condition, items may be look ed at b ut not changed in the menu. When set off, entries may be changed. OFF is the default setting.
-3- maximum stability, slowest stabilization time.
27
4.3 Good Laboratory Practices (GLP) Data
The GLP Data submenu enables the storage of a user identification number (7 digits) and/or a project number (digits). When entered into the balance, the identification number and project number are a v ailab le when printing providing they are turned on in the GLP Set submenu. A lock setting is also available which locks in the user identifi­cation and project number .
Procedure
To select any of the items in the GLP Data menu, pro­ceed as follows:
• Press the Setup button, CAL is displayed.
• Press or button until GLP DATA is displayed.
• Press Enter button to continue.
• Press or button until either USER NO, PROJ NO, LOCK or EXIT is displayed.
• Press Enter button to continue.
• Press or buttons as directed by the display and enter a 7 digit number for the user ID number.
• Press Enter button to save setting.
• Press or button until either USER NO, PROJ NO, LOCK or EXIT is displayed.
• Press Enter button to continue.
• Press or buttons as directed by the display and enter a 7 digit number for the project number.
• Press Enter button to save setting.
• Press or button to select next item or EXIT.
• Press Enter button to continue.
4.4 Good Laboratory Practices (GLP) Set
Good Laboratory Practices (GLP) Set submenu allows the selection and will permit printing of Time, Balance Identifi­cation Number, User Identification Number, Project Number, Diff erence and Name data to be printed. When the selected items are set to ON, these items are not displayed. The default setting is OFF. When an external printer is used, and all items are set ON and the balance is calibrated, the printer will print out calibration data for audit trail purposes and will indicate date, and time. (It should be noted that the User ID number and Project n umber must be entered in the GLP Data submenu before printed data is availab le).
Procedure
To select any of the items in the GLP Set menu, proceed as follows:
• Press the Setup button, CAL is displayed.
• Press or button until GLP SET is displayed.
• Press Enter button to continue.
• Press or button until either TIME, BAL ID , USER NO, PROJ NO, DIFF, NAME, LOCK or EXIT is displayed.
• Press Enter button to continue.
• Press or button and select either ON or OFF.
• Press Enter button to save setting.
• Press or button to select next item or EXIT.
• Press Enter button to continue.
28
4.5 Print
The Print menu provides a number of options which can be turned ON or OFF. It contains eight submenus: Auto Print, feature which includes selection of Off, Continuous , Interval and on Stability, Inter , specifies time interval f or automatic output of displayed data, Stable data-only feature, Numeric only or full display data f or output, GLPCont, GLPT are, Reference which prints ref erence w eight v alue and Loc k which enab le y ou to prog ram balance par am-
eters and to lock the settings.
Procedure
• Press the Setup button, CAL is displayed.
• Press or button until PRINT is displayed.
• Press Enter button to continue.
• Press or button until either AUTOPRT, INTER, STABLE, NUMERIC , GLPCONT, GLPTARE, REFEREN, LOCK or EXIT is displayed.
• Press Enter button to continue.
• Press or button and select either menu setting or ON or OFF.
• Press Enter button to save setting.
• Press or button to select next item or EXIT.
• Press Enter button to continue.
Auto Print Feature
When enabled, the Auto Print f eature causes the balance to automatically output displa y data in one of three w a ys: continuously, at user specified time intervals, or upon stability. Default settings are shown bold.
OFF when set on turns off the auto print feature
Cont when set on, outputs printed data continuously Inter provides a user specified printing interval On Stb provides printed data only when a stable reading is achieved
Interval
Can be set to provide a specified printing interval between 1 and 3600 seconds.
Print Stable Data Only
When set On , this feature permits only stable display data to be output. OFF is the default setting.
Print Numeric Data Only
When Numeric Data Only function is turned ON, this allows the balance to output numeric data only for RS232 output. OFF is the default setting.
29
4.5 Print (Cont.) GLP Continuously
When the GLP Continuously function is set ON, allows the balance to output the GLP selections each time a weight value is printed to the printer. OFF is the default setting. The following example is with GLP Cont On.
Sample Printout
GLP Set Menu Options Turned On
7/01/97 12:01:37 AM
429.5 g 7/01/97 12:01:52 AM
Bal Id
429.8 g 7/01/97 12:02:17 AM
Bal Id USER NO 1000001
429.8 g 7/01/97 12:02:43 AM
Bal Id USER NO 1000001 PROJ NO 2000002
429.5 g 7/01/97 12:02:43 AM
Bal Id USER NO 1000001 PROJ NO 2000002
Name . . . . . . . . .. . . . . .
429.5 g
Time = On
Time = On Balance ID = On
Time = On Bal ID = On User No. = On
Time = On Bal ID = On User No = On Proj No = On
Time = On Bal ID = On User No = On Proj No = Name = On
GLP Once After Tare
When the GLP Tare function is set ON, allows the balance to output the GLP selections once after tare when the weight value is printed to the printer. OFF is the default setting. The following e xample is with GLP Once After Tare.
GLP Set Menu Options Turned On
7/01/97 12:01:37 AM
429.5 g
429.6 g
429.7 g
429.7 g
429.7 g
Time = On
Reference
When the Reference function is set ON, prints the value of weight used as a ref erence in either Percent and Parts Counting modes. OFF is the default setting.
Lock
Lock ON/OFF can only be changed when the hardware Lockswitch is set OFF/unlocked. A menu is locked when the menu lock is set ON and the Lockswitch is ON. Loc k when selected and turned on, locks all of the entries made under the Print menu. In the locked condition, items may be looked at but not changed in the menu. When set off, entries may be changed. OFF is the default setting.
30
4.6 RS232
The RS232 menu provides communication parameters which can be set to accommodate e xternal printers or com­puters. It contains five submenus: Baud rate, Parity, Data, Stop bit and Lock ON or OFF which enable you to program balance parameters and to lock the settings .
Procedure
• Press the Setup button, CAL is displayed.
• Press or button until RS232 is displayed.
• Press Enter button to continue.
• Press or button until either BAU D, PARITY, DAT A, ST OP, LOCK is displayed.
• Press Enter button to select.
• Press or button and select the desired menu setting.
• Press Enter button to sav e setting.
• Press or button to select next item or EXIT.
• Press Enter button to continue.
Baud Rate
This submenu is used to select the desired baud rate. There are fiv e available baud rates to choose from: 300, 1200, 2400, 4800 and 9600. The default setting is 2400.
Parity
Parity can be set to Odd, Even or None . The default setting is None.
Data Bits
To set the number of data bits to 7 or 8. The default setting is 7.
Stop Bits
The number of stop bits can be set to 1 or 2. The default setting is 2.
Lock
Lock ON/OFF can only be changed when the hardware Lockswitch is set OFF/unlocked. A menu is locked when the menu lock is set ON and the Lockswitch is ON. Loc k when selected and turned on, locks all of the entries made under the RS232 menu. In the locked condition, items may be look ed at but not changed in the menu. When set off, entries may be changed. OFF is the default setting.
31
4.7 Legal for Trade (LFT)
Legal for trade (LFT) is a software controlled option which can be set LFTLOCK. When set to LFTLOCK, certain items in the Calibration, Readout, Print, Function and Units menus are automatically preset and locked to permit the balance to operate in a legal for trade application and works in conjunction with a Loc ks witch. Default setting is UNLOCKED . See default table.
DEF AUL T T ABLE
Procedure
• Press the Setup button, CAL is displayed.
• Press or button until LFT is displayed.
• Press Enter button to continue.
• Press or button and select LFTLOCK.
• Press Enter button to continue.
• Press or button and select either ON or OFF.
• Press Enter button to save setting.
• Press or button to select next item or EXIT.
• Press Enter button to continue.
NOTE: For legal for trade applications, the balance must be physically sealed.
LFT and Lockswitch Menu lock Default Value
Date Menu Unlocked
Time Menu Unlocked Calibration Menu
Span, Linearity, User Locked CalTest Unlocked Internal Calibration Unlocked
Readout Menu Stability Unlocked .5d (limited to .5d and 1d) Auto zero Unlocked .5d (limited to OFF and .5d) Filter Level Unlocked -1­GLP Data Menu Unlocked GLP Selections Unlocked Print Options Unlocked RS232 Menu Unlocked LFT Menu Lockswitch Locked Mode Menu Locked Weigh Units Menu Locked Grams* Global Menu Locked Custom Menu Unlocked
* Units oz and oz t cannot be simultaneously enabled. If Print Numeric Data is turned ON, then Print Stable Data Only is
locked ON.
When the balance is first turned ON and LFT has been previously set ON, the follo wing display will appear if LFT is set in the menu and the Lockswitch is set ON.
When the balance is first turned ON and LFT has been previously set ON, the f ollowing displa y will appear if LFT is set in the menu and Calibration menu is locked, and the Lockswitch is set ON.
The display check countdown appears only in the first 30 seconds after
plugging it in
and only when the balance has been previously set with Type Approved/Legal for Trade set ON.
32
4.8 Mode
The Mode submenu permits the selection of five modes which can be turned ON or OFF. These modes are: Weigh, Percent, Count, Animal and Lock. Weigh is turned ON and all others have a default setting of OFF. When any of
the modes are turned ON, they can be selected for operation from the Mode button.
Procedure
• Press the Setup button, CAL is displayed.
• Press or button until MODE is displayed.
• Press Enter button to continue.
• Press or button until either WEIGH, PER­CENT, COUNT, ANIMAL, LOCK or EXIT is displayed.
• Press Enter button to continue.
• Press or button and select either ON or OFF.
• Press Enter button to save setting.
• Press or button to select next item or EXIT.
• Press Enter button to continue.
Weigh
The Weigh submen u is always set to ON as a default.
Percent
Percent weighing permits you to place a reference load on the balance, then view other loads as a percentage of the reference. Selection is made using the Mode button. The def ault setting is OFF.
Count
Counting is used when counting quanties of parts. Selection is made using the Mode button. The default setting is OFF.
Animal
Animal weighing provides special settings to accommodate animal mov ements . Selection is made using the Mode button. The def ault setting is OFF.
Lock
Lock ON/OFF can only be changed when the hardware Lockswitch is set OFF/unlocked. A menu is locked when the menu lock is set ON and the Lockswitch is ON. Loc k when selected and turned on, locks all of the entries made under the Mode menu. In the locked condition, items may be look ed at but not changed in the menu. When set off, entries may be changed. OFF is the default setting.
33
4.9 Units
The Units submenu permits the selection of the measuring units which can be turned ON or OFF and locked.
Procedure
• Press the Setup button, CAL is displayed.
• Press or button until UNITS is displayed.
• Press Enter button to continue.
• Press or button until desired measuring unit is displayed.
• Press Enter button to continue.
• Press or button and select either ON or OFF.
• Press Enter button to save setting.
• Press or button to select next item or EXIT.
• Press Enter button to continue.
Units
Measuring units settings are made using the Units button. This menu permits the measuring units to be turned ON or OFF. The default setting is OFF.
Lock
Lock when selected and turned on, locks all of the entries made under the Units button. The default setting is OFF.
4.10 Global
This menu contains three functions which can be set to either a yes or no type of operation. These functions are: List, Reset and Version. The def ault setting are NO. Global List is a convienent method of e xamining which param- eters are set up in the balance. The parameters do not show up on the display but print out when selected. The Global menu contains the List function.
Procedure
• Press the Setup button.
• Press or button until GLOBAL is displa yed.
• Press Enter button to continue.
• Press or button until either LIST, RESET, VERSION, LOCK or LIST is displayed.
• Press Enter button to continue.
• Press or button and select either YES or NO.
• Press Enter button to save setting.
• Press or button to select next item or EXIT.
• Press Enter button to continue.
34
4.10 Global (Cont.) List
This submenu can be used to output a listing of current menu settings via the RS232 interface. When YES is selected, all menu settings will be output either to an external printer or computer. To use this feature, your balance must be connected to a computer or printer. The def ault setting is OFF.
The partial sample shown, indicates the status in the
SAMPLE PRINTOUTS
menus.
EXPLORER xxxxxx-xxx Sr #x.xx OS#x.xx G#x.xx
Time= 12hr 3:19:51 PM Date= m/d/y 7/09/97 Function = Weigh ReadOut Menu
Stb= 2 d AZT= .5 d Filter= 1
GLP Menu
Time/Date= On Bal Id= On User No.= On Proj No.= On DIFF= On Name= On
Print Menu
Auto Print= Off Interval= 7 Stable Print= On NU= On GLP Cont = Off GLP on Tare = Off Print Ref= On
RS232 = 2400: N: 7: 2
LFT is Off Mode Menu
WEIGH= On PERCENT= Off COUNT= Off
ANIMAL= Off Lock Switch is Off Menu Locks
RS232= Off
READOUT= Off
GLPSET= Off
MODE= Off
UNITS= Off
PRINT= Off
GLP Data = Off
CAL= Off
GLOBAL= Off
CUSTOM= Off Enabled Units:
g
custm C. Units:
1.000000 E0 x 1
Reset
Reset when set to yes will reset the balance to factory default settings . The default setting is NO.
Version
Displays software re vision number for servicing purposes. This number is installed with the balance .
Lock
Lock when selected and set tp YES, locks all of the entries made under the Global menu. The def ault setting is NO.
35
4.11 Custom Unit
Custom Unit is enabled when Custom Unit Setup under Units Menu is turned ON. This feature can be used to create your own custom weighing unit. It permits entering a conversion factor which the balance will use to conv ert grams to the desired unit of measure.
Conversion Weight W eight
Factor x in = in
grams custom unit
Conversion f actors are e xpressed in scientific notation and entered into the balance in three parts:
a number between 0.1 and 1.999999 called the mantissa
a power of 10 called the exponent
a least significant digit (LSD)
SCIENTIFIC NOTATION
Number
Between
Con v . 0.1 and Po wer Man-
Factor 1.999999 of 10 tissa Exp.
123.4 = .1234 x 1000 = .1234 x 10
12.34 = .1234 x 100 = .1234 x 10
1.234 = .1234 x 10 = .1234 x 10 .1234 = .1234 x 1 = .1234 x 10 .01234 = .1234 x .1 = .1234 x 10 .001234 = .1234 x .01 = .1234 x 10 .000123 = .123 x .001 = .123 x 10
EXPONENTS
E-3 Moves decimal point 3
places to the left.
E-2 Moves decimal point 2
places to the left.
E-1 Moves decimal point 1
place to the left.
E0 Leaves decimal point
in normal position.
E1 Moves decimal point 1
place to the right.
E2 Moves decimal point 2
places to the right.
E3 Moves decimal point 3
places to the right.
3
2
1
0
-1
-2
-3
Procedure
• Press the Setup button.
• Press or button until CUSTOM is displayed.
• Press Enter button to continue, FACT OR is displayed.
• Press Enter button, The mantissa of the current conversion is displa yed. The mantissa of the current conversion f actor is displayed. This is a number between 0.1 and 1.999999 with the first digit flashing. For conv ersion factors outside of this range, the exponent will be used to mov e the decimal point.
• Press or buttons as directed by the display and enter a 7 digit number for the conver­sion factor .
• Press Enter button, EXP (exponent) is displa y ed.
• Press Enter button, 0 (exponent) is displayed.
• Press or button and select exponent value either -3, -2, -1, 0, 1, 2, or 3.
36
4.11 Custom Unit (Cont.)
LSD’s
LSD .5 Adds one decimal
place display counts by 5’s.
LSD 1 Display counts by 1’s.
LSD 2 Display counts by 2’s. LSD 5 Display counts by 5’s. LSD 10 Display counts by 10’s. LSD 100 Display counts by
100’s.
Procedure (Cont.)
• Press Enter button to continue, LSD is displayed. There are 6 LSD (least significant digit) settings you can choose from (see table).
• Press Enter button, LSD 1 is displayed.
• Press or button and select LSD value either .5, 1, 2, 5, 10 or 100.
•Press Enter button, SAVED is momementary dis­played f ollow ed by LOCK.
•Press Enter button to continue.
•Press or button and select ON or OFF.
•Press Enter button, EXIT is displayed.
• Press Enter button to return to weighing mode.
37
4.12 Menu Lock-Out Protection
Access to the various menus can be disabled setting the Lockswitch located on the PC board inside the balance to OFF position. The Lockswitch locks out all menus which have had Lock turned ON. The default setting for the Lockswitch is OFF. The Lockswitch access is located on the bottom of the balance.
Type Approved/Legal for Trade Balance Sealing
AlL Vo y ager balances ma y be sealed f or type appro v ed/ legal for trade applications. Type Approved balances are sealed as shown in the figures.
For type approv ed balances consult local Weights and Measures officials to determine sealing method require­ments.
After the balance has been set up properly and LFT is set ON, proceed to seal the balance as shown below.
Procedure
• T urn the display off and unplug the po w er cord.
• T urn the balance ov er, make sure the pan does not fall off.
• The Lockswitch can be accessed through the hole nearest the front panel as shown in the illustration below .
• Remove the plastic cover from the hole nearest the front panel.
• Select the desired position on the Lockswitch and reassemble the balance. Use a small screwdriver to reach the switch.
• Replace the plastic cover in the switch access hole.
• T urn the balance ov er and plug in the po w er cord.
LFT Switch Location and Sealing Method
Top Sealing Location and Method
38
5 CARE AND MAINTENANCE
To keep the balance operating properly, the housing and platform should be kept clean and free from foreign material. If necessary, a cloth dampened with a mild detergent may be used. Keep calibration masses in a safe dry place.
5.1 T r oubleshooting
SYMPTOM
Unit will not turn on.
Incorrect weight reading.
Cannot display weight in desired unit.
Unable to store menu settings/ changes.
RS232 interface not working.
PROBABLE CA USE(S)
Po wer cord not plugged in or properly connected to balance.
Balance was not re-zeroed before weighing.
Balance not properly calibrated.
Desired unit not enabled.
Enter was not selected. Menu locked. Print menu settings not properly
set up.
REMED Y
Check power cord connections.
Press with no weight on the pan, then weigh item.
Recalibrate correctly .
Press Units button until desired measuring unit is shown.
Enable units in Setup menu. Press Enter when prompted. Menu not locked properly. V erify interface settings in RS232
menu correspond to those of peripheral device.
Random segments displayed or display locks up.
Unable to change settings.
Unstable readings.
Error message display.
Cannot access weighing mode.
Cable connections. Microprocessor locks up.
Menu locked (Lockswitch set ON)
Excessive air currents. Vibration on table surface.
Desired weighing mode is not enabled.
Check cable connections. Turn power off, then turn on again.
If condition persists, unit must be serviced.
Set menu Lock OFF. Set Lockswitch to OFF.
Check environmental conditions. Place balance on a stable surface
or change averaging le v el. See Error Codes list.
Press mode until desired weighing mode is displayed.
Enable weighing mode.
39
5.2 RS232 Interface
Explorer balances are equipped with a bi-directional RS232 compatible interface f or communication with printers and computers. When the balance is connected directly to a printer, displa y ed data can be output at any time by simply pressing PRINT, or b y using the Auto Print feature.
Connecting the balance to a computer enables you to operate the balance from the computer, as w ell as receive data such as display ed w eight, weighing mode , stability status, etc.
The following sections describe the hardware and software provided with the balance
Hardware
On the side of the balance, the left-hand, 9-pin male subminiature “D” connector is provided for interfacing to other devices. The pinout and pin connections are sho wn in the adjacent illustration. The balance will not output any data unless pin 5 (CTS) is held in an ON state (+3 to +15 V dc). Interfaces not utilizing the CTS handshake may tie pin 5 to pin 6 to defeat it.
Output Formats
Data output can be initiated in one of three ways: 1) By pressing PRINT; 2) Using the Auto Print feature; 3) Sending a print command (“P”) from a computer.
The output format is illustrated in the RS232 command table which follows .
RS232 Connection
Side of Balance
RS232 Commands
All communication is accomplished using standard ASCII format. Characters sho wn in the following table are acknowledged by the balance. Inv alid command re­sponse "ES" error indicates the balance has not recog­nized the command. Commands sent to the balance must be terminated with a carriage return (CR) or carriage return-line line feed (CRLF). F or example, a tare command should appear as shown in the adjacent diagram. Data output b y the balance is alw ays terminated with a carriage return - line feed (CRLF).
TARE *
PRINT *
RS232 COMMAND TABLE
CommandCommand
Command
CommandCommand
CharacterCharacter
Character
CharacterCharacter
??
? Print current mode
??
nnnnAnnnnA
nnnnA Set Auto Print feature to “nnnn”
nnnnAnnnnA
DescriptionDescription
Description
DescriptionDescription
(see table).
Field: Mode Stab CR LF
Length: 5 1 1 1
blank if stable “ ? ” if unstable
mg
GN
g
tael
kg
tael
dwt
tael
ct
momme
oz
lb
oz t
nnn = 0 Turns feature OFF nnn = S Output on stability nnn = C Output is continuous nnnn = 1-3600 Sets Auto Print
N tical custm Pcs %
Interval
40
1 N/C 2 Data Out (TXD) 3 Data In (RXD) 4* Tare (External signal) 5 Clear To Send (CTS) 6 Data Terminal Ready (DTR) 7 Ground 8 Request To Send (RTS) 9* Print (External signal)
* External PRINT and/or TARE
switches may be installed as shown in the diagram. Momentary contact switches must be used.
RS232 COMMAND TABLE (Cont.)
CommandCommand
Command
CommandCommand
CharacterCharacter
Character
CharacterCharacter
CC
C Begin span calibration
CC
xDxD
xD Set 1 second print delay (set x = 0 for OFF, or x = 1 for ON)
xDxD
FF
F Print current function.
FF
xx
II
x
I Set Averaging Filter Level to “x”,
xx
II
LL
L Begin linearity calibration
LL
xMxM
xM Places balance in mode “x”,
xMxM
DescriptionDescription
Description
DescriptionDescription
where x = 0 to 3 (see table). If LFT, level 0 to 1.
where x = 1 to 17 (see table). If unit or mode is not already
enabled, command will be ignored.
0 = minimum level 1= 2= 3 = maximum level
1 = (not used) 2 = grams 3 = kilo grams 4 = dwt 5 = Carats 6 = Ounces 7 = Ounces troy 8 = Grains
9 = Taels Hong Kong 10 = Taels Singapore 11 = Taels Taiwan 12 = Mommes 13 = Decimal Pounds 14 = (Not used) 15 = Newton's 16 = tical 17 = Custom Units
PP
P Print display data
PP
xSLxSL
xSL Set stable data only printing (set x = 0 to 3). If LFT 0 or 1.
xSLxSL
TT
T Same effect as pressing O/T button.
TT
VV
V Print EPROM version
VV
Esc VEsc V
Esc V Print balance ID (13 characters).
Esc VEsc V
xZxZ
xZ Set Auto Zero to “x”,where x = 0 to 3 ). 0=Off, 1=0.5d, 2=1d, 3=5d.If LFT,
xZxZ
Field: Weight Mode Stab CR LF Length: 9 1 5 1 1 1 When "numeric only"data is selected for output in the RS232 menu, the Mode field is not output.
Displayed weight sent right justified w/lead zero blanking. Nine characters include: decimal point (1) weight (7 max) polarity (1): blank if positive
"-" if negative
programs Auto zero level from 0 to 1.
Same as ? command
41
RS232 COMMAND TABLE (Cont.)
CommandCommand
Command
CommandCommand
CharacterCharacter
Character
CharacterCharacter
x%x%
x% Set % reference function. Uses x (Real Number) as current Percent
x%x%
x#x#
x# Set PC reference function. Uses x (Real Number) as current Percent
x#x#
Esc REsc R
Esc R Resets Setup and Print menus to factory defaults.
Esc REsc R
ONON
ON Turns balance on.
ONON
DescriptionDescription
Description
DescriptionDescription
Reference. Reference weigh must be entered in grams.
Reference. Reference weigh must be entered in grams.
CAUTION: This will reset RS232 configuration.
OFFOFF
OFF Turns balance off.
OFFOFF
% Print current Percent Reference Weigh.
xFxF
xF Set current function, set x for 0=None (normal weigh), 1=percent, 2=parts
xFxF
xAWxAW
xAW Set Animal Level from 0 to 3. 0= least amount of filtering.
xAWxAW
xExE
xE Set/Reset Auto Restart in Animal mode. Where x is 0=Off and 1=ON.
xExE
EE
E Start Animal cycle.
EE
xTxT
xT Download tare, tare weight must be entered in grams.
xTxT
IDID
ID Print Current User ID String.
IDID
xIDxID
xID Program User ID String, 1-8 characters.
xIDxID
ACAC
AC Abort Calibration.
ACAC
xUCxUC
xUC User Weight Calibration.
xUCxUC
ICIC
IC Internal calibration (InCAL
ICIC
LELE
LE Show Last Error Code. Response: Err: Error Number.
LELE
SNSN
SN Show Serial Number.
SNSN
xSxS
xS Print Stable Only. Where x =0 Off and x=1 On.
xSxS
Counting, 3=Animal Weighing. None
##
# Print current Parts Count Reference Weigh.
##
TM)
TIMETIME
TIME Print Current Time.
TIMETIME
mm/dd/yy SETDATEmm/dd/yy SETDATE
mm/dd/yy SETDATE Set Date Command and remove Invalid Indicator
mm/dd/yy SETDATEmm/dd/yy SETDATE
hh:mm:ss SETTIMEhh:mm:ss SETTIME
hh:mm:ss SETTIME Set Time Command and Remove Invalid Time Indicator
hh:mm:ss SETTIMEhh:mm:ss SETTIME
DATEDATE
DATE Prints Current Date. Note.
DATEDATE
W$TMW$TM
W$TM Write Clock Trim Value. The clock can be adjusted by + - 60 seconds a
W$TMW$TM
R$TMR$TM
R$TM Read Clock Trim value
R$TMR$TM
SETUPSETUP
SETUP Program Setup menu Options
SETUPSETUP
SWSW
SW Show Lockswitch status.
SWSW
month.
42
5.3 Error Codes List
Error Codes List
The following list describes the various error codes and which can appear on the displa y and the suggested remedy.
Data Errors
1.0 Tr ansient error (hardware error, probably static discharge). If error persists, the balance must be serviced.
1.4 Transducer data error (lost communications between logic boards).
Tare Errors
2.0 Balance is unable to stabilize within time limit after taring. Environment is too hostile or balance needs recalibration.
Calibration Errors
3.0 Incorrect or no calibration mass used for calibration. Recalibrate with correct masses.
RS232 Errors
4.4 RS232 buffer is full.
User Errors
7.0 User entry out of bounds.
7.2 Number outside of display capacity.
Over-Under Load Errors
8.0 Hardware error causing an internal weight signal which is too low. Check if pan is off. If not, the balance must be serviced.
8.1 Hardware error caused by an internal weight signal which is too high. Chec k load on the pan which may be excessive . If error persists, the balance must be serviced.
8.2 Power-on load out of specification (LFT only)
8.3 Rated capacity exceeded. Remove excessive weight from pan.
8.4 Underload condition on balance. Check that the proper pan is installed.
8.5 Internal calibration weight internal sensor indicated its weight on the pan.
CheckSum Errors
9.0 Bad factory checksum. If error persists, have the balance serviced.
9.1 Bad factory checksum. If error persists, have the balance serviced.
9.2 Bad factory checksum. If error persists, have the balance serviced.
9.3 Bad factory checksum. If error persists, have the balance serviced.
9.4 Factory internal calibration data failed checksum. This f ailure will disab le access to the InCAL installed).
9.5 Factory calibration data failed checksum.
9.8 User calibration data failed checksum.
9.9 Factory temperature compensation data failed checksum.
5.4 Information Messages
CAL NOW If InCALTM (internal calibration) is installed. Message to recalibrate the balance . The
message will remain until calibrated.
WARM UP The user tried to perform an internal or external calibration and this message will be
flashed in the 14 segment field. The balance requires a 30 minute warmup period. During warmup the user can not perform calibration..
SA VED This message is flashed when an item is changed in the menu and the new value is written
to the EEPROM.
LOCKED This message is flashed when an item can not be changed in the menu because the menu
is locked and the Lockswitch is set locked.
LOW REF The message is flashed in parts counting or percent when the calculated reference weight
is very low .
UNSTBLE This message is flashed when the balance was unable to aquire stable data during internal
calibration.
43
TM
feature (if
5.5 Service Information
If the Troubleshooting section does not resolve or describe your problem, you will need to contact an authorized Ohaus Service Agent. For Service assistance in the United States, please call Aftermarket, Ohaus Corporation toll­free at (800) 526-0659. An Ohaus Product Service Specialist will be available to help you.
5.6 Replacement par ts
Description Ohaus Part No. Po wer Pack, 100/120 V ac US Plug (Cord set part of power pack) 490202-01 Po wer P a ck, (Cord set required for UK, European and Australian) 490203-01 Cord Set, 230 V ac, UK Plug 76448-00 Cord Set, 230 V ac , European Plug 76212-00 Cord Set, 230 V ac, Australian plug 76199-01
5.7 Accessories
Description Ohaus Part No.
Calibration Masses - ASTM Class 1 T olerance:
1 kg 49016-11 2 kg 49026-11 4 kg 49046-11
In-Use Display Cover Kit 470003-01 Security Device Modular Display Upgrade Kit Auxilliary Display Kit
(T ab le Mount) 470009-01 (Wall Mount) 470009-02 (Tower Mount) 470009-03
Remote Display Kit
(T ab le Mount) 470010-04 (Wall Mount) 470010-05
(Tower Mount) 470010-06 RS232 Interface Cable , Blunt end (user defined) AS017-01 RS232 Interface Cable, IBM
® -
PC 25 Pin AS017-02 RS232 Interface Cable , (connects impact printer) AS017-06 RS232 Interface Cable, IBM RS232 Interface Cable, Apple
® -
PC 9 Pin AS017-09
®
llGS/Macintosh AS017-10
Printer AS142
44
5.8 Specifications
Capacity (g) 12,000 22,000 32,0000 Readability (g) 0.1 Weighing modes g, kg, oz , oz t, ct, dwt, taels (3), mommes, gn, ti, N, custom unit Functions Parts counting, animal weighing, percent Options GLP, time, date, lockswitch, LFT (U.S.)/type approved Repeatability (Std. dev.) (g) 0.1 Linearity (g) (+) 0.4 Tare range Full capacity by subtraction Safe overload capacity 150% of capacity Stabilization time <4 seconds Sensitivity drift PPM/°C 3
(10°C - 30°C) Operating temperature range:
w/internal calibration 10° to 40°C/ 50° to 104°F w/o internal calibration 10° to 30°C/50° to 86°F
Calibration InCALTM calibration Power requirements External Adapter, 100-120 V ac, 220 V ac, 50/60 Hz
Display (in/cm) 0.6/1.5 Pan size (in/cm) 11 x1 3/4 x 13/27.9 x 4.4 x 33 Dimensions (WxHxD) (in/cm) 14 x 5 1/4 x 16/35.5 x 13.3 x 40.6
Plug configuration for US, Euro, UK, Japan & Australia
NOTE: Not all weighing modes apply depending upon capacity and resolution of the balance.
45
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of deliv­ery through the duration of the warranty period. During the warranty period Ohaus will repair , or , at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus.
This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for any consequential damages.
As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for further details.
46
Ohaus Corporation 29 Hanover Road, Florham Park, NJ 07932, USA Tel: (973) 377-9000, Fax: (973) 593-0359
With offices worldwide. 21202611A Printed in Switzerland 9709/2.12
P/N 400040-512 R0997 © Ohaus Corporation 1997, 1999 all rights reserved.
47
Loading...