This product conforms to the EMC Directive 2004/108/EC and the
Low Voltage Directive 2006/95/EC. The complete Declaration of
Conformity is available from Ohaus Corporation
Este producto se conforma con el EMC 2004/108/EC directivo y la
baja tensión 2006/95/EC directivo. El declaración completo de la
conformidad está disponible de Ohaus Corporation.
Ce produit se conforme à l’EMC 2004/108/EC directif et la basse
tension 2006/95/EC directif. La déclaration complète de la
conformité est fournie par Ohaus Corporation.
Important notice for verified weighing instruments
Weighing Instruments verified at the place of manufacture bear one of the preceding mark on the packing label and
the green ‘M’ (metrology) sticker on the descriptive plate. They may be put into service immediately.
Weighing Instruments to be verified in two stages have no green ‘M’ (metrology) on the descriptive plate and bear
one of the preceding identification mark on the packing label. The second stage of the initial verification must be
carried out by the approved service organization of the authorized representative within the EC or by the national
weight & measures (W+M) authorities.
The first stage of the initial verification has been carried out at the manufacturers work. It comprises all tests according to the adopted
European standard EN 45501:1992, paragraph 8.2.2.
If national regulations limit the validity period of the verification, the user of the weighing instrument must strictly observe the re-verification
period and inform the respective W+M authorities.
Notificación importante para instrumentos de pesaje verificados
Los instrumentos de pesaje verificados en el sitio de fabricación llevan una de las marcas precedentes en el rótulo
del empaque y la etiqueta de la ‘M’ verde (metrología) en la placa descriptiva. Estos instrumentos se pueden poner
en funcionamiento inmediatamente.
Los instrumentos de pesaje a ser verificados en dos etapas no tienen ninguna ‘M’ verde (metrología) en la placa
descriptiva, y presentan una de las marcas de identificación precedentes sobre el rótulo del empaque. La segunda
etapa de la verificación inicial debe ser llevada a cabo por la organización de servicio aprobada del representante
autorizado dentro de la CE o por las autoridades nacionales de pesos y medidas.
La primera etapa de la verificación inicial ha sido llevada a cabo en el sitio de fabricación. Ésta comprende todas las pruebas estipuladas
por el estándar europeo adoptado: EN 45501:1992, párrafo 8.2.2.
Si las normas nacionales limitan el periodo de validez de la verificación, el usuario del instrumento de pesaje debe seguir estrictamente el
periodo de re-verificación e informar a las correspondientes autoridades de pesos y medidas.
Avis important pour les instruments de pesage vérifiés
La première étape de la vérification initiale a été exécutée sur le site du fabricant. Elle se compose des tests requis par la norme européenne
EN45501:1992, paragraphe 8.2.2.
Si des règlements nationaux limitent la durée de validité de la vérification, il incombe à l’utilisateur dudit instrument de pesage de respecter
strictement la période de re-vérification et d’informer les autorités de poids et mesures respectives.
Les instruments de pesage vérifiés sur le site de fabrication portent l’une des marques précédentes sur l’étiquette de
l’emballage avec un autocollant M (pour Métrologie) en vert sur la plaque descriptive. Ces instruments peuvent être
immédiatement mis en service.
Les instruments de pesage à vérifier en deux étapes ne portent pas d’autocollant M (pour Métrologie) en vert sur la
plaque descriptive et portent l’une des marques d’identification précédentes sur l’étiquette de l’emballage. La
deuxième étape de la vérification initiale doit être exécutée par l’organisation de service homologuée du représentant
agréé au sein de la CE ou par les autorités nationales de poids et mesures.
Disposal
In conformance with the European Directive 2002/96 EC on Waste Electrical and Electronic Equipment (WEEE) this device
may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic
equipment.
If you have any questions, please contact the responsible authority or the distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be
related.
Thank you for your contribution to environmental protection.
Eliminación de residuos
De conformidad con las exigencias de la directiva europea 2002/96 CE sobre residuos de aparatos eléctricos y
electrónicos (RAEE), este equipo no puede eliminarse como basura doméstica. Esta prohibición es asimismo válida para
los países que no pertenecen a la UE cuyas normativas nacionales en vigor así lo reflejan.
Elimine este producto, según las disposiciones locales, mediante el sistema de recogida selectiva de aparatos eléctricos y
electrónicos.
Si tiene alguna pregunta al respecto, diríjase a las autoridades responsables o al distribuidor que le proporcionó el equipo.
Si transfiere este equipo (por ejemplo, para la continuación de su uso con fines privados, comerciales o industriales),
deberá transferir con él esta disposición.
Muchas gracias por su contribución a la conservación medioambiental.
Elimination
En conformité avec les exigences de la directive européenne 2002/96 CE relative aux déchets d'équipements électriques et
électroniques (DEEE), cet appareil ne doit pas être éliminé avec les déchets ménagers. Logiquement, ceci est aussi valable
pour les pays en dehors de l'UE conformément aux règlementations nationales en vigueur.
Veuillez éliminer cet appareil conformément aux prescriptions locales dans un conteneur séparé pour appareils électriques et
électroniques.
Pour toute question, adressez-vous aux autorités compétentes ou au revendeur chez qui vous avez acheté cet appareil.
En cas de remise de cet appareil (p. ex. pour une utilisation privée ou artisanale/industrielle), cette prescription doit être
transmise en substance.
Merci pour votre contribution à la protection de l'environnement.
C
S
Compliance to the following standards is indicated by the corresponding marking on the product.
El cumplimiento con los siguientes estándares se indica mediante el correspondiente marcado sobre el
producto.
La conformité aux normes suivantes est indiquée par le marquage correspondant sur le produit.
Marking
Marcado
Marquage
Standard
Estándar
Normes
UL60950-1 : 2003
R
U
AS/NZS4251.1, AS/NZS4252.1
NSF / ANSI 169
NSF/ANSI /3-A 14159-1
FCC Note
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Industry Canada Note
This Class B digital apparatus complies with the Canadian ICES-003.
Cet appareil numérique de la classe B est conforme à la norme NMB-003 du Canada.
ISO 9001 Registration
In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI),
confirming that the Ohaus quality management system is compliant with the ISO 9001 standard’s requirements. On May 15, 2003,
Ohaus Corporation, USA, was re-registered to the ISO 9001:2000 standard.
Registro ISO 9001
En 1994, Bureau Veritus Quality International (BVQI) le otorgó a Ohaus Corporation, EE.UU., un certificado de registro ISO 9001 el cual
confirma que el sistema administrativo de calidad de Ohaus cumple con los requerimientos del estándar ISO 9001. En mayo 15 del
2003, Ohaus Corporation, EE.UU., fue registrada nuevamente al estándar ISO 9001:2000.
Enregistrement ISO 9001
En 1994, le Bureau Veritus Quality International (BVQI) a octroyé la certification d’enregistrement ISO 9001 à Ohaus Corporation, ÉtatsUnis d’Amérique, confirmant que le système de gestion de la qualité Ohaus était conforme aux conditions normalisées de l’ISO 9001. Le
15 mai 2003, Ohaus Corporation, États-Unis d’Amérique, a été ré-enregistrée à la norme ISO 9001:2000.
This manual contains installation, operation and maintenance instructions for the Valor 5000 Series
scale. Please read the manual completely before installation and operation.
1.1 Product Description
The Valor 5000 Series is an easy-to-clean and rugged compact scale engineered specifically for
hygiene-oriented environments and to enhance productivity. With its compact size and food safety
approvals, it is ideal for use in sanitary and food preparation areas such as kitchens, bakeries,
restaurants, packaging and processing facilities. Battery operation and a backlit LCD display make
for a portable scale that can be easily carried to various locations and used under a wide range of
lighting conditions. Its fast display update speed, straight forward 4-button operation and
configurable checkweighing mode make this a simple yet flexible scale for general weighing,
portioning and sorting applications. This precision weighing scale will provide years of service
when properly cared for.
1.2 General Features
• Capacity Models (V51Px, non-approved): 3kg x .5g / 6lb x 0.001lb, 6kg x 1g / 15lb x 0.002lb,
15kg x 2g / 30 lb x 0.005lb, 30kg x 5g / 60lb x 0.01lb
• Maximum Displayed Resolution (V51Px Series) 1:6000 to 7500
• Capacity Models (V51Px, approved): 3kg x 1g / 6lb x 0.002lb, 6kg x 2g / 15lb x 0.005lb, 15kg x
5g / 30 lb x 0.01lb, 30kg x 10g / 60lb x 0.02lb
• Type-approved Resolution 1:3000
• Capacity Models (V51PHx): 3kg x 0.1g / 6lb x 0.0002lb, 6kg x 0.2g / 15lb x 0.001lb, 15kg x 1g
/ 30lb x 0.001lb, 30kg x 1g / 60lb x 0.002lb
• Maximum Displayed Resolution (V51PHx Series) 1:30000
• Food Safety Certifications: NSF, USDA/AMS
• Stainless steel weighing pan: 209 x 209 mm (8.2" x 8.2")
• Multiple weighing units: kg, g, lb, oz (decimal or fractional), lb:oz (decimal or fractional oz)
• Display update speed: within 2 seconds
• 3-color LED with configurable audible alert checkweigh indications
• 4-button control panel, including dedicated Tare button
• Built-in level indicator and adjustable slip-resistant rubber leveling feet
• Built-in rechargeable lead-acid battery: 120-hour operation on a 15-hour charge
• Battery status indicator with power-saving Auto-off function
• Gross, Net, Tare operation
• Auto-zero tracking operation
• Auto-tare operation
• Enhanced digital filtering
• Connection for accessory Foot Switch for remote tare operation
Valor 5000 Series
1.3 Safety Precautions
For safe and dependable operation of this scale, please comply with the following
safety precautions:
• Verify that the input voltage printed on the AC Adapter’s data label and the plug type matches the
local AC power supply.
• Make sure that the power cord does not pose a potential obstacle or tripping hazard.
• Use only approved accessories and peripherals.
• Operate the scale only under ambient conditions specified in these instructions.
• Disconnect the scale from the power supply when cleaning the scale.
• Do not operate the scale in hazardous or unstable environments.
• Do not immerse the scale in water or other liquids.
• Avoid directly splashing the scale with water or other liquids.
• Do not drop heavy loads on the platform.
• Service should only be performed by authorized personnel.
EN-4
EN-5 Valor 5000 Series
2INSTALLATION
2.1 Unpacking
Unpack and verify that the following components have been included:
• Scale • Instruction Manual
• Weighing Pan• Warranty Card
• AC Adapter• Weights and Measures Sealing Kit
Save the packaging material. Repack the scale in the original packaging to ensure safe storage or
transport of the product.
2.2 Installing Components
Place the weighing pan securely over the four load pads on the top of the scale. If using the optional
foot switch, plug the foot switch connection into the appropriate receptacle on the underside of the
scale (refer to Figure 2-3).
2.3 Selecting the Location
Operate the scale on a firm, level surface. Avoid
locations with rapid temperature changes or
excessive dust, air currents, vibrations,
electromagnetic fields or heat.
Adjust the leveling feet so that the bubble is
centered in the circle of the level indicator.
Figure 2-1. Installing Scale Components.
Pan
Load pads x 4
Leveling feet x 4
Level Indicator
Figure 2-2. Leveling Indicator.
NOTE: The scale should be leveled each time its location is changed.
2.4 Connecting Power
2.4.1AC Power
Check to ensure that the voltage printed on the scale’s data label matches your local line voltage.
If this is not the case, DO NOT connect the scale to the power supply. Contact an OHAUS dealer for
assistance.
Connect the AC adapter to the wall outlet. Connect the plug into the power receptacle on the underside
of the scale.
Foot Switch Connector
Figure 2-3. Bottom of Scale.
Power Receptacle
Valor 5000 Series
Before using the scale for the first time, the internal rechargeable battery
should be fully charged for up to 15 hours. The scale can be operated
during the charging process. The battery is protected against over
charging and the scale can remain connected to the AC power line.
Plug the AC Adapter into the receptacle marked “Power In” located on the underside of the scale
and allow it to charge. While the battery is charging, the segments in the battery symbol will
alternately light. When the battery is fully charged, the battery symbol will disappear.
EN-6
2.4.2Battery Power
The scale can be operated on the internal rechargeable battery when AC power is not available. The
scale will automatically switch to battery operation if there is a power failure or the power cord is
removed. A fully charged battery can operate the scale for up to 120 hours (with the backlight off)
independent of the AC power line.
During battery operation, the battery symbol will light to show the current discharge status of the
battery (1 segment = 25% capacity). A flashing symbol indicates that the battery requires recharging.
The scale will automatically turn off when the battery is fully discharged.
CAUTION
BATTERY IS TO BE REPLACED ONLY BY AN AUTHORIZED OHAUS SERVICE DEALER.
RISK OF EXPLOSION CAN OCCUR IF REPLACED WITH THE WRONG TYPE OR
CONNECTED IMPROPERLY.
Dispose of the lead acid battery according to local laws and regulations.
2.5 Initial Calibration
When the scale is first installed, a SPAN calibration should be performed to ensure accurate
weighing results. Before performing the SPAN calibration, be sure to have the appropriate
calibration weights. Refer to Section 3.7.1.
EN-7 Valor 5000 Series
3OPERATION
3.1 Overview of Parts and Controls
Identify parts and controls using item numbers from Table 1.
21
20
3
2
4
5
6
7
8
1
Figure 3-1. Parts.
9
10
19
14
13
15
1718
16
Figure 3-2. Control Panel.
11
12
Valor 5000 Series
EN-8
24
23
22
36
38
Item Description
1Control buttons
2Display
3Data Label
4Weighing pan
5Adjustable leveling feet
6Connection for AC adapter
7Connection for optional foot switch
8Security Slide
9Power and Zero button
10Gross/Net/Tare and Unit button
11Target/Menu button
12Tare button
13Tare indicator
14Net indicator
15Leveling indicator
16Checkweigh mode indicator
17Center of Zero indicator
18Stable indicator
19Yellow Under LED
37
Figure 3-3. Display.
TABLE 3-1. PARTS AND CONTROLS.
25
35
34
Item Description
20Green Accept LED
21Red Over LED
22Not used
23Minus sign
24Not used
257-segment display
26Fractional ounce symbol(P in %mode)
27Battery charge symbol
28Kilogram unit symbol
29Gram unit symbol
30Ounce unit symbol
31Pound:Ounce unit symbol
32Pound unit symbol
33Tare indicator
34Net indicator
35Checkweigh mode indicator
36Pound:Ounce separator
37Center of Zero indicator
38Stable indicator
26 27
3233
28
29
30
31
EN-9 Valor 5000 Series
3.2 Control Functions
Four multi-function buttons are used to operate the scale.
Button
Primary
Function
(short
press)
Secondary
Function
(long press)
Menu
Function
Target
Setting
Function
ON / ZERO
• If scale is off,
turns scale on
• If scale is on,
sets zero
Off
If scale is on, turns
scale off
Yes
Accepts current
setting and
advances to the
next menu item
Shifts to next
digit
G/N/T
Momentarily
Displays
Gross Net and
Tare values
Units
Changes
weighing unit
No
Rejects current
setting or menu
item and
advances to the
next available
item
Increments
value of the
active digit
3.3 Basic Operation
3.3.1Turning On
With the scale off, press the ON/ZERO
zero.
3.3.2Turning Off
With the scale on, press and hold the ON/ZERO
Off
button. The scale performs a display test, then displays
TARGET
Initiates target
setting mode
Menu
Enters
menu mode
Back
Reverts back to
previous menu
item
Accepts current
displayed value
Off
button until OFF is displayed.
TARE
•Sets tare
• Clears tare, if
pan is empty
None
Exit
•Immediately
exits menu
mode
• Aborts
calibration
in progress
Changes sign
(+ / -) of
displayed value
3.3.3Zeroing
Remove the load from the pan and press the ON/ZERO
Center of Zero indicator turns on when the measurement is within ±1/4 d of the zero setting.
Off
button to set the display to zero. The
3.3.4Taring
Semi-automatic Tare (Tare Button)
To store or change the Tare value, place an empty container on the pan and press the TARE button.
The Net weight and NET indicator will be displayed.
To clear the Tare value, remove the container from the pan and press the TARE
indicator turns off and the gross weight will be displayed.
button. The NET
Valor 5000 Series
Automatic Tare (Refer to Section 3.7.3)
Auto-tare is a convenience feature that automatically tares the initial weight (such as an empty
container) placed on the empty pan, without having to press the TARE button. The tare value is cleared
automatically when the weight on the pan is fully removed.
Remote Tare (Foot Switch)
Taring with the optional foot switch is equivalent to pressing the Tare button.
EN-10
3.3.5Displaying Gross, Net and Tare Values
When a tare has been entered, pressing the G/N/T button successively displays the Tare (TARE
indicator turns on), the Gross, and the Net weights (NET indicator turns on). After a few seconds
the scale automatically reverts to displaying the NET weight.
3.3.6Changing Unit of Measure
Press and hold the
displayed, enable it as described in Section 3.7.4.
Units
button until the desired unit is displayed. If one of the available units is not
3.4 Menu
3.4.1Menu Structure
The FD Series menu structure consists of three levels: Main, Sub-menu, and Menu item. The Main
menu contains the Sub-menus and each Sub-menu contains several Menu items. Each Menu item
contains two or more settings.
TABLE 3-2. MENU STRUCTURE.
Main menu
Sub-menu
Menu item
P
Note: LFT menu item and LOCK sub menu will not appear in V51PHXX models
EN-11 Valor 5000 Series
3.4.2Menu Navigation
Use the four buttons marked with Yes, No, Back and Exit to navigate through the menus (refer to
Section 3.2).
• Yes - Enters the displayed sub-menu or menu item and accepts the displayed setting.
• No - Skips to the next sub-menu or menu item and changes the displayed setting.
• Back - Goes back to the previous sub-menu or menu item.
• Exit - Immediately exits the Menu mode.
3.4.3Changing Menu Settings
Enter the menu by holding down the
released, the first available sub-menu is displayed.
• Navigate to the desired sub-menu using the No button.
• Enter the sub-menu using the Yes button.
• Navigate to the desired menu item using the No button.
• Enter the desired menu item using the Yes button.
• Change the setting using the No
• Continue navigation using the Yes, No and Back buttons or return to weighing using the Exit button.
Menu
button until Menu is displayed. When the button is
button or accept the setting using the Yes
button.
3.5 Application Modes
Two application modes are available : Weigh and Checkweigh. To enable the Checkweigh mode,
Checkweigh must be turned on in the Mode menu (refer to Section 3.7.5).
3.5.1Weighing
Determines the weight of an item in the selected unit of measure. Scale is shipped with kilograms
(kg) and pounds (lb) enabled. To enable other units of measure, they must be turned on in the
Unit menu (refer to Section 3.7.4).
3.5.2C h e ckw e i gh i ng
Use this mode to compare the weight of an item to Under and Over target limits. When the checkweigh
mode is enabled, the “CHECK” indicator will turn on.
To set the target limits, refer to Section 3.6 Checkweigh Target Settings.
Place the sample on the weighing pan. If the sample weight is under the target weight range, the
yellow LED will light. If the sample is within the target weight range, the green LED will light. If the
sample is over the target weight range, the red LED will light. The LED indication is complemented by
an audible alert feature (refer to Section 3.7.5).
Normal Checkweighing
Set both targets as positive values, where the lightest acceptable weight is entered as the Under target
limit and the heaviest acceptable weight is entered as the Over target limit. Items placed on the pan
will be compared to the targets.
Valor 5000 Series
Negative Checkweighing
In “Negative” checkweighing, the part of an item subsequently removed from the pan is checked
against the target limits. This method may be preferred in portioning applications.
• Set both targets as negative values, where the lightest acceptable weight is entered as the Under
target limit and the heaviest acceptable weight is entered as the Over target limit.
• Place a quantity of items on the pan.
• Tare the items.
• Remove a portion of the items from the pan.
• The amount removed will be compared to the targets.
“Zero” Checkweighing
In ”Zero” checkweighing, displayed results will show the deviation of the sample’s weight from a
reference or ideal weight. This approach may be more preferred for grading or sorting applications.
• Set the targets using a negative value (negative tolerance) for the Under limit and a positive value
(positive tolerance) for the Over limit.
• Place the reference weight on the pan.
• Tare the reference weight and remove it from the pan.
• Place the item on the pan to be checked against the reference weight.
• The display will show a negative, positive or ‘0” value if the sample is less than, more than or equal
to the reference weight, respectively.
EN-12
3.6 Checkweigh Target Settings
To set the target limits, begin by pressing the TARGET button. The value of the Under target limit will
be displayed first, with the left-most digit blinking.
::
Note
: The scale must be in checkweigh mode to be able to enter or modify target values.
::
• To change the Under target limit, press the No button to increment the flashing digit, the Yes
to shift to the next digit and the Exit button to change the sign.
• Press the TARE button to toggle the display between a positive and negative value.
• To accept the Under target limit, press the TARGET button.
• Next, the value of the Over target limit will be displayed, with the left-most digit blinking.
• Repeat the procedure above to set the Over target limit.
• When the Over target limit has been set, press the TARGET button to store the settings and begin
Checkweighing.
button
EN-13 Valor 5000 Series
3.7 Scale Settings
Enter the Menu mode. Proceed to a menu item and use the Yes and No buttons to view and select
available scale settings. When the desired changes have been made, press Exit to return to weighing.
Refer to Section 3.4 for more information on how to enter and navigate the menus.
3.7.1Calibration sub-menu
Note: Calibration is not available when the scale has been set to Legal for Trade operation (LFT set to
ON in the Setup sub-menu, refer to Section 3.7.2).
Calibration Types – Calibration can be adjusted in two ways, Span and Linearity. Span calibration
uses two calibration points to adjust the sensitivity of the scale. Linearity calibration uses three
calibration points to correct for non-linear weighing results.
Calibration Weights – Before beginning calibration, make sure that the required calibration weights are
available. The model-specific calibration points are listed in the table below.
TABLE 3-3. CALIBRATION POINTS.
Model Calibration Unit Span Cal. points Linearity cal. points
V51P3 / V51PH3 kg 1, 2 or 3 kg 1 or 2 and 3 kg
V51P6 / V51PH6 kg 2, 5 or 6 kg 2 or 4 and 6 kg
V51P15 / V51PH15 kg 5, 10 or 15 kg 5 or 10 and 15 kg
V51P30 / V51PH30 kg 10, 20 or 30kg 10 or 20 and 30 kg
Notes:
1. Default calibration points are shown in bold print.
2. Calibration Units kg or lb are selected in the Setup sub-menu (Section 3.7.2.)
3. ASTM Class 4 or OIML Class F2 weights are required for calibration.
lb 2, 5 or 6 lb2 or 4 and 6 lb
lb 5, 10 or 15 lb5 or 10 and 15 lb
lb 10, 20 or 30 lb10 or 20 and 30 lb
lb 10, 30 or 60lb 20 or 30 and 60 lb
Valor 5000 Series
k
k
k
k
k
Span Calibration Procedure
• When CAL is displayed, press Yes to enter the calibration
sub-menu.
• When Span is displayed, press Yes to begin the Span
calibration process.
• Clear the pan if prompted to do so, then press Yes
• The display will show “--C--”, followed by the span
calibration weight value.
• If desired, press No to change to an alternate span
calibration weight value.
• Place the indicated calibration weight on the scale and
press Yes.
• The display will show “--C--”, then “done” and then
revert to weighing. Remove the calibration weight.
Note: The calibration procedure can be canceled at any
time by pressing the Exit button or turning the scale off.
Linearity Calibration Procedure
• When LIN is displayed, press Yes to begin the linearity
calibration process.
• Clear the pan if prompted to do so, then press Yes.
• The display will show “--C--”, followed by the first
calibration weight value.
• If desired, press No to change to an alternate first
calibration weight value.
• Place the indicated calibration weight on the scale
and press Yes.
• The display will show “--C--”, followed by the second
calibration weight value.
• Place the indicated calibration weight on the scale
and press Yes.
• The display will show “--C--”, then “done” and then
revert to weighing. Remove the calibration weight.
Note: The calibration procedure can be canceled at any
time by pressing the Exit button or turning the scale off.
..
.
..
EN-14
End
• Press Yes to leave this sub-menu and go to the next.
• Press No to return to the top menu item in this sub-menu.
EN-15 Valor 5000 Series
k
3.7.2Setup sub-menu
Press Yes to enter the Setup sub-menu.
Reset
Press Yes to enter the Reset menu.
• No - does not change the Setup sub-menu settings.
• Yes - reset the Setup sub-menu to the factory
default settings (bold and underlined text herein).
Calibration Unit
Press Yes to enter the Calibration Unit menu.
• kg - calibrate using kilogram weights.
• lb - calibrate using pound weights.
Legal for Trade
Press Yes
to enter the LFT menu.
• Off - set the scale for non-Legal for Trade operation.
• On - set the scale for Legal for Trade operation.
Note: See section 3.8 for information related to Legal
for Trade operation.
Beeper
Press Yes
to enter the Beeper menu.
• Off - beeper does not sound when a button is pressed.
• On - beeper sounds when a button is pressed.
Volume
Press Yes
to enter the Beeper Volume menu.
• Soft - volume of beeper is low.
• Loud - volume of beeper is high.
End
• Press Yes to leave this sub-menu and go to the next.
• Press No to return to the top menu item in this sub-
menu.
Valor 5000 Series
3.7.3 Readout sub-menu
YesYes
Press
Yes to enter the Readout sub-menu.
YesYes
Reset
Press Yes to enter the Reset menu.
• No - does not change the Readout sub-menu settings.
• Yes - reset the Readout sub-menu to the factory
default settings (bold and underlined text herein).
Averaging Level
Press Yes to enter the Averaging Level menu.
• Lo - less filtering (faster stabilization time).
• Hi - more filtering (longer stabilization time).
Light
Press Yes to enter the Backlight menu.
• Off - turn display backlight off.
• On - turn display backlight on.
• Auto - automatically turns the display backlight on when
the scale is active and off after 5 seconds of inactivity.
Auto Off
Press Yes to enter the Auto-off menu.
• Off - disable auto off.
• On - turn the scale off after 5 minutes of inactivity.
EN-16
Auto Zero tracking
Press Yes to enter the Auto Zero Tracking menu.
• 0.5 d - 1/2 division capture range.
• 1 d - 1 division capture range.
• 3 d - 3 division capture range.
• Off - disable Auto Zero Tracking.
Auto Tare
Press Yes to enter the Automatic Tare menu.
• Off - disable Automatic Tare.
• On - enable Automatic Tare.
End
• Press Yes to leave this sub-menu and go to the next.
• Press No to return to the top menu item in this sub-menu.
EN-17 Valor 5000 Series
k
P
3.7.4Unit sub-menu
YesYes
Press
Yes to enter the Unit sub-menu.
YesYes
Reset
Press Yes to enter the Reset menu.
• No - does not change the Unit sub-menu settings.
• Yes - reset the Unit sub-menu to the factory default
settings (bold and underlined text herein).
Gram unit
YesYes
Press
Yes to enter the Gram unit menu.
YesYes
• Off - disable grams.
• On - enable grams.
Kilogram unit
YesYes
Press
Yes to enter the Kilogram unit menu.
YesYes
• Off - disable kilograms.
• On - enable kilograms
Pound unit
YesYes
Press
Yes to enter the Pound unit menu.
YesYes
• Off - disable pounds.
• On - enable pounds.
Ounce unit
YesYes
Press
Yes to enter the Ounce unit menu.
YesYes
• Off - disable ounces.
• dEC - enable decimal ounces.
• FrAC - enable fractional ounces.
Pound:Ounce unit
NoNo
menu.
*Use P to measure the weight of a sample as a % of ref.
weight. When P is selected diplay will show S.rEF and
P.Press “No” to use ref. weight in memory,”Yes” to establish
new ref.After placing reference material, press “Yes” to
establish 100% ref.Replace ref. material with new sample
to read percentage of the sample compared to ref. weight
Valor 5000 Series
3.7.5Mode sub-menu
YesYes
Press
Yes to enter the Mode sub-menu.
YesYes
Reset
Press Yes to enter the Reset menu.
• No - does not change the Mode sub-menu settings.
• Yes - reset the Mode sub-menu to the factory
default settings (bold and underlined text herein).
Checkweigh mode
YesYes
Press
Yes to enter the Checkweigh menu.
YesYes
• Off - disable the Checkweighing mode.
• On - enable the Checkweighing mode.
Alert
Press Yes to enter the Checkweigh Alert menu.
• Off - disable alert beeping.
• OK - beep when the green Accept light is on.
• Lo - beep when the yellow Under light is on.
• Hi - beep when the red Over light is on.
• Lo-Hi - beep when the yellow Under or red Over light is on.
• Change - beep once when the green Accept light is on, twice
when the red Over light is on.
EN-18
End
• Press Yes to leave this sub-menu and go to the next.
• Press No to return to the top menu item in this sub menu.
EN-19 Valor 5000 Series
3.7.6Lockout sub-menu
YesYes
Press
Yes to enter the Lockout sub-menu.
YesYes
Reset
Press Yes to enter the Reset sub-menu.
• No - does not change the Lockout sub-menu settings.
• Yes - reset the Lockout sub-menu to the factory default
settings (bold and underlined text herein).
Lock Calibration
YesYes
Press
Yes to enter the Lock Calibration sub-menu.
YesYes
• Off - enable changes to the Calibration sub-menu.
• On - lock Calibration sub-menu settings.
Lock Setup
Press Yes to enter the Lock Setup sub-menu.
• Off - enable changes to the Setup sub-menu.
• On - lock Setup sub-menu settings.
Lock Readout
YesYes
Press
Yes to enter the Lock Readout sub-menu.
YesYes
• Off - enable changes to the Readout sub-menu.
• On - lock Readout sub-menu settings.
Lock Unit
Press Yes to enter the Lock Unit sub-menu.
• Off - enable changes to the Unit sub-menu.
• On - lock Unit sub-menu settings.
Lock Mode
Press Yes to enter the Lock Mode sub-menu.
• Off - enable changes to the Mode sub-menu.
• On - lock Mode sub-menu settings.
End
YesYes
• Press
Yes to leave this sub-menu and go to the next.
YesYes
NoNo
• Press
No to return to the top menu item in this sub-menu.
NoNo
Valor 5000 Series
EN-20
3.8 Legal for Trade Settings
The Scale is designed to comply with OIML, EEC, weights and measures regulations. These
approvals may be pending; contact Ohaus for further details about availability.
Note:
Consult your local Weights and measures office for their regulations before placing the scale into a
Legal for Trade application.
When the LFT menu item setting is On, the following conditions apply:
• The Calibration (CAL) sub-menu is hidden.
• The Calibration Lock (LCAL) menu item is hidden.
• The Auto Zero Tracking (AZt) menu item is set to 0.5 d.
• lb:oz and P units are disabled.
When the weights and measures official has approved the scale for use, it should be sealed as
shown in Section 3.9.
3.9 Sealing the Scale
When used in conjunction with the Lockout menu and the security switch, the scale may be sealed
to prevent or detect unauthorized changes to scale settings. For Legal for Trade applications, set
the LFT menu item to ON, set the security slide to the On position (by sliding the switch to cover the
screw), then seal the scale to prevent access to the metrological parameters. Seal the scale
according to the local weights and measures regulations.
Sliding the switch to the ON position in V51PHXX models will freeze the menu item settings in their
current state
To regain access to the scale settings, break the seal and set the security slide to the OFF position.
ON OFF
Security Slide
Figure 3-4. Security Switch Settings.
EN-21 Valor 5000 Series
4MAINTENANCE
4.1 Calibration
Verify calibration periodically by placing either a ASTM Class 4 weight or an OIML Class F2 weight
on the scale and compare the reading to the specifications. Refer to Section 3.7.1 for calibration
procedure.
4.2 Cleaning
• Disconnect the scale from the power supply before cleaning.
• The housing and pan should be kept clean and free from foreign material.
• Wipe the exterior components of the scale using a soft cloth dampened with water and a mild
detergent (do not use acids, alkalis or strong solvents).
• The leveling feet and load pads may be unscrewed and removed for cleaning.
• Do not allow liquids to enter the scale.
• Remove and clean the pan separately from the scale.
• The pan is dishwasher safe and may be placed with normal loads. All other parts must not be
placed in a dishwasher.
4.3 Troubleshooting
The following table lists common problems, possible causes and remedies. If the problem persists,
contact Ohaus or your authorized Ohaus dealer.
TABLE 4-1. TROUBLESHOOTING.
Symptom
Scale will not turn on.AC power not connected.
Battery indicator is low.Battery discharged.
Possible CausesRemedy
Connect scale to power.
Battery discharged.
Connect scale to power and charge battery.
Connect scale to power and charge battery.
Battery fails to charge fully.
Weight reading does not
stabilize.
Scale does not display
accurately.
Cannot calibrate the scale.
Battery is defective.
Unstable environment.
Interference under pan.
Improper calibration.Calibrate scale.
Calibration locked.
Have the battery replaced by an authorized
Ohaus service dealer.
Ensure a stable environment.
Ensure that the weighing pan is unobstructed
and free to move.
Check menu settings and recalibrate.
If error persists, service is required.
Calibrate using correct calibration
weights.
Remedy
EN-22
4.4 Service Information
If the Troubleshooting section does not resolve or describe your problem, contact your authorized
Ohaus Service Agent. For Service assistance in the United States and Canada, please call Ohaus
Corporation toll free at (800) 526-0659 from 8 AM to 5 PM Eastern Time. An Ohaus Product Service
Specialist will be available to help you.
4.5 Accessories
DescriptionPart Number
Foot Switch71169633
Loading...
+ 64 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.