OEC Mini 6600 User manual

Page 1
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance
MINI6600
TM
Digital Mobile C-Arm

Periodic Maintenance Procedure

00-879519-01
OEC Medical Systems, Inc.
384 Wright Brothers Drive
Salt Lake City, Utah 84116
Page 2
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance
This manual may not be reproduced, in whole or in part,
without the written permission of
OEC Medical Systems, Inc.
This procedure does not contain any steps which can be performed by the system operator.
The steps contained in this procedure should not be attempted by anyone
who is not specifically trained or authorized by OEC to work on the
MINI6600TM system.
The contents of this procedure are accurate at the time of publication. However, changes in design and additional
features may be incorporated in the hardware and software which are not reflected in this version of the proce-
dure. Contact OEC Technical Support for clarification, if discrepancies arise.
The MINI6600 Digital Mobile C-Arm is
manufactured under the following U.S. Patents:
4.768.216;
4.797.907;
4.209.706.
Other Patents Pending.
Revision History Rev Dash Date Change Description
1 -01 Jul 96 Initial Publication A -01 Aug 96 Manufacturing Release
Dec. 97 CD-ROM Release
Page 3
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

General Instructions

Use the following procedures to perform periodic maintenance on the MINI6600 Mobile Imaging System. The MINI6600 Periodic Maintenance Form must be completed as the procedures in this manual are performed. If the system does not pass a portion of the procedure, correct the problem before completing the PM Form. Note all failures and the corrective action taken in the comments section of the PM Form. Additional space is provided at the end of the PM Form.
The tools and test equipment that are required to perform Periodic Maintenance, are described in Appendix A. All Torx screws have recommended torque specifications. These specifications are described in Appendix A.
IMPORT ANT:
The PM Procedures do NOT list all the steps necessary to complete the PM. The service engineer WILL NEED the Service Manual (00-878315) to complete the PM.
Page 4
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

SAFETY

Ingress Of Water And Solutions

The system must never be operated or stored in locations where conductive fluids, like water and saline solution, might spill on the equipment.
Always unplug the AC power cable from the wall outlet before cleaning the equipment. Do not allow water, soap, or other liquids to drip into the equipment and cause short circuits, electric shock and fire hazards.

Mechanical Operation

The Articulating arm is manually adjusted.

T rained Service Personnel

Serious injury and property damage can result from incorrectly performed service procedures. Observe all operat­ing and safety procedures contained within this manual.
W ARNING:
Procedures contained in this manual should only be performed by service personnel specifically trained by OEC Medical Systems, Inc. to calibrate the MINI6600 System.
Page 5
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Electrical Shock

W ARNING:
This system is capable of generating lethal voltages. Observe safe electrical testing proce­dures when performing periodic maintenance procedures on the system.

X-Radiation Exposure

The X-ray tube assembly produces X-radiation when energized. Never operate this device without X-ray shielding in place. Use lead shielding and draping to protect personnel.
W ARNING:
Lead aprons, radiation monitors and appropriate radiation shields must be utilized by person­nel while performing tests.
CAUTION:
Living human anatomy should never be used as a phantom or demonstration aid.
Many of the PCB’ s in this system contain components which are sensitive to Electro-Static Discharge (ESD). Observe ESD safety procedures.
Page 6
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance
W ARNING: Labels
The following warning labels may be found on the system:
X-RAY SAFETY WARNING:

W ARNING: Symbols

The following symbols may be found on the MINI6600:
DANGEROUS VOL TAGE PRESENT
Dangerous voltages are present use safety precautions.
DANGEROUS VOL TAGE PRESENT
Dangerous voltages are present - use safety precautions.
Page 7
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance
EARTH GROUND
The protective earth ground should be the last electrical connection broken and the first electrical connection made during servicing procedures.
POTENTIAL EQUALIZATION
GRN/YEL wire is used to indicate protective earth conductors, accessible parts connected to earth parts, and potential equalization conductors.
EMITTING X-RAY SOURCE
This symbol indicates the presence or potential of producing ionizing X-radiation.
Page 8
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance
!
NON-A/P
This symbol indicates that equipment is non-anesthetic proof and should not be operated in the presence of flam­mable vapors, liquids or other substances.
ATTENTION
This symbol cautions you to refer to the service manual for additional information.
Page 9
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Completing the PM Form

1. At the top of the MINI6600 Periodic Maintenance Form, check whether you are performing a Semi­Annual or an Annual PM.
2. Completely fill in the customer information block, including the T ime Of Day (start time) measured in military time, System Serial Number and the Field Service Report (FSR) / Request ID Number . Indicate the system configuration with appropriate check marks.
NOTE:
T est equipment used must have current calibration dates. Do not use test equipment with expired calibration stickers.
3. While performing the Periodic Maintenance procedures, log all test equipment used in the Test Equip­ment Used block of the Periodic Maintenance Form.
4. Place a check mark on the PM Form as each test paragraph is completed. Fill in actual measured val­ues where indicated on the PM Form. If adjustments are made, note the adjustment in the comments section of the PM Form.
Page 10
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Begin Periodic Maintenance

Remove the Power And Covers

W ARNING:
Electrical circuits inside the equipment use voltages that are capable of causing serious injury or death from electrical shock. Use appropriate precautions.
1. If the system has power applied, position the System keyswitch/pushbutton to the OFF position. V erify that the System’s AC power cord is not plugged into the wall outlet.
2. Remove the System covers. Refer to the System Overview section in the MINI6600 Digital Mobile C­Arm Service Manual (00-878315) for cover removal procedures.

Elapsed Time Indicator

Read the elapsed time meter on the System and enter the number on the PM Form.

Inspect the Static Drag Wire

1. V erify that the static drag wire underneath the System is touching the floor.
2. V erify the static drag wire is securely attached.

Inspect the Articulating Arm and Brake

V erify that the articulating arm mechanics move freely and smoothly throughout the entire range of motion.
Page 11
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Inspect the Rotation Brake

1. Release the rotation brake and rotate the MINI-C assembly.
2. V erify smooth rotation.
3. V erify that the brake locks the mechanics in any position within the range of motion.

Inspect the Orbital Brake

1. Release the orbital brake.
2. Move the MINI-C throughout the orbital range. Ensure smooth operation.
3. V erify that the brake locks the mechanics in any position within the range of motion.

Verify the System Steering And Brake Operations

1. Select the LOCK mode by pressing the steering brake pedal completely down. V erify that the casters and wheels are locked in position and do not move.
2. Select the FREE mode by lifting the steering pedal up halfway . Verify that the casters rotate and the wheels roll freely .
3. Select the STEER steering mode by lifting the steering pedal completely up. V erify that the casters rotate into the steering position and lock, and the wheels roll freely.
Page 12
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Inspect/Clean the System Air Filter

Inspect and clean the System air filter attached to the rear panel.

Inspect the AC Power Cable

Inspect the System AC power cable for damage and wear .
Inspect the Cables, Connectors and PCB’s
1. Inspect all the cables for damage. Look for signs of abrasion to the cable jacket. If abrasions are de­tected, replace the cable.
2. Inspect all connectors. Ensure that all connectors are seated securely.
3. Ensure all PCB’s are seated securely.

Measure the Ground Continuity

W ARNING:
Do not perform this continuity test with the AC plug plugged into a power receptacle.
Verify the system power cord is disconnected from the AC receptacle.
From To Spec
System Equipotentiality Stud (Underneath the bottom right side) AC Power Plug Ground Pin < 0.15 Ω
TABLE 1
Page 13
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Power Up the System

1. Plug the AC power cord into the wall receptacle.
2. Turn or press the System keyswitch/pushbutton to ON.
3. The system will perform a Power ON self-test during power up.

Cart Progress and Error Codes

1. Failure of the self-test will be indicated by error codes on the System rear panel LED’s.
2. If any code remains and the system fails to complete the Power ON sequence, refer to your Service Manual (00-878315).
3. V erify that the elapsed time indicator is running.
NOTE:
The elapsed time indicator will increment every 6 minutes or 1/10th of an hour . If the indicator is not running, initiate repairs before continuing Periodic Maintenance.

System Virus Inspection

1. Turn or press the system keyswitch/pushbutton to the OFF position.
2. Access DOS by inserting a DOS disk in drive A and turn or press the keyswitch/pushbutton to the ON position.
3. When you see A:\> remove the DOS Disk and insert the Virus Scan Disk.
4. Run the virus scan program.
5. Repair/clean any files found infected. Remove the virus scan disk.
6. Reboot the system.
Page 14
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Measure the Line V oltage Regulation

W ARNING:
This procedure produces high energy X-rays. Even though the system radiation levels are lower than other OEC C-arms, wear protective radiation clothing. Because of the radiation produced, non-service personnel should not be present in the room during any of the following tests.
1. Measure the AC line voltage at the wall receptacle.
Vn=_____
2. V erify that the white/blue wire is connected to Primary 1 COM and the blue wire is connected to Pri­mary 2 COM.
3. Ensure that the brown wire and the white/brown wire are properly strapped to match the input line volt­age, measured in step 1, as indicated by the following table:
NOTES:
(A) Additional information regarding the transformer taps can be found in the
Power Distribution section of the Service Manual (00-878315).
(B) In the next step, the voltage measured at the wall receptacle may drop during the fluoroscopic exposure. Read the meter at this time.
(C) Place a lead apron over the I.I. to prevent any undo wear and tear to the image tube.
Transformer Tap 105V 1 15V 125V Input V olt Range 98 to 107 108 to 1 1 7 118 to 127
TABLE 2
Page 15
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance
4. Measure the line voltage at the wall receptacle while making a fluoroscopic exposure at 98µA @ 75kV (Max load voltage).
V1=_____
5. Use the following formula to calculate the percentage of line voltage regulation and record the results below.
Line Regulation = 100 X (
where: Vn = No load voltage
V1 = Max load voltage
Regulation = _________%
6. If the results above do not fall within the range of 0 to 15 percent, the hospital should be informed that an in-house wiring problem exists that should be corrected as soon as possible. Note this on the FSR. (This problem results in a violation of a provision of the Federal Performance Standards 21 CFR
1020.30)
Vn - V1)
V1
Page 16
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Inspect/Adjust the System Power Supply Voltages

W ARNING:
Electrical circuits inside the equipment use voltages that are capable of causing serious injury or death from electrical shock. Use appropriate precautions.

PS4 (24V)

1. Remove the plexiglass shield preventing access to PS4 (refer to the Power Distribution section,
FIGURE 2 of the service manual (00-878315) for the location of PS4). Record the values measured in
the following steps on the Periodic Maintenance Form.
2. Measure the +24V on the monitor circuit boards at J6 between the black (GND) and yellow (+24V) wires. Adjust R1 1 on PS4 to +24 VDC (±1V) if necessary. Refer to FIGURE 1 for the location of R11.
BLACK YELLOW
WHITE/REDWHITE
DO NOT ADJUST
R6
R11 V ADJ
FIGURE 1 - SYSTEM POWER SUPPLY PS4 +24VDC
Page 17
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

PS1 (5V and 12V)

1. Measure the voltages at the following test points on the Video Switching PCB and adjust as required with R53 (+5) and R21 (+12) on PS1. Refer to FIGURE 2 in the Power Distribution section of the service manual (00-
878315) for the location of PS1. Refer to FIGURE 2 below for locations of R53 and R21.
2. The Video Switching PCB is located on the inside of the hinged rear panel of the electronics box (Refer to
FIGURE 4 in the Power Distribution section of the service manual (00-878315)).
GND TO ADJUST MEASUREMENT TOLERANCE
TP25 TP24 Not Adjustable - 5.0 VDC ±0.75 VDC TP25 TP27 Adjust PS1 R53 (OUT1) + 5.0 VDC ±0.75 VDC TP25 TP28 Adjust PS1 R21 (OUT 2) +12 VDC ±1 VDC TP25 TP19 Not Adjustable - 12 VDC ±1 VDC
TABLE 3
F1
250V 4A
FAST BLOW
C5
C51
OUT 2 V ADJ
R21
OUT 1 V ADJ
C18
C32 C29 C42
C30
C28
R53
C31
R43
PS1 P3
DO NOT ADJUST
FIGURE 2 - System Power Supply PS1 (+5V, -5V, +12V, -12V)
Page 18
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Inspect/Adjust the C-Arm Power Supply Voltages (PS2)

W ARNING:
Electrical circuits inside the equipment use voltages that are capable of causing serious injury or death from electrical shock. Use appropriate precautions.
1. Refer to the PROCEDURES section of the service manual (00-878315) for accessing the 6600 Controller PCB. Refer to the Power Distribution section in the service manual (00-878315) for the location of PS2.
2. PS2 output voltages are adjusted using the voltage adjustment pot shown in FIGURE 3. Power supply voltages are measured on the 6600 Controller PCB at the test points listed below:
- Lead + Lead Supply Tolerance Reading Range
NOTE:
TP2 TP42 +24.0 VDC +10%/-5% +22.8 to +26.4 VDC TP2 TP37 +12.0 VDC ±3% +11.6 to +12.4 VDC TP2 TP23 +5.0 VDC ±2% +4.9 to +5.1 VDC (VCC) TP2 TP39 +5.00 VDC ±2% +4.9 to +5.1 VDC (+5.00 V Ref) TP2 TP43 -10.00 VDC ±2% -9.8 to -10.2 VDC (-10.00 V Ref) TP2 TP44 -12.0 VDC ±3% -11.6 to -12.4 VDC
TABLE 4
The 6600 Controller PCB contains a circuit breaker (CB1) to protect the +24 volt supply to the monoblock X-ray source.
Page 19
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance
VOLTAGE
PIN 1
J2
PIN 1
J1
F1
C6
ADJUSTMENT
POWER FAIL DETECT CIRCUIT
T1
L1
1 +5 2 +5 3 4 D-GND 5 D-GND 6 7 A-GND 8 +12 9 10 -12 11 KEY 12 +24
1 GROUND 3 NEUTRAL 5 LINE
FIGURE 3 - C-Arm Power Supply PS2 (+5V, +15V, -15V, +24V)
Page 20
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Inspect the Service Outlet Strip

1. Check for 115 V AC ± 5 VAC at the service outlet strip located below the monitors at the rear of the System.
2. Check for 230 VAC ± 10 VAC at the service outlet strip located below the monitors at the rear of the System.

Perform the System Diagnostic T ests

1. Insert the Service Disk into the disk drive. Same service disk as used for the Series 9600 C-Arm (00-
877444).
2. Press SETUP OPTIONS and access Level 2 diagnostics. Select the Setup Options menu, and then select Run System Diagnostics. The following tests will appear:
••
• Front Panel Test
••
• System Memory Tests (A T Motherboard Memory)
• Image Processor Tests (IP PCB Memory)
• Toggle Motion Artifact (Toggles ON/OFF), Non Functional
3. Select the Front
Panel Test and press the following switches and verify that a hex code is displayed and an audible tone is heard for each switch pressed: Negate, VCR Play , VCR Record, Auto, Help, Brightness and Contrast, Fluoro, One Shot, Enhanced. Averaging, L-R.
4. V erify that the remaining tests pass successfully. Press the ENTER key to toggle through tests.
5. Remove the Service Disk from the disk drive.
6. Pressing ESC will exit the System Diagnostic tests and reboot the System.
Page 21
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Verify the System Controls

W ARNING:
Steps within this procedure produce X-rays. Use appropriate precautions.
1. Connect the footswitch and the handswitch (optional) to the system. Confirm the operation of each switch.
2. Place the Beam Alignment Tool on the Image Intensifier .
3. V erify the CAMERA rotation operation.
4. V erify the camera REVERSAL operation.
5. V erify the NEGATE operation.
6. V erify the MANUAL BRIGHTNESS and CONTRAST operations.
7. V erify the AUT O BRIGHTNESS AND CONTRAST operations.
8. V erify the MANUAL MODE kV/ µA operation.
9. V erify the AUTO NORMAL MODE operation.
10.Verify AUTO OSTEO/PED MODE operation.
11. V erify the image SAVE operation. Not applicable if system is equipped with 4 image storage capability.
12.Verify the VCR RECORD and the VCR PLAY operations (if applicable).
13.Verify the ONE-SHOT MODE.
a. System will fluoro until the video stable condition is achieved and then terminate the exposure.
14. Verify the ENHANCED MODE.
a. System will fluoro until the video stable condition is achieved and then terminates the exposure as well as enhances edge.
Page 22
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance
15.Verify the A VERAGING operation.
No Lamps Lit Intrinsic Averaging 2 Frames 1 Lamp Lit Low Averaging 4 Frames 2 Lamps Lit Medium A veraging 8 Frames 3 Lamps Lit High Averaging 16 Frames
TABLE 5
16. V erify the PATIENT ANNOTATION operation.
17. Verify the DISK VIEW operation (16 image option systems).
NOTE:
18. Verify the TEST P ATTERN operation.
19. Verify the ALARM RESET operation.
20. Verify L-R operation.
21. Verify the HELP operation.
22. Verify the SET-UP OPTIONS operation.
The image directory , markers, X-ray summary, special apps and customize features are not operational. Attempts at selecting these features will produce a negative beep.
Page 23
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Perform the Accessory Operational Test

1. Test the operation of any accessories which may be configured with the system; such as a thermal printer or VCR.

Inspect/Adjust the Gen-Lock Calibration

1. Observe LED’s DS1 and DS2 on the V ideo Switching board. If both LED’s are illuminated, the gen-lock circuit is calibrated.
2. If both DS1 and DS2 are not equally illuminated, adjust tuning capacitor C89, on the Video Switching PCB, (using a non-conductive alignment tool) until both LED’s are illuminated equally. This verifies that the gen-lock circuit is tuned to the center of its operating range.
Page 24
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Perform the Camera And Beam Alignment

W ARNING:
Steps within this procedure produce X-rays. Use appropriate precautions.
Refer to the PROCEDURES section in the Service Manual (00-878315). Complete the following procedures:
CCD Camera Preparation Camera Focus Image Intensifier Focus Monitor/Camera Alignment Pattern Confirm Monitor Size & Linearity Camera Alignment Beam Alignment Beam Alignment V erification Window Placement V erification Set Camera Iris and Video Gain Align and Check Rotation Indicator V erify Resolution
Page 25
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Check the Entrance Exposure

W ARNING:
Steps within this procedure produce X-rays. Use appropriate precautions.
Refer to the PROCEDURES section in the Service Manual (00-878315).
Attention:
Stop here if you are performing a semi-annual PM. Proceed directly to Cleanup.
Complete the Annual Maintenance section if you are performing
an annual Periodic Maintenance.
Page 26
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Annual Maintenance

WARNING:
Steps within this procedure produce X-rays. Use appropriate precautions.
Complete all semi-annual PM procedures.
Refer to the PROCEDURES section of the Service Manual (00-878315). Complete the following sections:
Accessing 6600 Controller PCB Test Points Indicated vs. Actual T echnique V erification
Record Technique V erification readings on the PM Form.
Page 27
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Clean-up

WARNING:
Steps within this procedure produce X-rays. Use appropriate precautions.
1. Use the compressed air to remove any dust particles from inside the system.
2. Reboot and verify system operation.
3. Replace all covers and wipe down the System.
Page 28
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Complete the PM Form

1. V erify that the information blocks have been completed.
2. V erify that each procedural step has been checked.
3. Attach all films to the Periodic Maintenance Form cover .
4. Sign and date the PM Form.
5. Obtain the customers signature and date. Leave the customer’s copy with the customer.
6. Mail the form and film(s) to:
OEC MEDICAL SYSTEMS, INC. ATTENTION: TECHNICAL SUPPORT 384 WRIGHT BROTHERS DRIVE SALT LAKE CITY, UTAH 84116
Page 29
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

APPENDIX A: Tool & Equipment List

Required T ools

In addition to common hand tools the following tools will be required:
Torx Screwdriver Set
1.5mm allen wrench
Jewelers Screwdriver Set.

Required T est Equipment

The test equipment required to complete the PM include:
Oscilloscope Tektronix 2236 or equivalent Digital V olt-Ohm-Meter Dosimeter (Keithley Model 35050, PMX II non-invasive meter or equivalent) Model 96030 ion chamber Resolution Tool, Converging Line Pair (LP/mm), OEC P/N 00-900860 Resolution Tool, Mesh (LP/Inch), OEC P/N 00-900040 Ion Chamber Positioning Fixture Lead Apron 1 mm Copper Filter (3 each required), OEC P/N 00-877682 MINI6600/9600 Service Disk (software), OEC P/N 00-877444 Beam Alignment Tool, OEC P/N 00-879272 Canned Air, OEC P/N 88-299311 TP35 Cleaner , OEC P/N 88-299318.
Page 30
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

Paint Kits

PART NUMBER DESCRIPTION
900524-11 Paint Kit, Elite Gray , Texture 900524-12 Paint Kit, Rhinestone, Texture 900524-13 Paint Kit, Windswept, T exture
TABLE 6

T orx Crestcup Torque Specifications

TORX SCREW SIZE NOMINAL TORQUE VALUE ACCEPTABLE RANGE
4-40 6 lbs./in 5-8 lbs./in 6-32 15 lbs./in 13-16 lbs./in 8-32 22 lbs./in 21-24 lbs./in 10-32 30 lbs./in 28-31 lbs./in
TABLE 7
Page 31
MINI6600TM Mobile Digital C-Arm Service Manual - Periodic Maintenance

APPENDIX B: X-Ray Verification Requirements

X-radiation output must be verified, annually, during regular Periodic Maintenance procedures or when certified components are replaced. Certified components used in the MINI6600 Mobile Imaging System include the follow­ing:
• Monoblock Tube Assembly
• Image Intensifier Assembly
• Beam Limiting Device
• X-ray Control Assembly
*
*
*
* Complete Report of Assembly , FDA Form 2579, whenever any of the preceding assemblies are
replaced in their entirety, per 21 CFR 1020.30 (d) (1) and (2).
The measurement values used for these tests are not the design specifications. The test specifications used here take into account measurement limitations and their use will ensure that the design specifications are met.
Page 32
MINI6600
Print 3 Pages
Periodic Maintenance Inspection
Performed at:
(Hospital Name)
(Street Address)
(City) (State) (Zip)
Performed on: / / System S/N
(Date)
Performed by:
(Printed Name of Service Engineer) (Emp. # or Dealer Name)
Type of PMI Performed: Check One
❑ ❑
Semi-Annual Annual
Return this booklette with the Yellow copies intact and the Beam Alignment Films attached, to:
OEC Medical Systems, Inc. 384 Wright Brothers Dr. Salt Lake City, UT. 84116 Attn: Technical Support Dept.
✉ ✉
Page 33
Attach Beam Alignment Films Here
☛ (Tape or Staple)
REMINDER !
MINI6600 system exposures must clearly indicate the following:
1) Normal field exposure, six or four inch
2) Four inch collimated exposure (Six inch systems).
NOTES:
1. Exposures made on Multi-format cameras, i.e., Lenzar, Visiplex, etc., are NOT acceptable as proof of Radiographic alignment.
2. If any of the above requirements are not met, a re-visit to the hospital will be required, to re-do the films, until such time as they meet the requirements and specifications. In cases where the hospital or clinic does not own or have access to a film processor, beam films will not be required. However, this must be noted on the PMI form. If not noted, it will be cause for rejection, and a Discrepancy Report will be generated and distributed.
Page 34
Time of Day:
Semi-annual PM
Annual PM
MINI6600 IMAGING SYSTEM
PERIODIC MAINTENANCE
Customer Name
Street
City
State
PERIODIC MAINTENANCE PROCEDURES
1. GENERAL INSTRUCTIONS
2. SAFETY
3. COMPLETING THE PM FORM
4. BEGIN PERIODIC MAINTENANCE
Remove Power and Covers
Elapsed Time Indicator
Inspect Static Drag Wire
Inspect the Articulating Arm
Inspect the Rotation Brake
Inspect the Orbital Brake
Verify System Steering and Brake Operation
Inspect/Clean the Workstation Air Filter
Inspect the AC Power Cable
Inspect the Cables, Connectors and PCBs
Measure Ground Continuity
Power Up the System
System Virus Inspection
Measure Line Voltage Regulation
Inspect/Adjust Workstation Power Supply Voltages
Inspect/Adjust Mini C-Arm Power Supply Voltages
Inspect the Service Outlet Strip
Serial No.
Zip
FSR No.
COMMENTS
Semi-annual PM: Perform 1, 2, 3, 4, 6, 7 Annual PM Perform: 1, 2, 3, 4, 5, 6, 7
Elapsed Time Indicator : __________
Line Voltage Regulation = ______% Record Voltages Measured Below
Record Voltages Measured Below Record Voltages Measured Below
WORKSTATION VOLTAGES
+24 VDC (+/- 1.0) ________ VDC
+5 VDC (+/- 0.1) ________ VDC
-5 VDC (+/- 0.75) ________ VDC
+12 VDC (+/- 1.0) ________ VDC
-12 VDC (+/- 1.0) ________ VDC
SYSTEM CONFIGURATION
VCR
:
6" I.I.
S/N _______________ S/N _______________ S/N _______________
4" I.I.
Hard Copy Camera
Thermal Printer
COMMENTS
WORKSTATION POWER STRIP
115 VAC (+/- 5.0) ________ VAC
230 VAC (+/- 10.0) ________ VAC
+12 VDC (+/- 0.36) ________ VDC
+24 VDC (+2.4,-1.2) ________ VDC
TEST EQUIPMENT USED
TYPE
MODEL
MINI C-ARM VOLTAGES
+5 VDC (+/- 0.10) ________ VDC
-12 VDC (+/- 0.36) ________ VDC
-10 VDC (+/- 0.2)________ VDC
S/N
CAL DUE DATE
Customer Signature
12-040-9/96
Date
White - Customer Yellow - Factory
Service Engineer Signature
Pink - Regional Service Center Green - Service Engineer
Employee #
Date
Page 1
Page 35
MINI6600 IMAGING SYSTEM
Periodic Maintenance
PERIODIC MAINTENANCE PROCEDURES
Perform System Diagnostic Tests
Verify the System Controls
Perform Accessory Operational Test
Inspect/Adjust Gen-Lock Calibration
Perform Camera and Beam Alignment
Record Image Resolution
Check Entrance Exposure
RECORD NEW IMAGE RESOLUTION
FIELD SIZE
6" 4"
60 Hz MIN lp/mm
2.5 mm
3.0 mm
PERIODIC MAINTENANCE PROCEDURES
END SEMI-ANNUAL PM
5. ANNUAL PM
Accessing 6600 Controller PCB Test Points
Indicated vs. Actual Technique Verification
7. COMPLETE PM FORM
50 Hz MIN lp/mm
2.2 mm
2.6 mm
6. CLEAN-UP
COMMENTS
Circle One: 60 Hz or 50 Hz
Actual Dose Rate = _______
MEASURED
COMMENTS
Record X-ray Calibration Data
Attach Films
( 0.89 R/min)
£
TEST POINTS
TP14 (GND) TP14 (GND)
TP14 (GND) TP14 (GND) TP14 (GND)
TP14 (GND)
DOSIMETER
TP25
TP15
TP25
TP15 TP25
TP15
COMMENTS:
X-RAY VERIFICATION TEST DATA
TECHNIQUE DATA
75 kVp @ 98.0 micro amp
75 kVp @ 98.0 micro amp 58 kVp @ 35.2 micro amp
58 kVp @ 35.2 micro amp 40 kVp @ 19.6 micro amp 40 kVp @ 19.6 micro amp +0.94 VDC to + 1.02 VDC
75 kVp @ 98.0 micro amp
+7.20 VDC to +7.80 VDC
+4.70 VDC to +5.10 VDC
+5.57 VDC to +6.03 VDC +1.69 VDC to +1.83 VDC
+3.84 VDC to +4.16 VDC
72 kVp to 78.0 kVp (+/- 4%)
TOLERANCE
MEASUREMENTS
VDC VDC VDC
VDC VDC VDC
kVp
12-040-9/96
Page 2
Loading...