Materials:
Biocompatible Black Delrin body bonded with a
Non-leaching solvent acrylic adhesive
to a 50 micron UV transparent film.
Contained volume, 50 uL
Pathlength 0.75 mm
Dead Volume in channels: 1 uL for the inlet channel, 1 uL for outlet
channel
Solvent compatibility:
Ethanol ( up to 99.5%)
Isopropyl alcohol ( up to 91% in water)
Aqueous buffers with or without detergents
DO NOT FILL with Acetone or Aliphatic solvents such as Hexane. Wipe
the outside with lens tissue to avoid scratching the Optical Film.
Figure 1 Figure 2 FluoroVette
Shown below with the Adaptor assembly.
Inlet Outlet
Side Side
FluoroVetteTM Product Specifications
Connectors:
Hose barb connector (CFV-PUMP) is Polysulfone seated in acrylic
Pipette interface (CFV-PIP) is acrylic
FluoroVette Overall dimensions: 4.7 mm tall, 4.3 mm wide, and 1 mm thick.
Adaptor dimensions: 15 mm from base to the center of light path, 1 cm x 1 cm
outer dimension. Insert retaining ring with or without a lens to reduce light scatter
NOTE: The adaptor has a separate retaining ring to allow insertion of a planoconvex lens. Suggested lens; Edmund Optics: www.edmund optics.com
Light Source
LENS PCX 5 X 4.6 MGF2
COM
Directions for use:
The FluoroVette has an inlet sid e, which allows the flowcell to fill from the bottom to
minimize air bubble formation. The inlet side is shown on the left, the outlet side is
shown on the right in Figure 1.
NOTE Insert the FluoroVette
the light source of the fluorometer. Two notches at the top of the adaptor
indicate the light path as shown in Figure 2
Dia.
E.F.L.
5.0
TM
Type
4.6
PCX
TM
with Hose Barb Connections
FluoroVette
Comment
LAKN14
into the adaptor with the inlet side facing
Part #
NT32-403
Price
$19.00
TM
with Pipette Interface
ALine, Inc. PO Box 1068, Redondo Beach, CA 90278
830 Douglas Ave., Dunedin, FL 34698
Customer support: 877-707-8575 727-733-2447