Ocean Optics DT-Mini-2-GS User Manual

Mini Deuterium Halogen Light Source
DT-Mini-2-GS
Installation and Operation Manual
Document Number 000-10000-110-02-1209
Offices:
E-mail: Info@OceanOptics.com (General sales inquiries)
Ocean Optics, Inc. World Headquarters
830 Douglas Ave., Dunedin, FL, USA 34698
Fax 727.733.3962
8 a.m.– 8 p.m. (Mon-Thu), 8 a.m.– 6 p.m. (Fri) EST
Ocean Optics Mikropack
Maybachstraße 11, D73760, Ostfildern, Germ
Phone
Fax
Info@Mikropack.de (Mikropack sales inquiries) Orders@OceanOptics.com (Questions about orders) TechSupport@OceanOptics.com (Technical support)
WARNING
+49 (0)711 34 16 96-0
+49 (0)711 34 16 96-85
Protective Eye Wear Must Be Worn When Using This Instrument ­Intense Ultraviolet Radiation Present See Important Safety Notices inside.
any
000-00000-000-02-A
Copyright © 2009 Ocean Optics, Inc.
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, by any means, electronic, mechanical, photocopying, recording, or otherwise, without written permission from Ocean Optics, Inc. This manual is sold as part of an order and subject to the condition that it shall not, by way of trade or otherwise, be lent, re-sold, hired out or otherwise circulated without the prior consent of Ocean Optics, Inc. in any form of binding or cover other than that in which it is published.
Trademarks
Microsoft, Windows, Windows 95, Windows 98, Windows Me, Windows NT, Windows 2000, Windows XP and Excel are either registered trademarks or trademarks of Microsoft Corporation.
Limit of Liability
Every effort has been made to make this manual as complete and as accurate as possible, but no warranty or fitness is implied. The information provided is on an “as is” basis. Ocean Optics, Inc. shall have neither liability nor responsibility to any person or entity with respect to any loss or damages arising from the information contained in this manual.
1. Do not remove or modify any installed safety device on this equipment. Doing so will void
2. Dangerous voltages are present in this device. There are NO user serviceable parts inside.
3. Only allow qualified personnel to service this unit.
4. Do not use the unit if it is damaged in any way. Contact your dealer for repair or replacement
5. Always screw in the fiber optic cables before starting the instrument.
WARNING
Important Safety Notices
your warranty and create an unsafe operating environment.
information.
Protective eyewear must be worn when using this equipment - Intense ultraviolet radiation present. Never look directly into the light beam, as this can cause eye damage.
000-10000-110-02-1209 A
Important Safety Notices
B 000-10000-110-02-1209
Table of Contents
About This Manual .............................................................................................................iii
Document Purpose and Intended Audience..............................................................................iii
What’s New in this Document ................................................................................................... iii
Document Summary..................................................................................................................iii
Product-Related Documentation ............................................................................................... iii
Upgrades ............................................................................................................................iii
Chapter 1: Setup ...................................................................................1
Overview.............................................................................................................................1
Unpacking the DT-Mini-2-GS .................................................................................................... 2
Contents ....................................................................................................................................2
Chapter 2: DT-Mini-2-GS Specifications ............................................3
Operating Environment ......................................................................................................3
DT-Mini-2-GS Components................................................................................................3
Front Panel ................................................................................................................................ 4
Rear Panel................................................................................................................................. 5
Appendix A: Bulb Replacement ..........................................................7
Replacing the Bulb .............................................................................................................7
Index ......................................................................................................9
000-10000-110-02-1209 i
Table of Contents
ii 000-10000-110-02-1209
About This Manual
Document Purpose and Intended Audience
This document provides you with an installation section to get your system up and running.
What’s New in this Document
This version of the Mini Deuterium Halogen Light Source DT-Mini-2-GS Installation and Operation Manual
Document Summary
Chapter Description
updates the logo and the contact information.
Chapter 1: Setup
Chapter 2: DT-Mini-2-GS Specifications
Appendix A: Bulb Replacement
Contains a list of package contents and unpacking instructions.
Contains operating environment specifications, as well as other physical details of the product.
Provides instructions for changing the bulb.
Product-Related Documentation
You can access documentation for Ocean Optics products by visiting our website at
http://www.oceanoptics.com. Select Technical → Operating Instructions, then choose the appropriate
document from the available drop-down lists. Or, use the of the web page.
You can also access operating instructions for Ocean Optics products on the
Resources
Engineering-level documentation is located on our website at Technical → Engineering Docs.
CD included with the system.
Search by Model Number field at the bottom
Software and Technical
Upgrades
Occasionally, you may find that you need Ocean Optics to make a change or an upgrade to your system. To facilitate these changes, you must first contact Customer Support and obtain a Return Merchandise
000-10000-110-02-1209 iii
About This Manual
Authorization (RMA) number. Please contact an Ocean Optics Application Scientist for specific instructions when returning a product.
iv 000-10000-110-02-1209
Chapter 1
Setup
Overview
The following sections provide instructions on unpacking and setting up your Mini Deuterium Halogen Light Source (DT-Mini-2-GS).
Before using the DT-Mini-2-GS for the first time, check for transport damage. Be sure to adhere to all warnings on the unit and in this manual.
000-10000-110-02-1209 1
1: Setup
Unpacking the DT-Mini-2-GS
Procedure
1. Unpack your lamp assembly carefully. Although the lamp is rigidly mounted, dropping this instrument can cause permanent damage.
2. Inspect the outside of the instrument and make sure that there is no damage. Do not use the instrument if damage is present.
3. Use this instrument in a clean laboratory environment.
Contents
Your DT-Mini-2-GS package should contain the following:
DT-Mini-2-GS Mini Deuterium Halogen Light Source One IC-DB15-2 interface cable for shutter operation
2 000-10000-110-02-1209
Chapter 2
DT-Mini-2-GS Specifications
This section provides information on the operating environment and physical controls of the DT-Mini-2­GS.
Operating Environment
The following table provides information on optimizing the operating environment of your DT-MINI-2­GS.
Operating Environment The DT-Mini-2-GS Unit . . .
Moisture Is designed for operation in dry rooms only.
Ventilation Should be situated so that its location or position does not interfere
with proper ventilation.
Heat Should be situated away from any device that emits excessive heat.
Object and Liquid Entry Should be positioned so that objects do not fall on top of the unit.
Additionally, ensure that no liquids are spilled into the enclosure through openings.
Power Sources Should be connected to a power supply only of the type described in
the operating instructions or as marked on the unit.
DT-Mini-2-GS Components
The following sections describe the components located on the front and rear of the DT-Mini-2-GS.
000-10000-110-02-1209 3
2: DT-Mini-2-GS Specifications
Front Panel
Deuterium ON/OFF
Halogen ON/OFF
SMA Connector
Shutter ON/OFF/TTL
Power LED
Front View
SMA Connector
The SMA connector has a protection cap in place to prevent users from unintentionally looking directly into the light source. To avoid this occurrence, connect the fiber optic cable to the SMA connector before starting the DT-Mini-2-GS lamp. Use caution when handling other end of the fiber optic cable to ensure that light radiation does not come into contact with skin and/or eyes.
WARNING
Protective eyewear must be worn when using this equipment – Intense ultraviolet radiation present.
Never look directly into the light beam, as this can cause eye damage.
4 000-10000-110-02-1209
Rear Panel
2: DT-Mini-2-GS Specifications
Rear View
Power Input Connector
When a 12VDC connection is supplied to the DT-Mini-2-GS, the lamp automatically switches to the basic operating mode. The LED indicator on the front panel will light to reflect this condition.
Lamp ON/OFF/TTL Switch
When this switch is turned ON (when the lamp has power), you can operate the lamp by using the switches on the front panel.
Switch Position Function
ON The deuterium and halogen bulbs are illuminated.
OFF The deuterium and halogen bulbs are not illuminated.
TTL The lamp is powered up and can be controlled by the
SUB-D connector:
Pin 1 = Deuterium ON/OFF
Pin 5 = Halogen ON/OFF
Pin 10 = Ground
Pin 13 = Shutter Open/Close
000-10000-110-02-1209 5
2: DT-Mini-2-GS Specifications
6 000-10000-110-02-1209
Appendix A
Bulb Replacement
To order replacement bulbs for the DT-MINI-2-GS, order item number: DT-Mini-2-B.
Replacing the Bulb
Use the following procedure to replace the bulb.
Procedure
1. Remove the four screws on the bottom of the housing of the DT-Mini-2-GS and remove the cover.
2. Disconnect the four cables attached to the deuterium and halogen bulb.
000-10000-110-02-1209 7
A: Bulb Replacement
Gently
3. Loosen the screw located on the bulb housing.
4. Remove the bulb from the DT-Mini-2-GS by gently pulling the bulb backwards out of the internal bulb socket.
Bulb Housing Screw
Pull
Caution
Do not turn or wiggle the bulb during removal, as this can damage the bulb socket.
5. Reverse the steps above to insert the new Mini-D2-B into the bulb housing.
8 000-10000-110-02-1209
Index
B
bulb replacement, 7
C
components, 3
front panel, 4 rear panel, 5
D
document
audience, iii purpose, iii summary, iii
L
Lamp ON/OFF/TTL Switch, 5
O
operating environment, 3
P
package contents, 2 Power Input Connector, 5 product-related documentation, iii
S
setup, 1 SMA Connector, 4 specifications, 3
U
unpacking procedure, 2 upgrades, iii
W
what's new, iii
000-10000-110-02-1209 9
Index
10 000-10000-110-02-1209
Loading...