All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, by any means, electronic, mechanical,
photocopying, recording, or otherwise, without written permission from Ocean Optics, Inc.
This manual is sold as part of an order and subject to the condition that it shall not, by way of trade or otherwise, be lent, re-sold, hired out or otherwise
circulated without the prior consent of Ocean Optics, Inc. in any form of binding or cover other than that in which it is published.
Trademarks
Microsoft, Windows, Windows 95, Windows 98, Windows Me, Windows NT, Windows 2000, Windows XP and Excel are either registered trademarks or
trademarks of Microsoft Corporation.
Limit of Liability
Every effort has been made to make this manual as complete and as accurate as possible, but no warranty or fitness is implied. The information provided is
on an “as is” basis. Ocean Optics, Inc. shall have neither liability nor responsibility to any person or entity with respect to any loss or damages arising from
the information contained in this manual.
Page 3
Important Safety Notices
1. Do not remove or modify any installed safety device on this equipment. Doing so will void your
warranty and create an unsafe operating environment.
2. Dangerous voltages are present in this device. There are NO user serviceable parts inside.
3. Only allow qualified personnel to service this unit.
4. Do not use the unit if it is damaged in any way. Contact your dealer for repair or replacement
information.
5. Always screw in the fiber optic cables before starting the instrument.
Protective eyewear must be worn when using this equipment - Intense
ultraviolet radiation present.
WARNING
Never look directly into the light beam, as this can cause eye damage.
Warranty
Mikropack GmbH warrants to the original user of this instrument that it shall be free of any defects
resulting from faulty manufacture of this instrument for a period of 12 months from the original data of
shipment. There are no warranties for the DH-2000-CAL-B (standard bulb) or DH-2000-CAL-B-DUV
(deep ultraviolet bulb).
This instrument should not be used for any Clinical or Diagnostic purposes. Data generated in these
areas is not warranted in any way by Mikropack GmbH. Any defects covered by this Warranty shall be
corrected either by repair or by replacement, as determined by Mikropack GmbH.
There are no warranties that extend beyond the description herein.
This Warranty is in lieu of, and excludes, any and all other warranties or representations expressed,
implied, or statutory, including merchantability and fitness, as well as any and all other obligations or
liabilities of Mikropack GmbH including, but not limited to, special or consequential damages. No
person, firm, or corporation is authorized to assume for Mikropack GmbH. Any additional obligation or
liability not expressed provided for herein except in writing duly executed by an officer of Mikropack
GmbH:
Replacing the Deuterium Bulb...................................................................................................18
Replacing the Halogen Bulb ...................................................................................................... 19
Index ......................................................................................................21
ii000-10000-030-02-1209
Page 7
About This Manual
Document Purpose and Intended Audience
This document provides you with an installation section to get your system up and running.
What’s New in this Document
This version of the Deuterium Tungsten-Halogen Calibration Light Source DH-2000-CAL Installation
and Operation Manual
Document Summary
ChapterDescription
updates the logo and contact information.
Chapter 1: Setup
Chapter 2: DH-2000-CAL Specifications
Chapter 3: Operating Instructions
Chapter 4: Troubleshooting
Appendix A: Maintenance
Contains a list of package contents and unpacking
instructions.
Contains operating environment specifications, as
well as other physical details of the product.
Provides instructions for operating the DH-2000-CAL
Light Source.
Contains troubleshooting information for the power
supply and both the deuterium and halogen lamps.
Provides instructions for changing the bulb.
Product-Related Documentation
You can access documentation for Ocean Optics products by visiting our website at
http://www.oceanoptics.com. Select Technical → Operating Instructions, then choose the appropriate
document from the available drop-down lists. Or, use the
of the web page.
You can also access operating instructions for Ocean Optics products on the
Resources
Engineering-level documentation is located on our website at
CD included with the system.
Search by Model Number field at the bottom
Software and Technical
Technical → Engineering Docs.
000-10000-030-02-1209iii
Page 8
About This Manual
Upgrades
Occasionally, you may find that you need Ocean Optics to make a change or an upgrade to your system.
To facilitate these changes, you must first contact Customer Support and obtain a Return Merchandise
Authorization (RMA) number. Please contact an Ocean Optics Application Scientist for specific
instructions when returning a product.
iv000-10000-030-02-1209
Page 9
Chapter 1
Setup
Overview
The following sections provide instructions on unpacking and setting up your DH-2000-CAL Calibration
Light Source.
Before using the DH-2000-CAL for the first time check for transport damage. Be sure to adhere to all
warnings on the unit and in this manual.
000-10000-030-02-12091
Page 10
1: Setup
Unpacking the DH-2000-CAL
► Procedure
1. Unpack your lamp assembly and power supply carefully. Although the lamp is rigidly mounted,
dropping this instrument can cause permanent damage.
2. Inspect the outside of the instrument and make sure that there is no damage. Do not use the
instrument if damage is present.
3. Use this instrument in a clean laboratory environment (see
Contents
Your DH-2000-CAL package should contain the following:
DH-2000-CAL unit
Power cord
UV safety goggles
CC-3 Cosine Corrector
Barrel adapter for SMA 905 Connectors (fibers)
Calibration diskette
Four lamp calibration files:
Deuterium lamp with bare fiber
Deuterium lamp with CC-3 Cosine Corrector
Tungsten Halogen lamp with bare fiber
Tungsten Halogen lamp with CC-3 Cosine Corrector
Four calibration reports (one for each calibration file type)
Operating Environment).
Connecting the Fiber Optic Cable
Use the following procedure to connect the cable to the lamp. See the Sample Spectral Output section for
specific information on configuring your optical fiber or cosine corrector.
2000-10000-030-02-1209
Page 11
► Procedure
To connect the fiber optic cable to the DH-2000-CAL,
1: Setup
1. Locate the cap on the front of the DH-2000-CAL.
2. Lift the cap on the front of the DH-2000-CAL to expose the optical input.
3. Insert the SMA connector barrel or CC-3-UV
fully into the optical input.
4. Secure with the setscrew on the optical input.
000-10000-030-02-12093
Page 12
1: Setup
4000-10000-030-02-1209
Page 13
Chapter 2
DH-2000-CAL Specifications
This section provides information on the operating environment, physical controls, and dimensions of the
DH-2000-CAL. It also provides a graph of spectral lines.
Operating Environment
The following table provides information on optimizing the operating environment of your DH-2000CAL.
Operating EnvironmentThe DH-2000-CAL Unit . . .
MoistureIs designed for operation in dry rooms only.
VentilationShould be situated so that its location or position does not interfere
with proper ventilation.
HeatShould be situated away from any device that emits excessive heat.
Object and Liquid EntryShould be positioned so that objects do not fall on top of the unit.
Additionally, ensure that no liquids are spilled into the enclosure
through openings.
Power SourcesShould be connected to a power supply with the following
specifications:
Units manufactured since April 2003 are equipped with power
supplies that can handle voltage input of 90 to 240 VAC.
Units manufactured before April 2003 are equipped with
power supplies that can handle either 110 VAC or 240 VAC.
These units have serial numbers formatted as 02000XXX.
The power type should be listed on a sticker on the rear of the light
source.
DH-2000-CAL Components
The following sections describe the components located on the front and rear of the DH-2000-CAL unit.
000-10000-030-02-12095
Page 14
2: DH-2000-CAL Specifications
Connector
Front Panel
Tungsten-Halogen
On/Off
Mechanical Protection –SMA
Deuterium On/Off
Power LED
ComponentDescription
Press to turn the Deuterium lamp on or off. There is a delay of 20 seconds before
Deuterium On/Off
Power LEDIndicates the power state of the DH-2000-CAL.
Mechanical Protection –
Optical Output
Halogen On/Off
the Deuterium lamp is illuminated. You must allow the bulb to warm up for 20
minutes to receive accurate data from the lamp.
LED lights green upon successful illumination, or red to indicate lamp malfunction.
Covered to protect users from unintentionally looking directly at the beam of light.
Connect the fiber cable to the DH-2000-CAL
unnecessary exposure to UV radiation.
Always wear proper eye protection when using the DH-2000-CAL lamp.
Press to turn the Halogen lamp on or off. Requires a warm-up time of 20 seconds
before the Halogen lamp is illuminated. You must allow the bulb to warm up for 20
minutes to receive accurate data from the lamp.
LED lights green upon successful illumination, or red to indicate lamp malfunction.
BEFORE turning the lamp on to avoid
6000-10000-030-02-1209
Page 15
Rear Panel
Power Terminal Input
Fuse
Type Sign
Cooling Fan
Main Power Switch
2: DH-2000-CAL Specifications
ComponentDescription
Connect power cable to provide voltage to DH-2000-CAL:
Power Terminal Input
Main Power Switch
Fuse
Cooling FanCools the interior of the DH-2000-CAL. Do not obstruct.
Type Sign
Note: Only connect the power cable to the lamp when the Main Power Switch is in
the OFF position.
Turn on to supply power to the DH-2000-CAL. The Power LED illuminates when
this switch is in the On position.
Contains the fuse to protect the unit against overload:
European Fuse Type: Miniature fuse 5 x 20 mm, 1 Amp slow blow
USA Fuse Type: Miniature fuse 5 x 20 mm, 2.5 Amp slow blow
The following charts illustrate the spectral output of the DH-2000-CAL.
Deuterium Output
2: DH-2000-CAL Specifications
Deuterium bulb spectral output using a bare SMA-terminated fiber
Deuterium bulb spectral output using a CC-3 Cosine Corrector
000-10000-030-02-12099
Page 18
2: DH-2000-CAL Specifications
Tungsten-Halogen Output
Tungsten-Halogen bulb spectral output using a bare SMA-terminated fiber
Tungsten-Halogen bulb spectral output using a CC-3 Cosine Corrector
10000-10000-030-02-1209
Page 19
Chapter 3
Operating Instructions
Operating the Lamp
The following sections provide instructions on operating the Deuterium and Tungsten-Halogen lamps in
the DH-2000-CAL.
Starting the Lamp
Press the Deuterium or Halogen On/Off switch down to preheat the desired lamp. The bulb requires a 20
second preheating period. You must allow this warm up period in order to receive accurate data. After the
warm up period, the lamp will illuminate.
After successful illumination, the two-color LED beneath the lamp’s On/Off switch lights up green to
indicate that the lamp is on. Should the lamp fail to light, see
Protective eyewear must be worn when using this equipment - Intense ultraviolet radiation
present.
WARNING
Never look directly into the light beam, as this can cause eye damage.
Troubleshooting for more information.
Turning the Lamp Off
Turn the lamp off by pressing the appropriate On/Off switch.
Warming Up the Lamp
The Deuterium and Tungsten-Halogen lamps require 20 minutes of operation to reach a state of thermal
equilibrium. During this warm-up period, the intensity of the UV output power can vary substantially.
If applications require extreme intensity stability, the Halogen lamp should be warmed up for an
additional 15 minutes. Once warmed up for this amount of time, the lamp will reach specified drift values.
000-10000-030-02-120911
Page 20
3: Operating Instructions
Configuring Calibration with the DH-2000-CAL
The following sections provide instructions on configuring calibration with the DH-2000-CAL.
Configuring the DH-2000-CAL
► Procedure
Follow the steps below to configure the DH-2000-CAL:
1. If using a bare fiber with the DH-2000-CAL:
a.Screw the fiber end onto the SMA barrel adapter until secure.
b.Insert the SMA barrel adapter completely into the optical input on the DH-2000-CAL.
If using a CC-3 cosine corrector with the DH-2000-CAL:
a.Screw the CC-3 cosine corrector onto one end of the fiber until the connection is tight.
b.Insert the CC-3 cosine corrector completely into the input on the DH-2000-CAL.
2. Use the Allen wrench to tighten the setscrew on the optical input of the DH-2000-CAL.
3. Screw the other end of the fiber into the SMA 905 Connector of the spectrometer.
4. Plug the black power cord into the back of the DH-2000-CAL.
5. Plug the other end of the power cord into a standard outlet.
6. Turn the lamp power on using the black switch on the rear of the DH-2000-CAL.
7. Switch the individual lamps on using the buttons on the front of the DH-2000-CAL.
8. Allow the lamp to warm up for at least 20 minutes before proceeding.
12000-10000-030-02-1209
Page 21
3: Operating Instructions
Calibrating a Spectrometer with the DH-2000-CAL
► Procedure
Follow the instructions below to calibrate a spectrometer using the DH-2000-CAL:
1. Copy all calibration files supplied with the lamp to the application (software you will be using)
folder.
2. Follow the instructions in your software for UV range calibration, using the appropriate
Deuterium lamp file (bare fiber or CC-3 Cosine Corrector).
3. Follow the instructions in your software for VIS range calibration, using the appropriate Tungsten
Halogen lamp file (bare fiber or CC-3 Cosine Corrector).
Note
You can use either UV or VIS range calibration only. Note that if the spectrometer range
exceeds the calibration range, the data obtained outside the calibration range is inaccurate
and should be discarded. Deuterium calibration is not appropriate for the VIS range, and
Tungsten Halogen calibration is not appropriate for the UV range.
If a full range calibration is required, combine the two resulting calibration files (UV and
VIS) at 400 nm either manually or in your software.
000-10000-030-02-120913
Page 22
3: Operating Instructions
14000-10000-030-02-1209
Page 23
Chapter 4
Troubleshooting
Deuterium Lamp
If the power supply or Deuterium lamp does not seem to functioning properly, check the following:
IssueProbable CauseResolution
Power switches on, but no
LEDs light.
Deuterium lamp does not
ignite.
The two-color LED under the
Deuterium On/Off switch lights
up red, indicating an error.
Deuterium lamp turns off
during operation.
Line power not presentCheck line voltage
Fuse defectiveCheck fuse
Allow the Deuterium lamp to cool
down (20 minutes).
Deuterium lamp too hot
Deuterium lamp life exhaustedReplace Deuterium lamp
Deuterium lamp’s internal
connection plug is not closed
right
Deuterium lamp too hot
Press On/Off switch again to
reset the Deuterium lamp, then
press again to restart.
Open unit (see the
and close connector plug.
Turn off the unit.
Allow the unit to cool down for at
least 20 minutes.
Once the unit has cooled down,
turn the Deuterium lamp back on.
Maintenance)
000-10000-030-02-120915
Page 24
4: Troubleshooting
Halogen Lamp
If the Halogen lamp does not seem to functioning properly, check the following:
IssueProbable CauseResolution
Halogen lamp does not work
after pressing On/Off switch
LED does not light after
switching on the Halogen lamp
Deuterium lamp is warming up
Halogen lamp is defectiveReplace the Halogen lamp
Internal power supply is
defective
Wait until the Deuterium lamp
has lit and try again
Disconnect the unit from the
main power source and contact
your dealer for repair or
replacement
16000-10000-030-02-1209
Page 25
Appendix A
Maintenance
Overview
Maintenance of your DH-2000-CAL unit involves periodic recalibration and replacing the light source
bulbs, when necessary.
Recalibration
You should have the DH-2000-CAL recalibrated after every 50 hours of use. Contact Ocean Optics or
your DH-2000-CAL vendor for recalibration information. The recalibration order code is DH-2000RECAL.
Bulb Replacement
You can manually change the Deuterium and Halogen bulbs in the DH-2000-CAL. However, as a
recalibration is required when the bulb in the DH-2000-CAL is replaced, we recommend sending the DH2000-CAL in for bulb replacement and calibration service after 50 hours of use.
Contact an Ocean Optics Application Sales Engineer for information on bulb replacement service and
lamp recalibration.
To order replacement bulbs for the DH-2000-CAL, order the following item number(s):
Deuterium Spare Bulb (220 – 400 nm): D-2000-B
Halogen Spare Bulb (300 – 1050 nm): DH-2000-B
000-10000-030-02-120917
Page 26
A: Maintenance
Replacing the Deuterium Bulb
Note
You MUST recalibrate the lamp after replacing the bulb. See Recalibration for more
information.
► Procedure
1. Open the six slotted screws (14) and open the casing cover.
2. Open the screws (15) with the tool (18) that is delivered with the spare bulb (16).
3. Disconnect the old bulb and connect the new Deuterium lamp only with the originally supplied
connection plugs (17).
18000-10000-030-02-1209
Page 27
Replacing the Halogen Bulb
Note
You MUST recalibrate the lamp after replacing the bulb. See Recalibration for more
information.
1. Open the six slotted screws (14) and open the casing cover.
A: Maintenance
2. Remove the screw (21) with the tool (20) provided with the spare bulb (19).
3. Disconnect the old Halogen bulb from the connection plugs (18)
4. Open the screws of the cable-clamp on the lamp-side and remove the defective Halogen lamp
module.
5. Insert the new Halogen lamp module and replace the screw (21).
6. Reattach the two cables of the Halogen lamp module to the cable-clamp.