Ocean Optics Collimation tester User Manual

Collimation Tester Instructions

Description

Use shear-plate collimation testers to examine and adjust the collimation of laser light, or to measure the wavefront curvature and divergence/convergence magnitude of large-radius optical components.
The setting up of a laser beam expander to give a collimated or parallel beam can be a problem if the full capability of any precision laser optics is to be achieved. Methods using autocollimation or measuring beam diameter over some distance are inadequate to obtain diffraction-limited performance unless tedious measurements are made. To accomplish this task, Ocean Optics has an interferometer that is extremely simple to use, requires almost no adjustment and produces results in a few moments.

Method of Use

The method of use is so simple it maybe hard to realize that the device is a precision interferometer.
Procedure
1. Set up the beam expander with the lenses spaced according to their focal lengths.
2. Place the collimation tester in the beam so as to illuminate the front face as fully as possible,
and to reflect the beam by about 90° to a screen placed at right angles to the reflected beam (see figure below).
CT-10000-000-05-201401 1
Collimation Tester Instructions
3. The maximum sensitivity of the collimation tester is achieved when the reflected angle is
108°, but sensitivity is reduced by only 1% at 90°, so the angle is not critical. Two overlapping images will be seen on the screen, one from each face of the collimation tester. Fringes (parallel patterns of light and dark) will be seen in the overlap region.
4. Adjust the lens spacing until the fringes are parallel to the shadow of the cursor wire of the
tester. The beam is now collimated.
Tips
If the fringes are not straight, check to see that the lenses of the beam expander are centered, square to the beam and with the proper face towards the collimated light (if these adjustments are appropriate). Any residual fringe curvature or wiggles indicates aberrations or errors that cannot be focused out.
If no fringes are seen in the initial setup, it may be that the beam is so far from collimation that the fringes are too fine to be seen. In that case, redu ce the ref le cted beam angle to 45°or even 30°, and move the screen accordingly. This reduces the sensitivity of the tester. Make the initial adjustment for collimation and then return to 90° for the final tuning.

USEFUL RANGE, APERTURE, WAVELENGTH

Each size of collimation tester is designed to have 5 – 6 horizontal fringes across the aperture when illuminated by a parallel beam. In theory, that can be tested is about 1/5 full aperture. In practice, it is found that the fringe slope changes from positive to negative when going through collimation and that the fringe spacing decreases away from collimation so that smaller aperture beams can be collimated by finding the center position between two defocused positions. When testing smaller apertures, it may also be necessary to reduce the reflected beam angle to get an appropriate amount of overlap of the two beams. The fringes only appear in the overlap area.
The choice of 5 – 6 fringes across the aperture is a trade-off between sensitivity and utility. For sensitivity, it is desirable to have only one full fringe over the aperture. The actual design allows an individual tester to be usable over a range of beam sizes of at least 5 to 1. If a collimation tester is to be used at a specific wavelength for a single aperture with critical demands on collimation, you can either select a larger aperture such that one fringe will be seen across your aperture or a custom design can be made for optimum sensitivity. Collimation testers are usable in wavelength over the transmission range of the glass, that is, from about 350 nm to about 2500 nm. To observe the fringes in the UV, a fluorescent screen may be necessary while in the IR, an image converter is suitable out to 1200 nm and an IR phosphor card can be used further out.
to have at least one fringe to set against, the smallest beam
2 CT-10000-000-05-201401
Loading...
+ 4 hidden pages