Oceanic Pro Plus 2 User Manual

OCEANIC
Pro Plus 2
dive computer
operating manual
®
LIMITED TWO-YEAR WARRANTY
For details, refer to the Product Warranty Registration Card provided.
This operating manual is copyrighted, all rights are reserved. It may not, in whole or in part, be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine readable form without prior consent in writing from Oceanic / 2002 Design.
Pro Plus 2 Operating Manual, Doc. No. 12-2328
TRADEMARK NOTICE
Oceanic, the Oceanic logo, Pro Plus 2, the Pro Plus 2 logo, Smart Glo, OceanGlo, Graphic Diver Interface, Tissue Loading Bar Graph, Pre Dive Planning Sequence, Variable Ascent Rate Indicator, Set Point, Control Console, Turn Gas Alarm, and OceanLog are all registered and unregistered trademarks of Oceanic. All rights are reserved.
PATENT NOTICE
U.S. Patents have been issued, or applied for, to protect the following design features: Dive Time Remaining (U.S. Patent no. 4,586,136), Data Sensing and Processing Device (U.S. Patent no. 4,882,678), Air Time Remaining (U.S. Patent no. 4,586,136), and Variable Ascent Rate Indicator (U.S. Patent no. 5,156,055). User Setable Display (U.S. Patent no. 5,845,235) is owned by Suunto Oy (Finland).
DECOMPRESSION MODEL
The programs within the Pro Plus 2 simulate the absorption of nitrogen into the body by using a mathematical model. This model is merely a way to apply a limited set of data to a large range of experiences. The Pro Plus 2 dive computer model is based upon the latest research and experiments in decompression theory. Still, using the Pro Plus 2, just
as using the U.S. Navy (or other) No Decompression Tables, is no guarantee of avoiding decom­pression sickness, i.e. “the bends.” Every diver’s physiology is different, and can even vary from day to day.
No machine can predict how your body will react to a particular dive profile.
© 2002 Design 2002
San Leandro, Ca. USA 94577
2
CONTENTS
LIMITED TWO-YEAR WARRANTY ................................................................................................................................2
NOTICES ............................................................................................................................... ....................................... 2
DECOMPRESSION MODEL ......................................................................................................................................... 2
FEATURES AND DISPLAYS.............................................................................................. 7
INTERACTIVE CONTROL CONSOLE .......................................................................................................................... 8
BAR GRAPHS .............................................................................................................................................................. 9
Tissue Loading Bar Graph (TLBG) ............................................................................................................................. 9
Oxygen Bar Graph (O2BG) ...................................................................................................................................... 10
Variable Ascent Rate Indicator (VARI) ....................................................................................................................... 10
Air Time Remaining Bar Graph (ATRBG) .................................................................................................................. 11
DIVE TIME REMAINING .............................................................................................................................................. 11
ALPHA / NUMERIC DISPLAYS ................................................................................................................................... 12
Cylinder Pressure ................................................................................................................................................... 12
Depth Displays ....................................................................................................................................................... 12
Time and Date Displays .......................................................................................................................................... 13
Temperature Display ............................................................................................................................................... 13
AUDIBLE ALARM ....................................................................................................................................................... 14
LED Warning Light .................................................................................................................................................. 14
SMARTGLO BACKLIGHT ........................................................................................................................................... 16
POWER SUPPLY ........................................................................................................................................................ 16
Battery Consumption Indicator .................................................................................................................................. 16
Low Battery Condition ............................................................................................................................................. 17
FO2 MODE .................................................................................................................................................................. 18
FO2 50% Default .................................................................................................................................................... 19
3
CONTENTS (continued)
ACTIVATION AND SETUP............................................................................................... 21
ACTIVATION ............................................................................................................................................................... 22
SURFACE MODE ........................................................................................................................................................ 23
Accessing Other Modes .......................................................................................................................................... 23
SET MODES ............................................................................................................................................................... 24
ENTERING SETTINGS -SET MODE #1 ................................................................................................................... 25
ENTERING SETTINGS -SET MODE #2 ................................................................................................................... 29
PLAN AND DIVE MODES ................................................................................................ 45
PRE DIVE PLANNING SEQUENCETM....................................................................................................................... 46
TISSUE LOADING BAR GRAPH ................................................................................................................................. 48
OXYGEN ACCUMULATION BAR GRAPH ................................................................................................................... 49
VARIABLE ASCENT RATE INDICATOR ..................................................................................................................... 49
AIR TIME REMAINING BAR GRAPH BAR GRAPH ..................................................................................................... 50
CONTROL OF DISPLAYS ........................................................................................................................................... 50
NO DECOMPRESSION DIVE MODE .......................................................................................................................... 51
DECOMPRESSION DIVE MODE ................................................................................................................................54
VIOLATION MODES .................................................................................................................................................... 57
Conditional Violation Mode ...................................................................................................................................... 57
Delayed Violation Modes ......................................................................................................................................... 59
Immediate Violation Mode and Gauge Mode ............................................................................................................ 60
HIGH PO2 DIVE MODE ............................................................................................................................................... 62
HIGH OXYGEN ACCUMULATION ............................................................................................................................... 63
USER SET DIGITAL GAUGE MODE ........................................................................................................................... 64
UNEXPECTED LOSS OF DISPLAYED INFORMATION .............................................................................................. 64
4
CONTENTS (continued)
POST DIVE MODES ........................................................................................................ 65
POST DIVE SURFACE MODE .................................................................................................................................... 66
TRANSITION PERIOD ................................................................................................................................................ 66
AFTER THE TRANSITION PERIOD (THE FIRST 2 HOURS)....................................................................................... 68
To activate the Backlight - ........................................................................................................................................ 68
To access the Pre Dive Planning Sequence - ............................................................................................................ 68
To access the Time to Fly Countdown ....................................................................................................................... 69
To access the Time to Desaturate Countdown ........................................................................................................... 69
Log Mode ............................................................................................................................................................... 70
AFTER THE FIRST 2 HOURS ..................................................................................................................................... 72
WET CONTACTS ........................................................................................................................................................ 73
DOWNLOADING DATA TO A PC ............................................................................................................................... .. 74
SIMULATOR (DEMO) MODE ............................................................................................ 75
GENERAL ........................................................................................................................ 83
CARE AND CLEANING .............................................................................................................................................. 84
INSPECTIONS AND SERVICE ................................................................................................................................... 84
BATTERY REPLACEMENT ........................................................................................................................................ 86
SPECIFICATIONS ....................................................................................................................................................... 90
OCEANIC WORLD WIDE ............................................................................................................................................ 96
SERVICE RECORD .................................................................................................................................................... 97
RESET PROCEDURE ................................................................................................................................................ 99
Pay special attention to items marked
with this Warning symbol.
WARNING: Prior to diving with the Pro Plus 2, you must also read and understand the Oceanic Dive Computer Safety and Reference Manual.
5
Components: a. Select (right) Button
m
l
DECO
k
N
j
i
h
DEMO
MAX
g
O
2
O
f
2
10 20 30 40 50 6005
ATR
O
M FEET
MAX
TOO
FAST
2
PSI
BAR
n
o
p
q
r
b. LED Warning Light c. Smartglo Sensing Port d. Advance (front) Button e. Air Time Remaining Bar Graph f. O2 (oxygen) Bar Graph g. Icon - Descend Arrow
Icon - Decompression Ceiling
Icon - Ascend Arrow h. Graphic - Maximum Depth i. Tissue Loading Bar Graph j. Battery Indicator k. Icon - Alarm (Set Mode) l. Icon - Log Mode m. Graphic - Demo Mode n. Graphic - Depth o Graphic - Ascent Too Fast p. Icon - Time q. Variable Ascent Rate Indicator r. Icon - Operating Mode
e
d
c
a
b
PRO PLUS 2
6
WARNING: Prior to diving with the Pro Plus 2, you must also read and understand the Oceanic Dive Computer Safety and Reference Manual which provides Important Warnings and Safety Recom­mendations as well as general product information.
FEATURES and DISPLAYS
7
INTRODUCTION
Welcome to Oceanic and thank you for choosing the Pro Plus 2 !
It is extremely important that you read this Operating Manual in sequence and under­stand it completely before attempting to use the Pro Plus 2.
It is equally important that you read the Oceanic Dive Computer Safety and Reference Manual (Doc. No. 12-2262) provided with your Pro Plus 2. It contains information that you must become familiar with prior to diving with your Pro Plus 2.
Remember that technology is no substitute for common sense, and a dive computer only provides the person using it with data, not the knowledge to use it.
DECO
N
O
2
O
2
ATR
M FEET MAX
TOO FAST
DEMO
MAX
O
2
PSI
BAR
10 20 30 40 50 6005
Fig. 1 - Interactive Control Console
8
INTERACTIVE CONTROL CONSOLE
The Interactive Control Console consists of two Control Buttons that allow you to select display options and access specific information when you want to see it. They are also used to enter Settings, activate the Backlight, and acknowledge the Audible Alarm.
The Left button is referred to as Advance (Fig. 1a) and the Right button as Select (Fig. 1b).
ba
DECO
ATR
O
2
N
2
BAR GRAPHS
Tissue Loading Bar Graph (TLBG)
The Tissue Loading Bar Graph (Fig. 2a) represents tissue loading of nitrogen, showing your relative no decompression or decompression status. As your depth and elapsed dive time increase, segments will add to the Graph, and as you ascend to shallower depths, the Bar Graph will begin to recede, indicating that additional no decompression time is allowed for multilevel diving.
The Tissue Loading Bar Graph monitors 12 different nitrogen compartments simultaneously and displays the one that is in control of your dive. It is divided into a green No Decom­pression (normal) zone, a yellow Caution zone (also No Decompression), and a red Decompression (danger) zone.
While you cannot provide a guarantee against the occur­rence of decompression sickness, you may choose your own personal zone of caution based upon age, physique, exces­sive weight, etc., to reduce the statistical risk.
NOTE: Displays associated with oxygen and the O2 Bar Graph will only appear if FO2 has been set at a value other than 'Air' (e.g., a numerical value).
a
Fig. 2 - TLBG
9
Oxygen Accumulation Bar Graph (O2BG)
The O2 Bar Graph (Fig. 3a) represents oxygen loading, showing the maximum of either per dive accumulated oxygen, or 24 hour period accumulated oxygen.
As your oxygen exposure (accumulation) increases during the dive, segments will add to the Bar Graph, and as loading decreases, it will begin to recede, indicating that addi­tional exposure is allowed for that dive and 24 hour period.
Variable Ascent Rate Indicator (VARI)
The Variable Ascent Rate Indicator (Fig. 3b) provides a visual representation of ascent speed (i.e., an ascent speedometer). Green is a 'normal' rate, yellow a 'caution' rate, and red is 'Too Fast'. The segments of the VARI represent two sets of speeds which change at a reference depth of 60 feet (18 meters). Refer to the chart (Fig. 4) for seg­ment values.
DECO
N
2
Deeper than 60 feet (18 m)
Segments Ascent Rate =
b
TOO
FAST
a
Displayed FPM MPM 0 0-20 0 - 6 1 21-30 6.5-9 2 31-40 9.5-12 3 41-50 12.5-15 4 51-60 15.5-18 5 >60 >18
Fig. 4 - VARI Values
60 feet (18 m) & Shallower
Segments Ascent Rate = Displayed FPM MPM 0 0-10 0 - 3 1 11-15 3.5-4.5 2 16-20 5-6 3 21-25 6.5-7.5 4 26-30 8-9 5 >30 >9
O
2
O
2
10 20 30 40 50 6005
ATR
Fig. 3 - O2BG and VARI
10
WARNING: At depths greater than 60 feet (18 meters), ascent rates should not exceed 60 fpm (18 mpm). At depths of 60 feet (18 meters) and shallower, ascent rates should not exceed 30 fpm (9 mpm).
PSI
10 2005
ATR
FEET
O
2
N
2
Air Time Remaining Bar Graph (ATRBG)
The Air Time Remaining Bar Graph (Fig. 5a) provides a graphic representation of the time that you can remain at your present depth and then, following a safe ascent, surface with a predetermined reserve of breathing gas.
This calculation and display is based on your Breathing Gas Consumption Rate that is continuously monitored by the Pro Plus 2, and it takes into account the breathing gas required for a safe ascent including any required decompression stops.
The green, yellow, and red zones adjacent to the bar graph
Fig. 5 - ATRBG
enable you to quickly focus on remaining breathing gas times of 60 minutes and less, based on your pre selected End Pressure Alarm Set Point (described later). The bar graph is more precise as time decreases toward the red zone.
DIVE TIME REMAINING
The Pro Plus 2 constantly monitors three critical pieces of information; no decompression status, oxygen accumulation, and breathing gas consumption rate. Refer to the Oceanic Dive Computer Safety and Reference Manual which describes Dive Time Remaining which is a display that indicates the time that is more critical for you at that particular moment (i.e.; whichever time is the least amount available of the three).
Displayed will be either Time Remaining before reaching the No Decompression limit (Fig. 5b), or Time Remaining before reaching the limit for Oxygen Accumulation, or Time Remaining before reaching the End Pressure Alarm set point.
b
a
11
b
N
2
MAX
c
O
2
10 2005
ATR
a
Fig. 6 - Cylinder Pressure
and Depth Displays
12
FEET
ALPHA / NUMERIC DISPLAYS
Cylinder Pressure Display
Cylinder Pressure (Fig 6a) is displayed any time the Pro Plus 2 is connected to a pressurized cylinder and in an active operating mode.
Values of pressure are displayed numerically from 10 PSI (.5 BAR) up to 5000 PSI (352 BAR) in increments of 10 PSI (.5 BAR).
Depth Displays
During a dive, the Current Depth display (Fig. 6b), indi­cates depths from 0 to 330 feet (99.9 meters) in 1 foot (.1 meter) increments.
By pressing the Left (Advance) button, the Maximum Depth reached during that dive will be displayed in the center/left portion of the display (Fig. 6c).
During a Decompression Dive, the required Ceiling Stop
PSI
Depth is displayed in the center of the screen. Maximum Depth can be viewed by pressing the Left (Advance) button.
PSI
10 2005
ATR
FEET
O
2
N
2
Time and Date Displays
Time displays are shown in hour:minute format (i.e., 1:16 represents 1 hour and 16 minutes, not 116 minutes!). The colon that separates hours and minutes blinks once per second when the display is indicating real time (e.g., Elapsed Dive Time), and is solid (non-blinking) when times are calculated projections (e.g., Time to Fly).
The Main Time display is located in the lower/middle portion of the display (Fig. 7a) and a second time display (Fig. 7b) is located in the center/right. Both displays are identified by clock icons.
• Time of Day can be set for 12 hour format (Am/Pm) or 24 hour format.
Date is displayed in the center/left portion of the screen only to identify dive data while it is viewed in the Log Mode (see page 70). When Units of Measure are set for 'Imperial', the Month appears to the left of Day. When set for Metric, the Month appears to the right of Day.
Temperature Display
Ambient Temperature is displayed in the center/left portion of the screen (Fig. 7c) while in the Surface Mode and
c
Log Mode, and can be viewed as part of an Alternate Display when the Left (Advance) button is pressed while in a dive mode. If the Temperature exceeds a value of '99', two dashes ( - - ) will be displayed on the screen until the unit's tempera­ture decreases to '99'.
Fig. 7 - Time and Temperature
b
a
Displays
13
NOTE: Each numeric and graphic display represents a unique piece of informa­tion. It is imperative that you understand the formats, ranges, and values of the information represented to avoid any possible misunderstanding that could result in error. The Informational Displays are described in detail as the various operat­ing modes they appear in are presented throughout this manual.
AUDIBLE ALARM
LED Warning Light
A red LED Warning Light will light and a Speaker icon will appear when the Audible Alarm emits a tone. They will turn off when the Alarm is acknowledged, and will not activate if the Audible is Set OFF (a user setting).
When warning situations activate the Alarm, the unit will emit a continuous tone for 10 seconds, or until the situation is corrected, or it is acknowledged by the user pressing the Left (Advance) button for 2 seconds. If acknowledged by the user and the situation corrected, the Alarm will sound again upon reentry into the warning situation, or entry into another type of warning situation.
A single short beep (which cannot be disabled) is emitted for the following -
• After the Diagnostic countdown, if everything is okay.
• If the unit automatically returns to Surface Mode from Simulator Mode.
• Upon completion of a Hot Swap battery change with calculations saved.
• Change from Delayed to Full Violation 5 minutes after the dive.
14
Situations that will sound the Alarm, if it is turned ON (a user setting), include -
• Entry into Decompression Mode.
• Air Time Remaining Bar Graph Alarm (a user setting).
• Air Time Remaining = 5 minutes.
• Air Time Remaining = 0 minutes.
• Turn Pressure Alarm (a user setting).
• End Pressure Alarm (a user setting).
• PO2 => than the Max PO2 Alarm (a user setting), or => 1.60 ATA.
• Descent deeper than the Max Depth Alarm (a user setting).
• Tissue Loading Bar Graph Alarm (a user setting).
• Dive Time Remaining Alarm (a use setting).
• Elapsed Dive Time Alarm (a use setting).
• O2 Accumulation => allowable per dive limit, or limit for a 24 hour period.
• Ascending above a required Decompression ceiling stop depth for less than 5 minutes (referred to as a Conditional Violation).
• Ascent rate exceeds 60 feet/minute (18 meters per minute) if greater than 60 feet (18 meters), or 30 feet/minute (9 meters/minute) at 60 feet (18 meters) and shallower.
During the following situations, the 10 second continuous tone will be followed by a 5 second steady beep that will not turn off when acknowledged, even if it was user Set OFF -
• Ascending above a required Decompression ceiling stop depth for more than 5 minutes (referred to as a Delayed Violation).
• Decompression requires a ceiling stop depth of 70 feet/21 meters or deeper.
• Being on the surface for 5 minutes after a Conditional Violation (Permanent Violation).
15
a
N
2
O
2
ATR
Fig. 8 - Battery Indicator
16
SMARTGLO® BACKLIGHT
To activate the Backlight - press the Right (Select) button.
• The Smart Glo® feature senses the intensity of natural light present. If a low level of light is present, the Back­light will activate and illuminate the display for button depression time plus the user set additional Duration time (3 or 7 seconds).
• Press the button again to activate as desired.
NOTE: Extensive use of the Backlight reduces estimated Battery life. Also, the Backlight does not operate during a Low Battery Condition or when downloading data to a PC.
POWER SUPPLY
The Pro Plus 2 utilizes one (1) 3 volt, CR2 Lithium Battery that should provide from 50 dive hours of operation, if you conduct 1 - one hour dive each time the unit is activated, to over 150 dive hours of operation, if you conduct 3 or more one dives each time the unit is activated.
Battery Indicator
A Battery Indicator (Fig. 8a) provides an indication of Battery condition. The Battery Indicator will be displayed during Surface Mode. It will not be displayed during Dive Mode(s).
PSI
ATR
O
2
N
2
Low Battery Condition
• Voltage level is checked upon activation and every 10 minutes during operation.
• When 75 % of the Rated Power has been consumed, only the lower bar of the Indicator will be displayed and the full icon will flash (Fig. 9) once per second as a warning that the Battery is to be replaced prior to con­ducting any further dives.
• Upon decreasing to a voltage level that will no longer sustain proper operation, the Battery Indicator will flash 5 times followed by shutdown of the unit.
• If a Low Battery Condition exists when the unit is manu­ally activated (by pressing the button), the graphic bAT and the Battery Indicator will appear flashing for 5 seconds followed by shutdown of the unit.
• If the button is not pressed to activate the unit prior to a dive (e.g., the unit activated automatically by immersion in water), and a Low Battery Condition exists, the Battery Indicator will appear flashing as a warning upon descent past 4 feet (1.2 meters). No other information will be displayed and the unit will not enter Dive Mode.
Fig. 9 - Low Battery Condition
17
• If the unit did not display a Low Battery Condition prior to entering the Dive Mode, and a Low Battery Condition occurs during the dive, there will be sufficient Battery power to maintain unit operation for the remainder of that dive. The Battery Indicator will appear after the dive upon entry into Surface Mode.
• When the Battery is removed, settings and calculations for repetitive dives will be retained, if a new battery can be inserted within 8 seconds. Otherwise the calculations will reset to zero and settings must be reset.
FO2 MODE
N
2
O
2
ATR
Fig. 10 - FO2 set for AIR
18
After Activation, the Pro Plus 2 will operate as an Air com­puter without displaying information associated with oxygen calculations, unless it is set for a percentage of oxygen (FO2) other than Air (a numerical value between 21 and 50 %).
When set with an FO2 value of 'Air' (Fig. 10), the Pro Plus 2 will perform calculations the same as if FO2 were set for 21% oxygen, internally accounting for oxygen loading for any subsequent Nitrox dives. However, oxygen related displays, warnings, and the O2 Bar Graph will not appear on the display for that dive, or subsequent dives, unless FO2 is set for a numerical value (21 - 50).
Once a dive is made with the unit set as a Nitrox Computer
MAX
ATR
O
2
N
2
ATR
O
2
N
2
(FO2 set for a numerical value), it cannot be programmed to operate as an Air Computer until 24 hours after the last dive. 'Air' will not be displayed as an option in the FO2 Mode. However, you can set FO2 for 21% for use with Air.
When FO2 is set at a value of 21% (Fig. 11), the unit will remain set at 21% for subsequent nitrox dives until FO2 is set to a higher value, or until it automatically turns off and is reactivated.
FO2 50% Default
If the Default function is set to ON (Fig. 12) and FO2 is set to a value 'greater than 21%', the FO2 set point value will automatically revert to 50% 10 minutes after that dive. The Maximum Depth that can be achieved with a PO2 of
1.60 ATA will also be displayed.
• FO2 must therefore be reset for each repetitive nitrox dive, or the value will automatically 'default' to 50(%) and the dives will be calculated based on 50% O2 (50% nitrogen) for oxygen calculations and 21% O2 (79% nitrogen) for nitrogen calculations.
Fig. 11 - FO2 set for 21%
Fig. 12 - FO2 Default ON
19
N
2
O
2
ATR
Fig. 13 - FO2 Default OFF
20
If the Default function is set to OFF (Fig. 13), the FO2 value for repetitive dives remains the same as previously set until the set point is manually changed.
WARNING: When the unit shuts off, the FO2 set point will default to AIR regardless if the 50% Default is set ON or OFF. Always set, or verify, the FO2 set point prior to each nitrox dive.
Setting FO2 for a nitrox dive is described on page 25.
R
R
E
E
S
V
P
I
O
D
N
E
S
L
I
B
WARNING: Prior to diving with the Pro Plus 2, you must also read and understand the Oceanic Dive Computer Safety and Reference Manual which provides Important Warnings and Safety Recom­mendations as well as general product information.
ACTIVATION and SETUP
21
ACTIVATION
To Activate the Pro Plus 2, press and release the Left (Advance) Button.
• Upon manual activation, the unit will enter Diagnostic Mode (Fig. 14), displaying all segments of the LCD as 8's, followed by dashes (- -), then a countdown from 9 to 0. Diagnostic Mode checks the display and battery voltage to ensure that everything is within tolerance and functioning properly.
Backup Activation (only if Water Activation is set ON)
As a backup, the Pro Plus 2 will also automatically activate by water contact. This is accomplished by bridging the gap between contacts located on the Button stems and case. The graphic H2O that will be displayed as an indication is described later.
DECO
N
2
DEMO
MAX
O
2
O
2
10 20 30 40 50 6005
ATR
Fig. 14 - Diagnostic Mode
22
If no dive is made within 2 hours after initial activation, the unit
M FEET
will automatically deactivate. If the wet contacts are still
MAX
TOO
bridged, the unit will reactivate and display the H2O graphic.
FAST
WARNING: If the unit is manually activated at elevations higher than 14,000 feet (4,267 meters), it will perform a
O
2
PSI
BAR
diagnostic check followed by immediate shutdown.
SURFACE MODE
PSI
ATR
O
2
N
2
Surface Mode (Fig. 15), identified by the Surface Time icon, follows Diagnostic Mode after Activation. Information includes Dive Number '0' (no dive made yet), Temperature (and icon), Time of Day (with icon), the Battery Consumption Indicator, and Surface Time (with flashing colon).
NOTE: If the wet contacts are bridged, the graphic 'H2O' will appear in place of the dive number '0' (Fig.
16). After the unit is rinsed and dried, '0' will replace 'H2O'.
To activate the Backlight:
• press the Right (Select) button.
Accessing Other Modes
While in the Surface Mode you can access various other Modes that are described in detail throughout this manual.
• Press the Left (Advance) button to access a sequence that
includes Plan, Fly, DeSaturation, and Log Modes.
• Press the Right (Select) button to activate the Backlight.
• Press Both buttons (simultaneously) to access Set and
Simulator Modes.
Fig. 15 - Surface Mode
N
2
O
2
ATR
Fig. 16 - Surface Mode
(rinse and dry the unit)
PSI
23
N
2
SET MODES
Settings are divided into 2 categories. Set Mode #1 includes several settings that you would change more often and Set Mode #2 includes those items not likely to change once you set them. Set Mode 2 can be accessed by first entering settings in Set Mode 1, or by bypassing Set Mode 1.
O
2
ATR
Fig. 17 - Set Mode 1
N
2
O
2
ATR
Fig. 18 - Set Mode 2
24
Settings can be made one after the other, or you can access a specific item that you want to set, bypassing others. Set points can be advanced by scrolling (press/hold) or one increment at a time (press/release < 2 seconds)
Set Mode Access Timing
While in Surface Mode, press Both buttons simultaneously and hold -
• after 2 seconds, SET: 1 appears (Fig. 17)
• after 2 more seconds, SET: 2 appears (Fig. 18)
• Access is gained by releasing the buttons during the 2 second window in which SET: 1 or SET: 2 appears. then pressing the Left (Advance) button.
• If the buttons are held longer, and SET 1 and 2 are both bypassed, the unit will go to Simulator (Demo) Mode which is described on page 75.
• While in the Set Mode, if neither button is pressed during a 2 minute period, the unit will revert to Surface Mode.
MAX
ATR
O
2
N
2
ENTERING SETTINGS -SET MODE #1
ATR
O
2
N
2
TO SET - FO2 (while in the Surface Mode) Factory set for AIR, FO2 can also be set to values between 21 and 50% in increments of 1%.
• Press Both buttons simultaneously, release when SET: 1 appears (2 seconds).
• Press and release the Left (Advance) button, FO2 ap­pears with the set point value flashing (Fig. 19).
• Press and release the Right (Select) button repeatedly to increase the FO2 value from 21 to 50% in increments of 1%, then display AIR again; - or - Press and hold the Right (Select) button to scroll from AIR to 32%, then press and hold again to scroll from 32 to 50%, then AIR.
• For each FO2 value that appears, the display indicates the Maximum Depth that can be achieved for a PO2 of
1.60 ATA (Fig. 20a), or the PO2 Alarm set point. If FO2 is set for AIR, no Depth value is displayed.
• Press the Left (Advance) button to accept the setting and advance to Set Depth Alarm, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
Fig. 19 - Set FO2
a
Fig. 20 - FO2 set for 32%
25
N
2
O
2
ATR
Fig. 21 - Set Max Depth Alarm
26
FEET MAX
TO SET - MAX DEPTH ALARM
(while in the Surface Mode)
Factory set for 330 feet, the Alarm can be set to values between 30 feet (3 meters) and 330 feet ( 99 meters) in increments of 10 feet (3 meters).
• Press Both buttons simultaneously, release when SET: 1 appears (2 seconds).
• Press and release the Left (Advance) button, FO2 ap­pears with the value flashing.
• Press the Left (Advance) button 1 more time.
• The graphics FEET MAX and dEEP, and Alarm icon appear with the Max Depth set point value flashing (Fig. 21).
• Press and release the Right (Select) button until the desired Depth Alarm value appears; or press and hold to scroll through the set points.
• Press the Left (Advance) button to accept the setting and advance to Elapsed Dive Time Alarm, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
ATR
O
2
N
2
TO SET - ELAPSED DIVE TIME ALARM
(while in the Surface Mode)
Factory set for 3:00 (hr:min), the Alarm can be set to values between 0:10 and 3:00 (hr:min) in increments of 5 minutes.
• Press Both buttons simultaneously, release when SET: 1 appears (2 seconds).
• Press and release the Left (Advance) button, FO2 ap­pears with the set point value flashing.
• Press the Left (Advance) button 2 more times.
• The graphic EdT, and Alarm and Dive Time icons appear with the Elapsed Dive Time value flashing (Fig. 22).
• Press and release the Right (Select) button until the desired Alarm value appears; or press and hold to scroll through the set points.
• Press the Left (Advance) button to accept the setting and advance to PC Interface, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
Fig. 22 - Set Elapsed Dive
Time Alarm
27
NOTE: For more information regarding PC Interface, refer to page 74 of this manual and to documents provided with the PC download product.
PC INTERFACE
PC Interface is not a setting, it is included in the Set 1 menu for easy access when data in the unit's memory is to be downloaded (copied) to the PC download software program for storage and viewing.
To download data (while in the Surface Mode)-
• Press Both buttons simultaneously, release when SET: 1 appears (2 seconds).
• Press and release the Left (Advance) button, FO2 ap­pears with the set point value flashing.
N
2
O
2
ATR
Fig. 23 - PC Interface
28
• Press the Left (Advance) button 3 more times.
• The graphic PC appears with a 120 second countdown timer (Fig. 23). Download must be initiated before the timer reaches 00 (within 2 minutes).
• Download is initiated by the external device requesting data transfer (i.e., the PC download program).
• The unit reverts to Surface Mode after completion of the Download operation, or after 2 minutes if neither button is pressed.
ATR
O
2
N
2
ENTERING SETTINGS -SET MODE #2
These settings are ones that are not likely to change. To save time at the dive site, verify the set points and adjust them as desired prior to departing on the day's dive trip.
TO SET - UNITS OF MEASURE
(while in the Surface Mode)
Factory set for Imperial, Units of Measure can also be set for Metric.
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the graphics FT, F, and PSI (or M, C, and BAR) flashing (Fig. 24).
• Press and release the Right (Select) button to toggle between Imperial and Metric units.
• Press the Left (Advance) button to accept the setting and advance to Set Hour Format, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
To return to Sur­face Mode at any time while in Set Mode, press and hold Both buttons for 2 seconds.
Fig. 24 - Set Units of Measure
29
N
2
O
2
ATR
Fig. 25 - Set Hour Format
30
TO SET - HOUR FORMAT (while in the Surface Mode) Factory set for 12 Hr (12: Am to 11: Pm), the Format can also be set for 24 Hr (0: to 23: hours).
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 1 more time.
• The graphic Hour appears with 12 (or 24) flashing (Fig. 25).
• Press and release the Right (Select) button to toggle between 12 and 24.
• Press the Left (Advance) button to accept the setting and advance to Set Hour, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
ATR
O
2
N
2
ATR
O
2
N
2
TO SET - TIME (while in the Surface Mode) Set for factory local time, the Time can be set to values between 1:00 and 12:59 (AM/PM) or 0:00 and 23:59.
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 2 more times.
• Time appears with the Hour value flashing (Fig. 26).
• Press and hold the Right (Select) button to scroll through the Hour values in increments of one hour.
• Press the Left (Advance) button to accept the setting.
• The Minute value flashes (Fig. 27).
• Press and hold the Right (Select) button to scroll through the Minute values in increments of one minute.
• Press the Left (Advance) button to accept the setting and advance to Set Year.
Fig. 26 - Set Hour
Fig. 27 - Set Minute
31
N
2
Fig. 28 - Set Year
N
2
Fig. 29 - Set Month
N
2
TO SET - DATE (after setting Time) Factory set for the factory local Date, the Date can be set to values between 01/01/01 and 12/31/20.
• After having set and accepted the Time, the Date ap-
pears with the graphic dAY, and Year value flashing (Fig. 28).
• Press and hold the Right (Select) button to scroll through the Year values in increments of one Year.
• Press the Left (Advance) button to accept the setting.
• The Month value flashes (Fig. 29).
• Press and release the Right (Select) button to scroll through the Month values in increments of one month.
• Press the Left (Advance) button to accept the setting.
• The Day value flashes (Fig. 30).
• Press and hold the Right (Select) button to scroll through the Day values in increments of one day.
32
Fig. 30 - Set Day
• Press the Left (Advance) button to accept the setting.
TO SET - AUDIBLE ALARM (while in the Surface Mode) Factory set for ON, the Alarm can be also be set to OFF. This setting also applies to the red LED Warning Indicator that is synchronized with the Audible Alarm.
When set OFF, the Alarm will not sound during the condi­tions described on page 14.
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 7 more times.
• The graphic ALM and Alarm (speaker) icon appear with On (or OFF) flashing (Fig. 31).
• Press and release the Right (Select) button to toggle between ON and OFF.
• Press the Left (Advance) button to accept the setting and advance to Set Max TLBG Alarm, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
N
2
O
2
ATR
Fig. 31- Set Audible Alarm
33
DECO
N
2
O
2
ATR
Fig. 32 - Set Max TLBG Alarm
34
TO SET - MAXIMUM TISSUE LOADING BAR GRAPH (TLBG) ALARM (while in the Surface Mode)
Factory set for DECO (all 8 segments), the Maximum Tissue Loading Bar Graph (TLBG) Alarm can be set to values between DECO (all 12 segments) and 1 segment.
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 8 more times.
• The graphic ndc and Alarm (speaker) icon appear with the full TLBG flashing (Fig. 32).
• Press and release the Right (Select) button to decrease the number of segments one at a time.
• Press the Left (Advance) button to accept the setting and advance to Set Air Time Remaining Alarm, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
10 20 30 40 50 6005
ATR
O
2
N
2
TO SET - AIR TIME REMAINING BAR GRAPH ALARM
(while in the Surface Mode)
Factory set for 5 minutes, the Air Time Remaining Bar Graph Alarm can be set to values of 0, 5, and 10 to 60 minutes in increments of 5 minutes.
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 9 more times.
• The graphic Atr, and Alarm (speaker) and Cylinder icons appear with the Air Time Remaining Bar Graph value flashing (Fig. 33).
• Press and release the Right (Select) button to decrease the Alarm value in increments of 5 minutes. Or press and hold to scroll.
• Press the Left (Advance) button to accept the setting and advance to Set Turn Pressure Alarm, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
Fig. 33 - Set Air Time Remaining
Bar Graph Alarm
35
N
2
O
2
ATR
Fig. 34 - Set Turn Pressure Alarm
PSI
36
TO SET - TURN PRESSURE ALARM
(while in the Surface Mode)
Factory set for 00 PSI/BAR (disabled), the Turn Pressure Alarm can be set to values between 1000 and 3000 PSI (69 to 205 BAR) in increments of 250 PSI (17 BAR).
• Press Both buttons simultaneously, release when SET: 2 appears.
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 10 more times.
• The graphic trn and Alarm (speaker) icon appear with the Turn Pressure Alarm value flashing (Fig. 34).
• Press and hold the Right (Select) button to scroll through the Alarm values in increments of 250 PSI (17 BAR).
• Press the Left (Advance) button to accept the setting and advance to Set End Pressure Alarm, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
PSI
ATR
O
2
N
2
TO SET - END PRESSURE ALARM
(while in the Surface Mode)
Factory set for 00 PSI/BAR (disabled), the End Pressure Alarm can be set to values between 300 and 1500 PSI (20 to 104 BAR) in increments of 100 PSI (7 BAR).
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 11 more times.
• The graphic End and Alarm (speaker) icon appear with the End Pressure Alarm value flashing (Fig. 35).
• Press and hold the Right (Select) button to scroll through the Alarm values in increments of 100 PSI (6 BAR).
• Press the Left (Advance) button to accept the setting and advance to Set Dive Time Remaining Alarm, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
Fig. 35 - Set End Pressure Alarm
37
N
2
O
2
ATR
Fig. 36 - Set Dive Time
Remaining Alarm
38
TO SET - DIVE TIME REMAINING ALARM
(while in the Surface Mode)
Factory set for 5 minutes, the Dive Time Remaining Alarm can be set to values between 0:00 and 0:20 (minutes) in increments of 1 minute.
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 12 more times.
• The graphic dtr, and Alarm (speaker) and Dive Mode icons appear with the Dive Time Remaining value flashing (Fig. 36).
• Press and hold the Right (Select) button to scroll through the Alarm values in increments of one minute.
• Press the Left (Advance) button to accept the setting and advance to Set Max PO2 Alarm, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
MAX
ATR
O
2
N
2
TO SET - MAXIMUM PO2 ALARM
(while in the Surface Mode)
Factory set for 1.60 (ATA), the Maximum PO2 Alarm can be set to values between 1.20 and 1.60 (ATA) in increments of .10 (ATA).
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 13 more times.
• The graphics PO2 and MAX, and Alarm (speaker) icon appear with the PO2 value flashing (Fig. 37).
• Press and release the Right (Select) button to advance the Alarm value in increments of .10 (ATA).
• Press the Left (Advance) button to accept the setting and advance to Set FO2 50% Default, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
Fig. 37 - Maximum PO2 Alarm
39
N
2
O
2
ATR
Fig. 38 - Set FO2 50% Default
40
TO SET - FO2 50% DEFAULT
(while in the Surface Mode)
Factory set ON, the FO2 50% Default feature can be set to OFF. The effects of this feature being ON or OFF are described on page 19.
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 14 more times.
• The graphics FO2 and 50 appear with ON (or OFF) flashing (Fig. 38).
• Press and release the Right (Select) button to toggle between ON and OFF.
• Press the Left (Advance) button to accept the setting and advance to Set Backlight Duration, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
ATR
O
2
N
2
TO SET - BACKLIGHT DURATION
(while in the Surface Mode)
Factory set for 3 seconds, the Backlight Duration can be set to values of 0 (disabled), 3, or 7 seconds. This is the length of time the Backlight stays On after the button is released.
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 15 more times.
• The graphic GLO and Time icon appear with the Dura- tion value flashing (Fig. 39).
• Press and release the Right (Select) button to advance the Duration from :00 to :03 to :07 (seconds).
• Press the Left (Advance) button to accept the setting and advance to Set Sampling Rate, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
Fig. 39 - Set Backlight Duration
41
NOTE: Sampling Rate is the interval at which data samples are recorded during a dive for subse­quent download to the PC pro­gram. This setting has no effect on dis­played data or data in the unit's viewable Log.
N
2
O
2
ATR
Fig. 40 - Set PC Sampling Rate
42
TO SET - PC SAMPLING RATE
(while in the Surface Mode)
Factory set at 15 seconds, the Sampling Rate can be set at values of 2, 15, 30, or 60 seconds, or 2, 5, or 10 feet (.5,
1.5, or 3 meters).
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 16 more times.
• The graphics SR and SECS appear with the Sampling Rate value flashing (Fig. 40).
• Press and release the Right (Select) button advance the Rate one selection at a time.
• Press the Left (Advance) button to accept the setting and advance to Set Digital Gauge Mode, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
ATR
O
2
N
2
TO SET - USER SET DIGITAL GAUGE MODE
(while in the Surface Mode)
Factory set OFF, Digital Gauge Mode can also be set ON. The effects of this feature being ON or OFF are described on page 64.
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
• Press the Left (Advance) button 17 more times.
• The graphic GAU appears with OFF (or ON) flashing (Fig. 41).
• Press and release the Right (Select) button to toggle between ON and OFF.
• Press the Left (Advance) button to accept the setting and advance to Set Water Activation, or press and hold Both buttons for 2 seconds to revert to Surface Mode.
NOTE: Once a dive is made with this feature set ON, the setting will be locked ON for 24 hours after the dive. Set Digital Gauge Mode will not appear as a selection during that 24 hour period.
Fig. 41 - Set Digital Gauge Mode
43
WARNING: If this feature is set for OFF, the Pro Plus 2 must be manu­ally activated prior to commencing a dive.
TO SET - WATER ACTIVATION
(while in the Surface Mode)
Factory set for ON, the Water Activation feature can also be set for OFF (disabled) to prevent inadvertent activation during travel or storage. When set ON, the Pro Plus 2 will automatically Activate when immersed in water.
• Press Both buttons simultaneously, release when SET: 2 appears (4 seconds).
• Press and release the Left (Advance) button, the Units screen appears with the set point flashing.
N
2
O
2
ATR
Fig. 42 - Set Water Activation
44
• Press the Left (Advance) button 18 more times
(14 more times after a dive in Digital Gauge Mode).
• The graphics ACT and H20 appear with ON (or OFF) flashing (Fig. 42).
• Press and release the Right (Select) button to toggle between ON and OFF.
• Press the Left (Advance) button to accept the setting and revert to Surface Mode.
WARNING: Prior to diving with the Pro Plus 2, you must also read and understand the Oceanic Dive Computer Safety and Reference Manual which provides Important Warnings and Safety Recom­mendations as well as general product information.
PLAN and DIVE MODES
45
Depth NDL feet (meters) hours:mins
30 (9 ) 4:20 (4:43) 40 (12) 2:17 (2:24) 50 (15) 1:21 (1:25) 60 (18) :57 (:59) 70 (21) :40 (:41) 80 (24) :30 (:32) 90 (27) :24 (:25) 100 (30) :19 (:20) 110 (33) :16 (:17) 120 (36) :13 (:14) 130 (39) :11 (:11) 140 (42) :09 (:09) 150 (45) :08 (:08) 160 (48) :07 (:07) 170 (51) :07 (:06) 180 (54) :06 (:06) 190 (57) :05 (:05)
No Decompression Limits
for an Air Dive
(no dive made yet)
46
PRE DIVE PLANNING SEQUENCE
TM
Oceanic strongly recommends that you review the Pre Dive Planning Sequence (PDPS) prior to every dive to help you plan your dive as required to avoid exceeding no decompression or oxygen exposure limits.
This is especially important for repetitive dives when the PDPS indicates adjusted dive times that are available for the next dive, based on residual nitrogen or oxygen accumula­tion (whichever is in control) following the last dive and surface interval.
The PDPS provides a sequence of theoretical dive times available for depths ranging from 30 feet (9 meters) to 190 feet (57 meters) in 10 foot (3 meter) increments.
During the sequence, no decompression times are only displayed for depths where there is at least 1 minute of theoretical dive time available at the depth, taking into account a descent and ascent rates of 60 feet (18 meters) per minute.
To access the Pre Dive Planning Sequence
MAX
ATR
FEET
O
2
N
2
MAX
O
2
ATR
FEET
O
2
N
2
(while in the Surface mode) -
• Press the Left (Advance) button 1 time
• Press and release the Right (Select) button to advance through the Depths/Times, one screen at a time; and then to Surface Mode.
• Press the Left (Advance) button to access Fly Mode.
• The unit will revert to Surface Mode after 2 minutes if no button is pressed.
• With each depth displayed by the Pre Dive Planning Sequence, you will see either predicted no decompres­sion limits (NDLs) based upon your previous dive profiles (if calculated to be nitrogen controlled), or predicted oxygen tolerance limits (OTLs) based upon either a single dive exposure or your 24 hour accumulation of oxygen (if calculated to be oxygen controlled).
• The Maximum Depth allowed for the PO2 Alarm and FO2 set points will also be displayed.
• Depths greater than the maximum depth that can be achieved with a PO2 of 1.60 ATA will not be displayed.
• If the Tissue Loading Bar Graph is displayed (Fig. 43), that next dive is calculated to be controlled by nitrogen loading.
• If the O2BG and O2 symbol are displayed (Fig. 44), it is calculated to be controlled by oxygen loading.
Fig. 43 - Nitrogen Control
Fig. 44 - Oxygen Control
47
a
N
2
O
2
10 2005
ATR
Fig. 45 - Tissue Loading Bar Graph
FEET
PSI
48
NOTE: The Pro Plus 2 will store oxygen accumula­tion calculations for up to 10 dives conducted during a 24 hour period. If the maximum limit for oxygen loading has been exceeded for that day (24 hour period), all of the segments of the O2 bar graph will be displayed flashing .
Depth/Time values will not appear until the O2 bar graph recedes into the green (normal) zone (i.e., your daily oxygen dosage decreases an amount equivalent to the amount accumulated during the latest dive completed).
TISSUE LOADING BAR GRAPH (Fig. 45a)
As your Depth and Elapsed Dive Time increase, the Tissue Loading Bar Graph (TLBG) will add segments (green toward red) to represent the absorption of nitrogen.
While ascending to shallower depths, the number of seg­ments displayed will begin to recede, offering a graphic representation of your multilevel diving capability.
PSI
10 2005
ATR
FEET
O
2
N
2
OXYGEN ACCUMULATION BAR GRAPH (Fig. 46a)
If FO2 was set for a numerical value (nitrox), the Oxygen Accumulation (O2) Bar Graph will add segments (green toward red) to represent oxygen accumulation for that dive, or 24 hour period, whichever amount is greater.
VARIABLE ASCENT RATE INDICATOR (Fig. 46b)
The Variable Ascent Rate Indicator (VARI) shows how fast you are ascending. When you exceed the maximum recom­mended ascent rate for the depth you are at (see below), it will enter the red (Too Fast) zone and you will be alerted by the graphic TOO FAST, all segments of the bar graph flashing, an Audible alarm and the red LED Warning Indica­tor, unless set OFF (Fig. 47). The warnings will stop when your ascent rate is slowed.
• At depths greater than 60 feet (18 meters), maximum recommended rate is 60 feet/minute (18 meters/minute).
• At depths of 60 feet (18 meters) and shallower, maxi­mum recommended rate is 30 feet/minute (9 meters/ minute).
NOTE: Refer to page 10 for additional information pertaining to the Variable Ascent Rate Indicator.
a
Fig. 46 - O2 Bar Graph & VARI
N
2
O
2
10 20 3005
ATR
Fig. 47 - Ascent Alarm
FEET
FAST
b
TOO
PSI
49
N
2
AIR TIME REMAINING BAR GRAPH
FEET
The Air Time Remaining Bar Graph (Fig. 48a) displays remain­ing breathing gas times of 60 minutes and less. As time remaining decreases, segments are removed from the bar graph from the right (green) toward the left (red) in increments of 5 minutes (from 60 to 0 minutes).
O
2
10 2005
ATR
Fig. 48 - Air Time Remaining
Bar Graph
PSI
When the time decreases to the Alarm set point, the Audible Alarm will emit a beep and the red Warning LED will flash until
a
acknowledged. You should begin a safe controlled ascent to the surface.
CONTROL OF DISPLAYS
During No Decompression conditions, various displays of information (up to 4) are available. Each provides Depth, Dive Time Remaining, and additional information. The intent of this feature is to provide the diver control of how much information is on the display at any given time during the dive. The diver can change from one display to another as often as desired by pressing the Left (Advance) button, otherwise it does not change. An exception to this is during a 3 minute no decompression dive Safety Stop (described on page 53).
50
PSI
10 2005
ATR
FEET
O
2
N
2
During conditions in which Cautionary type information is displayed (e.g., No Decom­pression Safety Stop, Decompression, High PO2, etc.), there is a Main Display of impor­tant information relevant to the specific condition. The diver can access another display, but it will automatically revert to the Main Display after 3 seconds.
To activate the Backlight during a dive, press the Right (Select) button.
• The display will be illuminated as long as the button is depressed, plus it will remain illuminated for the Backlight Duration time that has been set (3 or 7 seconds).
• The Backlight will not activate during a Low Battery condition or if the Backlight duration is set for 0.
NO DECOMPRESSION DIVE MODE
The Pro Plus 2 will enter the No Decompression Dive Mode when you descend deeper than 4 feet (1.2 meters).
No Decompression Dive Mode - Display #1 (Fig. 49)
Information includes Current Depth, Dive Time Remaining and icon, Tank Pressure, and the applicable bar graphs.
• Press and release the Left (Advance) button to view Display #2.
No Deco
Mode icon
Fig. 49 - No Deco #1
51
N
2
MAX
O
2
10 2005
ATR
Fig. 50 - No Deco #2
N
2
(lower portion same as Fig. 50)
Fig. 51 - No Deco #3
N
2
(lower portion same as Fig. 50)
Fig. 52 - No Deco #4
52
FEET
FEET
FEET
No Decompression Dive Mode - Display #2 (Fig. 50)
Information includes Current Depth, Max Depth for that dive and icon, Elapsed Dive Time and icon, Dive Time Remaining and icon, Tank Pressure, and the applicable bar graphs.
• Press and release the Left (Advance) button again to view Display #3.
PSI
No Decompression Dive Mode - Display #3 (Fig. 51)
Information includes Current Depth, Temperature, Time of Day, Dive Time Remaining and icon, Tank Pressure, and the applicable bar graphs.
• Press and release the Left (Advance) button again to view Display #4 (only if FO2 is set for a numerical value ­Nitrox), or revert to Display #1 (if FO2 is set for Air).
No Decompression Dive Mode - Display #4 (Fig. 52)
Information includes - Current Depth, current value of PO2 (if a nitrox dive) and PO2 graphic, Dive Time Remaining and icon, Tank Pressure, and applicable bar graphs.
• press the Left (Advance) button to view Display #1.
No Decompression Dive Mode - Secondary Display (Fig. 53)
PSI
10 20 3005
ATR
FEET
O
2
N
2
PSI
10 2005
ATR
FEET
O
2
N
2
To view, press the Left (Advance) button for 2 seconds.
Information includes Current Depth, Air Time Remaining (and Mode icon), and applicable bar graphs.
The Display will remain on the screen for 3 seconds then revert back to the Display previously selected (#1, 2, 3, 4).
No Decompression Dive Mode - SAFETY STOP (Fig. 54)
Upon ascending to 20 feet (6 meters) on any No Decompres­sion dive in which depth exceeded 30 feet (9 meters), a short beep will be emitted and a Safety Stop at 15 feet (4.5 meters) will appear on the display with a 3 minute count­down timer that counts down from 3:00 to :00 (min:sec).
The Safety Stop will be displayed until the countdown times out, or the diver descends below 30 feet (10 meters), or the diver surfaces. There is no Penalty if the diver surfaces prior to completing the Safety Stop.
Information includes Current Depth, Stop Depth (15 feet or
4.5 meters), Stop Bar icon, Countdown Timer, Dive Time Remaining, Cylinder Pressure, and applicable bar graphs.
Fig. 53 - No Deco Secondary
Fig. 54 - No Deco Safety Stop
53
DECOMPRESSION DIVE MODE
The Pro Plus 2 is designed to help you by providing a repre­sentation of how close you are to entering Decompression. Decompression Dive Mode (Fig. 55) activates when theoreti­cal No Decompression time/depth limits are exceeded.
Upon entering Decompression Mode, the Audible Alarm will sound and the red LED Warning Light will flash for 10 seconds (unless set OFF), or until acknowledged.
Deco Mode icon
• Press and hold the Left (Advance) button for 2 seconds to acknowledge/silence the Audible Alarm (unless set OFF).
• The UP Arrow and Deco Bar will flash if you are greater
DECO
N
2
FEET
than 10 feet (3 m) deeper than the Required Stop Depth.
• Once you are within 10 feet (3 m) of, and below, the Stop Depth, both Arrows and the Bar appear solid.
a
O
2
1005
ATR
Fig. 55 - Entry into Deco Mode
54
Total Ascent Time
Total Ascent Time (Fig. 55a) includes Stop Times required at all required decompression ceilings and vertical Ascent Time
PSI
calculated at 60 feet (18 meters) per minute for depths deeper than 60 feet (18 meters), and 30 feet (9 meters) per minute for depths of 60 feet (18 meters) and shallower.
PSI
10 2005
DECO
ATR
FEET
O
2
N
2
Managing Decompression Stops
To fulfill your decompression obligation, you should make a safe controlled ascent to a depth slightly deeper than, or equal to, the Required Ceiling Stop Depth indicated (Fig. 56a) and decompress for the Stop Time indicated (Fig. 56b).
The amount of decompression credit time that you receive is dependent on depth, with slightly less credit given the deeper you are. You should stay slightly deeper (Fig. 56c) than the Required Stop Depth indicated until the next shallower Stop Depth appears. Then, you can slowly ascend to, but not shallower than that indicated ceiling Stop Depth.
• While in Decompression Mode, the Pro Plus 2 will automatically revert to the Main Display after 3 seconds unless the Left (Advance) button is pressed to view an­other display of information.
Decompression Dive Mode - Main Display #1 (Fig. 56)
Information includes - Current Depth, required Ceiling Stop Depth and Time, Total Ascent Time (and Mode icon), both Arrows and the Deco Bar, and the applicable bar graphs.
• Press and release the Left (Advance) button to view Display #2.
c
b
a
Fig. 56 - Deco #1
55
DECO
N
2
MAX
O
2
10 2005
ATR
FEET
PSI
Fig. 57 - Deco #2
DECO
N
2
(lower portion same as Fig. 57)
Fig. 58 - Deco #3
DECO
N
2
FEET
FEET
Decompression Dive Mode - Display #2 (Fig. 57)
Information includes - Current Depth, Maximum Depth for that dive (and icon), Elapsed Dive Time (and icon), Total Ascent Time (and Mode icon), both Arrows and the Deco Bar, and the applicable bar graphs.
• Press and release the Left (Advance) button again to view Display #3.
Decompression Dive Mode - Display #3 (Fig. 58)
Information includes - Current Depth, Temperature, Time of Day (and icon), Total Ascent Time, both Arrows and the Deco Bar, and the applicable bar graphs.
• Press and release the Left (Advance) button again to view
- Display #4 (only if FO2 is set for a numerical value ­Nitrox), or revert to Display #1 (if FO2 is set for Air).
Decompression Dive Mode - Display #4 (Fig. 59)
Information includes - Current Depth, the current value of PO2 (if a nitrox dive) and PO2 graphic, Total Ascent Time, Mode icon, both Arrows and the Deco Bar, and applicable bar graphs.
(lower portion same as Fig. 57)
Fig. 59 - Deco #4
56
• Press the Left (Advance) button to view Display #1.
PSI
10 2005
DECO
ATR
FEET
O
2
N
2
VIOLATION MODES
While in Violation Modes the Alternate Displays previously described can be accessed using the Left (Advance) button, and the Backlight can be activated using the Right (Select) button.
• The Pro Plus 2 will automatically revert to the Main Display after 3 seconds unless the Left (Advance) button is pressed to view another display of information.
Conditional Violation Mode (Fig. 60)
If you ascend shallower (Fig. 60a) than a Required Decompression Ceiling Stop Depth displayed (Fig. 60b), the Down Arrow, Deco Bar, and Total Ascent Time
will flash until you descend below the Required Stop Depth. Also displayed will be Current Depth and applicable bar graphs.
a
Unless set OFF, the Audible Alarm will emit a continuous tone and the red LED Warning Light will flash for 10 seconds, or until acknowledged by pressing the Left (Advance) button.
If you descend below the required decompression ceiling before 5 minutes have elapsed, the Pro Plus 2 will continue to function in Decompression Dive Mode. In this case, no off­gassing credit will be given, and for each minute above the ceiling 11/2 minutes of Penalty Time will be added to Required Stop Time.
b
Fig. 60 - Conditional Violation
57
The added Penalty (decompression) Time will have to be 'worked off' first, before obtaining off-gassing credit. Once the Penalty Time is worked-off, and off-gassing credit begins, required decompression Stop Depths and Time will decrease toward zero. The Tissue Loading Bar Graph will recede into the yellow Caution Zone and the Pro Plus 2 will revert to the No Decompression Dive Mode.
NOTE: Upon entry into the following 3 Delayed Violation Modes, the red LED warning light will light and the Audible Alarm will sound, even if Set OFF. The Alarm cannot be acknowledged (silenced) by pressing the Left (Advance) button.
DECO
N
2
O
2
10 2005
ATR
Fig. 61 - Delayed Violation #1
58
FEET
Delayed Violation Mode #1 (Fig. 61)
(This is a continuation of a Conditional Violation.)
If you remain above the Required Ceiling Stop Depth for more than 5 minutes, the full Tissue Loading
Bar Graph and Total Ascent Time will flash until you descend below the Required Stop Depth.
PSI
MAX
PSI
10 2005
ATR
FEET
O
2
N
2
PSI
10 20 30 40 5005
DECO
ATR
FEET
O
2
N
2
Delayed Violation Mode #2 (Fig. 62)
The Pro Plus 2 cannot calculate decompression times for Stop Depths much greater than 60 feet (18 meters) and offers no indication of how much time spent underwater would result in the need for a greater Stop Depth.
If the Decompression obligation requires a Ceiling Stop Depth between 60 feet (18 meters) and 70 feet (21 meters), the Tissue Loading Bar Graph will flash.
Total Ascent Time will still be displayed.
You must ascend to just deeper than, and stay as close as possible to, 60 feet (18 meters) without causing the Total Ascent Time display to flash. When the Required Stop Depth indicates 50 FT/ 15 M, etc., you can ascend to those depths and continue decompressing.
Delayed Violation Mode #3 (Fig. 63)
If you descend deeper than 330 feet (99.9 meters), the accumulated Tissue Loading Bar Graph seg-
ments will flash, and the Current Depth and Max Depth displays will only indicate 3 dashes ( - - -).
Fig. 62 - Delayed Violation #2
Fig. 63 - Delayed Violation #3
59
DECO
N
2
MAX
O
2
O
2
1005
ATR
Fig. 64 - Immediate Violation
Gauge Mode (underwater)
60
FEET
Upon ascending above 330 feet (99.9 meters), the Current Depth display will be restored, however Max Depth will only display 3 dashes for the remainder of that dive. The Log for that dive will display 3 dashes as the Max Depth achieved.
Immediate Violation Mode and Gauge Mode
During a Dive, if a ceiling much greater than 60 feet (18 meters) is required, an Immediate Violation Mode will be entered. This situation would be preceded by entering Delayed Violation Mode #2.
The Pro Plus 2 would then operate with limited functions in Gauge Mode during the remainder of that dive and for 24 hours after surfacing. Gauge Mode turns the Pro Plus 2 into a digital instrument without any decompression or oxygen monitoring functions.
Only Current Depth, Max Depth, Elapsed Dive Time, and the bar graphs will be displayed (Fig. 64). The full Tissue Load­ing Bar Graph and O2 Bar Graph will both flash as a warn-
PSI
ing of this condition.
• Press the Left (Advance) button to view Temperature and Time of Day.
• Press the Right (Select) button to activate the Backlight.
The Pro Plus 2 will also enter an Immediate Violation
Responsible
PSI
O
2
DECO
ATR
O
2
N
2
Mode 5 minutes after reaching the surface from a dive in which a Delayed Violation occurred.
On the surface, Gauge Mode displays the Dive Number, Temperature, Time of Day, and Surface Interval (Fig. 65). It does not provide the FO2, PDPS, or Time to Fly and Desaturate features.
The countdown timer that appears when you try to access Time to Fly does not represent 'Time to Fly'. It is only pro­vided to inform you of the time remaining before normal Pro Plus 2 operation can resume with full features and functions.
This condition is a Permanent Violation, and in the event that a dive is made during the 24 hour period, a full 24 hour surface interval must then be served before all functions are restored.
Responsible
Fig. 65 - Immediate Violation
Gauge Mode (after surfacing)
61
N
2
O
2
O
2
ATR
Fig. 66 - PO2 Warning
N
2
O
2
O
2
ATR
Fig. 67 - PO2 Alarm
62
FEET
HIGH PO2 DIVE MODE
When partial pressure of oxygen (PO2) becomes equal to, or greater than, 1.40 ATA, or 0.2 ATA less than the PO2 Alarm set point (a user setting); the red LED warning will flash, the Audible Alarm will sound (unless set OFF), and the current PO2 value, PO2 symbol, O2 segment of the O2 Bar Graph, and UP Arrow will appear on the
10 20 3005
PSI
Main Display as a warning until PO2 decreases. Current Depth and Dive Time Remaining will also be displayed (Fig.
66).
If PO2 continues to increase, the value displayed will in­crease toward a maximum value of 5.00 ATA in increments
FEET
of .01 ATA. When it reaches a value of 1.60 ATA, or the PO2 Alarm set point (a user setting), the Audible
Alarm will sound (unless set OFF) and the current PO2 value, PO2 symbol, O2 segment of the O2 Bar Graph, and UP Arrow will flash as a warning until PO2 decreases (Fig. 67).
• Press the Left (Advance) button to view the Alternate Displays.
10 20 3005
PSI
• The unit will revert to the Main Display after 3 seconds.
• Press the Right (Select) button to activate the Backlight.
PSI
1005
O
2
O
2
ATR
FEET
O
2
N
2
HIGH OXYGEN ACCUMULATION
The O2 Bar Graph displays either oxygen accumulated during that nitrox dive, or during the repetitive nitrox dives you conduct during that 24 hour period, whichever of the two is greater at that time.
The yellow Caution Zone of the O2 Bar Graph offers you a convenient way to consistently monitor how close you are coming to the limits of oxygen exposure. Use it as a
visual reference to place a wider margin of protec­tion between you and the Limits.
If the theoretical amount of oxygen accumulated equals, or exceeds, the limit for a single exposure, or the exposure limit for a 24 hour period, Oxygen Dive Time Remaining becomes zero (0:00) and the O2 Bar Graph will enter the red O2 (Danger) Zone (Fig. 68). The red LED warning will flash, the Audible Alarm will sound (unless set OFF), and the UP Arrow and the full O2 Bar Graph will flash as a warning until the level of oxygen decreases below the limit.
• Press the Left (Advance) button to view the Alternate Displays.
• The unit will revert to the Main Display after 3 seconds.
• Press the Right (Select) button to activate the Backlight.
Fig. 68 - High O2
63
N
2
MAX
O
2
1005
ATR
FEET
PSI
Fig. 69 - User Set Gauge Mode
USER SET DIGITAL GAUGE MODE
When Digital Gauge Mode is set for ON, the Pro Plus 2 will operate as a Digital Depth Gauge/Timer without performing nitrogen and oxygen calculations (Fig. 69).
Time of Day and Elapsed Dive Time will also be displayed.
• Press the Left (Advance) button to view Air Time Remain­ing for 3 seconds in place of Elapsed Dive Time (Fig. 70).
UNEXPECTED LOSS OF DISPLAYED INFORMATION
N
2
MAX
O
2
1005
ATR
Fig. 70 - User Set Gauge Mode
(Air Time Remaining)
64
FEET
If your Pro Plus 2 stops working for any reason, it is impor­tant that you have anticipated this possibility and are pre­pared for it. This is an important reason for not pushing the no decompression and oxygen accumulation limits, and a critical reason to avoid entering decompression. If you dive in situations where your trip would be ruined or your safety would be jeopardized by losing the use of your Pro Plus 2, a backup instrument system is highly recommended.
PSI
WARNING: Prior to diving with the Pro Plus 2, you must also read and understand the Oceanic Dive Computer Safety and Reference Manual which provides Important Warnings and Safety Recom­mendations as well as general product information.
POST DIVE MODES
65
N
2
O
2
ATR
Fig. 71 - Transition Period
66
POST DIVE SURFACE MODE
When you ascend to 3 feet (1 meter) or shallower, the Pro Plus 2 will enter Surface Mode and begin counting your surface interval.
TRANSITION PERIOD
The first 10 minutes is, in affect, a Transition Period during which time the following information is displayed (Fig. 71):
• Number of that dive (during that day)
• Battery Consumption Indicator
• Temperature (ambient)
• Time of Day and icon
• Surface Interval time (colon flashing) and icon (flashing)
• Tissue Loading Bar Graph indicating current nitrogen loading
• O2 Bar Graph indicating current oxygen accumulation (if nitrox)
• During the Transition Period, Log Mode can be accessed.
PSI
No other modes (e.g., PDPS, Fly, Desat, Set, PC) are accessible.
• Press the Right (Select) button to activate the Backlight.
To view that dive's Log (Fig. 72) -
ATR
O
2
N
2
Refer to page 64 for a description of the Log Mode displays.
• Press the Left (Advance) button 1 time.
• Press the Right (Select) button 1 time to view the Nitro-
gen data screen.
• Press the Right (Select) button again to view the Oxygen
data screen (if a nitrox dive).
• Press Both buttons simultaneously for 2 seconds to return to Surface Mode.
• The unit will revert to Surface Mode after 2 minutes if no button is pressed.
Log Data will not be stored in the unit's memory until the 10 minute Transition Period on the surface is completed.
Once 10 minutes have elapsed, the Surface Mode icon and Surface Interval time display colon stop flashing indicating that the Dive and Transition Period are completed, and a subsequent descent will be considered a new dive.
If you descend during the 10 minute Transition Period, time underwater will be considered a continuation of that dive. The time at the surface (if less than 10 minutes) will not be added as Dive Time.
Fig. 72 - Log Mode
(during Transition Period)
67
N
2
AFTER THE TRANSITION PERIOD (THE FIRST 2 HOURS)
For the remainder of the first 2 hours after surfacing, information will continue to be displayed as Surface Mode (Fig. 73) and you will have full access to other modes (e.g., PDPS, Fly, Desat, Log, Set, etc.).
O
2
ATR
Fig. 73 - Surface Mode (>10 min)
N
2
O
2
ATR
Fig. 74 - Adjusted NDLs
FEET
68
To activate the Backlight -
• Press the Right (Select) button.
To access the Pre Dive Planning Sequence -
(Also refer to page 46)
• Press the Left (Advance) button 1 time (while in Surface
Mode).
• Press and release the Right (Select) button to advance through the sequence of available 'adjusted' depths and times one screen at a time.
• The unit will revert to Surface Mode after 2 minutes, unless the Left (Advance) button is pressed to access the Fly Mode.
The Pre Dive Planning Sequence now shows adjusted No Decompression Limits (Fig. 74) based on residual nitrogen and accumulated oxygen calculated to be remaining from the previous dives.
To access the Time to Fly Countdown (while in Surface Mode)
ATR
O
2
N
2
ATR
O
2
N
2
• Press the Left (Advance) button 2 times.
• The Time to Fly counter (Fig. 75) begins counting down 10 minutes after surfacing from a dive (after the Transition Period) displaying the word 'FLY' and a countdown that begins at 23:50 (hr:min) and counts down to 0:00 (hr:min).
• If a Violation occurred during the dive a single dash ( - ) will appear instead of the letters FLY.
• The unit will revert to Surface Mode after 2 minutes, unless the Left (Advance) button is pressed to access the Desat Time Countdown.
To access the Time to Desaturation Countdown
(while in Surface Mode)
• Press the Left (Advance) button 3 times.
• The unit will revert to Surface Mode after 2 minutes, unless the Left (Advance) button is pressed to access the Log Mode.
• If a Violation occurred during the dive, Desaturation Time will not be displayed.
The Time to Desaturate counter (Fig. 76) provides calculated time for tissue desaturation at sea level. The countdown starts 10 minutes after surfacing at 23:50 (hr:min) maximum and counts down to 0:00 (hr:min).
Fig. 75 - Time to Fly
Fig. 76 - Time to Desaturate
69
LOG MODE
Information from your latest 24 dives is stored in the Log for viewing. The first dive of a new 'activation period' will be #1, then #2, etc. After 24 dives are accumulated, each subsequent dive will overwrite the oldest dive in the Log (i.e., the most recent dive deletes the oldest). Log information will not be lost when the battery is removed, but factory service will delete data.
Dives are displayed in a reverse sequence that starts with the dive most recently recorded back to the oldest of the 24 dives stored. Thus, your most recent dive will always be the first shown in the sequence. Each dive has three log screens - Date/Time started, Nitro­gen data, and Oxygen data (not displayed if the dive included a Violation). Dives are identified by the Date/Time started and 'number' that day.
Button Control in Log Mode -
• The Left (Advance) button is used to access a specific dive's Log.
• HINT: To bypass a dive's Log and search for another in the sequence, press the Left (Advance) button repeatedly. Do not press the Right (Select) button until you find the dive Log you wish to view.
• The Right (Select) button is used to view the second and third screens (Nitrogen and Oxygen related data) for that dive.
• To return to Surface Mode at any time while in Log Mode, press Both buttons simulta­neously for 2 seconds.
• The unit will automatically revert to Surface Mode after 2 minutes if no button is pressed while in the Log Mode.
70
To access the Log Mode and view the First Screen
ATR
O
2
N
2
ATR
FEET MAX
O
2
N
2
(While in Surface Mode) -
• Press the Left (Advance) button 4 times.
• The First Screen of the most recent dive will appear
displaying (Fig. 77) -
• The Log Mode icon
• Dive Number
• Time/ Date the dive started
To view the Second Screen of the Log (Nitrogen data) -
• Press the Right (Select) button 1 time (while viewing the
Time/Date). Displayed will be (Fig. 78) -
• Log Mode icon
• Maximum Depth - reached during the dive (and icon)
• Temperature - minimum during the dive (and icon)
• Surface Interval - prior to that dive (and icon)
• Elapsed Dive Time (and icon)
• Variable Ascent Rate Indicator - showing the maximum
ascent rate maintained for 4 consecutive seconds during the dive
• Tissue Loading Bar Graph - showing tissue nitrogen
loading at the time you surfaced at the end of the dive. Also, the segment that reflects the maximum loading will appear flashing.
Fig. 77 - Log (Date / Time)
Fig. 78 - Log (Nitrogen Data)
71
N
2
O
2
ATR
Fig. 79 - Log (Oxygen Data)
To view the Third Screen of the Log (Oxygen data)
• Press the Right (Select) button 1 time (while viewing the
Nitrogen data). Displayed will be (Fig. 79)-
• Log Mode icon
• FO2 graphic and value set for that dive
• Maximum PO2 level reached during that dive and PO2 graphic
• O2 Bar Graph - showing oxygen loading at the end of the dive
To access the first screen of the previous dive's Log
• Press the Left (Advance) button 1 time.
N
2
O
2
ATR
Fig. 80 - Surface Mode
72
AFTER THE FIRST 2 HOURS
Two hours after the last dive, Surface Mode will no longer be displayed, the Time to Fly and Desaturation countdowns will be displayed alternately for 3 seconds each until they count down to 0:00 or another dive is made.
To access other modes or enter settings -
• Press Either button to return to Surface Mode (Fig. 80).
• The unit will again revert to the Time to Fly and Desaturation countdowns after 2 hours, if no button is pressed.
WET CONTACTS
ATR
O
2
N
2
ATR
O
2
N
2
If the graphic H2O appears during the Fly Mode (Fig. 81) and Desaturation Mode (Fig. 82) countdowns that alternate 2 hours after a dive, it is an indication that the water activa­tion contacts are bridged (still wet) and the unit must be rinsed in fresh water and thoroughly dried.
One contact is located between the control buttons and the other at the base of the Right (Select) button's stem.
• Once the unit is dry, the graphic H2O will disappear
from the display.
• If the unit is not cleaned and dried prior to the count­downs reaching 0:00 (hr:min), or making another dive, it will shut off then automatically reactivate.
• The graphic H2O would then appear in place of Dive Number when Surface Mode is displayed during the Surface Mode.
• If no dive is made, the unit would shut off after 2 hours, then automatically reactivate again, repeating the action until cleaned and dried.
Fig. 81 - Fly Mode
(Activation Contacts Wet)
Fig. 82 - Desaturation Mode
(activation contacts wet)
73
N
2
O
2
ATR
Fig. 83 - PC Interface
74
DOWNLOADING DATA TO A PC
Using special linking hardware, dive data can be down­loaded (copied) from your Pro Plus 2 to an IBM compatible PC program running on a Windows® operating system. Compatibility requirements and instructions are provided with the optional download package that is available from your Authorized Oceanic Dealer. The software program provides dive tabular and graphic profile data sampled throughout the dives at the interval you set.
NOTE: Ensure that the download product that you acquire is compatible with the Pro Plus 2 and the PC equipment that you will be using.
The Interface Cable will be connected to the Data Port located on the side of the Pro Plus 2 housing.
Prior to attempting to download data from your Pro Plus 2, refer to the instructions provided in the User Manual that is incorporated into the CD for the download package.
Refer to page 28 of this manual for instructions regarding access to PC Interface (Fig. 83).
WARNING: Prior to diving with the Pro Plus 2, you must also read and understand the Oceanic Dive Computer Safety and Reference Manual which provides Important Warnings and Safety Recom­mendations as well as general product information.
SIMULATOR (DEMO) MODE
75
N
2
DEMO
O
2
ATR
Fig. 84 - Simulator Mode
SIMULATOR MODE
This mode provides you with the ability to practice various dive mode scenarios and computer functions while observing the various displays.
• The SmartGlo Backlight functions as it normally does.
• At any time while in Simulator Mode, pressing and holding both buttons simultaneously for 2 seconds shall revert operation to real Surface Mode.
• The real set points entered into the Pro Plus 2 do not affect the operation of the Simulator which has its own settings that allow Digital Gauge Mode to be set ON or OFF, calculations to be cleared, and FO2 to be set.
N
2
DEMO
O
2
ATR
Fig. 85 - Set Demo Gauge
76
Access and Setup (while in real Surface Mode)
• Press and hold Both buttons for 6 seconds.
• Release the buttons during the 2 second window when SIM and DEMO appear (Fig. 84).
• Press and release the Left (Advance) button to access SIM MODE. The graphics GAU and DEMO will appear with OFF (or ON) flashing (Fig. 85).
• When set ON, the SIMULATOR will operate as the unit would in User Set Digital Gauge Mode only displaying Depth and Elapsed Dive Time. When set OFF, it oper­ates as it would as an Air or Nitrox computer.
DEMO
ATR
O
2
N
2
• Press and release the Right (Select) button to toggle
between Gauge ON and OFF.
• Press and release the Left (Advance) button to accept the
setting and advance to DEMO: NI-O2 with CUR (or NEW) flashing (Fig. 86).
• When set for NEW, calculations are based on zero
residual nitrogen and oxygen loading (a clean dive).
• When set for CUR, calculations take into consideration
any residual nitrogen and oxygen remaining from previ­ous 'actual' dives.
• Subsequent (repetitive) Simulated dives do not take into consideration previous Simulated dives.
• Press and release the Right (Select) button to toggle between NEW and CUR.
• Press and release the Left (Advance) button to accept the setting and advance to DEMO: PSI (or bAr) with OFF, CYL, or 3000 flashing (Fig. 87).
• When set for OFF, no calculations are made and no displays appear relating to tank pressure.
• When set for CYL, Air Time Remaining is calculated based upon actual pressure of the tank that the Pro Plus 2 is connected to.
• When set for 3000, Air Time Remaining is calculated based upon a simulated tank pressure of 3000 psi.
Fig. 86 - Set Demo Calibration
N
2
DEMO
O
2
ATR
Fig. 87 - Set Pressure On/Off
PSI
77
N
2
DEMO
O
2
ATR
Fig. 88 - Set Demo FO2
N
2
DEMO
• Press and release the Right (Select) button to toggle
between Pressure ON and OFF.
• Press and release the Left (Advance) button to accept the
setting and revert to Demo Surface Mode with the DEMO icon flashing.
• Press and release the Left (Advance) button to access
DEMO: FO2 with the DEMO icon and Air (or a numeri­cal value) flashing (Fig. 88).
• Press and hold the Right (Select) button to scroll from Air to 21% through 32% in 1% increments. The scroll will stop when the button is released, or at 32%.
• Press and hold the Right (Select) button again to scroll from 32% through 50% in 1% increments, then to Air. The scroll will stop when the button is released, or at Air.
• Note that pressing and releasing the Right (Select) button advances the FO2 setting from AIR to 21 through 50, in increments of 1% per button depression (no scroll).
• Press and release the Left (Advance) button to accept the setting and revert to Demo Surface Mode with the DEMO icon flashing.
O
2
10 20 30 40 50 6005
ATR
Fig. 89 - Demo Dive Mode
78
PSI
Demo Dive Mode
• Press and hold the Right (Select) button for 2 seconds to access Demo Dive Mode. The DOWN Arrow will appear flashing (Fig. 89).
DEMO
PSI
10 2005
ATR
FEET MAX
O
2
N
2
Descending
DEMO
PSI
10 20 30 40 50 6005
ATR
O
2
N
2
Hint: Quick taps (<2 seconds) of the Right (Select) button starts/stops Descents and Time Acceleration. Quick taps of the Left (Advance) button accesses Alternate displays.
• While the Down Arrow is flashing, tap (press/release)
the Right (Select) button to begin a Descent at a rate of 5 feet (1.5 meters) per real second.
• Tap (press/release) the Right (Select) button during the Descent to stop the Descent.
• Tap (press/release) the Left (Advance) button during the Descent, or stopped, to access the Alternate Displays.
• Press and hold the Right (Select) button for 4 seconds to access Time Acceleration. The small clock icon begins flashing (Fig. 91).
• Tap (press/release) the Right (Select) button to increase Elapsed Dive Time 1 minute per real time second.
• Tap (press/release) the Right (Select) button during Time Acceleration to restore normal time rate of 1 second per real second.
• To start an Ascent, tap (press/release) the Right (Select) button to stop a Descent.
• To revert operation to real Surface Mode, press and hold Both buttons simultaneously for 2 seconds.
Fig. 90 - Dive Mode Setup
Fig. 91 - Time Acceleration
79
N
2
DEMO
O
2
1005
ATR
Fig. 92 - Ascent set at 1 fps
N
2
DEMO
O
2
10 2005
ATR
Fig. 93 - Ascent set at 3 fps
80
FEET
PSI
FEET
TOO
FAST
PSI
Ascending
Rate of Ascent can be set for 1 or 3 feet per seconds (fps) prior to start of the Ascent. To change the Rate during an Ascent, it must first be stopped.
• Pressing and holding the Left (Advance) button for 2
seconds will access an Ascent Rate of 1 fps. The green and yellow segments of the VARI and small UP Arrow will appear with the Arrow flashing (Fig. 92).
• To set the Rate at 3 fps (Too Fast), press and hold the Left (Advance) button for 2 seconds. The red segment of the VARI will appear together with the graphic TOO FAST and an outer Arrow around the small one (Fig. 93). The Alarm will sound and red LED will flash.
• Pressing and holding the Left (Advance) button for 2 seconds will return the Rate to 1 fps.
• Tap (press/release) the Left (Advance) button while the Arrow is flashing to begin an Ascent at the Rate set.
• Tap (press/release) the Left (Advance) button during the Ascent to stop the Ascent.
• Tap (press/release) the Left (Advance) button while the Ascent is stopped to access Alternate displays.
• Press and hold the button for 2 seconds while the Ascent is stopped to restart the Ascent.
• Pressing and holding both Buttons simultaneously for 2 seconds will revert operation to real Surface Mode.
DEMO
PSI
ATR
O
2
N
2
Surfacing from a Simulated Dive
• The Simulator will enter Demo Surface Mode (Fig.
94) upon ascending to 3 feet (1 meter) or shallower.
• Press and hold the Right (Select) button for 4 seconds to access Time Acceleration. The small Time Clock icon will begin flashing.
• Tap (press/release) the Right (Select) button to increase Surface Interval Time 1 minute per real time second.
• Tap (press/release) the Right (Select) button during the Time Acceleration to restore the time rate to normal.
• FO2 can be set 10 minutes after surfacing.
• Tap (press/release) the Left (Advance) button to access FO2 Set Mode. The FO2 graphic and value previously set will appear with the value flashing.
• Press/release the Right (Select) button to increase the FO2 value in increments of 1%, or press and hold it to scroll.
• Tap (press/release) the Left (Advance) button to accept the FO2 setting and return to Demo Surface Mode.
• To make another dive, press and hold the Right (Select) button for 2 seconds to access Dive Mode.
• To revert operation to real Surface Mode, press and hold Both buttons simultaneously for 2 seconds.
Fig. 94 - Demo Surface Mode
81
82
Be a -
RESPONSIBLE DIVER
at all times.
R
E
R
E
S
V
P
I
O
D
N
E
S
L
I
B
WARNING: Prior to diving with the Pro Plus 2, you must also read and understand the Oceanic Dive Computer Safety and Reference Manual which provides Important Warnings and Safety Recom­mendations as well as general product information.
GENERAL
83
a
X
X
b
Fig. 95 - Case Back
84
CARE AND CLEANING
Protect your Pro Plus 2 from shock, excessive temperatures, chemical attack, and tampering. Protect the lens against scratches with a transparent Instrument Lens Protector. Small scratches will naturally disappear underwater.
• Soak and rinse the Pro Plus 2 in fresh water at the end of each day of diving, and check to ensure that the areas around the Low Pressure (Depth) Sensor (Fig. 95a), Download Interface Port (Fig. 95b), and Buttons are free of debris or obstructions.
• To dissolve salt crystals, use lukewarm water or a slightly acidic white vinegar/water bath. After removal from the bath, place the unit under gently running water and towel dry before storing.
• Transport your unit cool, dry, and protected.
INSPECTIONS AND SERVICE
Your Pro Plus 2 should be inspected annually by an Authorized Oceanic Dealer who will perform a factory
X
prescribed function check and inspection for damage or wear. To keep the 2 year limited warranty in effect, this inspection must be completed one year after purchase (+/-
X
30 days). Oceanic recommends that you continue to have this inspection performed every year to ensure it is working properly. The costs of annual inspections are not covered under the terms of the 2 year limited warranty.
To Obtain Service:
Take your Pro Plus 2 to an Authorized Oceanic Dealer or send it to the nearest Oceanic Regional Distributor Facility.
To return your Pro Plus 2 to Oceanic:
• Record all dive data in the Log and/or download the data in memory. All data will be erased when it receives factory service.
• Package it using a protective cushioning material.
• Include a legible note stating specific reason for return, your name, address, daytime phone number, serial number, and a copy of your original sales receipt and Warranty Registration Card.
• Send freight prepaid and insured using a traceable method to the nearest Oceanic Regional Service Facility (see page 96), or to Oceanic USA.
• If shipping to Oceanic USA, obtain an RA (Return Autho­rization) number by contacting Oceanic at 510/562­0500 or send an e-mail to service@oceanicusa.com.
• Non-warranty service must be prepaid. COD is not accepted.
• Additional information is available at the Oceanic web site OceanicWorldWide.com
R
R
E
E
S
V
P
I
O
D
N
E
S
L
I
B
85
NOTE: The procedures that follow must be closely adhered to. Damage due to improper Battery replace­ment is not covered by the unit's 2 year warranty.
BATTERY REPLACEMENT
The Battery Compartment should only be opened in a dry and clean environment with extreme care taken to prevent the entrance of moisture or dust.
As an additional precautionary measure to prevent formation of moisture in the Battery Compartment, it is recommended that the Battery be changed in an environment equivalent to the local outdoor temperature and humidity (e.g., do not change the Battery in an air conditioned environment then take it outside during a hot sunny day).
Inspect the Buttons, Lens, and Housing to ensure they are not cracked or damaged. If there is any sign of moisture in the module, DO NOT use the Pro Plus 2 until it receives proper service by an Authorized Oceanic Dealer, or the Oceanic factory.
86
R
R
E
E
S
V
P
I
O
D
N
E
S
L
I
B
Hot Swap
If the new Battery can be inserted within 8 seconds after the old one is removed (Hot Swap), settings and nitrogen and oxygen calculations for repetitive dives will be retained.
Battery Removal
Examine the Case Back to locate the Battery Hatch (Fig. 96a):
• Remove the 4 screws that secure the Battery Hatch to the Housing by turning counter clockwise.
• Lift the Hatch up and away from the Housing.
• Lift the Battery, positive (+) end first, out of the Battery Compartment.
• Remove the Battery Hatch O-ring and inspect it for any signs of deterioration or deformity. DO NOT use tools to remove the O-ring.
• O-ring replacement is highly recommended to ensure proper sealing.
• Closely examine the sealing surfaces of the Battery Hatch and Housing for any signs of damage that might prevent proper sealing. If found, return the Pro Plus 2 to an Authorized Oceanic Dealer, and DO NOT attempt to use it until it has received factory service.
• Closely examine the inside of the Battery Compartment for any signs of corrosion indicating entrance of moisture into the unit.
• If corrosion is found, return the Pro Plus 2 to an Autho­rized Oceanic Dealer, and DO NOT attempt to use it until it has received factory service.
(continued on page 88)
X
X
a
Fig. 96 - Battery Hatch
X
X
87
+
a
Fig. 97 - O-ring Installation
+
Fig. 98 - Battery Orientation
88
Moisture in the Battery Compartment
• If moisture is found, it is best to have the unit inspected and cleaned by an Authorized Oceanic Dealer.
• If it is necessary to clean the Battery Compartment, flush the it and all components with a solution of 50% white vinegar and 50% fresh water. Rinse with fresh water, and allow to dry overnight, or blow dry with a hair dryer (set at 'no heat').
• Closely check all of the sealing surfaces for any signs of damage that might impair proper sealing.
• Inspect the Buttons, Lens, and Housing to ensure they are not cracked or damaged.
Battery Installation
• Replace the Hatch O-ring with a new one. This O-ring must be a genuine Oceanic part that can be purchased from an Authorized Oceanic Dealer. Use of any other O-ring will void the warranty.
• Lightly lubricate the new Hatch O-ring with silicone grease and place it on the beveled outer edge of the Battery Compartment (Fig. 97a). Ensure that it is evenly seated.
• Place a new 3 volt, CR2, .75 AHr, Lithium Battery
a
(Duracell model DL-CR2 or equivalent) negative end first into the Battery Compartment with the negative end toward the spring (Fig. 98a).
Battery Hatch Installation
X
X
X
X
• Ensure that the Battery is properly oriented and the Hatch O-ring is evenly seated.
• Carefully place the Battery Hatch into position so that it seats on top of the O-ring, and while holding it in place, secure it with the 4 screws by turning them clockwise. DO NOT attempt to use any other screws.
• Carefully tighten the screws by alternately turning them each one turn at a time (Fig. 99). Turn the upper left one (a), then the lower right (b), then the lower left (c), then the upper right (d).
• Repeat the sequence until all of the screws are evenly secure. The outer surface of the Battery Hatch should be flush with the outer surface of the Housing. DO NOT overtighten.
Inspection
• Activate the unit and watch carefully as it performs a full diagnostic and battery check, and enters Surface Mode. Observe the LCD display to ensure it is consistently clear and sharp in contrast throughout the screen.
• If any portions of the display are missing or appear dim, or if a Low Battery Condition is indicated, return your Pro Plus 2 to an Authorized Oceanic Dealer for a complete evaluation before attempting to use it.
da
c
Fig. 99 - Hatch Installation
b
89
SPECIFICATIONS
CAN BE USED AS
 Air Computer  Nitrox Computer  Digital Depth Gauge/Timer
NO DECOMPRESSION MODEL
Basis:
 Modified Haldanean Algorithm  12 tissue compartments
Data Base:
 Diving Science and Technology (DSAT) - Rogers/Powell
Performance:
 Tissue compartment halftimes (mins.) Spencers M values
5, 10, 20, 40, 80, 120, 160, 200, 240, 320, 400, 480  Reciprocal subsurface elimination  60 minute surface credit control for compartments faster
than 60 minutes  Tissue compartments tracked up to 24 hours after last dive
Decompression Capabilities (stop ceilings):
 10, 20, 30, 40, 50, and 60 feet
(3, 6, 9, 12, 15, and 18 meters)
Altitude Algorithm:
 Based on NOAA tables
Oxygen Exposure Limits:
 Based on NOAA tables
90
OPERATIONAL MODES (SURFACE)
 Activation/Diagnostic  Surface  Pre Dive Planning Sequence
(30 - 190 feet / 9 - 57 meters)  Time to Fly Countdown  Desaturation Countdown  Dive Log (Date / Time, Nitrogen, and Oxygen)  Set Mode 1:
 FO2 (Air, 21 to 50% O2)
 Max Depth Alarm (30 to 300 feet /9 to 99 meters)
 Elapsed Dive Time Alarm (0:10 - 3:00 hr:min)
 PC Interface (to Download data)  Set Mode 2:
 Units of Measure (Imperial / Metric)
 Hour Format (12 / 24)
 Time (Hour, Minute)
 Date (Year, Month, Day)
 Audible Alarm / LED Warning (On/Off)
 Max TLBG Alarm (1 to 12 segments)
 Air Time Remaining BG Alarm (0 to 60 minutes)
 Turn Pressure Alarm (1000 to 3000 PSI / 69 to 205 BAR)
 End Pressure Alarm (300 to 1500 PSI / 20 to 104 BAR)
 Dive Time Remaining Alarm (0:00 - 0:20 min)
 Max PO2 Alarm (1.20 - 1.60 ATA)
 FO2 50% Default (On/Off)
 Backlight Duration (:00 / :03 / :07 seconds)
 Sampling Rate - memory storage for download
(2, 15, 30, 60 seconds; 2, 5, 10 feet, .5, 1.5, 3 meters)
 Digital Gauge Mode (On / Off)
 Water Activation (On / Off)  Simulator Mode
SPECIFICATIONS (CONTINUED)
OPERATIONAL MODES (DIVE)
No Decompression Dive:  #1 - Current Depth, Dive Time Remaining, Cylinder Pressure, Bar Graphs  #2 - Current Depth, Dive Time Remaining, Max Depth, Elapsed Dive Time, Cylinder Pressure, Bar Graphs  #3 - Current Depth, Dive Time Remaining, Temperature, Time of Day, Cylinder Pressure, Bar Graphs  #4 - if a nitrox dive - Current Depth, Dive Time Remaining, Current PO2, Cylinder Pressure, Bar Graphs  Secondary - Current Depth, Air Time Remaining, Cylinder Pressure, Bar Graphs
 Safety Stop - Current Depth, Stop Depth (15 FT/5 M), Countdown Timer (3:00 to 0:00 min:sec), Dive Time Remaining,
Cylinder Pressure, Bar Graphs
Digital Gauge Mode:  Default - Current Depth, Max Depth, Time of Day, Elapsed Dive Time, Cylinder Pressure, Bar Graphs  Secondary - Current Depth, Air Time Remaining, Cylinder Pressure, Bar Graphs
Decompression Dive:  #1 - Default - Current Depth, Stop Depth / Time, Total Ascent Time, Cylinder Pressure, Bar Graphs  #2 - Current Depth, Stop Depth / Time, Total Ascent Time, Max Depth, Elapsed Dive Time, Cylinder Pressure,
Bar Graphs  #3 - Current Depth, Temperature, Time of Day, Total Ascent Time, Cylinder Pressure, Bar Graphs  #4 - if nitrox dive - Current Depth, Current PO2, Total Ascent Time, Cylinder Pressure, Bar Graphs
Violation - Conditional, Delayed, and Immediate/Gauge
High PO2 (1.20 to 1.60 ATA)
High Oxygen Accumulation (300 OTU per dive / 24 hr)
91
SPECIFICATIONS (CONTINUED)
DISPLAY RANGE/RESOLUTION
Numeric Displays: Range: Resolution:
 Dive Number 0 to 24 1  Depth 0 to 330 ft (0 to 99.9 m) 1 ft (.1 m )  Maximum Depth 330 ft (99.9 m) 1 ft (.1 m)  FO2 Set Point Air, 21 to 50 % 1 %  PO2 Value 0.00 to 5.00 ATA .01 ATA  Dive Time Remaining 0:00 to 9:59 hr:min 1 minute  Air Time Remaining 0:00 to 9:59 hr:min 1 minute  Total Ascent Time 0:00 to 9:59 hr:min 1 minute  Safety Stop Time 3:00 to 0:00 min:sec 1 second  Decompression Stop Time 0:00 to 9:59 hr:min 1 minute  Elapsed Dive Time 0:00 - 9:59 hr:min 1 minute  Surface Time 0:00 - 23:59 hr:min 1 minute  Dive Log Surface Interval 0:00 - 23:59 hr:min 1 minute  Time to Fly 23:50 - 0:00 hr:min* 1 minute
 Time to Desaturate 23:50 - 0:00 hr:min* 1 minute
 Temperature 0 to 99°F (-9 to 60°C) 1°  Cylinder Pressure 0 to 5000 psi (0 to 352 BAR) 10 psi (.5 BAR)
Special Displays: Occurrence  Diagnostic Display After Manual Activation  Out of Range (- - -) >330 feet (>99.9 meters)  Gauge Mode Countdown Timer 23:50 to 0:00 hr:min (after violation)
(* starting 10 min after the dive)
(* starting 10 min. after the dive)
92
SPECIFICATIONS (CONTINUED)
BAR GRAPHS
Tissue Loading Bar Graph: segments Oxygen (O2) Bar Graph: segments
 No Decompression zone (green) 9  Normal zone (green) 4  No Deco Caution zone (yellow) 2  Caution zone (yellow) 1  Decompression Warning zone (red) 1  Danger zone (red) 1
Variable Ascent Rate Indicator: 60 feet (18 m) & Shallower Deeper than 60 feet (18 m)
segments feet/min meters/min segments feet/min meters/min
 Normal Zone (Green) 1 11 - 15 3.5 - 4.5 1 21 - 30 6.5 - 9  Normal Zone (Green) 2 16 - 20 5 - 6 2 31 - 40 9.5 - 12  Normal Zone (Green) 3 21 - 25 6.5 - 7.5 3 41 - 50 12.5 - 15  Caution Zone (Yellow) 4 26 - 30 8 - 9 4 51 - 60 15.5 - 18  Too Fast Zone (Red - flashing) 5 > 30 > 9 5 > 60 > 18
OPERATIONAL PERFORMANCE
Function: Accuracy:
 Depth ±1% of full scale  Timers 1 second per day
Dive Counter:
 Displays Dives #1 to 24, 0 if no dive made yet  Resets to Dive #1, upon diving (new activation period)
Dive Log Mode:
 Stores 24 most recent dives in memory for viewing  After 24 dives, adds 25th dive in memory and deletes the first dive
0 0 - 1 0 0 - 3 0 0 - 20 0 - 6
93
SPECIFICATIONS (CONTINUED)
OPERATIONAL PERFORMANCE (continued)
Altitude:
 Operational from sea level to 14,000 feet (4,267 meters) elevation  Measures ambient pressure every 30 minutes and when manually activated (no when wet contacts are bridged)  Compensates for Altitude when manually activated (no compensation if activated by immersion in water)  Compensation begins at 2,000 feet (610 meters) elevation and every 1,000 feet (305 meters) higher
Power:
 Battery 1 - 3 vdc, CR2, .75 Ahr, Lithium battery (Duracell model DL-CR2 or equivalent)  Shelf life Up to 5 years  Replacement User replaceable (annual recommended)  Life expectancy 50 dive hours (if 1 - 1 hour dive per dive day) to over 150 dive hours (if 3 - 1 hour dives per dive day)
Battery Indicator: segments displayed estimated power remaining
Activation:
 Manual - push button (recommended)  Automatic - by immersion in water (as a backup if set ON)  H2O graphic indicates Wet Contacts are bridged (unit must be dried prior to transport or storage)  Cannot be manually activated deeper than 4 feet (1.2 meters), if the Water Activation feature is set OFF.  Cannot be activated at elevations higher than 14,000 feet (4,267 meters)
Shutoff:
 Automatically shuts off if no dive is made within 120 minutes after initial activation. Reactivation required.  Automatically shuts off 24 hours after last dive (will reactivate if the H2O graphic is displayed).  Cannot be shut off manually.
all 25 to 100% 1 (inside) < 25%
94
SPECIFICATIONS (CONTINUED)
OPERATIONAL PERFORMANCE (continued)
Setting FO2:
 Automatically set for 'Air' upon activation  Remains set for Air unless an FO2 numerical value is set  Nitrox set points from 21 to 50 %  If set for 21%, remains set for 21% until changed  If set for >21%, it reverts to 50% 10 minutes after the dive, if the FO2 Default is ON. If the FO2 Default is OFF, the value will
remain at the value set.
Operating Temperature:
 Out of the water - between 20 °F and 140 °F (-6 and 60 °C)  In the water - between 28 °F and 95 °F (-2 and 60 °C).
At extremely low temperatures, the LCD may become sluggish, but this will not affect it's accuracy. If stored or transported in extremely low temperature areas (below freezing), you should warm the unit and its battery with body heat before diving.
ACCESSORIES (optional items available from your Authorized Oceanic Dealer):  Lens Protector - adheres to lens face, prevents scratches  PC download package (hardware and software)  Battery Kit - includes 1 Battery, 1 Battery Hatch O-ring, Silicone Grease
95
OCEANIC WORLD WIDE
OCEANIC USA
2002 Davis Street
San Leandro, CA 94577
Tel: 510/562-0500
Fax: 510/569-5404
Web site: http://www.OceanicWorldWide.com
Oceanic Germany - Nurnberg, Germany
Tel: 49-911-324-6630 Fax: 49-911-312-999
E-mail: office@oceanic.de
Oceanic South Europe - Genova, Italy
Tel: 0039-010-834-51 Fax: 0039-010-834-52-50
E-mail: Shawne.Stanley@oceanicse.it
Oceanic SW, Ltd - Devon, United Kingdom
Tel: 44-1-404-89-1819 Fax: 44-1-404-89-1909
E-mail: info@oceanicuk.com
Oceanic France - Marseille, France
Tel: 33-491-25-27-45 Fax: 33-491-25-35-86
E-mail: oceanicfrance@wanadoo.fr
Oceanic International (Pacific) - Kapolei, Hawaii
Tel: 808-682-5488 Fax: 808-682-1068
E-mail: oceanichi@oceanicusa.com
96
Oceanic Diving Australia Pty. Ltd
Sorrento, Victoria, Australia
Tel: 61-3-5984-4770 Fax: 61-3-5984-4307
E-mail: sales@oceanicaus.com.au
Oceanic Asia-Pacific Pte. Ltd - Singapore
Tel: 65-779-3853 Fax: 65-779-3945
E-mail: info@oceanicasia.com.sg
Oceanic Japan - Yokohama, Japan
Tel: 045-575-6671 Fax: 045-575-6673
E-mail: oceanic@gol.com
Oceanic New Zealand
Wellington, New Zealand
Tel: 64-4-472-5335 Fax: 64-4-472-5334
INSPECTION / SERVICE RECORD
Serial Number ________________________
Date of purchase _____________________
Purchased from ______________________
R
E
S
P
V
I
O
D
N
E
S
L
I
B
Below to be filled in by an Authorized Oceanic Dealer:
Date Service Performed Dealer / Technician
R
E
97
98
NOTES
RESET FEATURE
This dive computer is configured with a RESET feature that allows data to be cleared, including Nitrogen and Oxygen calculations, FO2 set point, Log Mode entries, and Download data.
WARNING: Reset after a dive and subsequent use for a repetitive dive conducted by the same diver could result in serious injury to or death.
RESET PROCEDURE
• While in SURFACE MODE (new activation period or greater than a 10 minute post dive surface interval), press the Left (Advance) button one (1) time to access PLAN MODE.
• While 30 feet (or 9 meters) is displayed in the PLAN MODE, press and hold Both buttons until SET 2 appears, then release the buttons.
• Press and release the Left (Advance) button to display the first 2 digits of the assigned KEY CODE, flashing as 00.
• Press and release the Right (Select) button to increase the digits to 01.
• Press and release the Left (Advance) button again to display the second 2 digits of the KEY CODE, flashing as 00.
• Press and release the Right (Select) button to increase the digits to 01.
• Press and release the Left (Advance) button to complete the RESET operation and turn the unit OFF.
99
OCEANIC® USA
2002 Davis Street
San Leandro, CA 94577
Tel: 510-562-0500
http://www.OceanicWorldWide.com
© 2002 Design 2002 Doc. No. 12-2328-r03 (6/6/02)
Fax: 510-569-5404
Loading...