Oceanic DataTrans Plus Owner's Manual

Page 1
DataTrans Plus
owner's guide
Page 2
Page 3
DataTrans Plus
Pay special attention to items marked with this Warning symbol.
WARNINGS:
• The DataTrans Plus is intended for use by recreational divers who have successfully completed a nationally recognized course in scuba diving, and diving with enriched nitrogen-oxygen (nitrox) mixtures.
• It must not be used by untrained persons who may not have knowledge of the potential risks and hazards of scuba diving, and diving with enriched nitrogen-oxygen (nitrox) mixtures.
• You must obtain scuba certification, and certification in diving with enriched nitrogen-oxygen mixtures (nitrox) before using the DataTrans Plus if you have not already done so.
• It is NOT for use by military and commercial divers.
• It should NOT be utilized for any competitive, or repetitive square wave or decompression diving, as it is intended solely for recreational use and no decompression multilevel diving.
• As with all underwater life support equipment, improper use or misuse of this product can cause serious injury or death.
• Never participate in sharing or swapping of a dive computer.
• Conduct your dives in such a manner so as to insure that you continuously check the computer's proper function.
• Read and understand this owner’s guide completely before diving with the DataTrans Plus.
• If you do not fully understand how to use this dive computer, or if you have any questions, you should seek instruction in its use from your authorized Oceanic dealer before you utilize this product.
iii
Page 4
®
LIMITED TWO-YEAR WARRANTY
Oceanic guarantees, to the original purchaser only, that the DataTrans Plus will be free of defects in materials and/or craftsmanship under normal recreational multilevel scuba use for two years from date of purchase, provided proper care and annual service are per­formed as described within this owner’s guide. Should your DataTrans Plus prove to be defective for any reason (other than those listed in the limitations section below) it will be repaired or replaced (at Oceanic’s discretion) free of charge excluding shipping and handling charges.
This warranty will be considered void if the DataTrans Plus was purchased from anyone other than an Authorized Oceanic Dealer, and/or if the registration card is not filled out completely at the time of purchase and mailed to Oceanic within 30 days
of purchase, and/or if the annual inspection is not done according to this owner’s guide. This warranty is non-transferrable and applies to the original purchaser only. All correspondence concerning this warranty must be accompanied by a copy of the original sales receipt and a copy of the owner’s portion of the warranty registration card including the annual inspection record.
Once each year you must return the DataTrans Plus to an Authorized Oceanic Dealer within 30 days of the original purchase date anniversary to keep the two year limited warranty in force. Annual inspection includes verification of depth
accuracy and proper general function. Labor charges for the annual inspection are not covered by the warranty. You must provide a copy of the original sales receipt and a copy of the owner’s portion of the warranty registration card including the annual service record to obtain warranty service.
Statement of Limitations - General: Warranty does not cover damage from accident, abuse, battery leakage, tampering, lack of proper care and maintenance and/or proper annual servicing, or improper use of the DataTrans Plus. Modifications or repair by anyone other than an
Oceanic Sales & Service Center authorized to service the DataTrans Plus will void the warranty. Oceanic will not be responsible for recovery or replacement of the product in the event of loss or theft. Oceanic, its distributors, and retailers make no warranties, either expressed or implied, with respect to this product or its owner’s guide except those stated in the preceding paragraphs. In consideration
of the sale of the DataTrans Plus to you, you agree and understand that in no event will Oceanic, its distributors or retailers, be held liable for any personal injuries resulting from its operation, or for any other damages whether direct, indirect, incidental, or consequential even if Oceanic is advised of such damages. Warranty does not extend to plastic gauge face, o-rings, batteries, or damage due to accident, abuse, modification, or tampering.
Some states do not allow the exclusion or limitation of implied warranties or liabilities for incidental or consequential damages, so the above limitation may not apply to you.
iv
Page 5
DataTrans Plus
COPYRIGHT NOTICE
This owner’s guide is copyrighted, all rights are reserved. It may not, in whole or in part, be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine readable form without prior consent in writing from Oceanic / 2002 Design. The Surface Time/Mode, Plan Mode, Breathing Gas Time, No Decompression Time, Decompression Stop Time, Dive Log, Transmitter Low Battery, Receiver Low Battery, Link, Temperature, Elapsed Dive Time, Maximum Depth, Ascent Rate, Breathing Gas Consumption, and Caution Zone icons are protected by copyright, and are trademarks of Oceanic.
DataTrans Plus Owner's Guide, Doc. No. 12-2022
© 2002 Design 1997
2002 Davis Street
San Leandro, Ca. USA 94577
510/562-0500
TRADEMARK NOTICE
Oceanic, the Oceanic logo, DataTrans Plus, the DataTrans Plus logo, DX3 Integrated, Oceanglo, Graphic Diver Interface, Tissue Loading Bar Graph, Pre Dive Planning Sequence, Variable Ascent Rate Indicator, Breathing Gas Time Remaining, Message Box, Set Point, Control Console, and OceanLog are all registered and unregistered trademarks of Oceanic. All rights are reserved.
PATENT NOTICE
U.S. Patents have been issued, or applied for, to protect the following design features: Graphic Diver Interface, Pre Dive Planning Sequence, Dive Time Remaining, Breathing Gas Time Remaining Bar Graph, Variable Breathing Gas Consumption Bar Graph, DataTrans Plus Message Box, DataTrans Plus Mode Menu Structure, Breathing Gas Alarm Set Point, Depth Alarm Set Point, Breathing Gas Time Remaining (U.S. Patent no. 4,586,136), Data Sensing and Processing Device (U.S. Patent no. 4,882,678), Tissue Loading Bar Graph (U.S. Patent no. 4,882,687), and Variable Ascent Rate Indicator Bar Graph (U.S. Patent no. 5,156,055).
DECOMPRESSION MODEL
The programs within the DataTrans Plus simulate the absorption of nitrogen into the body by using a mathematical model. This model is merely a way to apply a limited set of data to a large range of experiences. The DataTrans Plus dive computer model is based upon the latest research and experiments in decompression theory. Still, using the DataTrans Plus, just as using the U.S. Navy (or other) No Decompression Tables, is no guarantee of avoiding decompression sickness, i.e. “the bends.” Every diver’s physiology is different, and can even vary from day to day. No machine can predict how your body will react to a particular dive profile.
v
Page 6
®
FCC ID: MH8A
FCC Compliance:
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: 1) this device may not cause harmful interference, and 2) this device must accept any interference received, including interference that may cause undesired operation.
FCC Interference Statement:
This equipment has been tested and found to comply with the limits for an Intentional Radiator, a Class B Digital Device, pursuant to Part 15 of FCC Rules, Title 47 of the Code of Federal Regulations. These limits are designed to provide reasonable protection against harmful interference in a commercial or residential installation. This equipment generates, uses, and can radiate radio frequency energy, and if not installed and used in accordance with the instructions, may cause interference to radio communications.
There is no guarantee that interference will not occur in a particular installation. If this equipment does cause interference to radio or television reception, which can be determined by turning the equipment on and off, the user is encouraged to try to correct the interfer­ence by one or more of the following measures:
a. Reorient or relocate the receiving antenna. b. Increase the separation between the equipment and the affected receiver. c. Connect the equipment and the affected receiver to power outlets on separate circuits. d. Consult the dealer or an experienced radio/TV technician for help.
WARNING: Changes or modifications not expressly approved by Oceanic could void the user's authority to operate the equipment.
vi
Page 7
DataTrans Plus
CONTENTS
FEATURES & DISPLAYS............................................................ 1
Introduction ..................................................................................................................2
Interactive Control Console .........................................................................................4
Informational Displays.................................................................................................5
Universal Graphic Diver Interface ..............................................................................6
Tissue Loading Bar Graph .......................................................................................7
Oxygen Accumlation Bar Graph..............................................................................8
Variable Ascent Rate Indicator ................................................................................9
Breathing Gas Consumption Indicator .................................................................10
Breathing Gas Time Remaining Bar Graph .........................................................11
Dive Time Remaining.................................................................................................11
Breathing Gas Time Remaining ............................................................................12
No Decompression Dive Time Remaining ............................................................14
Oxygen Accumulation Time Remaining ...............................................................15
Alpha Numeric Displays.............................................................................................15
Cylinder Pressure Display......................................................................................15
Depth Displays........................................................................................................16
Time Displays .........................................................................................................17
Ambient Temperature Display ..................................................................................18
vii
Page 8
®
CONTENTS (continued)
Audible Alarm.............................................................................................................19
Message Box................................................................................................................21
Backlight Feature .......................................................................................................21
Operating Temperature .............................................................................................22
Sharing the DataTrans Plus ......................................................................................23
Graphic Interface Legend ..........................................................................................24
ACTIVATION & SETUP............................................................. 25
Making the DataTrans Plus Personal.......................................................................26
Activating the Display ................................................................................................27
Surface Mode ..............................................................................................................31
Mode Menu System ....................................................................................................31
Entering Settings .......................................................................................................33
Set Time ..................................................................................................................33
Set Date...................................................................................................................34
Set Alternate Display .............................................................................................35
Set Units of Measure ..............................................................................................36
Breathing Gas & Depth Alarm Set Points ............................................................37
Depth Alarm Set Point.......................................................................................37
viii
Page 9
DataTrans Plus
CONTENTS (continued)
Breathing Gas Alarm Set Point .........................................................................38
Set Depth & Breathing Gas Alarms ..................................................................39
Turning Off the Audible Alarm .........................................................................40
Linking Procedure..................................................................................................41
Set Language ..........................................................................................................43
Language Correction Procedure ........................................................................44
PREDIVE & DIVE MODES ....................................................... 45
Positioning of the Display Module.............................................................................47
Link Interruption Underwater ..................................................................................47
Operational Modes......................................................................................................48
FO2 Mode....................................................................................................................49
FO2 Set for Air .......................................................................................................49
Settings FO2 for Nitrox .........................................................................................50
Plan Mode ...................................................................................................................52
No Decompression Dive Mode ...................................................................................55
Decompression Dive Mode .........................................................................................56
Violation Modes ..........................................................................................................57
Gauge Mode ................................................................................................................57
ix
Page 10
®
CONTENTS (continued)
Ascending to the Surface............................................................................................57
Altitude Diving ...........................................................................................................59
POST DIVE MODES .................................................................. 61
Post Dive Surface Mode..............................................................................................62
Transition Period....................................................................................................62
FO2 Mode....................................................................................................................63
Plan Mode ...................................................................................................................63
Time to Fly..................................................................................................................64
Dive Log Mode ............................................................................................................65
History Mode ..............................................................................................................68
External Access Mode.................................................................................................69
HANDLING THE EXTREMES.................................................. 71
Emergency Decompression ........................................................................................72
Gas Time Remaining During Decompression.......................................................75
Caution Zone (C.Z.).................................................................................................76
Decompression Dive Mode .....................................................................................77
x
Page 11
DataTrans Plus
CONTENTS (continued)
Violation Modes ..........................................................................................................79
Conditional Violation Mode ...................................................................................79
Delayed Violation Mode..........................................................................................80
Immediate Violation Mode.....................................................................................83
Gauge Mode ............................................................................................................84
Permanent Violation ..............................................................................................84
Exceeding Maximum Operating Depth.....................................................................85
Oxygen Exposure........................................................................................................86
Partial Pressure of Oxygen ....................................................................................87
High PO2 Dive Mode..............................................................................................87
Oxygen Accumulation ............................................................................................88
High Oxygen Accumulation ...................................................................................89
Message Box Warnings ...............................................................................................90
Unexpected Loss of Displayed Information ..............................................................90
CARE & MAINTENANCE.......................................................... 93
Care and Cleaning ......................................................................................................94
Annual Inspections & Service....................................................................................95
How to Obtain Service ...........................................................................................96
xi
Page 12
®
CONTENTS (continued)
Battery Life.................................................................................................................97
Low Battery Condition ...............................................................................................99
Battery Replacement ..............................................................................................99
Flooded Battery Compartment ............................................................................101
Transmitter Installation Instructions.....................................................................102
Transmitter Compatibilty with Nitrox ...................................................................103
REFERENCE ............................................................................ 105
More About Flying After Diving..............................................................................106
More About Altitude Diving.....................................................................................107
More About Nitrox Diving .......................................................................................109
Multiple Tissue Tracking .........................................................................................110
No Decompression Limits ........................................................................................112
Repetitive Decompression Diving............................................................................112
Specifications ............................................................................................................114
Responsible Computer Diving .................................................................................117
Language Cross Reference .......................................................................................118
Glossary.....................................................................................................................121
Accessories ................................................................................................................125
DataTrans Plus Service Record ...............................................................................126
xii
Page 13
DataTrans Plus
FEATURES
and
DISPLAYS
1
Page 14
®
INTRODUCTION
Welcome to Oceanic and thank you for choosing DataTrans Plus!
Your new DataTrans Plus is a two component, hoseless, integrated system that consists of a computer Display Module and a radio frequency Transmitter that will be installed into a high pressure port of your regulator first stage, or is built into an Oceanic DX3™ Integrated first stage regulator. The transmitter sends tank pres­sure data to the display module via a low frequency signal. In addition to nitrogen and oxygen loading data, breathing gas consumption is calculated and displayed in graphic and alpha/numerical formats on the computer’s screen. The Display Module can also be used without the transmitter as a stand alone, non-inte­grated computer, and will retain full use of all functions except those that related to tank pressure.
Your DataTrans Plus presents the information that you need before, during, and after your air (or nitrox) dives using Oceanic's intuitive combination of easy to read displays and unique identification icons. Tissue loading of nitrogen, accumulation of oxygen, ascent rate, breathing gas consumption rate, and breathing gas time remaining are presented as segmented bar graphs alongside color coded reference indicators that bring quick focus to these important status displays.
As you progress through this instructional guide, you will become familiar with all of the unique functions and features available and see examples of the displays that you could expect to see in the various operational modes. Refer to the Glossary of terms on page 121, and keep the waterproof Review Card handy during your dive trips. Although it will require an initial investment of time to become acquainted with the various icons
2
Page 15
DataTrans Plus
and bar graphs of the Graphic Diver Interface, you’ll soon agree that the DataTrans Plus is easy to under­stand and use.
The DataTrans Plus has a wide array of features described in detail throughout the pages that follow. Due to the importance that they be understood thoroughly prior to using the DataTrans Plus, information will be ex­panded upon and some refreshed as you proceed. Relax and read through the complete guide.
It is extremely important that you:
• Read this owner's guide in sequence and understand it completely before attempting to use the DataTrans Plus.
• Check the DataTrans Plus frequently during your dive.
• You must also be a trained diver, certified by a recognized training agency in Scuba diving.
• Prior to using the oxygen related features of the DataTrans Plus, you must also be trained and certified for diving with enriched nitrogen-oxygen (nitrox) mixtures by a recognized training agency.
Remember that the rules you learned in your basic scuba certification course still apply to the diving you will do while using a dive computer - some will become even more important. Technology is no substitute for com­mon sense, and a dive computer only provides the person using it with data, not the knowledge to use it.
Be a RESPONSIBLE DIVER at all times.
3
Page 16
®
INTERACTIVE CONTROL CONSOLE
(Surface Modes)
Oceanglo
Backlight
(Dive Modes)
Fig. 1 - Interactive
Control Console
4
®
®
SelectAdvance
Alternate
Display
The DataTrans Plus is a unique dive computer with interactive controls that allow you to select various display options and access specific information when you choose to see it. The Interactive Control Console consists of the Ad- vance (Left) button and Select (Right) button (Fig. 1).
The control buttons can be pressed repeatedly, releasing upon hearing a beep, or held in to scroll and continue as you set or access different display modes.
On the surface, before diving, you can perform the following operations using the Control Console:
• Activate the display module
• Select units of measure - English or Metric
• Select Message Box™ language - English, Italian, German, Spanish, or French
• Set depth & gas alarm warning level Set Points™
• Turn the audible alarm - on or off
• Turn the Alternate Dive Mode display - on or off
• Set the current date & time
• Set the display module/transmitter link code
Page 17
Also while on the surface, you can access the following modes with the Control Console:
FO2 mode - to program the percentage of oxygen in the nitrox mix.
Plan mode - to view no decompression limits and plan your next dive.
Log mode - to view data from your 12 most recent dives.
History mode - to view the total number of dives, maximum depth, etc.
Set mode - to establish prefered selections.
External Access mode - to download (copy) dive data from the DataTrans Plus to a unique PC log/profile program.
During the Dive mode, the Advance (Left) button can be used to activate the display's Oceanglo™ backlight, and the Select (Right) button can be used to ac­cess an Alternate Dive mode, that displays additional information including maximum depth, bottom time and temperature.
INFORMATIONAL DISPLAYS
DataTrans Plus
NO DECO
C.Z. C.Z.
DECO
O
2
Operational modes and status information is visually represented numerically and/or graphically and can be understood at a glance with the aide of universal icons (Fig. 2) that identify and bring quick focus to the displays. Also, segmented bar graphs will show how close you are to critical limits.
In critical situations, a Message Box™ flashes urgent messages, while an au-
5
10
30
0
20
®
60
40
50
Fig. 2 - Universal Icons
5
Page 18
®
C.Z. C.Z.
NO DECO
0
DECO
5
10
40
30
20
®
50
Fig. 3 - Message Box
C.Z. C.Z.
NO DECO
DECO
PSI BAR
MAX FT M
O
2
simple messages (Fig. 3), such as “TOO FAST” or “TOO DEEP" are displayed in the language that you choose during setup.
Each DataTrans Plus numeric and graphic display represents a unique piece of information. It is imperative that you understand the formats, ranges, and values of the information represented to
dible alarm sounds to alert you to check this information. These concise,
60
avoid any possible misunderstanding that could result in error.
NOTE: Throughout this owner's guide reference is made to the term 'breathing gas'. The rational being that the DataTrans Plus can be used for 'air' dives or 'nitrox' dives. For clarity these terms are defined as -
O
2
Breathing Gas - the gaseous mixture breathed during a dive. Air - a breathing gas that contains approximately 21% oxygen and 79% nitrogen (nature's common nitrogen-oxygen mixture). Nitrox - a nitrogen-oxygen breathing gas that contains a higher fraction of oxygen (22 to 50%) than air.
5
10
30
0
20
®
60
40
50
Fig. 4 - Bar Graph Displays
6
UNIVERSAL GRAPHIC DIVER INTERFACE
Five bar graphs referred to as the Universal Graphic Diver Interface™ appear around the perimeter of the DataTrans Plus screen (Fig. 4). These segmented
R2
Page 19
bar graphs are located next to green, yellow, and red color coded portions of the peripheral decal that denote normal, caution, and danger zones, respectively. When underwater, you can quickly focus on the bar graphs to make sure that they are in the green. You can quickly verify that you're not getting too close to the no decompression limit or the oxygen tolerance limit, or ascending too fast, or consuming breathing gas too fast, or running low on breathing gas.
DataTrans Plus
Tissue Loading Bar Graph
The Tissue Loading Bar Graph™ represents nitrogen loading (Fig. 5), showing your relative no decompression or decompression status. As your depth and elapsed dive time increase, segments will add to the graph beginning in the lower left portion of the screen. As you ascend to shallower depths, this bar graph will begin to recede, indicating that additional no decompression time is allowed for multilevel diving.
The Tissue Loading Bar Graph monitors 12 different nitrogen compartments simultaneously and displays the one that is in control of your dive. It is divided into a green No Decompression zone (NO DECO), a yellow Caution zone (C.Z.), and a red Decompression zone (DECO). The bar graph gives a visual represen­tation of just how close you are to the no decompression limit with a yellow Caution (C.Z.) Zone.
This Caution Zone portion of the bar graph (Fig. 5a) allows you to make a deci­sion regarding safety stop duration or necessity. While you cannot provide a
a
NO DECO
C.Z. C.Z.
DECO
O
2
b
5
10
30
0
20
®
60
40
50
Fig. 5 - Tissue Loading
Bar Graph
7
Page 20
NO DECO
®
C.Z. C.Z.
DECO
guarantee against the occurrence of decompression sickness, you may choose
O
2
your own personal zone of caution based upon age, physique, excessive weight, etc., to reduce the statistical risk.
FT
The Tissue Loading Bar Graph assists you with managing decompression by filling a large red 'ceiling stop required' segment (Fig. 5b). Decompression is explained in detail in the Handling the Extremes section.
5
10
30
0
20
®
60
40
50
Fig. 6 - Nitrogen Control
NO DECO
C.Z. C.Z.
DECO
O
2
a
5
10
30
0
20
®
60
40
50
Fig. 7 - Oxygen Accumulation
(O2) Bar Graph
8
Prior to a repetitive nitrox dive, if all segments of the Tissue Loading Bar Graph are displayed during the Pre Dive Planning Sequence™, and no seg­ments of the O2 bar graph are displayed (Fig. 6), that next dive is calculated to be controlled by nitrogen loading.
WARNING: Oceanic advocates responsible diving practices consistent with your individual level of formal training and ex­perience, and does not recommend decompression diving or diving below 130 feet (39 m).
Oxygen Accumulation (O2) Bar Graph
NOTE: Displays associated with oxygen and the O2 bar graph will only appear if FO2 has been set at a value other than 'Air'.
The Oxygen Accumulation (O2) Bar Graph (Fig. 7) represents oxygen loading,
Page 21
your relative oxygen tolerance dosage (OTU), showing the maximum of either per dive allowable oxygen, or 24 hour period allowable oxygen. As your expo­sure (accumulation of oxygen) increases during the dive, segments will add to the graph around the upper right perimeter of the screen. As accumulation decreases, the bar graph will begin to recede, indicating that additional expo­sure (accumulation) is allowed for that dive, and that 24 hour period.
DataTrans Plus
NO DECO
C.Z. C.Z.
DECO
O
2
FT
The O2 Bar Graph also assists you with managing high partial pressure of oxy­gen (PO2) by flashing the large red Danger zone segment (Fig. 7a) as a warn­ing when the level of PO2 exceeds the maximum allowed limit of 1.60 BAR (ATA). This is explained in detail in the Handling the Extremes section.
Prior to a repetitive nitrox dive, if the segments of the O2 Bar Graph are dis­played during the Pre Dive Planning Sequence™ (PDPS), and no segments of the Tissue Loading Bar Graph are displayed (Fig. 8), that next dive is calcu­lated to be controlled by oxygen loading.
Variable Ascent Rate Indicator
The Variable Ascent Rate Indicator™ (VARI), located along the lower/right portion of the screen (Fig. 9), is provided to help you to avoid excessive ascent rates by providing a visual representation of ascent speed, rather than just showing that you are ascending too fast.
5
10
30
0
20
®
60
40
50
Fig. 8 - Oxygen Control
NO DECO
0
5
C.Z. C.Z.
DECO
10
30
20
®
O
2
60
40
50
Fig. 9 - Variable Ascent
Rate Indicator
9
Page 22
®
VARI Segments = Speed (rate) 0 = 0 - 10 fpm (0 - 3 mpm) 1 = 11 - 20 fpm (3 - 6 mpm) 2 = 21 - 30 fpm (6 - 9 mpm) 3 = 31 - 40 fpm (9 - 12 mpm) 4 = 41 - 50 fpm (12 - 15 mpm) 5 = 51 - 60 fpm (15 - 18 mpm) 6 = 61 - 90 fpm (18 - 27 mpm) 7 = 91 - 120 fpm (27 - 36 mpm) 8 = >120 fpm (>36 mpm) (when > 5, the segments flash)
NO DECO
0
5
C.Z. C.Z.
DECO
10
30
20
®
O
2
60
40
50
The 8 triangular segments of the bar graph, located beside green, yellow, and red reference zones, appear beginning from the bottom and may be considered an ascent rate speedometer. Green is a 'normal' rate, yellow is a 'caution' rate, and red is 'Too Fast'. The actual speeds that the VARI segments represent are shown at the left.
When your ascent rate exceeds the maximum recommended rate of 60 feet (18 meters) per minute, the bar graph segments will enter the red zone and all dis­played segments will flash once per second until your ascent speed is slowed. When this occurs, you should immediately slow your ascent.
The Variable Ascent Rate Indicator is a unique feature of Oceanic dive com­puters that has been granted U.S. Patent No. 5,156,055.
Breathing Gas Consumption Indicator
The Breathing Gas Consumption Indicator bar graph located at the lower/left portion of the display (Fig. 10) is a true biofeedback monitor that indicates your current breathing rate as compared to your personally established breathing parameters. The comparison is based upon an average rate estab­lished during the first 70 seconds of breathing that is sensed.
Fig. 10 - Breathing Gas Consumption Indicator
10
After the comparison, the bar graph will provide you with continuous visual indication of your breathing rate as it slows or increases. Use of Oceanic's pat-
Page 23
ented (U.S. Patent No. 4,586,136) breathing gas consumption calculation method makes information accurate even during sudden changes in depth.
Breathing Gas Time Remaining Bar Graph
The Breathing Gas Time Remaining Bar Graph located along the bottom of the display (Fig. 11) provides a graphic representation of the time that you can remain at your present depth and then, following a safe ascent, surface with a predetermined breathing gas reserve. This calculation and display is based on your breathing gas consumption rate that is continually monitored by the DataTrans Plus, and it takes into account the breathing gas required for a safe ascent including any required decompression stops.
DataTrans Plus
The green, yellow, and red zones adjacent to the bar graph enable you to quickly focus on remaining breathing gas times of 60 minutes or less, based on your pre selected Gas Alarm Set Point. The bar graph is more precise as time remaining decreases toward the red zone.
DIVE TIME REMAINING
One of the most important pieces of information on the DataTrans Plus is the patented Dive Time Remaining numeric display. To numerically display Dive Time Remaining, the DataTrans Plus constantly monitors three critical pieces of information; no decompression status, oxygen accumulation status, and rate
NO DECO
0
5
C.Z. C.Z.
DECO
10
30
20
®
O
2
60
40
50
Fig. 11 - Breathing Gas
Time Remaining Bar Graph
11
Page 24
®
(No Decompression Time)
(Breathing Gas Time)
(Oxygen Accumulation Time)
Fig. 12 - Dive Time
Remaining Identification
NO DECO
C.Z. C.Z.
DECO
PSI
O
2
FT
of breathing gas consumption. The Dive Time Remaining display will indicate the time that is more critical for you at that particular moment (i.e.; whichever time is the least amount available of the three). The time being displayed is identified by the No Decompression Dive Time icon, or Gas Time Remaining icon displayed to the left of the numeric display, or the O2 symbol displayed to the right of the numeric display (Fig. 12).
Knowing that you have 45 minutes of no decompression time remaining is not as critical as knowing that you only have 40 minutes of breathing gas time re­maining. Or, knowing that you have 40 minutes of breathing gas time remain­ing is not as critical as knowing that you only have 35 minutes of oxygen accu­mulation time remaining. The DataTrans Plus presents the dive time remain­ing that is considered to be of primary importance. This unique feature of se­lect Oceanic dive computers has been granted U.S. Patent No. 4,586,136.
Breathing Gas Time Remaining
5
10
30
0
20
®
Fig. 13 - Breathing Gas
Time Remaining
12
a
Breathing Gas Time Remaining will appear as the numeric Dive Time Remain­ing display (Fig. 13) only when it is less than No Decompression Time Remain­ing and Oxygen Accumulation Time Remaining. Breathing Gas Time Remain-
60
40
50
ing of 60 minutes, or less, will be displayed continuously as the Gas Time Re­maining Bar Graph (Fig. 13a) regardless of which time is being displayed as the numeric Dive Time Remaining. The bar graph represents only Breathing Gas Time information and it will be the only indication of breathing gas time
Page 25
remaining if you are in a decompression or violation mode. The DataTrans Plus calculates Breathing Gas Time Remaining using a pat-
ented algorithm that is based on a diver's individual breathing gas consump­tion rate and depth. Tank pressure is measured once each second, and an av­erage rate of consumption is calculated over a 90 second period. This rate of consumption is then used in conjunction with a knowledge of the depth depen­dence to predict the breathing gas required for a safe ascent including any re­quired decompression stops.
Breathing gas consumption and depth are continuously monitored, and Breathing Gas Time Remaining reflects any change in your circumstances. For example, when a buddy starts breathing from your octopus or you sud­denly find yourself swimming against a strong current and begin breathing more rapidly, the DataTrans Plus will recognize this change and adjust your Breathing Gas Time Remaining accordingly.
DataTrans Plus
Remember, the Breathing Gas Time Remaining is the time you can remain at the present depth and still surface with the tank pressure reserve (Gas Alarm Set Point) that you set during setup. When Gas Time Remaining indicates zero, you should immediately initiate a controlled ascent. However, there is no reason to panic, the DataTrans Plus has allowed for the breathing gas neces­sary for a safe ascent including any emergency decompression stops.
R
E
S
P
O
N
R
E
V
I
D
E
S
L
I
B
13
Page 26
®
No Decompression Dive Time Remaining
No Decompression Dive Time Remaining is the maximum amount of time that you can stay at your present depth before entering a decompression situation. It is calculated based on the amount of nitrogen absorbed by twelve hypotheti­cal tissue compartments. The rates each of these compartments absorb and release nitrogen is mathematically modeled and compared against a maximum allowable nitrogen level. Whichever one of the twelve is closest to this maxi­mum level is the controlling compartment for that depth. Its resulting value will be displayed numerically along with the No Decompression Dive icon and graphically as the Tissue Loading Bar Graph (Fig. 14).
NO DECO
0
5
C.Z. C.Z.
DECO
PSI
10
30
20
®
O
2
FT
60
40
50
Fig. 14 - No Decompression
Dive Time Remaining
14
As you ascend from depth following a dive that has approached the no decom­pression limit, the Tissue Loading Bar Graph will diminish as control shifts to slower compartments. This is a feature of the decompression model that is the basis for multilevel diving, one of the most important advantages the DataTrans Plus offers. See the Reference section for more information per­taining to tissue tracking.
The no decompression algorithm is based upon Haldane’s theory using maxi­mum allowable nitrogen levels developed by Merrill Spencer. Repetitive diving control is based upon experiments designed and conducted by Dr. Ray Rogers and Dr. Michael Powell in 1987. Diving Science and Technology® (DSAT), a corporate affiliate of PADI®, commissioned these experiments and now uses
Page 27
the findings in the Recreational Dive Planner™ distributed by PADI.
Oxygen Accumulation Time Remaining
Oxygen accumulation (exposure) during a dive, or 24 hour period, appears graphically as the O2 Bar Graph. As time remaining before reaching the oxy­gen exposure limit decreases, segments are added to the O2 bar graph. When the amount of time remaining before reaching the oxygen limit becomes less than the No Decompression Dive Time Remaining or Breathing Gas Time Re­maining, calculations for the dive will be controlled by oxygen. Oxygen Time Remaining will then appear as the main numeric time display (Fig. 15a) as sig­nified by the O2 symbol appearing to the right of the display.
As oxygen accumulation continues to increase, the O2 bar graph will enter the yellow Caution Zone. High O2 Mode is explained in detail in the Handling the Extremes section.
ALPHA/NUMERIC DISPLAYS
DataTrans Plus
DECO
a
O
2
PSI
FT
b
NO DECO
C.Z. C.Z.
R2
Cylinder Pressure Display
The Cylinder Pressure display, located in the upper/left portion of the screen (Fig. 15b), indicates how much breathing gas is in your cylinder, up to 5000 PSI (352 BAR) to nearest 10 PSI (.5 BAR).
5
10
30
0
20
®
60
40
50
Fig. 15 - Oxygen Accumula-
tion Time Remaining
15
Page 28
®
The value of pressure will be displayed during all dive modes when the Display Module is linked to the Transmitter and within the operating range. Linking is explained in more detail in the Pre Dive and Dive Mode section.
Depth Displays
During a dive, the Current Depth display, located in the upper/right portion of the screen (Fig. 16a), indicates depths from 0 to 330 feet (99.5 meters) in 1 foot (.5 meter) increments.
C.Z. C.Z.
NO DECO
0
DECO
PSI
5
10
40
30
20
®
Fig. 16 - Depth Displays
(Alternate Dive Mode)
16
a
O
2
FT
The value of current depth will be displayed during all dive modes unless you descend deeper than 330 feet (99.5 meters), at which point the display will show three dashes ( - - - ) to indicate that you have gone 'out of range'. This is described in detail in the Handling the Extremes section.
A second depth display located in the lower/right portion of the LCD (Fig. 16b) indicates the Maximum Depth reached during that dive. If the Alternate Mode is turned 'off', the display will appear when the Select/Right button is
60
50
pressed. More critical information such as a message will override it.
b
In the event that you descend deeper than 330 feet (99.5 meters), this display will only show three dashes ( - - - ) as the maximum depth for the remainder of that dive, and as the Max Depth in the Dive Log for that dive. This is de­scribed in detail in the Handling the Extremes section.
Page 29
During a Decompression Dive the required Ceiling Stop Depth appears in the lower/left portion of the screen (Fig. 17a). The display toggles with the message CEILING once every 15 seconds while in the Decompression Mode. This is described in detail in the Handling the Extremes section.
Time Displays
DataTrans Plus
NO DECO
C.Z. C.Z.
DECO
PSI
O
2
FT
R2
The Main Time display, located in the center of the screen (Fig. 18a), has larger digits than the other numerial displays. Depending on the operating mode that the DataTrans Plus is in at the time, the display indicates elapsed Surface Time, theoretical Dive Time Available, Dive Time Remaining, or Total Ascent Time required.
A second Time display appears in the lower portion of the screen. Depending on the operating mode that the DataTrans Plus is in at the time, the display indicates Elapsed Dive Time (Fig. 18b), Decompression Stop Time required, Time of Day, or Time to Fly.
Each display is described in detail in subsequent sections of this owner's guide. Most of the time displays are shown in hour:minute format (i.e. 1:02 repre-
sents one hour and two minutes, not 102 minutes!). The colon that separates hours and minutes blinks once per second when the display is indicating real time such as elapsed Surface Time. Dive Time Available, No Decompression
5
10
30
0
20
®
60
40
50
b
Fig. 17 - Depth Displays
(Decompression Dive Mode)
a
NO DECO
0
5
C.Z. C.Z.
DECO
PSI
10
30
20
®
O
2
FT
60
40
50
b
Fig. 18 - Time Displays
(Alternate Dive Mode)
17
Page 30
®
Dive Time Remaining, Total Ascent Time required, or Time to Fly are calcu­lated projections of time and use a solid (non-blinking) colon to indicate that they are counting down, rather than counting up.
Elapsed Dive Time, that appears in the lower/center portion of the display (Fig. 19a) when the Alternate Mode is turned 'on', is in minute format with a maximum display value of 99 minutes. If the Alternate Mode has been turned 'off', Elapsed Dive Time can be displayed by depressing the Select/Right but­ton.
Ambient Temperature Display
NO DECO
0
5
C.Z. C.Z.
DECO
PSI
10
30
20
®
b
O
2
FT
60
40
50
a
Fig. 19 - Alternate Dive Mode
18
When the Alternate Mode is turned 'on', ambient Temperature will appear in the lower/left portion of the display(Fig. 19b). If the Alternate Mode has been turned 'off', Temperature can be displayed by depressing the Select/Right but­ton.
NOTE: More critical information such as Ceiling Stop Depth required or a message will override the lower displays.
Responsible
Page 31
AUDIBLE ALARM
When you are approaching dangerous situations, the DataTrans Plus alerts you to check the Message Box, Graphic Diver Interface, and numeric displays. There are four Audible Alarms.
Potential Danger – One Double Beep
During situations that may pose potential danger, one Double Beep is emitted from the DataTrans Plus. These situations are as follows:
• Entry into decompression.
• Decreasing to 5 minutes of Breathing Gas Dive Time Remaining.
• Partial pressure of oxygen equal to or greater than 1.40 ATA.
Immediate Danger – continuous One Beep per Second
DataTrans Plus
BEEP!
BEEP!
(Potential Danger)
When the DataTrans Plus senses immediate danger to you, it emits One Beep per Second until one of the following situations is corrected:
• Descent deeper than the Depth Alarm Set Point.
• Continuous interruption of signal link of more than 60 seconds.
• Ascent to a depth shallower than a required stop depth.
BEEP...
BEEP...
BEEP...
(Immediate Danger)
19
Page 32
®
• Ascent rate that exceeds 60 ft./min (18 m/min).
• Gas Time Remaining equals required Decompression time.
• Gas Time Remaining equals zero (0:00).
• Partial pressure of oxygen equal to or greater than 1.60 ATA.
• Oxygen accumulation greater than the allowed per dive or 24 hour limit.
Permanent Violations – Single Long Beep
If you enter a Delayed or Immediate Violation Mode, a Single Long Beep will be emitted. This will occur if one of these Violation rules are broken:
20
BEEEEEEP!
(Violation)
BEEP!
(Transition)
• Depth is shallower than the required stop depth for more than 5 minutes.
• Required Decompression exceeds a 60 FT/ 18 M ceiling.
Transition – Short Beep
To indicate that a command has been accepted, the DataTrans Plus will emit a Short Beep whenever you use the control console, and immediately following activation and the Diagnostic Mode.
Page 33
MESSAGE BOX
The Message Box, located in the lower portion of the screen (Fig. 20), provides a visual explanation of what is occurring when the audible alarm sounds, working in conjunction with the Depth and Gas Alarm Set Points, as well as Decompression and Violation Modes, and the Variable Ascent Rate Indicator. Messages such as, “TOO FAST”, “GAS ALARM”, and “VIOLATION” flash on the display in the language that you set. This is described in detail in the Han­dling the Extremes section. Also, a language cross reference is provided on page 118 for your convenience.
BACKLIGHT FEATURE
DataTrans Plus
C.Z. C.Z.
NO DECO
0
DECO
5
10
30
20
®
Fig. 20 - Message Box
O
2
60
40
50
In addition to using a high contrast LCD for easy readability in low light condi­tions, the DataTrans Plus Oceanglo™ backlight feature evenly and easily illu­minates the full display (Fig. 21). This means that on night dives, in caves, or any other low light situation, you illuminate the display when you wish to view it with the touch of a button.
Upon activation the Oceanglo™ backlight will illuminate the display while the unit performs its diagnostic check. To activate the backlight during the Dive mode, simply press the Advance/Left control button. Oceanglo will remain il­luminated as long as the button is depressed, plus 10 seconds after being re-
DECO
20
®
PSI
304050
O
2
FT
60
NO DECO
0
C.Z. C.Z.
5
10
Press
Fig. 21 - Backlight
(Oceanglo™)
21
Page 34
®
140°F (60°C)
NO DECO
0
2
DECO
FT
PSI
5
60
10
304050
20
®
O
C.Z. C.Z.
32°F (0°C)
Fig. 22 - Operating
Temperature Range
leased (for a maximum of 15 seconds).
Oceanic recommends that you always carry primary and backup dive lights when conducting dives that could include low light situations.
OPERATING TEMPERATURE
The DataTrans Plus will operate in almost any temperature diving environ­ment in the world (Fig. 22) between 32 and 140°F (0 and 60°C). At extremely low temperatures, the LCD may become sluggish, but this will not affect it's accuracy. If stored or transported in extremely low temperature areas (below freezing), you should warm the module and its batteries with body heat before diving.
Even though the DataTrans Plus will operate in this wide range of tempera­tures, it is possible to damage the electronics if left exposed to direct sunlight, or in a hot confined space (like a car trunk). After the dive, cover the Display Module and keep it out of the sun. If inadvertently left in the direct sunlight for a long period, the LCD display may become totally black. If this occurs, immediately immerse the Display Module in water. The display should recover its normal appearance after a few minutes. Damage
from excess heat, or cold, is not covered by the DataTrans Plus two year limited warranty.
22
Page 35
SHARING THE DATATRANS PLUS
WARNING: Never participate in sharing or swapping of a dive computer. Doing so may result in injury or death.
The DataTrans Plus provides information based upon a diver’s personal dive profile, and therefore must not be “shared” between divers. You should never, under any circumstances, swap your computer with another unit be­tween dives, or share your computer with another diver underwater. It is im­possible for two divers to stay precisely together underwater, and your computer's dive profile tracking of previous dives will be pertinent to you only. Nitrogen and oxygen loading of a second user may be significantly different and thus swapping dive computers could lead to inaccurate and po­tentially dangerous predictions of decompression and oxygen accumulation status. This rule applies to the use of all dive computers, but is especially im­portant when using the DataTrans Plus, due to the personal information it provides.
DataTrans Plus
R
E
S
P
O
N
R
E
V
I
D
E
S
L
I
B
23
Page 36
®
NO DECO
b
C.Z. C.Z.
DECO
PSI BAR
O
MAX FT M
2
c
h
g
Key: a - Gas Consumption Indicator b - Tissue Loading Bar Graph c - O2 Accumulation Bar Graph d - Variable Ascent Rate Indicator e - Gas Time Remaining Bar Graph f - Transmitter Link icon g - Operating mode icon (detail A) h - Oxygen mode symbol i - Low Battery icon (detail B) j - Maximum Depth icon k - Elapsed Dive Time icon l - Temperature icon m - PC Interface sensors
i
a
d
f
5
0
l
Fig. 23 - Graphic Interface
Legend
24
10
20
30
®
40
50
60
e
k
m
j
Page 37
DataTrans Plus
ACTIVATION
and
SETUP
25
Page 38
®
MAKING THE DATATRANS PLUS PERSONAL
Before you dive with the DataTrans Plus for the first time, you will need to be­come acquainted with its interactive features, and select your personal display settings using the Control Console and Mode Menu.
Fig. 24 - Standard
First Stage
Fig. 25 - DX3 Integrated
First Stage
26
The transmitter must first be installed into a high pressure port of your regu­lator first stage, facing to one side (Fig. 24) - unless you purchased the DX3
Integrated first stage in which the transmitter is a built-in component (Fig.
25). Oceanic strongly recommends that installation be performed by an Au­thorized Oceanic Dealer at the time of purchase. If this is not possible, refer to the instructions for this procedure on page 102.
NOTE: The DataTrans Plus transmitter is compatible with all Oceanic first stages, but cannot be guaranteed to fit certain models produced by other manufacturers. Check with your Au­thorized Oceanic Dealer to verify compatibility with your regu­lator first stage.
For the Display Module to receive a tank pressure signal from the Transmit­ter, the two devices must first be 'Linked'. The code (serial number) of the Transmitter must be entered as the 'Link' code in the Display Module.
Page 39
If your Display Module and Transmitter were packaged and shipped from the
®
DECO
C.Z. C.Z.
PSI
BAR
M
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
MAX
factory as a complete system, the code of your Transmitter has already been entered as the 'Link' code in your Display Module. If the two units have been purchased separately by you or your Authorized Oceanic Dealer, it will be nec­essary to set the Link code in the Display Module so that a tank pressure sig­nal can be received. If the Display Module has been purchased as a stand alone computer without the Transmitter, it has been preset at the factory as a non-linked unit, but can easily be reset at any time to 'Link' with a Transmit­ter code (serial number) using the Control Console.
ACTIVATING THE DISPLAY
Before activating the Display Module, the regulator containing the transmitter
R2
must be connected to a full tank and pressurized by slowly opening the tank valve. Tank pressure of 50 psi (3.5 BAR ), or greater, is required for the trans­mitter to activate. Position the display module within 3 feet (1 m) of and par­allel to the transmitter (Fig. 26), and hold it in this location during activation.
WARNING: Never activate the DataTrans Plus underwater. This may result in inaccurate depth and no-decompression time displays. If activated deeper than 4 feet (1 meter) underwater a message "TOO DEEP" will appear and the unit will shut off.
DataTrans Plus
Fig. 26 - Positioning
During Activation
27
Page 40
R2
®
C.Z. C.Z.
NO DECO
0
DECO
PSI BAR
5
10
40
30
20
®
50
Fig. 27 - Diagnostics
in Progress
C.Z. C.Z.
NO DECO
0
DECO
PSI
5
10
40
30
20
®
50
Fig. 28 - Diagnostics
Successfull
MAX FT M
FT
To activate the Display Module, press the Select/Right button once and release.
O
2
R2
The DataTrans Plus will immediately enter Diagnostic Mode, displaying all “8’s”, followed by “dashes”, and then a countdown from 9 to 0 (Fig. 27). The Message Box will read SELF - TEST and the Oceanglo backlight will illumi­nate the display throughout the diagnostic check. A single beep will be emit­ted to indicate successful completion of the operation.
60
While conducting diagnostics, the DataTrans Plus checks its display functions, coded frequency link to the transmitter, and battery voltage to ensure that ev­erything is working correctly.
If the Display Module is already set to the Transmitter's link code, the Link
O
2
icon will disappear from the screen and tank pressure will be displayed nu­merically (Fig. 28).
It will also check the ambient barometric pressure, and calibrate its present depth as zero. At elevations of 2,000 ft. (610 m) or higher, it will recalibrate itself to measure depth in feet of fresh water instead of feet of sea water.
60
28
Page 41
During the diagnostic mode, battery voltage level of both the Display Module and the Transmitter are measured to determine if sufficient voltage is avail­able to maintain operation for one day of diving. If there is not enough battery voltage in the Display Module, it will deactivate itself or would not have acti­vated at all. If there is not sufficient voltage in the Transmitter, the Link icon and a tank pressure of "00" PSI will flash on display (Fig. 29). This could also indicate one of the following conditions:
The Transmitter was not pressurized prior to activation of the Display Module.
The Display Module was not positioned in close proximity to the Trans­mitter during activation, or not correctly positioned parallel to it.
The Display Module is not linked to the same coded frequency as the Transmitter.
If the Display Module is out of the range of the Transmitter while in Surface Mode, the flashing will stop and the Link icon will disappear within 5 seconds when the Display Module is returned to its correct proximity to the Transmit­ter.
R2
In the latter case, it will be necessary to follow the prescribed linking proce­dure outlined on page 41 to set the Display Module to the Link code that matches that of the Transmitter's serial number, or to set a specific link code that will allow the Display Module to function as a stand alone unit.
DataTrans Plus
C.Z. C.Z.
NO DECO
0
DECO
PSI
5
10
30
20
®
Fig. 29 - Unsuccessful Link
O
2
FT
60
40
50
29
Page 42
®
R
E
S
P
O
N
R
E
V
I
D
E
S
L
I
B
Be a -
RESPONSIBLE DIVER
at all times.
Low Battery conditions and battery power conservation are described in more detail in the Care & Maintenance section.
WARNING: If either or both of the Low Battery icons remain on display following diagnostics, Oceanic strongly recommends that you DO NOT dive until the batteries are replaced. See the battery replacement procedure on page 99.
If no dive is made within 2 hours after initial activation, the Display Module will automatically deactivate to conserve its battery power. Check your Dis­play Module before entering the water to verify that it is functioning and doesn’t need reactivation.
To save its battery power while on the surface, the Display Module will stop searching for a Transmitter signal after 10 minutes. The signal Link can be restored by depressing the Select/Right button on the Display Module. It will also be restored automatically upon descent on a dive.
WARNING: During activation and diagnostics, if any display or message varies from the information presented here, return the DataTrans Plus to your Oceanic Dealer for inspection.
30
Page 43
SURFACE MODE
Surface Mode immediately follows Diagnostic Mode after initial activation (Fig. 30), or after the linking procedure has been performed. It also appears after a dive when you ascend shallower than 3 feet (1 meter). Surface Mode is identified by the Surface Time icon. Information displayed includes Tank Pressure, Depth (00 FT), Surface Time with flashing colon, Temperature, and Time of Day with colon flashing.
MODE MENU SYSTEM
The Mode Menu system allows you to set the various display options that will make the DataTrans Plus your personal computer. If you followed the Linking procedure, you have already developed a feel for how the Control Console works. The Advance/Left button is used to move through the Mode Menu and change each setting, and the Select/Right button is used to select (enter) the mode or setting that is currently on the screen. A brief glossary and hierarchy of the menu system is as follows:
DataTrans Plus
NO DECO
C.Z. C.Z.
DECO
PSI
O
2
FT
Mode - Each mode provides a different display of information, or access to a submenu or setting. Some modes, such as Dive Mode and Surface Mode, are entered into automatically after activation. Others, such as the FO2 Mode, Plan Mode and Alternate Dive Mode, are accessed using the Control Console, when you want to view specific information.
5
10
30
0
20
®
60
40
50
Fig. 30 - Surface Mode
31
Page 44
®
Menu - The main menu allows interactive access from the Surface Mode to various other modes used while on the surface.
Settings - These are display options (settings) that are determined by you be­fore going diving. For example, with 'alarms' you can set the values at which depth and tank pressure will alert you when you are going too deep or running low on breathing gas.
32
Responsible
Mode Menu Sequence
• Surface
• FO2
• Plan
• Log
• History
• Set
• External Access
Settings Sequence
• Time
• Date
• Alternate
• Unit
• Alarms
• Link
• Language
NOTE: If the DataTrans Plus is left unattended for five min­utes while in the Mode Menu, it will automatically revert to Surface Mode.
Page 45
ENTERING SETTINGS
®
0
5
10
20
304050
60
®
0
5
10
20
30
40
50
60
Before going diving, enter the general settings to be used for each of your dives. These include - Time, Date, Alternate Display, Units of Measure, Depth Alarm, Gas (tank pressure) Alarm, Link code, and Language.
Note that FO2 is a 'pre dive' setting that must be entered prior to each nitrox dive. Setting the FO2 value for the nitrox mixture being used is described in the Pre Dive and Dive Mode section.
Set Time
Your DataTrans Plus has been factory set for 12:00 AM. To change to the cur­rent Time, follow this procedure, beginning in Surface Mode:
DataTrans Plus
Fig. 31 - Set Mode
1. Press the Advance/Left button 5 times to advance to the Set Mode. MODE:SET will appear, with SET flashing (Fig. 31). If you accidentally pass the Set Mode, you will need to press the Advance/Left button repeat­edly until MODE:SET reappears.
2. Press the Select/Right button to select (enter) the Set Mode. SET:TIME will appear, with TIME flashing (Fig. 32).
3. Press the Select/Right button once to select (enter) the Time setting. Time of day will appear, with the first digit flashing (Fig. 33).
4. To set the time, press the Advance/Left button to change that digit until
Fig. 32 - Set Time
5
10
0
40
30
20
®
50
Fig. 33 - Time of Day
60
33
Page 46
®
it matches that of the current time, and press the Select/Right button to save the digit shown and move on to the next.
5. Press the Advance/Left button to toggle between AM or PM, and press the Select/Right button to save the one displayed.
After the time has been set, SET:DATE will appear with DATE flashing. To
5
10
304050
0
20
®
60
set the date, continue with step 4 of the following Set Date procedure, or to re­turn to the Surface Mode press the Advance/Left button 6 times.
Set Date
R2
Fig. 34 - Set Date
5
10
0
20
®
Fig. 35 - Set Month
34
Your unit has been factory set for JAN 1 96. To change to the current Date, follow this procedure, beginning with step 1 if in Surface Mode; or beginning with step 4, if you just set the Time:
1. Press the Advance/Left button 5 times to advance to the Set Mode.
2. Press the Select/Right button to select the Set Mode.
3. Press the Advance/Left button 2 times to advance to the Date setting. SET:DATE will appear, with DATE flashing (Fig. 34).
60
40
30
50
4. Press the Select/Right button to select the Date setting. The date will ap­pear, with the month flashing (Fig. 35).
5. To set the Date, press the Advance/Left button to change the month, and ress the Select/Right button to save it. Repeat for the day and year.
Page 47
After the year has been set, SET: ALT will appear, with ALT flashing. To set
®
0
5
10
20
30
40
50
60
the Alternate display continue with step 4 of the following Set Alternate Dis­play procedure, or to return to the Surface Mode press the Advance/Left but­ton 5 times.
Set Alternate Display
Your unit has been factory set so the values of Temperature, Elapsed Dive Time, and Maximum Depth will be displayed continuously during the Dive Mode. You can turn this Alternate display 'off' and have them appear only when you wish to see them by depressing the Select/Right button during the dive. To turn the Alternate Display 'off', follow this procedure, beginning with step 1 if in the Surface Mode, or beginning with step 4 if you just set the Date:
1. Press the Advance/Left button 5 times to advance to the Set Mode.
2. Press the Select/Right button to select the Set Mode.
3. Press the Advance/Left button 3 times to advance to the Alternate set­ting. SET: ALT will appear with ALT flashing (Fig. 36).
4. Press the Select/Right button to select the Alternate Display setting. ALT: ON will appear with ON flashing (Fig. 37).
5. Press the Advance/Left button to toggle between ON and Off, and press the Select/Right button to select the one displayed.
DataTrans Plus
Fig. 36 - Set Alt
5
10
30
0
20
®
60
40
50
After the Alternate display has been set, SET:UNITS will appear with UNITS flashing. To set the Units of Measure continue with step 4 of the following Set
Fig. 37 - Alt On/Off
35
Page 48
®
Units of Measure procedure, or to return to the Surface Mode press the Ad­vance/Left button 4 times.
Set Units of Measure
5
10
0
Fig. 38 - Set Units
C.Z. C.Z.
NO DECO
5
10
0
Fig. 39 - Units
36
20
DECO
20
®
60
40
30
®
50
You can choose between Imperial (PSI and Feet) and Metric (BAR and Meters) units of measure. Your unit has been factory set for FT and PSI. To change to metric units of measure, follow this procedure, beginning with step 1 if in the Surface Mode, or beginning with step 4 if you just set the Alternate Display:
1. Press the Advance/Left button 5 times to advance to the Set Mode.
R2
2. Press the Select/Right button to select the Set Mode.
3. Press the Advance/Left button 4 times to advance to the Unit setting.
O
2
PSI
FT
SET:UNIT will appear, with UNIT flashing (Fig. 38).
4. Press the Select/Right button to select the Unit setting. PSI and FT will appear, flashing (Fig. 39).
5. Press the Advance/Left button to toggle between Imperial and Metric units, and press the Select/Right button to accept the one displayed.
After the Units have been set, SET:ALRM will appear with ALRM flashing.
60
40
30
50
To set the alarm values continue with step 4 of the Set Gas and Depth Alarm procedure beginning on page 40, or to return to the Surface Mode press the Advance/Left button 3 times.
Page 49
BREATHING GAS & DEPTH ALARM SET POINTS
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
After planning each dive according to the no-decompression dive times shown to be available in the Pre Dive Planning Sequence, Oce­anic strongly recommends that you utilize one of the greatest safety features the DataTrans Plus offers - the Gas & Depth Alarm settings.
While the DataTrans Plus uses the Audible Alarm, Message Box, and Graphic Diver Interface to automatically alert you whenever you en­ter a potentially dangerous situation, such as Decompression Dive Mode, ascending too fast, running low on breathing gas, etc., the Alarm settings allow you to preset more conservative limits to better avoid these situations.
DataTrans Plus
Depth Alarm Set Point
The Depth Alarm will alert you whenever you reach or exceed the maximum depth Set Point value that you have chosen. Of course, if you set the Depth Alarm for a depth that is deeper than the no-decompression or decompression limits for that dive, you will first be alerted by other built-in alarms that you have exceeded those limits before the Depth Alarm is activated. When the Depth Alarm is activated by reaching or exceeding your preset maxi­mum depth, the audible alarm will sound once per second, while the Message Box flashes the words “TOO DEEP” (Fig. 40) until you ascend above the
Fig. 40 - Too Deep Message
(60 FT Set Point)
37
Page 50
®
Depth Alarm Set Point value. The Depth Alarm value may be set for depths ranging from 30-320 feet (9-97.5 meters), in 10 foot (3 & 3.5 meter) incre­ments. The setting that you choose for the Depth Alarm does not change the displayed limits of no-decompression dive time remaining.
NO DECO
0
5
C.Z. C.Z.
DECO
PSI
10
30
20
®
O
2
FT
60
40
50
Fig. 41 - Gas Alarm Message
38
Breathing Gas Alarm Set Point
The Breathing Gas Alarm is an acoustic alert that indicates you are approach­ing a critical Breathing Gas Time Remaining. The Breathing Gas Alarm Set Point refers to the surfacing tank pressure reserve of your choice, which may be set for tank pressures ranging from 300 to 1000 psi (21 to 70 bar).
You will recall that Breathing Gas Time Remaining is the time that you can remain at your present depth and, following a safe ascent, still surface with a prescribed breathing gas reserve (identified here as the Breathing Gas Alarm Set Point).
When your Breathing Gas Time Remaining reaches 5 minutes, the Gas Alarm will emit a double beep as a preliminary warning. If you allow your Breathing Gas Time Remaining to decrease to zero, the tank pressure display will flash and the Message Box flashes the words "GAS ALARM" (Fig. 41) until you as­cend to a depth of 5 feet (1.5 meters) or less.
While an immediate safe ascent is called for if the Breathing Gas Time Re-
Page 51
maining decreases to zero, there is no reason to panic. The DataTrans Plus
®
DECO
C.Z. C.Z.
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
®
0
5
10
20
30
40
50
60
has allowed for the breathing gas you will consume during a safe ascent, in­cluding decompression stops if they are required, and still provide the tank pressure reserve you have chosen, e.g., 300 psi (21 bar).
Set Depth and Breathing Gas Alarms
Your DataTrans Plus alarms have been factory set for 320 FT and 300 PSI. These set point values will be retained unless you change them.
To set your desired Depth and Breathing Gas Alarm values, or turn the au­dible alarm off, follow this procedure, beginning with step 1 if in the Surface Mode, or beginning with step 4 if you just set Units of Measure:
1. Press the Advance/Left button 5 times to advance to the Set Mode.
2. Press the Select/Right button to select the Set Mode.
3. Press the Advance/Left button 5 times to advance to the Alarm setting. SET:ALRM will appear, with ALRM flashing (Fig. 42).
4. Press the Select/Right button to select the Alarm setting. The current Depth Alarm Set Point value will appear, flashing (Fig. 43).
5. Press the Advance/Left button repeatedly to change the Depth Alarm Set Point value to the depth you choose, and press the Select/Right button to accept (save) that setting. The current Breathing Gas Alarm Set Point
DataTrans Plus
Fig. 42 - Set Alarm
Fig. 43 - Depth Alarm
39
Page 52
NO DECO
®
C.Z. C.Z.
DECO
PSI
value will appear, flashing (Fig. 44).
6. Press the Advance/Left button repeatedly to change the Breathing Gas
O
2
Alarm Set Point value to the pressure you choose, and press the Select/ Right button to save that setting. ALARM:ON will appear, with ON flashing (Fig. 45).
7. Press the Advance/Left button to toggle between 'on' or 'off', and press the Select/Right button to accept the one displayed.
5
10
0
20
®
Fig. 44 - Gas Alarm
C.Z. C.Z.
NO DECO
0
DECO
5
10
20
®
Fig. 45 - Alarm On
40
60
40
30
50
Turning Off the Audible Alarm
WARNING: Turning off the audible alarm disables an impor­tant tool that can help you avoid decompression diving or low breathing gas emergencies. Although possible, Oceanic does
O
2
not recommend the disablement of the audible alarm.
The audible portion of the alarm may not be desired by some divers in certain situations. Underwater photographers, for instance, may find that the alarm frightens off marine life at a close distance, and will therefore want to turn it off temporarily before they begin a dive involv-
60
40
30
50
ing that activity. For a reason such as this, the audible portion of the alarm feature may be turned off at your discretion. The DataTrans Plus will otherwise continue to display information according to the values that have been entered for the Alarm settings.
Page 53
After the Alarm values have been set, SET:LINK will appear with LINK flash­ing. To set the Transmitter Link code continue with step 4 of the following LINKING procedure, or to return to the Surface Mode press the Advance/Left button 2 times.
LINKING PROCEDURE
Your DataTrans Plus Display Module has been factory set with the Transmitter's serial number, or at serial number 999999 if no Transmitter was purchased. If the DataTrans Plus linked automatically immediately following activation, there is no need to perform the Linking procedure. However, if the Link icon and pressure value of 00 PSI remained flashing on the screen, the Linking procedure must be performed before the Display Module can receive tank pressure data from the Transmitter.
The Linking procedure may also need to be performed in the event that your DataTrans Plus Display Module or Transmitter has received factory service, and is returned to you with a different Linking code. You may also choose to “unlink” your Display Module from the Transmitter to use the Display Module as a stand alone computer, without its pressure integrated features, or to link it to a Transmitter that has been purchased separately at a time in the future.
To set the Display Module with the link code (serial number of the Transmit­ter), follow this procedure beginning with step 1 if in the Surface Mode, or be-
DataTrans Plus
R
E
S
P
O
N
R
E
V
I
D
E
S
L
I
B
41
Page 54
®
5
10
0
40
30
20
®
50
Fig. 46 - Set Link
5
10
0
40
30
20
®
Fig. 47 - Link Mode
ginning with step 4 if you just set the Alarms:
1. Press the Advance/Left button 4 more times to advance to the Set Mode.
60
2. Press the Select/Right button to select the Set Mode.
3. Press the Advance/Left button 6 times to advance to the LINK setting. SET:LINK will appear, with LINK flashing (Fig. 46).
4. Press the Select/Right button to select the Link Mode (Fig. 47).
5. Compare the 6 digit serial number code shown on the display to the first 6 digits of the serial number embossed on the Transmitter.
a. If the numbers are the same, press the Select/Right button 6 times to ac-
cept (save) the code number shown. DO NOT press the Advance/Left button, this will change the code to an incorrect number.
b. If the numbers are not the same, press the Advance/Left button to change
60
50
the first digit of the code as required, then press the Select/Right button to save that digit. Repeat until all 6 digits are set correctly.
If you would like to set the Display Module to function as a stand alone, non pressure integrated computer, use the control buttons as described above to set the code as 999999 (Fig. 48).
5
10
0
20
®
Fig. 48 - Stand Alone
42
60
40
30
50
After the Link code has been set, SET:LANG will appear with LANG flashing. To set your preferred Language continue with step 4 of the following Set Lan­guage procedure, or to return to the Surface Mode press the Advance/Left but­ton once.
Page 55
Set Language
®
0
5
10
20
30
40
50
60
®
0
5
10
20
30
40
50
60
The Message Box displays warning messages in the language that you choose ­either English, Italian, German, Spanish, or French. The Mode Menu sys-
tem is also displayed in the language selected, so it is very important that you do not accidentally change this setting to a language that you do not understand.
Your unit has been factory set for English. To change the language, follow this procedure beginning with step 1 if in the Surface Mode, or beginning with step 4 if you just set the Link code:
DataTrans Plus
1. Press the Advance/Left button 4 more times to advance to the Set Mode.
Fig. 49 - Set Language
2. Press the Select/Right button to select the Set Mode.
3. Press the Advance/Left button 7 times to advance to the Language set­ting. SET:LANG will appear, with LANG flashing (Fig. 49).
4. Press the Select/Right button to select the Language setting. The default setting ENGLISH will appear, flashing (Fig. 50).
5. Press the Advance/Left button to scroll through the Language setting op­tions until you arrive at the one you prefer.
6. Be careful to ensure that the language selection flashing is the one that you prefer before you press the Select/Right button.
7. Press the Select/Right button to save the Language chosen and return to the Surface Mode.
Fig. 50 - Language
43
Page 56
NO DECO
®
C.Z. C.Z.
DECO
Language Correction Procedure
If you accidentally selected a language that you do not prefer or understand, you may find it very confusing to navigate further in the Mode Menu or under­stand the Message Box warnings (Fig. 51). To correct the language, it is rec­ommended that you wait 5 minutes to allow the Display Module to default to the Surface Mode, then carefully perform the following procedure:
1. Press the Advance/Left button 5 times to arrive at the Set Mode.
2. Press the Select/Right button once to enter Set Mode.
3. Press the Advance/Left button 7 times to arrive at the Language sub­menu.
O
2
4. Press the Select/Right button once to select the Language submenu.
5. Press the Advance/Left button as needed until your preferred language appears.
6. Press the Select/Right button to save the language chosen and return to the Surface Mode.
5
10
30
0
20
®
Fig. 51 - Language
(Set: Link - German)
44
60
40
50
Responsible
Page 57
DataTrans Plus
PRE DIVE
and
DIVE MODES
45
Page 58
®
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
Poor reception distance. (Greater than 3 feet)
Poor reception area.
Poor reception area.
46
Best reception zone.
Page 59
POSITIONING OF THE DISPLAY MODULE
The Transmitter emits a low frequency signal that radiates outward in a semi­circular pattern that is parallel to the length dimension of the Transmitter. A coiled antenna inside the Display Module receives the signal when it is posi­tioned within a zone parallel to or at a 45 degree angle to the Transmitter as shown on page 46. The Display Module cannot effectively receive the signal when it is held out to the sides of the Transmitter, or held at distances greater than 6 feet (2 meters) in front of the Transmitter. Best reception is achieved when the Display Module is within 3 feet (1 meter) of the Transmitter.
DataTrans Plus
When installed into a high pressure port of your first stage regulator, the Transmitter must be positioned so that it faces horizontally outward from the tank valve.
LINK INTERRUPTION UNDERWATER
During a dive, you may at times move the Display Module out of the signal pattern resulting in a temporary interruption of the link signal.
An interruption greater than 15 seconds will cause the Link icon and tank pressure display to flash (Fig. 52). Also, an audible alarm will sound once per second until the link is restored. The link will be restored within 4 seconds
NO DECO
0
5
C.Z. C.Z.
DECO
PSI
10
30
20
®
O
2
FT
60
40
50
Fig. 52 - Underwater Link
Interruption
R2
47
Page 60
Responsible
®
after the Display Module is moved back into its correct position. An interruption may also occur while the Display Module is in an area 3 to 4
feet (1 meter) of a running dive propulsion vehicle. The link will be restored within 4 seconds after the vehicle motor is shut off or when the Display Mod­ule is moved out of this area.
When using a photo strobe, temporary link interruption may occur shortly af­ter the strobe flashes. The link will be restored in 4 seconds.
WARNING: During the period of link interruption, the Display Module will temporarily loose transmitted pressure related functions and displays. These will be regained 4 seconds after the link is restored.
OPERATIONAL MODES
The DataTrans Plus operates in different operational modes, including the Di­agnostic, Surface, and Set Modes which have already been explained in de-
tail in the previous section, “Activation & Setup.” This section describes the modes that the DataTrans Plus operates in before and during a dive.
48
Page 61
FO2 MODE
®
0
5
10
20
30
40
50
60
WARNING: The percentage of oxygen (FO2) in the nitrox mix being used must be set 'before each' nitrox dive.
The DataTrans Plus can be used either as an Air computer or a Nitrox com­puter. After activation, the Display Module will operate as an Air computer without displaying information associated with oxygen calculations, unless it is set for a percentage of oxygen (FO2) other than Air (numerical value between 21 and 50 %).
FO2 Set for Air
DataTrans Plus
Fig. 53 - FO2 Mode
If you are using 'Air' as your breathing gas, you can verify that 'Air' is the FO2 value by performing the following procedure beginning in the Surface Mode:
1. Press the Advance/Left button once to enter the Mode Menu. MODE: FO2 will appear, with FO2 flashing (Fig. 53).
2. Press the Select/Right button once to select the FO2 Mode. FO2 Air will appear, with Air flashing (Fig. 54).
3. Press the Select/Right once to accept Air as the setting. MODE:PLAN will appear with PLAN flashing.
4. Press the Advance/Left button 5 times to return to the Surface Mode.
C.Z. C.Z.
NO DECO
0
DECO
5
10
40
30
20
®
Fig. 54 - Air Setting
O
2
60
50
49
Page 62
NO DECO
®
C.Z. C.Z.
DECO
PSI
When set with an FO2 value of 'Air', the DataTrans Plus will perform calcula-
O
2
tions the same as if FO2 were set for 21% oxygen, internally accounting for oxygen loading for any subsequent Nitrox dives. However, oxygen related dis-
FT
plays, warnings, and the O2 bar graph will not appear on the LCD display for that dive (Fig. 55), or subsequent dives, unless FO2 is set for a numerical value (21 to 50 %).
5
10
0
40
30
20
®
Fig. 55 - Dive Mode (Air)
C.Z. C.Z.
NO DECO
0
DECO
5
10
40
30
20
®
Fig. 56 - FO2 Default
(after nitrox dive)
50
60
50
Setting FO2 for a Nitrox Dive
You can program the DataTrans Plus for nitrogen-oxygen (nitrox) mixtures of 21% to 50% oxygen (O2) before each nitrox dive. If FO2 is set at a value of 21%, the unit will remain set as a '21% nitrox computer' for subsequent nitrox dives until FO2 is set to a higher value, or the Display Module automatically
O
2
turns off. Once FO2 is set to a value 'greater than 21%' to match the nitrox mix being used for that nitrox dive, the FO2 value displayed during the FO2 Mode that is accessible 10 minutes after that dive will be 50% (Fig. 56).
FO2 must be reset for each repetitive nitrox dive, or the value will automati­cally default to 50 and the dives will be calculated based on 50% O2 for oxygen
60
50
calculations and 21% O2 (79% nitrogen) for nitrogen calculations. Once a dive is made with the DataTrans Plus set as a nitrox computer (FO2 set for a nu­merical value), the unit cannot be programmed to operate as an 'Air' computer until 24 hours after the last dive. 'Air' will not be displayed as an option in the FO2 Mode.
Page 63
WARNING: The DataTrans Plus will default to an FO2 value of
®
DECO
C.Z. C.Z.
O
2
NO DECO
0
5
10
20
30
40
50
60
50 after a 10 minute surface interval following a nitrox dive. Therfore, this procedure must be performed prior to descent on each and every nitrox dive, even if the percentage of oxygen in the nitrox mixtures used remains the same.
To set (enter) a numerical value for the percentage of oxygen (FO2) in your nitrox mix, perform the following procedure beginning in the Surface Mode:
1. Press the Advance/Left button once to enter the Mode Menu. MODE: FO2 will appear with FO2 flashing.
2. Press the Select/Right button once to select the FO2 Mode. If a previous dive was made that day with the unit set for Air, FO2 and Air will appear, with Air flashing. If a previous dive was made with the unit set for a nu­merical value of FO2, FO2 and 50 will appear, with 50 flashing.
3. Press and hold the Advance/Left button, or press it repeatedly, until the proper value of FO2 appears (Fig. 57). The percentage displayed will ad­vance 1 (%) per second from 21 to 50 (%), then display 'Air' again.
4. Once the proper value of FO2 is displayed, press the Select/Right once to save the value shown as the setting for that dive. MODE:PLAN will ap­pear with PLAN flashing.
5. Press the Advance/Left button 5 times to return to the Surface Mode.
DataTrans Plus
Fig. 57 - Setting FO2
51
Page 64
®
PLAN MODE
Oceanic strongly recommends that you access the Plan Mode prior to every dive to review the Pre Dive Planning Sequence (PDPS) that will help you plan your dive as required to avoid exceeding no de­compression, or oxygen exposure limits. This is especially important for
repetitive dives, when the Pre Dive Planning Sequence will indicate for you the adjusted no decompression bottom times that are available to you for your next dive, based on any residual nitrogen or oxygen accumulation (whichever is in control) following your last dive and surface interval.
WARNING: The Pre Dive Planning Sequence predicts only no decompression times for subsequent dives. Depending on cyl­inder size, breathing gas consumption, and oxygen accumula­tion, you may have less time available than indicated because of breathing gas quantity or other limitations.
5
10
0
20
®
Fig. 58 - Plan Mode
52
60
40
30
50
To access the Plan Mode, press the Advance/Left button twice beginning in the Surface Mode. MODE:PLAN will appear, with PLAN flashing (Fig. 58). Press the Select/Right button once to enter the Plan Mode.
If FO2 was set for 'Air', the PO2 screen will not appear and the DataTrans Plus will begin to scroll through the Pre Dive Planning Sequence.
Page 65
If FO2 was set for a numerical value, FO2 and the value set will appear as the
®
DECO
C.Z. C.Z.
O
2
NO DECO
0
5
10
20
30
40
50
60
®
DECO
C.Z. C.Z.
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
MAX
first screen (Fig. 59), followed by a screen showing the maximum depth that can be achieved for an oxygen partial pressure of 1.60 ATA for that FO2 (Fig.
60). The DataTrans Plus then scrolls through the Pre Dive Planning Sequence
(PDPS), displaying a sequence of depths from 30 to 160 feet (9 to 48 meters) in 10 foot (3 meter) increments. With each depth display, you will see either 'pre­dicted' no decompression limits based upon your previous dive profiles (if cal­culated to be nitrogen controlled), or 'predicted' oxygen tolerance limits based upon either a single dive oxygen dose or your 24 hour accumulation of oxygen (if calculated to be oxygen controlled).
DataTrans Plus
Fig. 59 - FO2 Value Set
Information displayed includes Previous Dive #, Depth, Dive Time available at that depth (which includes descent time at a rate of 120 ft/min), Plan Mode icon, and the message MODE:PLAN.
No decompression times are only displayed for depths where there is at least 3 minutes of dive time available at the depth, taking into account a descent rate of 120 feet (36 meters) per minute. Depths greater than the maximum depth that can be achieved with a partial pressure of oxygen (PO2) of 1.60 ATA will not be displayed.
R2
Fig. 60 - Max Allowed Depth
53
Page 66
NO DECO
®
C.Z. C.Z.
DECO
O
2
FT
After scrolling once through depth and dive times available, the DataTrans Plus will automatically return to the Surface Mode. Plan Mode can be ac­cessed and the sequence repeated as often as you choose. If you wish to inter­rupt the Pre Dive Planning Sequence to return to the Surface Mode, you may do so at any time by pressing either control button.
Fig. 61 - Nitrogen Controlled
NO DECO
C.Z. C.Z.
DECO
O
2
FT
Fig. 62 - Oxygen Controlled
NO DECO
C.Z. C.Z.
DECO
O
2
Fig. 63 - Max Daily O2
Limit Exceeded
54
Prior to a Repetitive Nitrox Dive
If the segments of the Tissue Loading Bar Graph are displayed during the Pre Dive Planning Sequence (Fig 61), that next dive is calculated to be controlled by nitrogen loading.
If the segments of the O2 bar graph are displayed Pre Dive Planning Sequence (Fig. 62), that next dive is calculated to be controlled by oxygen loading.
R2
The DataTrans Plus will store oxygen accumulation for up to 10 dives con­ducted during a 24 hour period. In the event that the maximum limit for oxy­gen loading has been exceeded for that day (24 hour period), all of the seg­ments of the O2 bar graph will be displayed (Fig. 63). Depth and Time dis­plays will not appear until the O2 bar graph recedes into the green (normal) zone (i.e., your daily oxygen doseage decreases an amount equivalent to the amount accumulated during the latest dive completed).
Page 67
WARNING: The DataTrans Plus must be manually activated
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
and be in an operating mode prior to start of a dive. The unit will not activate automatically by immersion in water. Also, FO2 must be set prior to commencing each nitrox dive.
NO DECOMPRESSION DIVE MODE
After it has been manually activated and completed its diagnostic checks, the DataTrans Plus will enter the No Decompression Dive Mode when you descend deeper than 5 feet (1.5 meters). No Decompression Dive Mode (Fig. 64) can be recognized by the No Decompression Dive Mode icon that is displayed to the left of the large numeric display of Dive Time Remaining. Also displayed are Tank Pressure, Current Depth, and the applicable bar graphs.
If you set the Alternate display option 'ON' during setup, the values of Tem­perature, Elapsed Dive Time, and Maximum Depth will be displayed continu­ously in the lower portion of the screen (Fig. 65), unless overridden by more critical information. If you set the Alternate display option 'OFF', the values can be viewed by depressing the Select/Right button. Time of Day will appear briefly when the button is released.
To activate the Oceanglo™ backlight during a dive, press the Advance/Left but­ton. The display will be illuminated as long as the button is depressed plus 10 seconds after it is released (for a maxmum illumination time of 15 seconds).
DataTrans Plus
Fig. 64 - No Decompression
Dive Mode
Fig. 65 - Alternate Display
55
Page 68
®
As your depth and elapsed dive time increase, the Tissue Loading Bar Graph will fill with segments (green to red) to represent the absorption of nitrogen; and if FO2 was set for a value other than 'Air', the O2 bar graph will fill with segments (green to red) to represent oxygen accumulation for that dive, or 24 hour period, whichever is greater. The Gas Time Remaining bar graph will recede from green to red as breathing gas time remaining decreases from 60 to 0 minutes. Segments of the Breathing Gas Consumption and Variable Ascent Rate Indicator bar graphs fill (and recede) as their respective rates increase (and decrease) throughout the dive.
Deco Zone
C.Z. C.Z.
NO DECO
0
DECO
PSI
5
10
40
30
20
®
Fig. 66 - Decompression
Dive Mode
56
WARNING: Every effort should be made to keep all of the bar graphs in the green throughout your dives to reduce your risk
O
2
FT
of exposure to decompression sickness and oxygen toxicity.
DECOMPRESSION DIVE MODE
The DataTrans Plus provides information that will help you avoid, or if neces­sary, manage emergency decompression.
60
50
The Decompression Dive Mode activates when the No Decompression Limits are exceeded and the Tissue Loading Bar Graph enters the red decompression zone (Fig. 66).
Page 69
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
VIOLATION MODES
The DataTrans Plus enters Violation Modes when it is unable to predict an as­cent procedure.
GAUGE MODE
If the DataTrans Plus enters a Permanent Violation Mode, it will not display information relating to nitrogen or oxygen loading for the remainder of that dive or for subsequent dives conducted during the 24 hour period after surfac­ing.
NOTE: Decompression Dive Mode, Violation Modes, and Gauge Mode are each described in the Handling the Extremes section.
ASCENDING TO THE SURFACE
While ascending to shallower depths, the segments that have filled up the Tis­sue Loading Bar Graph will begin to recede (Fig. 67), offering a graphic repre­sentation of your multilevel diving capability. If you entered Decompression Mode, you must not complete your ascent until the Tissue Loading Bar Graph is at least inside the yellow Caution Zone.
DataTrans Plus
Fig. 67 - Tissue Loading Bar
Graph Receding
57
Page 70
NO DECO
®
C.Z. C.Z.
DECO
PSI
If you have not entered Decompression Mode, a safety stop made be­tween 15-20 feet (5-6.5 meters) is strongly recommended as a stan­dard procedure before completing your ascent.
You should make every effort to complete all of your ascents with the Tissue Loading Bar Graph inside of the green zone.
While you cannot provide a guarantee against the occurrence of de­compression sickness, you may choose your own personal zone of caution based upon your individual age, physique, excessive weight, training, experience, etc. to reduce the statistical risk.
The Variable Ascent Rate Indicator™ shows how fast you are ascending. When
O
2
you exceed the maximum recommended ascent rate of 60 feet per minute (18 meters per minute), the bar graph will enter the red (Too Fast) zone (Fig. 68),
FT
and you will be alerted by the segments flashing. You will also be alerted by an audible alarm, and the Message Box will flash the message “TOO FAST". The warnings will stop when your ascent rate is slowed.
5
10
0
40
30
20
®
50
Fig. 68 - Ascent 'Too Fast'
58
60
Page 71
ALTITUDE DIVING
The mathematical model within the DataTrans Plus accounts for the reduced No Decompression dive time available at higher elevations based on NOAA (National Oceanic and Atmospheric Administration) guidelines. When diving in high altitude lakes or rivers from 2,000 to 14,000 feet (610 to 4,268 meters), the DataTrans Plus will adjust automatically, providing corrected depth and reduced No Decompression and Oxygen Exposure times.
Also, when above 2,000 feet (610 meters), depth calibration is automatically changed to read in feet of freshwater rather than feet of seawater.
NOTE: If activated above 14,000 feet (4,268 meters), a message "TOO HIGH" will appear and the unit will shut off.
DataTrans Plus
More about altitude diving is presented in the Reference section.
WARNING: Until it has shut itself off, you must not use the DataTrans Plus at a different altitude than the altitude where it was originally activated. Doing so will result in an error equal to the difference in barometric pressure, and possibly a false dive mode with erroneous data.
R
E
S
P
O
N
R
E
V
I
D
E
S
L
I
B
Be a -
RESPONSIBLE DIVER
at all times.
59
Page 72
®
(This page intentionally left blank.)
60
Page 73
DataTrans Plus
POST
DIVE
MODES
61
Page 74
®
NO DECO
0
5
C.Z. C.Z.
DECO
PSI
10
30
20
®
O
2
FT
60
40
50
Fig. 69 - Post Dive
Surface Mode (<10 min)
POST DIVE SURFACE MODE
When you ascend to 3 feet (1 meter) or shallower, the DataTrans Plus will en­ter Surface Mode (Fig. 69) and begin counting your surface interval.
TRANSITION PERIOD
The first 10 minutes is, in affect, a transition period during which time:
• The main time display starts counting Surface Interval (colon flashing).
• The Surface Mode icon will appear (flashing).
• The Tissue Loading Bar Graph will indicate current nitrogen loading.
• The O2 bar graph will indicate current oxygen loading, if the dive was a nitrox dive (FO2 set for a numerical value).
• Temperature and Time of Day will also be displayed.
If you descend during the 10 minute transition period, time underwater will be considered a continuation of that dive. The time at the surface (if less than 10 minutes) will not be added as Elapsed Dive Time. During the 10 minute tran­sition period, the Log Mode is accessible in the Mode Menu and will display that dive's data, however, the data will not be stored in the unit's memory until the 10 minute period on the surface is completed. No other modes are acces­sible in the Mode Menu during the 10 minute transition period.
62
Page 75
Once 10 minutes have elapsed, the Surface Mode icon and Surface Interval
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
time display colon will stop flashing (Fig. 70) indicating that the dive and tran­sition period are completed, and a subsequent descent will be considered a new dive. Other information will continue to be displayed as described above and you will have full access to the Mode Menu.
FO2 MODE
If the DataTrans Plus was set for FO2 of 'Air' or 21% prior to the dive, it will stay set for 'Air' or 21%, respectively, unless you reset it to a higher numeric value prior to the next dive. If it was set for a numeric value of FO2 greater than 21% (22 to 50 %) prior to the dive, the FO2 value displayed after the dive will be 50 % (Fig. 71) and subsequent dives will be calculated based on 50% oxygen for oxygen calculations and 21% oxygen (79% nitrogen) for nitrogen calculations, unless you set FO2 for another value.
Remember!! You must set the FO2 to match the specific nitrox mix for each nitrox dive, even if the percentage of oxygen in the mix is the same as the previous.
PLAN MODE
DataTrans Plus
Fig. 70 - Surface Mode
(> 10 min)
NO DECO
0
5
C.Z. C.Z.
DECO
10
30
20
®
O
2
60
40
50
When you access the Plan Mode after a dive, the FO2 screen will show the FO2 set point value that you entered for the next dive. The PO2 screen will show
Fig. 71 - FO2 Default
63
Page 76
®
the maximum depth allowed for the FO2 set point and your 24 hour oxygen accumulation (dose). Time to Fly, described next, will appear prior to the Pre Dive Planning Sequence.
The Pre Dive Planning Sequence will show 'adjusted' no decompression limits based on residual nitrogen calculated to be remaining from previous dives. Calculated dive times and the maximum allowed depth displayed will increase as the real time surface interval increases after completion of a dive. The Pre Dive Planning Sequence will only scroll to the maximum depth allowed by the nitrogen or oxygen limit, whichever is in control. The respective bar graph will be displayed to indicate which is in control.
C.Z. C.Z.
NO DECO
0
DECO
5
10
20
®
Fig. 72 - Time to Fly
(first 12 hours)
64
O
2
Time to Fly
As you should be aware from your own training, the longer you wait to fly (or travel to higher elevations) after diving, the more you will reduce your exposure to decompression sickness.
The Time To Fly counter begins counting down 10 minutes after the last dive
60
40
30
50
has ended to assist you with deciding when enough surface time has elapsed to fly. It appears prior to the Pre Dive Planning Sequence when you access the Plan Mode, and shows the word 'FLY' with a countdown (Fig. 72) that starts at 23:50 (hr:min) and counts down to 12:00 (hr:min).
Page 77
Twelve hours after the last dive, the Surface Mode will disappear from the screen, and the Fly Mode will be displayed continuously, with the final 12 hour countdown from 11:59 to 0:00.
WARNING: During the final 12 hours, the DataTrans Plus is in a countdown mode only, and must be 'activated' by depressing the Select/Right button prior to start of a dive.
After a surface interval of 12 hours, you may choose to fly (or travel to higher elevations), provided that your dive profile(s) did not enter decompression. If your diving involved decompression or a repetitive, multi day profile, it is strongly recommended that you wait a full 24 hours after your last dive to add a greater degree of protection. More About Flying After Diving and DAN's guidelines is presented in the Reference section.
DIVE LOG MODE
Dive Log Mode, identified by the Dive Log icon (Fig. 73a), can be accessed us­ing the Mode Menu while on the surface. Information from your 12 latest dives is stored in the log for viewing, giving you the opportunity to record data in your log book before it is eventually overwritten by subsequent data. After 12 dives are accumulated, each subsequent dive will overwrite the oldest dive that exists in the log, i.e. the DataTrans Plus will add the most recent dive while deleting the oldest. Dive log information will be retained when batteries
DataTrans Plus
a
C.Z. C.Z.
NO DECO
0
DECO
5
10
30
20
®
Fig. 73 - Log Mode Icon
O
2
60
40
50
65
Page 78
®
are removed from the Display Module. Dives are displayed in a reverse sequence that starts with the dive most re-
cently conducted back to the oldest of the 12 dives stored. Thus, your most re­cent dive will always be the first shown in the sequence.
5
10
304050
20
0
®
60
To eliminate confusion, each dive is separately 'stamped' (identified) with the date on which it was made and the time of day that the dive started.
Fig. 74 - Dive Log Mode
C.Z. C.Z.
NO DECO
0
DECO
5
10
40
30
20
®
Fig. 75 - Log Mode (first screen - date)
66
Each dive has four log screens that display date, time, nitrogen related data, and oxygen related data, respectively. To access the Dive Log Mode, follow this procedure, beginning in Surface Mode:
O
2
1. Press the Advance/Left button 3 times to advance to the Log Mode. MODE:LOG will appear with LOG flashing (Fig. 74).
2. Press the Select/Right button once to select (enter) the Log Mode. If you accidentally pass Log Mode, press the Select/Right button repeatedly un­til MODE:LOG appears.
3. The first screen to appear (Fig. 75) will display the most recent dive re-
60
50
corded, identified by the Log Mode icon, the dive Number and dive Date .
4. To bypass the dive currently being displayed to view an older dive, press the Select/Right button until the desired dive is displayed, identified by the dive number and date of the dive.
Page 79
5. Press the Advance/Left button once, to display the Time of Day (Fig. 76) when the dive started.
6. Press the Advance/Left button once more to view the information associ­ated with nitrogen (Fig. 77). Displayed will be: Log Mode icon, dive Num­ber, Surface Time between that dive and the one previous to it, lowest water Temperature during the dive, Elapsed Dive Time, and Maximum Depth. If the dive shown in the log display was the only one of the day, Surface Time will represent the time between initial activation and the beginning of the first dive. Also shown will be the tissue nitrogen loading at the time you surfaced from the dive (Tissue Loading Bar Graph), the maximum rate of breathing gas consumption (Gas Consumption Indica­tor), and the maximum rate of ascent at any time during the dive (Vari­able Ascent Rate Indicator).
7. Press the Advance/Left button once to view the information associated with oxygen (Fig. 78). Displayed will be: Log Mode icon, dive Number, FO2 symbol and the value that FO2 was set at for the dive. If FO2 was set for a value other than 'Air', the PO2 symbol and value of the maxi­mum partial pressure of oxygen attained during the dive will also be dis­played. Also shown will be the oxygen loading (O2 bar graph) at the time you surfaced from the dive (maximum for that dive or 24 hour period, whichever was greater at the time). The O2 bar graph will also display the segment that represents the maximum level of oxygen exposure (dose) that you achieved during that dive.
8. Press either button to advance to the first screen (Date) of the next dive
DataTrans Plus
C.Z. C.Z.
NO DECO
DECO
Fig. 76 - Log Mode
(second screen - time)
C.Z. C.Z.
NO DECO
DECO
Fig. 77 - Log Mode
(third screen - nitrogen)
C.Z. C.Z.
NO DECO
DECO
Fig. 78 - Log Mode
(fourth screen - oxygen)
O
2
O
2
O
2
67
Page 80
®
5
10
304050
0
20
®
60
Fig. 79 - History Mode
in the log's reverse sequence.
To exit the Log Mode, press the Select/Right button repeatedly to advance through all recorded dives and return to the Surface Mode.
HISTORY MODE
The History Mode offers a convenient summary of your DataTrans Plus com­puter since it was originally purchased, or last received factory service. Infor­mation provided on three sequential screens includes total number of all dives, total number of decompression dives, total elapsed dive time in hours, deepest maximum depth, and total number of violation dives.
C.Z. C.Z.
NO DECO
0
DECO
5
10
30
20
®
Fig. 80 - History Mode
(first screen - total)
68
O
2
To access the information provided in History Mode, follow this procedure, be­ginning in the Surface Mode:
1. Press the Advance/Left button 4 times to advance to the History Mode. MODE:HIST will appear, with HIST flashing (Fig. 79).
2. Press the Select/Right button once to select (enter) the History Mode.
60
40
50
The first screen (Fig. 80) will show the No Decompression icon, total
number of dives, total elapsed dive time, and maximum depth.
3. Press the Advance/Left button once to view the second screen (Fig. 81) that shows the Decompression Dive icon and the total number of Decom­pression dives. Total elapsed dive time and maximum depth will remain
Page 81
on display.
®
DECO
C.Z. C.Z.
O
2
NO DECO
0
5
10
20
30
40
50
60
4. Press the Advance/Left button once again to view the third and final
R2
screen (Fig. 82) that shows the Dive Mode icon and the total number of dives during which the DataTrans Plus entered a Violation Mode.
5. Press either button to return to the Surface Mode.
NOTE: Previous unit history will be erased whenever your DataTrans Plus receives factory service.
EXTERNAL ACCESS MODE (EACC)
Using special infrared linking hardware and a unique PC software program, data from your dives can be downloaded (copied) from your DataTrans Plus into an IBM compatible personal computer program running on a Windows
3.x or Windows® 95 operating system. Instructions for performing the inter­face and download are provided with the hardware and software package that is available separately from your Authorized Oceanic Dealer. Ask for OceanLog™ for DataTrans Plus.
The OceanLog™ software program provides dive profile data, and nitrogen and oxygen loading information that was sampled throughout the dives that are downloaded..
Once you have acquired your OceanLink™ package, to access the External Ac-
DataTrans Plus
C.Z. C.Z.
NO DECO
0
DECO
5
10
30
20
®
Fig. 81 - History Mode
(second screen - deco)
®
Fig. 82 - History Mode
(third screen - violation)
O
2
60
40
50
69
Page 82
®
5
10
30
0
20
®
60
40
50
Fig. 83 - External Access
Mode
cess Mode, follow this procedure, beginning in the Surface Mode:
R2
1. Press the Advance/Left button 6 times to advance to the External Access Mode. MODE:EACC will appear, with EACC flashing (Fig. 83).
2. If you are going to download information, follow the procedures described in the OceanLog™ User's Manual.
3. If you are not going to download information, press the Advance/Left but­ton once to return to the Surface Mode. If you press the Select/Right but­ton once, the letters EACC will appear with a countdown from '8' to '0' (Fig. 84). After counting down to '0', the DataTrans Plus will revert to the Surface Mode.
NO DECO
0
5
C.Z. C.Z.
DECO
10
30
20
®
O
40
50
Fig. 84 - EACC Countdown
70
2
R
E
60
S
P
O
Be a RESPONSIBLE DIVER at all times.
R
E
V
I
D
N
E
S
L
I
B
Page 83
DataTrans Plus
HANDLING
THE
EXTREMES
71
Page 84
®
DECOMPRESSION
RULES
ARE NOT MEANT
TO BE BENT
EMERGENCY DECOMPRESSION
There are few legitimate excuses for making unplanned Decompression dives, and the consequences of this type of diving can be severe. Decompression div­ing requires special training and support. The DataTrans Plus is intended
for use by recreational divers not engaged in intentional decompres­sion diving. Decompression features are provided only for emer­gency situations. By entering decompression, you automatically impose a
“ceiling” above you which you cannot immediately ascend beyond, denying you free access to the surface.
Professional military and commercial divers plan ahead for this situation by ensuring that they have complete surface support, including a redundant breathing gas supply for emergencies. They also navigate very carefully throughout their dive to ensure that they begin and complete their ascent while maintaining contact with a rope or a line to the surface. This is neces­sary for making a well controlled ascent. The DataTrans Plus is not in-
tended for use by military or commercial divers.
72
Responsible
By making an unplanned Decompression dive without the necessary prepara­tion and training, you will have placed yourself in an unnecessarily danger-
ous situation. Consider also, that one mistake can quickly be compounded by several others.
Page 85
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
The DataTrans Plus is a sophisticated instrument designed with capabilities that go beyond the range of recreational diving with compressed air. It
should not be considered, however, that these built-in capabilities provide any implied approval or consent from Oceanic for individu­als to exceed the defined limits of recreational dive profiles, as agreed on by all internationally recognized training agencies.
Decompression diving should therefore be strictly avoided. The
DataTrans Plus is designed to help you by providing a complete representation of how close you are to entering decompression. In the event that you do inad­vertently enter decompression, the red zone of the Tissue Loading Bar Graph (Fig. 85), the DataTrans Plus can provide you with limited information to help you ascend to the surface, if you follow the instructions given in this section.
Oceanic strongly recommends that you avoid entering decompres­sion, and reminds you that decompression diving requires special training.
DataTrans Plus
The DataTrans Plus cannot provide you with a backup breathing gas supply for emergencies or the ascent line you will need, and decom­pression diving greatly increases your risk of decompression sick­ness.
Fig. 85 - Decompression
Mode
73
Page 86
®
R2
NO DECO
0
5
C.Z. C.Z.
DECO
PSI
10
40
30
20
®
Fig. 86 - Decompression Stop
WARNING: Existing data for making planned decompression dives is extremely limited, and virtually nonexistent for repeti­tive decompression diving. You must therefore avoid decom­pression diving and allow a surface interval of at least 24 hours before reentering the water in the event a dive requiring emer­gency decompression is made.
If you're not careful, it is possible to enter decompression rapidly, whether at deep depths or during repetitive dives.
Upon entering decompression, you must immediately change the focus of your dive to getting safely back to the surface. Upon seeing the Tissue Loading Bar Graph enter the red decompression (DECO) zone, you should immediately be-
O
2
FT
gin a safe controlled ascent, 60 feet (18 meters) per minute or slower, to a depth slightly deeper than or equal to the required ceiling stop depth indicated (Fig. 86).
If you continue the dive at a depth more than a few feet (1 meter) deeper than the required ceiling stop depth, your exposure to decom-
60
50
pression sickness will increase, and you will risk entering violation mode and losing the information needed to ascend properly.
74
Page 87
GAS TIME REMAINING DURING DECOMPRESSION
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
Breathing Gas Time Remaining will fluctuate throughout your dive, decreas­ing as you dive deeper or breath heavier, and increasing as you ascend to shal­lower depths or relax your breathing efforts. If you inadvertently exceed no decompression limits, the DataTrans Plus will provide critical information re­garding your remaining breathing gas supply.
DataTrans Plus
NO DECO
C.Z. C.Z.
DECO
PSI
O
2
FT
Once you enter Decompression Mode, you will need to refer to the Gas Time Remaining bar graph for Breathing Gas Dive Time Remaining information.
When Breathing Gas Time Remaining decreases to 5 minutes, the tank pres­sure numerals will flash and a Double Beep will sound (Fig. 87). This means that only 5 minutes remain before your breathing gas supply will be reduced to the minimum level necessary to perform the required decompression stops and still provide a tank pressure reserve upon surfacing.
When the Breathing Gas Time Remaining decreases to zero minutes, the Au­dible Alarm will continuously beep once per second (Fig. 88), signaling the need for an immediate safe ascent to your first decompression stop. However, there is no need to panic. The DataTrans Plus has allowed for the breathing gas you will consume during a safe ascent including the required decompres­sion stops and still provide the surfacing tank pressure reserve you chose (en­tered) during setup, e.g. 500 psi.
5
10
30
0
20
®
60
40
50
Fig. 87 - Gas Time Remain-
ing (5 minutes)
Fig. 88 - Gas Time Remain-
ing (zero minutes)
75
Page 88
®
R2
NO DECO
0
5
C.Z. C.Z.
DECO
PSI
10
30
20
®
Fig. 89 - Caution Zone
(Tissue Loading Bar Graph)
CAUTION ZONE (C.Z.)
Your dive training taught you not to get too close to the No Decompression limits. The yellow Caution Zone (C.Z.) of the Tissue Loading Bar Graph (Fig.
89) offers you a convenient way to consistently monitor how close you are com­ing to the No Decompression limit. Oceanic suggests always leaving the water with the Tissue Loading Bar Graph in the green No Decompression (NO DECO) zone.
WARNING: Never exit the water with the Tissue Loading Bar Graph in the red decompression (DECO) zone. Doing so greatly increases the risk of decompression sickness, and may result in injury or death.
O
2
FT
Body metabolism varies from person to person, and even from day to day. If you are feeling slightly less than 100%, or you are in less than perfect physical shape, use the yellow Caution Zone as a visual reference to place a
wider margin of protection between you and the No Decompression limit.
60
40
50
Responsible
76
Page 89
DECOMPRESSION DIVE MODE
The DataTrans Plus will help you to avoid and manage decompression.
WARNING: Oceanic recommends the application of respon­sible diving practices and does not recommend decompression diving, or diving deeper than 130 feet (39 meters), as these practices will greatly increase your risk of decompression sick­ness.
DataTrans Plus
NO DECO
0
5
C.Z. C.Z.
DECO
PSI
10
30
20
®
O
2
FT
60
40
50
Decompression Dive Mode (Fig. 90), identified by the Decompression Dive icon located to the left of the Main Time Display, activates when the theoretical no decompression dive/depth limits are exceeded causing the Tissue Loading Bar Graph to pass the yellow caution (C.Z.) zone and enter the red decompression (DECO) zone. Also, the audible alarm will emit a double beep to alert you. The other bar graphs will continue to represent their respective information.
Information displayed includes Tank Pressure, current depth, and total ascent time that includes stop times required at all ceilings and vertical ascent time calculated at 60 feet (18 meters) per minute. The Message Box alternately flashes the required stop depth and time, and CEILING (Fig. 91).
The amount of decompression credit time that you receive is dependent on depth, with slightly less credit given the deeper you are. Still, you must
Fig. 90 - Decompression
Mode
NO DECO
0
5
C.Z. C.Z.
DECO
PSI
10
30
20
®
O
2
FT
60
40
50
Fig. 91 - Deco Mode (alternate message)
R2
77
Page 90
®
C.Z. C.Z.
NO DECO
0
DECO
PSI
5
10
40
30
20
®
Fig. 92 - Deco Mode
(managing a stop)
never ascend shallower than your decompression ceiling. Doing so will greatly increase your risk of decompression sickness, and place
the DataTrans Plus into a Conditional Violation Mode described later. When coping with surge and swells, it may be difficult to stay at an exact depth. You should stay slightly deeper (Fig. 92) than the required stop depth indi­cated until the next shallower stop depth appears. Then, you can ascend to, but not shallower than, that indicated ceiling stop depth.
Once you have performed the required decompression, the DataTrans Plus will switch to the No Decompression Dive Mode, allowing additional time un­derwater. Though more time is theoretically available, it is strongly recom­mended that you spend the remainder of the dive continuing to decompress at, or slightly deeper than, 10 feet (3 meters). This will let the Tissue Load-
O
2
ing Bar Graph recede further into the yellow caution (C.Z.) zone or green no decompression (NO DECO) zone, helping you reduce your tissue nitrogen
FT
loading as much as possible.
Warning: If you exceed certain limits, the DataTrans Plus will
60
50
not be able to tell you how to get safely back to the surface. These situations will make the DataTrans Plus enter Violation Modes and must be avoided at all costs. They exceed tested limits and can result in loss of some DataTrans Plus functions for 24 hours after the dive in which a violation occurred.
78
Page 91
VIOLATION MODES
The Violation Modes that the DataTrans Plus can enter, depending on the situation, are termed Conditional, Delayed, and Immediate. Gauge Mode and Permanent Violation Mode are continuations of these Violation Modes. It is important to understand each different Violation Mode and how to carry out emergency procedures in the event you enter one.
CONDITIONAL VIOLATION MODE
The DataTrans Plus will alert you to the possibility of losing decompression management abilities by entering the Conditional Violation Mode. If properly handled, the Conditional Violation Mode can assist you in getting back to the surface and allow continued use of the DataTrans Plus. The situation that will force the DataTrans Plus to enter a Conditional Violation Mode is: Ascent to a Depth Shallower than the Required Decompression Ceiling (Fig. 93).
A momentary rise above the ceiling, such as with a surge or swell, could cause this to happen. Therefore you should stay slightly deeper than the exact ceil­ing stop depth, watching the DataTrans Plus closely when managing decom­pression. The audible alarm will beep once per second and the Message Box will alternately flash CEILING and VIOLATION until you descend below the required decompression ceiling stop depth.
R2
DataTrans Plus
C.Z. C.Z.
NO DECO
0
DECO
PSI
5
10
30
20
®
Fig. 93 - Conditional Viola-
tion (above ceiling)
O
2
FT
60
40
50
79
R2
Page 92
®
If you descend below the required decompression ceiling before 5 minutes have elapsed, the DataTrans Plus will continue to function as if no violation had oc­curred. In this case, no off-gassing credit will be given, and for each minute above the ceiling 11/2 minutes of penalty time is added to decompression stop time.
The added penalty decompression time will have to be 'worked off' first, before obtaining off-gassing credit. Once the penalty time is worked-off, and off-gas­sing credit begins, required decompression stop depths and time will decrease and the Tissue Loading Bar Graph will recede into the caution (C.Z.) zone and revert to the No Decompression Dive Mode.
R2
NO DECO
0
5
C.Z. C.Z.
DECO
PSI
10
30
20
®
Fig. 94 - Delayed Violation
(above ceiling >5min)
80
If you stay above (shallower than) the required ceiling stop depth for more
O
2
than 5 minutes, the Tissue Loading bar Graph segments will flash and the De­layed Violation Mode will be entered.
FT
DELAYED VIOLATION MODE
Three conditions will cause the DataTrans Plus to enter the Delayed Violation
60
40
50
Mode:
1. You remain above the required Decompression Ceiling Stop Depth for more than 5 minutes (Fig. 94).
Page 93
The audible alarm will beep once per second and the Message Box will alter­nately flash CEILING and VIOLATION until you descend below the required ceiling stop depth.
As previously described, you would then need to follow the ceiling stop depths and times toward the surface as the Tissue Loading Bar Graph recedes into the caution (C.Z.) zone.
2. Your necessary decompression requires a ceiling stop depth be­tween 60 feet (18 meters) and 70 feet (21 meters).
In this situation the Audible Alarm will emit One Long Beep, the Tissue Load­ing Bar Graph will flash, and the Message Box alternately flashes EXCEEDED and CEILING (Fig. 95) four times. Total Ascent Time needed to get back to the surface will still be displayed numerically in the Main Time display.
To get back to the surface, you must safely ascend to just deeper than 60 feet (18 meters) staying as close to 60 feet (18 meters) as possible without causing the Message Box to flash. After waiting until the required ceiling stop depth display indicates 50 FT/ 15 M, you can ascend to, but no shallower than 50 feet (15 meters) and continue decompressing. As the required ceiling stop depth display indicates 40 FT/ 12 M, 30 ft/ 9 M, 20 FT/ 6 M, and then 10 FT/ 3 M, you can ascend to, but no shallower than the required ceiling stop depth indicated.
DataTrans Plus
C.Z. C.Z.
NO DECO
0
DECO
PSI
5
10
30
20
®
Fig. 95 - Delayed Violation
(>60 FT ceiling required)
O
2
FT
60
40
50
R2
81
Page 94
®
R2
NO DECO
0
5
C.Z. C.Z.
DECO
PSI
10
30
20
®
Fig. 96 - Delayed Violation
(>330 ft / 99.5 m)
After Total Ascent Time reaches zero and the Tissue Loading Bar Graph re­cedes into the yellow caution (C.Z.) zone, you can surface. However, to add
a greater margin of protection, Oceanic strongly recommends that you wait until the segments of the Tissue Loading Bar Graph are well within the 'green' no decompression (NO DECO) zone, unless a low tank pressure condition requires you to surface.
3. You descend deeper than 330 feet (99.5 meters).
NOTE: Oceanic reminds you that the DataTrans Plus is in­tended for no decompression diving at depths within 130 feet (39 meters). Expanded capabilities of the DataTrans Plus are provided as safety features to assist you with emergency situa-
O
2
tions.
Upon descending deeper than 330 feet (99.5 meters), the Tissue Loading Bar
FT
Graph will flash and the Current Depth and Maximum Depth displays will only indicate 3 dashes (Fig. 96) until ascent is made to a depth shallower than 330 feet (99.5 meters), at which time the Current Depth display will be re­stored. Max Depth will continue to display 3 dashes. Exceeding the maximum
60
40
50
operating depth is described in detail on page 85 of this section.
Five minutes after reaching the surface from a dive in which a De­layed Violation occurred, the DataTrans Plus will enter an Immedi­ate Violation Mode and revert to Gauge Mode for 24 hours.
82
Page 95
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
IMMEDIATE VIOLATION MODE
Warning: The DataTrans Plus enters Immediate Violation Mode when a situation totally exceeds its capacity to predict an ascent procedure. These dives represent gross excursions into decompression that are beyond the boundaries and spirit of the DataTrans Plus design. If you are following these dive profiles, Oceanic advises you not use a DataTrans Plus dive computer.
Immediate Violation Mode occurs when a Decompression Stop depth
much greater than 60 feet (18 meters) is required. This situation would be preceded by entering the Delayed Violation Mode previously described.
DataTrans Plus
The DataTrans Plus cannot accurately calculate decompression times for depths much greater than 60 feet (18 meters) and offers no indication of how much time spent underwater would result in the need for greater than a 60 foot (18 meter) decompression stop depth.
If a ceiling much greater than 60 feet (18 meters) is required, an Immediate Violation Mode (Fig. 97) will be entered, and you will be alerted by a Single Long Beep of the audible alarm. This situation would be preceded by the De­layed Violation Mode. The DataTrans Plus would then operate with limited functions (current depth, maximum depth, and elapsed dive time) in Gauge Mode during the remainder of that dive and for 24 hours after surfacing.
R2
Fig. 97 - Immediate
Violation
83
Page 96
C.Z. C.Z.
NO DECO
5
10
0
R2
Fig. 98 - Gauge Mode
(underwater)
C.Z. C.Z.
NO DECO
0
DECO
5
10
20
Fig. 99 - Gauge Mode
(10 min after surfacing)
DECO
20
®
®
®
GAUGE MODE
O
2
PSI
FT
Underwater, the Gauge Mode is a continuation of the Immediate Violation Mode that turns the DataTrans Plus into a digital instrument without any de­compression or oxygen monitoring functions (Fig. 98). The Tissue Loading Bar Graph and O21 Bar Graph flash, while the Message Box flashes VIOLA­TION. The numeric Dive Time Remaining will be absent from the screen.
60
40
30
50
The Alternate Dive Mode may still be accessed underwater by pressing the Se­lect/Right button.
After surfacing, Gauge Mode does not provide the FO2, Plan, or Time to Fly features. A countdown timer beginning 10 minutes after the dive at 23:50 (hr:min) with a “triple dash” display (Fig. 99) will inform you of the time re-
O
2
maining before normal DataTrans Plus operation can resume with full fea­tures and functions.
PERMANENT VIOLATION
Entering the Immediate Violation Mode, then Gauge Mode, will result in loss
60
40
30
50
of all DataTrans Plus decompression and oxygen monitoring functions for 24 hours after that dive. This condition is considered a Permanent Violation.
84
Page 97
EXCEEDING MAXIMUM OPERATING DEPTH
Although the DataTrans Plus will withstand the pressures found at 330 feet (99.5 meters), the depth that you can still use all of its features could be much shallower.
WARNING: The maximum recommended sport diving depth limit is 130 feet (39 meters). Any deeper dive should be avoided. Special training, equipment, and support are neces­sary for this type of diving. Oceanic does not advocate diving to depths below 130 feet (39 meters), the basis for which is purely theoretical.
The maximum depth the DataTrans Plus will display all of its features is 330 feet (99.5 meters). Upon exceeding a depth of 330 feet (99.5 meters), the Tis­sue Loading Bar Graph will flash, and Depth and Max Depth displays will only indicate and flash three dashes “ ” signifying that you are 'Out of Range'. Also, the message box will alternately display TOO DEEP with the Alternate
R2
displays (Fig. 100). The numeric display for current depth will not reappear until you ascend shallower than 330 feet (99.5 meters). You will also enter the Delayed Violation Mode. For the remainder of that dive, and in the log for that dive, only three dashes will be displayed as the value for Maximum Depth.
DataTrans Plus
C.Z. C.Z.
NO DECO
0
DECO
PSI
5
10
30
20
®
Fig. 100 - Out of Range
(>330 ft / 99.5 m))
O
2
FT
60
40
50
R2
85
Page 98
®
R
E
S
P
O
N
R
E
V
I
D
E
S
L
I
B
Be a -
RESPONSIBLE DIVER
at all times.
OXYGEN EXPOSURE
There are few legitimate excuses for exceeding the maximum limits for expo­sure to oxygen, and the consequences of CNS (Central Nervous System) oxy­gen toxicity can be severe, resulting in Gran Mal convulsions and drowning. Diving with enriched nitrogen-oxygen (nitrox) mixtures requires special train­ing and certification.
WARNING: The oxygen features of the DataTrans Plus are in­tended for use only by recreational divers trained for nitrox diving by an instructor certified by a recognized training agency to teach diving with nitrox. The DataTrans Plus is not intended for use by military or commercial divers.
By making a nitrox dive without the necessary training, preparation, and equipment, you will have placed yourself in an unnecessarily dangerous situ-
ation. The DataTrans Plus is a sophisticated instrument designed with capa­bilities that go beyond the range of recreational diving with compressed air. It
should not be considered, however, that these built-in capabilities provide any implied approval or consent from Oceanic for individu­als to exceed the defined limits of oxygen exposure, as agreed on by all internationally recognized nitrox training agencies. Nitrox diving should therefore be strictly controlled.
86
Page 99
PARTIAL PRESSURE OF OXYGEN
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
®
DECO
C.Z. C.Z.
PSI
FT
O
2
NO DECO
0
5
10
20
30
40
50
60
As depth increases during the dive, the partial pressure of oxygen increases. As you approach the depth limit for the FO2 value set before that dive, the DataTrans Plus will alert you and display the PO2 level while you reduce oxy­gen partial pressure according to your training.
High PO2 Dive Mode
DataTrans Plus
The DataTrans Plus enters the High PO2 Dive Mode when partial pressure of oxygen becomes equal to or greater than 1.40 ATA. The audible alarm will emit a double beep, and the PO2 value and the symbol 'PO2' will appear in the lower portion of the display (Fig. 101). They will remain on display until par­tial pressure of oxygen decreases below a value of 1.40 ATA.
If partial pressure of oxygen continues to increase, the value of PO2 displayed will increase from 1.40 toward a value of 5.00 ATA in increments of '.01' ATA. When PO2 reaches the maximum limit of 1.60 ATA, the audible alarm con­tinuously emits one beep per second, and the large red segment of the O2 bar graph, the PO2 value, and PO2 symbol will flash continuously as a warning (Fig. 102) until the level of PO2 decreases below 1.60 ATA . In the event that you enter High PO2 Dive Mode, you must immediately focus on reducing the partial pressure of oxygen by slowly ascending to a shallower
Fig. 101 - High PO2 Mode
( 1.40 BAR)
Fig. 102 - High PO2 Mode
( 1.60 BAR)
87
Page 100
®
NO DECO
0
5
C.Z. C.Z.
DECO
10
30
20
®
O
2
60
40
50
Fig. 103 - O2 Bar Graph
depth at a safe rate in accordance with your nitrox training. If you continue
the dive at your current depth, or descend deeper, your exposure to CNS oxygen toxicity will increase.
OXYGEN ACCUMULATION
It is also important that you understand that conducting repetitive dives using enriched nitrogen-oxygen (nitrox) mixtures can lead to oxygen buildup, reduc­ing oxygen tolerance while increasing the risk of pulmonary oxygen toxicity. The O2 bar graph (Fig. 103) provides a visual representation of oxygen accu­mulation for either that dive or 24 hour period, whichever is greater.
Oceanic strongly recommends that you avoid exceeding oxygen expo­sure limits, and reminds you that nitrox diving requires special train­ing and understanding of the effects of oxygen toxicity.
WARNING: In the event that you exceed the maximum per dive allowable oxygen exposure (dose), it is recommended that you allow a surface interval of at least 2 hours before reentering the water. If you exceed the maximum 24 hour period allow­able oxygen exposure (dose), you must allow a surface interval of at least 24 hours before reentering the water.
88
Loading...