• Datalogging for up to 1000 points on
Temp-100, 2000 on Temp-300
• USB output (Temp-300 only)
• Differential temperature
measurements
• Operator selection of Celsius or
Fahrenheit scale
• Resolution of 0.1° C/F from -199.9 to
999.9°
• Large backlit LCD with two lines of
four-digit display
• Hold feature for temporarily retain a
reading
• Displays min and max readings
• Field calibration capability
• Disabling of Auto-Off function
• Low battery warning
• Two blade female ANSI miniconnector input
• Operates with a wide selection of
probes
- 1 -
2. SAFETY PRECAUTIONS
WARNING:
1. This instrument is designed to
accept low level signals supplied by
standard Thermocouples. Under NO
circumstances should the input
voltage exceed the specified 50V
RMS.
2. To prevent ignition of a hazardous
atmosphere, batteries must only be
changed in an area known to be
non-hazardous.
CAUTION:
1. Do not use or store this instrument in
microwave ovens or any abnormally
hot or cold areas.
2. Weak batteries should not be left in
the instrument. Dead batteries can
leak and cause damage to unit.
DANGER:
1. Voltages present at the
Thermocouple may also be present
at the battery terminals. Always
disconnect the Thermocouple when
changing batteries.
- 2 -
3. SPECIFICATIONS
Thermocouple Thermometers
Type
Type J
Type K
Type T
Type E
Type R (300 only)
Type S (300 only)
Type N (300 only)
Type B (300 only)
Temperature range
–210°C to 1200°C
(–346°F to 2192°F)
–250°C to 1372°C
(–418°F to 2501°F)
–250°C to 400°C
(–418°F to 752°F)
-250°C to 1000°C
(-418°F to1832°)
0°C to 1768°C
(32°F to 3214°F)
0°C to 1768°C
(32°F to 3214°F)
-250°C to 1300°C
(-418°F to 2372°F)
200°C to 1800°C
(392°F to 3272°F)
ra
Out of
nge display: - - -
lution
Reso
0.1°/1° auto-ranging,
0.1° C/F from -199.9 to 999.9°,
1° outside this range
-
- 3 -
Accuracy
J,K,T,E & N
Below -150 °C (-238 °F):
±0.25% of reading ±1 °C (±0.25% ±0.7 °F)
Above -150 °C (-238 °F):
±0.1% of reading ±0.4 °C (±1% ±0.7 °F)
R,S & B
±0.1% of reading ±1 °C (±0.1% ±2 °F)
Display
Backlit Dot-matrix 50mm X 37.2mm
Data Logging
Temp-100 : 1000 points
Temp-300 : 2000 points
Logging Interval
1 sec to 60 min
Min/Max/Avg Function
Yes
Auto Off (adjustable time)
Enable/Disable option available
Stability Criteria
Yes, upon stability of 5 seconds
te rate
Display upda
0.6
sec per up
Input
Two thermocouple with ANSI connector
Input Protection
50V
rms
date.
.
- 4 -
Storage
–40°C to 65°C (–40°F to 149°F)
Humidity
10% to 90% (non-condensing)
Battery Life
Size: Three AA, 1.5V; Alkaline
Life: 400 hours continuous, typical,
(without backlighting)
Dimensions
Without Armor:
175mm (L) X 97mm (W) X 42mm (H)
With Armor:
180mm (L) X 102mm (W) X 52mm (H)
Weight with batteries
Without Armor: 267g
With Armor: 362g
Ingress protection:
Meets IEC-529 IP-54 for dust and water
resistant enclosures (probe attached)
CE Compliance
EN61326-1/A1: 1998 (EU EMC Directive)
- 5 -
4. BATTERY INSTALLATION
The total battery life without
backlighting and alarm is about 400
hours. The battery bar annunciator
represents the battery strength. An
empty battery annunciator indicates low
battery strength; a blinking battery
annunciator
should be replaced immediately
indicates that the batteries
.
Selected settings
and
will remain in memory even
power is turned
are being replac
are
stored in
off, or
ed.
while
memory
after
batteries
1. Before
2.
3.
4. Insert three
5.
changing
instrume
nt off
thermocoupl
battery, tur
and disconnect
e.
Loosen screw and
back
of case.
Remove
pol
Install
arity.
cover and
the t
hree
new
batteries
tighten
n
lift battery
cover
AA batteries.
observing
screw
off
.
- 6 -
5. OPTIONAL RUBBER ARMOUR
1. To insert thermometer into the
optional rubber armor, slide in from
the top of meter before pushing the
bottom edges of meter down to set it
into position. Lift up the stand at the
back of meter for bench top
applications if necessary.
2. To remove thermometer from armor,
push out from the bottom edges of
meter until it is completely out of boot.
- 7 -
6.
OPTIONAL HANDSFREE KIT
You can use the optional magnets and
strap in the Hands-free Kit accessories for
hands free operations.
- 8 -
7. CONNECTING A THERMOCOUPLE
Use
the
your instrume
incorr
erroneous readi
colour coded by type using
American
correct thermocouple type
ect th
ANSI
nt setting.
ermocouple
ngs. T
Colour Code as
Using an
type
hermocouples are
for
will result in
th
e North
follows:
TYPE COLOR
J
Black
K Yello
T Blue
E Purple
Thermocouple connectors have one wide
blade and one narrow blade. Do not force
connector in backwards. Connect
thermocouples to receptacles at top of
instrument as shown in the following
illustration.
w
- 9 -
Thermocouple wiring polarity must be
correct. If readings decrease as the
temperature increases, the thermocouple
wires may be reversed. The red wire is
negative for thermocouple wires
manufactured in North America.
If no probe is connected the display will
read “open”.
hermocouples are sensitive
T
measuring junction
,
measurements
to stabilize. Multip
of the
approx
probe by 5 will give you
imate time
allow
lying
required
at the tip or
. W
hen
time for the
taki
reading
the time constant
.
ng
the
- 10 -
8. KEY FUNCTIONS
F1
F2 Choose probe T1, T2 or T1-T2
F3
hold Freeze display
on/off
light
recall▲
log▼
Note: Function keys change in setup
mode to provide advanced operation
flexibility. Text above key will indicate
function.
Step through Min, Max and
Avg readings.
Toggle between menu and
measure mode
Turns meter on and off (press
and hold for 3 seconds to turn
off)
Press momentarily to turn on
backlight
Recalls and steps through
stored readings
Stores current measured
value to memory
- 11 -
9. DISPLAY OVERVIEW
1
7
6
5
2
3
4
The dot matrix display features a large
primary display, smaller secondary
displays for channel info or min/max/ave,
and helpful annuciators for added
measurement data
1 HOLD - Active
2 Data Logging is Active
3 Alarm Enabled – channel in alarm
indicated: T1 or T2 or T1&T2
4 MAX/MIN/AVG of Secondary
channel if MIN/MAX/AVG activated
5 MAX/MIN/AVG of Primary channel
if MIN/MAX/AVG activated
6 Min/Max hit time since Min/Max
activated. For Avg, it is continually
increments since activated
7 Current active Mode –
Min/Max/Avg
- 12 -
10. MEASUREMENT MODE
On initial start-up the meter will display
the measured value for input one in the
primary display and for input two in the
secondary display.
F2
Pressing the
key will toggle primary
display through input one (T1), input two,
and the delta (T1 – T2) values.
F1
Pressing the
key initiates and toggles
through Minimum, Maximum, and
Average reading modes.
F3
Pressing
enter accesses Setup mode.
- 13 -
11. HOLD FUNCTON
Press the hold key to retain the reading
on the display. Press the hold key again
for normal operation.
12. MIN, MAX, and AVE FUNCTION
Press the F1 key to toggle between the
minimum, maximum, and average
readings. The minimum and maximum
reading function is ideal for monitoring
unattended operations while continually
displaying every temperature change that
occurs. The minimum and maximum
values are sensed and automatically
stored.
To exit and clear this function, press the
F3 to access the Menu functions.
See the Clear Reset menu section for
more details.
13. DATA LOGGING
Press the
log
▼ key to store the current
reading to memory. The memory
indicator M = 1234 shows the memory
location for the next stored reading.
Press the
recall ▲
key to review stored
readings.
See section on Data Logging for timed
logging, and logging to a computer (300
model only).
See section on Clear/Reset for
information on clearing stored readings.
- 14 -
14. SETUP MODE
To access the setup mode from
measurement mode press the F3 key
(Menu).
Press ▲▼ keys on the meter key pad to
scroll through options.
To enter a setup screen press Select
key.
To return to measurement mode press
F3
Meas
are listed
key. Following menu options
1. General Setup
2. User field calibration
3. Alarm settings
4. Data logging settings
5. View user calibration report
6. Clear/Reset options
- 15 -
F1
15. GENERAL SETUP SCREEN
The first page of the General Setup
screens let you set probe type,
measurement units, time, and date.
Press F1 to indicate you want to change
the setting of the current parameter or
recall▲ or log▼ to move to the next
parameter.
Press recall▲ or log▼ to change the
options.
Press F2 to choose the next setting.
Whenever set the options, press F1 for
accepting the choice.
- 16 -
On the second page you can set auto-off
time, line frequency, and password.
This screen below is used to reset/change
password. In the event if uses forget
his/her password, 5586 can be used to
reset to a new value
- 17 -
16.
CALIBRATION SCREEN
The thermometer is factory calibrated and
does not require calibration before use.
The Calibration function allows single
point calibration of the thermometer, at
0°C (32°F) to compensate for
thermocouple off-set error. It is NOT
necessary to perform a field calibration to
obtain specified meter accuracy. Use the
field calibration feature to improve
thermometer/probe accuracy or to
compensate for thermocouple drift.
..
Before go into the calibration mode, must
enter the password. Press F2 to change to
the next digit. (Default Password is 9900)
There are three calibration options:
Offset – adjusts at a single point
Slope – adjusts at two points
Match – adjusts readings on T1 to match
those on T2. Or adjusts T2 to match T1.
- 18 -
Select calibration method by pressing F1
and the
▲ or ▼ keys.
channel you with to calibrate the
.
keys
Then select the
▲ or ▼
Offset Calibration
Use the ▲ or ▼ keys to adjust the value to
match known temperature standard. Press
F1 to accept.
- 19 -
Slope Calibration
Use the ▲ or ▼ keys to adjust the value to
match known temperature standard. .
Press F1 to accept. Then move to second
temperature point using the ▲ or ▼ keys
and repeat.
Match Calibration
Use the ▲ or ▼ keys to adjust the value to
match T1 readings and T2 readings.
- 20 -
17.
ALARMS SCREEN
Disable or enable the alarm for individual
probe by pressing recall▲ or log▼and F1
to accept. Increase or decrease individual
limit by pressing recall▲ or log▼.
- 21 -
18. DATA LOGGING SCREEN
Data Transfer from Meter to Computer
(Temp-300 only)
Press recall▲ or log▼to choose the
logging methods as auto or manual. If it is
auto logging, using recall▲ or log▼to set
time interval. Its range is from 0min to
60min.
Button “Page1” will appear only in Temp-
300.
Please refer to softcopy of the driver
manual in the CD for installation and
datalogging instructions.
***NOTE: METER MUST BE TURNED
ON PRIOR TO CONNECTING USB
CABLE OR COMPUTER MAY NOT
RECOGNIZE THE INSTRUMENT***
- 22 -
Once the USB connection is establish with
PC, press the Start button to download
data from Meter to PC using
HyperTerminal.
19.
CALIBRATION REPORT SCREEN
The Calibration report will show the time
and date along with results of the last user
calibration.
- 23 -
20. CLEAR / RESET SCREEN
Press F1 to choose which data you want
to clear or reset. For calibration, logged
data and reset all, you will have to enter
the password to proceed. (Default
Password is 9900)
- 24 -
21.
MAINTENANCE
Properly us
maintain calibr
require service
cleaning
ed, the th
ermome
ation indefinitely
other than
of the
housing and changing
ter
should
and
occasional
not
of
the batteries.
22.
CLEANING
WARNING:
TO PREVENT IGNITION OF A
HAZARDOUS ATMOSPHERE BY
E
LEC
TROSTATIC DISCHARGE,
CLEAN WITH DAMP CLOTH.
Do
not
clean wi
Use mild
use excessive
BATTERIES
23.
If there is no display when the
thermometer is turned on, check condition
of the three AA batteries. Also check that
the battery terminals are clean and
batteries are properly installed. If
replacement is necessary, refer to the
BATTERY INSTALLATION AND
REPLACEMENT section for replacement
procedure.
th
abrasives or solven
detergents,
fluid.
ts.
never immerse nor
- 25 -
24. TROUBLE SHOOTING
The following chart lists the most probable
faults. There are no internal adjustments
or user-replaceable parts.
FAULT ACTION
NO
Display
Display
shows
- - - -
Display
Shows
OPEN
Display
Shows
Err
Can not
connect
to PC
(Temp-
300 only)
Check condition of batteries.
Check that batteries are inserted
properly.
Out of range indication
No thermocouple connected in the
Connector
If display shows this message
other than during the field
calibration mode, please return the
instrument for servicing
Ensure that the meter is powered
on and measuring prior to
connecting to USB port. Before and
after the driver is installed, your PC
may not recognize the Temp-300 if
the Temp-300 is not powered on
prior to connecting the USB to the
computer.
- 26 -
25.
ACCESSORIES
Replacem
ent
Accessories
Meters and Meter
Item Part Number
Temp-100 thermometer
Temp-300 thermometer
Rubber Armour with Stand
Hands free Kit (Two Magnets and Strap)
USB cable (Temp-300 only)
General purpose probe, type K
Penetration probe, type K
Surface probe, type K 08516-60
Clip-on probe, type K
General purpose probe, type J
Penetration probe, type J
Surface probe, type J 08517-60
Clip-on probe, type J
91427-40
91427-50
35427-50
35427-80
35427-85
35427-86
08516-55
08516-65
08469-02
08517-55
08517-65
08469-00
- 27 -
26.
WARRANTY
The
Manufac
turer
warra
nts this product to
be free from significant deviations from
published
three years. If
necessary within
problem will be corrected
if it is not
your
Manufacture
warr
prod
invoiced
specifications
repair or adjus
the
due
to
part as deter
r.
anty
uct
Repair
period, or
misuse or abus
to
you
.
for a
period
tment
warranty period
at
no charge
misuse or abuse on
mined by
costs
those
the
outside
resulting
e,
may be
from
of
is
, the
the
27.
PRODUCT RETURN
To limit
seller or Manu
and shipping in
the
warranty peri
produc
return
carefully and insure
damage or
be responsible for damage resulting from
careless or insufficient packing.
charges and dela
fact
produ
ct,
either within or
od. W
t,
please
. For
your
structions
state the
protectio
urer
hen
it
for
retur
agai
ys, contact the
author
before returning
reason
n, pack
nst
loss. The Manufacturer will not
ization
outside of the
ning
the
for the
the
possible
- 28 -
28. INNOCAL® CALIBRATION AND
REPAIR SERVICES
Optimum performance of your
temperature-measuring instrument is not a
timeless condition. To ensure quality
measurements, have your instrument
calibrated regularly. Trust InnoCal® to
satisfy your calibration and equipment
repair needs. With over a decade of
service, we've helped thousands of
customers meet ISO, FDA, EPA,
GLPs/cGMPs and other quality standards.
Our substantial inventory of replacement
parts ensures a fast turnaround and
prevents costly downtime. Most
instruments serviced in five business days!
Excellent Value
Get quality at a fair price. Our InnoCal®
NIST-traceable certificates offer extensive
test data on a broad range of
measurement parameters without breaking
the bank!
Reliable Support
- 29 -
Trust in our free diagnostic support and
troubleshooting advice. Our factory-trained
metrologists and technicians are armed
with years of experience and extensive
technical data.
Convenient Reminders
It’s so easy to keep your instruments
functioning properly. Based on your
requirements, InnoCal will send you a
reminder when it’s time to re-certify or
service your instrument.
We provide you with the documentation
you need to meet your most stringent
quality requirements for the control of
inspection, measuring, and test equipment.
Certification includes certificate of
calibration with test data, including:
● description and identification of the item
certified
● condition of the item
● issue date
● identification of calibration procedure
● calibration date
● as found/as left test data (where
applicable)
● signature of technician
● statement of estimated uncertainty
● list of equipment used to perform
calibration (including their calibration
dates)
With today's high quality standards such
as ISO 9000, certification is becoming
- 30 -
increasingly important. Traceability is not a
timeless condition. It must be verified and
maintained over the life of the calibration to
ensure the highest accuracy possible.
When you have your calibration done by
InnoCal, we will send you an automatic
reminder when it is time to recalibrate your
instrument.
Are your calibration certificates good
enough?
InnoCal surpasses the competition by
providing the most complete certificates as
required by NIST. All of our certificates
include measured data and point-by-point
measurement uncertainty, and by request,
we’ll provide test accuracy and test
uncertainty ratios at no extra cost. Call us
today and see why InnoCal is The Choice
of Quality.
*See our Scope of Accreditation for any
limitations.
Calibration against NIST-traceable
standards
Four test points across range of instrument.
Meters: –270 to 2316°C (–454 to 4200°F)
Probes & Systems: –80 to 1000°C (–112 to 1832°F) **Actual range is dependent on type of probe
Meter only Probe only
System
(meter + probe)**
17000-10 17001-10 17002-10
InnoCal—The Choice of Quality
866-InnoCal (866-466-6225)
www.InnoCalSolutions.com
- 31 -
TECHNICAL ASSISTANCE
If
you have any questions abo
of this
prod
uct, contact the
Manufacturer or authorized se
ut the
ller.
For
more information on
OAKTON
Instruments products, please contact
your nearest distributor or visit our web
site listed below:
Oakton Instruments
625 E Bunker Court
Vernon Hills, IL 60061
USA
Tel: (1) 888-462-5866
Fax: (1) 847-247-2984
info@4oakton.com
www.4oakton.com
Distributed by
:
use
- 32 -
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.