Before you Begin:
Remove electrode cap. Switch unit on for 15 minutes to stabilize the batteries. Soak electrodes for
a few minutes in alcohol to remove oils.
Calibration:
Select a calibration standard appropriate for your
waterproof TDSTestr:
TDSTestr 1: between 300 ppm and 1990 ppm
TDSTestr 2: between 3 ppt and 10.00 ppt
TDSTestr 3: between 300 µS and 1990 µS
TDSTestr 4: between 3 mS and 19.90 mS
It is best to select a standard close to the test solution value, and one that has a similar chemical
make-up to the test solution.
1. Pour calibration standard into two separate
containers and tap or deionized water into a
third.
2. Rinse electrode in the deionized water, then
rinse it in first container of standard, then dip
it into the second container of standard.
3. Switch unit on (ON/OFF button). Wait several
minutes to allow
display to stabilize.
4. Using a small
screwdriver, adjust the
trimpot (located inside
the Testr ’s battery
compartment) until
the display reads the
same value as the
standard.
5. Rinse the electrode in the tap or deionized
water and proceed with testing.
TDS Testing:
1. Remove cap from electrode. Switch unit on
(ON/OFF button).
2. Dip the electrode into the test solution. Stir
once and let the reading stabilize.
3. Allow time for the Automatic Temperature
Compensation to correct the readings for solution temperature changes.
4. Note the reading once the display is stable. If
the solution is above the range of the TDSTestr,
the display will show
5. Press ON/OFF to turn off Testr.
Replace cap.
Instrument Maintenance:
•
To improve performance, clean the stainless
steel electrodes by periodically rinsing them in
alcohol for 10-15 minutes.
•
Replace all 4 batteries if the display becomes
faint or disappears, or if the readings are unstable or constant.
•
If drift is detected while electrodes are
continuously exposed to solution for longer
than one hour, allow electrode to fully dry off
periodically.
NOTE: Testr life is dependent on meter and electrode care. If the electrode is exposed to damaging materials, electrode life will be shortened.
When you need a new electrode, see “Electrode
Replacement” on insert in back of box.
Changing batteries
1. Open battery compartment lid (with attached
lanyard loop).
2. Remove old
batteries;
replace with
fresh ones.
Note polarity
(shown in
battery
compartment
and in picture at right).
3. Recalibrate Testr after battery change.
Specifications
Electrode replacement:
You can replace the electrode module at the fraction of
the cost of a new Testr. When the Testr fails to calibrate, gives fluctuating readings in buffers, shows
error messages ‘E2’ or ‘OR’in a buffer, and the procedures in the Maintenance section do not help, you
need to change the electrode.
1. With dry hands, grip the ribbed Testr collar
with electrode facing you. Twist the collar
counter clockwise. (see diagram A). Save the
ribbed Testr collar and O-ring for later use.
2. Pull the old electrode module away from the
Testr.
3. Align the four tabs on the new module
so they match the four slots on the testr.
(see diagram B).
4. Gently push the module onto the slots to seat
it in position. Push the smaller O-ring fully
onto the new electrode module. Push the collar over the module and thread it into place
by firmly twisting clockwise.
Warranty:
The waterproof TDSTestr 1-4 meter body is warranted against defects in materials and workmanship for a period of 12 months from the date of
purchase; the electrode module is warrantied for
a period of 6 months from the date of purchase. If
repair, adjustment or replacement is necessary
and has not been the result of abuse or misuse
within the time period, please return the Testr—
freight pre-paid—and correction will be made
without charge. Out of warranty products will be
repaired on a charge basis.
Return of Items:
Authorization must be obtained from your
OAKTON Distributor before returning items for
any reason. When applying for authorization,
please include information regarding the reason
the item(s) are to be returned.
Note: We reserve the right to make improvements in design, construction and appearance of
products without notice. Prices are subject to
change without notice.
Waterproof TDSTestrs 1-4 Instructions
trimpot
TDSTestr 1 2 3 4
0 to 0 to 0 to 0 to
Range 1990 10.00 1990 19.90
ppm ppt µS mS
Resolution
10 0.10 10 0.10
ppm ppt µS mS
Accuracy ±2% full scale
Calibration 1-point calibration with trimpot
Calibration
300 to 3 to 300 to 3 to
Standard
1990 10.00 1990 19.90
ppm ppt µS mS
Operating
32 to 122ºF / 0 to 50ºC
Temperature
ATC
32 to 122°F / 0 to 50°C
(1.11% per °F / 2% per °C)
Four 1.5V alkaline batteries
(Eveready A76BP; supplied)
Power
100 hrs. continuous use
Alternate replacement Model
Eveready 303 silver oxide,
70 hrs. continuous use.
Dimensions 6.5" L x 1.5" Dia. (165 x 38 mm)
Weight 3.25 oz (90 gms)
keypad
+
–
Small O-Ring
(hidden)
Rotate collar
away from you
Ribbed Collar
Large
O-Ring
Large Tab
Small
Tab
Insert Electrode
Electrode
module
B
A