Oakton CON550 Instruction Manual

CON550 Benchtop Conductivity Meter
Instruction Manual
OAKTON INSTRUMENTS
www.4oakton.com
v2.0
Table of Contents
1 Brief Introduction ............................................................................................................ - 1 -
1.1 Main Features ...................................................................................................................... - 1 -
2 Technical Specifications.................................................................................................... - 1 -
3 Instrument Description .................................................................................................... - 2 -
3.1 LCD Display .......................................................................................................................... - 2 -
3.2 Keypad ................................................................................................................................ - 3 -
3.3 Connectors .......................................................................................................................... - 4 -
3.4 Stable reading display mode ................................................................................................. - 4 -
3.5 Data storage, recall, and deletion ......................................................................................... - 4 -
3.6 Max/Min Function ............................................................................................................... - 5 -
4 Conductivity Measurement .............................................................................................. - 5 -
4.1 Preparation .......................................................................................................................... - 5 -
4.2 Conductivity Electrode Information ...................................................................................... - 5 -
4.3 Information regarding Calibration ......................................................................................... - 5 -
4.4 Conductivity Calibration (take 1413 μS as an example) .......................................................... - 6 -
4.5 TDS and Conductivity ........................................................................................................... - 7 -
4.6 Sample Test.......................................................................................................................... - 7 -
4.7 Parameter Setting ................................................................................................................ - 8 -
4.8 Conductivity Electrode’s Maintenance .................................................................................. - 9 -
5 Complete Kit .................................................................................................................. - 10 -
6 Warranty ....................................................................................................................... - 10 -
- 1 -
1 Bri ef In tro duc tion
Thank you for purchasing Oakton Instruments CON 550 Benchtop Conductivity Meter. This instrument is an outstanding combination of advanced electronics desin and sensor technology. It’s the
most economical choice of a reliable lab benchtop conductivity meter.
Please read this maual carefully to properly use and maintain the meter. Oakton Instruments reserves the right
to update the contents of this manual without giving prior notices.
1.1 Main Features
The meter has a built-in microprocessor chip, which enables intelligent functions such as automatic calibration, automatic temperature compensation, data storage, parameter setup, max/min reading display, etc.
The advanced automatic calibration mode has calibration guide and self-diganosis functions, automatically recognizing standard calibration solutions.
The advanced digital processing technology intelligently improves the meter’s response time and measuring accuracy with stable reading display mode available.
The kit comes with a combination conductivity electrode with temperature sensor, measuring conductivity and temperature simultaneously, and allows for automatic temperature compensation.
IP54 dustproof and spillproof. Connectors are protected by silicone seal caps, effectively ensuring the reliability and service life of the meter under harsh environment.
Detachable electrode holder and 3 bottles of conductivity calibration solutions come with the kit for convenient use.
2 Technical Sp ec ifi ca tio ns
Conduct
ivity
Range
Conductivity: 0-200.0 mS/cm, divided into 4 ranges: (0-199.99) μS/cm, (200-1999) μS/cm, (2.00-19.99) mS/cm, (20.0-
199.9) mS/cm
TDS: (0-100) g/L, devided into 4 ranges: (0-99.9) mg/L, (100-999) mg/L, (1.00-9.99) g/L, (10.0-100.0) g/L
Resolution
Conductivity: 0.1/1 μS/cm; 0.01/0.1 mS/cm
TDS: 0.1/1 mg/L, 0.01/0.1 g/L
Accuracy
±1.0% F.S ±1 digit
Temperature
Compensation
0 to 50 ˚C (Auto or Manual)
Electrode Constant
0.1/1/10 cm-1
Reference Temperature
25 ˚C
Temperature
Compensation
Coefficient
0.00-9.99%/ ˚C, default value: 2.00%/ ˚C
- 2 -
Calibration
1-4-point Auto Calibration
Temp.
Range
0 to 100˚C; 32 to 212°F
Resolution
0.1˚C; 0.1/1°F
Accuracy
±0.5˚C ±1 digit
Others
Data Storage
50 groups
Storage Content
Numberings, Measurement, Unit, Temperature, Temperature
Compensation Status
Power
DC9V/300mA
IP Ratings
IP54 dustproof and spillproof
Size and Dimension
(240*235*103) mm/1kg
3 Ins tru ment Des cr ipt ion
3.1 LCD Display
(1) — Measurement mode
(2) — Measurement value
(3) — Calibration icons and numberings
(4) — Measurement unit
(5) — Temperature unit
(6) — The unit displayed during calibration
(7) — The value displayed during calibration, numberings for storage and recall, and reminder icons
(8) — Icons for data storage and recall — M+: store measuring data; RM: Recall measuring data
(9) — Temperature and reminder icons
Figure-1
Loading...