Powers the meter on and off . Upon power on, the meter
automatically begins in the mode that was last used. Calibration
and memory values are retained even if meter is unplugged.
Toggle between available measurement modes; pH/Temp,
mV/Temp, or ppm/mV (Ion 700 only).
Also used to switch from pH to Temp during pH calibration
mode. Press and hold for 5 seconds to enter SETUP mode.
Toggles between measurement and calibration modes.
In SETUP mode, returns user to the measurement mode.
MI (Memory Insert) stores values into memory.
Increase value. Scroll up in SETUP mode.
MR (Memory Recall) recalls values from memory
Decrease value. Scroll down in SETUP mode.
Freezes measured reading. Press again to resume live reading.
Confi rms calibration values in CAL mode.
Confi rms selections in SETUP mode.
View recalled values in memory mode.
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Part of Thermo Fisher Scientifi c
n
LCD Annunciators
Measured
value is stable
During setup,
select auto hold
ON, ready ON or
Ready OFF
Error or invalid
key pressed
Prompts user
to select bu er
Setup
mode
Electrode error
when appears
with ERR
Measurement
mode
Primary
reading
14
Calibration
mode
Secondary
reading
Memory mode or
when data is stored
into memory
Measured
value is held
Current
measurement
unit / mode
of operation
Automatic
temperature
compensation
is active
During pH CAL,
secondary reading is
the auto recognized pH
bu er standard value
Page 2
n
Meter Connections
BNC connection for pH, ORP (Redox), or
pH
Ion Selective Electrodes (ISE).
Pin connection for half cell reference electrodes.
DC
ATC REF.
pH
Requires separate half cell BNC electrode.
REF
Note: REF is not commonly used and is not required.
For Automatic Temperature Compensation probe.
ATC
Power supply.
DC
CALIBRATION
n
pH Calibration
For best results, periodic calibration with known accurate standards is recommended. The 700
series meters can be calibrated with up to 5 buffers. The non-volatile memory retains all calibration
values upon meter shut down.
The following calibration standards are automatically recognized;
USA buer group1.68, 4.01, 7.00, 10.01, 12.45
NIST buer group1.68, 4.01, 6.86, 9.18, 12.45
To eliminate temperature errors associated with the pH electrode, attach the automatic
temperature compensation (ATC) probe for best accuracy.
1. Press
2. Dip the pH and ATC electrodes into pH buffer and press
as needed to select pH.
. The secondary display will lock on
the appropriate buffer value. Provide stirring for best results.
When the READY indicator appears, press
to accept. The primary reading will flash briefly
before the secondary display begins scrolling the remaining available buffers.
3. Rinse the pH and ATC electrodes then dip into the next pH buffer. The secondary display will
lock on the appropriate buffer value. When the READY indicator appears, press
to accept.
The primary reading will flash briefly then display the percent efficiency (slope) before the
secondary display begins scrolling the remaining available buffers.
4. To calibrate another buffer repeat step 3 or press
to return to the measurement mode.
Page 3
n
Temperature Calibration
The thermistor sensor used for automatic temperature compensation and measurement is
both accurate and stable. Temperature calibration is recommended upon electrode
replacement, whenever the temperature reading is suspect, or if matching against a certified
thermometer is desired.
C
1.
onnect the temperature probe to the meter and place into a solution with a known accurate
t
emperature such as a constant temperature bath.
N
ote: To adjust the manual temperature compensation (MTC) value, do not connect the
t
emperature probe.
2. Press
3. Press
secondar
4. Adjust the temperature using
allo
n
Millivolt (mV) Oset Adjustment
as needed to select pH or mV/R.mV.
followed by . The primary display shows the measured temperature while the
y display shows the factory default temperature.
or . Press to accept or to cancel. The meter
ws an adjustable maximum value of ± 5
º
C (or ± 0.9 ºF) from the factory default temperature.
Oxidization Reduction Potential (ORP or Redox) is not a precise measurement, but is useful as a
relative indicator. mV offset adjustment makes readings comparable to a reference.
1.
onnect an ORP electrode and press
C
2.
Dip the ORP elec
3. Press
trode into a solution with a known mV value and stir.
when the reading is stable. The primary display shows the relative millivolt value
as needed to select mV (or R.mV).
(R.mV) while the secondary display shows the factory default mV value.
4. Adjust the R.mV value using
or . Press to accept or to cancel.
The meter allows an adjustable maximum value of ± 150 mV from the factory default
mV value.
Note: When an offset has been stored successfully, R.mV replaces mV.
32
Page 4
n
Ion Calibration (Ion 700 Only)
The available ION calibration values are 0.10, 1.0, 10.0, 100.0, and 1000 parts per million (ppm).
Choose a minimum of 2 consecutive values for calibration and prepare the corresponding ion
calibration solutions. For best results always begin with your lowest standard value, followed by
the next lowest standard.
1. Connect the ISE and press
“- - - -” without a stored ion calibration. The secondary value is the corresponding mV reading
of the ISE.
2. Dip the ISE into your lowest standard solution value and stir. Press
reading is stable.
3. Press
value (0.10, 1.0, 10, etc).
4. Press
standard value.
5. Rinse the ISE then dip into the corresponding calibration standard and stir.
6. Press
calibration is successful. “SLP Err” indicates calibration for the current point was not successful.
7. Revert to step 3 to calibrate additional points or press
mode at any time.
or to match the primary display to your corresponding ion calibration
to accept or to abort. The primary display will show the next highest calibration
to accept or to cancel. The mV/decade slope value will be displayed briefl y if the
as needed to select ppm. Note: The primary reading will show
when the secondary
as needed to return to measurement
STORING AND RECALLING DATA
The 700 series meters can retain up to 100 points into memory for later retrieval.
1. In the measurement mode, press
The stored memory location value (StO) is briefl y displayed.
to insert the measured value into memory.
2. To recall data from memory, press
fi rst. Press
3. Press
For more information on our products, please contact our channel partner or visit our websites listed below:
or to select the location of the desired data, then press to accept.
to return to the stored data location. Press to return to measurement mode.
Visit us at www.TestEquipmentDepot.com
. The location of the most recent stored data is displayed
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.