Oakton 35660-92 Instruction Manual

Gel-filled pH sensor cartridge
CAUTION
Plug
O-ring
Nut
Liquid junction
Guard tube
Sensor body
Responsive glass membrane
Temperature sensor
Note
Note
Sensor head
Plug
Nut
35660-92
CODE:M003860-3200839746-GZ0000590201
December, 2019
Conformable standards
This equipment conforms to the following standards:
RoHS: EN50581
9. Monitoring and control instruments
Authorised representative in EU
Cole-Parmer UK 9 Orion Court, Ambuscade Road Colmworth Business Park St Neots Cambridgeshire PE19 8YX, United Kingdom Phone: +44-(0)1480-272279 Fax: +44-(0)1480-212111
Caution on use
Safety precautions
Chemical solution
If the internal material comes in contact with the hands or skin, wash immediately with water.
Glass fragments
Glass fragments can cause injury. The tip of the sensor cartridge are made of glass. Be careful not to break them.
Points of concern
Do not give an excessive shock to the sensor cartridge. Cracked or chapped sensor cartridge can not be used.
Replace such sensor cartridge with a new one.
Avoid continuous measurement soaking in the strong
acidic or alkaline solution concentrated more than 0.01 mol/L. These solution may reduce the performance and the lifetime of the
sensor cartridge.
Do not to use the sensor cartridge in a solution in which
hydrofluoric acid
(HF) is contained. Glassware will
corrode in the HF solution.
Avoid sudden temperature change over 50 °C on the
sensor cartridge. The the glassware might be cracked.
Solid particles of silver chloride might be observed
sensor body might be deformed or
floating in the inner solution of the responsive glass membrane. This is not a problem and this will not affect performances of the
sensor cartridge.
Do not use (store) the sensor cartridge in any place
where
the operation temperature (storage temperature)
is out of the specified range.
When disposing of the product, follow the related laws
and/or regulations of your country for disposal of the product.
Specifications and parts description
Parts name
Specifications
Measurement range pH 0 to 14
Operation temperature range 0°C to 80°C
Storage temperature range 0°C to 50°C
Liquid junction material Porous sintered polyethylene
Wetted material
Wetted parts dimensions Outer diameter 16 mm × Length 110 mm
Liquid junction height
Zero point pH7.0
Glass, silicone rubber, polycarbonate, polyethylene, Polyolefin
Approximately 15 mm (from the tip of sensor cartridge)
Installation procedure
1. Remove the rubber cap of the plug.
2. Connect the slot of the sensor cartridge with the marker of the sensor head and push into the sensor head.
3. Tighten the nut firmly until the nut is totally tight on the sensor head.
Push the plug into the sensor head keeping connecters dry.Avoid using a rusted plug.In the absence of O-ring or in the case where the nuts is not
tight enough, liquids might flow into the connection space and cause malfunction.
After replacement of sensor cartridge, perform calibration
before measurement. Refer to the instruction manual of the meter.
Maintenance and storage
Refer to the instruction manual of the meter for details.
Do not dry the responsive glass membrane and the liquid
junction.
Keep the liquid junction away from the water for storage.
Otherwise, the sensor cartridge life span will be shortened.
Oakton Instruments
625 E. Bunker Ct. Vernon Hills, IL. 60061, USA E-mail: info@4oakton.com Phone number: (1)888-462-5866
For any questions regarding this product, please contact your local agency, or inquire from the following website. http://www.4oakton.com/