Gel-filled pH sensor cartridge 35660-92
Conformable standards
This equipment conforms to the following standards: RoHS: EN50581
9.Monitoring and control instruments
Authorised representative in EU
Cole-Parmer UK
9 Orion Court, Ambuscade Road Colmworth Business Park
St Neots Cambridgeshire
PE19 8YX, United Kingdom Phone: +44-(0)1480-272279 Fax: +44-(0)1480-212111
Caution on use
Safety precautions
CAUTION
Chemical solution
If the internal material comes in contact with the hands or skin, wash immediately with water.
Glass fragments
Glass fragments can cause injury.
The tip of the sensor cartridge are made of glass.
Be careful not to break them.
Points of concern
Do not give an excessive shock to the sensor cartridge.
Cracked or chapped sensor cartridge can not be used. Replace such sensor cartridge with a new one.
Avoid continuous measurement soaking in the strong acidic or alkaline solution concentrated more than 0.01 mol/L. These solution may reduce the performance and the lifetime of the sensor cartridge.
Do not to use the sensor cartridge in a solution in which hydrofluoric acid (HF) is contained. Glassware will corrode in the HF solution.
Avoid sudden temperature change over 50 °C on the sensor cartridge. The sensor body might be deformed or the glassware might be cracked.
Solid particles of silver chloride might be observed floating in the inner solution of the responsive glass membrane. This is not a problem and this will not affect performances of the sensor cartridge.
Do not use (store) the sensor cartridge in any place where the operation temperature (storage temperature) is out of the specified range.
When disposing of the product, follow the related laws and/or regulations of your country for disposal of the product.
Specifications and parts description
Parts name
|
Plug |
|
|
|
|
Sensor body |
|
O- |
|
Guard tube |
|
Nut |
|
Responsive glass |
|
Liquid junction |
membrane |
||
|
Temperature sensor
Specifications
Measurement range |
pH 0 to 14 |
|
|
|
|
Operation temperature range |
0°C to 80°C |
|
|
|
|
Storage temperature range |
0°C to 50°C |
|
|
|
|
Liquid junction material |
Porous sintered polyethylene |
|
|
|
|
Wetted material |
Glass, silicone rubber, polycarbonate, |
|
polyethylene, Polyolefin |
||
|
||
|
|
|
Wetted parts dimensions |
Outer diameter 16 mm × Length 110 mm |
|
|
|
|
Liquid junction height |
Approximately 15 mm |
|
(from the tip of sensor cartridge) |
||
|
||
|
|
|
Zero point |
pH7.0 |
|
|
|
Installation procedure
1.Remove the rubber cap of the plug.
2.Connect the slot of the sensor cartridge with the marker of the sensor head and push into the sensor head.
Sensor head
Plug
Nut
3.Tighten the nut firmly until the nut is totally tight on the sensor head.
Note
Push the plug into the sensor head keeping connecters dry.
Avoid using a rusted plug.
In the absence of O-ring or in the case where the nuts is not tight enough, liquids might flow into the connection space and cause malfunction.
After replacement of sensor cartridge, perform calibration before measurement. Refer to the instruction manual of the meter.
Maintenance and storage
Refer to the instruction manual of the meter for details.
Note
Do not dry the responsive glass membrane and the liquid junction.
Keep the liquid junction away from the water for storage.
Otherwise, the sensor cartridge life span will be shortened.