You have just purchased one of the finest Laboratory Centrifuges available. With proper care,
maintenance and laboratory procedure, this centrifuge will provide you years of productive service. Please
read this manual carefully to familiarize you with proper installation, maintenance and operation of the
centrifuge. Other reference and guideline materials are available on our website, www.nuaire.com.
The NUWIND 3.0 is a laboratory benchtop series of centrifuges that maximizes productivity while minimizing laboratory
required bench space. The centrifuge functionality allows the separation of substances of different densities held in
suspension or emulsion in a liquid using centrifugal force.
The NUWIND 3.0 centrifuge is available in two models, one ventilated and the other refrigerated.
The NU-C300V ventilated model has a unique ventilation system to maximize performance within laboratory ambient
temperature conditions.
The NU-C300R and NU-C300RF refrigerated model adds precise temperature control to further maximize performance.
The refrigeration system refrigerant used is CFC-free to meet regulations of refrigerant use.
NUWIND Centrifuge
Centrifuges NUWIND 3.0: ventilated and refrigerated models
Intended use
The centrifuge is designed and manufactured for use with rotating accessories supplied by NuAire, Inc. It is suitable for
separating substances of different densities held in suspension, or emulsion in a liquid.
The maximum density of substances should be 1.2 g/ml.
The centrifuge must be used in an optimal state of operation and maintenance. The non-observance of the rule of use may
threaten the health of users or third persons, but it may also cause damages to the unit and/or surrounding equipment.
Users of the equipment must be trained on good centrifugation practices, as well specific usage of the NUWIND models and
Accessories. Any use outside the framework defined above is considered non-compliant.
NuAire, Inc. assumes no responsibility for injury due to improper use.
The operator must observe the following precautions when using the centrifuge:
• Ensure stability and perfect levelling of the centrifuge.
• Check the correct mounting of the rotor and its accessories before starting a centrifugation cycle.
• Check tube resistance at maximum applied centrifugal force: chemical resistance to centrifuged products and
mechanical strength at the centrifugal force applied thereto.
• Check the condition of the tubes and remove damaged tubes.
• Use only the rotors and accessories authorized by the manufacturer NuAire, Inc.
• Swing-out rotors: install four (or two) buckets; never perform any cycle with missing buckets.
• Maintain and control accessories.
• Balance the load of the rotor around the axis of rotation.
• Respect the maximum density of 1.2 g / ml, in particular in case of a cycle at full speed.
• Limit volume in case of excess density.
• Install the centrifuge in a ventilated area, on a horizontal rigid support to absorb the vibrations generated by the
centrifuge.
• During operation, the centrifuge must not be moved or be subject to impact.
• Do not attempt to open the lid while the rotor is spinning.
• Do not attempt to neutralize the motorized lid lock
• Do not lean over the centrifuge during the spin cycle.
• Do not stay within a space of 10 inch around the centrifuge longer than necessary.
• Do not leave potentially dangerous materials inside the free space.
• Use aerosol-barrier accessories when centrifuging bio hazardous material.
• Condensation may form inside the centrifuge when it is moved from a cold environment to a warm environment.
Allow the centrifuge to warm up for two hours before use.
Prohibitions:
The centrifuge is not designed for the following uses and environments.
The following are prohibited:
• Flammable, explosive, toxic and radioactive materials
• Materials which may react, causing a hazard
• Contaminated materials not contained in an aerosol-barrier container
• Rotating rotors and accessories that have exceeded their maximum use.
• Rotating rotors and accessories that are not maintained according to the instructions in this manual,
or showing signs of wear and/or corrosion.
• Rotating rotors and accessories not supplied by NuAire, Inc.
• Explosive environments
• Radioactive environments
Only operate a centrifuge after taking all necessary safety measures.
Rotors and buckets should be removed from use in case of mechanical fault
trace, or corrosion. These elements have a life cycle duration, engraved on
their visible part: To maintain safe conditions, it is imperative to replace them
when the recommended duration is reached.
Environment Temperature Range: : 60⁰F - 85⁰F (15⁰C - 30⁰C)
Environment Humidity: Maximum relative humidity 80% for
temperatures up to 31⁰C decreasing linearly to 50% relative humidity at 40⁰C
13,000 xg 16,660 xg 16,660 xg
3413 BTU/h 5801 BTU/h 5801 BTU/h
20⁰C ± 2⁰C20⁰C ± 2⁰C20⁰C ± 2⁰C
21.3 x 29.1 x 30.3 / inches
Motorized lid lock
21.3 x 35.4 x 31.5 inches
Motorized lid lock
NU-RX750 rotor can be
maintained at 3,600RPM at an
ambient temperature of
39.4 x 31.5 x 31.5 inches
Motorized lid lock
NU-RX750 rotor can be
maintained at 3,600RPM at an
ambient temperature of
NuAire, Inc. warrants NU-C300 Series NUWIND Centrifuge that it will repair F.O.B. its factory or furnish without charge
F.O.B. its factory a similar part to replace any material in its equipment within 60 months after the date of sale if proved to
the satisfaction of the company to have been defective at the time it was sold provided that all parts claimed defective shall
be returned, properly identified to the company at its factory, charges prepaid. Factory installed equipment or accessories
are warranted only to the extent guaranteed by the original manufacturer, and this warranty shall not apply to any portion
of the equipment modified by the user. Claims under this warranty should be directed to NuAire, Inc. setting forth in detail
the nature of the defect, the date of the initial installation and the serial and model number of the equipment.
This warranty shall not apply to any NuAire product or part thereof which has been subject to misuse, abuse, accident,
shipping damage, improper installation or service, or damage by fire, flood or acts of God. If the serial number of this
product is altered, removed or defaced as to be illegible, the Warranty shall be null and void in its entirety.
The warranty is for the sole benefit of the original purchaser and is not assignable or transferable. Prior to returning any
item, for any reason, contact NuAire for a Return Authorization Number. This number must accompany all returns. Any
product shipped to NuAire without this number will be returned refused shipment or collect freight.
4.0 Shipments
NuAire takes every reasonable precaution to assure that your NUWIND Centrifuge arrives without damage. Motor carriers
are carefully selected and shipping cartons have been specially designed to insure your purchase. However, damage can
occur in any shipment and the following outlines the steps you should take on receipt of a NuAire NUWIND Centrifuge to be
sure that if damage has occurred, the proper claims and actions are taken immediately.
4.1 Damaged Shipments
4.1.1 Terms are factory, unless stated otherwise. Therefore, it is important to check each shipment before acceptance.
4.1.2 If there is visible damage, the material can be accepted after the driver makes a notation on the consignee's copy
of the freight bill. Then an inspection must be made to verify the claim against the carrier. This inspection is the
basis of your filing the claim against the carrier.
4.1.3 If concealed damage is found it is absolutely necessary to NOTIFY THE FREIGHT AGENT AT ONCE and request an
inspection. Without this inspection, the transportation company may not accept a claim for loss or damage. If the
carrier will not perform the inspection, an affidavit must be prepared stating that he was contacted on a certain
date and that he failed to comply with the request. This along with other papers in the customer's possession will
support the claim.
The package containing the centrifuge needs to be handled by suitable means (pallet truck, etc.).
Please ensure that the people in charge of handling are qualified to handle the lifting equipment.
Maintain the following conditions during transport:
• Temperature: -20 to 50 ° C
• Relative humidity: <90%, under the dew point
5.2 Unpacking
Remove the cardboard packaging made of 2 parts (a lid and a bottom).
The NUWIND centrifuge was packed with great care to prevent transport hazards. Before disposing of the packaging of the
centrifuge, and its spinning accessories, please make sure you are in possession of the following:
1. Insert the allen wrench in the hole on the left side panel to manually
sitting at the bottom
1. Insert the release key into the hole in the front
is
1 2 3
2 13
2
1 3 5 4 7
6
The rotor and the 4 buckets may have been packed inside the centrifuge in order to reduce the volume of packing
material. However, before any utilization, make sure to remove the packing foam before turning the unit on for the
first time.
These packing foams allow the rotor to be centered inside the bowl without any constraint or forces applied directly
on the motor shaft or without damaging any internal part in the bowl.
open the centrifuge
2. Turn it clockwise until the lid opens up
3. Remove the allen wrench from the hole
4. Remove all the packing foam inside the bowl
5. Remove the buckets and put them on a table next to the centrifuge
6. Remove the rotor with both hands and put it on a table
(it shouldn’t be locked on the shaft) next to the centrifuge
7. Remove all the different layers of packing foam
8. If needed, clean the bowl if any residual packing material or dust is
5.2.3 Description
2.Turn down until the lid opens up
3.Remove the key from the hole
4. Remove all the packing foam inside the bowl
5. Remove the buckets and put them on a table next to the centrifuge
6. Remove the rotor with both hands and put it on a table
(it shouldn’t be locked on the shaft) next to the centrifuge
7. Remove all the different layers of packing foam
8. If needed, clean the bowl if any residual packing material or dust
Prepare a stable, clean and levelled support (e.g. laboratory bench). This support must be able to support the weight of the
centrifuge and its use, and be firm enough so as not to generate or maintain vibration. Such vibrations would degrade the
results of the centrifugation, and may result in an unbalanced error.
A clearance of 10 inch (250mm) around the centrifuge should be
arranged according to IEC 61010-2-020: No potentially hazardous
material, nobody or no object shall be deposited within the free
space.
Open space around the centrifuge
It is recommended that no less than two people are present
to manually lift the centrifuge onto the laboratory bench.
5.4 Electrical environment
The centrifuge requires must be 230V/50Hz or 120V/60Hz, single phase, (current rating varies per centrifuge model,
reference Electrical/Environmental Requirements).according to the manufacturer sheet on the back of the unit.
The centrifuge must be plugged into an outlet with protective earthing connection with the standard power cord.
The electrical outlet into which the centrifuge is connected should be readily accessible for maintenance purposes.
Do not position the centrifuge to prevent access to the power cord.
The power cord plug serves as the disconnect and should remain readily accessible. The electrical outlet should be on its
own branch circuit, protected with a circuit breaker at the distribution panel near the centrifuge.
Position the 4 numbered buckets in their corresponding positions
and/or the centrifuge.
Loading the buckets
Loading the rotor
Forbidden:
Asymmetrical loading
2
1
1. Touch screen for direct access to various functions.
2. Single button for easy operation.
This button allows the validation, the start of a cycle, but also
the interruption of a cycle, and the output of a menu.
The functions accessible through this single button are
activated as required by the microcontroller.
6.1.5 Opening the lid
On the first start, the lid may not open right away due to a long storage in a cold environment.
In this case, warm-up the jacks by repeated manual shifting of the lid downwards.
To open the lid, press the button .
6.1.6 Rotor loading and balancing
Allowed:
on the rotor. Each bucket should pivot freely in its slot.
Place the inserts corresponding to the tubes to be centrifuged in
the four buckets
The buckets must be symmetrically distributed.
For better balance, each numbered bucket must be placed in the
corresponding position marked on the rotor.
Each position on the rotor must be equipped with its bucket.
Do not use a rotor without its complete set of four buckets.
Maximum loads according to the type of rotor are specified in the
annex.
The centrifuge tolerates an imbalance of 15grams.
If the imbalance is higher than the tolerated threshold, an
imbalance detection system stops the centrifuge automatically.
The following message appears: "Imbalance" and the rotor stops
in free wheel mode without braking system.
Excessive imbalance is likely to damage the turning components
From an Admin profile, a program can be locked to prevent any modifications of the settings
Select the program to lock
Press on the open lock for 4 seconds.
The lock is now closed.
The program is now protected.
Note: To remove the protection, enter the
program through an Admin user
profile and press on the closed lock
for 3 seconds.
It will then re-open. You are now
able to modify the program.
In the list of programs, a lock appears in
front of the program
This operation must be performed by a person trained in centrifugation-
1. Switch off the power to the unit
2 13
1
2
3
CAUTION
!
During a centrifugation, press the Start/Stop
button.
The rotor will start slowing down until
complete stop
Open the lid.
6.2.8 Open after an electric power failure / Emergency Stop
related risks.
In case of power failure, the lid cannot be opened by the engine unlocking
control.
You must wait 15 minutes for the rotor to come to a complete stop before
opening manually.
2. Insert the unlocking key into the hole on the left side panel
Following a manual release, the wrench must be removed from its housing to prevent accidents related to the rotation of the
Visual
Instruction
Control
Select the pre-cooling program:
Rotor: Standard Precool
This program is set to cool the centrifuge
cannot be modified.
Air: Smart Precool
This program is set to cool the centrifuge
will stop when the temperature is reached.
wrench during the activation of the locking mechanism. In addition, the security indicator label must be replaced during a safety
check carried out by an authorized technician.
6.2.9 Precooling
On the Refrigerated model, a "Pre-cooling" cycle prior to centrifugation allows to cool the centrifuge down to
4 degrees Celsius so it is ready for temperature sensitive samples.
Access to the list of programs by pressing
this icon :
and accessories down to 4°C during a 22
minute cycle at 2500 RPM. Acceleration and
Deceleration slope are both set on 9 and
until the probe temperature reaches 4⁰C and
*Select Rotor or Air in user preferences
Start the cycle by pressing
The temperature inside the bowl will show
on the display
This menu is only accessible to certified service
companies.
Exit
Access the menu through this icon :
the Software and Hardware of the
centrifuge :
SERIAL NUMBER: Unique identification
number of the centrifuge
Centrifuge fw: Software version of the
microcontroller
Display fw: Software version of the display
µController Board Id: ID of the
microcontroller electronic board
Locker LT: Locking tempo of the lid
Locker UT: Unlocking tempo of the lid
Imbalance: Imbalance calibration setting
All maintenance actions on this equipment must be performed by a qualified
technician who is familiar with the proper maintenance procedures required
for this equipment
The device and its accessories may be contaminated.
Apply decontamination measures required before any maintenance.
Turn off the centrifuge power switch. Ensure you lock the switch when working on the
unit, to avoid any accident from a third party.
Special care must be provided to maintain the original terms of safety and performance,
to extend the life of accessories. A visual inspection can identify any signs of wear. An
annual inspection may be requested according to relevant regulations
7.1 Care and Cleaning
7.1.1 Authorized maintenance products
Users should not use cleaning or decontamination methods different from those recommended by the
manufacturer, as they may cause damage to the unit.
The following products and materials are allowed for maintenance:
- Distilled water,
- 70% IPA, peroxides or quaternary ammonium surface disinfectants.
Perform regular cleaning of the following parts to prevent any risk of corrosion in case of persistent impurity, and to
ensure the hygiene of the working equipment.
External parts lid & keyboard:
- Use fluids and tissues impregnated with hydro-alcoholic solution or quaternary ammonium
To ensure a proper swinging movement of the buckets,
Fig. 2 Trunions
Remove the leftover grease and dust.
Also clean the buckets, more specifically the area in contact
Use clear or distilled water to clean the rotor and buckets. A soft non-metallic brush or lint-free cloth can be used.
Dry the rotor and accessories ensuring the least accessible areas are wiped.
clean all 8 trunions of the rotor.
Apply a thin layer of food quality grease (provided).
Make sure to set the “greasing reminder” in the User Preferences menu.
To prolong the life of accessories, you can then place a thin film of silicone spray or lanolin.
Rotors, buckets, inserts can be autoclaved at 122.5 ° C / 252 ° F / 215 kPa for 15 minutes
Note: When autoclaving, buckets, inserts, and rotors must be separated from each other.
Autoclaving causes accelerated aging of plastics, and may change the color of the plastic.
Perform regular cleaning to prevent any risk of corrosion in case of persistent impurity,
and to ensure the hygiene conditions of working equipment.
7.1.4 Disinfection / Decontamination
Bio hazardous, radioactive and/or toxic materials:
Before any intervention, to ensure the safety of the operator, please refer to the laboratory’s decontamination
protocol. The centrifuge and its accessories may be contaminated and require remediation in this case.
When there is only 2,000 cycles left before the end
3
2
1. Open the lid.
2. Grab the red wheel on top of the rotor and turn it
counterclockwise while holding the rotor firmly. Once you hear
a “click” sound, the rotor is unlocked
3. Grab and lift the rotor with both hands once it’s unlocked.
4. Set the rotor on a flat surface or on one of NuAire’s rotor stand.
7.1.6 Limiting the duration of operation
The use of accessories and the rotors is limited in time to ensure the integrity of their mechanical strength.
These limits are marked on the accessories.
of the theoretical life time of the rotor, this screen
will be displayed. It shows the amount of cycles left
before the rotor should stop being used (rotor and
buckets). This value will decrease by 1 each time the
rotor is being used.
You can still use the centrifuge, but only as long as
the value is higher than 0. Anticipate the purchase of
a new set of rotor + buckets in the near future.
Once the rotor has reached the end of its theoretical
life time, the screen will inform you that the rotor
cannot be used.
The rotor and buckets are obsolete and have to be
replaced promptly. A certified technician has to
configure the centrifuge for the new rotor.
Check that no trace of corrosion appears. Check also the absence of any
mechanical damage (traces of impact, corrosion, cracking), or wear on the
rotor and bucket.
In order to ensure optimal performances, once the accessories have reached half of their theoretical life time, we
recommend inspecting them. This life time is different from one accessory to the other, and the value is engraved on
the accessory (i.e.: Max cycles: 55,000 means that the accessory has a theoretical life time of 55,000 cycles)
For safety reasons, the use of accessories that have reached the end of their life cycle is prohibited. These
components must be replaced.
7.1.8 Storage of accessories
Do not store the accessories on a wire rack. Oxidation build-up may appear and show premature wear of accessories
7.1.9 Annual maintenance
A periodic check of the centrifuge and accessories is required. Checks must also be carried out in accordance with the
regulations in force: Functional (lubrication of the drive shaft, the lock, readability of labels), security (continuity of
the protective ground connection, checking & maintenance of rotors and accessories) and centrifuge performance
checks are recommended at least once a year.
Replacement of consumables (grease, provided)
Check that the grounding of the device is always properly secured by its power cable and the wall outlet it is
connected to. The operator is not authorized to access the internal part of the equipment. A NuAire authorized
technician will ensure maintenance.
11. Press on the Next button to average the setting.
Check
The following steps are compulsory.
High threshold:
maximum threshold of 25gr.
12. Place the 15 gr + 10 gr weights in a nacelle.
Low threshold:
15. Leave only a weight of 15 gr in a nacelle.
should take place normally.
The centrifuge has a reaction to the load balancing faults which may be different depending on the media on which the
centrifuge is placed. A calibration of the imbalance sensitivity is therefore necessary during installation.
Required equipment: Use the reference balancing kit AFI-71122002.
Access the unbalance sensitivity calibration menu.
Equip a rotor swing-out and its 4 nacelles in the
centrifuge.
Remove the inserts of the nacelles.
Do not adjust this threshold with a weight
greater than 25 gr.
This would cause a breach of warranty.
The appliance has been dimensioned for a
nacelles and close the lid,
2. Press the button “15gr”
3. The rotor accelerates until it stabilizes.
4. The centrifuge shakes, then the rotor slows down.
5. The low threshold is then determined. It appears
under the button “15 gr”. (e.g. 480)
8. Proceed in the same way as for the 15 gram weight.
9. The low threshold is then determined. (e.g. “520”)
10. Open the lid and remove the 25 gram weight.
(e.g. [480+520]/2 = 500 )
The average value is displayed.
By default, the factory setting value is 500.
13. Start the cycle. 2000 rpm / 1 minute / 9
Acceleration / Braking 9
14. The centrifuge must stop during the acceleration,
and display the message ERROR1:
The 25 gr imbalance is not tolerated.
Repeat the High threshold check 3 times: ERROR1
must appear each time.
16. Start the cycle. 2000 rpm / 1 minute /
9 Acceleration / Braking 9
17. The centrifuge should achieve the desired speed
Problem of communication with the
temperature measurement.
The lid was manually opened or the hook was
Contact a Certified Technician.
Verify the power plug is connected properly.
Turn the switch on.
Verify that the temperature probe is properly
PT500)
ERROR OVERSPEED
SAFETY
The startup test of the overspeed safety has
failed.
Microcontrol board is defaulting: Replace the
control center.
07
NOT SEEN – CLOSED
The open position switch was released after the
lid closed.
The micro contact doesn’t receive a command.
The coil of the micro contactor is not working
The micro contact was seen as open when it
should have been seen as closed.
Contact a Certified Technician.
working improperly.
08 ERROR LID SAFETY
The connection to these parts is not good.
The sensor for the closed lid position was
NO SPEED SIGNAL
09
AT START
The board with tachometer / accelerometer is
working improperly.
The speed sensor is working improperly.
10 SPEED SIGNAL LOST
The board with tachometer / accelerometer is
working improperly.
The connection to these parts is not good.
The board with tachometer / accelerometer is
WRONG SPEED
11
SIGNAL
working improperly.
The connection to these parts is not good.
Magnets from the tachometer might have been
12 OVERSPEED Major error, the centrifuge will stop.
Bus I²C
13
Accelerometer
working improperly.
The connection to these parts is not good.
The lid slightly opened during the security auto
test when the cycle started.
The speed cannot be read within the first 5s of
the cycle.
Unit will be locked for 30 minutes. Acknowledge
error and after 30 minutes lid will be able to be
The speed signal is not correct.
maximum speed allowed for this rotor.
Problem of communication with the
accelerometer.
14
LID OPEN DURING
15
ROTATION
NO DISPLAY No power.
The microcontroller board is working improperly.
Manual lid opening during centrifugation.
Wrong calibration of locking mechanism.
released during the centrifugation.
Check the power in the laboratory.
installed under the lid and replace it if
TEMPERATURE
16
CONTROL ISSUE
Temperature probe out of service.
necessary.
Remove the probe and measure the resistance.
The value must be 500+ - 100 Ohm (Probe
Signal permanently lost:
Impossible to clear the Error and open the lid: 30
SPEED SIGNAL IS
18
LOST
SPEED SENSOR
19
ISSUE
minute safety timer is running,
Wait for 30mn before being allowed to open the
lid
Speed signal recovered during braking:
Error can be erased after full stop: lid can be
Signal permanently lost:
Impossible to clear the Error and open the lid: 30
minute safety timer is running,
Wait for 30mn before being allowed to open the
lid
Speed signal recovered during braking:
Error can be erased after full stop: lid can be
After erasing the error, cycle the power supply
and try to launch a new cycle. If Error 18 is
detected again, call technical service
After erasing the error, cycle the power supply
and try to launch a new cycle.
If Error 19 is detected again, call service
In case of malfunction, contact after-sales service if the previous table does not eliminate the error.
You will be asked for the type of centrifuge and the serial number for more effective troubleshooting.
Never try to disassemble or fix the unit on your own: the risks of getting injured or electrocuted are high. The manufacturer
warranty would then be void. Only a certified technician is authorized to perform such operations.
9.1 Electrical (Supply Voltage Fluctuations Not to Exceed +/- 10%)
9.2 Operational Performance (for indoor use only)
Environment Temperature Range: 60⁰F - 85⁰F (15⁰C - 30⁰C)
Environment Humidity: Maximum relative humidity 80% for temperatures up to 31⁰C decreasing linearly to
Standard Fluorescent Lighting @ 150 ft. candles (1614 LUX) maximum intensity.
9.4 Installation Category:2.0
Installation category (overvoltage category) defines the level of transient overvoltage, which the instrument is designed to
withstand safely. It depends on the nature of the electricity supply and its overvoltage protection means. For example, in
CAT II, which is the category used for instruments in installations supplied from a supply comparable to public mains such as
hospital and research laboratories and most industrial laboratories, the expected transient overvoltage is 2500 V for a 230 V
supply and 1500 V for a 120 V supply.
9.5 Pollution Degree: 2.0
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2
assumes that normally only non-conductive pollution such as dust occurs with the exception of occasional conductivity
caused by condensation.
9.6 Chemical Exposure
Chemical exposure should be limited to antibacterial materials used for cleaning and disinfecting.
CHLORINATED AND HALOGEN MATERIALS ARE NOT RECOMMENDED FOR USE ON STAINLESS STEEL SURFACES.
9.7 EMC Performance (classified for light industrial)
Emissions: EN61326
Immunity: EN61326
Class A equipment is intended for use in an industrial environment. In the
documentation for the user, a statement shall be included drawing attention to the fact
that there may be potential difficulties in ensuring electromagnetic compatibility in
other environments, due to conducted as well as radiated disturbances.
Before disposing the equipment, it must be decontaminated and cleaned to protect people, the environment and equipment.
Legal regulations must be observed when disposing of the unit.
Electrical appliances are governed by national regulations based on the 2002/96/EC European Directive
on electrical and electronic equipment waste WEEE.
According to this directive, no equipment supplied after August 13, 2005 in the Business-to-Business
sector - which includes this centrifuge - must be disposed of with household waste.
For easy identification, the appliances are marked with the symbol displaying a crossed out dustbin.
Centrifuges that are no longer in use and are ready for disposal contain reusable materials.
ALL components may be disposed and/or recycled after they are known to be properly disinfected.
Prior to any disassembly for disposal, the centrifuge must be decontaminated.
Note: Material type can be verified with use of a magnet with stainless and aluminum being non-magnetic.
Relative Centrifugal Force
In customary practices and usages of centrifugation, the quantization unit of angular velocity (denoted N), is the revolution
per minute (abbreviated rpm, RPM, r / min, or r • min-1). This is a measure of frequency of rotation. It corresponds to the
number of full rotations completed in one minute around a fixed axis.
The derived frequency unit for SI units is Hertz, with the symbol Hz. Its expression in terms of SI is s-1.
Laboratory centrifuges are used to separate solids of different densities in suspension, applying the relative centrifugal
force (RCF for short) on samples. The effective force increases with the square of the rotational speed and distance from
the axis of rotation. This force, known as "g-force" is quantified as the number of "g" applied to the sample. No SI units are
provided.
The following formula is used to convert the angular speed to "g-force"
RCF Relative Centrifugal Force, "g"
N: angular velocity, revolutions per minute
r: radius of the circular path of the sample, mm
g: acceleration of standard gravity, 9806 65 meters per second per second (ms-²)
The relative centrifugal force depends on the speed and the rotation radius.
An approximation, denoted F, is used in practice to calculate the value of the relative centrifugal force:
Speed, based on the relative centrifugal force is calculated by the following formula:
Integral spin
It is the accumulated relative centrifugal force acting on the centrifuged object at the corresponding time.
It is outlined by the colored surface of the chart below.
(*) The radius is the distance between the rotor shaft and the bottom of the insert holes when the bucket is spinning in a
horizontal position (during centrifugation)
Microplate carrier
Part number :
NU-BXDW4
Description:
Microplate carrier for NU-RX1000 rotor
(Set of 2 OR 4)
Maximum speed
3500 rpm
Maximum RCF
2655 xg
Microplate capacity
6 x 96 wells
Deepwell block capacity
1 x 96 wells
Maximum capacity :
800 grams
Lifetime
(Indicative duration)
50,000 cycles
8 years
Maximum Radius :
194 mm
Included :
Part number :
NU-TXDW4
Description:
Stainless Steel Extractor & Neoprene
(***)The radius is the distance between the rotor shaft and the bottom of the microplate carrier when the carrier is
spinning in a horizontal position (during centrifugation)
(*) The radius is the distance between the rotor shaft and the bottom of the inserts when the bucket is spinning in a
horizontal position (during centrifugation)
(**) Same insert for many types of tubes.
(***) Can only be used without a lid
Support micro plaque
Part number:
NU-BXDW3
Microplate bucket for rotor NU-RX750
(Set of 2 OR 4)
Maximum speed
4600 rpm
Maximum RCF
4564 xg
Capacity standard Microplates
4 plates per bucket (total of 16)
Capacity Deep well blocks
1 block per bucket (total of 4)
Maximum capacity
800 grams
50,000 cycles
8 years
Maximum Radius(***) :
193 mm
(***) The radius is the distance between the rotor shaft and the bottom of the microplate bucket (without pad and
extractor) when the bucket is spinning in a horizontal position (during centrifugation)