Ntron Yellow Box Specifications

Page 1
Yellow Box
Portable Oxygen Analyser
The Ntron Yellow Box is a compact,
intelligent, transportable Oxygen analyser
providing an affordable and versatile
solution for portable high purity oxygen
measurement applications.
Housed in a rugged high impact case and
with microprocessor based electronics,
makes the Ntron Yellow Box the
instrument of choice for demanding
applications where a portable, accurate,
fast and reliable High Purity Analysis is
required.
Applications
An Oxygen Analyser for use in:
• Glove box purge and leak detection
• Additive manufacturing
• Semiconductor wafer machines
• Wave soldering
• Membrane air separators
• Inert welding gases
• Inert ovens
• Medical gas applications
• Pharmaceutical industries
• Food and beverage industries
• Gas generation industries
Features
• Lightweight, rugged and highly portable
• Mains powered / Rechargable battery pack
• Dual Oxygen Analaysers for % and PPM Measurement
• Measurement ranges: 0 - 1000 PPM Oxygen 0 - 25% Oxygen 0 - 96% Oxygen
• Reliable, long life zirconia sensor technology with Electrochemical options
• Integral Flow Meter for positive pressure sample flow indication
• LCD Displays and 4 button multi-function user interface
EXCELLENCE THROUGH DIVERSITY
Page 2
Dual Oxygen Analysers
Two switch selectable Oxygen Analyser Modules give % and PPM range measurement respectively.
Zirconia Sensor Technology
At the heart of the Ntron Yellow Box Portable Analyser are two Solid State Zirconia Sensors for long life, high stability accurate ZRO2 measurement of % and PPM Oxygen Concentrations.
Minimal Sample requirements
Operating the Yellow Box Portable Analyser is made very simple by the internal sample system. Taking a gas measurement only requires the operator to connect the gas line, purging for a few seconds and switching the valve position. When not in use, the sensor is isolated from the ambient air and process sample.
Fast Response Time
Due to its reduced size and unique design, the Ntron Zirconia ZRO2 sensor has a T 90 of less than 10 seconds.
Page 3
Yellow Box
Product specifications
Gas Type Ranges
Oxygen / 0 - 1000 PPM Oxygen / 0 - 25% Oxygen / 0 - 96%
Sensor Type
Ntron Zirconia ZRO2 / Electrochemical
Sensor Life
>30,000 hours in 20.9% O (ambient air)
2
Response Time (T90)
<10 seconds. Warm up time 10mins from switch on.
Power Requirements
Optional internal rechargeable battery and charger. Charge via 3-pin IEC Mains connector @ 80 - 230VAC; 50Hz
Yellow Box
Environmental Rating
IP67 when case closed for transportation
Connection Ports
6mm ‘Bulkhead’ tube compression fittings
for sample in. Sample exhaust out via integral Flowmeter port.
Dimensions
410mm (h) x 340mm (w) x 190mm (d)
Weight
8kg
Application Setup / System Schematic
Flowmeter
Battery Level
Indicator
PPM & %
Analysers
% / OFF / PPM
Analyser Selector
Mains Power
Connector
Switch
www.ntron.com
Flow
Regulator
Sample In
Connection
Port
Sample In
Flow Regulator
Select
PPM
Analyser
Units
Select
%
Flowmeter
Sample
Exhaust
CONTACT: Mullaghboy Industrial Park, Navan, Co. Meath, Ireland PHONE:+353 (0)46907 1333| FAX:+353(0)46907 1331| EMAIL:sales@ntron.com