NSK VIVA ace Operation Manual

000
English
English
Thank you for purchasing VIVA ace Scaler Kit. Please read this Operation Manual and the VIVA ace Basic Set Operation Manual carefully before use for operating instructions and care and maintenance guidelines so that you can use it for many years to come through a correct use. Keep this Operation Manual within easy reach of users for future reference.
Contents
1. User and Indications for Use .....................2
2. Precautions for handling and
operation ..................................................2
3. Package Contents.....................................5
4. Preparation for Use ...................................6
4-1 Connecting the Scaler Cord and VIVA ace
Control Unit ..........................................................6
4-2 Connecting and Disconnecting
the Handpiece ...................................................... 6
4-3 Mounting the Tip ..................................................7
4-4 Using the Tip Card ................................................9
4-5 Check before treatment ...................................... 10
5. Operation Procedure ...............................10
6. Post-use Maintenance ............................10
6-1 Preparation for Maintenance ............................... 10
6-2 Maintenance for between each patient ................ 12
6-3 Maintenance after close .....................................13
6-4 Sterilization ........................................................29
6-5 Maintenance of the Water Line of the
Scaler Handpiece and the Scaler Cord ....................
(In case of Connecting the Control Unit) ..............31
7. Maintenance .......................................... 34
7-1 Replacing the O-ring ..........................................34
7-2 Periodical Maintenance Checks ..........................35
8. Troubleshooting ..................................... 35
8-1 Problems and Solutions ......................................35
9. Specifications ....................................... 38
9-1 Specifications .....................................................38
9-2 Classification of Equipment .................................38
9-3 Operation Principle .............................................38
9-4 Symbol ..............................................................39
10. After-sales Service ................................ 39
10-1 Warranty ..........................................................39
10-2 Disposing Product ............................................39
Operation procedure of this product is introduced in a movie.
Check the following URL or the QR code.
URL http://www.nsk-dental.com/support/videos/
*This product is used by connecting to VIVA ace Basic Set.
1
1
User and Indications for Use
User :
Dentist, Dental hygienist
Indications for Use :
The ultrasonic scaler is intended for use in dental applications such as scaling, root canal treatment, periodontal and cavity preparation by generating ultrasonic waves.
2
Precautions for handling and operation
Please read these precautions carefully and use only as intended or instructed. Safety instructions are intended to avoid potential hazards that could result in personal injury or
damage to the device.
Safety instructions are classified as follows in accordance with the seriousness of the risk.
Class Degree of Risk
WARNING
CAUTION
NOTICE
Hazard that could result in serious injury or damage to the device if the safety instructions are not correctly followed.
Hazard that could result in light or moderate injury or damage to the device if the safety instructions are not correctly followed.
General product specification information highlighted to avoid product malfunction and performance reduction.
2
Precautions for handling and operation
WARNING
Do not attempt to disassemble the product nor tamper with the mechanism except as recommended by NSK
in this Operation Manual. This may result in an injury, electric shock or fire.
Do not touch the connection of the Handpiece and the Scaler Cord. This could cause an electric shock.
Do not allow any impact on to the product. Do not drop the product. It could cause an electric shock and
damage.
Do not use the product when the patient or the operator is using cardiac pacemakers as there is a danger
that could affect the pacemaker.
Keep away from explosive substances and flammable materials. Also, keep away from patients whom a
flammable anesthetic (laughing gas) is administered as this may cause fire.
Electromagnetic waves may cause malfunction of the product. Keep away from devices that generate
electromagnetic waves. Turn off the product around ultrasonic generators (excluding the product) or electrode knife. Keep away from communication equipment or elevators that generate electromagnetic waves.
Apply water for regular use. Insufficient supply of water may result in generation of heat and damage on the
tooth surface.
English
CAUTION
When operating the product always consider the safety of the patient.
Use only as indicated.
The end user shall be responsible for any judgment that relates to the application of this product to a patient.
Users are responsible for the operational control, maintenance and periodical maintenance check of this
product.
Operators and all others in the area must wear eye protection, a mask, and gloves when operating this
product.
Should the product function abnormally, cease operation immediately and contact your Authorized NSK
Dealer.
Do not turn the lighting of the Handpiece or Scaler Cord directly to the eyes of the patients or the operators.
There is a danger that it may damage the eyes.
Do not clean, immerse, or wipe with electrolyzed-oxidizing water (strong acidic water, super acidic
water), strong acidic and strong alkaline liquid medicine, solvent containing chlorine, benzene, or thinner. (
Reference : 6 Post-use Maintenance
)
3
Precautions for handling and operation
CAUTION
The Handpiece, Tip, Tip wrench, and Tip Holder are delivered in a non-sterile condition and must be
autoclaved prior to use. (Reference : 6-4 Sterilization
Immediately wipe off any moisture, liquid medicine, solvent, or liquid disinfection on the Handpiece or the
Scaler Cord after use. Failure to do so may cause discoloration and deformation.
)
Perform regular function and maintenance checks. (Reference : 7-2 Periodical Maintenance Checks
If the product is not used for a long period, check for motor vibration, noise and overheating before using on
a patient.
Portable and mobile RF communications equipment may affect Medical Electrical equipment. Do not use RF
equipment near the product.
The temperature of the tip surface may reach up to 50 degrees depending on the condition of use. In that
case, leave the tip until it cools down before using it again.
)
NOTICE
The Handpiece and Scaler Cord during operation could interfere with the computers, LAN cables in vicinity of
use or could cause noise in radio receivers nearby.
To avoid clinical downtime, it is recommended that a spare is kept on hand in case of a breakdown during
surgery.
Contact the Authorized NSK Dealer for repairing. The service staff of the company will take care of the
product in accordance with the service manual.
No special training is required for this device.
EMC (Electromagnetic compatibility) has been evaluated with the VIVA ace Control Unit and conforms to the
requirements.
4
3
Package Contents
No. Part Name Order Code Quantity Remarks
1Handpiece E3510501 Optic
English
2 Scaler Cord Z1140055 1 Optic
3Tip
4Tip Wrench Z2210761 CR-10
5 Tip Holder Z221A080 1 -
6O-ring
7 Cleaning Wire Z144095 1 *Not in the illustration
8 Tip Card - 1 *Not in the illustration
9 Tip Guide - 1 *Not in the illustration
10 Power Guide - 1 *Not in the illustration
11 Operation Manual - 1 *Not in the illustration
Refer to the Tip Guide
D0310020080
2G6,G8
2For Scaler Cord
Please specify the order code listed above when purchasing the spare parts.
5
4
Preparation for Use
4-1 Connecting the Scaler Cord and VIVA ace Control Unit
No. Action Fig.
1
2
3
Aligning the mark of the Scaler
1
Cord plug and the shape of the terminal as shown below, insert the plug deep into the Scaler connector.
Turn the Plug Cover for tightening.
3
Put the Scaler Cord plug on the holder.
mark
4-2 Connecting and Disconnecting the Handpiece
Action Fig.
2
Loosen
6&$/(5
Tighten
02725
<Connecting>
Align the
mark on the Scaler Cord Plug and push
straight in.
mark on the Handpiece and the
<Disconnecting>
Holding the Handpiece and Scaler Cord Plug, pull straight apart.
Do not touch the connection (terminal section) of the Handpiece and the Scaler Cord. This could cause
anelectric shock.
Handpiece
WARNING
mark
Connection
Scaler Cord Plug
6
Preparation for Use
CAUTION
Insert the Handpiece firmly to the end of the Scaler Cord.
Do not connect Handpieces other than the attached Handpiece.
Remove the Tip before disconnecting the Handpiece to prevent injuries.
When connecting/disconnecting the Handpiece, be careful not to twist the Handpiece and the Scaler
Cord. The terminal at the connection could break and cause a failure.
4-3 Mounting the Tip
No. Action Fig.
Make sure that there is no dirt on the threads of the Handpiece and
1
threads
1
the Tip.
First, lightly tighten the Tip with
2
hand.
English
2
Get the hole of the Tip Wrench through the Tip, align the square part of the Tip and insert firmly.
3
Then tighten the Tip Wrench tight enough until hearing a clicking sound. Loosen it when removing the Tip.
Be careful not to touch the gum, mucous membrane, or the skin directly with the Tip to prevent injuries
and burns.
Be sure to use genuine Tips manufactured by NSK. Use of other products may cause problems listed
below and unexpected accidents.
Vibration problem caused by use of improper screws.
·
· Accidental ingestion caused by breakage of Tips manufactured by other companies.
· Handpiece breakage or early abrasion of the threads.
3
WARNING
Click, Click
Tighten
Loosen
7
Preparation for Use
CAUTION
Pay attention to the Scaler Cord not to be twisted when tightening the Tip.
Be careful of the pointed end of the Tip when replacing a Tip that is longer than the Tip Wrench.
Do not use a Tip which has a crack, a bend, or rust. The Tip may snap during use.
If the Tip Wrench is rotated at an angle, the threads may bend or vibration becomes insufficient.
If installed with the Handpiece or the threads of the Tip having dirt on, abnormal sound or weakening/
stopping of vibration may occur. Remove the dirt from the threads.
Do not sharpen or change the angle of the Tip. It could cause abnormal sound, Tip breakage, or stopping
of vibration during use.
Tip is a wear item. When it is worn, abnormal sound, weakening/stopping of vibration or breakage may
occur. When abnormal sound or vibration weakening is detected, replace the Tip with a new one.
Tighten the Tip with attached Tip Wrench until hearing a clicking sound. If the tightening is not enough, it
may cause abnormal sound or vibration weakening.
When replacing the Tip, be sure to wear sterilized gloves and use a sterilized Tip, Handpiece, and Tip
Wrench.
Be sure to use the Tip at or under the maximum power written on the Tip Case or the Tip Power Guide.
If it is used over the maximum power, a noise, breakage of the Tip, or damage to the tooth surface may occur.
Other than the prosthesis removing Tip (metal crown and porcelain crown), do not contact anything to the
prosthesis to prevent from coming off or being damaged.
When connecting/disconnecting the Handpiece, be sure to remove the Tip to prevent injuries.
If the Tip stops vibrating during use, take out the Handpiece from the patient’s mouth and step on the
foot control again. If the condition does not improve, check the connection of the Tip because there is a possibility that the Tip has become loose during use.
If looseness of the Tip, a noise, or vibration weakening does not improve even after tightening the Tip with
the Tip Wrench until hearing a clicking sound, the Tip Wrench might be worn. Replace the Tip Wrench after about one-year of use because it is a wear item.
When using a file as a Tip to enlarge the root canal, ensure you understand its characteristics and the
application area.
8
Preparation for Use
4-4 Using the Tip Card
Measure the length of the Tip end by comparing with the corresponding Tip illustration and check the replacement time. It is recommended to replace the Tip when the length of the Tip becomes “Yellow Line: 1mm worn” for safe and efficient use.
Green Line : Not worn. (
No replacement is needed.
Yellow Line : 1mm worn. (
Replacement is recommended.
Red Line : 2mm worn. (
Replacement is required.
)
)
)
English
CAUTION
Tip is a wear item, check the abrasion degree with the attached Tip Card and replace the Tip properly.
Tip is a wear item, the tartar removal
efficiency decreases by 25% when the Tip end is worn by 1mm, and 50% when the Tip end is worn by 2mm.
Also, the vibration condition may change and damage the tooth surface. Check the abrasion degree with the Tip Card regularly and replace the Tip with a new one early.
25% down
1mm
50% down
2mm
9
Preparation for Use
4-5 Check before treatment
Follow the instruction on operation manual for VIVA ace Basic Set. (Reference : 5-10 Check before treatment
5
)
Operation Procedure
Follow the instruction on the operation manual for VIVA ace Basic Set. (Reference : 6 Operation Procedure
6
Post-use Maintenance
6-1 Preparation for Maintenance
No. Action Fig.
Wear eye protection, a mask, and gloves to prevent
1
2
3
4
infection.
Turn off the Power Switch of the Control Unit.
Remove the Water Plug (IN) [Clear Tube] while pulling the Slide Ring.
Remove the Water Plug (OUT) [Blue Tube] pressing the Slide Ring.
3
Water Plug (IN)
[Clear Tube]
Slide Ring
)
5
6
7
Dispose of the water in the Water Bottle and place the empty bottle again in the Control Unit.
Turn on the Power Switch of the Control Unit.
Maximize Scaler Water Adjuster, activate the Scaler Handpiece to let out the water remaining in the Scaler Cord.
10
Post-use Maintenance
WARNING
Failure to properly maintain as below, the Scaler Handpiece and the Tip may cause infection, product
failure, overheating leading to burn injuries:
- Be sure to clean the Scaler Handpiece, and Tip immediately (within 30 minitues) after use to remove residue.
- Be sure to disinfect the Scaler Handpiece, Tip and the Tip Wrench, within 2 hours after cleaning.
CAUTION
Do not clean, immerse, or wipe with electrolyzed-oxidizing water (strong acidic water, super acidic water),
strong acidic and strong alkaline liquid medicine, solvent containing chlorine, benzene, or thinner.
Follow any additional local directives, standards, and guidelines for cleaning, disinfection, and sterilization.
NOTICE
English
Be sure to remove the Water Plug (IN) [Clear Tube] first.
Note that if the Water Plug (OUT) [Blue Tube] is removed first, the water in the bottle may blow out for a
moment.
When the Water Bottle (IN) [Clear Tube] is removed, the air in the bottle bursts out and makes a noise.
However, this is not a failure.
11
Post-use Maintenance
6-2 Maintenance for between each patient
6-2-1
No. Action Fig.
< Cleaning >
1
2
Cleaning and Disinfection of outside of the Scaler Handpiece and the Scaler Cord, and Replacing the Scaler Handpiece
Remove the all debris on the surface of the Scaler Handpiece and the Scaler Cord with a wipes (Minuten Wipes: ALPRO).
Remove the Tip from the Scaler
Handpiece.
* Reference: "4-3 Mounting the Tip".
1
Minuten Wipes: ALPRO
2
Wipe the debris on the surface of the Scaler Handpice and the Scaler
3
Cord with a cloth moistened with water.
≥ 500 lx
After cleaning, under appropriate lighting (500 lx or higher), inspect the Scaler Handpiece and the Scaler Cord for contamination.
4
< Disinfection >
1
2
If any visible contamination remains, repeat the process until it is visually clean. Be sure to repeat from the procedure No. 3.
Wipe the surface of the Scaler Handpiece and the Scaler Cord with disinfectant wipes (Minuten Wipes: ALPRO).
Disconnect the Scaler Handpiece from the Scaler Cord, then attach a sterilized Scaler Handpiece and place it on the Holder.
3 4
≥ 500 lx
1 2
Minuten Wipes: ALPRO
12