NSK Varios 970 Operation Manual

Multi Function Ultrasonic Scaler
OPERATION MANUAL
Please read this Operation Manual carefully before use, and file for future reference.
Varios 970
OM-E0847E 000
Classifications of equipment
• Type of protection against electric shock: – Class I equipment
• Degree of protection against electric shock: – Type BF applied part:
• Method of sterilization or disinfection recommended by the manufacture:
– See 12. Sterilization
• Degree of protection against ingress of water as detailed in the current edition of IEC 60529: – Foot Control: IPX1 (Protected against vertically falling water drops)
• Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide: – EQUIPMENT not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide.
• Mode of operation: – Continuous operation
English
Intended to Use
This product is designed only for dental clinic /dental office use. This device generates ultrasonic waves intended for use in dental applications such as scaling, root canal treatment, periodontal and cavity preparation.
1. Cautions for handling and operation
Read these cautions carefully and use only as intended or instructed. Safety instructions are intended to avoid potential hazards that could result in personal injury or damage to the device. Safety instructions are classified as follows in accordance with the seriousness of the risk.
Class Degree of Risk
WARNING CAUTION NOTICE
A hazard that could result in bodily injury or damage to the device if the safety instructions are not
followed.
A hazard that could result in light or moderate bodily injury or damage to the device if the safety
instructions are not followed.
General information needed to operate the device safely.
WARNING
• TO PREVENT ELECTRIC SHOCK Do not unplug the power cord with wet hands.
• TO PREVENT ELECTRIC SHOCK Be sure to prevent water on the Control Unit.
• TO PREVENT ELECTRIC SHOCK Do not touch the handpiece backend electrical connections.
• TO PREVENT ELECTRIC SHOCK Use an electrical outlet that is grounded.
• If you feel any abnormality such as vibration, heat generation, abnormal noise, etc., prior or during the use of the unit, stop
using it immediately.
• Do not turn the Power Switch without reason; it might blow out a fuse.
• This product is Medical Electrical equipment Electromagetic compatable (EMC).As described in the accompanying
documentation.
• Portable and mobile RF communications equipment can affect Electrical Medical equipment. Do not use RF equipment in
close proximity to the product.
• When installing the product, provide space of approximately 10cm around the Control Unit for easy access to the inlet and the Power Cord.
• When operating the product always consider the safety of the patient.
• Use by medical professional, such as doctor or dental hygienist, is intended.
• Check the vibration outside the patient’s oral cavity before use. If any abnormalities are found, stop using immediately and
contact your dealer.
• Do not drop, hit, or subject to excessive shock to the Control Unit/Handpiece.
• USE ONLY NSK genuine tips when using NSK Varios Ultrasonic Scaler (Varios 970 or Varios 970 LUX) problems such as damage, failure and accident of Handpieces resulting from use of Non-NSK Tips are not included in the warranty. The following are the possible failure that could happen when using the Non-NSK Tips;
· Vibration failure caused by using non conforming screws.
· Patients accidental ingestion of broken tips.
· Damage of thread ridge of handpiece.
· You must use the tip within the power range described on the Tip-Power Guide. If you use it out of the power range, the tip might break or damage an operative site.
• To prevent possible tooth plane damage and handpiece overheating, Always use with sufficient water.
• Do not sterilize by ultraviolet light. Handpiece could discolor.
• Sterilize the Tip, Handpiece, and Tip Wrench by autoclaving. Wipe the Control Unit, AC Power Cord, Foot Control, and Handpiece Cord including the cover.
• If chemical, solvent or antiseptic solution is deposited on this product, immediately wipe it away. Discoloration or
deformation may occur if left.
• Do not disassemble or alter the handpiece/Control Unit.
• Keep away from patients with cardiac pacemakers.
• Keep away from explosive substances and flammable materials. Do not use for patients anesthetized under laughter gas. (Nitrous Oxide)
• Use the Fuse of specified rating. (120V: T630mAL 250V, 230V: T315mAL 250V)
• This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information.
• The use of ACCESSORIES, transducers and cables other than those specified, with the exception of transducers and cables sold by the manufacturer of this product as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of this product.
• This product should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this product should be observed to verify normal operation in the configuration in which it will be used.
• If any water drops remain on the handpiece after autoclaving, wipe them off., Staining may result if left.
• There is the judgment that applies this product to a patient in the user side.
• Grounding reliability can only be achieved when the equipment is connected to an equipment receptacle marked "Hospital Only" or "Hospital Grade".
• Do not apply excessive power to the Tip. It may damage the teeth because of the ultrasonic vibration.
Power plug below is used in N orth America area.
Plug Type NEMA 5-15P (Hospital Grade Type)
CAUTION
• During operation, high frequency oscillations in the handpiece and handpiece cord may affect computer and LAN Noise may be heard during operation near a radio receiver.
• Be sure to turn off the Power Switch after use. Remove the power plug and water inside of the Control Unit before storage.
• Users are responsible for operational control, maintenance and inspection.
• Clean/sterilize the product immediately after using it. Then store it. Leaving it non-sterile might lead to failure.
• When cleaning, use MinutenWipes (ALPRO) to wipe the surface of the handpiece. Use of chemicals other than those may cause the handpiece to discolor, crack, etc.
• When you have not used the product for long time and use it again, check the operation before use.
• Eye damage may result if the LED is stared directly into, Do not look into or turn it to the eyes of the patient.
• This product does not consider patient’s age (except infants), gender, weight or nationality.
• No special training is required for this device.
• Applied parts for patient and/or operator are/is tip and Handpiece.
• Surface temperature of tip shall be more than 50 degree without using a tap water or bottle. To avoid this event, be sure to use a tap water or bottle.
NOTICE
• Repeated autoclaving may cause the handpiece to become discolored due to heat. However, this is due to properties of the product and is not a problem in terms of quality.
2. Component Names
Optic Non-OpticNon-OpticOptic
English
English
Option
15
12 13 14
16
No. Parts Name Quantity
1 Control Unit 1 2 VA Bottle 2 3 AC Power Cord 1 4 Varios2 Handpiece (Optic or Non-Optic) 1
Handpiece Cord (Unshielded 2M)
5
119 10
17
18
6 Foot Control 1 7 Sterilization Case 1 8 Tip Wrench 3 9 Tip G4 1
10 Tip G8 1
11 Tip G16 1 12 O-Ring (Thin section) (For VA Bottle) 2 13 O-Ring (Thick section) (For VA Bottle) 2 14 O-Ring (For Handpiece Cord) 2
15 Water Connector (Option) 1 16 Water Tube (Option) 1 17 Spanner Wrench (5x8) (Option) 2 18 Tip Cover S (Option) 1 19 Tip-Power Guide 1* 20 Tip Card 1* 21 Operation Manual 1*
* These are not on photo above.
(Optic or Non-Optic)
1
* Operation Principle
A sinusoidal electrical signal, at ultrasonic frequency ( f > 20Khz ), is delivered by the generator. This signal is applied to the ‘piezoelectric ceramic’ located inside the transducer. Piezoelectric ceramic converts this signal into mechanical vibrations. These vibrations are at the same ultrasonic frequency as the electrical signal. The mechanical vibrations are propagated towards the distal end of the transducer. The “TIP” insert, which is attached at the distal end of the transducer, vibrates at ultrasonic frequencies and makes it possible to achieve the aimed purpose.
3. Name and Function of each part
Left Bottle (L Bottle) Right Bottle (R Bottle)
* L and R bottle is common
Handpiece Holder
Bottle Selection Indicator L
Handpiece Cord Connector
Tap Water Adjustment Knob
(Left)
Bottle Selection Indicator R (Right)
Bottle Water Adjustment Knob
Power Switch
Foot Control Connector Tap Water Connector
AC Power Cord Connection Jack
<Bottom>
Operation Panel and Display
Mode Select Keys Power Level Keys
Endo Key
Perio Key General Key
Down Key
Up Key
Irrigation Select Key
Auto-cleaning Key
English
English
Operation Mode Display
Numerical Display Bar Graph Indicator
If you purchase the Optional products such as Water Tube and Water Connector you can use Tap Water.
Irrigation Mode Display
Mode Select Keys
You can select Operation Mode to pressing this key. (Perio, Endo and General) The Control Unit can resume power level, water volume and irrigation mode for each Operation Mode.
Power Level Keys
Press key to select Power o/p Level . There are 11 levels (0 to 10). There is no output vibration at level 0 (zero). (Fig.1). The Bar Graph Indicator and Numerical Display will change simultaneously.
Irrigation Select Key
Press key to select ‘R’ or ‘L’ Bottle. Front panel display and Bottle Selection Indicator simultaneously change in position. Pressing the Irrigation Select Key for more than one second will select Tap Water Mode.
Auto-Cleaning Key
Press key to select Auto Cleaning Mode, For detail refer to 11. (5).
Fig.1
Bottle Water Adjustment Knob
Water Volume Adjustment can be made prior to the tip vibrating you can adjust the Water Volume during Bottle Irrigation or during a wait before Tip vibration start. If the setting is not applicable (too low or too high) for the Control Unit, it may beep. During the operation, Front Panel displays the current Power Level. However, keep turning the knob more than a second; it may change the Water Volume.
CAUTION
• Do not turn the knob fast. It may not sense
the movement if it turn fast.
• Water Volume can set during 5ml/min to 45ml/min.
• Operation sound may different between Right and Left Bottle.
• During adjustment of water volume, Numerical Display indicate “–”.
Tap Water Adjustment Knob
You can adjust tap water volume by this Knob. (Even the tip vibration).
4. Prior to Operating System
(1) Water System Setup
• Use of Bottle
1) Remove the Dust Cover from the Bottle Base Connector. (Fig. 2)
2) Remove the cap of the VA Bottle and fill solution to the desired level.
3) Close the cap of VA Bottle and insert the Bottle Joint into the Bottle Base Connector until it clicks. (Fig. 3)
To remove the Bottle, pull it up.
CAUTION
• Use the VA Bottle Set 400 only for Varios 970.
• Before filling solution to the VA Bottle, check the Gasket inside the bottle cap is clean. (Fig. 4)
• Do not use a sharp tool to clean the Gasket or do not allow any impact on to the product. It may cause malfunction.
• Insert the Bottle straight. (Damage to seals may result).
• Keep the Gasket clean. When it becomes dirty by water or antiseptic solution, wash by clean water it immediately.
• The Gasket is consumable. *Optional Gasket: Order Code Z1047350
Dust Cover
Bottle Base Connector
Fig.2
Bottle Joint
NOTICE
• The Bottle calibrations are printed on both sides of the Bottle and can be read accurately from the fill position or mounted on the Control Unit.
• Mount the Dust Cover when not in use.
• Use of Tap Water (Option)
1) Remove the Cover from the Tap Water Connector. (Fig. 5)
2) Connect the filter side of the Water Tube deep into the Control Unit Tap Water Connector (Fig. 6).
3) Connect the water tube to the water outlet on the Dental Unit.
Cover
Water Tube
Fig.5
Gasket
Fig.3
Fig.4
Tap Water Connector
Water Filter Case
Fig.6
CAUTION
Ensure the water from the dental unit runs clear before connecting to the scalar.
English
NOTICE
• Insert the Water Tube firmly into Control Unit.
Pushing the White Ring, (the Quick Connector Release Ring) on the Tap Water Connector, gently pull the tube to remove. (Fig. 7)
• When the water tube is not connected, mount the cover on the Tap Water Connector.
(2) Foot Control Connection
Connect Foot Control Plug and the Control Unit with [ ] mark on the upper surface of the plug. (Fig.8)
(3) Handpiece Cord Connection
Insert Handpiece Cord Plug into Control Unit. [ ] Mark side is upper surface. Do not insert it up-side-down. (Fig.9)
Foot Control Plug
Foot Control Connector
Handpiece Cord Connector
[ ]
Mark
White Ring
Fig.7
[ ] Mark
Fig.8
Handpiece Cord Plug
CAUTION
Check that the Handpiece Cord Plug is clean & dry before connecting. (Fig.10)
Fig.9
Fig.10
(4) Mounting Power Cord
Insert the Power Cord into the AC Power Cord Connection Jack at the back of the Control Unit. (Fig. 11)
Power Cord
AC Power Cord Connection Jack
CAUTION
• Ensure Power is OFF on the Control Unit during the Power Cord Connection. It may cause Fuse to blow.
• Do not connect the cord in wall outlet before connecting AC Power Cord.
• Do not pull the AC Power Cord forcibly.
• Do not unplug the Power Cord or handpiece cord while pressing on the Foot Control.
5. Mounting and Removing the Handpiece
Align the Dots on the Handpiece and the Handpiece Cord. Push
handpiece into connector.
To remove the handpiece, grip the Handpiece and Handpiece Cord and pull to part handpiece and cord. (Fig. 12)
Dots
WARNING
To avoid Electrical Shock Do not touch the handpiece backend electrical contacts.
Handpiece backendHandpiece Handpiece Cord
Fig.11
Fig.12
CAUTION
• Always confirm that the handpiece is correctly seated and locked into place.
• Do not connect or use Handpiece other than included one (Varios2 handpiece).
6. Mounting and Removing Tip
1) Turn Tip lightly by hand, and install it. (Fig. 13)
2) Tip will insert from the bottom hole of Tip Wrench. Align the four corner of the Tip base area into the four corner of Tip Wrench. And turn it clockwise until it clicks. (Fig. 14)
Do not touch the top part of Tip to avoid an injury. (There is the case that is longer than height of TIP WRENCH)
To remove the Tip, turn it counterclockwise with the Tip Wrench.
Fig.13
Tip Wrench
Loosen
Tighten
Loosen
Tighten
Fig.14
Caution for Tip Usage
• Check the Tip before use. (Flush, Damage, Bending or Rust)
• Do not exceed Maximum Power Level for Tip. Damage to tooth structure and Tip may result.
• Do not hit ceramic prosthesis with Tip during scaling. Tip Damage may result.
• Do not hit metal or prosthetic crown except for removing them. Tip could break and fall into mouth.
• Do not hit gingival, mucosa and/or skin. It could cause damage and/or burn injury.
• Do not sharpen and/or bend the Tip. Tip may damage and not generate enough vibration during scaling.
• During cutting, Tip will gradually wear away, as the Tip wears the stroke will get smaller and decrease cutting efficiancy When level drops too far, change the Tip.(tip card check)
• DO ENSURE When securing tip to use the tip wrench as supplied, inefficient cutting will result.
• DO ENSURE before attaching Tip, Cleanliness of the tip screw, inefficient cutting will result.
• To avid personal injury DO ENSURE tip is removed prior to disconnecting the handpiece or the handpiece cord.
• If you feel the Tip is not vibrating, remove it from an operative site, and press the Foot Control again. If this does not improve the condition, Ensure the tip is secure, turn the power off and restart it.
• When mounting the Tip, always use groves and Tip Wrench as supplied.
• Ensure that water volume must be “0”, when you use Tip which does not appear of water.
• Tip Wrench is consumable For reliable operation replace annually.
7. Operating Procedures
(1) Water System Setup
• Use of Bottle
1) Check that the VA Bottle is filled to the proper level.
2) Make sure that the cap of the Bottle is secure and not leaking.
CAUTION
• DO ENSURE liquid temperature is below 35°C.
• Do not put liquid such as high acid water in the Bottle.
English
• Use of Tap Water
1) Ensure water tube is firmly connected.
2) Open the dental unit's water valve. (Set water pressure between 0.1-0.5MPa (1-5 kgf/cm
(2) Power On
Connect the AC Power Cord to the wall outlet. Turn the Power Switch on the Control Unit. Front Display will illuminate.
Power Switch
2
)).
Fig.15
(3) Power Level Setting
DO ENSURE Power setting does not exceed the recommended Power Level (Tip-Power Guide included in the package.)
1) Select the Operating Mode with the Mode Select Keys on the Front Panel. The Indicator light over the selected mode will illuminate. (Fig. 16)
Mode Select Keys
Power Level for each mode
Operating Mode Display
2) Set the power level with the Power Level Key on the Front Panel. The Bar Graph Indicator and Numerical Display will indicate the selected power level. (Fig. 17) Make sure the power level is set in the appropriate range for the attached Tip.
Fig.16
NOTICE
• Press & Hold the Power Level Key will increase or decrease the Power Level.
• If the Power Level is 0 (zero) and set the water volume, Tip will not oscillate but water comes out
from the handpiece.
(4) Irrigation setting
Select the Irrigation Mode (L Bottle, R Bottle or Tap Water) with the Irrigation Select Key on the Front Panel. (Fig. 18) The Indicator light over the selected mode will illuminate. Press & Hold the Irrigation Select Key to select Tap Water Mode.
(5) Operate Varios 970 / 970 LUX
By stepping on the Foot Control, the tip vibrates and spraying starts (except for tips that do not spray) and the handpiece LED lamp lights up (Varios 970 LUX). When you remove your foot from the foot control, tip vibration and water spraying stop and the handpiece LED lamp turns off. (Varios 970 LUX).
Power Level Key
Bar Graph IndicatorNumerical Display
Irrigation Select Key
L Bottle Tap Water R Bottle
Increase
Fig.17
Fig.18
Increase
• Water Supply Volume Adjustment
Turn the Water Adjustment Knob counterclockwise gradually to increase the supply volume. (Fig. 19) For detail, refer to P5 Bottle Water Adjustment Knob or P6 Tap Water Adjustment Knob.
10
Tap Water Adjustment Knob Bottle Water Adjustment Knob
Fig.19
CAUTION
• While pressing the Foot Control and switching the power ‘ON’. The Control Unit will display "F" and sound a beep, for your safety the Control Unit will not operate. Remove your foot from Foot Control to cancel and reset.
• Bar Graph Display (Fig.20)
Minimum Irrigation -> One white and blue LED. No Irrigation -> Blue LED only
• Always use the water supply. If water supply is insufficient, handpiece will overheat and patient’s tooth surface can be injured.
• Verify that the water spray is clean and of adequate volume before use.
• If irrigation volume set low, sometimes irrigation water is difficult to come out from the Tip. When it happened, set volume again after setting up high volume.
• During Water Adjustment Knob operation;
Numerical Display: Display "-"
Bar Graph: Display current volume of water
Minimum irrigation No Irrigation
(6) After the treatment
Release the Foot Control and Power off the Control Unit.
• Use of Bottle
Thoroughly wash the Bottle (s) Water Supply system. Refer to 11. (5) Auto Cleaning (Cleaning of Irrigation Tube).
CAUTION
When using medicated solutions, clean the entire Irrigation System thoroughly.
• Use of Tap Water
Close the dental unit's water valve.
English
Fig.20
NOTICE
• LED of the handpiece will remain ‘On’ for approx 5 seconds after Foot Control is released. (Varios 970 LUX)
• When the Control Unit is Power off, the last mode settings in use are automatically retained in memory.
Initialized Program (Factory Setting)
Press both Auto-Cleaning and Power Key on the Control Unit to initialize the Factory memory Setting. Do not release Auto-Cleaning Key until the beeping sound from the Control Unit. (Initial Mode is Perio)
Power Flow amount (L, R each) Irrigation Mode Initial Mode Perio 1 10 L Bottle Endo 1 10 L Bottle General 1 10 L Bottle
During the Handpiece operation :
Possible: Power Level and Water Volume adjustment. Impossible: Operation Mode and Irrigation Mode setting, Auto Cleaning.
11
8. Provided Scaler Tips
G4
Apply the top of the Tip on the tooth plane and move it sideways finely in the same way as G8 Tip. (Fig. 21)
The end of the Tip is thin and for supragingival fine scaling and interdental scaling. The round cross-section allows tooth surfaces to be finished without causing damage.
Fig.21
G8
Apply the top of the Tip on the tooth plane and move it sideways finely along the neck of tooth. (Fig. 22)
G16
Insert the top of the Tip into the periodontal pocket and move it slowly. The top of the Tip is sharp so that it could remove tartar on long coroner and retracted gingival. (Fig. 23) Clean periodontal pocket at low power.
Removal of supragingival and interdental calculus. This Tip can be used in all quadrants and is very useful for the removal of hard calculus.
Removal of supra and subgingival calculus. It provides easy access to interdental
spaces and narrow pockets.
Fig.22
Fig.23
CAUTION
Tip is article of consumption. We recommend periodical replacement. About time of replacement, check the Tip Card.
12
WARNING
Do not connect the supplied G16 Tip to a handpiece other than a Varios2 handpiece (VA2-LUX-HP / VA2-HP). Doing so may cause fracturing of the Tip or scratches on the surface of the dentition, prosthesis, etc. due to an increase in vibration amplitude (Fig. 24).
Fig.24
How to use the Tip Card
1) Place the neck of the Tip in the cut out.
2) Check wear of the Tip.
3) See the green, yellow and red line to check wear of the Tip. *See below what each color means. At NSK we recommend to replace a Tip when the Tip meets the yellow line (wear of 1mm) to guarantee safe and effective use.
Tip Card
Green: No wear - Tip is OK Tip replacement is not necessary.
English
CAUTION
Tips are consumables. The efficiency of dental scaling decreases approximately 25% when the top of the Tip wears 1 mm and approximately 50% when it wears 2 mm. In addition, the vibration condition changes owing to the wear, which may damage a patient’s tooth surface. Check the Tip wear condition with the Tip Card periodically, and replace the Tip with a new one in good time.
13
Yellow: Wear of 1mm - Tip is showing some wear Tip replacement is recommended.
Red: Wear of 2mm - Tip is badly worn Tip replacement is necessary.
Fig.25
1mm
25% Decrease
Efficiency
2mm
50% Decrease
Fig.26
9. How to Use Tip Cover S (Option)
Grip the Tip Cover S and insert it to the Tip. To remove, grip the Tip Cover S and the handpiece & pull. (Fig. 27)
The Tip Cover S is not designed for use as a Tip changing tool.
CAUTION
Carefully insert the Tip into the Tip Cover S. Avoid
injuring the fingers.
Slit
10. Handpiece Holder
While the Handpiece is not in use, put the Handpiece in the Handpiece Holder. The Handpiece Holder is adjustable. (Fig. 28)
CAUTION
Do not put excessive load to the Handpiece Holder to prevent from breaking down and deformation.
NOTICE
To prevent injury, always mount Scaler Tip Cover (S).
Slit
Tip
Tip Cover
Tip-Handpiece Joint
Fig.27
Handpiece Holder
Fig.28
11. Care and Maintenance
(1) Cleaning of Optic Fiber (Varios 970 LUX)
Wipe the debris off the end of the Optic Fibers at the handpiece with alcohol soaked cotton swab. (Fig. 29)
CAUTION
Do not use any sharp pointed tools to clean the Optic Fiber End Face. In case the light degridation,
contact your dealer.
Optic Fiber End Face
Fig.29
14
(2) Cleaning the Handpiece cord
Remove the handpiece after use on each patient and clean it
as described below.
1) Wipe the surface of the handpiece cord with a cloth soaked
in alcohol.
2) Carefully wipe the handpiece cord plug with an alcohol-immersed cotton swab. If it is difficult to use a cotton swab, carefully wipe with a towelette wound around a thin stick-shaped object.
CAUTION
Do not use a sharp pointed stick or push the terminal part when cleaning the handpiece cord plug. Doing so may cause damage, resulting in contact failure (Fig. 30).
Terminal part
English
Fig.30
(3) Changing O-Ring
• Handpiece Cord
An O-Ring is located in the Handpiece Cord Connector. Use a pointed tool to remove, and mount new O-Ring into the groove. (Fig. 31) * Optional O-Ring: Order Code D0310020080
• VA Bottle
Remove two O-Rings at the Bottle Joint with a pointed tool, and mount new O-Rings into the grooves. (Fig. 32) * O-Ring (Thick section) : Order code D0310075150 O-Ring (Thin section) : Order code D0312090100
(4) Changing the Irrigation Pump
1) Remove the Bottle, the Power Cord, the Handpiece Cord and the Foot Control from the Control Unit.
2) Turn back the Control Unit. Hang a finger on “ up the bottom cover to remove.
” point and pull
O-Ring
Fig.31
O-Ring (Thick section)
O-Ring (Thin section)
Fig.32
Fig.33
15
Picture below is shows inside of the Control Unit.
Front Panel Side
R Pump Tube Joint
L Pump Tube Joint
R Bottle Bottle Joint
R PumpL Pump
Bottom Cover Hook
Bottom Cover (Back Side)
L Bottle Bottle Joint
3) Remove the Irrigation Tube from the Control Unit. (Bottle side and Front Panel side.) (Fig. 35, 36)
4) Remove the Connector Ring from the Irrigation Tube. Do not dispose it. You can use the Rings to the replacement Irrigation Pump.
5) Turn the Irrigation Pump counterclockwise until it clicks
Irrigation Tube
and pull it out. (Fig. 37)
6) Mount the Connector Ring to the new Irrigation Pump. Observing Ring direction. (Fig. 38)
7) Align the replacement Irrigation Pump with the Drive Shaft. Turn clockwise until it clicks. (Fig. 37)
8) Mount the Irrigation Tube opposite procedure of removing
Remove
Remove
Mount
Mount
Irrigation Tube
Connector Ring
Connector Ring
(Fig.35). Connector Ring should firmly into the Control Unit until it stops. (Fig.39)
* Bottle side
Fig.34
Fig.35
* Front Panel side
Irrigation Tube
L/R Pump side
Irrigation Tube
Irrigation Tube
Control Unit side
Connector Ring
Fig.36
Fig.38
16
Release LockRelease Lock
Irrigation Pump
Irrigation Pump
Connector RingConnector Ring
* Front Panel side
Fig.37
Fig.39
9) Align the Bottom Cover Hook and hole on the Control Unit. Mount the Bottom Cover.
* Optional Irrigation Pump: Order Code 10000643 (Not included the Connecter Ring.)
CAUTION
• If water is spilled out the irrigation pump, wipe it off and allow drying completely prior to use. If water gets inside
the irrigation pump the roller may slip and fail to pump.
• Before replacing the Irrigation Pump, wipe off excess water on pump and Drive Shaft. The wet drive shaft and rollers can be slippery and cause improper operation.
• Wipe dirt and water off the Drive Shaft from bottom up. (Fig.40)
• Insert the replacement Irrigation Pump into the Drive Shaft straight (slow and soft) to prevent damaging rollers
in pump.
• Run the replaced Irrigation Pump about 10 seconds on largest setting of Water Volume before operation to adopt Irrigation Tube to new pump.
• Ensure Irrigation tube has no kink or twists If tube is set incorrectly, Irrigation Water may not come out.
• Do not pull the tube when the bottom cover is closed.
Drive ShaftDrive Shaft
NOTICE
• Perform periodical cleaning for the Drive Shaft with socked alcohol cloth. Dirt on Drive Shaft may cause an
incorrect pump operation.
• The pump is consumable. If the irrigation volume decreases markedly, replace pump.
(5) Auto Cleaning (Cleaning of Irrigation Tube (Use of Bottle))
NOTICE
• After each use, remove all the disinfectant solution and perform "Auto Cleaning" procedure. If you have not cleaned the system, it may become dirt disinfectant. And it is stuck in the tubing or some of the metal parts may be rusted.
• During Auto Cleaning, water comes out from the handpiece. Perform cleaning after turning handpiece into a cup.
English
Fig.40
1) Remove the 2 Bottles from the Control Unit.
2) Clean inside of the Bottle.
3) Half fill the bottle with purified water (DO NOT USE SALINE)
CAUTION
Use only distilled water for cleaning.
4) Install the cap on the Bottle. Install the Bottle Joint into the Bottle Base Connector. Install it until clicks into place. Improper connection may cause water leakage. Make sure the connection is tight.
CAUTION
• Perform Auto-Cleaning without tip.
• Make sure the handpiece and handpiece cord are firmly attached.
17
5) To perform the Auto-Cleaning, keep pressing the Auto-Cleaning Key (more than 1 second).It takes 30 seconds per bottle to clean The Numerical Display will alternately displays “A” and “C”, the Bar Graph displays time remaining. Single display (Bar Graph Display) is 6 second. When five displays of Bar Graph disappeared, Bottle will be changing
the other side.
To cancel the Auto-Cleaning, press Auto-Cleaning Key again.
Auto-Cleaning Key
NOTICE
During Auto-Cleaning, LED of the handpiece does not illuminate. (Varios 970 LUX)
Numerical Display Bar Graph
6) When the Auto-Cleaning is finished, the Control Unit returns to the settings prior to cleaning. Remove the both bottles from the Control Unit by pulling straight up. Clean thoroughly rinse and dry.
Following method is also available for cleaning. (Manual Cleaning)
1) Remove the Bottle from the Control Unit.
2) Open the cap of the cleaned Bottle and fill it with distilled water.
3) Close the cap firmly and insert the Bottle Joint into the Bottle Base Connector on the Control Unit until it clicks.
4) Operate the Control Unit about 30 seconds with water supply at maximum setting.
NOTICE
The Control Unit does not perform in Auto-Cleaning in Tap Water.
(6) Changing Water Filter (Option)
If you use Tap Water, change the Water Filter as it may necessary.
1) Close the water valve of the dental unit.
2) Mount two Spanner Wrenches (5x8) and turn those as shown in Fig.42.
3) When the Water Filter case is separated, the Water Filter can be removed as shown in Fig.43.
4) Replace with new ( Order Code U387042 ) and reassemble the filter in the reverse order.
Varios970/ Varios970 LUX
O-Ring
Fig.41
Water Filter Case
Dental Unit
Fig.42
Water Filter
Varios970/Varios970 LUX Dental Unit
Fig.43
18
12. Sterilization
Only handpiece can be washed via Thermo Disinfector.
• Autoclave sterilization is recommended.
• Autoclave sterilization required first time you use and after each patient as noted below. Take handpiece out of the packing bag before sterilization.
• ONLY the Tip, Handpiece and Tip Wrench can be autoclaved.
Autoclave Procedure
1) Remove the Tip after use. (Refer to 6. Mounting and Removing Tip)
2)
Wipe dirt and debris from the products, and wipe clean with alcohol-immersed cotton swab or cloth. For handpiece, use
MinutenWipes (ALPRO) to wipe. Do not use a wire brush.
3) Insert those into the Sterilization Case or an autoclave pouch. Seal the pouch.
4) Autoclavable up to max. 135˚C.
Ex.) Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C.
5) Keep the products in the Sterilization Case or autoclave pouch to keep it clean until you use it.
* Sterilization at 121°C for more than 15 minutes is recommended by ISO17664 and EN ISO17665-1.
CAUTION
• Do not sterilize by ultraviolet ray. The handpiece could discolor.
• If autoclaved with other instruments stained with chemical solution, it could strip the plating and make the surface black.
• Do not autoclave any parts (the Control Unit, Power Cord, Bottle, Foot Control, Handpiece Cord, O-Ring). Other than those that can be subjected to autoclave sterilization. Perform alcohol disinfection to the Control Unit, Power Cord, Foot Control, Handpiece Cord including after every patient.
• Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.
English
NOTICE
Repeated autoclaving may cause the handpiece to become discolored due to heat. However, this is due to properties of the product and is not a problem in terms of quality.
Sterilization Case
The Handpiece, Tip and Tip Wrench can be sterilized together using Sterilization Case.
1) Remove the Tip after use. (Refer to 6. Mounting and Removing Tip)
2) Set the Tip Wrench with Tip into the Sterilization Case. (You can set four Tip Wrenches and Tips at once).
3) Remove handpiece from the Handpiece Cord, and clean.
4) Set the handpiece into the Sterilization Case.
5) Autoclavable up to max. 135˚C.
ex.) Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C.
6) Keep the products in the Sterilization Case or autoclave pouch to keep it clean until you use it.
19
Fig.44
13. Troubleshooting
When trouble is found, please check the followings prior to consulting your dealer.
Problem Probable Cause Cause Solution
The Power Cord or the Jack is
disconnected.
The Fuse is burned out. Contact dealer.*
The Tip is not tightened firmly. Tighten the Tip until the Tip Wrench clicks. Worn Tip. Replace the Tip. Power has not been correctly
adjusted for the Tip. The Foot Control is disconnected. Connect the Foot Control correctly. Failure of vibrator in the handpiece. Contact dealer.* Failure of internal components of
the Foot Control. Power has not been properly
adjusted for the Tip.
Power has not been properly adjusted for the Tip.
The Tip is not tightened firmly. Tighten the Tip until the Tip Wrench clicks. Failure of vibration in the handpiece
or the Control Unit. Power has not been properly
adjusted for the Tip. The Tip is not tightened firmly. Tighten the Tip until the Tip Wrench clicks. Failure of vibration in the handpiece
or the Control Unit.
The tube twisted. Straighten the twisted Irrigation Tube.
Time to replace Irrigation Pump. (Approx. 500hours after used.)
The Water Adjustment Knob is
closed.
Disconnected Irrigation supply at low volume range. (less than 10ml/ min.)
The Water Filter is clogged.
Correctly insert the Power Cord or the Jack.
Adjust the power on the Power Guide or Tip case label. Do not exceed.
Contact dealer.*
Adjust the power level the Power Guide or Tip case label. Do not exceed.
Adjust the power level on the Power Guide or Tip case label. Do not exceed.
Contact dealer.*
Adjust the power level on the Power Guide or Tip case label. Do not exceed.
Contact dealer.*
Replace with new Irrigation Pump (Refer to 11. (4) Changing the Irrigation Pump).
Check the water circuitry and supply to the Control Unit. Water pressure : 0.1-0.5MPa (1-5kgf/cm
Turn the Water Adjustment Knob and adjust to
the appropriate volume.
No problem. Turn the Water Adjustment Knob and increase the Irrigation volume.
Replace with new Water Filter (Refer to 11. (6) Changing Water Filter (Option) ).
2
)
No / Poor vibration.
The Tip is bent or broken.
The Tip is flying
away.
Noise from the
handpiece.
The handpiece is
overheating.
No Irrigation supply and/or unstable Irrigation supply (Use of Bottle)
No / Poor water. (Use of Tap Water)
The Front Panel does not light, even if the Power Switch is ON.
The Tip does
not generate
vibration, in spite
of depressing
the Foot Control.
The Tip is not tightened firmly. Tighten the Tip until the Tip Wrench clicks.
The Irrigation Pump is running.
The Irrigation Pump is
stopping.
The water does
not reach to the
Control Unit.
Check to see if
water reaches
the Control Unit.
20
Problem Probable Cause Cause Solution
Water is leaking
from the joint
between the Irrigation Tube
The Irrigation Tube is not connected
correctly.
Firmly insert the Irrigation Tube into the Irrigation
Connector inmost. and the Irrigation Connector.
Water leakage.
Water is leaking
from the joint
between the
handpiece and
O-Ring at the handpiece cord is
worn or damaged.
Replace with new O-Ring (Refer to 11 (3)
Changing O-Ring •Handpiece Cord).
the cord.
Water is leaking from the Control Unit.
Tip oscillates, but
Handpiece LED does not
illuminate.
(Varios 970 LUX)
Handpiece LED
turns on and off.
Tip oscillates, but Handpiece LED
does not turn on.
Beeping while
power on.
Start Beeping
Beeping while
stopping
vibration of Tips.
* Repairs cannot be made by the customer.
The water circuitry in the Control Unit is damaged.
The handpiece is not connected into the Handpiece Cord correctly.
Disconnection in the Handpiece Cord, or failure in the Control Unit.
Contact dealer.*
Firmly insert the handpiece into the Handpiece
Cord inmost.
Contact dealer.*
Depress Foot Control. Release the Foot Control.
Abnormal heating of the Control Unit.
Stop the operation and leave until Control Unit
becomes cool.
English
21
14. Specifications
Type NE255
Power Source
AC120V 50/60Hz
AC230V 50/60Hz Vibration Frequency 28-32kHz Maximum Output 11W Rated Power 29VA
2
Water Pressure 0.1-0.5MPa (1-5kgf/cm
Lighting
Varios 970 : No
Varios 970 LUX : Yes
)
Bottle Volume 400mL (Per Bottle) Dimensions W160 x D270 x H190mm (Including Bottle) Weight 2.1kg (Except attachment)
Temperature 0 - 40 ˚C (The liquid must not freeze up) Use Environment
Humidity 30 - 75 %
Atmospheric pressure 700 - 1060 hPa
Temperature -10 - 60 ˚C Store Environment
Humidity 10 - 85 %
Atmospheric pressure 500 - 1060 hPa
15. Protection Circuit
It may overheat inside when you use this Control Unit in more than Power 8 at G mode for long time. In this case, Protection Circuit reduces the Power automatically. (Power 7) Bar Graph Indicator from 8 to 10 flashes. (Fig.45)
After Protection Circuit is released, the flashes stop. However, Power Level can not automatically increase. If needed, increase manually.
NOTICE
• During Protection Circuit function (during Bar Graph Indicator flash), the Control Unit can not increase the Power Level.
• If Power Level decreases less than 7, Bar Graph Indicator stops flashing. However, it the Power increase more than 8, flashes it again.
22
Fig.45
16. Error Code
If an operational problem occurs numerical Display shows the error code to allow an immediate problem diagnosis.
Error Code Error Check / Remedy
E 0 Self-Check Error Contact dealer. E 1 Circuit Failure Contact dealer. E 7 Does not vibrate Contact dealer.
Confirm connection of the handpiece.
E 9 Handpiece Self Check Error
E 10 Circuit Failure Contact dealer.
*“E” and the number alternately display on the Display.
Power on the Control Unit again. Leave the Control Unit until it become cool down and powers it again. When an error can not be eliminated, Contact dealer.
17. Spare Parts
Model Products Oeder Code Model Products Oeder Code
VA Bottle Set 400 Z1047002
VA Bottle 400 20000947
Sterilization Case Z1035001
Water Tube Set U387040
VA Bottle Cap 400 10000652
Gasket Z1047350
Irrigation Pump 10000643
Tip Wrench
(CR-10)
English
Z221076
Water Connector U387030
Water Filter U387042
Spanner Wrench
(5x8)
O-Ring
(for Handpiece
Cord)
Autoclavable at 135˚C max.
Y1001301
D0310020080
Tip Holder Z221A080
Tip Cover S Z217851
O-Ring (Thick section) (For VA Bottle)
O-Ring
(Thin section)
(For VA Bottle)
23
D0310075150
D0312090100
18. Disposing product
Consult with dealer from whom you purchased it about waste disposal.
19. Warranty
Manufacturer warrants its products to the original purchaser against defects in material and workmanship under normal practices of installation, use and servicing. Such expendable items as O-Rings and Irrigation Pump are not covered by this
warranty.
Symbols
TUV Rhineland of North America is a Nationally Recognized Testing Laboratory (NRTL) in the United States and is accredited by the Standards Council of Canada to certify electro-medical products with Canadian National Standards.
Follow the waste of electric and electronic equipment (WEEE) Directive (2012/19/EU) to dispose of the product and
accessories.
Consult operation instructions.
Manufacturer.
This conforms to CE European Directive of “Medical equipment directive 93/42/EEC.”
Type BF applied part. Authorised representative in the European community.
Protected against vertically falling water drops. Autoclavable up to Max.135°C. *for detail see Sterilization.
This product can be cleaned and disinfected with a Thermo-Disinfector.
Marking on the outside of Equipment or Equipment parts that include RF transmitters or that apply RF electromagnetic energy
for diagnosis or treatment.
GS1 DataMatrix for Unique Device Identifier.
Guidance and manufacturer's declaration - electromagnetic emissions
The Varios 970 / Varios 970 LUX is intended for use in the electromagnetic environment specified below. The customer or the user of the Varios 970 / Varios 970 LUX
should assure that is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR11/EN55011
RF emmissions CISPR11/EN55011
Harmonic emissions EN/IEC61000-3-2
Voltage fluctuations/flicker emissions EN/IEC61000-3-3
Group 1
class B
class A
Complies
The Varios 970 / Varios 970 LUX uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The Varios 970 / Varios 970 LUX is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supply network that supplies buildings used for domestic purposes.
24
Guidance and manufacturer's declaration - electromagnetic immunity
The Varios 970 / Varios 970 LUX is intended for use in the electromagnetic environment specified below. The customer or the user of the Varios 970 / Varios 970 LUX
should assure that it is used in such an environment.
Immunity test EN/IEC60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD) EN/IEC61000-4-2
Electrical fast transient/burst EN/IEC61000-4-4
Surge
EN/IEC61000-4-5 Voltage dips, short
interruptions and voltage variations on power supply input lines
EN/IEC61000-4-11
Power frequency (50/60Hz)
magnetic field
EN/IEC61000-4-8
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic immunity
The Varios 970 / Varios 970 LUX is intended for use in the electromagnetic environment specified below. The customer or the user of the Varios 970 / Varios 970 LUX
should assure that it is used in such an environment.
Immunity test EN/IEC60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF EN/IEC61000-4-6
Radiated RF EN/IEC61000-4-3
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobiles radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Varios 970 / Varios 970 LUX is used exceeds the applicable RF compliance level above, the Varios 970 / Varios 970 LUX should be observed to verity normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Varios 970 / Varios 970 LUX.
b Over the frequency range 150kHz to 80MHz, field strengths should be less than 3 V/m.
±6kV contact ±8kV air
±2kV for power supply lines ±1kV for input/output
±1kV line(s) to line(s) ±2kV line(s) to earth
<5% Ut (>95% dip in Ut) for 0.5 cycle
40% Ut (60% dip in Ut) for 5 cycles
70% Ut (30% dip in Ut) for 25 cycles
<5% Ut (>95% dip in Ut) for 5 secs
3 A/m 3 A/m Power frequency magnetic fields should be at levels
3Vrms 150 kHz to 80MHz
3V/m 80MHz to 2.5 GHz
±6kV contact ±8kV air
±2kV for power supply lines ±1kV for input/output
±1kV line(s) to line(s) ±2kV line(s) to earth
<5% Ut(>95% dip in Ut) for 0.5 cycle
40% Ut (60% dip in Ut) for 5 cycles
70% Ut (30% dip in Ut) for 25 cycles
<5% Ut (>95% dip in Ut) for 5 sec
3Vrms
3V/m
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Varios 970 / Varios 970 LUX requires continued operation during power mains interruptions, it is recommended that the Varios 970 / Varios 970 LUX be powered from an uninterruptible power supply or a battery.
characteristic of a typical location in a typical commercial or hospital environment.
Portable and mobile RF communications equipment should be used no closer to any part of the Varios 970 / Varios 970 LUX, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 P
d = 1.2 P 80MHz to 800MHz d = 2.3 P 800MHz to 2.5GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
English
25
Cables and accessories Maximum length Complies with
Handpiece cord Foot Control AC Power Cord
Recommended separation distances between portable and mobile RF communications equipment and the Varios 970 / Varios 970 LUX.
The Varios 970 / Varios 970 LUX is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Varios 970 / Varios 970 LUX can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Varios 970 / Varios 970 LUX as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
2 m 2 m 2 m
W
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73 1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
RF emissions, CISPR11, EN55011 Harmonic emissions, Voltage fluctuations/ flicker emission, Electrostatic discharge (ESD) Electric fast transient / burst
Surge
Voltage dips, short interruptions and voltage variations on power supply input lines Power frequency(50/60Hz) magnetic field Conducted RF Radiated RF
150kHz to 80MHz
d=1.2 P
Separation distance according to frequency of transmitter
m
80MHz to 800MHz
d=1.2 P
Class B/ Group 1 EN/IEC61000-3-2 EN/IEC61000-3-3 EN/IEC61000-4-2 EN/IEC61000-4-4 EN/IEC61000-4-5 EN/IEC61000-4-11 EN/IEC61000-4-8 EN/IEC61000-4-6 EN/IEC61000-4-3
800MHz to 2.5GHz
d=2.3 P
26
Klassifizierung der Geräte
• Schutzart gegen Stromschlag :
– Geräteklasse I
• Schutzart gegen Stromschlag : – Anwendungsteil Typ BF:
• Vom Hersteller empfohlenes Verfahren zum Sterilisieren oder Desinfizieren : – Siehe 12. Sterilisation
• Schutzart gegen Eindringen von Wasser gemäß der Beschreibung in der aktuellen Ausgabe von IECD 60529: – Fußschalter : IPX1 (gegen senkrecht herunterfallende Wassertropfen geschützt)
• Grad der Anwendungssicherheit bei Verwendung einer entflammbaren Betäubungsmittelmischung mit Luft oder mit Sauerstoff oder Lachgas : – GERÄT ist nicht zur Verwendung mit einer entflammbaren Betäubungsmittelmischung mit Luft oder mit Sauerstoff oder
Lachgas geeignet.
• Betriebsart :
– Dauerbetrieb
Deutsch
Bestimmungsgemäßer Gebrauch
Dieses Gerät ist nur zum Gebrauch in Zahnkliniken / Zahnarztpraxen bestimmt. Dieses Gerät erzeugt Ultraschallwellen, die für Dentalanwendungen wie zum Beispiel Scaling, Wurzelkanalbehandlung, Paradontalbehandlung und Zahnpräparationen
bestimmt sind.
1. Vorsichtsmaßregeln für Handhabung und Bedienung
Lesen Sie diese Vorsichtsmaßregeln sorgfältig durch und verwenden Sie das Gerät nur bestimmungsgemäß bzw. gemäß der Anleitung. Die Sicherheitsvorschriften dienen zum Vermeiden möglicher Gefahren, die zu Verletzungen oder einer Beschädigung des Geräts führen könnten. Die Sicherheitsvorschriften werden entsprechend des Risikogrades wie folgt eingestuft.
KLASSE RISIKOGRAD
WARNUNG ACHTUNG HINWEIS
Eine Gefahr, die zu Verletzungen oder zu einer Beschädigung des Geräts führen können, wenn die Sicherheitsvorschriften nicht befolgt werden. Eine Gefahr, die zu leichten oder mittelschweren Verletzungen oder einer Beschädigung des Geräts führen können, wenn die Sicherheitsvorschriften nicht befolgt werden.
Allgemeine Informationen für den sicheren Betrieb des Geräts.
WARNUNG
• Stecken Sie das Anschlusskabel nicht mit nassen Händen aus, um einen Stromschlag zu vermeiden.
• Achten Sie darauf, dass die Steuereinheit nicht mit Wasser in Berührung kommt, da dies zu einem Kurzschluss und einem
Stromschlag führen kann.
• Berühren Sie das hintere Ende des Handstücks nicht, wo elektrische Anschlüsse mit dem Kabel verbunden sind. Dies
könnte zu einem Stromschlag führen.
• Wenn Sie vor oder während des Betriebs des Geräts eine Anormalität wie z.B. Vibrationen, Wärmeentwicklung, unnormale
Geräusche etc. feststellen, schalten Sie das Geärt sofort ab.
• Verwenden Sie eine geerdete Steckdose. Es kann zu einem Stromschlag kommen, wenn Sie eine andere verwenden.
• Betätigen Sie den Hauptschalter nicht grundlos, dies könnte eine Sicherung auslösen.
• Dieses Gerät ist ein medizinisches Elektrogerät. Die EMK (elektromagnetische Kompatibilität) wird in der
Begleitdokumentation beschrieben.
• Tragbare und mobile RF-Kommunikationsgeräte können das medizinische Elektrogerät beeinträchtigen. Verwenden Sie keine RF-Geräte in der Umgebung des Geräts.
• Sehen Sie beim Installieren des Geräts Platz von circa 10 cm um die Steuereinheit herum vor, damit der Zulauf und das
Anschlusskabel einfach zugänglich sind.
27
• Denken Sie beim Verwenden des Geräts stets an die Sicherheit des Patienten.
• Es ist zur Verwendung durch medizinisches Fachpersonal wie zum Beispiel durch einen Arzt/eine Ärztin oder einen Dentalhygieniker /eine Dentalhygienikerin bestimmt.
• Überprüfen Sie vor dem Verwenden die Vibrationen außerhalb des Mundes des Patienten. Sollte Ihnen etwas unnormal vorkommen, stellen Sie die Verwendung sofort ein und setzen Sie sich mit Ihrem Händler in Verbindung.
• Die Steuereinheit / das Handstück darf nicht fallen gelassen oder starken Erschütterungen ausgesetzt werden.
• Verwenden Sie nur echte NSK-Aufsätze für den NSK Varios Ultraschallscaler (Varios 970 oder Varios 970 LUX). Probleme wie zum Beispiel eine Beschädigung, ein Ausfall oder eine Störung von Handstücken aufgrund der Verwendung von anderen als NSK-Aufsätzen werden von der Garantie nicht abgedeckt. Im Folgenden finden Sie mögliche Fehler, die beim Verwenden von anderen als NSK-Aufsätzen auftreten können.
– Schwingungsbruch, verursacht durch die Verwendung nicht konformer Schrauben.
– Patient verschluckt versehentlich beschädigte Aufsätze. – Beschädigung des Gewindes am Handstück.
– Sie müssen den Aufsatz innerhalb des in der Leistungsrichtlinie für Aufsätze beschriebenen Leistungsbereichs
verwenden. Wenn Sie ihn außerhalb des Leistungsbereichs verwenden, könnte der Aufsatz abbrechen oder eine
Operationsstelle geschädigt werden.
• Verwenden Sie immer ausreichend Wasser (Kühlmittel), da es sonst zu einer Schädigung der Zahnoberfläche und einer Überhitzung des Handstücks kommen kann.
• Sterilisieren Sie es nicht mit ultraviolettem Licht. Das Handstück könnte sich verfärben.
• Sterilisieren Sie den Aufsatz, das Handstück und den Drehmomentschlüssel mit dem Autoklaven. Wischen Sie die Steuereinheit, das Wechselstrom-Anschlusskabel, den Fußschalter und das Handstückkabel mit DSH gelisteter Desinfektionslösung ab.
• Wenn chemische Lösungen, Lösungsmittel oder antiseptische Lösung an dieses Gerät gelangen, wischen Sie es sofort ab.
Sonst kann es zu einer Verfärbung oder Verformung kommen.
• Das Handstück/die Steuereinheit darf nicht auseinandergenommen oder verändert werden.
• Halten Sie das Gerät von Patienten mit einem Herzschrittmacher fern.
• Halten Sie das Gerät von explosiven Stoffen und entflammbarem Material fern. Verwenden Sie es nicht für Patienten, die
mit Lachgas betäubt werden.
• Verwenden Sie eine Sicherung mit entsprechender Bemessung (120 V: T630 mAL 250 V, 230 V: T315 mAL 250 V).
• Für dieses Gerät gelten besondere Vorsichtsmaßregeln bezüglich der EMK und es muss entsprechend den EMK-Daten
installiert und in Betrieb genommen werden.
• Die Verwendung von anderen ZUBEHÖRTEILEN, Wandlern und Kabeln als den hier angegebenen kann, mit Ausnahme von Wandlern und Kabeln, die vom Gerätehersteller als Ersatzteile für Einbauteile verkauft werden, zu einer vermehrten EMISSION oder einer verringerten STÖRFESTIGKEIT dieses Geräts führen.
• Dieses Gerät sollte nicht direkt neben, auf oder unter anderen Geräten aufgestellt werden, und wenn es direkt neben, unter oder auf anderen Geräten verwendet werden muss, muss sichergestellt werden, dass das Gerät in der Konfiguration,
in der es verwendet werden soll, normal funktioniert.
• Wenn nach dem Autoklavieren noch Wassertropfen am Handstück, wischen Sie sie ab. Wenn Sie sie nicht abwischen, können sich Flecken bilden.
• Dieses Gerät darf nicht vom Patienten benutzt werden.
• Eine zuverlässige Erdung kann nur erreicht werden, wenn die Ausrüstung an einer Anschlussdose mit der Kennzeichnung "Nur Krankenhaus" oder "Krankenhaus-Grad" angeschlossen wird.
• Wenden Sie keine zu hohe Leistung an der Spitze an, denn dies könnte den Zahn beschädigen.
Netzstecker unten wird in Nordamerika verwendet.
Steckertyp NEMA 5-15P (Typ Krankenhaus-Grad)
ACHTUNG
• Während des Betriebes können das Handstück und das Handstückkabel Computer und LAB-Kabel beeinflussen. Es kann
zu einem Rauschen kommen, wenn es neben einem Rundfunkgerät betrieben wird.
• Stellen Sie sicher, dass der Hauptschalter am Gerät nach der Benutzung ausgeschaltet wird. Ziehen Sie den Netzstecker und lassen Sie das Wasser aus dem Inneren der Steuereinheit ab, wenn sie für längere Zeit nicht verwendet wird.
• Der Benutzer ist für die Bedienung, Wartung und Inspektion verantwortlich.
• Reinigen/ sterilisieren Sie das Gerät direkt nach dem Verwenden. Dann lagern Sie es ein. Wenn Blut etc. darauf verbleibt,
kann dies zu einem Ausfall führen.
• Verwenden Sie zum Reinigen MinutenWipes (ALPRO) , um die Oberfläche des Handstücks abzuwischen. Die Verwendung anderer Chemikalien als dieser kann dazu führen, dass sich das Handstück verfärbt, bricht usw.
28
• Wenn Sie das Gerät längere Zeit nicht verwendet haben und es erneut einsetzen möchten, überprüfen Sie es vor dem Einsatz auf seine Funktionstüchtigkeit.
• Schauen Sie nicht in die LED-Lampe und lassen Sie die Patienten nicht hineinschauen. Dies kann zu einer Schädigung der Augen führen.
• Dieses Gerät kann für Patienten jeden Alters (außer Kleinkinder), Geschlechts, Gewichts und jeder Staatsangehörigkeit
verwendet werden.
• Für dieses Gerät ist keine besondere Schulung erforderlich.
• Anwendungsteile, die mit dem Patienten bzw. Bediener in Berührung kommen, sind Aufsatz bzw. Handstück.
• Oberflächentemperatur der Spitze ist mehr als 50 Grad, ohne einen Leitungswaßer zu verwenden. Um dieses Ereignis zu
vermeiden, seien Sie sicher einen Leitungswaßer zu benutzen.
HINWEIS
• Das wiederholte Autoklavieren kann dazu führen, dass sich das Handstück durch die Hitze verfärbt. Dies ist jedoch auf die Eigenschaften des Produkts zurückzuführen und stellt kein Qualitätsproblem dar.
Deutsch
29
Loading...
+ 137 hidden pages